Electrical Safety
Electrical safety is a topic of paramount importance in medical device quality assurance best practices. Elec-
trical shock can cause disruptions during health care procedures, and result in injury or death. Physiological
effects range from a tingling sensation to serious burns and electrocution. Excitable human tissue is very
sensitive to current in the frequency range of electrical power systems worldwide (50-60 Hertz). Figure 6-1
shows the effects of current flowing from one skin contact point to another. The term macroshock describes
electrical current applied externally.
Burns
Cardiac fibrillation
Let-go current
Sensitivity limit
Electrical safety takes on added significance in electrically-susceptible patients. For cardiac procedures,
electrically-conductive catheters may be placed into the heart while the patient is connected to medi-
cal equipment. Skin exhibits a high electrical resistance, but internal body components such as blood and
muscle have a low electrical resistance. Currents as low as 20 microamps have caused ventricular fibrillation
in experiments conducted with dogs when a conductor made direct contact to the heart. Microshock is the
term used to describe direct shocks to the cardiac muscle.
As a result of data collected regarding macroshock and microshock, worldwide standards have been estab-
lished to limit leakage current. In equipment designed for low resistance direct contact with patients, (such
as indwelling catheters) electrical isolation techniques are used to reduce the current flowing to the patient
to minimal levels even at line-voltage levels. In the event of a device failure or short-circuit condition, the
patient is protected from microshock. These techniques may utilize isolation transformers and optical cir-
cuits. Thus, electrical safety standards specify low microampere limits for direct patient contact equipment.
To reduce leakage current to negligible levels, chassis grounding is utilized to shunt any leakage or fault
230 V
50 Hz
Fuses 230 V
10
~22 A
~11 V
10,000
0.015 F
Milli-
Input 1,000 voltmeter
Leakage measurement (IEC 60601.1, Section 3, Clause 19) limits have been developed for equipment types
and measurements. NC is normal conditions and SFC is single fault conditions. Some of the measurements are
only applicable to manufacturer design testing.
Patient
leakage
Patient Patient current Patient Patient Patient
Earth Touch leakage leakage mains on auxiliary auxiliary auxiliary
Leakage leakage current current AC current DC applied current current current
current (A) current mA (A) (A) (A) (A) (A) (A) (A)
Type B NC 5 100 100 10 100 10 100
SFC 10 500 500 50 500 50 500
Type BF NC 5 100 100 10 100 10 100
SFC 10 500 500 50 5000 500 50 500
Type CF NC 5 100 10 10 10 10 10
SFC 10 500 50 50 50 50 50 50
Additional important points regarding IEC 60601.1 include: the use of up to 25 amperes ac for protective
earth testing, leakage current is measured at 100 percent of mains voltage, and performance of dielectric
strength/insulation testing is measured at 110 percent of mains voltage. A new IEC standard, IEC 62353, is
used for medical device testing in hospitals. IEC 62353 was developed because IEC 60601.1 is a type-testing
standard with no risk management criteria and is impractical for testing in the hospital environment.
Start
Equipment information
Control number:_________________________________ Hospital:_________________________________________
Manufacturer:___________________________________ Model:___________________________________________
Serial number:__________________________________ Location:_________________________________________
Test information
Technician:_____________________________________ Date:____________________________________________
Test type: Incoming_________ Post repair_________
Test equipment needed: Fluke Biomedical ESA615 or equivalent
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Fuse rating is appropriate
Power cord, accessory cables, charger, patient cables, connectors
Integrity of mechanical parts
Electrical safety Criteria IEC 62353
Ground wire resistance/protective earth 0.3 W
Current in A Applied part
Type Type Type
B BF CF
Equipment leakage alternative method
For accessible conductive parts of Class I equipment con-
1000 1000 1000
nected or not connected to the protective earth conductor
For Class II ME equipment 500 500 500
Equipment leakage direct or different method
Equipment leakage current for accessible conductive parts
of Class I ME equipment connected or not connected to the 500 500 500
protective earth conductor
Equipment Leakage current for Class II ME equipment (NC) 100 100 100
Applied part leakage current alternative method (ac)
Applied part leakage current of applied part < 5000 < 50
Applied part leakage current direct method (ac)
Total patient leakage current mains voltage on applied part 5000 100
Insulation test (optional) 500 V dc applied < 2 MW
APPLIED
PART
DUT_L2
MD
DUT_PE
FE
CONDUCTIVE PART
Figure 6-9. Mains to protective-earth
insulation resistance test schematic.
APPLIED
PART
DUT_L2
DUT_PE
FE
MD CONDUCTIVE PART
APPLIED
PART
DUT_L2
DUT_PE
FE
CONDUCTIVE PART
MD
TEST LEAD
Figure 6-11. Ground wire (protective
earth) resistance measurement
schematic.
APPLIED
PART
DUT_L2
MD
DUT_PE
FE
CONDUCTIVE PART
TEST
LEAD Figure 6-12. Mains to non-earth
accessible conductive points
schematic.
APPLIED
PART
DUT_L2
MD
DUT_PE
FE
CONDUCTIVE PART
APPLIED
PART
DUT_L2
DUT_PE
FE
CONDUCTIVE PART
TEST
LEAD
MD
L1
APPLIED
PART
MAINS
DUT_L2
L2
L2
REV
POL
DUT_PE
MD FE
CONDUCTIVE PART
APPLIED PART
PE
Figure 6-15. Earth leakage current
test schematic.
L1
APPLIED
PART
MAINS
DUT_L2
L2
L2
EARTH
DUT_PE
FE
CONDUCTIVE PART
MD TEST LEAD
PE APPLIED PART
Figure 6-16. Enclosure leakage
current test schematic.
L1
APPLIED
PART
MAINS
DUT_L2
L2
REV
POL
DUT_PE
FE
CONDUCTIVE PART
TEST
LEAD
PE
MD
MAP (ISOLATION)
TRANSFORMER LEAD
SELECT
RELAY*
MAP
REV
*Leads not selected are open. Figure 6-17. Lead isolation (mains
on applied parts) leakage test
schematic.
CONDUCTIVE PART
MD
LEAD
PE SELECT
RELAY*
LEAD GND
PE SELECT RELAY
(REMOTE ONLY)
*Leads not selected are open.
Figure 6-18. Lead-to-ground
(patient) leakage current test
schematic.