MAINTENANCE
For the analyser to work at optimum level, we recommend
you perform regular preventive maintenance and checks
2. Cleaning the optical filters
on certain parts of the instrument.
The following table indicates the different parts to be Remove the filters from their casing with the appliance
checked and the corresponding frequency. turned off. If you use the service programme option to
change the filter, the integration times will be reset.
To clean the filters, remove the filter from its casing using
Task Annual Bi-annual
a pair of pliers and blow on both sides of the filter with a
Check the rotor centring bolt
1 x dry air syringe or spray. Replace the filter. Repeat this
2 Clean the filters x
operation with each filter.
3 Clean the optical slit x
Check the operation of the
3 x
fan
Change the seal and re-
4 x
grease the ceramic pump
Perfom the service
5 x
programme tests
Photograph n 2
Photograph n 1 Photograph n 3
Photograph n 6
Photograph n 4
Photograph n 7
WARNING
When it must change to a cpu board (PC11551,
Microprocessor Board), by failure of some
component of the board, not to lose the information of
the adjustment and number of cycles of each element,
proceed with the following way:
Switch on the analyser with the failed board. Save to
a file the adjustments values and number of cycles.
Use the service programme with the Utilities/Read and
load Adjustments and Cycles menu, push the button
read adjustments and automatically will request if
it is wanted to save in a file. Write a file name and
push accept button.
Change the fail board for a new one
Initialise the analyser and go to the previous menu of
service programme. Push the button Load
adjustment. The programme will request a file name.
Select the previous file name and accept. This way
the values of adjustments and the number of cycles
have been loaded.