PREVENTIVE

MAINTENANCE
For the analyser to work at optimum level, we recommend
you perform regular preventive maintenance and checks
2. Cleaning the optical filters
on certain parts of the instrument.
The following table indicates the different parts to be Remove the filters from their casing with the appliance
checked and the corresponding frequency. turned off. If you use the service programme option to
change the filter, the integration times will be reset.
To clean the filters, remove the filter from its casing using
Task Annual Bi-annual
a pair of pliers and blow on both sides of the filter with a
Check the rotor centring bolt
1 x dry air syringe or spray. Replace the filter. Repeat this
2 Clean the filters x
operation with each filter.
3 Clean the optical slit x
Check the operation of the
3 x
fan
Change the seal and re-
4 x
grease the ceramic pump
Perfom the service
5 x
programme tests

1. Check the torque on the


rotor centring bolt
Check that the bolt holding the rotor centring unit is as
tight as possible. Use a 2.5 mm Allen wrench and fasten
the bolt as tightly as possible (do this only on analysers
with a serial number below 831011030).

Photograph n 2

Photograph n 1 Photograph n 3

Preventive maintenance 1 TEUS00028-01-ENG


3. Cleaning the optical slit
Remove the photodiode plate to access the optical slit
hole.
Use a dry air syringe or spray to blow inside the whole
and clean any dust from the interior. See photographs 4
and 5.

Photograph n 6

Photograph n 4

Photograph n 7

5. Changing the seal and re-


greasing the ceramic pump
For the ceramic pump to be in correct working condition
and not cause excessive noise, perform a maintenance
procedure every two years or every 20,000 million pump
Photograph n 5 operation steps, whichever comes first. This maintenance
consists of changing the seal and re-greasing the
transmission system.
4. Check the functioning of The number of pump cycles can be checked with the service
programme on the Utilities/Read and load Adjustments
the fans and Cycles menu. Select the Cycles tab and check the
step value of the Dispensing pump. See Screen 1.
Check that all the fans work correctly. Place your hand
To change the AC11527 Dispensing pump seal:
behind each fan and check that it expels air.
a) Remove the dispensing pump from the instrument.
The instrument has four fans:
b) Remove the fluidic chamber.
Two at the rear of the instrument.
c) Remove the seal support.
One at the bottom of the optical unit; the air is expelled
d) Replace the seal.
under the instrument.
e) Fit the seal support by tightening the 4 bolts gradually.
One on the needle. To check this fan, remove the arm
f) Refit the fluidic chamber on the pump.
casing and feel to check that air flows under the needle
unit. The fan should take in the air. If it expels it, remove
the fan and turn it so that it takes air in instead of
expelling it, as indicated by the arrows in photograph 6.

Preventive maintenance 2 TEUS00028-01-ENG


6. Performing the service
programme tests
Before performing the basic tests to check the analyser,
the following is necessary:
Update the firmware and service version. Read the
attached Quick Guide document included on the CD-
ROM, which explains the changes made to the version.
Make the new adjustments as indicated in that
document.
Clean the exterior of the needle with 70 alcohol,
together with the L-plate in the needle adjustment hole.
Remove the needle and apply the plunger.
Check the racords of tightens of the ceramic pump
Screen n 1 and the three-way valve and on the spiral pipe
connector.
To re-grease the transmission system: Perform 15 cycles of prime with air, 15 with wash
a) Remove the motor together with the transmission solution and 15 with system liquid from the utilities
system and the ceramic piston from the pump body. menu.
b) Separate the transmission from the ceramic piston. Check that there are no air bubbles in the tubes after
c) Remove the old grease from the transmission system the loading cycles.
using absorbent paper. Check that the lamp-holder is not cracked or broken. If
d) With a syringe, inject the new grease inside the unit you see that it has deteriorated, replace it. Check that
using the AC13079 Refill ceramic pump spare. the lamp voltage is between 11.5 and 12 V on the lamp
Remove the old grease that comes out of the holes on terminals.
the unit. Clean the inside of the rotor cover hole with alcohol.
e) Insert the transmission into the unit. Remove the excess Remove the wash station and clean the inside supply
grease that comes out of the holes. source with alcohol.
f) Fit the transmission unit with the ceramic piston in the Adjust the manipulator arm, the rotor position and the
pump body. filter wheel.
g) Refit the ceramic pump on the instrument. Perform the photometry tests.
h) Reset the dispensing pump step value to zero with the Perform the rotor temperature test.
service programme. Select the Cycles tab on the
Utilities/Read and load Adjustments and Cycles menu,
select the Dispensing pump checkbox and load a zero
into the number of steps. Remember that to modify
the number of cycles, you must have started the service
programme as a technical service operator.

WARNING
When it must change to a cpu board (PC11551,
Microprocessor Board), by failure of some
component of the board, not to lose the information of
the adjustment and number of cycles of each element,
proceed with the following way:
Switch on the analyser with the failed board. Save to
a file the adjustments values and number of cycles.
Use the service programme with the Utilities/Read and
load Adjustments and Cycles menu, push the button
read adjustments and automatically will request if
it is wanted to save in a file. Write a file name and
push accept button.
Change the fail board for a new one
Initialise the analyser and go to the previous menu of
service programme. Push the button Load
adjustment. The programme will request a file name.
Select the previous file name and accept. This way
the values of adjustments and the number of cycles
have been loaded.

Preventive maintenance 3 TEUS00028-01-ENG