You are on page 1of 13

INSTALLATION PROCEDURE

8810 - ORP

CALCIUM HARDNESS

0 - 500 mg/l CaCO3


(0 - 50 TH)

May 20, 2010 621=997=106


Revision: J
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

INDEX

1 APPLICATION: WATER FOR BOILERS AND REFRIGERATION CIRCUITS .......... 3


1.1 BASIC CONFIGURATION ....................................................................................................................... 3
1.2 MEASURING PRINCIPLE ....................................................................................................................... 3
1.3 CALIBRATION PRINCIPLE .................................................................................................................... 3

2 PROGRAMMING ........................................................................................................ 4
2.1 ACCESS TO PROGRAMMING: .............................................................................................................. 4
2.2 PM XXXX MENU ...................................................................................................................................... 4
2.3 TITRATION PARAMETERS: ................................................................................................................... 5
2.4 CYCLE PARAMETERS: .......................................................................................................................... 5
2.5 SAMPLE CONDITIONING:...................................................................................................................... 5
2.6 CHEMICAL CLEANING: ......................................................................................................................... 5
2.7 GENERAL TEST ...................................................................................................................................... 5
2.8 SEQUENCE ............................................................................................................................................. 6
2.9 ANALOGUE OUTPUTS........................................................................................................................... 7
2.10 PROGRAMMING AUTOMATIC CALIBRATION..................................................................................... 7

3 CALIBRATION............................................................................................................ 7
3.1 PRIMARY CALIBRATION ....................................................................................................................... 7
3.2 STARTING THE ANALYZER .................................................................................................................. 8
3.3 REGULAR CALIBRATION ...................................................................................................................... 8
3.4 READING THE CALIBRATION PARAMETERS .................................................................................... 8

4 PREPARING THE REAGENTS .................................................................................. 8


4.1 R1: TITRATION REAGENT (FOR 10 LITRES) ....................................................................................... 9
4.2 R2: CONDITIONING REAGENT (FOR 10 LITRES): .............................................................................. 9
4.3 R4: SOLUTIONS FOR AUTOMATIC CALIBRATION (10 LITRES): ...................................................... 9
4.4 SOLUTIONS FOR MANUAL CALIBRATION ....................................................................................... 10

5 PREVENTIVE MAINTENANCE ................................................................................ 11


5.1 DETERMINING THE END OF TITRATION ........................................................................................... 11
5.2 CHECKPOINTS ..................................................................................................................................... 11

6 ALL ABOUT THE ELECTRODES ............................................................................ 12


6.1 STORAGE, RECONDITIONING OF THE ELECTRODES .................................................................... 12
6.2 DIFFERENT ELEMENTS IN THE MEASURING CELL ........................................................................ 13

Page 2 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

INSTALLATION PROCEDURE - 8810 ORP

CALCIUM HARDNESS: 0 - 500 mg/l CaCO3 (0 - 50 TH)

1 APPLICATION: WATER FOR BOILERS AND REFRIGERATION CIRCUITS


1.1 BASIC CONFIGURATION

Range 0 500 mg/l (ppm) CaCO3


Lowest detection limit 10 mg/l (ppm) CaCO3
Working electrode Calcium selective electrode, Reference: 125=000=002
Reference electrode Type 8483 with Ag/AgCl, Gel electrolyte, Reference: 368483, 00000
Software version ORP N 6.13 or higher
Necessary equipment Conditioning pump (Several reference numbers possible), chemical cleaning
(368810, 56000).
Options Automatic calibration (368810, 60000), External calibration measure (368810,
65000), Thermostatting (368810, 76000)

1.2 MEASURING PRINCIPLE 1.3.1 Non-automatic calibration


We recommend manual calibration be carried out
The measurement cell of the Analyzer is
not more than once a week, but at least once a
automatically filled with a constant and
month.
reproducible sample volume. After conditioning
with a triethanolamine buffer, calcium hardness is
titrated directly by an EGTA solution. Two possibilities:

The titration is followed by a calcium selective


electrode, of which the raw signal (mV) as well as a. MANUAL = Manual introduction of two
the derivative (mV/s) allow for the automatic laboratory prepared calibration solutions in the
determination of the maximum inflection of the measuring cell. The analyzer will determine two
curve (so-called AutoEndPoint). In this way the calibration parameters, slope and asymmetry.
initial values of the potentials may vary by 10-50
mV in the course of the year, while the precision b. PROCESS = The analyzer carries out a double
remains within the specifications. analysis cycle using rinse (sprinkler) water as its
It is mandatory to carry out a 2-point calibration, first measuring point and then the process
the first one to be carried out on softened water, sample as its second point. Then take the
the other on a solution of known hardness. process sample and determine the nominal
concentration following the laboratory method.
The analyzer will determine two calibration
1.3 CALIBRATION PRINCIPLE parameters, slope and asymmetry.
On the first day of installation, a so-called
PRIMARY calibration is carried out, which serves As a consequence, to use this calibration method,
as a basis for all other calibrations. As a principle, the rinse water must be exempt of hardness
the PRIMARY CALIBRATION must never be (softened or demineralised water) and the sample
repeated (except in case of a loss of the whole must contain at least 200 mg/l CaCO3 during
programming due to a software update). The user sampling.
has access to the so-called SYSTEM calibration,
which he will be required to carry out periodically.

Page 3 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

1.3.2 Automatic calibration


2.2.2 Options
The Analyzer carries out a double analysis cycle With the arrow keys, select "OPTION" and
using rinse (sprinkler) water as its first measuring press ENTER.
point, and a CALIBRATION solution, introduced
automatically by means of a volumetric pump, as Using the same method, enter or check the
its second point. The Analyzer will determine two following information:
calibration parameters, slope and asymmetry.
As a consequence, to use this calibration method, AUTO SYS CAL : NO / YES only if the equipment
the rinse water must be exempt of hardness for the automatic calibration option is available.
(softened or demineralised water). The (press ENTER)
concentration of the CALIBRATION solution must
be near to 80% of the site range (e.g. 400 mg/l for CLEANING : YES (press ENTER)
a maximum 500 mg/l)
SAMPLE COND. : YES (press ENTER)
CAL 2 PTS : YES (press ENTER)
2 PROGRAMMING GRAB SAMPLE : NO / YES only if the equipment
of the External sample measure is available
(press ENTER)
2.1 ACCESS TO PROGRAMMING:
SAMPLE TIME : XXX usually 010 ~
Press the red button on the front panel. depends on the flow of the volumetric
Press SELECT to obtain the main menu. pump ~ (press ENTER)

Press the arrow keys to select ZINC : NO (press ENTER)


"PROGRAMMING". HEATING : NO / YES only if the equipment for
the Thermostatting option is available.
Press ENTER and keep it pressed for 5 (press ENTER)
seconds to enter the "PROGRAMMING" menu.
Tc : XXX.X C ~ the highest possible
sample temperature ~ (press ENTER)
NITROGEN : NO (press ENTER)
2.2 PM XXXX MENU
DILUTION : NO (press ENTER)
With the arrow keys, select "PM XXXX".
Press ENTER.
With the arrow keys, enter 4000. 2.2.3 Setting the analogue outputs:

Press SELECT to change the figures.


First of all, connect a multimeter in series to the
With the arrow keys, enter 4600. analogue output and select ADJUST I out .
Press SELECT to change the figures. With the arrow keys, select the range of
analogue output :
With the arrow keys, enter 4670.
OUT: (0 or 4) mA...20 mA (press ENTER)
NOTE With the arrow keys, set the simulation of the
loop current at the beginning of the scale :
If the access code is wrong, the Appliance
will return to the main menu. OUT: (0 or 4) mA XXX (press ENTER)
With the arrow keys, set the simulation of the
loop current at the end of the scale :

2.2.1 Application OUT : 20 mA XXX (press ENTER)

Press ENTER to select "APPLICATION".


NOTE
Press ENTER to select CAL TEMPERATURE AND ADJUST
"MEASURE DEC. POT. 1PT" and press ADC" are POLYMETRON reserved menus.
ENTER.
For this application, the position of the S1
to S5 jumpers should be : 0 . 0 . 0 . 1 . 0

Page 4 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

2.2.4 Units: 2.5 SAMPLE CONDITIONING:


With the arrow keys, select "UNIT" and press With the arrow keys, select "SAMPLE
ENTER. COND.", and press ENTER
With the arrow keys scroll through the pre- Enter or check the following information :
programmed units (..., mg/l, degF, ...) or
compose your own measuring unit (USER /
REAGENT 1 : YES (press ENTER)
Press ENTER / XXXXX) and press ENTER.
INJ. REAG. 1 : 120 s ~ depends on the initial pH
Exit PM XXXX by pressing SELECT. condition of the sample. See chapter 4.2 ~
(press ENTER)
TIME REAG. 1 : 030 s (press ENTER)
2.3 TITRATION PARAMETERS: REAGENT 2 : NO (press ENTER)

With the arrow up key, select


"MEASUREMENT" and press ENTER.
Enter or check the following information : 2.6 CHEMICAL CLEANING:
AUTO END PT (press ENTER) With the Arrow down key, select "CLEANING"
DEVIATION : 015 s is recommended (press and press ENTER.
ENTER) Enter or check the following information :
END PT 1 : - 0045 mV is recommended but may FREQUENCY : XXX usually 003 ~ depends on
vary according to the application (press ENTER) the rate of fouling ~ (press ENTER)
NOTE SPRINKLER t. : XX s usually 010 ~ depends on
The configuration of END PT may vary in the rinse water flow rate ~ (press ENTER)
the course of time. We recommend REAG. INJ. t. : XXX s usually 003 ~ depends on
normalizing the exact determination of the rate of fouling ~ (press ENTER)
this value preferably once a month. (see
the procedure described in the chapter RESIDENC t. : XXX usually 030 ~ depends on
Error! Reference source not found. ) the rate of fouling ~ (press ENTER)

2.4 CYCLE PARAMETERS:


2.7 GENERAL TEST
With the arrow keys, select "TIMING" and
press ENTER. With the Arrow down key, select "SERVICE"
and press ENTER
Enter or check the following information :
SAMPLE TIME : XXX s usually 030 ~depends on With the arrow keys, start and stop the different
the sample flow rate ~ (press ENTER) functions to test them.

SPRINKLER N : X usually 3 ~ depends on the You may pass from one function to the next by
rate of fouling ~ (press ENTER) pressing ENTER.
SPRINKLER t : XX s usually 10 ~ depends on NOTE
the rinse water flow rate ~ (press ENTER)
While testing pumps 1 and 2 let the pumps
INJ. TIME : 01 s (press ENTER) function until the liquid arrives in the cell.
DELAY TIME : 01 s (press ENTER) 2.7.1 Sample volume in the cell
CYCLE TIME : XXX mn usually 015 ~ depends
on the required analysis frequency ~ To refine the sample-taking in this application,
(press ENTER) reduce the siphon entrance diameter with the aid
of a ring (length 2 cm) cut from a 6 x 8 mm tube,
LIM. TIT. t : < or = to CYCLE TIME (press inserted into the siphon inlet.
ENTER)

Page 5 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

Using a graduated cylinder, adjust the siphon 2.7.4 Connecting the electrodes:
height in the cell to ensure a minimum volume of
75 ml 5 ml. Place the electrodes in the cell according to the
Exact measurement Empty the reactor diagram in chapter 6.2
(DRAIN), fill it with rinse water (SPRINKLER)
until it overflows through the siphon, weigh the Connect the working electrode to the cable
empty cylinder and place it under the drain valve with the biggest diameter.
connector, empty the reactor (DRAIN) and weigh
the full cylinder. The difference between the two Connect the reference electrode to the cable
weights (in grams) corresponds to the volume with the smallest diameter.
captured in the cell (in ml). It is mandatory to orient the injectors in such a
Repeat this operation 5 to 10 times and check way that the solutions are injected above the
that the reproducibility of the sample taking is sample presence detector. Otherwise the
1%. values will be scattered.

2.7.2 Flow rate of the volumetric pump:


This doesnt require calibration, as its flow rate is 2.8 SEQUENCE
never precisely known. However, check the time With the arrow down key, select "SEQUENCE"
necessary for the pump to fill the cell and activate and press ENTER
the siphon (usual value from 10 to 20 seconds).
With the arrow down key, choose "FIX TIME"
Enter this value in INJ. TIME: XXX (in seconds) in and press ENTER
the CALIBRATION menu (see chapter 2.2.2
Options) or in "SAMPLE TIME : XXX" of the PM Enter and check the injection times of each
XXXX / OPTION menu / GRAB SAMPLE (see pump that is used during the sequences "FIX-
chapter 2.2.2 Options). TIME" and "STAND-BY" :
INJ. TIME P1 : 00 SEC (press ENTER)
INJ. TIME P2 : 10 SEC (press ENTER)
2.7.3 Micro-piston pump: INJ. TIME P3 : 00 SEC (press ENTER)
The standard operating procedure for setting, INJ. TIME P4 : 00 SEC (press ENTER)
measuring and programming the flow rate of this
kind of pump is:
If "FIX-TIME" has been chosen, the time
Setting : Turn the ring of the pump (with the FMI
interval between two measurements is the time
label). From a front view, the whole label must fit
programmed in the submenu "CYCLE TIME"
between the in and out connectors. Loosen the
of the "TIMING" menu.
metal screw of the small ring, then rotate it until
the zero matches the graduated axe of the large If the "LOOP" mode is required, again press
ring. Tighten the screw. The flow rate of the pump ENTER, to adjust the SEQUENCE menu.
should now be around 2 ml/mn. If not, rotate the
large ring over 360. With the arrow down key, select LOOP and
press ENTER.
To measure the micro-piston pump flow with
precision : Weigh an empty 50 ml beaker, then
place the pump outlet tube (rigid, FEP type) in the
recipient. Set PUMP 1 = ON for the duration of 10
minutes and then weigh the full recipient. The
difference (divided by 10) should be near to 2 g,
the equivalent of 2 ml/mn.

Page 6 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

2.9 ANALOGUE OUTPUTS 3 CALIBRATION


With the arrow keys, select "ANALOG OUT"
and press ENTER. Enter or check the following 3.1 PRIMARY CALIBRATION
information : (e.g. 0-500 mg/l) This is the first operation to be carried out after the
programming. This application requires a two-
START I1 : 0.000 (press ENTER)
point calibration, the first one on softened or
END I1 : 500.0 (press ENTER) demineralised water, the second on a solution of
known hardness (e.g. 40F)
OUT : Conc

With the arrow down key, select


"CALIBRATION" and press ENTER

With the arrow keys, select "ZERO" and With the arrow down key, select "PRIMARY
position your recorder at "ZERO". CAL" and press ENTER

With the arrow keys, select "FULL" and With the arrow down key, select "AUTO. SYS
position your recorder at total deviation. CAL" or "PROCESS SYS CAL" or "MANUAL
SYS CAL" and press ENTER.
With the arrow keys, select "CONC" and press
ENTER. - If "AUTO. SYS CAL" has been chosen, press
ENTER and the screen will read
"IMMEDIATE : NO". With the arrow keys,
select "IMMEDIATE : YES". The analyzer will
automatically link together a double analysis
2.10 PROGRAMMING AUTOMATIC cycle using rinse (sprinkler) water as its first
CALIBRATION point, and then the automatic calibration
solution as its second point.
With the Arrow down key, select - If "MANUAL SYS CAL" has been chosen,
"CALIBRATION" and press ENTER follow the various instruction indicated on the
screen :
The screen reads SYSTEM CAL. Press
ENTER. If "FILL CAL SOL 1" appears, pour 200 ml of
the manual calibration solution No. 1 in the
The screen reads, "AUTO SYS CAL". Press cell (siphon activation) and press ENTER.
ENTER After titration, enter or check the value of
Enter or check the following information : "CONC. 1 = XX.XX" ( ~1/2 of CONC.2) and
press ENTER.
If "FILL CAL SOL 2" appears, pour 200 ml of
IMMEDIATE : NO (press ENTER) the manual calibration solution No. 2 in the
cell (siphon activation) and press ENTER.
FREQUENCY : XXX ~ usually 096 to have one After titration, enter or check the value of
Cal.Auto. a day (if one measurement is 15 mn). "CONC. 2 = XX.XX" and press ENTER.
To be defined according to your application
(expressed in number of cycles) ~ (press ENTER) - If "PROCESS SYS CAL" has been chosen,
the system will automatically take rinse water
CONC. 1 : 00.00 E 00 ~ Must be equal to for its first point and the process solution for
zero, because the water of the sprinkler must be its second point. After titration, enter the
exempt of hardness, like softened or nominal concentration value of the process
demineralised water ~ (press ENTER) solution.
CONC. 2 : XX.XX E XX ~ Concentration NOTE: If the rinse water is not softened or
of the automatic calibration solution. Expressed in demineralised water, never select "AUTO. SYS
the same unit as chosen in chapter 2.2.4 Units:. CAL" or "PROCESS SYS CAL".
For the preparation of this solution see chapter 4.3
~ (press ENTER) After calibration, copy the values of the slopes
and asymmetries onto a follow-up sheet and
INJ. TIME : XXX s ~ Refer to chapter 2.7.2 press ENTER.
to determine the exact value ~
(press ENTER) Follow the instructions of chapter Error!
Reference source not found. Error!
Reference source not found. to start the
measurement mode.

Page 7 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

3.2 STARTING THE ANALYZER 3.4 READING THE CALIBRATION


Now you are ready to start your measurements. PARAMETERS

Press SELECT to exit the PROGRAMMING At any moment it is possible to read the
menu. calibration parameters, without stopping the
analyzer.
With the arrow keys, select "START". The
analyzer will start and do measurement cycles Press SELECT to obtain the main menu.
on its own. Press the arrow down key to visualize the
message "READING".
Press ENTER and keep it pressed for 5
seconds to enter the "READING" menu.
3.3 REGULAR CALIBRATION
Press the arrow down key to obtain the "CAL
If the AUTOMATIC CALIBRATION option has PARAMETER" menu and press ENTER.
not been acquired, we recommend you
proceed with a maximum of one calibration per Press ENTER again to read the information
week and a minimum of one calibration per from the "LAST CAL" list:
month. To do so, always choose a SYSTEM A / P / M / N indicate the calibration modes. They
calibration to allow the analyzer to compare the stand for automatic, process, manual, and
values with the initial parameters of the programmed respectively.
PRIMARY calibration. S= XXX.X for the slope
Press ENTER twice to stop the analyzer. E= XXX.X for the asymmetry
Copy these parameters onto a follow-up sheet
Press SELECT to obtain the main menu. of the Analyzer.
Press the arrow down key to visualize the Press SELECT twice to go back to the main
message "PROGRAMMING". functioning menu.
Press ENTER and keep it pressed for 5
seconds to enter the "PROGRAMMING"
menu.
Press the arrow up key to visualize the 4 PREPARING THE REAGENTS
message "CALIBRATION". Then press This procedure allows for the preparation of two
ENTER. 10 litre canisters of the reagents R1 and R3 and a
The message "SYSTEM CAL" will appear. 5 litre canister of reagent R2.
Press ENTER.
With the arrow down key, select "PROCESS THE PREPARATION OF THE CONDITIONING
SYS CAL" or "MANUAL SYS CAL" and press AND CHEMICAL CLEANING REAGENTS
ENTER. MUST BE CARRIED OUT WITH THE
PRECAUTIONS FOR USE INHERENT TO
Follow the instructions of chapter 3.1 primary THESE PRODUCTS, AMONG WHICH THE
calibration until the measurement mode is WEARING OF SAFETY GLASSES.
launched

Note: The Estimated Shelf Life (ESL) of each


canister is indicated at the end of the preparation
instructions.

Page 8 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

4.1 R1: TITRATION REAGENT (FOR 10 2.5 Sample Conditioning:). Increase or diminish
LITRES) the injection time INJ. REAG so as to obtain an
after titration solution pH 8,3 0.2 pH units.
Triethanolamine > 97% 20 ml / l
EGTA 10 g/l

To eliminate the sources of variation, the Consumption: With a 120 second conditioning
preparation of this solution must be very time and a measurement every 12 minutes, 10
reproducible and very simple. Therefore it is
litres provide for 40 days of functioning.
preferable to use a scale for measuring 10 kg,
instead of multiple volume adjustments. ESL, sealed canister: 6 months
ESL, open canister: 2 months
Fill a 10 litre can (rinsed and tared),
with ~4 litres of demineralised water. To facilitate
the dissolving of the EGTA, use a graduated R3: CHEMICAL CLEANING (for 10 litres):
cylinder to measure 200 ml of Triethanolamine
HCl 37% 10 % v/v
(C6H15NO3, reagent quality > 97%), and pour it
Sodium benzoate 2g/l
into the canister. Then rinse the cylinder (at least
3 times) and pour the rinse water into the can. In a In a 10 litre canister, pour 9 litres of demineralised
clean and dry beaker, weigh 100 g of ethylene water. While shaking, add, portion-wise, 1 litre of
glycol-O-O-bis (2-aminoethyl) -N,N,N,N- tetra concentrated hydrochloric acid (HCl, 37%, d240 =
1.182). Then pour in 20 g of sodium benzoate
acetic acid (C14H24N2O10 97.0%) and pour it
directly into the canister. Put the top on, shake to (C7H5NaO2, puriss > 99%). Put the top on, shake
mix (or mix with a magnetic mixer) until the EGTA to mix, label, date and install the canister on the
is completely dissolved. analyzer.

Add ~5 litres of demineralised water to the The R3 reagent is added during the chemical
canister, then adjust to exactly 10.00 kg. Put the cleaning by activating the cleaning pump (situated
top on, shake to mix, label, date and install the on the back of the analyzer). The duration of the
canister on the analyzer. injection is programmable (see 2.6 chemical
cleaning:) so as to keep the measuring cell and
The R1 reagent is used by pump No. 1 for titration its electrodes perfectly clean.
until the programmed potential is obtained (See
chapter 2.3 TITRATION ParametErs:). Consumption: With one injection every 3
seconds, one cleaning per measurement, and
Consumption: depends on the sample a measurement every 12 minutes, the 10 litre
concentration. canister will last 40 days.
ESL, sealed canister : 6 months, ESL, sealed canister: 1 year,
ESL, open canister: 2 months. ESL, open canister: 6 months.

4.2 R2: CONDITIONING REAGENT (FOR 10 4.3 R4: SOLUTIONS FOR AUTOMATIC
LITRES): CALIBRATION (10 LITRES):
Triethanolamine > 97% 150 ml / l CaCl2 (expressed in mg/l equivalent CaCO3,
HCl 37% 20 ml / l according to the range table below)
In a graduated flask of 5 000 ml, introduce
Directly in a 10 l canister (rinsed), fill with about 6 precisely with a pipette Y ml of the STOCK
litres of demineralised water. Add slowly and in
solution 100 g/l equivalent CaCO3 (see chapter
portions 1.5 l of Triethanolamine (C6H15NO3,
reagent quality > 97%). Put the top on and shake 4.4.1 STOCK Solution: 100g/l equivalent CaCO3).
(or mix with a magnetic mixer) until the Top off to 5 litres with demineralised water and
triethanolamine is completely dissolved. transfer it into a 10 litre canister.
Finally, pour in 200 ml of concentrated Again fill the graduated flask with 5 litres of
Hydrochloric acid (HCl, 37%, d240 = 1.182). Top demineralised water and add it to the previous
off to 10 litres with demineralised water. Put the volume. Put the top on, shake to mix, label, date
top on, shake to mix, label, date and install on the and install the canister on the analyzer.
analyzer.
The R2 reagent is added automatically during the
conditioning phase, by activating pump N 2. The
duration of the injection is programmable (Chapter

Page 9 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

Target range [AUTO. CAL. SOL.] 4.4.3 Two CALIBRATION solutions prepared
Y
with demineralised water:
(mg/l CaCO3) in mg/l CaCO3 in ml
0 120 100 10
Manual calibration solution _No. 2
0 250 200 20 Z mg/l equivalent CaCO3 (for 1 litre)
0 500 400 40
In a graduated flask of 1 litre, introduce precisely,
0 1 000 800 80 with a pipette, Z ml of the intermediate solution at
10g/l equivalent CaCO3.

The R3 reagent is added during the automatic Top off to 1 litre with demineralised water.
calibration by activation of the volumetric pump.
The injection time and the concentration of the
Target range [CALIBRATION No. 2] Z
automatic calibration solution ("CONC 2 : XX. XX
E - XX") are programmable (see chapter 2.10 (mg/l CaCO3) in mg/l CaCO3 in ml
PROGRAMMing automatic calibration). 0 120 100 10
0 250 200 20
Consumption: With 250 ml consumed per 0 500 400 40
calibration and one calibration a day, 10 litres
will last 40 days. 0 1 000 800 80

ESL in a sealed canister : 6 months.


ESL in an open canister: 2 months.
Manual calibration solution No. 1
*Z mg/l equivalent CaCO3 (for 1 litre)
In a graduated flask of 1 litre, pour 500 ml of the
Manual calibration solution No. 2 and top off to 1
litre with demineralised water.

4.4 SOLUTIONS FOR MANUAL CALIBRATION


For all the 1 litre solutions:
4.4.1 STOCK Solution: 100g/l equivalent
CaCO3 ESL, sealed canister : > 1 year
In a graduated flask of one litre, pour 147 g of ESL, open canister : 1 year
calcium chloride dihydrate (CaCl2.2H2O,
crystallized, >99%). Dissolve and fill up to one
litre with demineralised water.

4.4.2 Intermediate Solution: 10g/l equivalent


CaCO3
In a graduated flask of one litre, pour 100 ml of the
STOCK solution at 100 g/l. Fill up to 1 litre with
demineralised water.

Page 10 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

5 PREVENTIVE MAINTENANCE Titrate with "PUMP 1". After several minutes,


the potential slowly decreases, then
accelerates to a new setting.
5.1 DETERMINING THE END OF TITRATION Keep "PUMP 1" set to ON until the variation in
The configuration of END PT can vary in the potential is less than ~2mV 1n 10s.
course of time. We recommend normalizing the Note down the value of the stabilized potential
exact determination of this value, ideally once a (in mV), because it marks the end of titration.
month.
A follow-up notebook of these manipulations will
Enter or check this same value in the menu
"MEASUREMENT / END PT XXXX mV"
allow you to note down the points underlined
below, to start with the date, time, and operators
initials.
Press ENTER twice to stop the analyzer. 5.2 CHECKPOINTS
Press SELECT to obtain the main menu. During an adjustment phase, the results obtained
Press the arrow down key to visualize the with the 8810 analyzer will be compared to regular
message "PROGRAMMING". laboratory measurements.
As a general rule, one can count on the necessity
Press ENTER and keep it pressed for 5
for maintenance once a month. We then
seconds to enter the "PROGRAMMING" menu.
recommend that the following operations be
With the arrow down key, select "SERVICE" carried out:
and press ENTER
Check the chemical product contents of the
With the arrow keys, activate and deactivate canisters and fill them up if necessary.
the various functions as follows:
Only for the peristaltic pumps with Tygon
Empty the cell with "DRAIN". tubing: every 4 weeks disconnect the tubing
and put it back in the opposite direction. Every
Rinse the cell with "SPRINKLER" while leaving 8 weeks replace the tubing.
"DRAIN" set to ON.
Check the fouling of the measuring cell and the
Stop "SPRINKLER", then stop "DRAIN". electrodes. If necessary clean manually.
Take a sample with "SAMPLE" letting it flow for Make sure that the analysis process runs
about 10s. perfectly (Error! Reference source not
Stop "SAMPLE" making sure that the level in found. Error! Reference source not found.)
the cell is reached correctly (activation of the Calibrate according to the locally chosen
siphon). method (Error! Reference source not found.
Set the "MIXER" to ON, until the end of the Error! Reference source not found.).
manipulation.
Condition with "PUMP 2" for the duration
defined in the menu "SAMPLE COND." (about
90 s), then OFF.
Around 15-20 seconds after conditioning, note
down the potential value (in mV), because it
marks the beginning of titration.

Page 11 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

6 ALL ABOUT THE ELECTRODES


6.1 STORAGE, RECONDITIONING OF THE ELECTRODES

Electrode for ISE CALCIUM Reference : 8483 Previous reference : 8447


Type of Ion exchange surface Internal element Ag/AgCl. Internal element Ag/AgCl.
electrode Internal electrolyte Gel electrolyte Electrolyte KCl (3M, 1M or
Diffusion through open 0.1M possible).
junctions. Diffusion by sintering.
Storage:
Conditions Dry Soaked in KCl, 1 or 3 M. Soaked in KCl, 1 or 3 M.
Duration Longer than 2 years Less than 6 months. Less than 6 months.
COMMISSION- Place it in the reactor of the Dab it on paper. Dab it on paper.
ING of a new 8810 (see chapter 6.2). Rinse with demineralised Rinse with demineralised
electrode (or Connect it to the cable with water. water.
release from the largest diameter. Shake it down (like a Shake it down (like a
storage) mercury thermometer). mercury thermometer).
Place it in the reactor of the Place it in the reactor of the
8810 (see chapter 6.2) 8810 (see chapter 6.2)
Connect it to the cable with Connect it to the cable with
the smallest diameter. the smallest diameter.
Calibration No calibration of the electrode See chapter 3 calibration . See chapter 3 calibration .
See chapter 3 calibration .
Specifications See chapter 1.1 See chapter 1.1 See chapter 1.1
BASIC CONFIGURATION BASIC CONFIGURATION BASIC CONFIGURATION
Simple 10 mn in HNO3 diluted 10% 10 mn in HNO3 diluted 10% 10 mn in HNO3 diluted 10%
cleaning v/v of the commercial 65%. v/v of the commercial 65%. v/v of the commercial 65%.
Follow Commissioning. Follow Commissioning. Follow Commissioning.
Regeneration See Operating mode of the Unplug the open junctions Rinse several times then fill
electrode. with the help of a metal the electrode with KCl
Follow Commissioning. filament (trombone type). solution [variable].
Follow Commissioning. 10 mn in KCl 0.1 M at 60-
80C.
Rinse then fill the electrode
with KCl solution [variable]
Follow Commissioning.

Page 12 / 13
8810 ORP CALCIUM HARDNESS: 0 - 500 mg/l CaCO3

6.2 DIFFERENT ELEMENTS IN THE MEASURING CELL

Reactor of the 8810


Overflow pipe Mixer
Injectors
Working or
Large combined electrode
Siphon section
cable
> 1 cm

> 1 cm
Small
section
cable

Sprinkler PT 100

Heating cartridge
Reference
electrode
Level detector

Page 13 / 13

You might also like