READ THIS MANUAL BEFORE USING THIS PRODUCT

Important Safety Information Inside

Keep this manual in a safe location


for future reference

OPERATING INSTRUCTIONS

HIVAMAT 200 Evident

01625 US
Legal Notice

Distribution, reproduction and translation of the software and its documentation (or excerpts thereof) are
prohibited without the prior written consent of PHYSIOMED ELEKTROMEDIZIN AG.

PHYSIOMED ELEKTROMEDIZIN AG reserves the right to change the software and associated data as well
as the documentation without notice. All other rights reserved.

Naimco, Inc. Dba Richmar


4120 South Creek Road
Chattanooga TN 37406
USA
PHONE: 423-7604194
FAX: 423-667-2325
E-MAIL:jeremybarnes@richmarweb.com
WEB: www.richmarweb.com

HIVAMAT is a registered trademark of PHYSIOMED ELEKTROMEDIZIN AG. The therapy mode is based on
patented technology.

HIVAMAT 200 Evident is made in Germany in compliance with the quality requirements of DIN ISO
13485:2003, FDA regulations, and the applicable safety standards:
IEC 60601-1: Electronic medical devices, Part 1: General safety requirements
IEC 60601-1-4 Electronic Medical Devices - Part 1-4: General Safety Requirements - Accompanying
Standard: Programmable Electronic Medical Devices

Operating Instructions last updated on August 1, 2016.

THIRD-PARTY SOFTWARE STATEMENTS

This product contains a Linux kernel (based on Linux kernel 2.6.15), Busybox version 1.1.3, MPlayer, all of which are distributed un-
der the GPL V2.

This product contains glibc, SDL, SDL_ttf, termcap, all of which are distributed under the LGPL.

This product contains e2fsprogs, usbutils, mtd, all of which are distributed under the GPL.

This product contains libgcc_s.so and libstdc libraries which are distributed under the GPL with run-time exception.

This product contains zlib 1.2.2 software.

Portions of this software are copyright 2008 The FreeType Project (www.freetype.org). All rights reserved.

The mass-storage-card included in the device includes source codes and the individual third-party-software-licences.
Table of Contents PHYSIOMED ELEKTROMEDIZIN AG

Table of Contents

Chapter 1 Important Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . 1

Chapter 2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.3 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.4 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Chapter 3 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.1 Definition of Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.2 Instrument Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.3 General Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.4 Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.5 Instrument Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Chapter 4 Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8


4.1 Function of Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1.1 Screen <1> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1.2 Data Selector <2> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1.3 Intensity Control <3> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1.4 Card Reader <4> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1.5 Connecting Socket <5> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.6 Power Module <6> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.7 Control Light <7> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Chapter 5 Instrument Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11


5.1 Starting up the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.2 Automatic Function Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.3 Instrument Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.4 Power Saving Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.5 Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.6 Cable Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.7 Hand Applicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

HIVAMAT 200 Evident i


PHYSIOMED ELEKTROMEDIZIN AG Table of Contents

Chapter 6 Overview of the Menus. . . . . . . . . . . . . . . . . . . . . . . . . . . 16


6.1 TREATMENT Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.1.1 Performing the Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.1.2 Setting the Therapy Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.1.3 TREATMENT EXPERT Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.1.4 Entering the Name of a Treatment Program . . . . . . . . . . . . . . . . . . 20
6.1.5 Saving a Treatment Program . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.1.6 Saving a Treatment Program on the Therapy Card . . . . . . . . . . . . . . . 22
6.2 TREATMENT EXAMPLES / TREATMENT PROGRAMS Menu . . . . . . . . . . . 23
6.3 INDIVIDUAL Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.3.1 Calling up an Individual Treatment Program . . . . . . . . . . . . . . . . . . 28
6.3.2 Saving an Individual Treatment Program on the Therapy Card . . . . . . . . . 28
6.3.3 Deleting an Individual Treatment Program . . . . . . . . . . . . . . . . . . . 29
6.4 THERAPY CARD Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.4.1 Reading a Therapy Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.4.2 Storing a Treatment Program from the Therapy Card in the HIVAMAT 200 Evident
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.5 SETTINGS Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.6 HELP Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Appendix A Reference Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 33


A.1 Service, Repairs, Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 33
A.2 Cleaning and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
A.3 Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
A.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
A.5 Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
A.6 UL 60601-1-2 Electromagnetic Compatibility Declaration . . . . . . . . . . . 34
A.7 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
A.7.1 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
A.8 Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
A.9 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Appendix B Scope of Supply and Accessories . . . . . . . . . . . . . . . . . . . . . 36


B.1 Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
B.2 Available Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

ii HIVAMAT 200 Evident


Table of Contents PHYSIOMED ELEKTROMEDIZIN AG

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

HIVAMAT 200 Evident iii


PHYSIOMED ELEKTROMEDIZIN AG Table of Contents

iv HIVAMAT 200 Evident


1 Important Safety Instructions PHYSIOMED ELEKTROMEDIZIN AG

Chapter 1 Important Safety Instructions


WARNING

Federal law restricts this instrument to sale by or on the order of a medical practitioner li-
censed by the law of the State in which he/she practices to use or order the use of the instru-
ment.

The instrument must only be operated by medical practicioners who have undergone special
training. Use by unqualified personnel or consumers bears the risk of serious injury.

You must operate the instrument properly, i.e. in accordance with the Operating Instructions.

If the patient experiences any unpleasant sensation during treatment, the practitioner must
stop the treatment immediately. Prior to treatment, the practitioner has to explicitely inform
the patient about this.

CAUTION

United States law restricts this instrument to sale by or on the order of a licensed healthcare
practitioner.

When making any swivel or rotary movements of the screen, make sure that no body parts
(especially fingers) or accessories are pinched.

The instrument is to be used exclusively with original accessories. Otherwise, treatment results
might be insufficient.

The instrument is to be used exclusively with original accessories. Otherwise, treatment results
might be insufficient.

Operating the instrument in the proximity (e.g. 1 m) of a short-wave or micro-wave therapy


unit may cause output irregularities and should be avoided for this reason, as well as simul-
taneous connection of the patient to high-frequency surgical instrument.

Adjust the intensity before each treatment and observe the contraindications (refer to Con-
traindications on page 2).

HIVAMAT 200 Evident 1


PHYSIOMED ELEKTROMEDIZIN AG 2 Introduction

Chapter 2 Introduction

2.1 Indications for Use

WARNING

The instrument must only be operated by medical practicioners who have undergone special
training. Use by unqualified personnel or consumers bears the risk of serious injury.

The function of HIVAMAT 200 Evident is based on a pulsed electrostatic field which is built up in the pa-
tient's body region to be treated. Its frequency changes between 5 - 250 Hz, according to the selected
program. Due to the movement of the hand applicator, a vibrating or pumping effect with deep impact is
induced in the patient's tissue. The treatment is carried out with the supplied hand applicators.

Treatment with HIVAMAT 200 Evident has the following positive effects on the treated tissue:
muscle relaxation
relief of muscle aches and pains
increased local blood circulation

2.2 Contraindications
Therapy with HIVAMAT 200 Evident is not indicated in the following cases:
acute infections
acute inflammations with participation of pathogenic agents
active tuberculosis
acute venous diseases (untreated thromboses)
untreated malignant processes
erysipelas
patients and therapists with cardiac pacemakers and other electronic implants
heart disorders and diseases, especially cardiac insufficiency, decompensated cardiac edema, and car-
diac arrhythmia
pregnancy
hypersensitivity to electrostatic fields
infectious skin diseases

2.3 Warnings
(a) Pulsed electrostatic fields should not be applied over swollen, infected, or inflamed areas or skin erup-
tions, e.g., phlebitis, thrombophlebitis, varicose veins, etc.
(b) Pulsed electrostatic fields should not be applied over, or in proximity to, cancerous lesions.
(c) Set the line voltage at the mains module to match the incoming line voltage.
(d) HIVAMAT 200 Evident must not be used in the operating room, or in the presence of flammable an-
esthetic gases mixed with air, oxygen or nitrous oxide.

2 HIVAMAT 200 Evident


2 Introduction PHYSIOMED ELEKTROMEDIZIN AG

Please also note the contraindications to pulsed static electromagnetic fields in Contraindications on page
2.

2.4 Precautions
(a) Caution should be used for patients with suspected or diagnosed heart problems.
(b) Caution should be used for patients with suspected or diagnosed epilepsy.
(c) Caution should be used:
When there is a tendency to hemorrhage following acute trauma or fracture;
following recent surgical procedures when vibration may disrupt the healing process;
over the menstruating uterus; and
over areas of the skin which lack normal sensation.
(d) Instrument settings should be based on the guidance of the prescribing practitioner.
(e) Massage-intensifying systems should be kept out of the reach of children.
(f) Massage-intensifying systems should be used only with the accessories recommended by the manufac-
turer.
(g) Massage-intensifying systems should not be used while driving, operating machinery, or during any
activity in which vibration may put the user at undue risk of injury.
(h) Massage-intensifying systems emit electromagnetic waves: Maintain sufficient separation from other
electronic devices, and at least one-meter separation from any short wave or microwave therapy unit.
(i) HIVAMAT 200 Evident should only be used under medical supervision.

HIVAMAT 200 Evident 3


PHYSIOMED ELEKTROMEDIZIN AG 3 Introduction

Chapter 3 Introduction
With HIVAMAT 200 Evident you have acquired an extremely versatile deep oscillation system. The instru-
ment will only show its true potential, however, if you are well informed about its functions. For this reas-
on, carefully read these Operating Instructions and familiarize yourself with the use of the instrument.

3.1 Definition of Symbols Used


Please note the following typographical symbols in these Operating Instructions:
Cross references and important terms used for the first time in this document are written in italics.
Names of menus and symbols on the display are written in bold typeface.

Paragraphs that deserve special attention are highlighted in the following way:

Type Meaning
DANGER Alerts of extremely dangerous situations that might lead to death or serious injury.
WARNING Warns of other situations that might also lead to death or serious injury.
CAUTION For risks that might lead to less serious injury.
NOTICE Only for dangerous situations that might lead to damage to the product or other assets.

3.2 Instrument Symbols


The meanings of the symbols placed on the instrument are as follows:

Symbol Meaning
CAUTION! Refer to the Operating Instructions and consider the physiological effects
of the parameters.

Application part Type BF, not connected to the protective conductor

Ensure environment-friendly disposal. Instruments that have reached the end of


their life cycle (see Disposal on page 33) must by any means be disposed of separ-
ately from household garbage.

4 HIVAMAT 200 Evident


3 Introduction PHYSIOMED ELEKTROMEDIZIN AG

3.3 General Notes

CAUTION

Operating the instrument in the proximity (e.g. 1 m) of a short-wave or micro-wave therapy


unit may cause output irregularities and should be avoided for this reason, as well as simul-
taneous connection of the patient to high-frequency surgical instrument.

The instrument may only be operated in dry rooms designed for medical purposes. It complies with the
technical specifications of IEC 60601-1.

It is not intended for operation in explosion hazard zones or hydrotherapy rooms. Drastic temperature
changes should be avoided, since condensation could be caused within the instrument. Do not start up
the instrument until it is in thermal equilibrium with its environment.

The instrument is to be operated properly, i.e. in accordance with the Operating Instructions.

The instrument housing must not be opened.

HIVAMAT 200 Evident 5


PHYSIOMED ELEKTROMEDIZIN AG 3 Introduction

3.4 Instrument Overview

6 HIVAMAT 200 Evident


3 Introduction PHYSIOMED ELEKTROMEDIZIN AG

Legend
1 Screen 2 Data Selector
3 Intensity Control 4 Card Reader
5 Connecting Socket 6 Power Module
7 Control Light

3.5 Instrument Description


HIVAMAT 200 Evident is a deep oscillation system which is successfully used in various therapeutic fields
of medicine. It is designed as a therapy device mainly for use in clinical and therapeutic environment.

You have a total of 150 memory locations available for individual treatment programs (see INDIVIDUAL
Menu on page 27).

You can store patient-specific treatment programs on a therapy card (one treatment program per card),
which can be read by HIVAMAT 200 Personal. You can thus allow for individual, patient-specific therapy
even outside the treatment rooms (see THERAPY CARD Menu on page 29).

The instrument complies with all the valid safety standards, especially:
IEC 60601-1: Electronic medical devices, Part 1: General safety requirements
IEC 60601-1-4 Electronic Medical Devices - Part 1-4: General Safety Requirements - Accompanying
Standard: Programmable Electronic Medical Devices

HIVAMAT 200 Evident 7


PHYSIOMED ELEKTROMEDIZIN AG 4 Controls and Indicators

Chapter 4 Controls and Indicators


The design of HIVAMAT 200 Evident allows for easy operation, in combination with a variety of func-
tions.

All screen and control elements are integrated in the casing, so its surface is easy to clean and the product
is protected from dust.

The instrument's microprocessor monitors the safety-related components, prevents erroneous operation
and checks the instrument after switching it on.

4.1 Function of Controls and Indicators


The individual screen and control elements of the HIVAMAT 200 Evident are presented in the following
sections. The numbers in the angle brackets refer to the Instrument Overview on page 7.

4.1.1 Screen <1>

CAUTION

When making any swivel or rotary


movements of the screen, make sure
that no body parts (especially fingers)
or accessories are pinched.

The screen <1> shows the selectable menu items or


the therapy parameters, depending on the current
context.

The parameters are selected by using the data se-


Screen <1>
lector <2>.

The screen <1> can be rotated and swiveled, thus


allowing an optimum positioning, depending on the
treatment situation.

8 HIVAMAT 200 Evident


4 Controls and Indicators PHYSIOMED ELEKTROMEDIZIN AG

4.1.2 Data Selector <2>

The data selector <2> below the screen <1> is for selecting the
treatment parameters.

Rotate the data selector <2> to change the individual menu items
and press to confirm selection.

The data selector <2> is also used for switching over to the standby
mode and back, by keeping it pressed until the instrument starts up
or goes in the standby mode.

Data Selector <2>

4.1.3 Intensity Control <3>

Using the intensity control <3> on the side, to the left of the screen
<1> you can increase the intensity. Increasing the intensity also starts
up the therapy timer and the treatment time countdown starts.

4.1.4 Card Reader <4>

The card reader <4> on the side to the right of the screen <1> is for
inserting the therapy card, on which the specific treatment paramet-
ers can be stored. The therapy card must be inserted with the chip
facing back.
Card Reader <4>

HIVAMAT 200 Evident 9


PHYSIOMED ELEKTROMEDIZIN AG 4 Controls and Indicators

4.1.5 Connecting Socket <5>

The connecting socket <5> on the rear of the instrument is to be


used for inserting the supplied patient cable DEEP OSCILLATION (Ref.
No. 00260).

The patient cable DEEP OSCILLATION sockets are for inserting the
supplied connecting cable DEEP OSCILLATION (Ref. No. 00261) for
Connecting Socket <5>
the hand applicator and the titanium neutral element.

Tips for connecting the cable are given under How to Prepare for
Treatment on page 16.

4.1.6 Power Module <6>

NOTICE

For mains supply only use the mains cable provided by


the manufacturer!

The power module <6> with the mains switch is on the rear of the
instrument. The power switch is for switching the mains supply to
Power Module <6>
HIVAMAT 200 Evident on and off.

4.1.7 Control Light <7>

The control light <7> on the front of the instrument lights up as


soon as the intensity is turned up. At a low therapy frequency, the
light flickers in the rhythm of the impulses. If the electrostatic therapy
field is short-circuited, the light goes off even if the intensity is turned
up.
Control Light <7>
The control light <7> can also be used for checking the supplied
connecting cable DEEP OSCILLATION (Ref. No. 00261) (see Cable
Check on page 13).

10 HIVAMAT 200 Evident


5 Instrument Operation PHYSIOMED ELEKTROMEDIZIN AG

Chapter 5 Instrument Operation


CAUTION

The screen <1> can be rotated and swiveled.

When making any swivel or rotary movements of the screen, make sure that no body parts
(especially fingers) or accessories are pinched.

NOTICE

Before beginning the first treatment with the HIVAMAT 200 Evident, you must first put the
instrument through the proper initial start-up.

5.1 Starting up the Instrument


Before you can execute the first treatment with the HIVAMAT 200 Evident, you must put the instrument
through the proper initial start-up.

How to Start up the Instrument


Proceed as follows, if you wish to start up the instrument for the first time:

(1) Make sure that the instrument is switched off (the switch on power module <6> is in the position
0).
(2) Insert the supplied mains cable in the power module <6> firmly on the rear side of the instrument
and connect it to the plug socket. A secure connection is made through a socket outlet with earthing
contact with protective conductor.
(3) Align the screen <1> in a position, which ensures good visibility.
(4) Put the switch on the power module <6> in the ON position.
(5) The instrument completes an automatic function test, checking all functions and initialization values.
The instrument is now ready for operation. You are now in the MAIN MENU.

5.2 Automatic Function Test


If you are not sure that your HIVAMAT 200 Evident is working properly, you can run an automatic func-
tion test.

How to Check the Function of the Instrument


Proceed as follows, if you wish to run an automatic function test:

(1) Switch on the instrument by putting the switch on power module <6> off and on again.
The instrument completes an automatic function test, checking all functions and initialization values.
The instrument is now ready for operation. You are now in the MAIN MENU (or in the start menu
defined by you, see SETTINGS Menu on page 31).

HIVAMAT 200 Evident 11


PHYSIOMED ELEKTROMEDIZIN AG 5 Instrument Operation

5.3 Instrument Errors

NOTICE

Inform your service partner if the monitoring note does not disappear even after several func-
tion tests! Never perform a treatment when the proper function of the instrument is not as-
sured!

You will find the manufacturer's address under the menu item Help or at the beginning of
this manual.

If a functioning error is detected in the automatic function test or during operation, screen <1> shows a
message. It is easy to pinpoint and rectify faults on the basis of these messages. If a fault occurs, the con-
sequences are:
A message is shown on the screen <1> in the bottom half (e.f. for a communication error: Monitor-
ing Note 110).
The intensity (and therefore the voltage on the connecting socket <5>) is lowered to zero.
The operation of the instrument is blocked; pressing the data selector <2> will hide the message.

If the message was triggered by a brief malfunction that does not persists, the instrument is ready to oper-
ate immediately, without switching on and off.

If the fault however persists, the message appears again. Switch the instrument on and off repeatedly. The
instrument will then run an automatic function test (see Automatic Function Test on page 11). Check
whether the message continues to appear.

5.4 Power Saving Mode

NOTICE

The instrument never goes into the power saving mode if there is voltage at the connecting
socket <5> or while the instrument is in operation.

If the HIVAMAT 200 Evident has not been used for a long time, it switches over into the power saving
mode. You can see whether the power saving mode is active, as well as the interval after which it switches
over, in the menu SETTINGS (see SETTINGS Menu on page 31).

If the instrument is is in the power saving mode, you can reactivate it by pressing or rotating the data se-
lector <2>. You can immediately continue to operate it.

5.5 Standby Mode


If you would like to briefly deactivate the HIVAMAT 200 Evident (e.g. between two operating sessions),
you can put it in the standby mode.

You can do so by pressing the data selector <2> for a few seconds. The instrument then continues to run

12 HIVAMAT 200 Evident


5 Instrument Operation PHYSIOMED ELEKTROMEDIZIN AG

(without output voltage), only the monitor and operating panel are switched off. Pressing or rotating the
switch again switches on the instrument again in the active operating state.

5.6 Cable Check


If you get the impression during treatment that a connecting cable is defective, you can have the instru-
ment check the cables.

How to Check a Cable

(1) Select a treatment and adjust the intensity by using the intensity control <3> to 100%. The treat-
ment frequency should be at least 100 Hz.
(2) Insert the patient cable DEEP OSCILLATION (Ref. No. 00260) in the connecting socket <5>.
(3) Insert both the ends of the connecting cable to be checked into the patient cable DEEP OSCILLATION.
(4) Move the connecting cable and observe the blue control light <7>:
The control light <7> lights up continuously: The cable is defective.
The control light <7> does not light up: The cable is OK
The control light <7> flashes when the cable is moved: The cable is defective.
(5) Change the defective cable immediately.

5.7 Hand Applicator

CAUTION

Replace the membrane immediately if damaged. Any damage to the oscillator membrane
might lead to harmless (but unpleasant) sensations on the skin.

The hand applicator consists of the following parts:

HIVAMAT 200 Evident 13


PHYSIOMED ELEKTROMEDIZIN AG 5 Instrument Operation

(1) Handle (Ref.-No. 00379)

(2) Cover

(3) Oscillator Plate

(4) Membrane

(5) Clamp Ring

Important

(2) - (5) make up the oscillator head in


the respective size.

Hand Applicator (Exploded View)

The oscillator head is screwed into the handle. To remove the head from the handle, you must rotate it to
the left.

How to Replace the Membrane

(1) Take off the clamping ring and remove the membrane.
(2) Insert the new membrane with the foil facing down loosely into the clamping ring, so that the mem-
brane foil protrudes laterally.

Changing the Membrane - Inserting


(3) Press the oscillator plate firmly onto the foam rubber area of the membrane until the clamping ring
engages with the oscillator plate. Make sure that the foil is tightly fixed with the clamping ring on the
outside, and that the clamping ring is engaged!

14 HIVAMAT 200 Evident


5 Instrument Operation PHYSIOMED ELEKTROMEDIZIN AG

Changing the Membrane - Engaging the Oscillator Plate

HIVAMAT 200 Evident 15


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

Chapter 6 Overview of the Menus

6.1 TREATMENT Menu

CAUTION

Adjust the intensity before each treatment and observe the contraindications (refer to Con-
traindications on page 2).

The HIVAMAT 200 Evident offers you the following options for treatment:
Direct access through the menu TREATMENT (see How to Perform a Treatment on page 18).
Via the treatment parameters in the menu TREATMENT EXPERT (see How to Set the Therapy Para-
meters on page 20).
Via the menu TREATMENT PROGRAMS (see How to Call a Treatment Program on page 24)
Via the menu INDIVIDUAL (see How to Rename an Individual Treatment Program on page 28)
Via the menu THERAPY CARD (see How to Read a Therapy Card on page 30)

How to Prepare for Treatment

(1) Dry the skin area to be treated or powder if necessary.


(2) Insert the patient cable DEEP OSCILLATION in the connecting socket <5>.
(3) Insert the DEEP OSCILLATION connecting cables for the hand applicator and for the titanium neutral
element into the patient cable DEEP OSCILLATION. Since the instrument works in dual phase, you
need not watch out for the polarity.
(4) Place the titanium neutral element in the hands of the person to be treated.

6.1.1 Performing the Treatment

NOTICE

The intensity should be selected according to the minimum dosage principle i.e. in the lower
effective range.

The hand applicator should always remain in contact with the patient's skin and be guided
over the body surface without removing it or stopping. Differentiate between therapy move-
ment and guided movements: The proximal therapy movement (for the body center) is ex-
ecuted with suitable pressure, the distal guided movement (from the body center outwards) is
done without pressure.

For low frequencies (5 ... 40 Hz) the therapy movement should be distinctly slow. The guided
movement can always be faster.

After switching over to a new therapy phase, it may be necessary to adjust the intensity in or-
der to reach the same resonant vibration strength.

16 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

In the menu TREATMENT you can select the following options:

Symbols in the TREATMENT Menu

Back to MAIN MENU

Menu TREATMENT EXPERT (see How to Set the Therapy Parameters on page 20)

Storing individual treatment program locally in the instrument (see How to Save a Treat-
ment Program in the Instrument on page 22)

Storing therapy parameters on the therapy card (see How to Save a Treatment Program
on a Therapy Card on page 22)

Important

It is possible to store on the card only if a therapy card has been inserted
with the chip side facing back. The color of the symbols changes from
white to blue.

TREATMENT Menu

HIVAMAT 200 Evident 17


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

How to Perform a Treatment

(1) Put the instrument in operation as described under Starting up the Instrument on page 11 below.
(2) Select the menu item Treatment . You come to the menu TREATMENT and the program steps of
the respective treatment program are displayed.

Tip

If so desired, you can raise the intensity at this time.

(3) Click on the symbol . You come to the TREATMENT EXPERT menu and can define the paramet-
ers for the therapy sections if you wish (see How to Set the Therapy Parameters on page 20).

Tip

You can already set the therapy time in the TREATMENT menu.

(4) By rotating the Intensity control <3> set the intensity. When the intensity is increased, the treat-
ment starts; the treatment time starts to count down and the control light <7> lights up.
When using with several therapy steps, the therapy parameters are switched over automatically after
the end of the first therapy phase. As soon as the time for the last treatment phase has expired, the
intensity returns to zero and an audible signal is emitted.

6.1.2 Setting the Therapy Time


In the TREATMENT menu, the preset therapy times for all therapy steps are displayed, as well as the re-
maining therapy time for the active therapy section:
The preset therapy time for the current section is always displayed to the right, next to the therapy
section.
The remaining therapy time for the active therapy section is displayed on the screen <1> at top right.

To change the therapy time of the individual therapy sections, proceed as follows:
(1) On the data selector <2> click the timer symbol behind the respective therapy section.
(2) Rotate the data selector <2> and shorten the therapy time to the desired value.
(3) Press the data selector <2> to confirm the selected therapy time.

Important

The total therapy time is always the sum of the therapy times of the individual therapy sec-
tions.

You can generally prolong or shorten the therapy time in the TREATMENT menu for all ther-
apy sections. In case of the active therapy section (that therapy section, in which the instru-
ment currently is, with the ongoing therapy and intensity set greater than zero), the new ther-
apy time is included only for the current treatment, if it is shorter than the originally set ther-
apy time.

18 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

6.1.3 TREATMENT EXPERT Menu


In the TREATMENT EXPERT menu you can set the following therapy parameters:

Treatment duration Duration of the therapy section in question (can also be set in the TREATMENT
menu)
Oscillation Ratio between impulse time and pause time
Burst Burst frequency for the application of the impulses in packets
Mode Selection: Fixed frequency or frequency band
Frequency Treatment frequency for fixed frequency or lower and upper frequency for fre-
quency band
Sweep Attack time, sustain time and decay time (only for frequency band)

TREATMENT EXPERT menu

Important

In the current context, the non-selectable parameters are displayed in light grey.

Example for a Frequency Cycle

Assuming that you have selected the frequency band 5 ... 250 Hz and the three times 1s, 2s
and 3s are selected:

The frequency is initially at the maximum value (250 Hz) for one second. Thereafter the fre-
quency reduces to two seconds long, continuously up to 5 Hz. This frequency is kept for three
seconds. Finally, the frequency lengthens continuously, until after two seconds, the upper lim-

HIVAMAT 200 Evident 19


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

it frequency (250 Hz) is attained and the cycle begins anew.

Example for Using a Burst Frequency

The selection of a burst frequency emits impulses in packets. The burst value here corresponds
to the number of the compounded packets per second. For example, a burst of the value 1 is
set for a selected fixed frequency of 10 Hz (i.e. 10 impulses per second), the impulses are
emitted in the initial half of a second; no impulses are emitted in the second half. This se-
quence continuously repeats itself.

How to Set the Therapy Parameters

(1) Click on the symbol . The TREATMENT EXPERT menu is displayed.


(2) Click on the symbol abc, if you wish to rename the treatment program (see How to Rename a Treat-
ment Program on page 21).
(3) Rotate the data selector <2> by one more position. Select the desired therapy section on which the
settings are to be applied (1, 2 or 3).
(4) Rotate the data selector <2> by one more position. Now select the treatment duration for the re-
spective therapy section (0 ... 60 min).
(5) Proceed to the field Oscillation. Select here the modulation (possible values are 1:3, 1:2, 1:1, 2:1 or
3:1).
(6) Proceed to the field Burst. Select here the burst frequency (0 ... 10 Hz).
(7) Proceed to the field Mode. Select here the type of frequency (either fixed frequency Selective or fre-
quency band. Low |<< F>>| High ).
(8) Proceed to the field Frequency. Select here the upper and lower frequency for the frequency band
or the fixed frequency (5 ... 250 Hz).

Important

The HIVAMAT 200 Evident automatically ensures that the lower frequency is not great-
er than the upper frequency. The lower frequency must always be 5 Hz smaller than the
upper frequency.

(9) On the data selector <2> select the Sweep field (only for frequency band; not possible for fixed fre-
quency). Set here the time spans (respectively 1 ... 60 s).
(10) Repeat the steps discussed here if necessary, for the other therapy sections.
(11) Click on to confirm. Your settings are saved and you return to the TREATMENT menu.

6.1.4 Entering the Name of a Treatment Program


You can name the treatment programs which you have defined by a name of your choice.

The following characters are available to you for naming the treatment program:
Capital letters A ...Z
Numbers 0 ... 9
Special characters ., _, +, -

20 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

Space SPACE

Important

The treatment programs are always sorted alphabetically according to their names.

How to Rename a Treatment Program

(1) Click on the symbol abc.


(2) Rotate the data selector <2> up to the desired character (e.g. A) and press to confirm. The selected
character is added to the name displayed in the top line.

Menu abc
(3) Continue until the desired name is displayed in the top line.
(4) Click on the symbol . The new name is saved and you return to the respective initial menu.

Tip

You can use the symbol to delete the last character of the entered name.

6.1.5 Saving a Treatment Program


The HIVAMAT 200 Evident provides you with 150 memory locations for individual treatment programs.
The percentage share of the available memory space is displayed in the menu. The programs stored can be
called again through the INDIVIDUAL menu.

HIVAMAT 200 Evident 21


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

NOTICE

If no memory space is available, the message Attention: No more memory space avail-
able. is displayed. You can delete an existing treatment program (see How to Delete an Indi-
vidual Treatment Program on page 29).

How to Save a Treatment Program in the Instrument

(1) In the menu TREATMENT or TREATMENT EXPERT click the symbol . The message Treatment is
being updated is displayed and the treatment is saved, provided at least one memory location is
free.

Tip

You can save a treatment program stored on a therapy card from the THERAPY CARD menu
locally in the instrument (also refer to How to Store a Treatment Program from the Therapy
Card in the HIVAMAT 200 Evident on page 30).

NOTICE

If a treatment program with the same name has already been saved in the instrument, the
message The name is already being used by a treatment program. Overwrite existing
treatment program? is displayed. You can overwrite the existing program or rename the
treatment program to be saved (see How to Rename a Treatment Program on page 21).

6.1.6 Saving a Treatment Program on the Therapy Card


You can also save specific therapy forms on a therapy card with the HIVAMAT 200 Evident.

Tip

The therapy card can also be read with the HIVAMAT 200 Personal, making patient treat-
ment at home possible.

NOTICE

You can overwrite the parameters saved on the card. However, the original parameters can
no longer be retrieved.

How to Save a Treatment Program on a Therapy Card

(1) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(2) Now click on the symbol . If the card is not empty, the message Caution: There is already a
treatment program on the therapy card! Would you like to overwrite the treatment pro-

22 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

gram? is displayed.
(3) Click on . The message CAUTION: Treatment program is being written on the therapy card!
Please do not remove the card! is displayed and the individual treatment program is saved on the
therapy card.

Tip

You can save treatment programs from the INDIVIDUAL menu on a therapy card (see How
to Save an Individual Treatment Program on the Therapy Card on page 28).

6.2 TREATMENT EXAMPLES / TREATMENT PROGRAMS Menu


With HIVAMAT 200 Evident you can perform treatments according to preset programs which are stored
together with the corresponding therapy parameters. For each displayed program you can refer to the re-
commended therapy instructions. Each program features a sequence of different frequencies.

The TREATMENT EXAMPLES / TREATMENT PROGRAMS menu offers you information on how to per-
form treatment at the individual body regions, as well as access to the preset treatment programs.

The treatment examples consist of:


A list of body regions
Instruction text
Animations
Videos

Symbols in the TREATMENT EXAMPLES / TREATMENT PROGRAMS Menu


Back to MAIN MENU or to the list of body regions

Carrying out the therapy for the current therapy program (see How to Perform a Treat-
ment on page 18)

Browse down to the next step or browse up to the previous step

Invoke animation

Start animation

Pause animation

HIVAMAT 200 Evident 23


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

Stop animation and return to the beginning

How to Call a Treatment Program

(1) In the MAIN MENU select the entry Treatment Examples / Treatment Programs . The TREAT-
MENT EXAMPLES / TREAMENT PROGRAMS menu is displayed.
(2) In the TREATMENT EXAMPLES / TREATMENT PROGRAMS menu, click TREATMENT
PROGRAMS. A list of available treatment programs is shown.
(3) Select the desired program, e.g. Program 1. The parameters of the program are displayed.
(4) Insert the applicator of your choice into the connecting cable DEEP OSCILLATION.
(5) Click on the symbol . You come to the menu TREATMENT from where you can begin the therapy
(see How to Perform a Treatment on page 18).

How to Display a Treatment Example


Every treatment example is completed by further information in the form of text, animations and films:

(1) In the TREATMENT EXAMPLES / TREATMENT PROGRAMS menu, click TREATMENT EXAMPLES.
A list of available treatment examples is shown:

Treatment Examples
(2) Rotate the data selector <2> until you have selected the treatment example of interest (e.g. Lumbar
region / gluteal region / hip) and press. The treatment instructions for the selected treatment example
are displayed:

24 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

Treatment Example: Lumbar region, gluteal region, hip - Step 1/3


(3) Click the symbol 1/3 and rotate the data selector <2> one step to the right to proceed. On the
second page, an animation shows you the treatment area, the recommended treatment path and the
direction of movement for the applicator:

Treatment Example: Lumbar region, gluteal region, hip - Step 2/3


(4) On the last page, you can run a short video showing the course of the treatment and the use of the
hand applicator:

HIVAMAT 200 Evident 25


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

Treatment Example: Lumbar region, gluteal region, hip - Step 3/3

Important

If the applicator head has a green background in the video, the movement is to be per-
formed with pressure; if it has a white background, without pressure.

26 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

6.3 INDIVIDUAL Menu


In the menu INDIVIDUAL you will find individual treatment programs stored locally according to your in-
puts. You can call, delete or save them on the therapy card.

INDIVIDUAL menu

Symbols in the INDIVIDUAL Menu


Back to MAIN MENU

Call current treatment program in the menu TREATMENT (see How to Rename an Indi-
vidual Treatment Program on page 28)

Saving current treatment program on therapy card (see How to Save an Individual
Treatment Program on the Therapy Card on page 28)

NOTICE

It is possible to store on the card only if a therapy card has been inserted
with the chip side facing back. The color of the symbols changes from
white to blue.

Delete current treatment program (see How to Delete an Individual Treatment Program
on page 29)

HIVAMAT 200 Evident 27


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

6.3.1 Calling up an Individual Treatment Program


The INDIVIDUAL menu allows you to access the locally stored treatment programs in the instrument.

How to Rename an Individual Treatment Program

(1) In the MAIN MENU select the menu item Individual. The INDIVIDUAL menu is displayed.
(2) On the data selector <2>, select the treatment program which you wish to call.
(3) Select the menu item Treatment . You come to the menu TREATMENT and the program steps of
the respective treatment program are displayed.
The process is further described in section How to Perform a Treatment on page 18.
(4) Increase the intensity to the desired level with the intensity control <3>. The therapy time starts the
countdown.

6.3.2 Saving an Individual Treatment Program on the Therapy Card


You can save an individual treatment program from the menu INDIVIDUAL on a therapy card for the pa-
tient to use at home.

Tip

The therapy card can also be read with the HIVAMAT 200 Personal, making patient treat-
ment at home possible.

NOTICE

You can overwrite the parameters saved on the card. However, the original parameters can
no longer be retrieved.

How to Save an Individual Treatment Program on the Therapy Card

(1) In the MAIN MENU select the menu item Individual. The INDIVIDUAL menu is displayed.
(2) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(3) Now click on the symbol . If the card is not empty, the message Caution: There is already a
treatment program on the therapy card! Would you like to overwrite the treatment pro-
gram? is displayed.
(4) Click on . The message CAUTION: Treatment program is being written on the therapy card!
Please do not remove the card! is displayed and the individual treatment program is saved on the
therapy card.

Tip

You can save treatment programs from the TREATMENT menu on a therapy card (see How
to Save a Treatment Program on a Therapy Card on page 22).

28 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

6.3.3 Deleting an Individual Treatment Program


You can always delete individual treatment programs which you no longer need to free up memory space.

NOTICE

Deleted treatment programs cannot be restored.

How to Delete an Individual Treatment Program

(1) In the MAIN MENU select the menu item Individual. The INDIVIDUAL menu is displayed.
(2) On the data selector <2> select the treatment program which you wish to delete.
(3) Click on the symbol . The message Do you really want to delete the selected treatment?
menu is displayed.
(4) Click on the symbol to confirm. The treatment program is deleted and the memory space is free
again.

6.4 THERAPY CARD Menu


You can use the HIVAMAT 200 Evident to save an individual treatment program on a therapy card, con-
sisting of maximum three therapy sections.

The therapy card can also be read with the HIVAMAT 200 Personal, making patient treatment at home
possible.

Important

Therapy Cards programed for the HIVAMAT 200 Personal for the various fields of application
cannot be read with the HIVAMAT 200 Evident. A corresponding error message is shown if
you try.

The following messages are displayed in the THERAPY CARD menu:


No therapy card is inserted! You must first insert a therapy card.
There is no treatment program on the therapy card! A treatment program has not been saved on
the therapy card. Treatment must be saved from the TREATMENT or INDIVIDUAL menu (see How to
Save a Treatment Program on a Therapy Card on page 22 and How to Save an Individual Treatment
Program on the Therapy Card on page 28)
There is a treatment program on the therapy card! You can read the stored treatment program
and perform a treatment with the corresponding therapy parameters. In addition you can overwrite
the treatment program stored on the card from the menu TREATMENT or INDIVIDUAL.

Symbols in the THERPAY CARD Menu


Back to MAIN MENU

HIVAMAT 200 Evident 29


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

Starting the therapy that is saved on the therapy card (see How to Read a Therapy Card
on page 30)

Saving treatment program stored on the card locally in the instrument (see How to
Store a Treatment Program from the Therapy Card in the HIVAMAT 200 Evident on
page 30)

6.4.1 Reading a Therapy Card


If you read the parameters stored on a therapy card, you can perform a treatment with the patient-specific
settings.

How to Read a Therapy Card

(1) In the MAIN MENU select the Therapy card menu. The THERAPY CARD menu is displayed.
(2) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(3) Now click on the symbol to go to the TREATMENT menu (see How to Perform a Treatment on
page 18).


6.4.2 Storing a Treatment Program from the Therapy Card in the HIVAMAT 200
Evident
You can also store a treatment program that was previously saved on a therapy card as an individual treat-
ment program in the HIVAMAT 200 Evident.

How to Store a Treatment Program from the Therapy Card in the HIVAMAT 200 Evident

(1) In the MAIN MENU select the Therapy card menu. The THERAPY CARD menu is displayed.
(2) Insert a therapy card with the chip side facing back in the card reader <4>. The card symbol
changes from white to blue to indicate that the card is readable and writable.
(3) Click on the symbol . The treatment is displayed, if at least one memory location is still free.

NOTICE

If a treatment program with the same name has already been saved in the instrument, the
message The name is already being used by a treatment program. Overwrite existing
treatment program? is displayed. You can overwrite the existing program or rename the
treatment program to be saved (see How to Rename a Treatment Program on page 21).

30 HIVAMAT 200 Evident


6 Overview of the Menus PHYSIOMED ELEKTROMEDIZIN AG

6.5 SETTINGS Menu


In the menu SETTINGS you can set the following instrument settings:
Symbol Meaning

Brightness of the screen <1>

Volume of the sound signal

Pitch of the sound signal

Time after which the instrument will go into the power saving mode (1 ... 60 min or off)

Dialog language

Start menu (possible values: Main menu, Individual, Treatment Examples / Treatment
Programs, Last treatment, Standard treatment, Therapy card)

You can change the settings this way

(1) In the MAIN MENU select the menu item Settings. The SETTINGS menu is displayed.

SETTINGS menu

HIVAMAT 200 Evident 31


PHYSIOMED ELEKTROMEDIZIN AG 6 Overview of the Menus

(2) Rotate the data selector <2> to select the parameters you wish to change (e.g. Brightness) and
press to confirm.
(3) Rotate the data selector <2> until the parameter assumes the desired value and press it to confirm.
(4) Change all the desired parameters and click on the symbol to include the settings.
The new settings are effective immediately.

6.6 HELP Menu


Through the menu HELP you can find important information about the instrument, which is of interest
primarily for customer service.

How to Call the HELP Menu

(1) In the MAIN MENU select the menu item Help. The HELP menu is displayed.

HELP menu
(2) Click the arrow symbol to browse through the help menu.

32 HIVAMAT 200 Evident


Reference Information PHYSIOMED ELEKTROMEDIZIN AG

Appendix A Reference Information

A.1 Service, Repairs, Maintenance


The manufacturer guarantees the safety of the instrument only in its original state. The instrument must
be operated in accordance with the Operating Instructions.

Repairs to the instrument may only be performed by parties duly authorized by PHYSIOMED
ELEKTROMEDIZIN AG. Any repairs performed by an authorized agent must be accompanied by written
certification, describing the nature and extent of the repairs undertaken, as applicable with details regard-
ing changes to nominal operating values or the operational range. The certification must also contain the
date performed, the name of the repair company and the signature of the service person. When defective,
components affecting the safe operation of the instrument must be replaced by manufacturers original
parts. Upon request, wiring diagrams, parts lists and service instructions can be made available to qualified
technical personnel employed by the customer.

We recommend having the instrument, including all accessories, serviced at regular intervals.

A.2 Cleaning and Disinfection

NOTICE

The oscillator plate must only be disinfected using an agent, but not be sterilized using heat.

All parts of the oscillator head must be absolutely dry before using them next time.

Clean the device and the accessories regularly with an aldehyde-based disinfectant. Switch off the instru-
ment by any means before cleaning.

Clean the instrument and the handle by using a soft foam cloth. Use caution so that no liquid substances
penetrate into the instrument.

Use a new or disinfected oscillator head on the hand applicator for every patient. Clean, disinfect or steril-
ise the hand applicator regularly. The clamping ring, oscillator plate and handle are resistant against disin-
fecting agents and can be sterilised up to 275F (135C).

A.3 Limited Warranty


Two-year limited warranty covers parts and labour on HIVAMAT 200 Evident.

A.4 Disposal
After the service life of the instrument, dispose of it in conformance with the applicable regulations for en-
vironment protection.

Handle, clean, and dispose of components and accessories that have come in contact with bodily fluids ac-

HIVAMAT 200 Evident 33


PHYSIOMED ELEKTROMEDIZIN AG Reference Information

cording to national, local and facility rules, regulations and procedures.

Environment Protection Symbol

A.5 Training
Training is given in the course of the introduction to the instrument.

Should the need arise however, continued training can be provided by a medical device consultant. Please
contact PHYSIOMED ELEKTROMEDIZIN AG.

A.6 UL 60601-1-2 Electromagnetic Compatibility Declaration


Medical electrical devices are subject to particular precautions regarding electromagnetic compatibility and
must be used in accordance with the instructions on electromagnetic compatibility contained in the ac-
companying documents.

Portable and mobile HF communication devices can affect medical electrical devices (see the supplement
on electromagnetic compatibility, technical description).

A.7 Technical Data


Protection class 1, Type BF (according to IEC 60601-1)
Mains connection 110 ... 240V
Power frequency 50 ... 60 Hz
Power consumption 35 VA
Current consumption 0.15 A (at 230 V) or 0.3 A (at 115 V)
Output voltage 400Vpeak
Load impedance 10 MOhm
Outlet frequency 5 ... 250 Hz
Impulse-pause ratio 3:1, 2:1, 1:1, 1:2, 1:3
Curve form Rectangular impulse, dual phase
Dimensions (W x H x D) 10.2 x 13.8 x 14.6 inch (26.0 x 35.0 x 37.0 cm; screen set ver-
tical)
Weight 18.7 lb (8.5 kg)

34 HIVAMAT 200 Evident


Reference Information PHYSIOMED ELEKTROMEDIZIN AG

A.7.1 Environmental Conditions


Operation
Temperature 50F ... + 104F (+10C ... +40C)
Relative humidity 30% ... 75%
Atmospheric pressure 23.6 ... 31.3 inHg (800 ... 1060 hPa)

Storage and Transport


Temperature 14F ... 122F (-10C ... +50C)
Relative humidity 10% ... 90%
Atmospheric pressure 14.8 ... 31.3 inHg (500 ... 1060 hPa)

A.8 Manufacturer
PHYSIOMED ELEKTROMEDIZIN AG
Hutweide 10
91220 Schnaittach
Germany

A.9 Customer Service


When you have questions concerning HIVAMAT 200 Evident please refer to the following address or call
the phone number below:
Naimco, Inc. Dba Richmar
4120 South Creek Road
Chattanooga TN 37406
USA
PHONE: 423-7604194
FAX: 423-667-2325
E-MAIL:jeremybarnes@richmarweb.com
WEB: www.richmarweb.com

HIVAMAT 200 Evident 35


PHYSIOMED ELEKTROMEDIZIN AG Scope of Supply and Accessories

Appendix B Scope of Supply and Accessories


CAUTION

The instrument is to be used exclusively with original accessories. Otherwise, treatment results
might be insufficient.

B.1 Scope of Delivery


HIVAMAT 200 Evident is supplied with the following accessories:

Item No. Name Quantity


00379 Applicator handhold 2
00261 Connection cable DEEP OSCILLATION 2

01100 Deep oscillation instrument HIVAMAT 200 Evident 1
00778 Mains cable 1
00391 Membrane set, 5.0 cm 5
00392 Membrane set, 9.5 cm 5
00381 Oscillator head, 5.0 cm 2
00386 Oscillator head, 9.5 cm 1
00260 Patient lead DEEP OSCILLATION 1
00348 Powder 1
00229 Therapy card PHYSIOMED DEEP OSCILLATION 3
00382 Titanium neutral element 1
01004 Trolley Evident 1
01625 Operating Instructions (English, USA) 1

B.2 Available Accessories


The following accessories are available for HIVAMAT 200 Evident:

Item No. Name


00379 Applicator handhold
00261 Connection cable DEEP OSCILLATION
00370 Hand applicator set for DEEP OSCILLATION
00778 Mains cable
00393 Membrane set, 1.5 cm
00391 Membrane set, 5.0 cm
00392 Membrane set, 9.5 cm

36 HIVAMAT 200 Evident


Scope of Supply and Accessories PHYSIOMED ELEKTROMEDIZIN AG

Item No. Name


00387 Oscillator head, 1.5 cm
00381 Oscillator head, 5.0 cm
00386 Oscillator head, 9.5 cm
00260 Patient lead DEEP OSCILLATION
00396 Pin applicator
00348 Powder
00229 Therapy card PHYSIOMED DEEP OSCILLATION
00382 Titanium neutral element
01004 Trolley Evident

01091 Upgrade package 2 for HIVAMAT 200 Evident, includes:

2 x 01099 HIVAMAT 200 Personal with standard accessories,

6 x 00229 Therapy cards PHYSIOMED DEEP OSCILLATION.

ATTENTION: This upgrade package is only available when purchased together with at least
one HIVAMAT 200 Evident.

HIVAMAT 200 Evident 37


PHYSIOMED ELEKTROMEDIZIN AG Index

function test 11
Index
H
A
hand applicator 13 , 16
accessories 36 help 32
available 36 HIVAMAT 200 Personal 29
supplied 36 hypersensitivity
ambient temperature 34 to electrostatic fields 2
application 2
I
B
implants
blood circulation electronic 2
inscrease 2 important user information 1
brightness 31 indicators 8
function 8
C individual program
delete 29
cable check 13 invocation 28
card reader 9 infections
cardiac arrhythmia 2 acute 2
cardiac insufficiency 2 inflammation
cardiac oedemata acute 2
decompensated 2 instrument description 7
cardiac pacemaker 2 instrument errors 12
caution symbols 4 instrument overview 7
cleaning 33 instrument symbols 4
compatibility intensity control 9
electromagnetic 34 invocation
connecting socket 10 individual program 28
contraindications 2
controls 8 L
function 8
curve form 34 language 31
limited warranty 33
D load impedance 34

data selector 9 M
definitions 4
delete maintenance 33
individual program 29 malignant diseases 2
dimensions 34 membrane
disinfection 33 replace 13
disposal 33 modulation 34
muscle aches 2
E muscle relaxation 2

electrostatic field 2 N
EMC 34
environment protection 33 name
error codes 12 entering 20
erysipelas 2 notes
general 5
F
O
fields
electrostatic 2 operating principle 2

38 HIVAMAT 200 Evident


Index PHYSIOMED ELEKTROMEDIZIN AG

outlet frequency 34
output voltage 34 V

P venous diseases
acute 2
pain reduction 2 volume 31
pitch 31
power module 10 W
power saving mode 12 , 31
precaution 3 warnings 2
pregnancy 2 warranty 33
protection class 34 weight 34
write
R therapy card 30

read
therapy card 30
repairs 33

save
treatment program 21
screen 8
brightness 31
service 33
settings 31
skin diseases
infectious 2
standby mode 12
start menu 31
starting up 11

technical data 34
therapeutic fields 7
therapy card 29
read 30
write 30
therapy parameters
setting 19
therapy time
setting 18
thromboses
untreated 2
training 34
treatment 16
treatment examples 23
TREATMENT EXAMPLES / TREATMENT PROGRAMS
menu 23
treatment program
save 21
treatment programs 23
individual 27
tuberculosis
active 2

HIVAMAT 200 Evident 39