Professional Documents
Culture Documents
126883
Standards Used
Asset # Description Traceability Number Cal Date Due Date
MLM-ELE-058 Fluke 5522A/SC1100 Multi-Product Calibarot with Oscilloscope option 127168 2016-11-11 2017-11-11
MLM-TEM-017 Extech Instruments HW30 Humidity/Temperature Stopwatch 126885 2017-01-10 2018-01-10
Martin Posada
Calibration Tech./Tecnico de Calibracion
-Any questions, comments, suggestions or complaints regarding this service, or the content of this calibration report please send to sugerencias@masmetrologia.com
-Cualquier duda comentario, sugerencia o queja en relacion con este servicio, o el contenido de este informe de calibracion favor de dirigirlo a sugerencias@masmetrologia.com
MET-06.7-1/11 Pgina 1 de 3
No. 126883
Test Results
Test Description True Value Test Result Lower limit Upper limit Units Status Uncert ()
UUT IDENTIFICATION
DC VOLTAGE TESTS
100 mV Range
0.000 mV 0.0000 0.002 -0.006 0.006 mV Pass 9.4e-007 V
90.000 mV 90.0000 90.003 89.971 90.029 mV Pass 2.2e-006 V
1000 mV Range
900.00 mV 900.000 900.01 899.71 900.29 mV Pass 1.0e-005 V
300 mV Range
0.00 mV 0.000 0.00 -0.02 0.02 mV Pass 5.7e-006 V
300.00 mV 300.000 300.00 299.90 300.10 mV Pass 7.7e-006 V
3V Range
3.0000 V 3.00000 3.0000 2.9990 3.0010 V Pass 6.2e-005 V
-3.0000 V -3.00000 -3.0000 -3.0010 -2.9990 V Pass 6.2e-005 V
30 V Range
30.000 V 30.0000 30.000 29.990 30.010 V Pass 6.3e-004 V
300 V Range
300.00 V 300.000 300.01 299.90 300.10 V Pass 7.0e-003 V
1000 V Range
1000.0 V 1000.00 1000.0 999.5 1000.5 V Pass 5.8e-002 V
DIODE TEST
0.0000 Ohm 0.00000 0.0002 -0.0008 0.0008 Pass
Open Pass
AC VOLTAGE TESTS
300 mV Range
0.00 mV 0.000 0.07 0.00 0.75 mV Pass 5.7e-006 V
15.00 mV @ 1 kHz 15.000 15.01 14.87 15.13 mV Pass 8.4e-006 V
15.00 mV @ 100 kHz 15.000 14.68 13.75 16.25 mV Pass 4.9e-005 V
300.00 mV @ 1 kHz 300.000 300.02 299.30 300.70 mV Pass 3.9e-005 V
300.00 mV @ 100 kHz 300.000 303.97 284.50 315.50 mV Pass 2.0e-004 V
3 V Range
3.0000 V @ 1 kHz 3.00000 3.0005 2.9930 3.0070 V Pass 3.9e-004 V
30 V Range
30.000 V @ 1 kHz 30.0000 30.005 29.930 30.070 V Pass 3.9e-003 V
300 V Range
300.00 V @ 1 kHz 300.000 300.07 299.30 300.70 V Pass 4.5e-002 V
750 V Range
750.0 V @ 1 kHz 750.00 750.2 747.5 752.5 V Pass 1.8e-001 V
FREQUENCY TESTS
10.0000 kHz @ 1 V 10.00000 10.0000 9.9940 10.0060 k Hz Pass 5.9e-002 Hz
RESISTANCE TESTS
-Any questions, comments, suggestions or complaints regarding this service, or the content of this calibration report please send to sugerencias@masmetrologia.com
-Cualquier duda comentario, sugerencia o queja en relacion con este servicio, o el contenido de este informe de calibracion favor de dirigirlo a sugerencias@masmetrologia.com
MET-06.7-1/11 Pgina 2 de 3
No. 126883
Test Results
Test Description True Value Test Result Lower limit Upper limit Units Status Uncert ()
0.00 Ohm 0.000 0.00 0.00 0.04 Pass 5.7e-003 W
3 kOhm Range
0.0000 kOhm 0.00000 0.0002 0.0000 0.0002 k Pass 5.6e-002 W
30 kOhm Range
30.000 kOhm 30.0000 30.000 29.983 30.017 k Pass 9.7e-001 W
3 MOhm Range
3.0000 MOhm 3.00000 3.0001 2.9980 3.0020 M Pass 1.6e+002 W
30 MOhm Range
30.000 MOhm 30.0000 29.993 29.922 30.078 M Pass 7.6e+003 W
DC CURRENT TESTS
30 mA Range
30.000 mA 30.0000 30.000 29.982 30.018 mA Pass 2.5e-006 A
100 mA Range
100.00 mA 100.000 100.01 99.93 100.07 mA Pass 1.1e-005 A
10 A Range
10.000 A 10.0000 9.993 9.975 10.025 A Pass 4.2e-003 A
AC CURRENT TESTS
30 mA Range
30.000 mA @ 1 kHz 30.0000 30.002 29.840 30.160 mA Pass 1.0e-005 A
100 mA Range
100.00 mA @ 1 kHz 100.000 100.01 99.40 100.60 mA Pass 4.5e-005 A
10 A Range
10.000 A @ 1 kHz 10.0000 10.000 9.890 10.110 A Pass 9.1e-003 A
-Any questions, comments, suggestions or complaints regarding this service, or the content of this calibration report please send to sugerencias@masmetrologia.com
-Cualquier duda comentario, sugerencia o queja en relacion con este servicio, o el contenido de este informe de calibracion favor de dirigirlo a sugerencias@masmetrologia.com
MET-06.7-1/11 Pgina 3 de 3