You are on page 1of 2

Applicator Sizing - Diaphragm Seal

Project :
Customer: C.Project No.:
Contact person: Phone: Fax:
eMail:

TAG : ---
Timestamp: --- Review number: ---
Sales order number:

Sizing sheet
Operating Conditions
minimum nominal maximum
Process temperature -10 60 100 C
Ambient temperature -10 25 60 C
Static pressure (abs) 13.053 29.008 72.519 psi

Pressure transmitter : Cerabar S PMP75


Adjusted span 58.015 psi
Sensor 4bar/400kPa/60psi gauge
Membrane material 316L
Diaphragm seal NPS 1-1/2" Cl 150 RF, 316/316L
Transmitter mounting direct
Fill fluid Silicone oil
Calibration unit psi
Date 09.10.2015

Order Code
Cantidad Item Description Opciones de cdigo de pedido
1 Transmitter Cerabar S PMP75 PMP75-***1M61AE


Sizing and Calculated Results

Error due to change in ambient temperature 0.122 % span /50F 0.071 psi/50F
Error due to change in process temperature 0.065 % span /50F 0.038 psi/50F
Maximum offset at minimum conditions -0.2 %span -0.1 psi
Maximum offset at nominal conditions 0.1 %span 0 psi
Maximum offset at maximum conditions 0.3 %span 0.2 psi

Print date: 09.10.2015 10:06 PM -1/2-


Applicator: 10.21.01 / 150

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors and
consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A change in
these parameters could lead to different results. Mandatory data are in the according technical information (TI).
Applicator Sizing - Diaphragm Seal
Project :
Customer: C.Project No.:
Contact person: Phone: Fax:
eMail:

TAG : ---
Timestamp: --- Review number: ---
Sales order number:

Sizing sheet

Charts

Response time curves Diaphragm deflection

minimum nominal maximum


Response time Tau (T63) 0.2 0.2 0.2 s
Diaphragm deflection -25 4 18 %

Note: Process temperature range / Process pressure limits


The specified range may be reduced by the selected process connection. Observe the pressure temperature dependency. The corresponding values can be
found in the standards.

Warnings / Messages

Print date: 09.10.2015 10:06 PM -2/2-


Applicator: 10.21.01 / 150

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors and
consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A change in
these parameters could lead to different results. Mandatory data are in the according technical information (TI).