You are on page 1of 48

TECHNICAL MANUAL

OPTIMA
PT/ST/VS
Page 2 NT1789 Rv.A0
TABLE OF CHANGES

The information contained in this document only concerns :


OPTIMA PT/ST/VS type,
MCM 440 PT/OT type,
MCM550 ST type.

Technical Manual reference :......................... : NT 1789


Revision date :......................................... ...... : 27 May 2003
Applicable from serial no :.............................. : 18180100

Date Revision no. Pages concerned Changes


27/05/2003 A0 All First release

NT1789 Rv.A0 page : 3


Page 4 NT1789 Rv.A0
3.3.10. Display of software version number (TestA)......... 28
3.3.11. Motor clamp Test (TestC) .................................... 29
TABLE OF CONTENT 3.3.12. Calibration values display (TestE) ........................ 29
3.3.13. Infusion motor Test (TestF).................................. 30
3.3.14. Display of last 10 alarms before a failure occurs
1. OVERVIEW .................................................. 7 (TestJ) ............................................................................. 30
1.1. Ge n e ra l ......................................................7 3.4. Ma in te n a n c e S e rvic e Co n fig u ra tio n . ... 31
3.4.1. Service configuration Menu.................................... 31
1.2. Blo c d ia g ra m s ...........................................8
3.4.2. Preventive maintenance (SAV1) ............................ 31
1.3. P re c a u tio n s to b e ta ke n b e fo re u s e .......9 3.4.3. Set number Configuration (SAV2).......................... 31
1.4. In te rn a l s a fe ty fe a tu re s ............................9 3.5. Te c h n ic a l fa ilu re s m e s s a g e s . ............... 32
1.5. Te c h n ic a l c h a ra c te ris tic s ........................9 4. REP LACING S UB-AS S EMBLIES ................34
1.5.1. Biologica l cha ra cte ris tics .......................................... 9
4.1. Mo d u le o p e n in g . .................................... 34
1.5.2. Me cha nica l cha ra cte ris tics ....................................... 9
1.5.3. Dime ns ions / We ight................................................ 9 4.2. Re p la c in g CP U b o a rd . ........................... 34
1.5.4. Ele ctrica l cha ra cte ris tics .......................................... 9
4.3. Dis m o u n tin g p u m p in g m e c h a n is m ...... 34
1.5.5. Ele ctronic cha ra cte ris tics ......................................... 9
1.5.6. Operators guide ...................................................... 9 4.4. Re p la c e m e n t o f p re s s u re s e n s o rs . ...... 34
1.5.7. Registration cards .................................................... 9 4.5. Re p la c e m e n t o f d o o r o r o f Ha ll e ffe c t
1.5.8. Material used to build the OPTIMA / MCM ............... 9
s e n s o r ............................................................ 34
1.5.9. Conformity and -symbols ......................................... 9
5. MAINTENANCE ...........................................35
2. ELECTRONIC BOARDS ............................. 11
5.1. Re c o m m e n d a tio n s ................................. 35
2.1. CP U b o a rd ...............................................11
2.1.1. Functional description. ........................................... 11 5.2. Cle a n in g a n d d is in fe c tin g ..................... 35
2.1.2. Connectors description .......................................... 12
5.3. S to ra g e .................................................... 35
2.1.3. Electronic drawings ................................................ 14
2.1.4. Components layout ................................................ 14 5.4. Ro u tin e in s p e c tio n s .............................. 36
5.4.1. Checking before use .............................................. 36
2.2. Dis p la y b o a rd ..........................................15 5.4.2. Occlusion alarm checking ...................................... 36
2.2.1. Functional description ............................................ 15
5.4.3. Battery autonomy test ............................................ 37
2.2.2. Connector description ............................................ 15
5.4.4. Fix clamp checking ( VS/PT and ST pumps)......... 37
2.2.3. Electronic drawings ................................................ 17
5.4.5. Air detector check .................................................. 37
2.2.4. Components layout ................................................ 17
5.4.6. Continuity test ........................................................ 37
2.3. P OWER S UP P LY BOARD ......................18 5.5. Te c h n ic a l c h e c k p ro c e d u re .................. 38
2.3.1. Functional description ............................................ 18
2.3.2. Connector description ............................................ 18 6. ANNEXE 1 : Illu s tra te d p a rts lis ts ..............40
2.3.3. Electronic drawings ................................................ 19 6.1. S u b a s s e m b lie s tra c e a b ility ta b le ......... 40
2.3.4. Components layout ................................................ 19 6.1.1. Introduction ............................................................ 40
2.4. Air d e te c to r b o a rd ..................................19 6.1.2. Spare parts table.................................................... 40
2.4.1. Electronic drawings and Components layout.......... 19 6.1.3. Upper housing........................................................ 41
6.1.4. Base ...................................................................... 42
3. Co n fig u ra tio n , c a lib ra tio n a n d c h e c k ....... 20 6.1.5. Brace ..................................................................... 43
3.1. Co n fig u ra tio n ..........................................20 7. Circ u it d ia g ra m ............................................45
3.2. Ca lib ra tio n ...............................................22 8. Ad d e n d a a n d in fo rm a tio n b u lle tin s ..........46
3.2.1. Manual calibration menu ........................................ 22
3.2.2. Door calibration (Etal1) .......................................... 22 9. Us e fu l a d d re s s e s ........................................48
3.2.3. Correcting coefficient adjustment (Etal2)................ 23
3.2.4. Up and Down stream pressure sensors calibration
(Etal9)..................................................................... 24
3.3. Te s ts : Ma in te n a n c e te s ts .....................26
3.3.1. Manual Service test menu...................................... 26
3.3.2. Maintenance tests list : .......................................... 26
3.3.3. Working duration (Test1)........................................ 26
3.3.4. LEDs, LCD screen, 7 segments display and Buzzers
Test (Test2) ............................................................ 27
3.3.5. Keyboard test (Test3)............................................. 27
3.3.6. Battery voltage display (Test4)............................... 27
3.3.7. Display of the last 10 alarms codes (Test5) ........... 28
3.3.8. Total working duration (Test6)................................ 28
3.3.9. Display analog values of sensors (Test9)............... 28

NT1789 Rv.A0 page : 5


P a ge 6 NT1789 Rv.A0
1. OVERVIEW

1.1. Ge n e ra l
The OPTIMA Pump is a new volumetric pump based on the same mechanical pumping mechanism as the MVP
pump.
The battery is automatically recharged when the pump is connected to the mains..
A choice of easily accessible configurations ensures optimum use of functions according to the needs of each depart-
ment.
3 types of OPTIMA exist : VS / PT/ ST
OPTIMA VS / MCM 440 OT: PVC Type Volumetric pump with standard PVC Set. The Drip sensors is optional. An
automatic control of the occlusivity of the pump ( OCS Occlusivity Control System) is implemented.
OPTIMA PT / MCM 440 PT :PVC Type Volumetric with standard PVC Set. The Drip sensors is compulsory.
OPTIMA ST / MCM 550 ST Silicone Type Volumetric with standard Silicone Set. The Drip sensors is compulsory
Mains differences between the types :
Description VS / OT PT ST
Functional differences
Tube Standard PVC Standard PVC Standard Silicone
Drip Sensor Optional Standard Standard
OCS Test (Occlusivity Yes No No
Control System)
Tube Clamping A mechanical clamp A mechanical clamp A mechanical clamp
on the device assure on the device assure on the device assure
the clamping of the the clamping of the the clamping of the
tube when the door is tube when the door is tube when the door is
opened. opened. opened.
Maximum Rate 1000 ml/h 1000 ml/h 1500 ml/h
OCS A dc motor assure the NO OCS -> no DC motor
clamping of the tube
during OCS
Tube clamping during A mechanical clamp on the device assure the clamping of the tube
door opening when the door is opened.

NT1789 Rv.A0 page : 7


1.2. Blo c d ia g ra m s

Ba ttery

RS 232
Nurs e Call
Key board
P owe r s upply
board
Dis pla y board
AIR conne ctors

bubble s Drip
Keyboa rd/dis pla y
S ens or
interface
s e ns or CP U (option)

P roce s s or function
OCS DC Motor
(MS /VS only)

Motor drive r
S e ns or interface

Motor
Rota tion
dete ctor

Trans mis s ion


Door P e ris taltic Me ca nis m
Openning
dete ctor
DOOR

L0OCK

Counte r P res s ure fla s k

downs tre am
occlus ion Up-s trea m
occlus ion
De tector
Detector
Clamp dete ctor
(MS only)

Luer de rivation Down s tre am exte ns ion P UMP clamp Up-s tre am extens ion Drip chamber

P erfus ion s et

Figure 1. Ge ne ra l dia gra m of the OP TIMA

P a ge 8 NT1789 Rv.A0
1.3. P re c a u tio n s to b e ta ke n b e fo re u s e
Please consult the user guide.

1.4. In te rn a l s a fe ty fe a tu re s
The pump has a continuous functions inspection system as soon as the pump is ON. . Nevertheless, the qualified per-
sonnel in your establishment or our Maintenance Department should always be notified of any abnormal functioning
when no specific cause can be found. (See useful addresses, chapter 9)
For any mains failure the pump is powered automatically from its internal battery.
Opening the pump or the battery cover must only be carried out by qualified staff with all the necessary technical pre-
cautions. Non-compliance with these procedures may be dangerous for the staff and may damage the Module.

1.5. Te c h n ic a l c h a ra c te ris tic s

1.5.1. Biological characteristics


The infusion liquid is only in contact with the infusion set.

1.5.2. Mechanical characteristics


The OPTIMA Pump operates with a linear peristaltic system.

1.5.3. Dimensions / Weight


q H x L x D ........................................................................... 135 x 175 x 145 mm
q We ight ............................................................................... 2.9 kg a pproxima te ly

1.5.4. Electrical characteristics


q Exte rna l powe r s upply........................................................ 100 240 Vac
q Power ................................................................................. 50 VA
q Battery................................................................................ 6V 2.7/ 3 Ah NiMH.
q Fuse ................................................................................... 2 x 630 mA T IEC 127

1.5.5. Electronic characteristics


The Device contains 4 PCBs:
CP U P CB.
Dis pla y / ke yboa rd P CB.
P owe r S upply P CB.
Air de te ction P CB
1.5.6. Operators guide
The operators guide is available upon request to our Maintenance Department (see useful addresses at the
end of this document)
1.5.7. Registration cards
Registration information is available upon request to our Maintenance Department.

1.5.8. Material used to build the OPTIMA / MCM


q Top a nd bottom hous ing .................................................... P olyca rbona te / ABS
q Ke yboa rd ............................................................................ P olye s te r
q S ticke rs .............................................................................. P olye s te r
q P ump me mbra ne ............................................................... Ela s tome r (la te x fre e )
q Drip s e ns or......................................................................... P olyca rbona te / ABS or polya mide
q Re a r hous ing ha ndle .......................................................... P C + 10%

1.5.9. Conformity and -symbols


q Conform EN60 601-1 a nd EN60 601-2-24.s ta nda rds

NT1789 Rv.A0 page : 9


CE0459 ma rking in complia nce with EEC 93/42 Me dica l De vice dire ctive .
IP 31 Liquid proje ctions prote cte d.

CF type de vice

Ba tte ry.

CLAS S I de vice

q Ea ch s ymbol s hows be low is vis ible on the de vice or in this docume nt.

Re fe r to a ccompa nie d docume nts .

Equipote ntia lity

P rote ctive e a rth

56 RS 232 Communica tion Inte rfa ce 4000V ins ula tion

Nurs e Ca ll 4000V ins ula tion / 24Vdc-1A

P e rma ne nt ma gne t a llowing the fixing of drops s e ns or

P a ge 10 NT1789 Rv.A0
2. ELECTRONIC BOARDS

2.1. CP U b o a rd

2.1.1. Functional description.


2.1.1.1 CP U

Ma in Microproce s s or .................. T80c5112 / 512 Ko of Fla s h / 8Ko of RAM


Exte rna l RAM .............................. 32 Ko
EEP ROM ..................................... 16 Ko
S e conda ry Microproce s s or ......... MS P 430
Control fre que ncy of the ma in microproce s s or, time ke e pe r function, ON/OFF function a nd control the
cha rge of the ba tte ry.
2.1.1.2 S te ppe r motor control
The motor is powe re d with 12 or 20 volts a ccording to the flow ra te .
The signal command is performed either by step or step according to the requested flow rate.
2.1.1.3 Rota tion control
An optical sensor verifies the rotation of the came shaft.
This sensor generates an alarm for any difference higher than 5 % with the programmed value
2.1.1.4 Door de te ctor
A Hall effect sensor placed in the Module and a magnet located in the door
This sensor detects door opening and set missing when door is closed.
2.1.1.5 Drip s e ns or (optiona l on Optima MS a nd VS or MCM 440 OT)
The drip detection is performed by an optical sensor.
An infra-red filter avoids interference due to external light.
This sensor generates an alarm for any flow rate difference higher than 40% of the programmed value
2.1.1.6 Downs tre a m pre s s ure s e ns or
This sensor detects downstream occlusion on the infusion line.
It also detects the right positioning of the infusion set when the door is closed.
It generates an alarm in both cases.
2.1.1.7 Up-s tre a m pre s s ure s e ns or
This sensor detects up-stream occlusion on the infusion line when the drip sensor is absent.
This sensor detects the right positioning of the infusion set when the door is closed.
It generates an alarm in both cases.
2.1.1.8 Buzze r
The device is equipped with 2 buzzers triggered by the microprocessor.
2.1.1.9 On/Off function
When the device is connected to the mains, it is in stand by mode .
S witching ON is pe rforme d
Eithe r by pre s s ing the ON/OFF ke y
Or by ope ning the door, a ccording to the s oftwa re configura tion.

S witching OFF is pe rforme d :


Eithe r ma nua lly by pre s s ing the ON/OFF ke y
Or a utoma tica lly whe n ba tte ry is low.

The On/off function is pe rforme d by the s e conda ry microproce s s or


2.1.1.10 Ba tte ry cha rge ma na ge me nt

NT1789 Rv.A0 page : 11


The ba tte ry cha rge is ma na ge d by both prima ry a nd s e conda ry microproce s s ors .
2.1.1.11 Air de te ctor
An ultra - s ound s e ns or a llows to de te ct a ir bubble s in the infus ion s e t.
It ge ne ra te s a n a la rm if a ir bubble s a re la rge r tha n progra mme d va lue .
2.1.1.12 Automa tic OCS cla mp
The OP TIMA MS a nd VS or MCM MS a nd OT a re e quippe d with a DC motor a ctua te d cla mp. Afte r e a ch infu-
s ion s ta rt or door clos ing, the infus ion s e t is cla mpe d by this s ys te m a nd the de vice che cks the n pumping
me cha nis m occlus ivity.
2.1.1.13 Nurs e Ca ll / RS 232
The RS 232 a nd Nurs e Ca ll ha ve both 4000 V die le ctric s tre nght. The Nurs e ca ll re la y is ra te d 24 VDC/1A.
2.1.2. Connectors description
2.1.2.1 Ge ne ra l

Co n n e c to r P in n u m b e r Fu n c tio n
J7 2x17 For de ve lopme nt purpos e
J3 2x7 For ma nufa cturing
J9 1x4 Air S e ns or
J1 1x20 CP U/Dis pla y inte rconne ction
J 12 1x4 Drip S e ns or
J4 1x4 S te ppe r Motor
J6 1x3 P owe r S upply
J8 1x4 Ba tte ry
J5 1x6 Not Us e d
J 11 1x8 RS 232 / NURS E CALL
J 10 1x2 DC motor
J2 1 x 20 S e ns ors
2.1.2.2 J9 Air Sensor

P in De s c rip tio n
1 /TS T S e ns or che ck
2 5V P owe r S upply
3 GND Ground
4 /DETECT Air output s igna l

2.1.2.3 J1 CPU/Display board interconnection

P in De s c rip tio n
1 5V P owe r S upply
2 VBAT Ba tte ry volta ge
3 GND
4 MOS I-AFF S e ria l output
5 GND
6 MIS O_AFF S e ria l Input
7 GND
8 CLK-AFF S e ria l clock
9 GND
10 CS LCD Chip S e le ct LCD
11 GND
12 CS AFCL Chip S e le ct Dis pla y

P a ge 12 NT1789 Rv.A0
P in De s c rip tio n
13 GND
14 A0AFF LCD Adre s s
15 GND
16 RES ET/
17 GND
18 TON/OFF ON/OFF s igna l
19 VBAT Ba tte ry Volta ge
20 5V P owe r s upply
2.1.2.4 Drip S e ns or

P in De s c rip tio n
1 CDLED-goutte Le d comma nd 5V / 40 mA
2 GND
3 GND
4 P uls e -goutte Output drop s igna l
2.1.2.5 J 4 - S te ppe r motor

P in De s c rip tio n
1 + Bobine A + Coil A
2 - Bobine A - Coil A
3 + Bobine B + Coil B
4 - Bobine B - Coil B

2.1.2.6 J6 Power Supply

P in De s c rip tio n
1 ALIM EXT Not Us e d
2 GND
3 ALIM S ECT P owe r S upply 9V / 1.3A

2.1.2.7 J8 Battery

P in De s c rip tio n
1 VBATNC + Ba tte ry
2 ANA-CTN Te mpe ra ture s e ns or output
3 GND
4 VBATNP + Ba tte ry be fore the rma l fus e (fus e bre a kdown de te ction)

2.1.2.8 J11 RS232/Nurse Call

P in De s c rip tio n
1 TxD-P C Da ta Re ce ption from P C
2 RTS -P C Re que s t to s e nd from P C (S hould be pos itione d a t high s ta te )
3 DTR-P C Da ta Te rmina l Re a dy of P C (S hould be pos itione d a t high s ta te )
4 RXD-P C Da ta Tra ns mis s ion to P C
5 GND
6 Nurs e Ca ll Norma lly Ope ne d Conta ct
7 Nurs e Ca ll Common Conta ct

NT1789 Rv.A0 page : 13


8 Nurce Ca ll Norma lly clos e d Conta ct

2.1.2.9 J 10 - D.C. motor

P in De s c rip tio n
1 + DC motor.
2 - DC motor.
2.1.2.10 J2 Sensors

P in De s c rip tio n
1 GND
2 P re s -Cla mp S witch cla mp output
3 CD-Cla mp S witch cla mp powe r s upply
4 Up-Force S e ns or Ups tre a m force s e ns or output
5 Up-Force S e ns or Upstream force sensor input
6 Up-Force Sensor Upstream force sensor output
7 Up-Force Sensor Upstream force sensor + input
8 GND
9 Hall effect sensor Hall effect power supply
10 Hall effect sensor GND
11 Hall effect sensor Hall effect output
12 Dw-Force Sensor Downstream force sensor output
13 Dw-Force Sensor Downstream force sensor input
14 Dw-Force Sensor Downstream force sensor output
15 Dw-Force Sensor Downstream force sensor + input
16 GND
17 ET-OPTO Outlet opto signal
18 CD-LED LED driver
19 GND Opto ground
20 5V LED Power supply

2.1.3. Electronic drawings


On request at Fresenius Vial After Sales Service department
2.1.4. Components layout
On request at Fresenius Vial After Sales Service department

P a ge 14 NT1789 Rv.A0
2.2. Dis p la y b o a rd

2.2.1. Functional description

The dis pla y boa rd is cons ide re d a s a pe riphe ra l boa rd of the CP U boa rd.
The ke yboa rd ma trix is conne cte d to this boa rd.

Dis pla y de s cription


Fu n c tio n Dis p la y
Flow ra te m/l 4 gre e n 7-s e gme nt LED dis pla y
Ala rm 1 re d LED
P re -a la rm 1 ora nge LED
Perfusion signal LEDs 4 green LED
Drop LEDs 1 green LED
Battery LEDs 1 green LED
Mains LEDs 1 orange LED
Perfusion start LEDs 1 green LED
LCD 33x100 points LCD display

2.2.2. Connector description


2.2.2.1 Ge ne ra l
J1 1 x 20 Pins CPU/Display boards Interconnection
J2 18 Pins LCD connector
J3 10 Pins Keyboard connector

NT1789 Rv.A0 page : 15


2.2.2.2 J 1 - CP U/Dis pla y Boa rd inte rconne ction
P in De s c rip tio n
1 5V P owe r S upply
2 VBAT Ba tte ry volta ge
3 GND
4 MOS I-AFF S e ria l output
5 GND
6 MIS O_AFF S e ria l Input
7 GND
8 CLK-AFF S e ria l clock
9 GND
10 CS LCD Chip S e le ct LCD
11 GND
12 CS AFCL Chip S e le ct Dis pla y
13 GND
14 A0AFF LCD Addre s s
15 GND
16 RES ET/
17 GND
18 TON/OFF ON/OFF s igna l
19 VBAT Ba tte ry Volta ge
20 5V P owe r s upply

P a ge 16 NT1789 Rv.A0
2.2.2.3 J 2 LCD Conne ctor

P in De s c rip tio n
1 VBAT + ba ck light
2 -le d - ba ck light
3 +5V
4 /RES ET
5 VS S Gnd
6 LD0 Da ta Write
7 GND Write S e le ct
8 LD6 Da ta Write
9 A0AFF LCD Addre s s
10 LD5 Da ta Write
11 LD4 Da ta Write
12 LD3 Da ta Write
13 LD2 Da ta Write
14 LD1 Da ta Write
15 LD0 Da ta Write
16 - Ena ble
17 LCDCS 2 Chip s e le ct
18 LCDCS 1 Chip s e le ct
2.2.2.4 J 3 ke yboa rd Conne ctor

P in De s c rip tio n
1 COL2 Ke yboa rd Column
2 COL1 Ke yboa rd Column
3 COL0 Ke yboa rd Column
4 Ligne 3 Ke yboa rd Line
5 Ligne 2 Ke yboa rd Line
6 Ligne 1 Ke yboa rd Line
7 Ligne 0 Ke yboa rd Line
8 TON/OFF On / Off
9 CONTRLE Keyboard Control line
10 GND

2.2.3. Electronic drawings


On request to Fresenius Vial After Sales Service department

2.2.4. Components layout


On request to Fresenius Vial After Sales service department

NT1789 Rv.A0 page : 17


2.3. P OWER S UP P LY BOARD

2.3.1. Functional description


An AC / DC 12W Module convertor (switching mode power supply) convert the mains input voltage ( 100
240 Vac) to 9Vdc 1.3A. This voltage is used to charge the battery and power the device.

ECG/EEG noise rejection filter : 2 Class X2 Capacitor and a 10Kohms resistance connected to a functional
earth garanteed the EEG/EEG noise rejection
2.3.2. Connector description
2.3.2.1 Ge ne ra l

Co n n e c to r P in Fu n c tio n
J1 1x1 Functional Earth
J2 1x2 Main supply connector
J3 1x3 Output power supply
J4 1x4 Battery Output
J5 2x5 Battery Input

2.3.2.2 J 1 Functiona l Ea rth

P in De s c rip tio n
1 - Functional Earth

P a ge 18 NT1789 Rv.A0
2.3.2.3 J 2 Ma in S upply Conne ctor

P in De s c rip tio n
1 PH Phase (100 240 Vac)
2 N Neutral (100 240 Vac)

2.3.2.4 J 3 Output P owe r s upply

P in De s c rip tio n
1 9V Sect 9V Output from AC/DC converter
2 GND
3 9V Ext 9V From external power Supply (not used)

2.3.2.5 J 4 Ba tte ry Output ( this conne ctor goe s to the CP U boa rd)

P in De s c rip tio n
1 VBAT + Battery
2 + Capteur Temp Temperature Sensor output
3 Masse GND
+ Battery before thermal fuse (fuse breakdown de-
tection)
4 VBAT NP

2.3.2.6 J 5 Ba tte ry Input ( this conne ctor come s from the ba tte ry pa ck)

P in De s c rip tio n
1 VBAT + Battery
+ Battery before thermal fuse (fuse breakdown de-
2 VBAT NP tection)
3 + Capteur Temp Temperature Sensor output
4 - Capteur Temp Temperature Sensor GND
5 Masse Battery GND

2.3.3. Electronic drawings


On request to Fresenius Vial After Sales Service department

2.3.4. Components layout


On request to Fresenius Vial After Sales Service department

2.4. Air d e te c to r b o a rd
The air sensor is based on an ultrasonic sensor.
2.4.1. Electronic drawings and Components layout
The air sensor is based on an ultrasonic sensor.

NT1789 Rv.A0 page : 19


3. Co n fig u ra tio n , c a lib ra tio n a n d c h e c k
The configura tion a nd ca libra tion of the de vice a re a cce s s ible :
us ing the s e ria l link (RS 232) a nd the Fre s e nius Via l P C s oftwa re to configure a nd ca libra te the pump.
Us ing the ke yboa rd of the pump through a s pe cia l e ncode d me nu.
For furthe r informa tion, ple a s e conta ct our Ma inte na nce De pa rtme nt (S e e a ddre s s e s a t the e nd of this docu-
me nt)
Any cha nge , in the configura tion le a ds to ma jor cha nge s in Module running, S ta ff tra ine d on the s oftwa re a nd
de vice s a re the only pe ople a uthoris e d to pe rforme d the s e s ope ra tions .
3.1. Co n fig u ra tio n
This mode is a ctive d by pre s s ing MODE ke y whe n s witching ON the pump/
The ma in me nu dis pla y :
- Us e r s e tting
- Wa rd s e tting
- Ma inte na nce
Each menu is composed with some sub-menu , press confirm key or enter to get through.
Some menu need a code to access to the sub-menu

The access code of user setting can be defined or disabled in the ward setting.
Once the access code is typed it is not necessary to type it again as long as the device is not switched off.
The 7 segments displays are off during the display of the main menu and codes typing.

Code
Ma in me nu S ub-me nu S ub-me nu
Typing
Ma in te n a n c e
This code will be give n a t the e nd of the ma inte na nce s ta ff tra ining
Te s t
Ma inte na nce te s t me nu
Ma in te n a n c e c o n fig u ra -
tio n
Ma inte na nce configura tion
me nu
Ca lib ra tio n
Ma inte na nce ca libra tion me nu
Us e r s e ttin g ( see operators guide)
Code XXXX
User configuration menu
Wa rd s e ttin g ( see operators guide)
Code 0200
Service configuration menu

P a ge 20 NT1789 Rv.A0
Test 7VW Test1 Work duration display
7VW Test2 LEDs, LCD screen, 7- segment displays and buzzers test.
7VW Test3 Keyboard test.
7VW Test4 Battery
Battery tension display
voltage display
7VW Test5 Last 10 alarms display.
7VW Test6 Total work duration display.
7VW Test9 Analog sensors value display.
Ma inte na nce

7VW$ TestA Software version number display.


7VW& TestC Clamp
Clamp motor test.test.
DC motor
7VW( TestE Calibration values display.
7VW) TestF Infusion stepper motor test.
7VW- TestJ Before failure last 10 alarms codes display
Configuration 6$9 SAV1 Preventive maintenance
6$9 SAV2 Tubing Set configuration
Etalonnage
Calibration Etal1 Door calibration.
(7$
(7$ Etal2 Infusion correction coefficient configuration.
(7$ Etal9 Battery
Up-stream and down-stream pressure sensors calibration.

Ma in m e n u

Us e r S e tting
Wa rd S e tting
Ma inte na nce

Acce s s code type

Us e r S e tting

Code 0000

Ma inte na nce

Code 0000

Wa rd S e tting

Code 0000

Code s e le ction with S e le ction ke ys ( or )


P roce e d to s ub-me nus with c o n firm ke y or ENTER ke ys .
Code ba ck to 0 if incorre ct code va lida tion, a fte r 15 s e conds non va lida tion = be e p.
S TOP = ba ck to ma in me nu.

Ma in te n a n c e s u b -m e n u

Te s t
Configura tion
Ca libra tion

NT1789 Rv.A0 page : 21


3.2. Ca lib ra tio n
Ca lib ra tio n is p o s s ib le o n :
Down-s tre a m pre s s ure s e ns or.
Up-s tre a m pre s s ure s e ns or.
Door Ha ll e ffe ct s e ns or..

Ca lib ra tio n is c o m p u ls o ry in th e fo llo win g c a s e s


Down-s tre a m pre s s ure s e ns or :
Afte r cha nge of s e ns or, of its prote ctive dia phra gm, of the door or CP U boa rd
Up-s tre a m pre s s ure s e ns or :
Afte r cha nge of s e ns or, of its prote ctive dia phra gm, of the door or CP U boa rd
Door Ha ll e ffe ct s e ns or :
Afte r cha nge of s e ns or, of the door or CP U boa rd
q The CP U boa rd is configure d with s pe cific pa ra me te rs for e a ch De vice , s o it is forbidde n to cha nge CP U
boa rds from a de vice to a n othe r.

q For the s a me re a s ons , it is forbidde n to cha nge door from a de vice to a nothe r

3.2.1. Manual calibration menu


S ELECTION ke ys = ca libra tion choice
CONFIRM or ENTER ke ys = ca libra tion proce dure
S TOP ke y = ba ck to ma inte na nce me nu. Va lue s a re not me moris e d.
Ca libra tion numbe r is dis pla ye d on 7 s e gme nts dis pla y a nd s imulta ne ous ly dis pla ye d in inve rs e vide o on LCD
s cre e n.

(W$ Eta l1-Door


Eta l2-Ra te
Eta l9-P re s s ure

Ca libra tions lis t :


Eta l1 : door.
Eta l2 : Infus ion corre cting coe fficie nt configura tion.
Eta l9 : Up a nd Down-s tre a m pre s s ure s e ns or.

3.2.2. Door calibration (Etal1)

q 3 le ve ls a re ca libra te d
q Note : P re s s ing S TOP a llows to e s ca pe from the me nu without me moris a tion.
q Door clos e d without s e t :

(W$
Clos e the door without s e t, re cord the va lue by pre s s ing o n or ENTER ke y.

q Door ope n without s e t :

P a ge 22 NT1789 Rv.A0
(W$
Ope n the door without s e t, re cord the va lue by pre s s ing on or ENTER ke y.
q Door clos e d with s e t :

(W$
P la ce a s e t with the cla mp corre ctly pos itione d a nd clos e the door, re cord the va lue by pre s s ing on
or ENTER ke y.

Va lue s :
Door clos e d without s e t :
1196 < X < 3750 mVolts

Door ope ne d without s e t :


2197 < X < 2695 mVolts

Door clos e d with s e t :


1699 < X < 3398 mVolts

3.2.3. Correcting coefficient adjustment (Etal2)


This function a llows to a djus t the flow ra te of the de vice a ccording to the us e d s e t.
The adjustment range allows a correction up to 20% +10%
Measure the flow rate for a flow rate of 100 ml/h. Install the tube full with non mineral water, the container
should be at 50cm above the device. Place a weight balance (0.01 g resolution and accuracy) at the same
level as the device. Place a needle at the end of the tubing set. Collect on the balance the infused water. Se-
lect the rate and infuse for around 15 min. Press STOP on the device. Reset the balance (000g). Press
START on the device (the pump infused at 100 ml/h). Infused for at least 15 min. Stop the infusion. Note the
infused volume and time. Calculate the measured flow rate ( volume collected on the balance / infusion time)

The Correcting coefficient adjustment is performed in Etal2.

Selected 100ml/h
(7$ Measured..100ml/h

Selected 100ml/h
(7$ Measured..105ml/h

Selected 100ml/h
(7$ Measured..125ml/h
Error

S ELECTION key ( or ) : selects the programmed flowrate(100, increment 1 ml/h) and actual flow rate
(80 120 ml/h).
ENTER key : to select a field value.
c o n firm ke y : ca lcula te re cord the ne w coe fficie nt if included between 0.8 and 1.1 times the programmed
one, back to the configuration menu otherwise Error is displayed.

NT1789 Rv.A0 page : 23


S TOP ke y : ba ck to the me nu without ca lcula tion of the coe fficie nt.

Note : Afte r flowra te a djus tme nt a control of the flowra te s hould be done
3.2.4. Up and Down stream pressure sensors calibration (Etal9)
This calibration should be done with the set selected on the pump in SAV2. sea 3.4.3, Set number
Configuration (SAV2)
The set has a relaxation time when the door is closed, so follow precisely the following procedure to guaran-
tee the occlusion alarm level.

Ba g

1m P UMP

Ma nome te r
3 wa ys ta p

50 cm

ta nk

Te s t s e t

The manometer and the device should be placed at the same level

RAZ Presure
q Ca lib ra tio n calibration NO
Complete calibration choice YES or NO through the selection key . If Yes, the force sensor calibration values
are set to zero for all the tubes
YES choice is mandatory for any replacement of the force sensor door or CPU board
NO choice does not set to zero the force sensor calibration value for the other tubes, only the one selected on
the pump is set to zero. See 3.4.3, Set number Configuration (SAV2)
The following procedure allows to carry out the calibration of the active tube.
3.2.4.1 Door ope n

1. Open the door of the device with o u t disposable set.


2. Record the values door open down-stream pressure and door open up-stream pressure by
pressing on ENTER

(W$
Values :
door open down-stream pressure threshold :
1997 < X < 4453 mVolts

door open up-stream pressure threshold :


1997 < X < 4297 mVolts
3.2.4.2 Ca libra tion of the down-s tre a m pre s s ure s e ns or a t pre s s ure ze ro (0 mmHg), Door clos e d.

P a ge 24 NT1789 Rv.A0
1. Place a disposable infusion set as described here above and open the 3 ways tap between the pump and
the manometer.

2. To relax the infusion set, press on the confirm key -> the pump run at 200 ml/h.

3. When the pressure reaches about 750 mmHg, open the 3 way tap to release the pressure.

4. Let infuse at zero pressure for about 5 minutes.

5. Stop the device by pressing on STOP.

6. Record the pressure by pressing ENTER;

(W$
Va lue s :
down-s tre a m pre s s ure s e ns or a t pre s s ure ze ro (0 mmHg), Door clos e d :
1499 < X < 3696 mVolts

3.2.4.3 Ca libra tion of the down s tre a m pre s s ure s e ns or a t 750 mmHg, Door clos e d;

1. Clos e the ta p be twe e n the pump a nd the ma nome te r.

2. S ta rt the pump by pre s s ing c o n firm ke y.

3. Ope n a nd clos e the ta p until the ma nome te r pre s s ure re a che s 750 mmHg 30 mmHg

4. Stop the pump by pressing on S TOP key.

5. Record the pressure by pressing on ENTER key.

(W$
Values :
down stream pressure sensor at 760 mmHg, Door closed
898 < X < 2998 mVolts

3.2.4.4 Ca libra tion of the up s tre a m pre s s ure s e ns or a t 0 mmHg

1. Open the tap between the pump, the manometer and tank.

2. Wait until the pressure set to zero.

3. Infusion set still in place and door closed, press ENTER key to record the value.

Values :
(W$
up stream pressure sensor at 0 mmHg
1499 < X < 3696 mVolts

NT1789 Rv.A0 page : 25


3.3. Te s ts : Ma in te n a n c e te s ts

3.3.1. Manual Service test menu


Choice with S ELECTION ke ys
P roce e d with c o n firm or ENTER ke ys .
Ba ck to ma inte na nce me nu with S TOP . (no pa ra me te r s tore d)
Te s t numbe r is dis pla ye d on 7 s e gme nts dis pla y a nd s imulta ne ous ly on LCD s cre e n.
3.3.2. Maintenance tests list :

 

Te s t1 : Dis pla y of working dura tion a nd numbe r of s witching on.


Te s t2 : LEDs, LCD screen, 7 segments display and buzzers test.
Test3 : Keyboard test .
Test4 : Battery voltage display.
Test5 : Last 10 alarms codes display.
Test6 : Total working duration and total number of switching on.
Test9 : Analog sensor value display.
TestA: Software version number display.
TestC : Clamp dc motor test.
TestE : Calibration values display.
Test F: Infusion motor test.
TestJ : Last 10 alarms before fail codes display.

3.3.3. Working duration (Test1)


Dis pla y the cumula te d pump working time in hours (<72 hours ) in da ys (<120 da ys ) or in months a nd the
ma inte na nce da te .
Fla s hing dis pla y &WUO, if pump working time is time be fore ma inte na nce (s e e S AV1) or if a ctua l da te
ma inte na nce da te (S AV1).

W6W

ENTER : selects date


S ELECTION : changes maintenance date in actual date or restores configured maintenance date.
(if maintenance date = actual date then display = 0 h and Ctrl = erasing)
Co n firm and ENTER : saves changes and goes back to test menu.
S TOP : goes back to test menu with erasing of change of maintenance date.
If date of maintenance is changed then working duration is set to zero.

P a ge 26 NT1789 Rv.A0
3.3.4. LEDs, LCD screen, 7 segments display and Buzzers Test (Test2)
Lighting of all the LEDs, display of  plus decimal point on the 7 segments display, LCD screen completely
black; wait for pressing on ON or ENTER.


Scrolling of LED, one by one, scrolling of segments one by one on all the display then scrolling of  and
decimal point on each display. Scrolling of LCD screen image.

Back to test menu (manual test).

S TOP : quick return to test menu (manual test).

3.3.5. Keyboard test (Test3)

W6W

Display of pressed keys in inverse video.


Co n firm or ENTER : back to test menu.
3.3.6. Battery voltage display (Test4)

W6W
W6W
Co n firm o r ENTER : back to test menu.

NT1789 Rv.A0 page : 27


3.3.7. Display of the last 10 alarms codes (Test5)

W6W
W6W
S ELECTION : Code s s crolling.
Co n firm or ENTER : ba ck to te s t me nu.
3.3.8. Total working duration (Test6)
Display the Total pump working duration independent of Test 1. The Display duration cant be set to zero.

3.3.9. Display analog values of sensors (Test9)

W6W
W6W
Display of analog input values in form 0-5000 for 0-1024 Lsb.
No display for up-stream sensor if not used.
Co n firm or ENTER: back to test menu
3.3.10. Display of software version number (TestA)

W6W$

W6W$
Co n firm or ENTER : proceeds to language version then back to test menu.
S ELECTION : language scrolling.

P a ge 28 NT1789 Rv.A0
3.3.11. Motor clamp Test (TestC)

W6W&
S ELECTION : function choice .
Co n firm or ENTER : Run cla mp or uncla mp.
S TOP : s tops motor a nd re turn to te s t me nu if motor a lre a dy s toppe d.
Cla mp: the dc motor turns forwa rd to cla mp.
Uncla mp : the dc motor turns ba ckwa rd to uncla mp.
If problem on clamp motor driver then LCD screen black and Er31 displayed on 7 segment display.
S TOP : to return to test menu.

3.3.12. Calibration values display (TestE)

W6W(
S ELECTION : scrolling of calibration values.
Co n firm or ENTER : back to test menu
Display of calibration values in form 0-5000 for 0-1024 Lsb.
The calibration values list is as following:
Por1 : Hall effect sensor, door closed, without set.
Por2 : Hall effect sensor, door opened.
Por3 : Hall effect sensor maximum value , door closed, with set.
Pam1 : Upstream pressure, door open.
Pam2 : Upstream pressure, 0 mmHg.
Pav1 : Downstream pressure, door open.
Pav2 : Downstream pressure, 0 mmHg.
Pav3 : Downstream pressure, 750 mmHg.

The value between brackets gives the number of calibrations done.


Display of upstream and downstream pressure calibration values in accordance with active set number.

NT1789 Rv.A0 page : 29


3.3.13. Infusion motor Test (TestF)
This te s t a llows to s e t the motor running without a la rm ma na ge me nt.

W6W)
S ELECTION : selection of infusion flow rate (0.1 ml/h 1000 ml/h).
ENTER : flow rate validation.
S TOP : stops motor and return to test menu if motor already stopped
If motor problem then LCD screen blank and Er01, Er11, or Er21 displayed on 7 segment display.
S TOP : to return to test menu.

3.3.14. Display of last 10 alarms before a failure occurs (TestJ)

W6W-
W6W-
S ELECTION : scrolling number of switching on then error and codes.
Co n firm or ENTER: back to test menu.

P a ge 30 NT1789 Rv.A0
3.4. Ma in te n a n c e S e rvic e Co n fig u ra tio n .

3.4.1. Service configuration Menu.


Dis pla y of configura tion numbe r on 7 s e gme nts dis pla y a nd coloure d inve rs e on the LCD s cre e n a t the s a me
time .

S ELECTION : configura tion s e le ction.


Co n firm or ENTER: configura tion e xe cution.
S TOP : re turn to te s t me nu.

6$9
S e rvice configura tions lis t :
S AV1 : pre ve ntive ma inte na nce .

S AV2 : s e t numbe r configura tion.


Dis pla y of configura tion numbe r on 7 s e gme nts dis pla y a nd coloure d inve rs e on the LCD s cre e n a t the s a me
time .

3.4.2. Preventive maintenance (SAV1)

6$9

S ELECTION : numbe r of hours be fore ma inte na nce or da te of ma inte na nce s e le ction.


ENTER : cha nge s one fie ld to a nothe r.
Co n firm : ba ck to S e rvice configura tion me nu with re cording of ne w dura tion a nd da te .
S TOP : e ra s e s the la s t cha nge s a nd goe s ba ck to S e rvice configura tion me nu.
Choice of numbe r of hours be fore ma inte na nce be twe e n 1 a nd 9999 hours .
0 inhibits the re que s t of pre ve ntive che ck with the working dura tion.
Choice of the da te be twe e n 01/01/1990 a nd 31/12/2079.
One of the thre e numbe rs = 0, inhibits the re que s t of pre ve ntive che ck with the da te .

3.4.3. Set number Configuration (SAV2)

6$9
Dis pla y of diffe re nt s e ts a va ila ble for the pump type .
A che ck ma rk s hows the s e ts a lre a dy ca libra te d.
S ELECTION : move s up or down in the s e t lis t.
Co n firm or ENTER: s a ve s the s e t s e le ction (coloure d inve rs e s e le ction)
S TOP : ca nce l the la s t s e le ction a nd re turn to ca libra tion me nu.
If a non-ca libra te d s e t is s e le cte d, a t the ne xt s witching on, the de vice is locke d in occlus ion a nd/or ups tre a m
pre s s ure a la rm, re quiring ca libra tion.

NT1789 Rv.A0 page : 31


3.5. Te c h n ic a l fa ilu re s m e s s a g e s .
Any te chnica l fa il us e initia te s a n e rror code a s we ll a s a continuous s ound a la rm.
An e rror code is dis pla ye d on the 7 s e gme nts dis pla y.

Erro r c o d e s Err xx De s c rip tio n


10 Inte rna l RAM proble m
20 Exte rna l RAM proble m
30 Incohe re nt EP ROM che ck s um
40 EEP ROM proble m
50 ADC proble m
60 Incohe re nt infus ion s e t pa ra me te r
70 Motor fre que ncy incompa tibility.
80 Ke yboa rd proble m.
01 Motor rota tion e rror.
11 Motor rota tion e rror.
21 Motor rota tion e rror.
31 DC Motor.
41 Motor rota tion e rror.
03 S e ria l link tra ns mis s ion e rror
14 Error on motor pe riod ca lcula tion pa ra me te rs ve rifica tion
24 Motor rota ry dire ction incompa tibility.
44 Qua rtz fre que ncy incompa tibility.
54 dis pla y boa rd mis s ing.
74 No ba tte ry conne ction.
84 Ba tte ry cha rge e rror
56 S oftwa re proble m.
28 Non compa tible la ngua ge file .
55 clock fa ilure
94 Ba tte ry te mpe ra ture e rror
55 S e conda ry microproce s s or e rror
85 S e conda ry microproce s s or e rror
16 Da te / time che ck e rror

Ala rm c o d e s A xx De s c rip tio n


10 Ba tte ry
12 End of infus ion
16 Downs tre a m Occlus ion
17 Air de te ction
18 Air de te ction
19 Ups tre a m Occlus ion
20 No tubing s e t
21 No cla mp

P a ge 32 NT1789 Rv.A0
Ala rm c o d e s A xx De s c rip tio n
22 Door
23 No drip s e ns or
24 Drip s e ns or : low flowra te
25 Drip s e ns or : high flowra te
26 Fre e flow
27 Occlus ion control

NT1789 Rv.A0 page : 33


4. REP LACING S UB-AS S EMBLIES

4.1. Mo d u le o p e n in g .
Whe n ope ning the OP TIMA, a nti-s ta tic pre ca utions mus t be s trictly a pplie d.
The us e of a nti-s ta tic ca rpe t a nd grounde d bra ce le ts is highly re comme nde d.
Dis conne ct ba tte ry a s s oon a s the Module is ope ne d.
4.2. Re p la c in g CP U b o a rd .
In ca s e of CP U boa rd re pla ce me nt, s ys te ma tica lly pe rform configura tion a nd ca libra tion a s de s cribe d in
cha pte r 3.
4.3. Dis m o u n tin g p u m p in g m e c h a n is m .
Re tra ct the door a xis a nd re move the door. Re move the s ticke r up to the pumping me cha nis m to uns cre w
a xis holde r.
Uns cre w the 4 fixing s cre ws of the pumping me cha nis m.
4.4. Re p la c e m e n t o f p re s s u re s e n s o rs .
In ca s e of re pla cing pre s s ure s e ns or, s ys te ma tica lly pe rform configura tions a nd ca libra tions de s cribe d in
cha pte r 3.
Be e xtre me ly ca re ful within the dire ction of the s e ns or pos itioning (le a d output).
4.5. Re p la c e m e n t o f d o o r o r o f Ha ll e ffe c t s e n s o r
In ca s e of re pla cing the s e e le me nts , s ys te ma tica lly pe rform configura tions a nd ca libra tions de s cribe d in
cha pte r 3

P a ge 34 NT1789 Rv.A0
5. MAINTENANCE

5.1. Re c o m m e n d a tio n s
The qualified technicians in your establishment or our Maintenance Service should be notified of any abnormal opera-
tion of the device.
For further information concerning troubleshooting or usage procedure, please contact our Maintenance Department or
our Sales Representative. (see Useful Addresses, chapter 9)
If the device has to be returned to our Maintenance Service, it must be disinfected and packed very carefully, if possible
in its original packaging.
FRESENIUS VIAL INFUSION SYSTEMS is not liable for loss or damage to the device during transport to our Mainte-
nance Service.

5.2. Cle a n in g a n d d is in fe c tin g


The Module DPS and the FRESENIUS VIAL Infusion Station form part of the patients immediate environment. It is
advisable to clean and disinfect the devices external surfaces on a daily basis in order to protect patient and staff.
q Dis conne ct the FRES ENIUS VIAL OP TIMA P ump from the ma ins s upply be fore s ta rting to cle a n.
q Do not pla ce in a n AUTOCLAVE nor IMMERS E the de vice . Do not le t liquids e nte r the de vice s ca s ing.
q Us e a cloth s oa ke d in DETERGENT-DIS INFECTANT, pre vious ly dilute d with wa te r if re quire d.
Avoid a bra s ive s crubbing which could s cra tch the ca s ing
Do not rins e or wipe s urfa ce s .
q If the de vice is pla ce d in a high conta mina tion ris k unit, it is a dvis a ble to le a ve it in the room during fu-
miga tion dis infe cting, a fte r ha ving dis infe cte d it with a mois t cloth.
q Do not us e :
TRICHLOROETHYLENE-DICHLOROETHYLENE,
AMMONIA,
AMMONIUM CHLORIDE,
CHLORINE a nd AROMATIC HYDROCARBON,
ETHYLENE DICHLORIDE-METHYLENE CHLORIDE,
CETONE
These aggressive agents could damage the plastic parts and cause device malfunction.
q Ta ke ca re a ls o with ALCOHOL BAS ED S P RAYS (20% - 40% a lcohol). The y le a d to ta rnis hing a nd
s ma ll cra cks in the pla s tic, a nd do not provide the ne ce s s a ry cle a ning prior to dis infe cting.
q For furthe r de ta ils , conta ct the s e rvice in cha rge of cle a ning a nd dis infe cting products of your e s ta b-
lis hme nt .

5.3. S to ra g e
The device should be stored in a dry and cool place. In case of prolonged storage time, disconnect the battery through
the battery door located on the rear panel . This operation should be made by qualified technician.
When storage a full recharge of the battery is recommended before restarting the pump, in order to avoid any risk
caused by micro mains supply cuts and to insure maximum autonomy.
Storage place should be :
Temperature between -10c and 60c
Maximum relative humidity 90% no condensation.

NT1789 Rv.A0 page : 35


5.4. Ro u tin e in s p e c tio n s

5.4.1. Checking before use


(Refer to operating instructions)
In order to ensure optimal work of the device, regular servicing inspections and tests are recommended.
A servicing check procedure should include the points listed below. These service checks are not covered by any con-
tract or agreement with FRESENIUS VIAL INFUSION SYSTEMS and are the responsibility of the Users technical
staff. For further information, please contact our Maintenance Service.
q Note: Failure to comply with thes e maintenance procedures could damage the device and lead
to a functional failure.

5.4.2. Occlusion alarm checking


1. Ma ke the following ins ta lla tion.

1m 50cm
P UMP

Ma nome ter
bypa s s

2. P rogra m a 125 ml/h flow ra te a nd a counte r pre s s ure a la rm a t 500 mmHg.


3. S ta rt infus ion with bypa s s clos e d be twe e n ma nome te r a nd pump
4. Afte r 5 minute s of infus ion a t 0 pre s s ure , ope n bypa s s be twe e n the ma nome te r a nd the pump.
5. Whe n a la rm goe s , che ck the pre s s ure P dis pla ye d by the ma nome te r is corre ct :
P = 500 mmHg 75 mmHg -> 0.65 + 0.1 ba r
6. Do the te s t a ga in with a progra mme d counte r pre s s ure of 750 mmHg.
P = 750 mmHg 110 mmHg -> 1 + 0.15 ba r

P a ge 36 NT1789 Rv.A0
5.4.3. Battery autonomy test
1. The battery autonomy is minimum 4h for usual flow rates (< 125 ml/h).
2. The battery(when properly charged) pre alarm gives a warning to the user about 30 minutes before total stop of
the device.

q Note : The ba tte ry cha rge dura tion is minimum 16 hours ( 100 % of its ca pa city) whe n the pump is ON.

q Note : The ba tte ry cha rge dura tion is minimum 5 hours ( 100 % of its ca pa city) whe n the pump is off.

5.4.4. Fix clamp checking ( VS/PT and ST pumps)


1. Ma ke the following ins ta lla tion.

Ma nome te r De vice Infus ion s e t

S yringe Drip cha mbe r

2. Ope n the door of the de vice , the in line me cha nica l cla mp s hould cla mp the infus ion s e t.
3. Apply a 0.5 ba r pre s s ure with the s yringe a nd che ck n o liq u id goe s in the drip cha mbe r.

5.4.5. Air detector check


Note : Do the te s t with the tubing s e t dis pla ye d on the de vice . We re comme nd to do the te s t with the
diffe re nts type s of tubing s e ts us e d in the hos tipa l.
1. P la ce a n infus ion s e t full of wa te r in the de vice .
2. For a n a ir bubble s volume progra mme d a t 0.20 ml.
3. P rogra m a 60 ml/h flow ra te .
4. Ge t drip cha mbe r ups ide down for a bout 20 s e conds ( to infus e a bout 0.5 ml a ir bubble )
5. Che ck the a ir a la rm goe s whe n a ir bubble s pa s s through the a ir de te ctor. The le ngth pa s s ing through
be fore a la rm s hould a pproxima te ly corre s pond to the s ize s e le cte d (1.3cm / 0.1ml).

5.4.6. Continuity test


1 Us ing a Ohmme te r, ve rify the pre s e nce of the e le ctrica l re s is tor highe r tha n 10 M ohms be twe e n

Ne utra l a nd drip s e ns or plug (a ny pin)


P ha s e a nd drip s e ns or (a ny pin)

NT1789 Rv.A0 page : 37


5.5. Te c h n ic a l c h e c k p ro c e d u re
In order to ensure follow-up within the framework of a preventive maintenance, a routine inspection is recommended. It
is recommended to use the technical manual as a complement, for correct run of the procedure.
A checking software allows to automatically perform the following tests.
IMPORTANT : recharge battery ( 16 hours) prior to the check procedure

De vice : OP TIMA / MCM Type : Code : Device serial N :

Check date : ........ / ........ / ........ Depart : Technician :

Ch e c k Co n - YES NA
N Description fo rm NO
FUNCTIONAL CHECK ( see 3.3.)
1
q Che ck ge ne ra l outs ide ca s ing a ppe a ra nce , ma ins powe r cord a nd s ticke rs .
a nd pumping me mbra ne

2 q Dis pla y of working duration (test1) H/D/M

q Last maintenance date (test1) ..D/M/Y


3

4 q Total working duration ( test 6) H/D/M

5 q Lights, buzzers and displays check (test2)

6 q Keyboard check (test 3 )

7 q Door check
Put on the pump without infusion set, door closed, connect drip sensor.
Check the message on the alarm screen : Insert Tube
Place an infusion set full of water in the device and closed the door:
Check the absence of alarm.
Open the door:
Check the message : Door opened

8 OCS Te s t ( for pump OP TIMA MS a nd VS or MCM MS a nd OT)


Ins ta ll a tubing s e t a nd put on the de vice
Check that the OCS test passes correctly

Clamp Check ( all pumps)

Install a tubing set full of water (roller clamp opened) and bottle place at least 50 cm above
the pump. Close the door and then open the door.
Check the effective clamping of the tube (by the device for VS,PT,ST or by the clamp
on the MS sets for the MS pump) :no drips at the end of the infusion sets.

De te ction of the cla mp of the MS s e ts ( for MS type pumps )


Ins ta ll the MS s e ts prope rly in the pump. Turn on the MS pump
Che ck tha t the re is no a la rm dis pla y
Ope n the door a nd move the cla mp on the MS s e t outs ide the de vice . Clos e the door.
Check that no clamp.

P a ge 38 NT1789 Rv.A0
Ch e c k Co n - YES NA
N Description fo rm NO
9 q Drip s e ns or che ck (if pre s e nt)
Che ck s e ns or inte grity (chock ma rks , cle a nline s s of the de te ction zone , drip
cha mbe r ma inta ining s pring, good condition of the cord)
Conne ct the cord to the Module
Ins ta ll a n infus ion s e t full of wa te r
S ta rt infus ion
Che ck tha t the gre e n indica tor fla s he s whe n a drop is de te cte d.
Dis conne ct the s e ns or:
Check that the drip sensor alarm goes. ON
Connect the drip detector , set the infusion at 60 ml/h remove the chamber from the
drip detector . Verify the alarm should set : 40 sec < <2 min 30 sec
10
q Air bubbles sensor check (see 5.4.6)
When switching on, without infusion set:
Check that the red indicator and the air sensor of LCD display flash
Install an infusion set full of water
Check that the red indicator and the air sensor of LCD display switch off.
Select a 60-ml/h flow rate and start infusion.
Keep drip chamber up side down for, at least 20 sec. to generate a > 0.25 ml air
bubble for a device configured for a cumulated air detection of 0.25 in 15 min.
Check the alarm air bubble goe s on whe n bubble pa s s e s in the s e ns or.

11
q Down-s tre a m pre s s ure de te ctor.
Install an infusion set full of water with a manometer at the end (see 5.4.2)
Select a pressure threshold of 750 mmHg.
Start infusion at 100 ml/h.
Check that manometer indicates 750 mmHg 110 mmHg when alarm is triggered.
Do it again for a pressure threshold of 500 mmHg (. 75 mmHg)
12 Ave ra ge Flow ra te che ck (ca n a ls o be pe rform with a s ca le & IS De bit progra m te s t)
P la ce the infus ion s e t in a 100ml te s t tube .
S e le ct a 80ml/h flow ra te a nd infus e for 1 hour
Note the infus e d volume
The flow ra te ca n be a djus te d ( s e e 3.2.3)
76 ml < infused volume :.ml< 84 ml
13
q Ba tte ry che ck
Before checking, charge the battery for at least 16 hours
Start an infusion at 125 ml/h, m a in s d is c o n n e c te d
Check working autonomy is > 4h
IMP ORTANT : if autonomy < 3h30 , CHANGE battery.

14
q Continuity te s t
Neutral and Drip sensor connector (any pin) > 10Mohms
Phase and drip sensor connector (any pin) > 10 Mohms

Earth and equipotentiality < 0.1 Ohms


Earth and screw below the device < 0.1 Ohms

Checking result conform no conform

Visa :

NT1789 Rv.A0 page : 39


6. ANNEXE 1 : Illu s tra te d p a rts lis ts

6.1. S u b a s s e m b lie s tra c e a b ility ta b le

6.1.1. Introduction
The aim of this chapter is to help the technician to find spare parts so that the module can be serviced.

6.1.2. Spare parts table


The table below lists the mains modifications made to improve the product. The modules serial number should be
used when looking for components.

OP TIMA / MCM
1
e quiva le nce ta ble
From :18180100
Ins trume nt s e ria l no.
To:
Dis pla y boa rd 159602
CP U boa rd 159601
P owe r S upply boa rd 159603

P a ge 40 NT1789 Rv.A0
6.1.3. Upper housing

100

101

102 103
104 105 106 107

108 109 110 111

112

Re fe re n c e Dig . re f Qu a n tity Co m p o n e n t
159611 101 1 Uppe r hous ing
159620 103 1 Ke yboa rd
159602 102 1 Dis pla y boa rd
159601 112 1 CP U boa rd
159646 107 1 OCS motor
168402 110 1 P T / VS / OT a ir bubble boa rd
168403 110 1 S T a ir bubble boa rd
168938 111 1 Air s e ns or
174252 108 1 OCS foa m ring
168442 104 1 Air s e ns or hous ing
159899 100 1 Drop de te ctor

NT1789 Rv.A0 page : 41


6.1.4. Base

2 11

2 00 2 01

2 03

2 05 2 04

2 06 20 7

20 8

2 09

2 10

Re fe re n c e Dig . Re f. Qu a n tity Co m p o n e n t
159895 204 1 Ba s e
168480 206 2 P re s s ure s e ns or
168428 2 S e ns or clip
168439 208 2 S e ns or me mbra ne
651211 1 Me cha nica l bloc me mbra ne
159621 205 1 Le ft hook
168437 209 1 Right hook
159604 200 1 Fla t ca ble
159648 203 1 Motor
168479 211 1 Me cha nica l bloc
168482 1 Be lt
168483 1 P ump s ide ge a r
168489 1 Motor s ide ge a r
168485 3 S ile nt bloc 40 s hore s
159618 207 2 Fe e t
199261 210 1 P T / VS / OT door kit
199262 210 1 S T door kit

P a ge 42 NT1789 Rv.A0
159632 1 S witch cla mp holde r

6.1.5. Brace
300
301
306

302

303 307

304

305

306

Re fe re n c e Dig . Re f. Qu a n tity Co m p o n e n t
159612 303 1 Bra ce Optima
159613 304 1 S he lf Optima
159603 302 1 P owe r s upply boa rd
159622 305 1 Ba tte ry pa ck
159642 301 1 Nut
182108 306 1 S cre w ca p
159643 307 1 M8 s cre w
159617 300 1 Ha ndle

NT1789 Rv.A0 page : 43


P a ge 44 NT1789 Rv.A0
7. Circ u it d ia g ra m

Cicuit da igra ms a re a va ila ble ne a rby our s e rvice de pa rtme nt, ple a s e re fe r to us e ful a ddre s s e s cha pte r.

NT1789 Rv.A0 page : 45


8. Ad d e n d a a n d in fo rm a tio n b u lle tin s

P a ge 46 NT1789 Rv.A0
NT1789 Rv.A0 page : 47
9. Us e fu l a d d re s s e s
All reques t for information or documentation (technical file, tubing catalogue or commercial documentation)
s hould be addres s ed to :

CUSTUMER SERVICE INTERNATIONAL

Fre s e n iu s Via l Tel. : 33 (0)4 76 67 10 14 or 10 03


Le Gra nd Che min, 38590 Brzins FRANCE or 1067
Fax : 33 (0)4 76 67 11 12
E-ma il : comme rcia l.via l@fre s e nius -
he moca re .com

MAINTENANCE SERVICES

INTERNATIONAL Fre s e n iu s Via l Tel. : 33 (0)4 76 67 10 76


Le Grand Chemin, 38590 Brzins FRANCE Fax : 33 (0)4 76 67 11 22

BELGIUM Fre s e n iu s NV/S A Be lg iq u e Tel. : 32/388.07307


DIVIS ION VIAL MEDICAL Fax : 32/388.05007
Molenberglei 7
2627 Schelle
BELGIUM

GERMANY FRES ENIUS MCM Tel. : 49/60 23 97 22-0


AM-NEUNEN BERG 8 Fax : 49/60 23 43 06
63749 ALZENAU
GERMANY

It is pos s ible that this document contains typing errors or mis takes . Changes may occur at any time in s ubs equent editions .

COPYRIGHT 2002 Fresenius Vial

This technical manual may not be reproduced in whole or in part without the written consent of Fresenius Vial

Fresenius Vial SAS - Head Office: Le Grand Chemin - 38590 Brzins (FRANCE)
Capital: 13,744,520 EUR - SIREN Grenoble B 408 720 282.

P a ge 48 NT1789 Rv.A0