You are on page 1of 10

Clinical Chemistry 61:2 Automation and Analytical Techniques

420429 (2015)

Digital Microuidic Platform for the Detection of


Rubella Infection and Immunity:
A Proof of Concept
Alphonsus H.C. Ng,1,3 Misan Lee,2,3 Kihwan Choi,3,4 Andrew T. Fischer,5 John M. Robinson,6
and Aaron R. Wheeler1,3,4*

BACKGROUND: Whereas disease surveillance for infec- Vaccine-preventable infectious diseases such as measles,
tious diseases such as rubella is important, it is critical to mumps, and rubella continue to be a global threat be-
identify pregnant women at risk of passing rubella to cause many developing countries have low or nonexistent
their offspring, which can be fatal and can result in con- immunization coverage (1 ). Although rubella is manage-
genital rubella syndrome (CRS). The traditional central- able for most adults and minors (rash, fever, and flu-like
ized model for diagnosing rubella is cost-prohibitive in symptoms), rubella infection during pregnancy has sig-
resource-limited settings, representing a major obstacle nificant risk of causing fetal death and/or a debilitating
to the prevention of CRS. As a step toward decentralized suite of defects known as congenital rubella syndrome
diagnostic systems, we developed a proof-of-concept dig- (CRS)7 (2 ); 100 000 infants are born with CRS each
ital microfluidic (DMF) diagnostic platform that pos- year (3 ).
sesses the flexibility and performance of automated im- As is the case for many infectious diseases, screening
munoassay platforms used in central facilities, but with a for rubella susceptibility, immunity, and infection is
form factor the size of a shoebox. challenging, requiring the selective detection of rubella
virus (RV)-specific IgG and IgM. Because the test com-
METHODS: DMF immunoassays were developed with in- ponents of RV IgM immunoassays can cross-react with
tegrated sample preparation for the detection of rubella other biogenic proteins [e.g., RV IgG or rheumatoid fac-
virus (RV) IgG and IgM. The performance (sensitivity tor (RF) IgM], there is a high occurrence of false-negative
and specificity) of the assays was evaluated with serum or false-positive results, depending on the quality of the
and plasma samples from a commercial antirubella assays and the experience of the operators (4 6 ). Thus,
mixed-titer performance panel. although there are portable tests available (7 ), rubella is
almost always diagnosed in central clinical laboratories.
RESULTS: The new platform performed the essential pro-
In resource-limited settings, transportation from rural
cessing steps, including sample aliquoting for 4 parallel communities to remote laboratories is often cost prohib-
assays, sample dilution, and IgG blocking. Testing of itive (8 ), meaning that rubella infections (and attendant
performance panel samples yielded diagnostic sensitivity cases of CRS) routinely sweep through rural populations
and specificity of 100% and 100% for both RV IgG and without testing or treatment (9 ).
RV IgM. With 1.8 L sample per assay, 4 parallel assays Given the challenges described above, there is great
were performed in approximately 30 min with 10% enthusiasm for moving to decentralized systems for
disease diagnosis and surveillance. This model requires
mean CV.
portable assays, and the microfluidic (or lab on a
chip) community has attempted to address this chal-
CONCLUSIONS: This proof of concept establishes DMF-
lenge primarily through systems that rely on fluid flow
powered immunoassays as being potentially useful for the through enclosed microchannels (10 12 ) or lateral flow
diagnosis of infectious disease. through a paper-based absorptive matrix (1315 ).
2014 American Association for Clinical Chemistry
Microchannel-based systems have excellent analytical
sensitivity (16 ) and proven diagnostic performance in

1
Institute of Biomaterials and Biomedical Engineering, 2 Innis College, and 4 Department Previously published online at DOI: 10.1373/clinchem.2014.232181
of Chemistry, University of Toronto, Toronto, ON, Canada; 3 Donnelly Centre for Cellular 2014 American Association for Clinical Chemistry
and Biomolecular Research, Toronto, ON, Canada; 5 Abbott Diagnostics, Irving, TX; 6 Ab- 7
Nonstandard abbreviations: CRS, congenital rubella syndrome; RV, rubella virus; RF,
bott Diagnostics, Abbott Park, IL. rheumatoid factor; DMF, digital microuidics; Cal, calibrator; HRP, horseradish peroxi-
* Address correspondence to this author at: University of Toronto, 80 St. George St. To- dase; PMT, photomultiplier tube; LOD, limit of detection; LOQ, limit of quantication;
ronto, ON, Canada M5S 3H6. Fax 416-946-3865; e-mail aaron.wheeler@utoronto.ca. s/co, signal-to-cutoff ratio; EIA, enzyme immunoassay; RLU, relative light unit.
Received August 27, 2014; accepted November 13, 2014.

420
Microuidic Rubella Assay

resource-limited settings (1719 ), but often require ex- cluded Superblock Tris-buffered saline and SuperSignal
tensive ancillary equipment to operate (e.g., pumps, flow ELISA Femto chemiluminescent substrate, comprising
meters, and valves) and intricate device fabrication for stable peroxide (H2O2) and luminol-enhancer solution,
integrated sample preparation (20 ). In contrast, paper- from Thermo Fischer Scientific; antirubella mixed-titer
based systems can be simple and inexpensive to manufac- performance panel (PTR201-00-0.5) from SeraCare Life
ture and use (21 ), but adequate analytical performance in Sciences; and defibrinated sheep blood from Quad Five.
such devices has been demonstrated for only a few ana- We prepared custom digital microfluidic compati-
lytes (15 ), and multistep liquid manipulation with uni- ble wash buffer and conjugate diluent as described previ-
form flow profiles requires careful material manipulation ously (32, 33 ). Conjugate working solutions for RV IgG
and device design (2224 ). or RV IgM assays were formed by diluting horseradish
An alternative to (channel-based or lateral) fluid peroxidase (HRP)-conjugated goat polyclonal anti-
flow for miniaturized analysis systems is a technology human IgG (21 ng/mL) or HRP-conjugated goat anti-
known as digital microfluidics (DMF) (2527 ), in which human IgM (23 ng/mL), respectively, in conjugate di-
sample and reagents are manipulated as discrete droplets luent. Pretreatment working solution was formed from a
on a hydrophobic surface. DMF systems actuate droplets 3 dilution of IgM pretreatment reagent in conjugate
through the application of electrical potentials on a ge- diluent. The microparticle working suspension was
neric array of insulated electrodesa format that enables formed in an Eppendorf tube by removing the original
software-reconfigurable, concurrent droplet operations diluent (with particles immobilized with a neodymium
including merging, mixing, splitting, and metering from magnet), washing the particles twice in Superblock, and
reservoirs (28 ). DMF devices can be cost-effectively fab- resuspending the washed particles in Superblock at 1.5
ricated on paper by inkjet printing (29 ) and can be 108 particles/mL (approximately 10 the stock concen-
operated with simple and compact instrumentation tration). Simulated whole blood samples were formed
(30, 31 ) with no need for pumps, interconnects, valves, from a 1:1 mixture of RV IgG calibrator and sheep blood.
or fittings. Although these advantages have motivated the Before running immunoassays, we diluted calibrators,
development of DMF-powered immunoassays for small- simulated whole-blood samples, or patient samples (from
molecule and protein biomarkers (3235 ), we are not performance panel) 10 in Dulbecco PBS containing
aware of DMF immunoassays being used for diagnosis of 4% BSA. All reagents used on digital microfluidic devices
infectious diseases. or on-chip were supplemented with Pluronic L64 at
Here, we report the development of DMF-powered 0.05%.
RV IgG and RV IgM immunoassays, with a focus on
evaluating the diagnostic performance of DMF immuno- DEVICE FABRICATION AND OPERATION
assays relative to central laboratory methods. These assays We used a custom automation system with approximate
use RV-immobilized magnetic particles to capture the dimensions 7 9 12 inches to manage droplet oper-
analyte from the sample and enzyme-linked detection ation, magnet and photomultiplier tube (PMT) position,
antibodies to transduce analyte-binding events to chemi- and data collection (Fig. 1A) (30, 32 ). Digital microflu-
luminescent signal. All assay steps, from sample to anal- idic devices, comprising a top plate and bottom plate,
ysis, were performed by DMF with a shoebox-sized pro- were fabricated in the University of Toronto Nanofabri-
totype instrument. We propose that this represents a cation Centre cleanroom facility and assembled as de-
useful step toward the development of decentralized di- scribed previously (32 ) (Fig. 1B). The bottom plate de-
agnostic tools for the diagnosis of rubella and other in- vice design featured an array of 80 actuation electrodes
fectious diseases. (2.54 2.54 mm) connected to reservoir electrodes for
sample and reagent storage and waste removal. Unit
Materials and Methods droplet volumes on the actuation electrodes were approx-
imately 900 nL, as determined by the area of each actu-
REAGENTS AND MATERIALS ation electrode and the gap spacing (approximately 140
Unless otherwise specified, we purchased reagents from m) between the top plate and bottom plate.
Sigma-Aldrich. Deionized water had a resistivity of 18 Droplets were actuated by applying a prepro-
M cm at 25 C. Pluronic L64 (BASF Corp.) was gen- grammed sequence of driving voltages (80 100 VRMS 10
erously donated by Brenntag Canada. kHz sine wave) between the top plate (ground) and elec-
We adapted most immunoassay reagents from the trodes in the bottom plate through a Pogo pin interface
Architect rubella IgM and IgG assay kits from Abbott (90 pins). For on-chip particle separation, a motorized
Laboratories, including RV IgG calibrators, RV IgM cut- magnet system called a magnetic lens (32 ) is positioned
off Calibrator 1 (Cal 1), RV IgM pretreatment reagent approximately 150 m underneath the device. In this
containing goat antihuman IgG, and virus-coated para- activated state, 600 N of magnetic force is sufficient
magnetic microparticles. Reagents from other vendors in- to focus particles into a pellet, immobilizing them on the

Clinical Chemistry 61:2 (2015) 421


A B
Motors 2 in
3 in Magnetic
lens
Top Electrode
plate array

PMT Bottom
Electrical Electrical
inputs plate
interface
30cm to device
Interface to
instrument
Magnetic
g Top plate
lens Glass
140 m Droplet Air
~ Electrode
Motor
Hydrophobic
18cm
23cm Bottom plate Insulator
Front view (photograph) Cross-section view (rendering) Magnetic lens

C D
1 2 1 2 3
Particle
IgM suspension IgG Magnetic lens active
2.54 mm 4 5 6
3 4
Waste
Separation zone Particle pellet 7 8 9

Fig. 1. Automated digital microuidic platform.


(A), Left: Photo of automation system (front view) showing the position of electrical inputs from the DropBot (30 ) and external motors that
control the PMT and enclosure. Right: Cross-section computer-aided design rendering showing the interior components of the system,
including the PMT, the magnetic lens, and the motor system that controls the position of the magnetic lens. (B), Three-dimensional schematic
and side view of digital microuidic device positioned above the magnetic lens. (C), Video sequence of digital microuidic magnetic separa-
tion of paramagnetic particles in parallel RV IgM and RV IgG immunoassays (see online Supplemental Video 1). (D), Video sequence of
resuspending and mixing of particles in wash buffer solution. Red arrows indicate the direction of droplet movement.

surface (Fig. 1C; also see Supplemental Video 1, which cluding (a) 2 sensors to ascertain the position of the
accompanies the online version of this article at http:// PMT, (b) a shutter mechanism to prevent light from
www.clinchem.org/content/vol61/issue2). With digital saturating the sensor when it is not in use, and (c) deep
microfluidic actuation, the supernatant can be separated grooves in the enclosure to ensure that the instrument is
from the particles, provided that the immobilization light-tight during measurements.
force exceeds the minimal threshold (approximately 500
N) required to overcome the surface tension of the RV IgG IMMUNOASSAY
droplet (32, 33 ). The particle pellet is resuspended by RV IgG antibodies were detected on-chip with an indi-
deactivating the magnetic lens (i.e., magnet is positioned rect ELISA comprising 7 steps. (a) A droplet of virus-
3.8 cm below the device) and moving and mixing a drop- coated paramagnetic particle suspension (1.8 L) was
let over the particles (Fig. 1D). dispensed from a reservoir, and the particles were sepa-
To measure on-chip chemiluminescence, the reac- rated from the diluent. (b) A droplet of sample (1.8 L,
tion droplet is moved to the center of the electrode array containing RV IgG) was dispensed, delivered to the par-
and a motorized PMT is positioned several hundred ticles, mixed for 3 min, and separated from particles. (c)
micrometers above the device to collect light from the The particles were washed in 4 successive droplets of
droplet. When the PMT is inactive (i.e., the PMT is wash buffer (i.e., 4 1.8 L), each time mixing and then
positioned 5 cm above the device), an integrated light- separating from the supernatant. (d ) A droplet of HRP
emitting diode and webcam assembly is used to monitor conjugate solution (1.8 L, containing HRP-conjugated
on-chip droplet movement. We implemented 3 design antihuman IgG) was dispensed, delivered to the particles,
measures to prevent users from damaging the PMT, in- and mixed for 2 min. (e) Step c was repeated. (f ) One

422 Clinical Chemistry 61:2 (2015)


Microuidic Rubella Assay

droplet each of H2O2 (900 nL) and luminol-enhancer these experiments, at least 2 replicates were evaluated for
solution (900 nL) were dispensed and merged with the each sample by absorbance, measured at 450 nm (refer-
particles; the solutions and particles were mixed for 2 ence 650 nm) with a Sunrise microplate reader (Tecan) in
min. (g) The mean chemiluminescent signal over 10 s was Accuracy read mode. We estimated the RF IgM con-
obtained with the integrated PMT. centrations of the samples from a calibration curve, gen-
By aliquoting 4 sets of samples and reagents (see erated by fitting the absorbance of the ELISA calibrator
online Supplemental Fig. 1), 4 parallel tests were run solutions to a 4-parameter logistic equation.
simultaneously (i.e., steps a g were performed in parallel
for 4 samples). Applying this protocol, a calibration curve RV IgM IMMUNOASSAY
for RV IgG was established in 4 replicates (intraassay), The DMF RV IgM immunoassay was identical to the
with calibrator solutions containing 0, 5, 10, 15, 75, and 7-step protocol (a g) described above, except that the
250 IU/mL RV IgG. The limits of detection (LOD) and conjugate solution contained HRP-conjugated anti-
quantification (LOQ) for this assay were the concentra- human IgM (instead of IgG) and the sample was pre-
tions corresponding to the position on the calibration treated with antihuman IgG (before step a). In the pre-
curve of the mean signal generated from blank measure- treatment procedure, a droplet of sample (1.8 L) and a
ments plus 3 (LOD) or 10 (LOQ) times the SD of the droplet of pretreatment reagent (0.9 L) were merged
blank measurements. A set of calibration data was used and mixed for 7 min. The assay was evaluated with and
for repeated experiments until a new batch of devices or without pretreatment with 4 test samples: (i) 0 IU/mL
magnetic particles were introduced, or when the control RV IgG calibrator (control), (ii) 250 IU/mL RV IgG
sample measurements were observed to be out of range. calibrator (IgG), (iii) 1:1 mixture of Cal 1 and 0 IU/mL
In practice, this allowed the use of each set of calibration RV IgG calibrator (IgM), and (iv) 1:1 mixture of Cal 1
data for 200 experiments over 50 days. and 500 IU/mL RV IgG calibrator (IgM IgG). For
each sample, 4 intraassay replicates were performed for
INTEGRATED ON-CHIP DILUTION AND RV IgG IMMUNOASSAY each condition tested. The results (for IgG, IgM, and
As shown in online Supplemental Fig. 2, samples were IgG IgM) were reported as fold change, which was
diluted on-chip in 3 iterating steps. (1) A droplet of sam- obtained by normalizing the signal intensity of each assay
ple (1.8 L) and a droplet of diluent (1.8 L, Dulbecco to the average signal intensity of the respective control
PBS 4% BSA) were dispensed, merged, and mixed. (2) experiments.
The pooled sample droplet (2 diluted) was split into 2 Applying the integrated RV IgM protocol (includ-
subdroplets. (3) One sample subdroplet was stored for ing both pretreatment and immunoassay), we established
later analysis, and the second sample subdroplet was used the cutoff calibration data in 3 replicates with unmodi-
to repeat steps 13 until the desired dilutions were fied Cal 1. In addition, the background signal of the RV
achieved. For example, 4 repetitions generate 4 droplets IgM assays was evaluated in 4 replicates (intraassay) with
with 2, 4, 8, and 16 dilution of original sample. a 0 IU/mL RV IgG calibrator.
After dilution, RV IgG immunoassays were performed,
with the 7-step (a g) procedure above, by delivering the ANTIRUBELLA MIXED-TITER PERFORMANCE PANEL
diluted samples to virus-coated particles. When the par- We estimated the diagnostic accuracy of DMF RV IgG
ticles were washed, the regions on the device used for and RV IgM immunoassays with 25 plasma/serum sam-
dilution and sample storage were also washed with addi- ples (PTR201-01-25, SeraCare Life Sciences). Serum or
tional wash buffer to prevent cross-contamination. plasma samples were thawed and aliquoted into single-
A fully integrated protocol, including dilution in 4 use vials for DMF immunoassays. Working solutions in-
iterations of steps 13 followed by RV IgG immunoas- cluding microparticles, conjugates, and chemilumines-
says in steps a g, was evaluated with a RV IgG calibrator cent substrates were made fresh each day from stock
and a simulated whole-blood sample (1:1 mixture of RV solutions.
IgG calibrator and sheep blood), both containing 250 DMF immunoassay analyses for the panel samples
IU/mL of RV IgG. Here, on-chip dilution was per- were completed by 3 different operators over a 1-month
formed to generate droplets containing 15.6, 31.3, 62.5, period. Each assay was carried out in 3 4 replicates in
and 125 IU/mL RV IgG. For each sample (calibrator and parallel, and operators were blinded to the reference re-
whole blood), the protocol was repeated on 3 different sult until all of the data were collected. The DMF assay
devices (i.e., 3 interassay replicates). results were reported as signal-to-cutoff ratio (s/co) val-
ues, calculated by normalizing the mean signal intensity
OFF-CHIP RF IgM SCREENING of each assay to the mean signal intensity of the respective
We used an RF IgM ELISA (Orgentec Diagnostika) ac- cutoff control. The cutoff controls for RV IgG and RV
cording to manufacturers instructions to screen for RF IgM assays were the 10 IU/mL RV IgG calibrator and the
IgM in the RV IgG calibrators and RV IgM Cal 1. In RV IgM cutoff Cal 1 (both included in the respective

Clinical Chemistry 61:2 (2015) 423


kits), respectively. A DMF s/co 1 was interpreted as may be suitable for portable testing in the field [i.e., low
positive, a DMF s/co in the gray zone (0.69 0.99 for cost, user friendly, battery-powered, and compact (39 )].
IgG and 0.75 0.99 for IgM) was interpreted as equivo- These new designs will likely benefit from the develop-
cal/indeterminate, and a DMF s/co below the gray zone ment of low-cost device fabrication methods (29 ), simple
(0.69 for IgG and 0.75 for IgM) was interpreted as instrumentation (30, 31 ), and integrated electrochemi-
negative. cal sensors (40, 41 ).

EVALUATION OF DIAGNOSTIC ACCURACY RV IgG IMMUNOASSAY AND SAMPLE DILUTION


We calculated the diagnostic sensitivity and specificity Whether acquired naturally or induced by vaccination,
for the DMF immunoassays with the Abbott enzyme the persistence of RV IgG antibody confers protection
immunoassay (EIA) as a reference testing method (36 ). against rubella infection. The WHO interpretation of
For cases in which the calculated sensitivity or specificity RV IgG serological results is dependent on the antibody
was equal to 1, a P value of 0.999 was used to calculate concentration: a sample is positive if antibody concentra-
95% CIs, as recommended by WHO (37 ) and others tion is 10 IU/mL, equivocal if it is between 5 and 10
(17, 18 ). IU/mL, and negative if it is 5 IU/mL. Individuals with
antibody concentrations 10 IU/mL are susceptible to
Results and Discussion clinical illness on exposure to rubella virus. As such, test-
ing women for RV IgG before conception or at their first
DIGITAL MICROFLUIDIC INSTRUMENTATION antenatal visit can help minimize the risk of CRS.
Motivated by the need to identify patients at risk for Using the particle separation and resuspension tech-
CRS, we developed DMF immunoassays for RV-specific niques described above, we developed a 7-step DMF in-
immunoglobulins. The challenging nature of this appli- direct ELISA for RV IgG, in which samples or reagents
cation, which includes the need to distinguish between are aliquoted to the particles for 4 parallel assays (see
RV IgG and RV IgM, necessitates a sophisticated system online Supplemental Fig. 1). As shown in Fig. 2A, RV
capable of handling a large number of reagents and per- IgG antibodies present in the sample bind to the RV-
forming a complex series of processing steps. To achieve coated paramagnetic particles. After incubation and
suitable levels of processing sophistication in a miniaturized washing, HRP-conjugated antihuman IgG is added to
instrument, we adapted an in-house designed computer- form immunocomplexes. After another incubation and
controlled instrument (32 ) powered by DropBot (30 ), wash cycle, hydrogen peroxide and luminol-enhancer so-
an open-source DMF automation system. lutions are added to the particles, and the resulting
The shoebox-sized instrument comprises a motor- chemiluminescence is measured as relative light units
ized PMT for chemiluminescent detection, an electronic (RLUs). This assay procedure, including determination
interface to the DropBot for droplet movement, and a of diluent compositions, reagent concentrations, incuba-
magnet assembly (called a magnetic lens) for particle sep- tion times, droplet operations, and chemiluminescent
aration (Fig. 1A). Movement of protein-containing signal generation, was developed from design of experi-
droplets through an air-filled device (Fig. 1B), which al- ment optimization (32 ). The use of chemiluminescence
lows for more efficient particle-concentration and ana- is particularly advantageous because few optical compo-
lyte washing than the more common methods relying on nents are required (e.g., no excitation source, filters, or
oil-filled devices (33 ), is facilitated by the inclusion of lenses).
Pluronic additives in samples and reagents, which re- The assay described here was designed for detection
duces protein adsorption and limits cross-contamination of RV in patient serum and plasma samples. Thus, to
(38 ). During particle separation, particles in the droplet optimize and calibrate the method, we used commercial
are immobilized on the surface in a region above the calibrators designed to simulate human serum (buffer
magnet, referred to as the separation zone (Fig. 1C; on- containing 4% BSA). As shown in the calibration curve
line Supplemental Video 1). Subsequently, the droplet is (Fig. 2B), a direct relationship exists between the RLUs
actuated away to waste, leaving behind a particle pellet. detected by the integrated PMT and the amount of RV
For particle resuspension, the magnetic lens is deacti- IgG in the calibrators. The calibration curve had a dy-
vated, and a new droplet is actuated over the particle namic range of approximately 2 orders of magnitude, the
pellet and mixed in a circular motion; this is sufficient to intraassay CVs ranged from 0.3% to 6.5%, and the LOD
break up the particle pellet and completely reconstitute and LOQ of the assay were 0.03 and 0.1 IU/mL, respec-
the particles in suspension (Fig. 1D). tively, below the clinical cutoff concentrations described
The DMF instrumentation and devices described above. Each set of 4 assays requires approximately 25 min
here are prototype designs used to demonstrate the diag- to complete.
nostic performance of DMF immunoassays. In ongoing In some instances, samples with high analyte con-
work, we are developing next-generation systems that centration must be diluted before analysis to facilitate

424 Clinical Chemistry 61:2 (2015)


Microuidic Rubella Assay

A
H2O2 (0.9 L)
Sample (1.8 L) Conjugate (1.8 L) Luminol (0.9 L)

Incubate (3 min) Incubate (2 min) Incubate (2 min)


Wash (4) Wash (4) Enzymatic
Rubella virus on chemiluminescent
magnetic particle Rubella virus IgG
Peroxidase-conjugated reaction
antihuman IgG

B 1.0 10 7 C 1.2 10 7
Standard P < 0.05*
Blood P < 0.05*
8.0 10 6
Signal (RLU)

Signal (RLU)
8.0 10 6 P = 0.08
6.0 10 6
P = 0.09
4.0 10 6
4.0 10 6

1/32 Blood

1/16 Blood
2.0 10 6

1/8 Blood

1/4 Blood
Diagnostic cutoff
(10 IU/mL)
0 0
4 8 16 32 64 128 256 15.6 31.3 62.5 125
Rubella virus IgG (IU/mL) Rubella virus IgG (IU/mL)

Fig. 2. RV IgG immunoassay.


(A), Scheme and procedure of RV IgG immunoassay on paramagnetic particles. (B), Calibration data (red diamonds) and curve (dotted line) for
RV IgG immunoassay implemented in the digital microuidic system. (C), Comparison of RV IgG immunoassay results for samples in calibrator
solution (gray) and blood (white) for dilutions formed on-chip. Each value represents the mean (SD) from 3 replicates. Statistical signicance
(*P < 0.05) was evaluated by Student 2-tailed t-test.

accurate quantification; this technique is relatively trivial sition, the signal intensities of the calibrator samples dif-
for central laboratories, but is not commonly used in fered significantly from the diluted samples containing
portable tests. Hence, we evaluated whether the DMF high blood content (i.e., P 0.049 for the 1/8 dilution
platform could reliably perform a range of dilutions on and P 0.050 for the 1/4 dilution). But when these
samples containing 250 IU/mL RV IgG, which was be- samples were diluted further (i.e., 1/16 and 1/32 dilu-
yond the linear range of the calibration curve (Fig. 2B). It tions of blood), the results for blood and calibrator sam-
is straightforward to dilute such samples on-chipfor ples could not be distinguished (P 0.05). These obser-
example, to form 4 droplets of 125, 62.5, 15.6, and 31.3 vations suggest that sample dilution may be useful for
IU/mL (see online Supplemental Fig. 2). Subsequently, alleviating interferences arising from the blood matrix for
RV IgG immunoassays can be performed on the diluted application in the field, provided that the dilution does
samples in parallel. not reduce the analyte concentration below the LOQ of
The assays described here were designed for applica- the assay.
tion to patient serum and plasma samples. But for porta-
ble analysis in the field, it would be useful if the assay RV IgM IMMUNOASSAY AND IgG BLOCKING
could also be used with whole blood, negating the re- Whereas testing for RV IgG (described above) is useful
quirement of phlebotomy (with finger stick) and frac- for determining susceptibility and immune status, the
tionation. To evaluate this possibility, simulated whole critical test for diagnosing rubella-infected patients (in-
blood samples (calibrator mixed with sheeps blood) were cluding the identification of pregnancies at risk for CRS)
diluted on-chip and tested with the methods described is the detection of RV IgM. Unlike RV IgG, there is no
above. As shown in Fig. 2C, the interassay CVs for these agreed cutoff concentration for RV IgM to distinguish
assays were only marginally higher in the blood samples between negative or positive test results. As such, assay
(10% to 15%) than in the control (standard) samples vendors develop their own cutoff calibrators on the basis
(6% to 9%). Because of dissimilarities in matrix compo- of serological testing of samples obtained from healthy

Clinical Chemistry 61:2 (2015) 425


A

Magnetic particle
Rubella virus (RV)
RV IgM
RV IgG
Detection anti-IgM
Rheumatoid factor IgM
1) RV IgM assay 2) False negative 3) False positive

B 25 C 1.2 10 6
No pretreatment
P = 0.0003*
20 Pretreatment

Signal (RLU)
8.0 10 5
Fold change

15
P = 0.6

10
4.0 10 5
P = 0.5
5

0 0
IgG IgM IgM + IgG 0 IgM cutoff

Fig. 3. RV IgM immunoassay.


(A), Scheme of RV IgM immunoassay (panel 1) and potential interferences or nonspecic interactions arising from the presence of RV IgG
(panel 2) and RF IgM (panel 3). (B), Bar graph illustrating the effect of on-chip sample pretreatment (gray bars) relative to no pretreatment
(white bars) in RV IgM immunoassays. (C), Diagnostic cutoff calibrator signal for RV IgM immunoassays implemented in digital microuidics
with sample pretreatment. Each value represents the mean (SD) from 3 4 replicates. Statistical signicance (*P < 0.05) was evaluated by
Student 2-tailed t-test.

and infected donors. In the assays reported here, the RV Complicating matters, vendor-specific calibrators, which
IgM Cal 1 reagent was adapted from the Abbott Archi- are formed from pooled human sera, often contain vary-
tect rubella IgM assay kit to define the cutoff between ing amounts of RF IgM or other cross-reactive species,
positive and negative/equivocal results with digital depending on the source. To ascertain the presence of RF
microfluidics. IgM in the calibrators used here, they were tested with an
Similar to the RV IgG immunoassays described off-chip ELISA kit (see online Supplemental Table 1)
above, the on-chip RV IgM assays relied on RV- the RV IgG calibrators have low concentrations of RF
immobilized paramagnetic particles to implement indi- IgM (1.4 1.85 IU/mL), and the RV IgM cutoff calibra-
rect ELISA. In an ideal assay, RV IgM antibodies from tor has high concentrations of RF IgM (27.0 IU/mL).
the sample will bind to the particles, and the detection Fortunately, errors related to the presence of endog-
antibodies (HRP-conjugated antihuman IgM) will bind enous RV IgG and RF IgM (Fig. 3A) can be alleviated by
to the RV IgM to form immunocomplexes (Fig. 3A, pretreating the sample with exposure to exogenous anti-
panel 1). However, there are at least 2 potential sources of human IgG before analysis, which serves to block un-
error in this assay scheme caused by the presence of RV wanted binding of RV IgG (and RV IgG/RF IgM com-
IgG and/or RF IgM in human serum (42 ). If the sample plexes) onto the virus-laden particles (4 6 ). Emulating
contains RV IgG, these antibodies will compete for bind- the Architect rubella IgM assay procedures, we developed
ing sites on the particles, leading to false-negative results an on-chip pretreatment method in which a droplet of
(Fig. 3A, panel 2). If the sample contains both RV IgG sample (1.8 L) is mixed with a droplet of pretreatment
and RF IgM, immunocomplexes will form on the parti- reagent (900 nL, containing goat antihuman IgG) for 7
cles, leading to false-positive results (Fig. 3A, panel 3). min before analysis by on-chip ELISA. To evaluate this

426 Clinical Chemistry 61:2 (2015)


Microuidic Rubella Assay

A 10 1.0
AUC = 1.00
C
Rubella virus IgG Rubella virus IgM

Sensitivity
DMF s/co Abbott EIA DMF s/co Abbott EIA
DMF rubella IgG (s/co)

0.5
Sample (result) s/co (result) (result) s/co (result)
0.0 PTR201-01 1.3 (pos) 1.1 (pos) 0.2 (neg) 0.2 (neg)
0.0 0.5 1.0
1 Specificity PTR201-02 2.0 (pos) 2.4 (pos) 0.7 (neg) 0.5 (neg)
1 PTR201-03 4.6 (pos) 3.1 (pos) 2.8 (pos) >2.3 (pos)
PTR201-04 2.2 (pos) 2.8 (pos) 0.1 (neg) 0.2 (neg)
PTR201-05 0.2 (neg) 0.1 (neg) 0.2 (neg) 0.1 (neg)
PTR201-06 1.2 (pos) 1.2 (pos) 0.2 (neg) 0.1 (neg)
PTR201-07 2.7 (pos) 2.5 (pos) 2.5 (pos) 1.2 (pos)
PTR201-08 2.0 (pos) 1.7 (pos) 0.1 (neg) 0.2 (neg)
0.1
Positive Negative PTR201-09 2.4 (pos) 2.7 (pos) 0.2 (neg) 0.1 (neg)
PTR201-10 1.7 (pos) 3.0 (pos) 0.1 (neg) 0.1 (neg)
Abbott EIA rubella IgG
PTR201-11 1.7 (pos) 2.7 (pos) 3.4 (pos) >2.3 (pos)
PTR201-12 3.1 (pos) 2.7 (pos) 0.5 (neg) 0.2 (neg)
B 10 1.0
AUC = 1.00 PTR201-13 2.4 (pos) 3.2 (pos) 0.2 (neg) 0.7 (neg)
Sensitivity
DMF rubella IgM (s/co)

0.5 PTR201-14 1.1 (pos) 2.4 (pos) 0.5 (neg) 0.1 (neg)
PTR201-15 2.7 (pos) 3.3 (pos) 0.4 (neg) 0.1 (neg)
0.0 PTR201-16 0.2 (neg) 0.1 (neg) 0.4 (neg) 0.2 (neg)
0.0 0.5 1.0
1 Specificity PTR201-17 2.0 (pos) 2.0 (pos) 0.5 (neg) 0.2 (neg)
1 PTR201-18 1.9 (pos) 1.1 (pos) 1.3 (pos) 0.9 (equ)
PTR201-19 2.2 (pos) 1.7 (pos) 1.7 (pos) 2.1 (pos)
PTR201-20 3.2 (pos) >3.4 (pos) 0.2 (neg) 0.1 (neg)
PTR201-21 2.7 (pos) 2.2 (pos) 0.3 (neg) 0.2 (neg)
PTR201-22 1.1 (pos) 1.4 (pos) 0.5 (neg) 0.1 (neg)
PTR201-23 1.6 (pos) 2.2 (pos) 0.3 (neg) 0.1 (neg)
0.1
PTR201-24 1.5 (pos) 1.3 (pos) 0.8 (equ) 1 (pos)
Positive Equivocal Negative
PTR201-25 1.8 (pos) 2.5 (pos) 0.2 (neg) 0.3 (neg)
Abbott EIA rubella IgM
Result differs from Abbott EIA (determined by supplier)

Fig. 4. Performance of RV IgG and RV IgM immunoassay on commercial plasma/serum panel.


Plots of s/co values for DMF RV IgG (A) and RV IgM (B) immunoassays from RV mixed-titer panel (Seracare PTR201). Each DMF s/co value
represents the mean from 3 4 replicates and is categorized as positive (pos), equivocal (equ), or negative (neg) on the basis of results from a
leading commercial RV antibody immunoassay (Abbott EIA) determined by the supplier. DMF s/co values in the shaded region were inter-
preted as equivocal. The inset plots show the ROC curves for DMF assays (red lines) and random guesses (dashed lines). The area under the ROC
curve (AUC) is reported in each plot. (C), Table comparing the results of the DMF and Abbott EIA assays. The s/co and interpretation of the Abbott
EIA assays were supplied by Seracare. Shaded box indicates discrepancy in DMF assay with Abbott EIA.

method, we used test samples (formed from calibrators) to implement and to establish cutoff values (Fig. 3C) in
containing RV IgG, RV IgM, or both RV IgG and RV digital microfluidics.
IgM. As expected, the IgG-alone and IgM-alone sample
results were unaffected by the pretreatment (Fig. 3B) EVALUATION OF DIAGNOSTIC ACCURACY
because these samples did not contain high concentra- The diagnostic accuracies of the digital microfluidic RV
tions of RF IgM and RV IgG (see online Supplemental IgG and IgM immunoassays were estimated with 25
Table 1). In contrast, the pretreatment significantly (P plasma/serum samples from a commercial rubella mixed-
0.0003) suppressed the nonspecific signal arising in the titer performance panel (Fig. 4). These are undiluted ali-
combined IgG plus IgM sample, which, as shown in on- quots of plasma or serum collected from individual do-
line Supplemental Table 1, contains substantial concen- nors between 1994 and 1995 by the vendor, used by
trations of RV IgG (250 IU/mL) and RF IgM (approxi- diagnostic laboratories and manufacturers to evaluate
mately 14 IU/mL). Thus, accurate diagnosis of rubella their rubella tests. For each sample, the vendor pro-
infection requires IgG blocking, which is straightforward vided reference results (s/co values and test interpreta-

Clinical Chemistry 61:2 (2015) 427


tion) from tests performed by the vendor and indepen- (44 ). This is not surprising, since DMF, like automated
dent reference laboratories (between 1995 and 1996) immunoassay platforms, is automated and software pro-
with several leading ELISA platforms at that time. In grammable, allowing for integration of operations such as
this study, the Abbott EIA test was chosen as the ref- aliquoting, dilution, or IgG blocking as needed in an
erence standard because test results were provided for immunoassay work flow. More importantly, when com-
both RV IgM and IgG. pared with the only rapid rubella IgM test that we are
In estimating diagnostic performance, all 25 samples aware of (Alere ImmunoComb Rubella IgM), for
were divided and tested for both RV IgG and RV IgM in which the manufacturer reports diagnostic sensitivity
3 4 replicates to ascertain assay reproducibility. Typi- and specificity [sensitivity/specificity 87%/99% (7 )], the
cally, 4 parallel immunoassays, requiring approximately DMF method has the added value of built-in, automated
25 (for RV IgG assays) or 35 (for RV IgM assays) min, replicates and digital readout (i.e., no user interpretation
were carried out in single (only RV IgG or RV IgM in of results). Of course, to further validate the diagnostic
quadruplicates) or duplex (both RV IgG and RV IgM in performance of the new method, additional testing must
duplicates) format. The mean intraassay CV (range) was be performed in countries such as Vietnam, where rubella
4.6% (0.4%17.5%) for RV IgG assays and 7.5% has not been introduced into the national immunization
(1.5%13.4%) for RV IgM assays. In the RV IgG assays program (45 ). Nevertheless, these initial results suggest
(Fig. 4A), the DMF tests identified 23 positives, 0 equi- that a portable DMF-based system, perhaps combined
vocal, and 2 negatives, which represents perfect agree- with inexpensive, paper devices (29 ), may represent a
ment with the Abbott EIA results. The diagnostic sen- useful new tool for identification of patients at risk for
sitivity and specificity (with 95% CIs) were 100% CRS in low-resource settings. More generally, we pro-
(98.6%100%) and 100% (95.5%100%), respec- pose that similar methods might be useful for distrib-
tively (see online Supplemental Table 2). In the RV uted diagnostics for a wide range of infectious diseases.
IgM assays (Fig. 4B), the DMF tests identified 5 pos-
itives, 1 equivocal, and 19 negatives, which represents
close agreement with the Abbott EIA results (2 mis-
matches: 1 DMF equivocal/Abbott positive and 1
DMF positive/Abbott equivocal). In the calculation of Author Contributions: All authors confirmed they have contributed to
the intellectual content of this paper and have met the following 3 require-
diagnostic performance, the equivocal results (a clas- ments: (a) significant contributions to the conception and design, acquisi-
sification not accounted for) were interpreted as pos- tion of data, or analysis and interpretation of data; (b) drafting or revising
itive, yielding a sensitivity and specificity (with 95% the article for intellectual content; and (c) final approval of the published
CIs) of 100% (97.4%100%) and 100% (98.4%100%), article.
respectively (see online Supplemental Table 3). Authors Disclosures or Potential Conflicts of Interest: Upon man-
The measured s/co for all 25 samples tested are uscript submission, all authors completed the author disclosure form. Dis-
shown in Fig. 4C. For the 2 mismatches in the RV IgM closures and/or potential conflicts of interest:
assay, the DMF values (s/co 0.9 and 1) were at or Employment or Leadership: None declared.
near the cutoff value (s/co 1). Thus, we hypothesize Consultant or Advisory Role: None declared.
that the discrepancy between the DMF and the stan- Stock Ownership: A. Fischer, Abbott.
dard results may be caused by a combination of assay Honoraria: None declared.
variability (CV 15%) and degradation of the 20- Research Funding: This study was supported by the Natural Sciences
and Engineering Research Council of Canada (NSERC) and Abbott
year-old samples. Regardless, these types of positive- Diagnostics. A. Ng, graduate fellowship from NSERC; A.R. Wheeler,
equivocal discrepancies represent a relatively minor Canada Research Chair Program.
problem, as in standard practice, all patients who test Expert Testimony: None declared.
either positive or equivocal are retested to ascertain Patents: None declared.
final diagnosis (43 ). Role of Sponsor: The funding organizations played no role in the
The performance of the new DMF test (Fig. 4) was design of study, choice of enrolled patients, review and interpretation of
similar to that of central laboratory based RV testing data, or preparation or approval of manuscript.

References
1. Gahr P, DeVries AS, Wallace G, Miller C, Kenyon C, 4. McKenzie KG, Laeur LK, Lutz BR, Yager P. Rapid protein 6. Meurman OH, Ziola BR. IgM-class rheumatoid factor in-
Sweet K, et al. An outbreak of measles in an undervac- depletion from complex samples using a bead-based terference in the solid-phase radioimmunoassay of
cinated community. Pediatrics 2014;134:e220 e8. microuidic device for the point of care. Lab Chip 2009; rubella-specic IgM antibodies. J Clin Pathol 1978;31:
2. WHO. WHO-recommended standards for surveillance 9:3543 8. 4837.
of selected vaccine-preventable diseases. Geneva, 5. Champsaur H, Fattal-German M, Arranhado R. Sensitiv- 7. Alere(TM). ImmunoComb rubella IgM. http://www.
Switzerland: World Health Organization, 2003. ity and specicity of viral immunoglobulin M determi- alere.com/ww/en/product-details/immunocomb-rubella-
3. WHO. Rubella. http://www.who.int/mediacentre/factsheets/ nation by indirect enzyme-linked immunosorbent as- igm.html (Accessed July 2014).
fs367/en/ (Accessed September 2013). say. J Clin Microbiol 1988;26:328 32. 8. Kiyaga C, Sendagire H, Joseph E, McConnell I, Grosz J,

428 Clinical Chemistry 61:2 (2015)


Microuidic Rubella Assay

Narayan V, et al. Ugandas new national laboratory sam- Spicar-Mihalic P, Singhal M, et al. Progress toward mul- form for magnetic-particle-based immunoassays with
ple transport system: a successful model for improving tiplexed sample-to-result detection in low resource set- optimization by design of experiments. Anal Chem
access to diagnostic services for early infant HIV diagno- tings using microuidic immunoassay cards. Lab Chip 2013;85:9638 46.
sis and other programs. PLoS One 2013;8:e78609. 2012;12:1119 27. 33. Ng AHC, Choi K, Luoma RP, Robinson JM, Wheeler AR.
9. Reef SE, Strebel P, Dabbagh A, Gacic-Dobo M, Cochi S. 21. Fridley GE, Holstein CA, Oza SB, Yager P. The evolution Digital microuidic magnetic separation for particle-
Progress toward control of rubella and prevention of of nitrocellulose as a material for bioassays. MRS Bull based Immunoassays. Anal Chem 2012;84:880512.
congenital rubella syndrome: worldwide, 2009. J Infect 2013;38:326 30. 34. Vergauwe N, Vermeir S, Wacker JB, Ceyssens F, Corna-
Dis 2011;204:S24 S7. 22. Apilux A, Ukita Y, Chikae M, Chailapakul O, Takamura Y. glia M, Puers R, et al. A highly efcient extraction pro-
10. Ng AHC, Uddayasankar U, Wheeler AR. Immunoassays Development of automated paper-based devices for se- tocol for magnetic particles on a digital microuidic
in microuidic systems. Anal Bioanal Chem 2010;397: quential multistep sandwich enzyme-linked immu- chip. Sens Actuator B-Chem 2014;196:28291.
9911007. nosorbent assays using inkjet printing. Lab Chip 2013; 35. Emani S, Sista R, Loyola H, Trenor CCI, Pamula VK,
11. Shriver-Lake L, Golden J, Bracaglia L, Ligler F. Simulta- 13:126 35. Emani SM. Novel microuidic platform for automated
neous assay for ten bacteria and toxins in spiked clinical 23. Lutz B, Liang T, Fu E, Ramachandran S, Kauffman P, lab-on-chip testing of hypercoagulability panel. Blood
samples using a microow cytometer. Anal Bioanal Yager P. Dissolvable uidic time delays for program- Coagul Fibrinolysis 2012;23:760 8.
Chem 2013;405:5611 4. ming multi-step assays in instrument-free paper diag- 36. Banoo S, Bell D, Bossuyt P, Herring A, Mabey D, Poole F,
12. Phurimsak C, Yildirim E, Tarn MD, Trietsch SJ, Hanke- nostics. Lab Chip 2013;13:2840 7. et al. Evaluation of diagnostic tests for infectious
meier T, Pamme N, Vulto P. Phaseguide assisted liquid 24. Whitesides GM. Viewpoint on Dissolvable uidic diseases: general principles. Nat Rev Microbiol 2008;
lamination for magnetic particle-based assays. Lab time delays for programming multi-step assays in 8:S16 S28.
Chip 2014;14:2334 43. instrument-free paper diagnostics. Lab Chip 2013;13: 37. WHO. Operational characteristics of commercially
13. Martinez AW, Phillips ST, Butte MJ, Whitesides GM. 4004 5. available assays to determine antibodies to HIV-1 and
Patterned paper as a platform for inexpensive, low- 25. Choi K, Ng AHC, Fobel R, Wheeler AR. Digital microu- or HIV-2 in human sera. Report 11. Geneva, 1999.
volume, portable bioassays. Angew Chem Int Ed 2007; idics. Annu Rev Anal Chem 2012;5:413 40. 38. Au SH, Kumar P, Wheeler AR. A new angle on pluronic
46:1318 20. 26. Sista RS, Eckhardt AE, Wang T, Graham C, Rouse JL, additives: advancing droplets and understanding in
14. Pollock NR, Rolland JP, Kumar S, Beattie PD, Jain S, Norton SM, et al. Digital microuidic platform for mul- digital microuidics. Langmuir 2011;27:8586 94.
Noubary F, et al. A paper-based multiplexed transami- tiplexing enzyme assays: implications for lysosomal 39. Fu E, Yager P, Floriano PN, Christodoulides N, McDevitt
nase test for low-cost, point-of-care liver function test- storage disease screening in newborns. Clin Chem JT. Perspective on diagnostics for global health. IEEE
ing. Sci Transl Med 2012;4:152ra29. 2011;57:1444 51. Pulse 2011;2:40 50.
15. Fu E, Liang T, Spicar-Mihalic P, Houghtaling J, Ram- 27. Srinivasan V, Pamula VK, Fair RB. An integrated digital 40. Dryden MD, Rackus DD, Shamsi MH, Wheeler AR. Inte-
achandran S, Yager P. Two-dimensional paper network microuidic lab-on-a-chip for clinical diagnostics on hu- grated digital microuidic platform for voltammetric
format that enables simple multistep assays for use in man physiological uids. Lab Chip 2004;4:310 5. analysis. Anal Chem 2013;85:8809 16.
low-resource settings in the context of malaria antigen 28. Cho SK, Moon H, Kim C-J. Creating, transporting, cut- 41. Shamsi MH, Choi K, Ng AHC, Wheeler AR. A digital mi-
detection. Anal Chem 2012;84:4574 9. ting, and merging liquid droplets by electrowetting- crouidic electrochemical immunoassay. Lab Chip
16. Tekin HC, Gijs MAM. Ultrasensitive protein detection: based actuation for digital microuidic circuits. J Micro- 2014;14:54754.
a case for microuidic magnetic bead-based assays. Lab electromech Syst 2003;12:70 80. 42. Wilson D. Rheumatoid factors in patients with rheuma-
Chip 2013;13:471139. 29. Fobel R, Kirby AE, Ng AHC, Farnood RR, Wheeler AR. toid arthritis. Can Fam Phys 2006;52:180 1.
17. Chin CD, Cheung YK, Laksanasopin T, Modena MM, Paper microuidics goes digital. Adv Mater 2014;26: 43. Tipples GA. Rubella diagnostic issues in Canada. J In-
Chin SY, Sridhara AA, et al. Mobile device for disease 2838 43. fect Dis 2011;204:S659 S63.
diagnosis and data tracking in resource-limited set- 30. Fobel R, Fobel C, Wheeler AR. DropBot: an open-source 44. Dimech W, Panagiotopoulos L, Francis B, Laven N, Mar-
tings. Clin Chem 2013;59:629 40. digital microuidic control system with precise control ler J, Dickeson D, et al. Evaluation of eight anti-rubella
18. Chin CD, Laksanasopin T, Cheung YK, Steinmiller D, of electrostatic driving force and instantaneous drop virus immunoglobulin G immunoassays that report re-
Linder V, Parsa H, et al. Microuidics-based diagnostics velocity measurement. Appl Phys Lett 2013;102: sults in international units per milliliter. J Clin Micro-
of infectious diseases in the developing world. Nat Med 193513. biol 2008;46:1955 60.
2011;17:10159. 31. Peng C, Zhang Z, Kim CJ, Ju YS. EWOD (electrowetting 45. Pham VH, Nguyen TV, Nguyen TTT, Dang LD, Hoang NH,
19. Jokerst JV, McDevitt JT. Programmable nano-bio- on dielectric) digital microuidics powered by nger Nguyen TV, Abe K. Rubella epidemic in Vietnam: char-
chips: multifunctional clinical tools for use at the point- actuation. Lab Chip 2014;14:111722. acteristic of rubella virus genes from pregnant women
of-care. Nanomedicine 2009;5:14355. 32. Choi K, Ng AHC, Fobel R, Chang-Yen DA, Yarnell LE, and their fetuses/newborns with congenital rubella
20. Laeur L, Stevens D, McKenzie K, Ramachandran S, Pearson EL, et al. Automated digital microuidic plat- syndrome. J Clin Virol 2013;57:152 6.

Clinical Chemistry 61:2 (2015) 429

You might also like