PRECISION CALIBRATION LABORATORY

NO.: G44CPI. EKAMRA VILLA. JAYDEV VIHAR. BHUBANESWAR -751 015 Ph.. 0674 -2558250(0).0674-6523411 (Lab).
Mobile 9937058250 Fax 0674 -2558250, E-mail meher. cal'bral,on@yahoocoln
CALIBRATION CERTIFICATE
WELDING RECTIFIER

Calibration Certificate No. PCL / 514/ 13-14 Issue No : peL / BB / 543 / 13 Date of
ue: 26_04.2013
Date of Calibration CaUbration Pag No of Pages
valid u to e
24.04.2013 23.04.2014 1/1 1
1.NAME & ADDRESS OF CUSTOMER : Mis THERMAX ENGG CONSTRUCTION CO LTD. TATA STEEL PROJECT,

KALINGA NAGAR, JAJPUR.

2.CUSTOMER REF NO : NIL, Dt: 20.04.2013

3 . DATE OF RECEIVED : 24.04.2013

4 . DETAILS OF INSTRUMENT UNIT UNDER CAUBRAnON (UUC) : Welding Rectifier


Make : Ajit Singh & Sons
ID. NO :
Range : 0-500AMP/0 100V
PCL ID.NO :514/13-14

5. CONDITION OF EQUIPMENTS : OK
6. TRACEABILITY : Calibration Is traceable to National Standard thrpugh NABl Accredited laboratory.

7. DETAILS OF THE REFERENCE STANDARD: 1. . DIGITAL CLAMP METER Make : lUTRON, Model: OM6056, Sr. No : 1138471
is traceable to National Standard through NABl Accredited Laboratory, Vide Calibration Certificate no. 5010608, Ot 22.11.2011, Valid upto
21.11.2013.

8 .ENVIRONMENTAL
CONDITIONS : Temperature : 270C :t 20C,

Relative Humidity : 40 to 60% R.H.

9. PROCEDURE : According to Procedure PCL /PROC/DOC No: 3/ ET.

10. CALIBRATION RESULTS DC CURRENT


Reference Standard In Avera e Value of UUC In A Error Expanded
A Uncertainty
100 99 -1
190 189 -1 1.08%
280 268 -2
380 377 -3

DC VOLTAGE

Reference Standard Avera e Value of UUC In A Error Expanded


In A Uncertainty
40 40 -
1 0.75%
78 77 -
4
11. Uncertainty measurement at 95"10 of confidence level with coverage factor k = 2

ISO 9001 :2006 CERTIFIED CALIBRATION LABORATORY


PRECISION CALIBRATION LABORATORY
PLOT NO.: 644(P), EKAMRA VILLA, JAYDEV VIHAR, BHUBANESWAR - 751 015
Ph.: 0674 - 2558250(0). 0674-6523411 (Lab). Mobile _ 9937058250.
Fax 0674 -2558250. E-mail meher calibration@yahoo co In

ISO 9001 :2006 CERTIFIED CALIBRATION LABORATORY


10.
CALIBRATION CALIBR
CERTIFICATE ATION
PORTABLE OVEN RESULT
Calibration Certificate No. PCL / 514/ 13-14 Issue No : peL / S
3 / 13 Date of Issue:

26_04.2013
Date of Calibration CaUbration valid u to Pag e
24.04.2013 23.04.2014 1/1
1.NAME& ADDRESS OF CUSTOMER : Mis THERMAX
ENGG CONSTRUCTION CO
LTD. TATA STEEL
PROJECT,

KALINGA NAGAR,
JAJPUR.

2.CUSTOMER REF NO : NIL, Dt: 20.04.2013

3 . DATE OF RECEIVED : 24.04.2013

4 . DETAILS OF INSTRUMENT UNIT UNDER CAUBRAnON


(UUC) : Temparature
Controller with sendor
Make : ORGAN
ID. NO : TEE/PO/035
Range : 50 to 300c
PCL ID.NO :553/13-14
Resolution : 10 c

5. CONDITION OF EQUIPMENTS : OK
6. TRACEABILITY : Calibration Is traceable to National Standard
thrpugh NABl Accredited laboratory.

7. DETAILS OF THE REFERENCE STANDARD: RTD WITH


INDICATOR
MAKE: TEMPSENS, SR. NO:
22907 and Indicator SR. NO :
08030422
Is traceable
to National
Standard
NABL
Accredited
Laboratory, C-
0321, VIde
Calibration
Cert. no. Tl
/ 212 /
376.1.1
Dt. 31.01.2013, Valid unto
30.01.2015

8 .ENVIRONMENTAL
CONDITIONS : Temperature : 270C :t 20C,

Relative Humidity : 40 to 60%


R.H.

9. PROCEDURE : Comparison Method with


reference to 1.s. 2848;1984

ISO 9001 :2006 CERTIFIED CALIBRATION LABORATORY


Average Value of
Parameter/Range UUC in oC Reference Standard in Error in oC
oF

'+ 1.3
50 51.3

'+ 2.5
100 102.5
Temparature
Controller with '+ 3.2
Sensor 50 to 300 oC 150 163.2

'+ 3.5
200 203.5

'+ 4.6
280 284.6

11. Uncertainty measurement at 95"10 of confidence level with coverage factor k = 2 ; +-6.28oC

ISO 9001 :2006 CERTIFIED CALIBRATION LABORATORY


PRECISION CALIBRATION LABORATORY
PLOT NO.: 644(P), EKAMRA VILLA JAYDEV VIHAR, BHUBANESWAR - 751 015
Ph.: 0674 - 2558250(0). 0674-6523411 (Lab). Mobile 9937058250.
Fax 0674 -2558250. E-mail meher calibration@yahoo co In

CALIBRATION CERTIFICATE
HOLDING OVEN
Calibration Certificate No. pel / 553 / 13-14
Issue No : PCl / BB / 5B2 / 13 Date of Issue: 26.04.2013

Date of Calibration CaUbration valid u to Pag e No of Pages


09.03.2013 08.03.2014 1/1 1
1.NAME & ADDRESS OF CUSTOMER : Mis THERMAX ENGG CONSTRUCTION CO
LTD. TATA STEEL PROJECT,

KALINGA NAGAR, JAJPUR.

2.CUSTOMER REF NO : NIL, Dt: 06.03.2013

3 . DATE OF RECEIVED : 09.03.2013

4 . DETAILS OF INSTRUMENT UNIT UNDER CAUBRAnON (UUC) : DIGITAL TEMPERATURE CONTROLLER WITH SENSOR
MAKE : PERFACT
RANGE : 50to4000e
ID.NO TEE/HO/002
PCL ID.NO. :7680/12-13
RESOLUTION : lOC
5. CONDITION OF EQUIPMENTS : OK
6. TRACEABILITY : Calibration Is traceable to National Standard thrpugh NABl Accredited laboratory.

7. DETAILS OF THE REFERENCE STANDARD: RTD WITH INDICATOR


MAKE: TEMPSENS, SR. NO: 22907 and Indicator SR. NO : 08030422
Is traceable to National Standard NABL Accredited Laboratory, C-0321, VIde
Calibration Cert. no. Tl / 212 / 376.1.1
Dt. 10.03.2013, Valid unto 09:.03:.2015

8 .ENVIRONMENTAL
CONDITIONS : Temperature: 270C: t 20C,

Relative Humidity: 40 to 60% R.H.

9. PROCEDURE : Comparison method with reference to I.S. 2053:1974

10. CALIBRATION RESULTS

Average Value of Reference


Parameter/Range UUC in oC Error in oC
Standard in oF

50 49.0 '+ 1.0

100 99.0 '+ 1.0


Digital Temparature
Controller with Sensor 50 200 198.0 '- 2.0
to 400 oC
250 248.0 '- 2.0

350 347.0 '- 3.0

11. Uncertainty measurement at 95"10 of confidence level with coverage factor k = 2 ; +-1.43oC

ISO 9001 :2006 CERTIFIED CALIBRATION LABORATORY


PRECISION CALIBRATION LABORATORY
PLOT NO.: 644(P), EKAMRA VILLA JAYDEV VIHAR, BHUBANESWAR - 751 015
Ph.: 0674 - 2558250(0). 0674-6523411 (Lab). Mobile 9937058250.
Fax 0674 -2558250. E-mail meher calibration@yahoo co In
CALIBRATION CERTIFICATE
BAKING OVEN
Date of Calibration CaUbration valid u to Pag e No of Pages
09.03.2013 08.03.2014 1/1 1

ISO 9001 :2006 CERTIFIED CALIBRATION LABORATORY