You are on page 1of 2

Coincidence/Anti-coincidence

Detector Applications Information Note

Coincidence and anti-coincidence are detection modes (or techniques) used to produce a simplified
spectrum from certain types of detector systems. These systems consist of two (or more) detectors, a
segmented detector, or a combination of these. Each detector or segment produces separate signals.

In coincidence, only those signals produced The spectrum is produced by the coincident events
simultaneously in at least two of the detectors or from positron pair creation in the spectrometers
segments are counted. In anti-coincidence, the central detector and absorption of the 511 keV
signals produced simultaneously in at least two of annihilation quanta in the surrounding shield detector.
the detectors or segments cancel or veto each other, This coincidence effectively isolates the double escape
leaving the non-coincident signals to be counted. The peak. However, the full energy peak is suppressed
single most important benefit of using coincidence or along with the Compton edge and background,
anti-coincidence is a greatly improved signal to noise reducing the counting efficiency.
ratio in the detector system. This leads to much greater
A typical pair spectrometer combines a highly efficient
accuracy in the determination of full energy peaks in
NaI(Tl) detector - or a Ge detector with very good
the spectrum.
energy resolution - with a NaI(Tl) annulus detector. In
The effectiveness of a coincidence/anti-coincidence some systems, the annulus may have optically isolated
system can be calculated by dividing the noise segments, or two separate detectors can be used. A
reduction factor by the factor by which the total pair spectrometer can be used as an anti-Compton
absorption peak height changes to get a number spectrometer by simply connecting the annulus in
relative to the net reduction in the signal to noise anti-coincidence with the central detector. However,
ratio. This quick assessment method points out the using a larger central detector is recommended for
importance of maintaining total peak height as the this application.
background continuum is reduced.
Detecting Positron Emitters
Coincidence is applied in sum-coincidence and pair
spectrometers and systems for detecting positron A large number of artificially produced radioactive
emitters. Anti-coincidence is used with Compton samples emit -rays in coincidence with positron
suppression spectrometers and can be set up in particles (+) decay. In this process the positron can
phoswich detectors as well. annihilate into two 511 keV rays. These -rays can be
detected in a spectrometers shield detector. If there
Sum-Coincidence Spectrometers is coincidence between the 511 keV gamma events and
the events detected in the central detector, the -rays
Two or more adjacent detectors (or one detector
emitted after + decay can be detected with a very
with two or more optically-isolated segments) form
low background.
a sum-coincident spectrometer. The signals from
events coincident in at least two detectors (or two Positron Emission Tomography (PET) is essentially a
segments) are summed to produce the spectrum, large multi-detector (thousands) pair spectrometer
which is essentially the full energy peak. that detects the two 511 keV gamma rays emitted from
radiopharmaceuticals. The known positron of the two
Pair Spectrometers detectors that are in coincidence define a line along
which the source originated. Many such events are
When samples emit -rays with energies large enough
then reconstructed into a 3-D tomographic image of
for pair production, it is possible to use this effect
a patient.
for Compton and background suppression. In pair
production the incident gamma photon loses at least
1022 keV of energy. This energy converts into kinetic
energy and mass for the simultaneously created
positron-electron pair. When the positron loses this
kinetic energy, it annihilates into two, diametrically-
emitted 511 keV gamma rays.

CRYSTALS
Coincidence/Anti-coincidence

The new Saint-Gobain Crystals materials The basic principle behind Compton suppression is the
BrilLanCe380*(LaBr3:Ce)and BrilLanCe350 (LaCl3:Ce) detection of photons scattered by the central detector
are very fast, with BrilLanCe 380s timing being as good into a large, high-efficiency detector configured to
as or better than BaF2. BrilLanCe 380 can be used to surround it . The escape of a Compton scattered -ray
add time-of-flight gating capability to a PET system. generates a signal in the surrounding shield detector.
The signal is coincident with the signal generated by
Compton Suppression Spectrometers the Compton electron in the central detector.
In -ray spectrosopy, the detection of low intensity The signal from the guard detector can therefore
peaks is often complicated by the Compton scattering be used to veto the registration of Compton events
of higher energy -rays. In the energy range between occurring in the central detector. In this way the
several hundreds of keV and several MeV, the Compton continuum can be suppressed.
interaction between -rays and a detector takes place
mainly through Compton scattering. Phoswich Detectors
The detection of Compton scattered -rays gives rise The two scintillators produce pulses with different
to a continuous background in the energy spectrum shapes and are viewed by the same photomultiplier.
and consequently raises the lower detection limits for With pulse shape discriminating (PSD) electronics,
-ray energies in that region. Anti-Compton shields a separate peak is obtained for each scintillator. By
(Compton suppressors) provide a tool to suppress the setting a window on a peak, the signals from the
unwanted background. The combination of a central appropriate scintillator are selected, either primary or
NaI(Tl) or germanium detector and an active, NaI(Tl) or guard crystal. This can be used to reduce background
BGO anti-Compton shield detector is called a Compton and Compton contributions to the region of interest.
suppression spectrometer. The combination of NaI(Tl) and BrilLanCe 380 can
produce an overall good performance because of the
fast decay time of BrilLanCe 380 (16ns) as well as the
other exceptional qualities of that material.

For additional product information, request the following SGC literature:

Compton Suppressor Product data sheet;


Phoswich Detectors Product data sheet;
A Faster Phoswich: A BrilLanCe Application
Low-Level/Low Background Counting Applications data sheet;
Time Coincidence Applications data sheet;
Individual Materials data sheets.
All are available as PDF files on www.crystals.saint-gobain.com

Protected under patents US7,067,816B2, US7250609B2, EP1257612B1*,EP1516078B1,


ZL03813659.7,UA75066C2,US7067815B2,US7233006B2,EP1255796B1*, EP1516078B1,-
ZL01805267.3,ZL03813659.7,EP1516078B1,UA75591C2,UA75066C2
*These original patents were granted to Stichting Voor de Technische Wetenschappen. Inven-
tors are P. Dorenbos, C.W.E. van Eijk, H.U. Gudel, K.W. Kraemer, E.V.D. van Loef.
Technology is licensed to Saint-Gobain Cristaux & Detecteurs.

is a registered trademark of Saint-Gobain Ceramics & Plastics, Inc.

The data presented are believed to be correct but are not guaranteed to be so.
Saint-Gobain Crystals
Manufacturer reserves the right to alter specifications.
www.crystals.saint-gobain.com 2004-2016 Saint-Gobain Ceramics & Plastics, Inc. All rights reserved. (08-16)

You might also like