With a single-point adjustment, this value is 100 C plus the difference between the reference value
entered beforehand and the corresponding actual value. Example: actual value = 37.0 C,
reference value = 37.4 C, difference = 0.4 C.
(Sample values)
Overview:
The adjustment is carried out in thermal steady-state condition
When the adjustment temperature is equilibrated, the first step in the adjustment menu is resetting the
former adjustment. This is followed by the entry of the adjustment value (reference value).
Procedure:
Let the chamber equilibrate to the adjustment temperature for at least 1 hour (example: 37 C).
If the calibration has been performed immediately beforehand and the chamber is still equilibrated,
no further equilibration is required.
Chamber with adjustable fan speed (BF, FED): Fan speed 100%
Read out and note down the reading of the reference instrument.
Enter value Example
Actual temperature value of the controller C 37,0 C
Display reading of the reference instrument C 37,4 C
Enter the controllers adjustment menu by pressing simultaneously the arrow buttons for 5 seconds
No reset of the former Reset the former adjustment Show the last adjustment
adjustment values, leave the values and continue with the values
adjustment menu adjustment
The menu shows the adjustment value (flashing). At the beginning, the actual value is displayed.
Confirm with OK
The entered adjustment value is displayed
Confirm with OK
The controller shows normal display. The single-point adjustment is completed.
(Sample values)
Overview:
Adjustment is performed first for the lower adjustment value and then for the upper one. The
adjustment is carried out in thermal steady-state condition.
When the lower adjustment temperature is equilibrated, the first step in the adjustment menu is
resetting the former adjustment. This is followed by the entry of the lower adjustment value (reference
value).
Then set the upper adjustment temperature and let the chamber equilibrate.
When the upper adjustment temperature is equilibrated, enter the upper adjustment value (reference
value) in the adjustment menu.
Procedure:
Let the chamber equilibrate to the lower adjustment temperature for at least 1 hour
(example: 50 C).
Chamber with adjustable fan speed (BF, FED): Fan speed 100%
Read out and note down the reading of the reference instrument.
No reset of the former Reset the former adjustment Show the last adjustment
adjustment values, leave the values and continue with the values
adjustment menu adjustment
The menu shows the lower adjustment value (flashing). At the beginning, the actual value is displayed.
Confirm with OK
The menu shows the upper adjustment value (no function).
Confirm with OK
The entered lower adjustment value is displayed
Confirm with OK
The lower adjustment value is displayed (no valid value entered yet)
Confirm with OK
The controller shows normal display. The adjustment menu is still active.
Let the chamber equilibrate to the upper adjustment temperature for at least 1 hour
(example: 150 C).
Chamber with adjustable fan speed (BF, FED): Fan speed 100%
Read out and note down the reading of the reference instrument.
Note: The lower and upper adjustment values must differ by at least 1 C.
Confirm with OK
The entered upper adjustment value is displayed
Confirm with OK
The controller shows normal display. The dual-point adjustment is completed.
(Sample values)