Professional Documents
Culture Documents
Sheesh Gulati
MeasureTest Corporation
MULTIPLICITY OF STANDARDS / GUIDELINES
ISO 1 10 2
ISO 2 100 24 10 4
Points to note:
For cleaner areas below ISO Class 8/Class 100K, the chart
in BS 5295 is marked as "NS" for "No Specified Limit".
Certification: ISO 14644-1 versus FS 209E
Minimum number of
2 with at least 5 1 with at least 3
samples at each samples total samples total
location
Note: Typical sample volume may be larger than minimum listed above
especially for smaller size particles in very clean areas (better than ISO
Class 5 or FS 209E Class 100)
ISO 14644-1 Minimum Sample Time at 1 CFM
Time required (in minutes) at 1 cfm (28.3 lpm) flow rate with 1-minute limit imposed
Three states
As-built
At Rest
Operational / Dynamic
A B
for 2 positions in ISO
Class 5 (FS 209E
Class 100) ...
125 3
119 8
120 12
364 >> 121 FDA 8 << 23
says
65
No !
Placement of Sample Probes
No regulatory standards for
monitoring
Not controlled earlier by FS 209E or
now by ISO 14644-1, when
conducting monitoring of the process
Costly and not practical to establish
monitoring points based on the ISO
14644 formula of square root of area
in sq.metres. Risk assessment is very
important in determining where to
monitor
ISO Class 5: ISO 14644-1 Classification
Calculations
Vial Freeze
Washing Dryer
System 5m 1
Freeze
Dryer
8m 2
5m
Calculations for Number of Points:
Freeze
Area of clean zone = 80 m Dryer
Take the SQRT (80) = 8.94 3
Rounding up to next integer = 9 sample
positions 4m
Vial Freeze
Washing Dryer
System 1
1 2 3 4 5 6
7
Freeze
Dryer
2
Calculations for Number of Points: 8
We might place them as shown. But this does not take into
account the reality of what is in the room: entrances, exit and
machinery. So we need to adjust for these.
ISO Class 5: ISO 14644-1 Calculations
1 2 3 4 5 6 7 8 9 10
Vial Freeze
Washing Dryer
System 1
Freeze
Dryer
2
Need to adjust for equipment in room.
Under ISO 14644-1, if you sample at 10 or more
positions, you can avoid the added calculation of
the UCL (Upper Confidence Limit). Calculation of
the UCL is only mandated when the number of Freeze
positions used is between 2 and 9. Dryer
Best to sample near potential problem spots 3
which are near entrances and exits and near
operator positions.
ISO Class 5: ISO 14644-1 Calculations
9
1
Vial 10
Freeze
Washing Dryer
8
System 1
2
11
3 4 5 6 7
Freeze
12
Dryer
2
Need to adjust for equipment in room.
Under ISO 14644-1, if you sample at 10 or more positions, 13
you can avoid the added calculation of the UCL (Upper
Confidence Limit). Calculation of the UCL is only
Freeze
mandated when the number of positions used is between
2 and 9. 14 Dryer
3
Best to sample near potential problem spots which are
near entrances and exits and near operator positions.
Here could be a distribution of sample points that would both
provide a good view of the cleanroom particulate values, and,
importantly, be defended against any regulatory challenge.
Note that there are more points suggested than the minimum
calculation.
Placement of Isokinetic Probes in a
Pharmaceutical Filling Area
for the Purpose of Monitoring
7
3 4
Freeze
Dryer
2
Presence of lyophilizers indicate vials may not be
fully stoppered so the holding position near 5
represents some risk 8
3 4
But this is not an easy number to set up for a grid pattern in a rectangular area
so forty-two positions might be a better choice.
ISO Class 7: ISO 14644-1 verification
The sample positions would be at work height in the middle of each rectangle.
Example: ISO 14644-1 Calculations
If the number of points sampled is more than 1 but less than 10,
then the UCL factor must be applied:
Calculate the standard deviation
Use Students T-factor from tables
Calculate UCL
Compare to classification limit
UCL must not exceed the applicable limit
ISO Class 7: Selecting Monitoring
Positions
Work Station 1
Work Station 2
100 ft
(30 m)
Storage
Work Station 3
Work Station 4
In the real world, monitoring positions will be affected by the physical layout of
the area and the activities that occur within it.
Also entry and exit points should be considered.
ISO Class 7: Selecting Monitoring
Positions
Work Station 1
Work Station 2
100 ft
(30 m)
Storage
Work Station 3
Work Station 4
FDA GMP:
Measurements should be taken with the particle
counting probe oriented in the direction of oncoming
airflow and at the sites where there is most potential
risk to the exposed product
Where to monitor?
Where to monitor in ISO Class 5
(Class 100) / Grade filling room
Isokinetic probe on an
Accumulation Turntable
provides monitoring of
rotary in-feed turntable
after sterilization zone.
Adjustable mount
(optional) allows fine
tuning of sampling
position.
At Rest In Operation
3
Grade Maximum permitted number of particles/m equal to or above
0.5 m 5 m 0.5 m 5 m
New
A 3 520 20 3 520 20 Effective Mar. 2009
Section 12 states:
"It is not necessary for the sample volume to be the same as that used
for formal classification of clean rooms and clean air devices.
i.e. you do not need to sample minimum 1 cu metre during monitoring
DEALING WITH 1 CU.METRE REQUIREMENT
DURING CLASSIFICATION
Increased sampling frequency of low air volume is
preferable to high air volume at low frequency
In other words, 35 readings of 1 minute at 1 cfm are
preferable to 1 reading of 35 minutes
Action limit for 1 cu.metre
Single readings: If all readings are below 1/35 of the
limjt, the limit will never be exceeded
Multiple readings: If some of the single readings exceed
the 1/35 of cu.metre limit, it has to be checked
whether the result of sampling 1 cu.metre air volume
would have exceeded the limit
DEALING WITH 1 CU.METRE REQUIREMENT
DURING CLASSIFICATION
Action limit for 1 cft readings:
Higher sample
probe for
monitoring during
capping operation
30 tha s
s
Sample probe to 5 n
Le
m
m
demonstrate air
quality before
capping process
Interpreting EU GMP Annex 1
FDA cGMP
areas where product is at most potential risk
not more than 1 foot away from the work site
Advice on Best Practice for
Cleanroom Monitoring
What was lacking was practical advice on how to implement these
continuous monitoring systems. There is none in Annex 1 and
ISO14644 is for room classification only.
Continuous what, for example, does the word continuous mean?
Monitoring where should we locate the monitoring points and how
many should there be?
5micron what do we do if we see 5micron counts. How many are
acceptable before we initiate and action limit
1 cubic metre it is not clear as to whether we should try to sample a
complete cubic meter during each manufacturing batch. Some
aseptic manipulations are complete in a matter of minutes and it
currently takes at least 20 minutes to capture 1cubic metre of air
with a modern counter.
Powder are we really supposed to monitor for particles during a
powder fill?
Scope & Aims of PHSS Special Interest
Group
Scope:
Cleanroom non-viable air particle monitoring
EU GMP Annex 1
Aims:
Collate best practice from Industry, Healthcare and regulatory
bodies
Publish monograph :
AstraZeneca
Bio Products
Boehringer-Ingelheim
Boots Contract Manufacturing
Cardinal Health
GlaxoSmithKline
Hach Ultra Analytics
Ipsen Biopharm
Particle Measurement Techniques
Wyeth
MHRA Regulatory Inspectors (EMeA)
Best Practice Document Contents
Changes to EU GMP Annex 1 published 2008, live March
2009
System Design
Operations
Maintenance and Cleaning
Training
Appendix A Worked example
Appendix B Manifold and Remote Particle Monitoring
Systems
Appendix C Examples of particle loss in transport tubing
Appendix D Isokinetic probes
Appendix E Validation and risk assessment standards
and guidelines
UK PHSS Best Practice monitoring
Grade A areas monitored continuously using dedicated particle
counters.
Vial
Sterilizing
Tunnel
Key
Remote
Counter
Vacuum
Tubing,
Power
& Data
Central Central
Software Vacuum
System Pump
Appendix B dedicated counter Grade A
Vial
Sterilizing 0.5
Tunnel
0.5
Key
Remote 0.5
(Built-in
Pump)
Power
& Data
Vacuum
Tubing
Central
Software
System
Appendix B Manifold for Grade B & C
Vial
Sterilising
Tunnel
Key
Sample
Probe
Vacuum
Tubing
Comments:
Attaining a sterile state in an aseptic processing
facility is impossible. Personnel are always present in
manned cleanrooms performing various activities
including microbial sampling. Detection of micro-
organisms occasionally is inevitable and need not be
a cause for action against product.
AIR CHANGES
Describing airflow in terms of air changes per hour is common
for non-unidirectional flow rooms (ISO classes 6 through 9)
and high-bay installations.
Since the airflow in these rooms is non-uniform, attempting to
directly measure the average air velocity is not feasible. The
average velocity may be calculated, however, using
volumetric measurements from the terminal filters. This
velocity is then converted into an equivalent room air
changes per hour (AC/H).
It is worth noting that in the UK GMP ('Orange Guide')
the room air change requirements have been removed
in the latest edition.
EU Annex 1 vs. FDA Guideline
EU Annex 1 FDA Guideline