You are on page 1of 3

Calibration Certificate / Certificado de Calibracin No.

126883

Metrologa Aplicada y Servicios S de RL de CV


Blvd. Manuel Gmez Morn 8009, Partido Senec C.P. 32470
Cd. Jurez, Chih., Mxico
Tels. (656) 617-6617, 207-3949 y 01-800-CALIBRA (225-4272)
www.masmetrologia.com
Accreditation # 22580
INSTRUMENT INFORMATION/INFORMACION DEL INSTRUMENTO CAL. CONDITIONS/CONDICIONES DE LA CALIBRACION

I.D.: MLM-ELE-024 PERFORMED ON: 16 enero 2017


DESCRIPTION: Dual Display Multimeter DUE DATE: 16 enero 2018
MANUFACTURER: Fluke TEMPERATURE: 24.00 C
MODEL: 45 HUMIDITY: 24 % H.R.
SERIAL NUMBER: 6769041 DATA TYPE: AS FOUND-AS LEFT
PROCEDURE NAME: Fluke 45: (1 year) CAL VER RS-232 /5520

CUSTOMER INFORMATION/INFORMACION DEL CLIENTE: INSTRUMENT CONDITIONS/CONDICIONES DEL INSTRUMENTO:

Metrologia Aplicada y Servicios, S de RL de CV Buenas Condiciones


Blvd. Gomez Morin #8009
Cd. Jurez, Chihuahua 32470
Calibration statement / Declaraciones Generales de la calibracin
Metrologa Aplicada y Servicios S de RL de CV certifies that the above listed instrument met or exceeds the published specifications unless otherwise noted and has
been calibrated in accordance with Metrologia Aplicada y Servicios Quality assurance manual that is accredited under ISO /IEC 17025:2005 requirements, This
Calibration report documents the traceability to National Institutes (NIST & CENAM) which realize the units of measurements according to the International System of
Units (SI), this report shall not be used to claim product endorsement by the accrediting body.
-This report may not be reproduced, except in full, unless permission for the publication of an approved abstract is obtained in writing for Metrologia Aplicada y
Servicios S. de R.L. de C.V. the user may be obliged to have the object recalibrated at appropriated intervals we recommend the use of OIML D 10, NCSL RP-1 or ISO
10012.
-Measurement uncertainties at the time of the test, are given on the following pages, where applicable. They are calculated in accordance with ISO /IEC Guide 98:1995
Guide to the expression of Uncertainty in Measurement (GUM).
-This calibration certificate may contain data that is not covered by our scope of accreditation. Unaccredited data, where applicable is indicated by an asterisk (*).
Metrologa Aplicada y Servicios S de RL de CV Certifica que el instrumento arriba descrito cumple o excede las especificaciones publicadas a menos que se indique
lo contrario y fue calibrado de acuerdo con el manual de aseguramiento de la calidad de Metrologia Aplicada y Servicios acreditado segun los requisitos de
ISO/IEC 17025:2005, Este informe de calibracion documenta la trazabilidad a Institutos Nacionales (NIST & CENAM) mismos que resguardan las unidades de
medida de acuerdo con el Sistema Internacional de Unidades (SI) El presente informe no podra ser usado para reclamar aprobacion de producto por parte de la
compaia que nos acredita.
-Este informe de calibracion no debe ser reproducido de forma parcial sin la autorizacion escrita de Metrologia Aplicada y Servicios S. de R .L. de C.V. , El usuario
puede verse obligado a tener por objetivo calibrar a intervalos apropiados para lo cual se recomienda se base en las recomendaciones OIML D 10, NCSL RP-1 o la
norma ISO 10012.
-Donde sea aplicable la estimacion de las incertidumbres al momento de la calibracion es expresada en las siguientes paginas y fueron estimadas en base a
ISO/IEC Guide 98:1995 Guide to the expression of Uncertainty in Measurement (GUM).
-Este Certificado de calibracin puede contener datos que no estn cubiertos por nuestro alcance de acreditacin. En su caso los datos no acreditados estn
marcados con un asterisco (*).

Standards Used
Asset # Description Traceability Number Cal Date Due Date
MLM-ELE-058 Fluke 5522A/SC1100 Multi-Product Calibarot with Oscilloscope option 127168 2016-11-11 2017-11-11
MLM-TEM-017 Extech Instruments HW30 Humidity/Temperature Stopwatch 126885 2017-01-10 2018-01-10

Calibrated By:/Calibrado por: Authorized By:/Autorizado por:

Martin Posada
Calibration Tech./Tecnico de Calibracion

-Any questions, comments, suggestions or complaints regarding this service, or the content of this calibration report please send to sugerencias@masmetrologia.com
-Cualquier duda comentario, sugerencia o queja en relacion con este servicio, o el contenido de este informe de calibracion favor de dirigirlo a sugerencias@masmetrologia.com

MET-06.7-1/11 Pgina 1 de 3
No. 126883

Test Results
Test Description True Value Test Result Lower limit Upper limit Units Status Uncert ()

UUT IDENTIFICATION

Serial Number: 6769041


Firmware Level: 1.7 D1.0

DC VOLTAGE TESTS

100 mV Range
0.000 mV 0.0000 0.002 -0.006 0.006 mV Pass 9.4e-007 V
90.000 mV 90.0000 90.003 89.971 90.029 mV Pass 2.2e-006 V

1000 mV Range
900.00 mV 900.000 900.01 899.71 900.29 mV Pass 1.0e-005 V

300 mV Range
0.00 mV 0.000 0.00 -0.02 0.02 mV Pass 5.7e-006 V
300.00 mV 300.000 300.00 299.90 300.10 mV Pass 7.7e-006 V

3V Range
3.0000 V 3.00000 3.0000 2.9990 3.0010 V Pass 6.2e-005 V
-3.0000 V -3.00000 -3.0000 -3.0010 -2.9990 V Pass 6.2e-005 V

30 V Range
30.000 V 30.0000 30.000 29.990 30.010 V Pass 6.3e-004 V

300 V Range
300.00 V 300.000 300.01 299.90 300.10 V Pass 7.0e-003 V

1000 V Range
1000.0 V 1000.00 1000.0 999.5 1000.5 V Pass 5.8e-002 V

DIODE TEST
0.0000 Ohm 0.00000 0.0002 -0.0008 0.0008 Pass
Open Pass

AC VOLTAGE TESTS

300 mV Range
0.00 mV 0.000 0.07 0.00 0.75 mV Pass 5.7e-006 V
15.00 mV @ 1 kHz 15.000 15.01 14.87 15.13 mV Pass 8.4e-006 V
15.00 mV @ 100 kHz 15.000 14.68 13.75 16.25 mV Pass 4.9e-005 V
300.00 mV @ 1 kHz 300.000 300.02 299.30 300.70 mV Pass 3.9e-005 V
300.00 mV @ 100 kHz 300.000 303.97 284.50 315.50 mV Pass 2.0e-004 V

3 V Range
3.0000 V @ 1 kHz 3.00000 3.0005 2.9930 3.0070 V Pass 3.9e-004 V

30 V Range
30.000 V @ 1 kHz 30.0000 30.005 29.930 30.070 V Pass 3.9e-003 V

300 V Range
300.00 V @ 1 kHz 300.000 300.07 299.30 300.70 V Pass 4.5e-002 V

750 V Range
750.0 V @ 1 kHz 750.00 750.2 747.5 752.5 V Pass 1.8e-001 V

FREQUENCY TESTS
10.0000 kHz @ 1 V 10.00000 10.0000 9.9940 10.0060 k Hz Pass 5.9e-002 Hz

RESISTANCE TESTS

300 Ohm Range

-Any questions, comments, suggestions or complaints regarding this service, or the content of this calibration report please send to sugerencias@masmetrologia.com
-Cualquier duda comentario, sugerencia o queja en relacion con este servicio, o el contenido de este informe de calibracion favor de dirigirlo a sugerencias@masmetrologia.com

MET-06.7-1/11 Pgina 2 de 3
No. 126883

Test Results
Test Description True Value Test Result Lower limit Upper limit Units Status Uncert ()
0.00 Ohm 0.000 0.00 0.00 0.04 Pass 5.7e-003 W

300.00 Ohm 300.000 299.95 299.83 300.17 Pass 9.7e-003 W

3 kOhm Range
0.0000 kOhm 0.00000 0.0002 0.0000 0.0002 k Pass 5.6e-002 W

3.0000 kOhm 3.00000 3.0001 2.9983 3.0017 k Pass 9.7e-002 W

30 kOhm Range
30.000 kOhm 30.0000 30.000 29.983 30.017 k Pass 9.7e-001 W

300 kOhm Range


300.00 kOhm 300.000 300.02 299.83 300.17 k Pass 1.0e+001 W

3 MOhm Range
3.0000 MOhm 3.00000 3.0001 2.9980 3.0020 M Pass 1.6e+002 W

30 MOhm Range
30.000 MOhm 30.0000 29.993 29.922 30.078 M Pass 7.6e+003 W

300 MOhm Range


300.0 MOhm 300.00 300.2 294.0 306.0 M Pass 7.6e+005 W

DC CURRENT TESTS

30 mA Range
30.000 mA 30.0000 30.000 29.982 30.018 mA Pass 2.5e-006 A

100 mA Range
100.00 mA 100.000 100.01 99.93 100.07 mA Pass 1.1e-005 A

10 A Range
10.000 A 10.0000 9.993 9.975 10.025 A Pass 4.2e-003 A

AC CURRENT TESTS

30 mA Range
30.000 mA @ 1 kHz 30.0000 30.002 29.840 30.160 mA Pass 1.0e-005 A

100 mA Range
100.00 mA @ 1 kHz 100.000 100.01 99.40 100.60 mA Pass 4.5e-005 A

10 A Range
10.000 A @ 1 kHz 10.0000 10.000 9.890 10.110 A Pass 9.1e-003 A

End of Certificate/Fin del Certificado

-Any questions, comments, suggestions or complaints regarding this service, or the content of this calibration report please send to sugerencias@masmetrologia.com
-Cualquier duda comentario, sugerencia o queja en relacion con este servicio, o el contenido de este informe de calibracion favor de dirigirlo a sugerencias@masmetrologia.com

MET-06.7-1/11 Pgina 3 de 3