You are on page 1of 12

5.

14 Vacuum Sensors PI
PE

B. G. LIPTK (1969, 1982, 1995), REVIEWED BY J. WELCH (1995) (Ion or Th)

R. A. GILBERT (2003)

Flow Sheet Symbol

Design Variations: A. Mechanical


A1. Liquid manometers (Section 5.9)
A2. Capacitance diaphragm manometers (Section 5.7)
A3. Quartz helix gauge (Section 5.4)
A4. Spinning ball viscous friction gauge
A5. Molecular momentum gauge
A6. Bourdon gauge
A7. Bellows gauge
B. Thermal
B1. Pirani (thermistor) gauge
B2. Thermocouple, thermopile gauge
B3. Convectron gauge
C. Ionization
C1. Hot cathode (Bayard-Alpert)
C2. Hot cathode (Suhultz-Phelps)
C3. Cold cathode (Philips)
C4. Cold cathode (Redhead)
D. Vacuum gauge calibration
D1. McLeod gauge
D2. Calibration reference tubes

Range: See Figure 5.14a

Inaccuracy: A1. Visibility limited to 0.1 mm


A2. Usually 0.05% or reading at 1 torr, increasing as vacuum rises to about 3% of
4
reading at 10 torr
A3. 0.01% of full scale
A4. Usually 3 to 4% of actual reading
A5. Usually 1 to 3% of actual reading
A6, A7. Limited range (only about 1% of atmosphere)
B1, B2, B3. From 2% of reading at point of calibration to over 10%; reading changes
with composition
C. Reading changes with composition
C1, C2. 10% of reading or less
C3, C4. 10 to 20% of reading

Cost: A1. $150 to $500


A2. $400 to $1000
A3. Over $5000
A4. Sensing head $2000; with controller, $11,000 to $14,000
A6. $50 to $100
A7. $75 to $300
B1. $350 to $700; with controller, $900 to $1100
B2. $300 to $800; with controller, $800 to $1200
B3. $120; tube and indicator, $500; tube, readout, and controller, $1100 to $1500
C1. $150 to $350; tube and controller, $1500
C2. $400

795
2003 by Bla Liptk
796 Pressure Measurement

C3, C4. $200 to $600


D1. $200 to $600
D2. $300

Partial List of Suppliers: Ametek Inc. (A6, A7) (www.ametekapt.com)


Cooke Vacuum Products (B, C) (www.cookevacuum.com)
Dresser Industrial Instruments (A6, A7) (www.dresserinstruments.com)
Dwyer Instruments Inc. (A1) (www.dwyer-inst.com)
Edwards High Vacuum International (B, C, D1) (www.edwards.boc.com)
Fountain Valley Instruments Inc. Televac Div. (hand-held) (www.telvac.com)
Fredericks Co. (www.frederickscom.com)
GFV Associates Inc. (B, C) (www.gfvassociates.com)
Granville-Phillips Co. (B, C) (www.helixtechnology.com)
Hastings (A, B, C, D2) (www.hastings-inst.com)
International Pressure Products Leopold Inc. (A6, A8, B1, B2, C)
Leybold Inficon Inc. (A) (www.inficon.com)
Marshalltown Instruments Inc. (A)
Meriam Instrument, a Scott Fetzer Co. (A5) (www.meriam.com)
MKS Instruments Inc. (A, B, C, D2) (www.mksinst.com)
Moeller Instrument Co. (A5) (www.moellerinstrument.com)
Rosemount Inc. Div. of Emerson (A2) (www.rosemount.com)
Stokes Div. of PermaH (D1)
Uehling Instrument Co. (A5, C1) (www.uehling.com)
Vacuum Technology Inc. (C) (www.vacuumtechnology.com)
Varian, Vacuum Products Div. (B, C) (www.varianinc.com)
Weksler Instruments Corp. (A6, A7) (www.monarchinstrcorp.com)

INTRODUCTION People new to vacuum gauges and vacuum technology


should also be prepared to deal with potentially confusing
From a working definition perspective, atmospheric pressure vocabulary when describing various pressure situations. Peo-
is considered to be 1.01 bar, 14.7 psia, 29.9 in. Hg, 760 ple with different technical backgrounds may use the terms
mmHg absolute, 760 torr, 760,000 m, or 101,300 Pa. An low vacuum, low pressure, and/or high vacuum, and high
obvious impact of this statement is the potential for confusion pressure imprecisely at different times. The reader should
due to the various unit options people have when it comes refer to a pressure that is very much lower than atmospheric
to reporting pressure. Fortunately, it is not an issue that the pressure as a low pressure or a high vacuum environment.
atmospheric pressure at sea level is not one fixed value, nor For a process situation that is very close to but below atmo-
is there any deleterious impact of defining 1 atmosphere of spheric pressure, the terms to use would be low vacuum or
pressurethe pressure at sea level on a clear cold nonstormy high pressure. In all situations, when a vacuum is referenced
2
day, to be 1,013,250 dynes/cm . Industrial vacuum measure- or described, the maximum pressure value being considered
ments are expressed in pressure units that reflect a pressure is atmospheric pressure. Therefore, vacuum chambers located
range and historic preference of the industrial practice. Thus, in Denver, CO, or Geneva, Switzerland, have maximum pres-
it is common for vacuum applications that are near atmo- sure values that are less than 1 atm (760 mmHg).
spheric pressure to be described in inches of water or mer-
cury. However, it is also expected that the pressure associated VACUUM GAUGE CLASSIFICATIONS
with a semiconductor metalization industrial application that
uses physical vapor deposition be described in Pascals, There are a variety of ways to categorize vacuum gauges.
microns, or perhaps millitorr units. Grouping gauges by operational principle is one common
The reader should also be prepared to deal with pressure way. Thus, the Bourdon tube and the bellows gauge are
unit selection variations with respect to the pressure units classified as mechanical gauges, but the spinning element
that appear on various vacuum pressure gauges. The unfor- gauge is know as a momentum transfer gauge. Figure 5.14a
tunate realities when dealing with industrial applications that provides a list of vacuum gauges and their operating range
require a vacuum environment is that there is no universal grouped by operational principle. Vacuum gauges that func-
vacuum gauge that will work for all applications and that tion because of a mechanical principle typically perform a
different types of gauges will use different pressure units. differential pressure measurement.
The normal measurement ranges for various gauges are sum- Gauges that operate within the same pressure range can
marized in Figure 5.14a. A variety of pressure units will be also be grouped together. Pirani gauges work in the same
used in this section. general pressure range as thermocouple (TC) gauges but the

2003 by Bla Liptk


5.14 Vacuum Sensors 797

Redhead

Bayard-Alpert

Philips

Momentum
Thermocouple
ionization
Pirani gauge

Viscous Friction thermal


gauge
Capacitance
mechanical
Convectron
gauge
Quartz

Diaphragm

Bellows
Medium Vacuum
Bourdon Tube Range

10 0.1 0.001
10+2 100 102 104 106 108 1010
Pressure in Torr

FIG. 5.14a
Ranges of the different types of vacuum gauges.

ion gauge would not be grouped with these two. A third way the vacuum is measured against an internal vacuum reference,
to categorize vacuum gauges is to identify if they actually the instrument is called an absolute pressure sensor; when the
make a direct or an indirect pressure measurement. Thus, a vacuum is compared to the barometric pressure, the instrument
direct vacuum pressure measurement gauge will provide a new will measure gauge pressure and is sometimes identified as a
measurement signal value simply because the pressure being compound pressure gauge. Examples of absolute pressure vac-
measured changes. The indirect pressure measurement gauge uum sensors include the copper-beryllium alloy bellows ele-
is actually monitoring some other parameter of the system that ments shown in Figures 5.3a, 5.3b, and 5.3c. Diaphragm-type
proportionally responds as the pressure of the system changes. vacuum gauges (Figure 5.5f ) use hermetically sealed, thin-
A vacuum gauge that uses a TC as the pressure-sensing element walled, copper-beryllium alloy, circular capsules. Similarly to
is one example of an indirect pressure gauge. The increase or the diaphragm and bellows elements, thin-walled, multiturn
decrease in pressure near the thermocouple does not directly spiral Bourdon tube elements can also be used to measure
alter the electromagnetic force generated by that TC. However, vacuum when a second reference Bourdon tube, containing a
that pressure change does reflect a change in the molecular near absolute zero vacuum, is also provided as reference.
population near the TC. This molecular density change affects
the heat transfer near that TC, which, in turn, alters the value MECHANICAL VACUUM GAUGES
of voltage, measured across the TC.
Finally, it is also important to identify if the vacuum gauge Manometers
performs an absolute pressure measurement. Similar to pressure
gauges designed for above atmospheric pressure measurement Manometers (Section 5.9) that use a liquid working fluid rep-
applications, vacuum pressure gauges may provide an absolute resent the classical differential pressure measurement device.
or gauge pressure reading. Naturally, it is important for instru- They are simple and inexpensive and some of their design
mentation and controls personnel to know which type of pres- variations are also shown in Figures 5.9a, 5.9c, 5.9h, and 5.9i.
sure measurement is being taken. The adjustment from gauge The simplest design for a vacuum pressure measurement
to absolute pressure is straightforward and may be provided as application is a U-tube with one leg connected to an evacuated
an electronic correction within the gauges. Elastic element reference chamber, while the other is exposed to the unknown
based vacuum gauges are simple and inexpensive examples. If process pressure. The difference between the two column

2003 by Bla Liptk


798 Pressure Measurement

heights indicates the process vacuum. The precision of reading Tensioned Inconel
the manometer is limited to about 0.1 mm when detected by Diaphragm
the naked eye. Inclining the manometer tube and using lower-
density filling fluids (low-vapor-pressure oils) can improve the
sensitivity of the readout to about 10 mtorr. When the units Capacitance
are operated at constant temperatures and are provided with Signals
micrometer readouts, vacuums down to the 1 mtorr level can
Process Pressure
be detected. Manometers are not used in vacuum related pro-
cesses that require medium to high vacuum or a clean vacuum
environment. This limitation reflects the possible vaporization Dual Electrodes
of the filling fluids at high vacuums or at high temperatures.
Density variation and light refraction can also make it difficult
to read the gauge. Although liquid manometers have limited High Vacuum
use in todays process environment, the term manometer has Reference Cavity
become ubiquitous. It is now generally applied to vacuum
FIG. 5.14b
gauges that make differential measurements, but do not nec- One-sided capacitance manometer detector can also be used on
essarily contain a working fluid. corrosive services.
1

Capacitance Manometers

The capacitance-type pressure detectors are examples of


manometers that do not have a moving fluid as a sensor element. Top (Bottom)
Vertical
Capacitance manometers have already been described in con- Suspended Ball Stabilization
nection with Figure 5.7i. These gauges provide high precision Rotation Permanent Magnets Coils
Driving
differential measurements because of their ability to detect Coils
extremely small diaphragm movements. The thinnest dia-
5
phragms are capable of measurement down to the 10 torr Gas Entry Port
level, while for the thicker diaphragms the measuring range from Vacuum
Chamber
can be extended to atmospheric or higher pressures.
A single capacitance manometer can report vacuum pres- Suspended Ball Suspended
sures values over a four- or five-decade range. Their inaccuracy in Sampling Tube Permanent Magnets Ball Rotation
in units of percent reading rises from about 0.1 to about 3% as Driving Coils
4
the vacuum rises from 1.0 torr to 10 torr. These devices, also
called micromanometers, are differential pressure detectors that
use a flexible diaphragm in an electrical capacitance circuit. FIG. 5.14c
One plate of the capacitor is the pressure-sensitive diaphragm Spinning ball element type vacuum gauge.
in the bridge circuit. The DC balancing voltage connected
across the plates of the capacitor exerts an electrostatic force. and the mirror used to detect its deflection have already been
When a pressure change occurs, the diaphragm deflects and the discussed in connection with Figure 5.4i. When used for vac-
electrostatic force restores the balance that opposes the initial uum detection, two quartz Bourdon elements are formed into
diaphragm movement. The magnitude of this rebalance voltage a helix. The reference tube contains a sealed vacuum, while
is a measure of the pressure exerted on the flexible diaphragm. the measuring tube is connected to the unknown process
The materials used in single capacitance manometer vacuum. The pressure difference between the two (at constant
detectors include Monel, Inconel, high-nickel alloys, and temperature) results in an angular deflection, which is
ceramics. The single-sided design illustrated in Figure 5.14b detected optically. The optical readout is used to eliminate
is corrosion-resistant because only one side of the diaphragm friction effects and to guarantee high resolution (about one
is exposed to the process vapors. Some models can operate part in 100,000). The main advantages of this detector are its
at temperatures up to 750F (400C). The two main sources precision and the corrosion resistance of quartz; however, it
of error are nonlinearity and temperature effects. As an alter- is an expensive gauge.
native to temperature compensation, the sensor can also be
heated and operated at a constant temperature. Viscous Friction of Spinning Ball

Quartz Helix Vacuum Gauge At high vacuums, viscosity and friction are pressure-depen-
7
dent. This instrument detects vacuum pressures down to 10
The quartz helix vacuum gauge is an example of a mechanical torr by detecting the deceleration caused by molecular friction
vacuum gauge that makes a differential pressure measurement on a levitating and spinning ball in a magnetic field. As suggested
without a working fluid element. The quartz helix element in Figure 5.14c, the vacuum is measured in this instrument by

2003 by Bla Liptk


5.14 Vacuum Sensors 799

first driving the ball until it reaches a rotational speed of about Mechanical Linkage Vacuum Gauge
425 r/s. When the drive is turned off, the rotational speed drops
as a function of the viscous friction of the process vapors. The only vacuum gauges that operate by a mechanical link-
When the speed has dropped to 405 r/s, the ball is accelerated age are the ones used for vacuum measurements, such as the
again. The pressure of the gas in the gauge is related to the Bourdon gauge and the bellows gauge. The Bourdon gauge
time that it takes for the speed to drop from 425 to 405 r/s. is also know as a C tube gauge and operates by the curving
As the process pressure decreases, the time needed to reach and uncurving of a hollow tube that is connected to the
process under vacuum. The movement of this C shaped
the lower rotational speed increases.
tube alters the position of a needle pointer connected to the
This instrument has several process advantages. Since
end of the sensing tube by a ratchet and spring mechanism.
its wetted parts are made of stainless steel, the gauge is
Although inexpensive and robust, the Bourdon gauge will
suitable for corrosive services. It is also suited for operation
demonstrate a memory effect and does not fit into automated
at baking temperatures up to 750F (400C). When cali-
data recording or process control situations. In addition, it
brated, the inaccuracy (uncertainty) can be as low as 1.5%
has a limited range and is not used in vacuum processes that
of the reading. Without calibration, the uncertainty is 4%
operate at pressures below 1 torr.
or more.
The Bellows vacuum gauge is another example of a
mechanical linkage type gauge. In this case, the bellows
Molecular Momentum Vacuum Gauges expands or contracts based on the pressure difference across
the inside and outside of the bellows unit. The gauge readout
Molecular momentum type gauges have two basic working is also the result of a mechanical interaction between the
parts: a rotating and a restrained cylinder. The gas molecules bellows and the needle pointer. These gauges do not have
from the vacuum chamber come in contact with the rotating electronic readouts.
cylinder (at a constant speed of 3600 rpm), experience a
momentum change, and are set in motion in the direction of
rotation. These molecules acquire energy from contact with
THERMAL VACUUM GAUGES
the spinning cylinder, and then, in turn, strike and transfer
that energy to the restrained cylinder. The pressure measure-
Heat transfer is a useful characteristic of a gas with respect
ment is possible because these molecular collisions move the
to the operational basis for a grouping of vacuum gauges.
restrained cylinder a distance that is proportional to the energy If a heated element with constant power input is placed in
transferred and represents a function of the number of gas a vacuum environment, the surface temperature of that ele-
molecules in that space. The number of molecules is related ment will be a function of the heat conductance of the
to the absolute pressure of the gas. The pointer attached to process gas, which also relates to the pressure of the process
the restrained cylinder indicates the gas pressure on the scale, gas near the heating element. Thermal vacuum gauges con-
somewhat similar to the operation of the viscous drag gas sist of three basic elements: a heater, a temperature sensor,
density sensor in Chapter 6. and a compensator for process temperature variations. There
The energy transferred in momentum transfer gauges is are two basic designs, depending on the type of temperature
not just related to the number of molecules (pressure) and sensor used: the resistance wire and the TC. In general,
the velocity of molecules. The molecular weight of the gas thermal vacuum gauges measure pressures down to 1 mtorr
is also a factor. Thus, the full-scale range of the gauge 3
(10 mmHg, 0.13 Pa) absolute pressures. Although this limit
depends on the type of gas to be detected. For air, the range can be extended with special designs involving special
3 4
is 20 to 10 mmHg (2.7 to 1.3 10 kPa), while for hydro- amplifiers and liquid nitrogen cooling around the gauge tube
gen, the maximum reading on the instrument is 280 mmHg to reduce radiation losses, an ion gauge is the single gauge
(37 kPa). Thus, molecular momentum transfer vacuum gauges 4
of choice when the pressure drops below 10 torr. Usual
have to be calibrated for each application. applications for thermal vacuum gauges include refrigera-
Molecular momentum transfer gauges give continuous tion, vacuum furnaces, freeze-drying, air conditioning, phar-
direct readout, but are not usually available as a signal trans- maceutical productions, and the manufacture of thermal food
mitter for remote indication or control. The inaccuracy of the containers.
unit is between 5 and 25%, with accuracy decreasing at Thermal vacuum gauges may be orientation sensitive.
lower pressures. Additional inaccuracy can be caused by Once attached to the vacuum system and calibrated, the
process temperature variations, which in the range of 50 to pressure readings from the gauge are dependent on that
100F (28 to 56C), can amount to 2%. External vibration physical location and orientation. If the gauge is remounted
in the range of 50 cps should be protected against by the use at the same location but rotated from its original position,
of bellows couplings on the process connection. The gauge is the gauge may need to be recalibrated. Most thermal vacuum
not damaged by exposure to atmospheric pressure, but the gauge manufacturers will indicate if their gauge is position
sample from the process has to be kept clean, free of dust, oil, sensitive and also indicate on the gauge housing the recom-
or other particles. mended gauge orientation.

2003 by Bla Liptk


800 Pressure Measurement

Compensating Cell Hot Junction Process


is Welded to the Connection
Heater
Output Midpoint of the Heater
Voltage Voltage
Supply
Process
Thermocouple
Pressure
Measuring Cell

FIG. 5.14d
Resistance wire (Pirani) vacuum detector.

Pirani Vacuum Gauge D.C.


Potentiometer
The Pirani gauge is a popular example of a thermal vacuum Readout
gauge. In this device, a constant current heats a filament
that is cooled by the gas present in the vacuum environment.
Since the amount of heat transferred changes with the
number of molecules, the equilibrium temperature, which 110V, 60C A.C.
is detected as the resistance of the metal wire, ribbon, or
FIG. 5.14e
thermistor element, is a function of the gas pressure near Single thermocouple vacuum gauge.
the wire. Figure 5.14d outlines the components of opera-
tion for the Pirani gauge. The heating and sensing elements Thermocouple Vacuum Gauge
are often combined as part of a Wheatstone bridge type
sensing circuit. A second resistance wire, which is enclosed As shown in Figure 5.14e, the single TC detector consists of
in a reference vacuum, is used to compensate for process a wire heated by the passage of constant AC or DC current.
temperature variations. The readout device detects the A TC is welded to the center of this heated filament, thereby
amount of current or voltage that is necessary to return the providing means to measure the temperature of the filament
bridge circuit to balance after a change in the vacuum being directly. An opening in the tube is provided for connection
measured. to the vacuum system being measured.
The standard pressure-sensing range for a Pirani gauge In operation, the constant current passing through the
3
is between 10 and 1 mmHg (0.13 to 133 Pa) absolute. The heater wire is in the order of 20 to 200 mA, and the TC sensor
inaccuracy of the gauge is about 2% at the calibration pres- develops a full-scale output in the order of 20 mV DC. For
sure and 10% over the operating range. Maximum output any constant value of current through the filament, the tem-
signal is about 0.1 mA at full-scale reading. perature increases as the pressure in the tube is reduced. The
Pirani gauges are relatively inexpensive and convenient temperature detected by the TC depends on the thermal con-
to use. They are not used for reliable pressure measurements ductivity of the gas surrounding the junction. For the same gas,
higher than 1.0 torr, because at higher pressures the thermal thermal conductivity is related to pressure when the pressure
conductivity of gases does not decrease with the decrease is at or below 1.0 torr. Like the Pirani gauge, TC gauge response
in pressure. The scale of the Pirani gauge is linear between is process gas dependent and the gauge must be calibrated
3 2
10 to about 2 10 torr. Above these pressures it becomes for specific applications.
roughly logarithmic. The two scales, when provided, are The available readout devices are as varied as the receiving
usually calibrated in terms of air. However, the gauge is gas instruments that measure TC output signals. Sensitive millivolt
dependent since different gases have different thermal con- meters and potentiometers are common choices. At higher
2
ductivity. Therefore, the Pirani gauge is composition-depen- ranges, such as 5 to 10 mmHg (665 to 0.13 Pa), dual scales
3
dent and needs to be calibrated for the process gas in the are usually used, while for the standard range, 0.1 to 10 mmHg
application. (133 to 0.13 Pa), a single scale is sufficient. The reading
Pirani gauges are not highly accurate. As they depend on inaccuracy at midscale is 2%, while 10% covers the full
the detection of thermal conductivity, the surface conditions range. On multistation installations, a single readout device
and emmissivity of the heated elements do affect the reading. can serve several TC gauges through a manual selector switch.
As these surfaces are coated, oxidized, or carbonized, their Because all filaments are on at the same time, the readings
temperatures rise and the sensor is likely to read high by as are instantaneous.
2
much as 2 10 torr. This represents a very large error at The filament temperature is kept below 400F (250C) to
the low end of the pressure range. Another substantial error reduce the possibility of the sample gases decomposing or form-
source is changes in the bridge voltage. ing a deposit on the gauge tube elements. Compensation can

2003 by Bla Liptk


5.14 Vacuum Sensors 801

be provided for process temperature variations, which other- metal thermopile. A change in process pressure results in a
wise would introduce an error by affecting the filament change of thermopile temperature causing a new DC output
temperature. This is achieved by the use of a reference gauge from the TCs. A third unheated thermocouple (C) is included
that has been fully evacuated and sealed. The same current in the circuit to compensate for operating process temperature
and voltage are applied to both the measuring and the ref- variations. This couple is the same size as the heated ones,
erence gauges. The temperature of the two heater wires is but it is connected in opposite polarity. A change in process
then compared, and the difference is used as a measure of temperature develops voltages in all the TCs, but the transient
process vacuum. The reference tube compensates for ambi- effects are equal and opposite in the heated and unheated
ent temperature changes because the two cells are at the elements. Therefore, compensation is achieved.
same temperature. Generally, the inherent low accuracy of Thermopile vacuum gauges have the same features and
these sensors does not make it worthwhile to compensate accessories normally associated with a single TC gauge.
for ambient temperature variations. However, the following additional characteristics should
It is also important that the filament current and voltage be noted. First, because of the noble metals used in the ther-
be carefully maintained for stable calibration. Filaments should mopile, oxidation of the couples does not occur. Second, the
not become dull or tarnished due to contamination because operating temperature of the heated thermopile is lower than
this would cause radiation losses affecting calibration. One that of the hot filament used in the single couple design.
way to overcome this problem is to precoat the filaments so Therefore, the probability of sample gas decomposition or
that further contamination would have no effect. Unfortu- deposit formation is remote. Third, thermopile gauges that
nately, precoating tends to impair gauge sensitivity. However, withstand several thousand PSIG overpressures are avail-
several precoated TCs connected in series, a thermopile, pro- able. Fourth, the gauges are corrosion resistant, rugged under
vide extra potential output that compensates for sensitivity demanding conditions, and provide stable calibrations. The
losses due to filament precoating. cold junctions are kept at ambient temperature by heavy
mounting studs. Thus, the amplified electromagnetic force
generated between the hot and cold junctions tends to be a
Thermopile Vacuum Gauge stable signal that reflects the temperature change at the hot
To increase sensitivity, thermopiles can be used to detect junctions. Finally, accurate readings are obtained in the
1 3
heater temperature. A thermopile vacuum gauge is a series range of 10 to 10 mmHg (13 to 0.13 Pa), but extended
of TCs, and a typical design is shown in Figure 5.14f. In this coarse detection is feasible over a broader range from 100
4 2
design, the TCs (A and B) are heated by low-voltage alter- mmHg (13 KPa) to 10 mmHg (1.3 10 Pa) absolute
nating current so that the heater and the temperature-sensing pressures.
functions are combined in the same noble (noncorrosive)
Convectron Vacuum Gauge

Process Connection A convectron gauge is similar to thermal conductivity type


to Vacuum System
Heated
vacuum detectors, except that it provides a wider (six-decade)
3
Thermocouples range, from 10 to 1000 torr within a single gauge. The heat
loss sensor is a temperature-compensated gold-plated tung-
A C B
sten wire that detects both conduction and convection cooling
effects. At lower pressures (higher vacuums), it operates as
Unheated a thermal conductivity vacuum gauge, while at higher pres-
Compensator
sures (in excess of 1.0 torr), it depends on convective cooling
Thermocouple
as the sensing principal. In this pressure regime, the process
gas molecules are circulated in the sensor tube by gravita-
tional forces.
In most respect, the features and limitations of convectron
vacuum gauges are similar to the Pirani and TC gauges, but
in addition to the extended range, there are some other nota-
D.C. ble differences. These include a lower sensor wire tempera-
Readout ture, which is likely to reduce the coating and carbonization
on the sensor wire surface. The gauge tube is stainless steel
and can be baked to up to 300F (150C), when not operating.
The maximum ambient or operating temperature is 122F
(50C). The bridge circuit is an integral part of the gauge
110V, 60C A.C.
tube, which is vibration and position-sensitive (it must be
FIG. 5.14f horizontal). The gauge is calibrated for nitrogen, and must
Multiple thermocouple (thermopile) vacuum gauge. be recalibrated when used on other gases.

2003 by Bla Liptk


802 Pressure Measurement

IONIZATION VACUUM GAUGES produced. These ions are then attracted to the negatively
charged collector electrode in the center of the grid to form an
Ionization vacuum detectors have been available since 1916. ion current. At constant accelerating voltage, the number of
However, practical, reliable, and sturdy ionization vacuum ions formed is proportional to the gas pressure if it is below
3
gauges did not come into common use until the 1930s. Today, 10 mmHg (0.13 Pa). Variations on the design include capping
ionization gauges are routinely used to detect pressure levels the end of the grid to prevent electron escape. A fine wire
4 9
from 10 to 10 torr. All ionization vacuum gauges detect collection electrode reduces the x-ray-generated photo current.
an electric current that is generated from the ionization of A spare emission filament is also often provided. At higher
the gas whose pressure is being measured. Ion gauges are pressures, the relationship between plate current and pressure
distinguished by the method applied in producing the ions. is not linear. This is because the mean free path becomes so
To convert a gas molecule into a positive ion, an electron short that an ionized molecule may pick up a free electron to
must be removed. If an atom or a molecule is supplied with become a molecule before it reaches the collector plate.
energy equal to its ionization potential, an outer orbit electron The accuracy of a hot cathode ionization gauge is poor
will escape and a positive ion will be created. The approxi- because the number of gas molecules to be measured is very
5 3
mate energy values required for ionization range from 5 to small. At an absolute pressure of 10 mmHg (1.3 10 Pa),
30 eV. If this energy is supplied at a constant rate and suffi- the inaccuracy would be about 10%. The vacuum range
3 11
cient atoms and molecules are available, the ions will be detectable by the hot cathode gauge is 10 to 10 mmHg
1 9
produced at a similarly constant rate. This ion stream is (1.3 10 to 1.3 10 Pa). The minimum span of the
directed to the gauges cathode, and the current that flows readout device is one decade, and it can be furnished with
through that cathode is proportional to the pressure of the five or six ranges. Range switching can be automatic or
gas in the gauge. The initial ion current is also proportional manual from the instruments front panel.
to molecular weight. The ionic current generally rises with The sensitivity of the hot cathode gauge is 100 A/m
3
molecular weight. In any case, calibration is usually done (10 mmHg) pressure. The readout device can be combined
with dry nitrogen or air. The various designs for ionization with thermocouple readout to extend its coverage to vacuums
3
gauges can be grouped by the method used to generate and in the range of 1 to 10 mmHg (133 to 0.13 Pa). This is a
direct the ion current. One broad grouping for ion gauges common practice, but a protective relay circuit is also fur-
includes hot cathode and the cold cathode gauges. nished to prevent filament burnout by keeping it off at pressures
3
higher than 10 mmHg. Besides multirange indicators, the
Hot Cathode Ionization Gauges readout device can provide signals for actuate electrical control
and alarm circuits. When several ionization sensors are
In the Bayard-Alpert hot cathode filament vacuum gauge illus- involved, the output signals can be multiplexed to reduce the
trated in Figure 5.14g, the ionization energy is supplied by number of signal detectors and pressure monitors, reducing
electron bombardment. The electrons are derived from thermi- total system cost.
onic emission from the hot filament. As these electrons are The hot filament vacuum gauge provides a wide pressure
attracted to and passing through the helix shaped grid, they reading range with fast response. Its application is limited to
acquire kinetic energy. When the electrons finally collide with gases that will not decompose on the hot filament. The glass
the gas molecules from the vacuum system, positive ions are ionization tube, by its nature, is subject to mechanical dam-
age. However, gauges for industrial process applications are
commercially available. The filament current is controlled in
Negatively Biased Positive Gas Entry Port such a way that a constant flow of electrons is emitted from
Ion Collection Electrode from Vacuum
Chamber it. In some designs, calibration is maintained by controlling
the grid charge so that if emissivity of the filament is decreas-
Hot Electron
ing, the grid receives a correspondingly greater charge to
Spare
Filament Emitting Filament maintain the entire circuit in equilibrium.
The limitations of the hot-cathode design are related to
the high filament temperatures (about 4000F, or 200C)
involved. At such temperatures, the incandescent filament
tends to deteriorate and is also susceptible to chemical attack.
The sorption of gas by the hot filaments and the degassing
that follows is another potential error source. These effects
are addressed by alternative gauge designs that use large
Helix-Shaped
Acceleration Grid diameter tubes and minimize in vacuum connection lengths.
As suggested above, various modified Bayard-Alpert hot
cathode vacuum gauges are available from different ion
FIG. 5.14g gauge manufacturers. One example is the Schultz and Phelps
Hot cathode ionization vacuum gauge. (Courtesy of Bayard-Alpert.) modification shown in Figure 5.14h. This gauge covers the

2003 by Bla Liptk


5.14 Vacuum Sensors 803

5
vacuum range of 1 to 10 torr and can be obtained in an which ions are produced. In the hot filament unit, the elec-
integrated unit with the Bayard-Alpert unit, when a wider trons are derived from thermionic emission. In the cold cath-
vacuum range needs to be covered. ode design, a high-potential field withdraws the electrons
from the electrode surface. Because the rate of electron emis-
Cold Cathode Ionization Gauges sion is lower in the cold cathode units, the collision frequency
between gas molecules and electrons would also be lower if
The cold cathode vacuum gauge is a composition sensitive the electrons traveled in a straight path. To increase the path
gauge traditionally known as Philips gauges, after its first length of the electrons, a magnetic field is created around the
manufacturer, or a Penning gauge after F.M. Penning, who tube to deflect the electrons. Thereby, the emitted electrons
outlined its operating principle in 1937. The basic difference spiral as they move across a magnetic field to the anode
between cold and hot cathode gauges is in the method by (Figure 5.14i). This spiraling action greatly increases the
electron path of travel and the corresponding chance of elec-
Exhaust System tron collision with gas molecules from the vacuum chamber.
The overall result is greater sensitivity of the cold cathode
gauge than that of the hot cathode.
The inaccuracy of the cold cathode unit is about 20% at
5 3
an absolute pressure of 10 mmHg (1.3 10 Pa). The gauge
High Pressure
Gauge response is gas molecular weight dependent and its detectable
2 7 5
vacuum range is from 10 to 10 mmHg (1.3 to 1.3 10 Pa).
As with the hot cathode gauges, the minimum span is one
Electron decade and readout devices are available with one, two, or
Collector three ranges, which are selected automatically or by manual
action at the front of the instrument. The gauge sensitivity is
Filament 5 mA/m micron pressure.
There are several advantages of the cold cathode gauge
relative to the hot cathode gauges. First, they cost less. Sec-
Ion
ond, they are more robust and do not burn out as easily. Third,
Collector
they do not subject the process gas to thermal stress. Cold
cathode gauge disadvantages include lower accuracy, nonlin-
ear output signal, and gas take-up caused by the high-voltage
operation. In order to remove polymerized organic contami-
nants, periodic cleaning of the electrodes and the vacuum
FIG. 5.14h chamber is required.
2
Hot cathode vacuum gauge. (Courtesy of Schultz-Phelps.)

Cathode

Process Micro-
Vacuum Anode Ammeter +
Reads the
~4000 V
Vacuum

Cathode

~1500 Gauss
Magnetic Field

FIG. 5.14i
Philips cold cathode ionization vacuum gauge. (The electrons travel from the cathode to the anode through a path of multiple spiral
oscillations which increase the opportunity for them to encounter and ionize molecules.)

2003 by Bla Liptk


804 Pressure Measurement

Similar to the hot cathode gauge, manufacturers have To Vacuum System


offered modified designs for the cold cathode gauge. One
Trap
example is a modification developed by Paul Redhead. In the
Redhead gauge, a cylindrical electrode is configured around
the anode that has a probe shape at the center axis. This
arrangement is like an inverted magnetron: the electrons ion-
Well
ize the gas molecules while they are spiraling in toward the
central anode, while the gas ions are collected on the cylin- Small Bulb Capillary
drical cathode surface. The Redhead gauge operates at around A
B
5000 V in a 1000-Gauss magnetic field and can detect vacu-
6 12 C
ums in the 10 to 10 torr range. Large Bulb
During the 1990s, a combination ionization gauge was
designed. This gauge is basically a cold cathode unit but is
Dual
furnished with a hot cathode serving to trigger the discharge Scale D
from the cold cathode. This design extends the detectable
4 14
pressure range to 10 decades, or from 10 to 10 mmHg Reservoir
2 12
(1.3 10 to 1.3 10 ) absolute.

Piston

VACUUM GAUGE CALIBRATION

Vacuum gauge calibration is an issue for two reasons. First, As a Reading is


Taken
a single vacuum gauge type does not work over the total Before Measurement
vacuum pressure range. Second, most vacuum gauges do not
directly measure pressure, and the pressure reading provided FIG. 5.14j
by the gauge is dependent on a property of the gas molecules Piston McLeod vacuum gauge.
being measured. Some common ways to calibrate vacuum
gauges are provided below. by capillary effects. These gauges are laboratory instruments
that measure on a sampling rather than on a continuous basis.
McLeod Vacuum Gauges They are not recommended for industrial installations.

The McLeod vacuum detector, or barometer gauge, is the Calibration Reference Tubes
classic reference or calibration gauge. Figure 5.14j shows one
version of this gauge. In this design, the unit is stationary, A relatively new method of vacuum gauge calibration
and a piston is used to trap the rarefied gas. The filling of the involves the use of calibration reference tubes. These tubes
gauge with mercury is done through the process connection. are available from most gauge manufacturers for application
This improved design does not use a dead-ended capillary, with their specific gauges. These reference tubes are funda-
and, therefore, the problems associated with keeping the cap- mentally evacuated, sealed vacuum tubes accurately cali-
illary clean are eliminated. The diagram on the left side of brated to precisely simulate an operating pressure environ-
Figure 5.14j shows the unit just prior to the taking of a ment of the gauges to be calibrated. The tubes permit quick
measurement. As the piston with micrometer adjustment is and easy gauge recalibration by simply connecting the gauge
moved up, the rarefied gas is trapped when the mercury to be calibrated to the appropriate reference tube and adjust-
reaches point D; thus, the initial volume (V1) is the volume ing the gauges pressure reading to reflect the value stated
between points A and D. When the instrument is connected on the reference tube.
to the vacuum system, the mercury level in the reservoir is
below point D to allow trapped gases to be liberated. As the
piston is moved up, the mercury fills the large bulb up to VACUUM CONTROLLERS
point C. At this point, a reading can be taken on the dual
scale if the pressure to be detected is in the mmHg range. If Vacuum gauges are one part of a vacuum process instrument
the vacuum is higher, the piston is moved further up, increas- system. In the complete instrument system, the process vac-
ing the compression ratio until the mercury reaches point B uum pressure is compared to a desired set point value, and,
above the small bulb. In this case, the reading is taken on the if necessary, a corrective signal is sent to a final control
micron side of dual scale. element. The vacuum controller is the component of the system
McLeod gauges can cover the vacuum range between 1 that will alter a control signal to adjust the vacuum pressure
6 4
and 10 mmHg (133 and 1.3 10 Pa). At pressures below when a process disturbance occurs. Process controllers do
4 2
10 mmHg (1.3 10 Pa), the reading accuracy is limited not have to have a pressure readout system; they only require

2003 by Bla Liptk


5.14 Vacuum Sensors 805

Vent
To
Restriction
Vacuum Pump
or Other
Vacuum Source Control Process
Connection to Port
Vacuum Source Diver
Connection

Valve Control
Port Air Supply
Spring
for Positive
Filtered Air Pressure
Service
Seal Fluid

Evacuated Controlled
Reference Vacuum Set Pressure
Bellows Sensing Connection
Line
FIG. 5.14l
FIG. 5.14k Cartesian diver.
Aneroid manostat.

rise as pressure varies. If the unit is to control a vacuum process,


a way to determine if the current vacuum is at the desired a vacuum pump (or other vacuum source) is connected to the
set point condition. Three types of vacuum controllers are unit. A process pressure increase causes the diver to sink,
briefly discussed. opening the control port and connecting the vacuum pump to
the process to lower its pressure. If the unit is installed for
Aneroid Manostats positive pressure control, a pressure source, not a vacuum
source, is connected to the control port, which is closed by an
The aneroid manostat illustrated in Figure 5.14k is a self- increase in process pressure and opened by its reduction.
contained bellows-type vacuum controller. The bellows are This device is capable of maintaining process pressures
fully evacuated to provide a zero absolute pressure reference between 1 mmHg absolute to 100 psia (0.13 to 690 kPa) to
unaffected by barometric changes. The spring is temperature- an approximate inaccuracy of 0.1% of set point. The unit is
compensated, and its tension is adjustable over the entire range available in both glass and metal, requires no external power
of the controller, which is 1 to 60 in.Hg (3.4 to 200 kPa) source, and is simple to operate or to change its set pressure.
absolute pressure. The sensitivity of setting is about 0.5 mmHg Because of its limited flow capacity, it can control small
(0.07 kPa) and the inaccuracy of control is 2% of set pressure. volume systems only.
The aneroid manostat works at above and below atmo-
spheric process conditions. When a below-atmospheric pres- Analog Electronic Controllers
sure is to be maintained, connections are made both to the
vacuum source (usually a vacuum pump) and to the controlled Continuous reading, relay action, analog gauge, electronic
system. The spring tension is set for the desired set pressure, controllers are also available and used for vacuum control.
and expansion or contraction of the bellows moves the valve These instruments provide dependable control with simplic-
port to control the airflow through the manostat. The manostat ity and economy. The desired set point is established by
is primarily designed for dead-ended service, but it will handle positioning the gauge need set point indicator at the desired
small flows. For example, at a setting of 300 mmHg (2000 kPa) value. The systems integrated vacuum gauge is stable and
without any restrictions in the vacuum source line, it will pass rugged and provides reliable measurements over the 1 to 50
3
0.2 ft (6 l) of air each minute. When an above-atmospheric torr range commonly provided. If the pressure-indicating
pressure is to be controlled, the filtered air supply is not atmo- needle passes the set point, the relay is activated. The relay
spheric any longer, but connected to a pressurized supply, and action is typically available as 115 V 5A SPDT @VAC nor-
the vacuum source connection is left open to vent the unit. mally open or normally closed contacts. The relay automat-
ically drops out when set point vacuum value returns.
Cartesian Diver Regulators
Mass Flow Controllers
Cartesian divers are self-contained pressure, or vacuum, reg-
ulators operating on principles somewhat similar to those of Many vacuum related processes require vacuum control, while
the aneroid manostats. As shown in Figure 5.14l, the set pres- process gases are admitted on a continuous basis. A (PID)
sure for this controller is sealed in under the diver. The process process control loop for such circumstances is possible with
pressure acts on the outside of the diver causing it to sink or a mass flow controller. Vacuum process mass flow controllers

2003 by Bla Liptk


806 Pressure Measurement

usually detect the temperature change as the gas passes Bibliography


through a small, typically 3-in. flow chamber. The temperature
is monitored with a TC gauge, and chambers can have stream Alvesteffer, W.J., Jacobs D.C., and Baker, D.H., Miniaturized Thin Film
splitter bypass elements to allow adaptation to different pro- Thermal Vacuum Sensor, Journal of Vacuum Science and Technology,
Vol. 13(6), 1995, p. 2980.
cess gas flow ranges. The differential temperature measure-
von Beckerath, A., Eberlein, A., Julien, H., Kerstein, P., and Kreutzer, J.,
ment signal becomes the input to an electronic PID controller WIKA Handbook on Pressure and Temperature Measurement, U.S. ed.,
and the resulting control signal is sent to the final control Lawrenceville, GA: Wika Instrument Corp., 1998.
element. Brombacher, W.G., 40 Years of Pressure Measurement, Instruments and
There are several final control element choices for mass Control Systems, September 1967.
Comber, J. and Hockman, P., Pressure Monitoring: Whats Happening?
flow controllers dedicated to vacuum process gas control. The
Instruments and Control Systems, April 1980.
solenoid proportional valve is one widely used cost effective Delajoud, P. and Girard, M. The Need for Evaolution in Standards and
control value that is coupled to a mass flow controller. This Calibration to Improve Process Measurement and Control of Low Mass
final control element uses a solenoid coil, a spring-supported Flow, 1996 National Conference of Standards and Laboratories Work-
armature, plug, and orifice. The unit can be configured with shop and Symposium, Monterey, CA, 1996.
an elastomer sealed valve. The assembly also usually contains Dyer, S.A., Wiley Survey of Instrumentation and Measurement, New York:
John Wiley & Sons, 2001.
both external seals, which isolate the gas flow path from Elliott, T.C., Temperature, Pressure, Level, Flow: Key Measurements in
atmosphere, and an internal seal for the valve seat. All metal Power and Process, Power, September 1975.
versions of the value are available when the process gas is Hall, J., Monitoring Pressure with Newer Technologies, Instruments and
also corrosive. Control Systems, April 1979.
Other control value options include a pieazoelectric con- Harvey, G.F., ISA Transducer Compendium, Instrumentation, Systems,
and Automation Society, Research Triangle Park, NC, 1969.
trol valve and moving coil actuator valve. In the former, a
Hughes, T.A., Pressure Measurement, EMC series, downloadable PDF,
stack of piezoelectric elements drives the value. This valve Instrumentation, Systems, and Automation Society, Research Triangle
type has a fast response time with high valve action force. Park, NC, 2002.
The counter valve action can be spring or diaphragm driven. Instrumentation, Systems, and Automation Society, Industrial Measure-
Piezo valves are all constructed from metal. For the latter ment Series: Pressure (video VHS, PAL & NTSC), Research Triangle
valve option, the actuator is a moving coil that surrounds a Park, NC, 2002.
Instrumentation, Systems, and Automation Society, Pressure: Indicators
permanent magnet. As current passes through the coil, it is and Transmitters, CD-ROM, Research Triangle Park, NC, 2002.
displaced. The displacement is transferred by a pivot arm and Johnson, D., Pressure Sensing: Its Everywhere, Control Engineering,
diaphragm action to the wetted volume on the other side of April 2001.
the diaphragm. This type of valve action produces a low valve Marrano, S.J., How to Choose and Apply Pressure Transmitters, Control,
sweep volume. This all-metal construction valve is good for March 2000.
OHanlon, J.F., Users Guide to Vacuum Technology, 2nd ed., New York:
contamination-sensitive applications.
John Wiley & Sons, 1989.
Roper, D.L. and Ryans, J.L., Select the Right Vacuum Gauge, Chemical
Engineering, March 1989.
References Sullivan, J.J., Journal of Vacuum Science and Technology, Vol. 3, 1985, p.
1721.
1. Welch, J., Capacitance Manometers, Measurements and Control, Sullivan, J.J. and Uttaro, F.L., Performance and Calibration of Capacitance
December 1989. Diaphragm Gauges, Measurement Science Conference, Irvine, CA, 1985.
2. Melling, R.J., Ionization Vacuum Gauge Measures Absolute Pres- Tompkins, H.G., Vacuum Technology: A Beginning, New York: American
sure, Instruments and Control Systems, September 1964. Vacuum Society, AVS Education Committee Book Series, July 2002.

2003 by Bla Liptk

You might also like