OPERATING INSTRUCTIONS
01625 US
Legal Notice
Distribution, reproduction and translation of the software and its documentation (or excerpts thereof) are
prohibited without the prior written consent of PHYSIOMED ELEKTROMEDIZIN AG.
PHYSIOMED ELEKTROMEDIZIN AG reserves the right to change the software and associated data as well
as the documentation without notice. All other rights reserved.
HIVAMAT is a registered trademark of PHYSIOMED ELEKTROMEDIZIN AG. The therapy mode is based on
patented technology.
HIVAMAT 200 Evident is made in Germany in compliance with the quality requirements of DIN ISO
13485:2003, FDA regulations, and the applicable safety standards:
IEC 60601-1: Electronic medical devices, Part 1: General safety requirements
IEC 60601-1-4 Electronic Medical Devices - Part 1-4: General Safety Requirements - Accompanying
Standard: Programmable Electronic Medical Devices
This product contains a Linux kernel (based on Linux kernel 2.6.15), Busybox version 1.1.3, MPlayer, all of which are distributed un-
der the GPL V2.
This product contains glibc, SDL, SDL_ttf, termcap, all of which are distributed under the LGPL.
This product contains e2fsprogs, usbutils, mtd, all of which are distributed under the GPL.
This product contains libgcc_s.so and libstdc libraries which are distributed under the GPL with run-time exception.
Portions of this software are copyright 2008 The FreeType Project (www.freetype.org). All rights reserved.
The mass-storage-card included in the device includes source codes and the individual third-party-software-licences.
Table of Contents PHYSIOMED ELEKTROMEDIZIN AG
Table of Contents
Chapter 2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.3 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.4 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Chapter 3 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.1 Definition of Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.2 Instrument Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.3 General Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.4 Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.5 Instrument Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Federal law restricts this instrument to sale by or on the order of a medical practitioner li-
censed by the law of the State in which he/she practices to use or order the use of the instru-
ment.
The instrument must only be operated by medical practicioners who have undergone special
training. Use by unqualified personnel or consumers bears the risk of serious injury.
You must operate the instrument properly, i.e. in accordance with the Operating Instructions.
If the patient experiences any unpleasant sensation during treatment, the practitioner must
stop the treatment immediately. Prior to treatment, the practitioner has to explicitely inform
the patient about this.
CAUTION
United States law restricts this instrument to sale by or on the order of a licensed healthcare
practitioner.
When making any swivel or rotary movements of the screen, make sure that no body parts
(especially fingers) or accessories are pinched.
The instrument is to be used exclusively with original accessories. Otherwise, treatment results
might be insufficient.
The instrument is to be used exclusively with original accessories. Otherwise, treatment results
might be insufficient.
Adjust the intensity before each treatment and observe the contraindications (refer to Con-
traindications on page 2).
Chapter 2 Introduction
WARNING
The instrument must only be operated by medical practicioners who have undergone special
training. Use by unqualified personnel or consumers bears the risk of serious injury.
The function of HIVAMAT 200 Evident is based on a pulsed electrostatic field which is built up in the pa-
tient's body region to be treated. Its frequency changes between 5 - 250 Hz, according to the selected
program. Due to the movement of the hand applicator, a vibrating or pumping effect with deep impact is
induced in the patient's tissue. The treatment is carried out with the supplied hand applicators.
Treatment with HIVAMAT 200 Evident has the following positive effects on the treated tissue:
muscle relaxation
relief of muscle aches and pains
increased local blood circulation
2.2 Contraindications
Therapy with HIVAMAT 200 Evident is not indicated in the following cases:
acute infections
acute inflammations with participation of pathogenic agents
active tuberculosis
acute venous diseases (untreated thromboses)
untreated malignant processes
erysipelas
patients and therapists with cardiac pacemakers and other electronic implants
heart disorders and diseases, especially cardiac insufficiency, decompensated cardiac edema, and car-
diac arrhythmia
pregnancy
hypersensitivity to electrostatic fields
infectious skin diseases
2.3 Warnings
(a) Pulsed electrostatic fields should not be applied over swollen, infected, or inflamed areas or skin erup-
tions, e.g., phlebitis, thrombophlebitis, varicose veins, etc.
(b) Pulsed electrostatic fields should not be applied over, or in proximity to, cancerous lesions.
(c) Set the line voltage at the mains module to match the incoming line voltage.
(d) HIVAMAT 200 Evident must not be used in the operating room, or in the presence of flammable an-
esthetic gases mixed with air, oxygen or nitrous oxide.
Please also note the contraindications to pulsed static electromagnetic fields in Contraindications on page
2.
2.4 Precautions
(a) Caution should be used for patients with suspected or diagnosed heart problems.
(b) Caution should be used for patients with suspected or diagnosed epilepsy.
(c) Caution should be used:
When there is a tendency to hemorrhage following acute trauma or fracture;
following recent surgical procedures when vibration may disrupt the healing process;
over the menstruating uterus; and
over areas of the skin which lack normal sensation.
(d) Instrument settings should be based on the guidance of the prescribing practitioner.
(e) Massage-intensifying systems should be kept out of the reach of children.
(f) Massage-intensifying systems should be used only with the accessories recommended by the manufac-
turer.
(g) Massage-intensifying systems should not be used while driving, operating machinery, or during any
activity in which vibration may put the user at undue risk of injury.
(h) Massage-intensifying systems emit electromagnetic waves: Maintain sufficient separation from other
electronic devices, and at least one-meter separation from any short wave or microwave therapy unit.
(i) HIVAMAT 200 Evident should only be used under medical supervision.
Chapter 3 Introduction
With HIVAMAT 200 Evident you have acquired an extremely versatile deep oscillation system. The instru-
ment will only show its true potential, however, if you are well informed about its functions. For this reas-
on, carefully read these Operating Instructions and familiarize yourself with the use of the instrument.
Paragraphs that deserve special attention are highlighted in the following way:
Type Meaning
DANGER Alerts of extremely dangerous situations that might lead to death or serious injury.
WARNING Warns of other situations that might also lead to death or serious injury.
CAUTION For risks that might lead to less serious injury.
NOTICE Only for dangerous situations that might lead to damage to the product or other assets.
Symbol Meaning
CAUTION! Refer to the Operating Instructions and consider the physiological effects
of the parameters.
CAUTION
The instrument may only be operated in dry rooms designed for medical purposes. It complies with the
technical specifications of IEC 60601-1.
It is not intended for operation in explosion hazard zones or hydrotherapy rooms. Drastic temperature
changes should be avoided, since condensation could be caused within the instrument. Do not start up
the instrument until it is in thermal equilibrium with its environment.
The instrument is to be operated properly, i.e. in accordance with the Operating Instructions.
Legend
1 Screen 2 Data Selector
3 Intensity Control 4 Card Reader
5 Connecting Socket 6 Power Module
7 Control Light
You have a total of 150 memory locations available for individual treatment programs (see INDIVIDUAL
Menu on page 27).
You can store patient-specific treatment programs on a therapy card (one treatment program per card),
which can be read by HIVAMAT 200 Personal. You can thus allow for individual, patient-specific therapy
even outside the treatment rooms (see THERAPY CARD Menu on page 29).
The instrument complies with all the valid safety standards, especially:
IEC 60601-1: Electronic medical devices, Part 1: General safety requirements
IEC 60601-1-4 Electronic Medical Devices - Part 1-4: General Safety Requirements - Accompanying
Standard: Programmable Electronic Medical Devices
All screen and control elements are integrated in the casing, so its surface is easy to clean and the product
is protected from dust.
The instrument's microprocessor monitors the safety-related components, prevents erroneous operation
and checks the instrument after switching it on.
CAUTION
The data selector <2> below the screen <1> is for selecting the
treatment parameters.
Rotate the data selector <2> to change the individual menu items
and press to confirm selection.
The data selector <2> is also used for switching over to the standby
mode and back, by keeping it pressed until the instrument starts up
or goes in the standby mode.
Using the intensity control <3> on the side, to the left of the screen
<1> you can increase the intensity. Increasing the intensity also starts
up the therapy timer and the treatment time countdown starts.
The card reader <4> on the side to the right of the screen <1> is for
inserting the therapy card, on which the specific treatment paramet-
ers can be stored. The therapy card must be inserted with the chip
facing back.
Card Reader <4>
The patient cable DEEP OSCILLATION sockets are for inserting the
supplied connecting cable DEEP OSCILLATION (Ref. No. 00261) for
Connecting Socket <5>
the hand applicator and the titanium neutral element.
Tips for connecting the cable are given under How to Prepare for
Treatment on page 16.
NOTICE
The power module <6> with the mains switch is on the rear of the
instrument. The power switch is for switching the mains supply to
Power Module <6>
HIVAMAT 200 Evident on and off.
When making any swivel or rotary movements of the screen, make sure that no body parts
(especially fingers) or accessories are pinched.
NOTICE
Before beginning the first treatment with the HIVAMAT 200 Evident, you must first put the
instrument through the proper initial start-up.
(1) Make sure that the instrument is switched off (the switch on power module <6> is in the position
0).
(2) Insert the supplied mains cable in the power module <6> firmly on the rear side of the instrument
and connect it to the plug socket. A secure connection is made through a socket outlet with earthing
contact with protective conductor.
(3) Align the screen <1> in a position, which ensures good visibility.
(4) Put the switch on the power module <6> in the ON position.
(5) The instrument completes an automatic function test, checking all functions and initialization values.
The instrument is now ready for operation. You are now in the MAIN MENU.
(1) Switch on the instrument by putting the switch on power module <6> off and on again.
The instrument completes an automatic function test, checking all functions and initialization values.
The instrument is now ready for operation. You are now in the MAIN MENU (or in the start menu
defined by you, see SETTINGS Menu on page 31).
NOTICE
Inform your service partner if the monitoring note does not disappear even after several func-
tion tests! Never perform a treatment when the proper function of the instrument is not as-
sured!
You will find the manufacturer's address under the menu item Help or at the beginning of
this manual.
If a functioning error is detected in the automatic function test or during operation, screen <1> shows a
message. It is easy to pinpoint and rectify faults on the basis of these messages. If a fault occurs, the con-
sequences are:
A message is shown on the screen <1> in the bottom half (e.f. for a communication error: Monitor-
ing Note 110).
The intensity (and therefore the voltage on the connecting socket <5>) is lowered to zero.
The operation of the instrument is blocked; pressing the data selector <2> will hide the message.
If the message was triggered by a brief malfunction that does not persists, the instrument is ready to oper-
ate immediately, without switching on and off.
If the fault however persists, the message appears again. Switch the instrument on and off repeatedly. The
instrument will then run an automatic function test (see Automatic Function Test on page 11). Check
whether the message continues to appear.
NOTICE
The instrument never goes into the power saving mode if there is voltage at the connecting
socket <5> or while the instrument is in operation.
If the HIVAMAT 200 Evident has not been used for a long time, it switches over into the power saving
mode. You can see whether the power saving mode is active, as well as the interval after which it switches
over, in the menu SETTINGS (see SETTINGS Menu on page 31).
If the instrument is is in the power saving mode, you can reactivate it by pressing or rotating the data se-
lector <2>. You can immediately continue to operate it.
You can do so by pressing the data selector <2> for a few seconds. The instrument then continues to run
(without output voltage), only the monitor and operating panel are switched off. Pressing or rotating the
switch again switches on the instrument again in the active operating state.
(1) Select a treatment and adjust the intensity by using the intensity control <3> to 100%. The treat-
ment frequency should be at least 100 Hz.
(2) Insert the patient cable DEEP OSCILLATION (Ref. No. 00260) in the connecting socket <5>.
(3) Insert both the ends of the connecting cable to be checked into the patient cable DEEP OSCILLATION.
(4) Move the connecting cable and observe the blue control light <7>:
The control light <7> lights up continuously: The cable is defective.
The control light <7> does not light up: The cable is OK
The control light <7> flashes when the cable is moved: The cable is defective.
(5) Change the defective cable immediately.
CAUTION
Replace the membrane immediately if damaged. Any damage to the oscillator membrane
might lead to harmless (but unpleasant) sensations on the skin.
(2) Cover
(4) Membrane
Important
The oscillator head is screwed into the handle. To remove the head from the handle, you must rotate it to
the left.
(1) Take off the clamping ring and remove the membrane.
(2) Insert the new membrane with the foil facing down loosely into the clamping ring, so that the mem-
brane foil protrudes laterally.
CAUTION
Adjust the intensity before each treatment and observe the contraindications (refer to Con-
traindications on page 2).
The HIVAMAT 200 Evident offers you the following options for treatment:
Direct access through the menu TREATMENT (see How to Perform a Treatment on page 18).
Via the treatment parameters in the menu TREATMENT EXPERT (see How to Set the Therapy Para-
meters on page 20).
Via the menu TREATMENT PROGRAMS (see How to Call a Treatment Program on page 24)
Via the menu INDIVIDUAL (see How to Rename an Individual Treatment Program on page 28)
Via the menu THERAPY CARD (see How to Read a Therapy Card on page 30)
NOTICE
The intensity should be selected according to the minimum dosage principle i.e. in the lower
effective range.
The hand applicator should always remain in contact with the patient's skin and be guided
over the body surface without removing it or stopping. Differentiate between therapy move-
ment and guided movements: The proximal therapy movement (for the body center) is ex-
ecuted with suitable pressure, the distal guided movement (from the body center outwards) is
done without pressure.
For low frequencies (5 ... 40 Hz) the therapy movement should be distinctly slow. The guided
movement can always be faster.
After switching over to a new therapy phase, it may be necessary to adjust the intensity in or-
der to reach the same resonant vibration strength.
Menu TREATMENT EXPERT (see How to Set the Therapy Parameters on page 20)
Storing individual treatment program locally in the instrument (see How to Save a Treat-
ment Program in the Instrument on page 22)
Storing therapy parameters on the therapy card (see How to Save a Treatment Program
on a Therapy Card on page 22)
Important
It is possible to store on the card only if a therapy card has been inserted
with the chip side facing back. The color of the symbols changes from
white to blue.
TREATMENT Menu
(1) Put the instrument in operation as described under Starting up the Instrument on page 11 below.
(2) Select the menu item Treatment . You come to the menu TREATMENT and the program steps of
the respective treatment program are displayed.
Tip
(3) Click on the symbol . You come to the TREATMENT EXPERT menu and can define the paramet-
ers for the therapy sections if you wish (see How to Set the Therapy Parameters on page 20).
Tip
You can already set the therapy time in the TREATMENT menu.
(4) By rotating the Intensity control <3> set the intensity. When the intensity is increased, the treat-
ment starts; the treatment time starts to count down and the control light <7> lights up.
When using with several therapy steps, the therapy parameters are switched over automatically after
the end of the first therapy phase. As soon as the time for the last treatment phase has expired, the
intensity returns to zero and an audible signal is emitted.
To change the therapy time of the individual therapy sections, proceed as follows:
(1) On the data selector <2> click the timer symbol behind the respective therapy section.
(2) Rotate the data selector <2> and shorten the therapy time to the desired value.
(3) Press the data selector <2> to confirm the selected therapy time.
Important
The total therapy time is always the sum of the therapy times of the individual therapy sec-
tions.
You can generally prolong or shorten the therapy time in the TREATMENT menu for all ther-
apy sections. In case of the active therapy section (that therapy section, in which the instru-
ment currently is, with the ongoing therapy and intensity set greater than zero), the new ther-
apy time is included only for the current treatment, if it is shorter than the originally set ther-
apy time.
Treatment duration Duration of the therapy section in question (can also be set in the TREATMENT
menu)
Oscillation Ratio between impulse time and pause time
Burst Burst frequency for the application of the impulses in packets
Mode Selection: Fixed frequency or frequency band
Frequency Treatment frequency for fixed frequency or lower and upper frequency for fre-
quency band
Sweep Attack time, sustain time and decay time (only for frequency band)
Important
In the current context, the non-selectable parameters are displayed in light grey.
Assuming that you have selected the frequency band 5 ... 250 Hz and the three times 1s, 2s
and 3s are selected:
The frequency is initially at the maximum value (250 Hz) for one second. Thereafter the fre-
quency reduces to two seconds long, continuously up to 5 Hz. This frequency is kept for three
seconds. Finally, the frequency lengthens continuously, until after two seconds, the upper lim-
The selection of a burst frequency emits impulses in packets. The burst value here corresponds
to the number of the compounded packets per second. For example, a burst of the value 1 is
set for a selected fixed frequency of 10 Hz (i.e. 10 impulses per second), the impulses are
emitted in the initial half of a second; no impulses are emitted in the second half. This se-
quence continuously repeats itself.
Important
The HIVAMAT 200 Evident automatically ensures that the lower frequency is not great-
er than the upper frequency. The lower frequency must always be 5 Hz smaller than the
upper frequency.
(9) On the data selector <2> select the Sweep field (only for frequency band; not possible for fixed fre-
quency). Set here the time spans (respectively 1 ... 60 s).
(10) Repeat the steps discussed here if necessary, for the other therapy sections.
(11) Click on to confirm. Your settings are saved and you return to the TREATMENT menu.
The following characters are available to you for naming the treatment program:
Capital letters A ...Z
Numbers 0 ... 9
Special characters ., _, +, -
Space SPACE
Important
The treatment programs are always sorted alphabetically according to their names.
Menu abc
(3) Continue until the desired name is displayed in the top line.
(4) Click on the symbol . The new name is saved and you return to the respective initial menu.
Tip
You can use the symbol to delete the last character of the entered name.
NOTICE
If no memory space is available, the message Attention: No more memory space avail-
able. is displayed. You can delete an existing treatment program (see How to Delete an Indi-
vidual Treatment Program on page 29).
(1) In the menu TREATMENT or TREATMENT EXPERT click the symbol . The message Treatment is
being updated is displayed and the treatment is saved, provided at least one memory location is
free.
Tip
You can save a treatment program stored on a therapy card from the THERAPY CARD menu
locally in the instrument (also refer to How to Store a Treatment Program from the Therapy
Card in the HIVAMAT 200 Evident on page 30).
NOTICE
If a treatment program with the same name has already been saved in the instrument, the
message The name is already being used by a treatment program. Overwrite existing
treatment program? is displayed. You can overwrite the existing program or rename the
treatment program to be saved (see How to Rename a Treatment Program on page 21).
Tip
The therapy card can also be read with the HIVAMAT 200 Personal, making patient treat-
ment at home possible.
NOTICE
You can overwrite the parameters saved on the card. However, the original parameters can
no longer be retrieved.
(1) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(2) Now click on the symbol . If the card is not empty, the message Caution: There is already a
treatment program on the therapy card! Would you like to overwrite the treatment pro-
gram? is displayed.
(3) Click on . The message CAUTION: Treatment program is being written on the therapy card!
Please do not remove the card! is displayed and the individual treatment program is saved on the
therapy card.
Tip
You can save treatment programs from the INDIVIDUAL menu on a therapy card (see How
to Save an Individual Treatment Program on the Therapy Card on page 28).
The TREATMENT EXAMPLES / TREATMENT PROGRAMS menu offers you information on how to per-
form treatment at the individual body regions, as well as access to the preset treatment programs.
Carrying out the therapy for the current therapy program (see How to Perform a Treat-
ment on page 18)
Invoke animation
Start animation
Pause animation
(1) In the MAIN MENU select the entry Treatment Examples / Treatment Programs . The TREAT-
MENT EXAMPLES / TREAMENT PROGRAMS menu is displayed.
(2) In the TREATMENT EXAMPLES / TREATMENT PROGRAMS menu, click TREATMENT
PROGRAMS. A list of available treatment programs is shown.
(3) Select the desired program, e.g. Program 1. The parameters of the program are displayed.
(4) Insert the applicator of your choice into the connecting cable DEEP OSCILLATION.
(5) Click on the symbol . You come to the menu TREATMENT from where you can begin the therapy
(see How to Perform a Treatment on page 18).
(1) In the TREATMENT EXAMPLES / TREATMENT PROGRAMS menu, click TREATMENT EXAMPLES.
A list of available treatment examples is shown:
Treatment Examples
(2) Rotate the data selector <2> until you have selected the treatment example of interest (e.g. Lumbar
region / gluteal region / hip) and press. The treatment instructions for the selected treatment example
are displayed:
Important
If the applicator head has a green background in the video, the movement is to be per-
formed with pressure; if it has a white background, without pressure.
INDIVIDUAL menu
Call current treatment program in the menu TREATMENT (see How to Rename an Indi-
vidual Treatment Program on page 28)
Saving current treatment program on therapy card (see How to Save an Individual
Treatment Program on the Therapy Card on page 28)
NOTICE
It is possible to store on the card only if a therapy card has been inserted
with the chip side facing back. The color of the symbols changes from
white to blue.
Delete current treatment program (see How to Delete an Individual Treatment Program
on page 29)
(1) In the MAIN MENU select the menu item Individual. The INDIVIDUAL menu is displayed.
(2) On the data selector <2>, select the treatment program which you wish to call.
(3) Select the menu item Treatment . You come to the menu TREATMENT and the program steps of
the respective treatment program are displayed.
The process is further described in section How to Perform a Treatment on page 18.
(4) Increase the intensity to the desired level with the intensity control <3>. The therapy time starts the
countdown.
Tip
The therapy card can also be read with the HIVAMAT 200 Personal, making patient treat-
ment at home possible.
NOTICE
You can overwrite the parameters saved on the card. However, the original parameters can
no longer be retrieved.
(1) In the MAIN MENU select the menu item Individual. The INDIVIDUAL menu is displayed.
(2) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(3) Now click on the symbol . If the card is not empty, the message Caution: There is already a
treatment program on the therapy card! Would you like to overwrite the treatment pro-
gram? is displayed.
(4) Click on . The message CAUTION: Treatment program is being written on the therapy card!
Please do not remove the card! is displayed and the individual treatment program is saved on the
therapy card.
Tip
You can save treatment programs from the TREATMENT menu on a therapy card (see How
to Save a Treatment Program on a Therapy Card on page 22).
NOTICE
(1) In the MAIN MENU select the menu item Individual. The INDIVIDUAL menu is displayed.
(2) On the data selector <2> select the treatment program which you wish to delete.
(3) Click on the symbol . The message Do you really want to delete the selected treatment?
menu is displayed.
(4) Click on the symbol to confirm. The treatment program is deleted and the memory space is free
again.
The therapy card can also be read with the HIVAMAT 200 Personal, making patient treatment at home
possible.
Important
Therapy Cards programed for the HIVAMAT 200 Personal for the various fields of application
cannot be read with the HIVAMAT 200 Evident. A corresponding error message is shown if
you try.
Starting the therapy that is saved on the therapy card (see How to Read a Therapy Card
on page 30)
Saving treatment program stored on the card locally in the instrument (see How to
Store a Treatment Program from the Therapy Card in the HIVAMAT 200 Evident on
page 30)
(1) In the MAIN MENU select the Therapy card menu. The THERAPY CARD menu is displayed.
(2) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(3) Now click on the symbol to go to the TREATMENT menu (see How to Perform a Treatment on
page 18).
6.4.2 Storing a Treatment Program from the Therapy Card in the HIVAMAT 200
Evident
You can also store a treatment program that was previously saved on a therapy card as an individual treat-
ment program in the HIVAMAT 200 Evident.
How to Store a Treatment Program from the Therapy Card in the HIVAMAT 200 Evident
(1) In the MAIN MENU select the Therapy card menu. The THERAPY CARD menu is displayed.
(2) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(3) Click on the symbol . The treatment is displayed, if at least one memory location is still free.
NOTICE
If a treatment program with the same name has already been saved in the instrument, the
message The name is already being used by a treatment program. Overwrite existing
treatment program? is displayed. You can overwrite the existing program or rename the
treatment program to be saved (see How to Rename a Treatment Program on page 21).
Time after which the instrument will go into the power saving mode (1 ... 60 min or off)
Dialog language
Start menu (possible values: Main menu, Individual, Treatment Examples / Treatment
Programs, Last treatment, Standard treatment, Therapy card)
(1) In the MAIN MENU select the menu item Settings. The SETTINGS menu is displayed.
SETTINGS menu
(2) Rotate the data selector <2> to select the parameters you wish to change (e.g. Brightness) and
press to confirm.
(3) Rotate the data selector <2> until the parameter assumes the desired value and press it to confirm.
(4) Change all the desired parameters and click on the symbol to include the settings.
The new settings are effective immediately.
(1) In the MAIN MENU select the menu item Help. The HELP menu is displayed.
HELP menu
(2) Click the arrow symbol to browse through the help menu.
Repairs to the instrument may only be performed by parties duly authorized by PHYSIOMED
ELEKTROMEDIZIN AG. Any repairs performed by an authorized agent must be accompanied by written
certification, describing the nature and extent of the repairs undertaken, as applicable with details regard-
ing changes to nominal operating values or the operational range. The certification must also contain the
date performed, the name of the repair company and the signature of the service person. When defective,
components affecting the safe operation of the instrument must be replaced by manufacturers original
parts. Upon request, wiring diagrams, parts lists and service instructions can be made available to qualified
technical personnel employed by the customer.
We recommend having the instrument, including all accessories, serviced at regular intervals.
NOTICE
The oscillator plate must only be disinfected using an agent, but not be sterilized using heat.
All parts of the oscillator head must be absolutely dry before using them next time.
Clean the device and the accessories regularly with an aldehyde-based disinfectant. Switch off the instru-
ment by any means before cleaning.
Clean the instrument and the handle by using a soft foam cloth. Use caution so that no liquid substances
penetrate into the instrument.
Use a new or disinfected oscillator head on the hand applicator for every patient. Clean, disinfect or steril-
ise the hand applicator regularly. The clamping ring, oscillator plate and handle are resistant against disin-
fecting agents and can be sterilised up to 275F (135C).
A.4 Disposal
After the service life of the instrument, dispose of it in conformance with the applicable regulations for en-
vironment protection.
Handle, clean, and dispose of components and accessories that have come in contact with bodily fluids ac-
A.5 Training
Training is given in the course of the introduction to the instrument.
Should the need arise however, continued training can be provided by a medical device consultant. Please
contact PHYSIOMED ELEKTROMEDIZIN AG.
Portable and mobile HF communication devices can affect medical electrical devices (see the supplement
on electromagnetic compatibility, technical description).
A.8 Manufacturer
PHYSIOMED ELEKTROMEDIZIN AG
Hutweide 10
91220 Schnaittach
Germany
The instrument is to be used exclusively with original accessories. Otherwise, treatment results
might be insufficient.
ATTENTION: This upgrade package is only available when purchased together with at least
one HIVAMAT 200 Evident.
function test 11
Index
H
A
hand applicator 13 , 16
accessories 36 help 32
available 36 HIVAMAT 200 Personal 29
supplied 36 hypersensitivity
ambient temperature 34 to electrostatic fields 2
application 2
I
B
implants
blood circulation electronic 2
inscrease 2 important user information 1
brightness 31 indicators 8
function 8
C individual program
delete 29
cable check 13 invocation 28
card reader 9 infections
cardiac arrhythmia 2 acute 2
cardiac insufficiency 2 inflammation
cardiac oedemata acute 2
decompensated 2 instrument description 7
cardiac pacemaker 2 instrument errors 12
caution symbols 4 instrument overview 7
cleaning 33 instrument symbols 4
compatibility intensity control 9
electromagnetic 34 invocation
connecting socket 10 individual program 28
contraindications 2
controls 8 L
function 8
curve form 34 language 31
limited warranty 33
D load impedance 34
data selector 9 M
definitions 4
delete maintenance 33
individual program 29 malignant diseases 2
dimensions 34 membrane
disinfection 33 replace 13
disposal 33 modulation 34
muscle aches 2
E muscle relaxation 2
electrostatic field 2 N
EMC 34
environment protection 33 name
error codes 12 entering 20
erysipelas 2 notes
general 5
F
O
fields
electrostatic 2 operating principle 2
outlet frequency 34
output voltage 34 V
P venous diseases
acute 2
pain reduction 2 volume 31
pitch 31
power module 10 W
power saving mode 12 , 31
precaution 3 warnings 2
pregnancy 2 warranty 33
protection class 34 weight 34
write
R therapy card 30
read
therapy card 30
repairs 33
save
treatment program 21
screen 8
brightness 31
service 33
settings 31
skin diseases
infectious 2
standby mode 12
start menu 31
starting up 11
technical data 34
therapeutic fields 7
therapy card 29
read 30
write 30
therapy parameters
setting 19
therapy time
setting 18
thromboses
untreated 2
training 34
treatment 16
treatment examples 23
TREATMENT EXAMPLES / TREATMENT PROGRAMS
menu 23
treatment program
save 21
treatment programs 23
individual 27
tuberculosis
active 2