You are on page 1of 14

Fluidics UNIT 1.

Pearlson P. Austin Suthanthiraraj1 and Steven W. Graves1


1
Department of Chemical and Nuclear Engineering, University of New Mexico,
Albuquerque, New Mexico

ABSTRACT
The use of fluidics is implicit in a technology named flow cytometry, which flows a cell or
particle through a sensing volume to obtain serial analysis of particles on a one by one basis.
This flow of particles enables flow cytometry to collect information on multiple particle popu-
lations, giving it a distinct advantage over bulk analysis approaches. Moreover, flow cytometers
can analyze thousands of particles per second in a single flowing stream. Additionally, use of
volumetric sample delivery makes it possible for flow cytometers to accurately count cells and
particles. Furthermore, the analysis results can be coupled with a fluidic diversion mechanism to
sort and collect particles based on desired properties. Finally, when high-throughput sampling
technologies are employed to rapidly change the input of the sample stream, a flow cytometer
can become an integral tool for high-throughput screening. The above properties have made
flow cytometry useful in a wide range of biomedical applications. In this unit we will present
an overview of fluidic systems that make flow cytometry possible. This will introduce historical
approaches, explanations of the commonly implemented current fluidics, and brief discussions
of potential future fluidics where appropriate. Curr. Protoc. Cytom. 65:1.2.1-1.2.14. C 2013 by
John Wiley & Sons, Inc.
Keywords: cell analysis r cell sorting r high-throughput screening r fluidic systems r
flow cytometry r FACS r microfluidics

INTRODUCTION The second function for the fluidics is to


The primary purpose of the fluidics of a position the sample stream for optimal analy-
flow cytometer is to deliver a sample contain- sis (Fig. 1.2.1). To ensure that only one particle
ing suspended cells or particles to the flow or cell is measured at a time, an optimal fluidic
cytometer. This has historically been done us- system minimizes coincident events within the
ing either a simple pump, such as a syringe analysis volume. This basic principle was first
pump, or a pressure head above the sample suggested as a method to count microscopic
(Fig. 1.2.1). Initially, flow cytometers deliv- cells (Moldavan, 1934). This system used a
ered single samples to the instrument and were pressure differential to drive cells suspended
manually operated (Moldavan, 1934). How- in a liquid medium through a glass capillary
ever, improved technologies have resulted in drawn to narrow dimensions. The capillary
a number of automated sampling approaches was placed under a microscope objective, and
that allow a flow cytometer to sample from cells that obscured the field of view could be
many samples without operator intervention automatically counted using a photodetector.
and at very high sampling rates (Edwards The fluidics of this effort was improved upon
et al., 2001). Most flow cytometers sold to- through the use of a fast moving sheath stream
day come with a microplate feeder option or to focus the sample stream in the center of the
an automatic tube delivery system. The fastest analysis capillary (Fig. 1.2.2), which allowed
sampling systems today can deliver discreet the use of wider capillaries that alleviated clog-
samples at rates as high as two per second ging and retained tight focal positioning of the
(Edwards et al., 2004). Beyond rapid deliv- sample stream. This approach also reduced co-
ery, there have been numerous inline sample incident events in the focal region of the optics,
preparation approaches to automate reagent as the narrow sample core limited the number
addition, rapidly mix samples, and automate of particles that could be present in the analysis
incubation times. These have included the use volume (Crosland-Taylor, 1953).
of conventional fluidics and the use of mi- These early instruments counted cells in
crofluidic preparation approaches. capillaries and did not use refined flow cells.
Flow Cytometry
Instrumentation

Current Protocols in Cytometry 1.2.1-1.2.14, July 2013 1.2.1


Published online July 2013 in Wiley Online Library (wileyonlinelibrary.com).
DOI: 10.1002/0471142956.cy0102s65 Supplement 65
Copyright C 2013 John Wiley & Sons, Inc.
optical
objective

positioning

pressure sorting

automated syringe pump


valve
sampling

to
waste

aqueous
buffer

Figure 1.2.1 A conceptual diagram of the three primary functions of the fluidics of a flow cy-
tometer. Many options are possible to accomplish each of these functions. Potential examples are
shown.

The first effort to collect detailed optical prop- mogeneous wash-free fluorescence assays in
erties from flowing cells used a syringe pump general (Sklar et al., 2002). When used solely
to draw cells through a 100 m 100 m flow for analysis, this approach is typically config-
channel that had been ultrasonically cut into a ured to flow upward, which ensures that any
glass slide (Kamentsky and Melamed, 1969), air bubble in the system simply floats upward
which foreshadowed the field of microfluidic out of the flow cell (Fig. 1.2.4A). Though hy-
flow cytometry. The fluidics of this instrument drodynamic focusing provides many benefits,
delivered the cells with enough precision to it also increases the use of fluids, the amount
enable the collection of absorption and scat- of waste, and the linear velocity of the par-
ter properties of cell as they passed through a ticles, which has led to many efforts to de-
tightly focused illumination spot. The eventual velop sheathless flow cytometers that retain
coupling of flow cytometry with fluorescence the benefits of hydrodynamic focusing with-
measurements (Bonner et al., 1972) gave new out the need for large volumes of sheath.
importance to the use of hydrodynamic focus- The third fluidic function is to efficiently
ing. Within the flow cell of a typical mod- sort and collect particles of interest based on
ern flow cytometer, hydrodynamic focusing the results of the particle analysis that occurs
constrains the sample stream to a very nar- just upstream of the sorting process. Sorting
row diameter. This ensures that the particle has been most commonly performed using
largely fills the optical interrogation volume, electrostatic droplet deflection. Electrostatic
which is defined by the intersection of the deflection of droplets was originally developed
sample stream with focused excitation light for use in conjunction with Coulter volume
(Fig. 1.2.3). Thus, particle-associated fluores- measurements (Fulwyler, 1965); it was subse-
cence normally dominates the optical mea- quently adapted for use with scatter and fluo-
surement, as solution-based fluorophores are rescence measurements (Bonner et al., 1972).
largely excluded from the analysis volume. In this approach the focused sample stream
This enables a flow cytometer to resolve flu- exits the flow cell or nozzle and the exiting
orophores bound to the surface of a particle jet-in-air is broken into droplets via a stand-
Fluidics versus those in solution, and allows for ho- ing wave imposed upon the flow stream by a

1.2.2
Supplement 65 Current Protocols in Cytometry
Figure 1.2.2 The principle of hydrodynamic focusing. Sheath fluid (black arrows) flows at a high
linear velocity compared to the slow moving sample stream (gray with white arrows). The fast
moving sheath accelerates and focuses the sample stream to a narrow core as it exits the flow
cell or nozzle, or the capillary tapers towards the exit orifice.

piezoelectric drive (Fig. 1.2.4B,C). The parti- variations on this theme have been developed
cles of interest are entrapped within droplets including piezo-based diversion of the flow
formed by the jet-in-air mechanism, which are stream (Gray et al., 1989), directed electroos-
charged electrostatically and, based on their motic pumping to selectively change flow di-
fluorescence and scatter properties, which rections (Fu et al., 1999), and MEMS-based
were determined at the point of analysis, de- valves (Yang et al., 2006).
flected by highly charged metal plates to be While, as the above introduction suggests,
collected into discrete sorted samples. Anal- flow cytometry is certainly an established tech-
ysis of such systems can be accomplished in nology, research into optimal fluidic systems
a flow cell, which is followed by droplet for- for flow cytometry continues. Recently, re-
mation in an exiting jet in air (Fig. 1.2.4B). search has focused on (i) improving sample
Alternatively, the analysis can be performed delivery systems to reduce system cost, retain
directly in the stream as it exits a flow nozzle volumetric delivery, and increase the sample to
(Fig. 1.2.4C). As compared to the use of a flow sample delivery rates; (ii) developing alternate
cell, this approach sacrifices optical sensitiv- focusing techniques to eliminate or minimize
ity but allows for closer analysis to the droplet the need for sheath fluid; and (iii) developing
breakoff point and more accurate sorting deci- new sorting mechanisms for safe and gentle
sions. Regardless of the approach taken, elec- handling of biological samples. The remain-
trostatic deflection can reach rates of 50,000 der of this unit discusses in detail the basics of
per second, and it has become the dominant flow cytometry fluidics and introduces recent
sorting technology used today in commercial advances with regards to sample delivery, par-
instruments. ticle focusing, and sorting. While most of these
However, droplet-based sorting does have advances have been implemented in commer-
some drawbacks such as required high pres- cial flow cytometers, some are still early stage
sures that can be damaging to fragile cells technologies that are expected to be seen in
and the generation of aerosols, which can be wider use in the coming years.
a significant biohazard. Therefore, it is also
becoming common to use actuators to selec- SAMPLE DELIVERY
tively divert the sample stream into collection Sample delivery has been accomplished us-
channels to sort particles of interest. The earli- ing either pumps or an increase of pressure dif-
est example of this used the same microchan- ferential between the sample and the system.
nel approach developed above, and sorting The use of pumps adds complexity, but of-
was accomplished via triggering a second sy- fers control over delivery rates, which enables
ringe pump to withdraw cells of interest as direct counting of particles per unit volume
they passed the intersection of a crossed chan- without the addition of a particle standard of a Flow Cytometry
nel (Kamentsky and Melamed, 1969). Many known concentration. Beyond these two basic Instrumentation

1.2.3
Current Protocols in Cytometry Supplement 65
Figure 1.2.3 Optical analysis of a focused sample stream in a typical hydrodynamic focusing flow cytometer.
The sample stream is surrounded by a clean sheath fluid that is not pictured.

approaches, sample delivery systems can be which poses challenges in delivering samples
broadly classified into the following types. at precisely known rates and prevents them
from being truly volumetric. Therefore, mi-
1. Traditional single sample methods.
crosphere standards that contain particles at
2. Rapid mixing systems that quickly pro- known concentrations must be added to the
cess for rapid single sample delivery. samples to precisely measure the particle or
cell concentrations within the sample (Stewart
3. High-throughput sample delivery for se- and Steinkamp, 1982). A further disadvantage
rial delivery of multiple samples. of the pressure differential delivery system is
that pressure head at the sample could cause
Traditional Single Sample Methods sample to be introduced into the air due to tube
failure or operator error, which is a concern for
Pressure differential many biological samples.
In this method of sample delivery, a pres-
sure differential is maintained across the reser- Syringe pumps
voirs containing the sample and sheath fluid. In In a typical syringe pumpbased delivery
practice, the sample is contained in a tube that system, an arrangement of capillary tubes and
is pressed onto the system, forming a sealed valves are used to draw the sample into a fill
pressure head above the sample (Fig. 1.2.5A). line and then pump the sample to the system
The seal has an inlet to drive an increase of (Fig. 1.2.5B). The precise action of the step-
pressure above the sample. Once the pressure per motor in a syringe pump enables sample
of the sample has increased above that of the delivery to the flow channel at precisely deter-
system, the sample is driven into the system. mined rates, which makes these systems truly
In these systems, the rate of sample delivery volumetric. Despite these advantages, syringe
depends upon the pressure gradient, as well as pump systems are problematic in that they can
the pressure drop that occurs along the length induce pulsations in flow, they deliver defined
of the capillary tubing. Once they have been sample volumes that are predetermined by fac-
carefully calibrated, such systems enable very tors such as the size of the fill line, and they
smooth sample delivery, as required for most are typically more expensive to construct.
flow cytometers. Another advantage of pres-
sure differential systems is that removal of Peristaltic pumps
the sample can reverse the pressure differen- In a peristaltic pump, the sample is drawn
tial between a sheath fluid and the delivery from an open vessel into a flexible tubing in
point. This causes sheath fluid to be driven contact with the rotor. As the rotor spins, the
through the sample line, which washes the line tubing is subject to alternate cycles of com-
and limits carryover between samples. Despite pression and relaxation. The compression cy-
these advantages, pressure differential systems cle moves the fluid within the tubing while
Fluidics are subject to pressure variations over time, the relaxation cycle draws a small volume of

1.2.4
Supplement 65 Current Protocols in Cytometry
A B C
sample
to waste inlet

optical sheath
interrogation inlet
(focused
laser) flow
nozzle
quartz
optical
flow
interrogation
cell
(focused exit orifice
laser)

exit orifice

droplet
sheath exit line breakoff
inlet distance
sample
inlet

Figure 1.2.4 Typical configurations of flow cytometer flow cells for (A) analysis, (B) electrostatic droplet
sorting after analysis in a flow cell, and (C) analysis directly in the jet-in-air system after ejection from a
nozzle.

sample into the tubing (Jaffrin and Shapiro, Peristaltic pump-based delivery systems also
1971). The combination of these cycles pumps do not require a pressure head to deliver the
the sample through the tubing and delivers the sample, which again reduces biosafety con-
sample to the flow cytometer (Fig. 1.2.5C). cerns during sample delivery. Furthermore,
Owing to its volumetric nature, the peristaltic they provide variable sample delivery rates,
pump offers measurement of cell concentra- do not require multiple draws to deliver large
tion, which further enables accurate measure- volumes of sample, and are relatively inex-
ment of fluorescence and scatter properties at pensive and simple to implement. The largest
the interrogation point. These advantages have issue regarding the use of peristaltic pumps is
enabled integrating peristaltic pumps in even that care must be taken to avoid pulsations in
low cost commercial flow cytometers (Rich flow. This can be accomplished in a variety of
and Howes, 2010). A pair of peristaltic pumps, ways, and instruments have been successfully
one each for the sheath and sample, has been constructed using these pumps.
used in push/pull configuration. One pump
pushes the sheath fluid through the flow cell Rapid Mix for Rapid Single Sample
while the other pump draws both the sheath Delivery
and sample by suction. The pressure differen- In the sample delivery systems described
tial so created enables precise focusing of the thus far, the cells are brought to the analysis
sample for optical analysis. However, as in the point only many seconds after injection. Such
case for all peristaltic systems, tubing wear at delays make them unsuitable for studying ki-
the pump head can cause sample delivery rate netics of fast reactions, such as ligand-receptor
variation over time. Therefore, care must be binding, which require subsecond resolution
taken to maintain the system and ensure ac- between mixing and analysis. These applica- Flow Cytometry
curate and reproducible sample delivery rates. tions require delivery systems that not only Instrumentation

1.2.5
Current Protocols in Cytometry Supplement 65
A B C

to cytometer

to cytometer
pressure inlet

automated
pressure valve
seal

peristaltic pump rotor

to cytometer
automated syringe pumps
sample container

sample container
sample container

drive fluid
supply

Figure 1.2.5 Traditional delivery methods. (A) Pressure differential, where a pressure head is introduced into a sealed
sample tube to drive sample to the flow cytometer. (B) Syringe pumps in combination with automatic valves that can first
fill a fill line and then drive the sample to the cytometer. (C) A simple peristaltic pump that pumps an unsealed sample to
the flow cytometer.

mix the samples rapidly but also deliver them sheath flow and expensive syringe drives, pres-
at low flow rates with reduced dead time (time sure disturbances at the nozzle affect the sta-
delay between successive cells at the interro- bility of the focused streams, i.e., the sample
gation point). Toward this aim, two types of streams tend to widen resulting in increased
systems have been developed: background signal (Seamer et al., 1999; Sklar
et al., 1998; Graves et al., 2002).
1. Syringe pump driven systems
2. Coaxial flow systems. Coaxial flow systems
In these systems, the samples are mixed via
Syringe pump driven systems a coaxial jet mixer constructed from two or
In a syringe pumpdriven system, samples more capillaries. This mixer is directly inserted
are loaded into holding lines that are then into a sheath stream, which enables the mixed
rapidly driven through a simple mixer into a sample to be immediately focused for analysis.
delay line. The mixed sample is then driven by This type of mixing allows studying kinetics
a drive syringe at a controlled rate into the flow within 60 msec of the sample injection (Scam-
cell. This process requires high-speed valves pavia et al., 1995). Discrete time points are
to switch the delivery from waste to the flow collected by moving the mixing point relative
cytometer just after mixing. This is a critical to the analysis point. As such, this approach
step for hydrodynamic systems, as the high requires discreet time points to be measured
fluidic rates that occur during mixing are in- individually, from which the complete time
compatible with laminar flow. While mixing, course can be reconstructed, unlike syringe-
the valves in the fluidic line prevent sample driven systems (Blankenstein et al., 1998).
flow into the flow cell. The mixed samples
are then delivered to the delay line where flu- High-Throughput Sample Delivery for
orescence measurements can be made within Serial Analysis of Many Samples
120 msec of the reaction (Graves et al., 2002; Traditionally, high-throughput screening
Nolan et al., 2005). This approach has the (HTS) for lead compounds is performed in
advantage of permitting continuous monitor- small volumes (l) in multi-well plates to iden-
ing of reaction kinetics for two or more reac- tify target compounds. However, the ability
Fluidics of flow cytometry to both directly analyze
tants. However, even with precisely controlled
1.2.6
Supplement 65 Current Protocols in Cytometry
cells and perform multiplex microsphere as- peristaltic pump to rapidly deliver plugs of
says has led to the use of flow cytometry in samples to the flow cytometer. In this system,
HTS (Edwards et al., 2007). As HTS requires the autosampler moves between samples as
the processing of millions of potential com- the peristaltic pump continuously pumps sam-
pounds, as such, sample-to-sample throughput ple to the flow cytometer. This results in small
is critical. plugs of sample (a few l) that are separated
by air bubbles being loaded into capillary tub-
ing that is connected to the flow cytometer.
Traditional high-throughput delivery
The plugs of sample and the air bubbles are
methods
delivered to the flow cytometer and the plugs
Traditional high-throughput sample deliv-
of sample are analyzed serially. The bubbles
ery systems have included tube loaders, which
have been shown to not significantly affect
simply feed tubes on the flow cytometer
the fluidics of the flow cell (Edwards et al.,
(Shapiro, 2003), and plate feeders that pro-
2004). This approach enables samples to be
cess micro-well plates in an automated fash-
drawn from multi-well plates, delivering sam-
ion (Edwards et al., 2007). In plate feeding
ples continuously at rates of up to 40 sam-
systems, the plates were typically moved so
ples per minute, again with minimal carry-
that the micro-wells make proper contact of
over (Edwards et al., 2009). This system has
the tip of the delivery tube with the sam-
been successfully developed and used in many
ple. The method of sample delivery, once the
HTS efforts for therapeutic compounds of
tip has made contact, is consistent with the
interest.
manual delivery systems. These systems do
ease the burden on the operator but are typi-
cally considered too slow for use in large HTS Parallel microchip
efforts. High content screening has also been per-
formed in microfluidic formats. A notable
device consisting of 384 parallel channels fab-
Plug flow cytometry
ricated into a glass plate was used to simul-
To speed sample delivery, plug flow meth-
taneously analyze samples delivered from a
ods of sample delivery have been developed
384-well micro titer plate. The samples were
(Edwards et al., 1999; UNIT 1.17). In these sys-
drawn by suction into these channels by using
tems, samples were originally drawn from sy-
a syringe pump. The channels were conical
ringes into a one of two fill lines within a
in shape and hemispherical in cross-section,
reciprocating valve. The valve could change
capable of handling as little as 1 to 3 l
position quickly and allowed simultaneously
of the sample. Such low volumetric capacity
filling of one sample and pressure-based de-
permitted continuous measurement of sample
livery of another sample into a commercial
concentration. Periodical cleaning using chlo-
flow cytometer. This resulted in the rapid in-
rine bleach prevents clogging while making
jection of small volume plugs of samples into
these channels re-useable for further assays
the flow cytometer, hence the name, plug flow
(McKenna et al., 2008, 2011). However, this
cytometry. About 5 l of the sample arrived
system required complex scanning optics to
at regular intervals at the interrogation point,
analyze the 384 channels.
allowing nearly nine plugs to be analyzed per
minute. Owing to plug flow, these systems al-
lowed precise measurement of sample concen- POSITIONING OF CELLS FOR
tration, and the use of the reciprocating valve ANALYSIS
made it possible to draw sample from unpres- The confinement of the sample into a nar-
surized samples (Edwards et al., 1999, 2001; row stream or channel for optical analysis en-
UNIT 1.17). This approach did not wash the flow hances the accuracy and sensitivity of mea-
cell or the delivery line between plugs. There- surements. Precise positioning of particles in
fore, carryover between plugs was carefully flow cytometry can be obtained using a variety
monitored and found to be only <2% between of methods that include:
samples, which is suitable for most HTS ap- 1. Simple micro channels or capillaries
plications (Edwards et al., 1999).
The concept of delivering small plugs of 2. Hydrodynamic focusing
sample was refined in the development of the
3. Acoustic focusing
HyperCyt delivery system (Edwards et al.,
2001, 2004). In this system, a computer- 4. Inertial particle migration and structured Flow Cytometry
controlled autosampler is combined with a microchannels for particle focusing Instrumentation

1.2.7
Current Protocols in Cytometry Supplement 65
No Focusing/Microcapillary Flow age plane (Howell et al., 2008). Regardless of
Cytometry the type of sample core desired and the poten-
Akin to the earliest flow cytometers, many tial disadvantages, many successful flow cy-
efforts have continued to analyze particles that tometer systems have been constructed using
are simply positioned by the physical con- capillary-based approaches for hydrodynamic
straints of a microchannel. This approach al- focusing or, more recently, the use of modern
lows simple coupling to the sample delivery microfluidics approaches.
system, but the optical interrogation volume is
usually quite large as particles can pass any-
where within the channel. Additionally, as the Capillary-based hydrodynamic focusing
dimensions used typically dictate the devel- For decades, commercial flow cytometers
opment of a parabolic flow profile, the rela- have used a nozzle and tubing arrangement to
tive linear velocity of particles varies based achieve hydrodynamic focusing of the sample.
on their position in the channel, which results The sheath fluid flows into a tubing of rela-
in reduced accuracy of optical measurements. tively large diameter than the nozzle through
Both of these issues can be minimized by the which the sample is injected. The channel
use of smaller channel cross-sections, but as tapers into a narrow region (Fig. 1.2.4). At
the channels are reduced in size, the likelihood this region, the hydrodynamic pressure of the
of clogging increases. Thus, this approach typ- sheath fluid focuses the sample into a stream of
ically must sacrifice accuracy to be usable for particle-dimension width, which, as described
any length of time. Furthermore, as the sample above, provides improved optical positioning
comes into direct contact with the optical sur- and enables the resolution of free versus bound
faces, the channels must be routinely cleaned fluorophore (Fig. 1.2.3).
or replaced for long-term use. Regardless of This conventional design for hydrodynamic
these disadvantages, these systems are sim- focusing can be modified suitably depending
ple and inexpensive to construct, and many on its application. In an extreme example, a
successful systems are based on this approach capillary tubing of 100-m o.d. was inserted
(King et al., 2010). into a 250-m square flow chamber. The sam-
ple was delivered into the capillary tube while
the sheath fluid flowed into the square cham-
Hydrodynamic Focusing ber. The sheath flow from the periphery of the
Most modern flow cytometers tightly posi- capillary tubing focused the single molecules
tion the sample for optical analysis via hydro- precisely, yielding high sensitivity of detection
dynamic focusing. Here, a carrier fluid called for DNA molecules (Keller et al., 1996).
the sheath fluid is used to position the sample Despite being able to focus cells and single
of cells into a single file for optical interroga- molecules, large particles are problematic for
tion. As both the sheath and sample are incom- this approach. As particles increase in size,
pressible liquids, the hydrodynamic pressure the flow channel cross-sections must corre-
of the fast moving sheath fluid simply accel- spondingly increase to prevent clogging. The
erates and focuses the slow moving sample flow of fluids is characterized by a dimension-
stream into a tight sample core that can be as less number called the Reynolds number (Re),
narrow as a few microns (Kachel et al., 1990). which is the ratio of inertial force to the vis-
This approach can be used to axially focus the cous force acting on the fluid. The Reynolds
core by simply inserting a narrow capillary number is a function of the linear flow velocity
tube into the center of the fast moving sheath, vf as well as the physical properties, such as
which results in a tightly focused core in two density and viscosity of the fluid and given
dimensions. This is the traditional approach by the equation;
taken by most flow cytometers and this enables
the alignment of particles for single file optical rvf D
analysis. Alternatively, if sheath fluid is deliv- Re =
ered from only the sides of the flow stream or m
at disproportionate rates from the top and bot-
tom flows, focusing occurs primarily in only Here, D refers to the diameter for a cylin-
one dimension, which can provide a ribbon- drical channel. When Re is less than 2300,
like core stream. For some applications, this flow is smooth and laminar, but at Re greater
would be problematic, but for imaging flow than 2300, flow can be turbulent. The Reynolds
cytometers, this approach can be desirable, as number equation predicts that, as the channel
Fluidics particles can be constrained to a single im- cross-section increases, turbulence will occur
1.2.8
Supplement 65 Current Protocols in Cytometry
at lower linear velocities. The very low linear Fluorescence-activated cell sorting by hy-
velocities required for laminar flow has lim- drodynamic focusing on a microscale was
ited the use of hydrodynamic focusing systems demonstrated by Wolff et al. A sample con-
at channel cross-sections wider than about sisting of beads and chicken red blood cells in
400 m. Though this is problematic for larger the ratio of 2.4*105 (depicting a rare event)
particles, which include Drosophila embryos, was focused by introducing buffer solution
tumor microspheroids, and many others, a va- through the side inlets. While the through-
riety of hydrodynamic focusing flow systems put was much higher than that of a commer-
have been designed to both analyze and sort cial cell sorter, particles were confined only in
such particles (Pulak, 2006). Of course, in all one dimension, i.e., the particles were focused
cases, the restricted linear velocities dramati- only at the sides but not at the top and bottom.
cally reduce the particle analysis rates. However, coaxial sample sheathing precisely
focused particles in two dimensions. In this
modified cell sorter, the sheath fluid moved
Microfluidic hydrodynamic focusing
the sample introduced from the bottom of the
Many groups have been exploring microflu-
microchannel just like smoke from a chim-
idic flow cytometry to facilitate integration of
ney. Hence, this focusing approach was called
flow cytometry into a laboratory-on-a-chip ap-
the smoking chimney method. Lower variation
plication (Huh et al., 2005). Hydrodynamic
in fluorescence intensity was observed with a
focusing in such systems has been developed
sample of fluorescein coaxially sheathed using
using a number of approaches. A brief review
phosphate-buffered saline (PBS) (Wolff et al.,
of the various microfabrication-based hydro-
2003). Sheath fluid injection from as many as
dynamic focusing is discussed below.
seven ports at varying heights in a sandwich
Among the various microchannel designs
type device microfabricated by soft lithogra-
for hydrodynamic focusing, the simplest ones
phy techniques precisely focused particles in
consisted of two inlets for sheath flow. These
both vertical and lateral directions (Sundarara-
inlets were located on either side of the sam-
jan et al., 2004).
ple inlet and were oriented either perpendicu-
lar or at an angle. By using high aspect ratio
(height/width) channels, the sample was fo- Acoustic Focusing
cused into a very narrow stream. Moreover, the Acoustic focusing uses ultrasound to form
parabolic flow profile in these channels pro- an acoustic standing wave inside the flow
vided some degree of focusing of the particles channel, which has been detailed for a cylin-
in the vertical direction (Lee et al., 2006). In an drical focusing geometry in UNIT 1.22. In the
attempt to use gas instead of liquid as sheath, presence of an acoustic field, the acoustic ra-
Huh et al. pumped air into a microchannel with diation force acts on the particles and moves
hydrophobic walls. The sample was well fo- them either to the pressure node (pressure min-
cused at low flow rates, but at higher flow rates, imum) or pressure antinode (pressure maxi-
pressure disturbances in flow resulted in drop mum) depending on the sign of the acoustic
formation (Huh et al., 2002). contrast factor, which is a function of the den-
Simmonet et al. designed a device from a sity and compressibility of the particle and
single mold of polydimethylsiloxane that was the surrounding medium. Particles that have
created with three inlet ports for the sheath a positive value of the contrast factor (most
fluid to hydrodynamically focus the particles. cells and particles) migrate to the node while
The sample was introduced from the bottom. the negative contrast particles migrate to the
Varying hydrostatic pressure of the sheath fluid antinode of the standing wave (Laurell et al.,
from each inlet resulted in particle diffusion to 2007; Cushing et al., 2013). The acoustic radi-
the geometric center (with respect to width and ation force is also directly proportional to the
height) of the channel (Simonnet and Grois- volume of the particle, which confers a size-
man, 2005). Precision of focusing was further dependence to this approach. Nonetheless, this
improved with introduction of sheath at dif- approach has been shown to be effective for
ferent depths, which enabled high-resolution particles as small as bacteria (Hammarstrom
images to be obtained using a charge-coupled et al., 2012).
device (CCD). Three-level introduction of As described in UNIT 1.22, this approach
sheath not only focused the particles precisely, was first developed for flow cytometry using a
but also increased throughput and precision, cylindrical geometry (Goddard and Kaduchak,
comparable to a commercial flow cytometer. 2005; Goddard et al., 2006). In this form, Flow Cytometry
(Simonnet and Groisman, 2006). the cylinder forms an axially positioned node, Instrumentation

1.2.9
Current Protocols in Cytometry Supplement 65
which makes it very analogous to the central force acts perpendicular to it and causes the
positioning provided by hydrodynamic focus- particle to migrate in the lateral direction. The
ing. However, it does not accelerate particles inertial lift force is composed of two com-
and it also provides a measure of inline par- ponents namely the wall lift and the shear-
ticle concentration as the particles are simply induced lift. The wall lift moves the particle
moved to the center of the channel. Both of away from the wall while the shear-induced
these effects make it possible to analyze sam- lift acts down the shear gradient. The super-
ples at slower particle velocities, but with sim- position of these two components gives rise
ilar overall particle analysis rates. to lateral migration of the particle to a shear
Acoustic focusing has also been success- region of high inertial lift, which is away from
fully adopted in a planar mode where the the channel centerline and closer to the wall
acoustic theory is perhaps simpler (Austin but a little away from it. Such equilibrium po-
Suthanthiraraj et al., 2012; Piyasena et al., sitioning of particles depends on the geometry
2012). In this single dimensional example, of the channel and the size of the particle. Fur-
a standing wave can be established in a mi- thermore, such migration occurs over a spe-
crochannel whose width equals half the wave- cific time scale and hence over a characteris-
length of the applied ultrasound. Using this tic length (Di Carlo, 2009). As the magnitude
relation, either the flow channel can be fab- of lift force depends on particle dimension,
ricated to match the driving frequency or a it is possible to separate particles of different
suitable driving frequency be chosen to match sizes to different lateral positions within mi-
the width of the flow channel. Thus, on appli- crochannels, thus achieving an effective size-
cation of the acoustic field, a pressure node is based separation.
formed at the center and two pressure antin- With regard to particle positioning, the most
odes at the walls of the flow cell (Wiklund important aspect to consider is how the parti-
et al., 2004). The use of higher harmonics cles reach a predicted position. While this is
of standing wave across the channel also en- complex in nature, some basic rules are known.
ables the development of multiple streams for In a rectangular channel, the inertial lift forces
parallel flow cytometry (Nilsson et al., 2004; result in particle migration into four equilib-
Austin Suthanthiraraj et al., 2012; Piyasena rium positions near the center of each wall
et al., 2012). face, of which only two are stable. In cylindri-
As acoustic focusing is a sheathless ap- cal channels, the particles have been observed
proach, it does minimize waste and also re- to move to an annular region, 0.6 times the
duce particle acceleration, thereby increasing radius R of the channel. For a square chan-
particle transit times at the interrogation point. nel, there are eight equilibrium positions, four
Moreover, the elimination of sheath reduces along the corners and four at the center of the
complexity of fluidics and improves the porta- faces of each wall. While multiple positions
bility and cost of a commercial flow cytometer. are potentially valuable, the use of additional
However, direct contact of the sample with the microfluidic structures can result in unique and
walls of the flow chamber makes them suscep- consistent spacing of particles for flow cytom-
tible to fouling. Hence, periodical cleaning or etry analysis (see below).
the use of a small amount of sheath is required; Another important aspect of inertial
otherwise, fouled surfaces may interfere with focusing is that it can induce particle order-
optical detection. ing. Under certain conditions, the particles ar-
range themselves in trains with uniform inter-
Inertial Particle Migration and particle spacing (Hur et al., 2010). This has
Structured Microchannels for been shown to eliminate particle coincidences
Particle Focusing in flow cytometry (Oakey et al., 2010). Uni-
The principle of inertial focusing in mi- form ordering of particles observed in inertial
crochannels has been well studied and re- focusing has also been used to trap beads and
viewed by the Di Carlo group (Di Carlo et al., cells within picoliter drops (Edd et al., 2008).
2007, 2008; Di Carlo, 2009). In inertial focus- Such precise ordering may eventually help in
ing, particles migrate to precise equilibrium overcoming drawbacks with stochastic parti-
positions based on the principles of fluid dy- cle delivery commonly observed in droplet
namics. The particles in flow experience two sorting.
types of forces namely the inertial lift force Inertial focusing in curving microchannels
and the drag force. The drag force accelerates is assisted by Dean flow, where mismatch
the particle in the direction of flow, and the lift of inertia between the fluid elements at the
Fluidics

1.2.10
Supplement 65 Current Protocols in Cytometry
channel center and the wall results in a vortex biosafety when performing fluidic sorting, as
at the curving edges (Di Carlo, 2009). Careful the generation of aerosols and other opera-
design of such inertial microfluidic channels tor exposure paths can occur (Schmid et al.,
have stabilized focusing to a single equilib- 2007). While there are many methods of sort-
rium position with efficiencies comparable to ing, we will only discuss those that are based
that of a low-end commercial flow cytometer on manipulation of the fluid stream, which ex-
(Di Carlo et al., 2008; Oakey et al., 2010). cludes optical techniques such as laser abla-
Both parallelization of microfluidic chan- tion or laser trapping (Shen and Price, 2001;
nels (Hur et al., 2010), as well as the use of ad- Applegate et al., 2004, 2009). Even so, there
vanced imaging techniques (Goda et al., 2012), are many variations on fluidic sorting, which
have enabled achieving high analysis rates re- can be broken down into the following subcat-
quired for the detection of circulating tumor egories:
cells.
1. Jet-in-air sorting
Despite not using any external field, inertial
focusing coupled with Dean flow precisely po- 2. Pneumatic sorting
sitions the particles in two dimensions in flow
channels of different geometric symmetries. 3. Mechanical sorting
Moreover, uniform interparticle spacing ob- 4. Piezo diversion sorting
served in inertially focused streams avoids par-
ticle coincidences at the interrogation region. 5. Microfluidic mechanisms
However, such spacing may not be observed in
low concentrated samples (Oakey et al., 2010). Jet-in-Air Sorting
It also requires cleaning or disposal of the op- This topic has been described in detail in
tics surfaces, and the dependence of inertial UNIT 1.7, but it will be summarized here. In
lift and secondary forces on particle dimen- this type of cell sorting, the fluid is ejected
sions may not only position particles based on through a nozzle into air. Just prior to exiting
their size, but also limit the range of particle the flow cell or just after exiting a flow noz-
dimensions that can be handled in a specific zle (Fig. 1.2.4), the optical properties of indi-
device. vidual cells are read using a laser excitation
In addition to Dean flow, structure-induced source and optical collection system. Droplet
flow effects using several pairs of grooves generation is made more uniform by vibrating
and chevrons, with or without sheath fluid, the flow cell using a piezo ceramic transducer
have also been used to position particles to coupled externally. The distance between the
the channel center for flow cytometry appli- interrogation point and the droplet break-off
cations (Howell et al., 2008). While for the point is called the droplet delay. It is deter-
groove design the degree of sheathing de- mined by the dimensions of the orifice, as well
pends on the ratio of flow rates of the sheath as by the flow and physical properties of the
and sample, the chevrons provide complete sheath fluid, and is critical to the success of
sheathing at all flow rates. Furthermore, ver- a flow sorter. Once the droplets are formed,
tical compression of the sample stream ob- a charge is placed on each droplet depending
served with chevrons find use in image an- on the optical parameters of the cell within it.
alyzers. Flow induced microvortices, without These charged droplets, on entering a static
using the sheath fluid, have also been demon- electric field of kV voltage maintained be-
strated in microfluidic channels patterned with tween two plates, are attracted towards oppo-
herringbone structures. In these channels, the sitely charged plates and deflected towards col-
balance between fluid drag, gravity and buoy- lection tubes or multi-well plates. The strength
ancy forces determines equilibrium position- of the applied electric field depends on the
ing of particles (Hsu et al., 2008). number of populations of interest, and hence
the magnitude of charge on individual drops
SORTING and the distance through which they are de-
After positioning and optical analysis, flected also vary accordingly. Though the cells
those particles having desired optical parame- are entrained within drops even at high linear
ters can be sorted (den Engh, 2000). The suc- velocities, stochastic cell arrival may result in
cess of a flow cytometer is partly determined either no cell or more than one cell within
by its ability to effectively sort cells of interest droplets, and hence not all cells of interest
at high rates and within the safety envelope are sorted. In the event of cell coincidence
required for a specific application. It is very at the interrogation point, the droplets are not Flow Cytometry
important to emphasize the need to consider charged and hence collected as waste, due to Instrumentation

1.2.11
Current Protocols in Cytometry Supplement 65
the delay in processing such events by the de- unit we will just mention two notable efforts
tector. These are termed as hardware aborts. that have garnered interest.
Similarly, when two cells travel too close to Particles have also been sorted in microflu-
be entrained in separate drops, they may be idic channels fabricated using soft lithogra-
charged as unwanted events. This is called phy. Owing to small sample volumes, these
a software abort. However, purity of sorting channels offer the ability to handle very small
can be compromised by adjusting the sort- samples and are also less hazardous than
ing logic suitably. Besides these drawbacks, droplet approaches, as samples are handled
droplet generation in air renders this type of in enclosed structures. Sorting in microfab-
sorting unsuitable for handling biohazardous ricated structures has been demonstrated us-
substances (Davies, 2007). ing a T-shaped device consisting of one inlet
well and two outlet wells. Cells were sorted
Pneumatic Sorting by electro-osmosis using three platinum elec-
The pneumatic sorting mechanism is with trodes placed below the inlet and outlet wells.
large flow cells that are capable of handling The fluorescence emission from the labeled
cells up to 1500 m in size. In this type of cells was directed by the microscope into a
sorting mechanism, while the cells of interest photomultiplier tube. Those cells emitting flu-
are allowed to flow undisturbed into a collec- orescence higher than the threshold limit were
tion receptacle or a well plate, the unwanted selectively diverted to the collection outlet by
events are gently pushed by a puff of air from a an automated increase in voltage, while the
solenoid valve located below the flow cell into remaining cells were allowed to flow into the
the waste collection chamber. Besides cell and waste outlet well. Using this technique, cells
tissues, even live organisms have been non- can be sorted both forward and backward. For
destructively handled in this fashion (Pulak, example, on identification of a cell of inter-
2006). est, the flow may be stopped temporarily to
prevent that cell from entering the waste out-
Mechanical Sorting let and the flow may be reversed to collect
This is another type of commonly used that cell. Despite effective sorting, throughput
closed sorting mechanism. If the cell cross- is limited in electro-osmotic flows. However,
ing the analysis point is identified as an event pressure-switching mechanisms may also be
of interest, it is collected into a catcher tube used, which not only increases throughput but
that moves into and out of the flow cell at also avoids high voltages required in electro-
rates up to 300 times per second; otherwise, osmosis. Throughput can also be increased
it is directed into the waste tank. Despite be- through the use of parallel microchannels (Fu
ing safe for handling biohazardous substances, et al., 1999).
sorting rates are low and the sorter is capable Cells have also been sorted using a micro-
of handling only a single population of interest fabricated polymer-based cell sorting device.
(Davies, 2007). This device consists of serpentine-shaped inlet
channels for sheath and sample fluids. These
channels provide the necessary hydrodynamic
Piezo Diversion Sorting
In this type of cell sorting, upon identifica- force for focusing the sample. A single electro-
tion of a cell of interest, the flow cell is ac- magnetic valve was used to adjust the pump-
tuated by a diamond-shaped piezo that drives ing rate of the pneumatic micropumps. The
acoustic pressure into it and deflects that cell width of the focused stream may be adjusted
along with some sheath fluid into the collec- by tuning the control frequency of these mi-
tion channel. The unwanted cells flow into the cropumps. The labeled cells were excited by a
waste chamber. Besides being a closed type laser diode and the fluorescence signals were
of sorting technique that is safe for biohaz- read using avalanche photodiode and a CCD
ardous substances, its gentle cell handling has camera. Cells of interest were sorted based on
allowed sorting of fragile but large cells such fluorescence signals using three microvalves
as the islets of Langerhans (Gray et al., 1989). downstream of the detection region to direct
those cells to the corresponding outlets (Yang
et al., 2006).
Microfluidic Sorting Approaches
Due to the breadth of the field of microflu-
idics, there has been an enormous amount LITERATURE CITED
Applegate, R. Jr., Squier, J., Vestad, T., Oakey,
of work towards cellular manipulation (Huh J., and Marr, D. 2004. Optical trapping, ma-
Fluidics et al., 2005; Patel and Perroud, 2010). In this nipulation, and sorting of cells and colloids in

1.2.12
Supplement 65 Current Protocols in Cytometry
microfluidic systems with diode laser bars. Op- E.R., Graves, S.W., and Sklar, L.A. 2007. High-
tics Express 12:4390-4398. throughput flow cytometry for drug discovery.
Applegate, R.W.Jr., Marr, D.W.M., Squier, J., and Expert Opin. Drug Discov. 2:685-696.
Graves, S.W. 2009. Particle size limits when us- Edwards, B.S., Young, S.M., Ivnitsky-Steele, I., Ye,
ing optical trapping and deflection of particles R., Prossnitz, E.R., and Sklar, L.A. 2009. High-
for sorting using diode laser bars. Optics Express content screening: Flow cytometry analysis. In
17:16731-16738. Methods Molecular Biology, vol. 486. pp. 151-
Austin Suthanthiraraj, P.P., Piyasena, M.E., Woods, 165. Springer.
T.A., Naivar, M.A., Lopez, G.P., and Graves, Fu, A.Y., Spence, C., Scherer, A., Arnold,
S.W. 2012. One-dimensional acoustic standing F.H., and Quake, S.R. 1999. A microfabri-
waves in rectangular channels for flow cytome- cated fluorescence-activated cell sorter. Nature
try. Methods 57:259-271. Biotechnol. 17:1109-1111.
Blankenstein, G., Scampavia, L., Ruzicka, J., and Fulwyler, M.J. 1965. Electronic separation of bio-
Christian, G. 1998. Coaxial flow mixer for real- logical cells by volume. Science 150:910.
time monitoring of cellular responses in flow Goda, K., Ayazi, A., Gossett, D.R., Sadasivam,
cytometry. Cytometry 25:200-204. J., Lonappan, C.K., Sollier, E., Fard, A.M.,
Bonner, W., Hulett, H., Sweet, R., and Herzenberg, Hur, S.C., Adam, J., and Murray, C. 2012.
L. 1972. Fluorescence activated cell sorting. Rev. High-throughput single-microparticle imaging
Sci. Instrum. 43:404-409. flow analyzer. Proc. Natl. Acad. Sci. U.S.A.
109:11630-11635.
Crosland-Taylor, P. 1953. A device for counting
small particles suspended in a fluid through a Goddard, G. and Kaduchak, G. 2005. Ultrasonic
tube. Nature 171:37-38. particle concentration in a line-driven cylindri-
cal tube. J. Acoust. Soc. Am. 117:3440-3447.
Cushing, K.W., Piyasena, M.E., Carroll, N.J., Maes-
tas, G.C., Lopez, B.A., Edwards, B.S., Graves, Goddard, G., Martin, J.C., Graves, S.W., and
S.W., and Lopez, G.P. 2013. Elastomeric nega- Kaduchak, G. 2006. Ultrasonic particle-
tive acoustic contrast particles for affinity cap- concentration for sheathless focusing of parti-
ture assays. Anal. Chem. 85:2208-2215. cles for analysis in a flow cytometer. Cytometry
A 69:66-74.
Davies, D. 2007. Cell sorting by flow cytometry. In
Flow Cytometry pp. 257-276. Springer. Graves, S.W., Nolan, J.P., Jett, J.H., Martin, J.C.,
and Sklar, L.A. 2002. Nozzle design parameters
den Engh, G. 2000. High-speed cell sorting. In and their effects on rapid sample delivery in flow
Emerging Tools for Single-Cell Analysis: Ad- cytometry. Cytometry 47:127-137.
vances in Optical Measurement Technologies,
vol. 2. p. 21. Wiley-Liss, New York. Di Carlo, D. Gray, D., Gohde, W., Carter, N., Heiden, T.,
2009. Inertial microfluidics. Lab Chip 9:3038- and Morris, P. 1989. Separation of pancreatic
3046. islets by fluorescence-activated sorting. Dia-
betes 38:133.
Di Carlo, D., Irimia, D., Tompkins, R.G., and Toner,
M. 2007. Continuous inertial focusing, order- Hammarstrom, B., Laurell, T., and Nilsson, J. 2012.
ing, and separation of particles in microchan- Seed particle enabled acoustic trapping of bac-
nels. Proc. Natl. Acad. Sci. U.S.A. 104:18892- teria and nanoparticles in continuous flow sys-
18897. tems. Lab Chip 12:4296-4304.
Di Carlo, D., Jon, F., Irimia, D., Tompkins, R.G., Howell, P.B. Jr., Golden, J.P., Hilliard, L.R., Erick-
and Toner, M. 2008. Equilibrium separation and son, J.S., Mott, D.R., and Ligler, F.S. 2008. Two
filtration of particles using differential inertial simple and rugged designs for creating microflu-
focusing. Anal. Chem. 80:2204-2211. idic sheath flow. Lab Chip 8:1097-1103.

Edd, J.F., Di Carlo, D., Humphry, K.J., Koster, S., Hsu, C.H., Di Carlo, D., Chen, C., Irimia, D., and
Irimia, D., Weitz, D.A., and Toner, M. 2008. Toner, M. 2008. Microvortex for focusing, guid-
Controlled encapsulation of single-cells into ing and sorting of particles. Lab Chip 8:2128-
monodisperse picolitre drops. Lab Chip 8:1262- 2134.
1264. Huh, D., Tung, Y.C., Wei, H.H., Grotberg, J.B.,
Edwards, B.S., Kuckuck, F., and Sklar, L.A. 1999. Skerlos, S.J., Kurabayashi, K., and Takayama,
Plug flow cytometry: An automated coupling de- S. 2002. Use of air-liquid two-phase flow in hy-
vice for rapid sequential flow cytometric sample drophobic microfluidic channels for disposable
analysis. Cytometry 37:156-159. flow cytometers. Biomed. Microdevices 4:141-
149.
Edwards, B.S., Kuckuck, F.W., Prossnitz, E.R.,
Ransom, J.T., and Sklar, L.A. 2001. HTPS flow Huh, D., Gu, W., Kamotani, Y., Grotberg, J.B., and
cytometry: A novel platform for automated high Takayama, S. 2005. Microfluidics for flow cy-
throughput drug discovery and characterization. tometric analysis of cells and particles. Physiol.
J. Biomol. Screen. 6:83-90. Meas. 26:R73.

Edwards, B.S., Oprea, T., Prossnitz, E.R., and Sklar, Hur, S.C., Tse, H.T.K., and Di Carlo, D. 2010.
L.A. 2004. Flow cytometry for high-throughput, Sheathless inertial cell ordering for extreme
high-content screening. Curr. Opin. Chem. Biol. throughput flow cytometry. Lab Chip 10:274-
8:392-398. 280.
Flow Cytometry
Edwards, B.S., Young, S.M., Saunders, M.J., Jaffrin, M. and Shapiro, A. 1971. Peristaltic pump- Instrumentation
Bologa, C., Oprea, T.I., Ye, R.D., Prossnitz, ing. Annu. Rev. Fluid Mech. 3:13-37.
1.2.13
Current Protocols in Cytometry Supplement 65
Kachel, V., Fellner-Feldegg, H., and Menke, E. croflow Cytometer. p. 53. Pan Stanford Publish-
1990. Hydrodynamic properties of flow cytom- ing, Stanford, Calif.
etry instruments. Flow Cytom. Sorting 27-44. Scampavia, L., Blankenstein, G., Ruzicka, J., and
Kamentsky, L. and Melamed, M. 1969. In- Christian, G. 1995. A coaxial jet mixer for rapid
strumentation for automated examinations of kinetic analysis in flow injection and flow injec-
cellular specimens. Proc. IEEE 57:2007- tion cytometry. Anal. Chem. 67:2743-2749.
2016. Schmid, I., Lambert, C., Ambrozak, D., Marti, G.E.,
Keller, R.A., Ambrose, W.P., Goodwin, P.M., Moss, D.M., and Perfetto, S.P. 2007. Interna-
Jett, J.H., Martin, J.C., and Wu, M. 1996. tional Society for Analytical Cytology biosafety
Single-molecule fluorescence analysis in solu- standard for sorting of unfixed cells. Cytometry
tion. Appl. Spectrosc. 50:12A-32A. A 71:414-437.
King, D., Cappione, A., Ilkov, F., Goldman, B., Seamer, L.C., Kuckuck, F., and Sklar, L.A. 1999.
Lefebvre, R., Pittaro, R., and Dixon, G. 2010. Sheath fluid control to permit stable flow in rapid
Progress in capillary flow cytometry. In The mix flow cytometry. Cytometry 35:75-79.
Microflow Cytometer. p. 69. Pan Stanford Shapiro, H.M. 2003. Practical Flow Cytometry.
Publishing, Stanford, Calif. Wiley-Liss, Hoboken, New Jersey.
Laurell, T., Petersson, F., and Nilsson, A. 2007. Shen, F. and Price, J.H. 2001. Laser ablation
Chip integrated strategies for acoustic separation cell sorting in scanning cytometry. BiOS 2001:
and manipulation of cells and particles. Chem. The International Symposium on Biomedical
Soc. Rev. 36:492-506. Optics.
Lee, G.B., Chang, C.C., Huang, S.B., and Yang, Simonnet, C. and Groisman, A. 2005. Two-
R.J. 2006. The hydrodynamic focusing effect dimensional hydrodynamic focusing in a sim-
inside rectangular microchannels. J. Micromech. ple microfluidic device. Appl. Phys. Lett. 87,
Microeng. 16:1024. 114104.
McKenna, B.K., Selim, A., Bringhurst, F.R., and Simonnet, C. and Groisman, A. 2006. High-
Ehrlich, D.J. 2008. 384-Channel parallel mi- throughput and high-resolution flow cytometry
crofluidic cytometer for rare-cell screening. Lab in molded microfluidic devices. Anal. Chem.
Chip 9:305-310. 78:5653-5663.
McKenna, B.K., Evans, J.G., Cheung, M.C., and Sklar, L.A., Seamer, L.C., Kuckuck, F., Posner, R.,
Ehrlich, D.J. 2011. A parallel microfluidic Prossnitz, E., Edwards, B., and Nolan, J.P. 1998.
flow cytometer for high-content screening. Nat. Sample handling for kinetics and molecular as-
Methods 8:401-403. sembly in flow cytometry. SPIE Proc. 3256:144-
Moldavan, A. 1934. Photo-electric technique for the 153.
counting of microscopical cells. Science 80:188- Sklar, L.A., Edwards, B.S., Graves, S.W., Nolan,
189. J.P., and Prossnitz, E.R. 2002. Flow cyto-
Nilsson, A., Petersson, F., Jonsson, H., and Laurell, metric analysis of ligand-receptor interactions
T. 2004. Acoustic control of suspended particles and molecular assemblies. Annu. Rev. Biophys.
in micro fluidic chips. Lab Chip 4:131-135. Biomol. Struct. 31:97-119.
Nolan, J.P., Posner, R.G., Martin, J.C., Habbersett, Stewart, C.C. and Steinkamp, J.A. 1982. Quantita-
R., and Sklar, L.A. 2005. A rapid mix flow cy- tion of cell concentration using the flow cytome-
tometer with subsecond kinetic resolution. Cy- ter. Cytometry 2:238-243.
tometry 21:223-229. Sundararajan, N., Pio, M.S., Lee, L.P., and Berlin,
Oakey, J., Applegate, R.W.Jr., Arellano, E., Carlo, A.A. 2004. Three-dimensional hydrodynamic
D.D., Graves, S.W., and Toner, M. 2010. Parti- focusing in polydimethylsiloxane (PDMS)
cle focusing in staged inertial microfluidic de- microchannels. J. Microelectromechan. Syst.
vices for flow cytometry. Anal. Chem. 82:3862- 13:559-567.
3867. Wiklund, M., Toivonen, J., Tirri, M., Hanninen, P.,
Patel, K.D. and Perroud, T.D. 2010. Miniaturized and Hertz, H.M. 2004. Ultrasonic enrichment of
sorters: Optical micro fluorescence activated cell microspheres for ultrasensitive biomedical anal-
sorter. In The Microflow Cytometer. pp. 221- ysis in confocal laser-scanning fluorescence de-
242. Pan Stanford Publishing. tection. J. Appl. Phys. 96:1242-1248.
Piyasena, M.E., Austin Suthanthiraraj, P.P., Ap- Wolff, A., Perch-Nielsen, I.R., Larsen, U., Friis, P.,
plegate, R.W.Jr., Goumas, A.M., Woods, T.A., Goranovic, G., Poulsen, C.R., Kutter, J.P., and
Lopez, G.P., and Graves, S.W. 2012. Multinode Telleman, P. 2003. Integrating advanced func-
acoustic focusing for parallel flow cytometry. tionality in a microfabricated high-throughput
Anal. Chem. 84:1831-1839. fluorescent-activated cell sorter. Lab Chip 3:22-
Pulak, R. 2006. Techniques for analysis, sort- 27.
ing, and dispensing of C. elegans on the Yang, S.Y., Hsiung, S.K., Hung, Y.C., Chang, C.M.,
COPASTM flow-sorting system. Methods Mol. Liao, T.L., and Lee, G.B. 2006. A cell count-
Biol. 351:275-286. ing/sorting system incorporated with a micro-
Rich, C. and Howes, G. 2010. The Accuri C6 fabricated flow cytometer chip. Meas. Sci. Tech-
flow cytometerA small revolution. In The Mi- nol. 17:2001.
Fluidics

1.2.14
Supplement 65 Current Protocols in Cytometry

You might also like