You are on page 1of 5

2010 International Conference on Industrial Electronics, Control and Robotics

Power Quality Issues With Medical


Electronics Equipment In Hospitals
Uma Rao, S.N.Singh Chandan Kumar Thakur
Department of Electronics and Communication Engineering
National Institute of Technology, Jamshedpur (India) - 831014
imumarao@gmail.com , snsnitjsr@gmail.com ex.chandu@gmail.com

Abstract- This paper describes power quality audits grounding, high wattage operating equipment,
conducted at hospitals in India. Voltage ,current and testing of emergency generators, physical plant
power measured at various points were used to assess renovation. Some equipment such as X-ray,
the quality of the power. The main power quality computerized axial tomography and magnetic
problems found in hospitals were voltage flicker, resonance imaging are considered as polluting ones
neutral currents , and harmonic distortion (THD
)value . Measurement, analysis and conditioning of due to their high amount of harmonics and voltage
power quality has been studied in few different transients produced on the electrical power supply
hospitals as a case study. Recommendation have been when in operation, while other equipment,
made to mitigate the issue
. connected to the same power supply, are considered
sensitive or victim ones, presenting failure or
Keywords - Power quality, harmonic distortion, voltage producing wrong. In order to deal with these
disturbances, Flickering, Neutral Current, Total problems, the electrical system should be improved
Harmonic Distortions. as the number of sensitive, nonlinear loads at a
hospital increases. However, many buildings were
not required to implement system improvements
1. INTRODUCTION suggested in new or revised regulations.
Power quality is the term used for overall voltage Compounding the problem is the fact that many
quality, current quality and nature of output codes or regulations deal with safety issues, but
waveform which directly or indirectly affect the they do not address power quality concerns .
loads as well as source. Power quality has been a Frequently, large-demand devices as x-ray machies
major issue in todays industrial, commercial, are installed without a complete power quality
residential and military and medical environments. survey. Since power quality problems are
Poor power quality may lead to disastrous cumulative; small power quality events (detectable
consequences like break-downs, production in an audit) can lead to loss-of-life and eventually
interruptions, excess energy consumptions etc. In premature equipment failure.
this paper power quality issues have been discussed
in the hospitals where these issues are becoming DESCRIPTION OF THE SYSTEM
more important due to greater impact on sensitive The first step of the power quality audit was an
medical equipment which utilizes microprocessor inspection of the hospitals electrical system. Most
controls and switched DC power supplies. Medical of the Hospitals receive their electric power from:
devices driven by electric power have come to be Normal Supply (NS) is a direct utility
commonly used in hospitals and rapid changes of (Western Power) supply for non-essential
voltage or current can easily cause them to fail or areas such.
does not respond as per the desired control action. Essential Supply (ES) is used for areas of
Externally and internally induced power surges and medical importance that are not critical to
harmonics can have a dramatic effect on equipment patients in the case of supply interruption.
performance. A stable and high quality power The essential supply is backed up by an
supply is indispensable in order to maintain safety in emergency generator.
the modern clinical setting. The widespread use of Uninterruptible Power Supply (UPS) is
sensitive, microprocessor /microcontroller based used for emergency lighting, operating
equipment at hospitals requires that the power theatres, patient monitoring and other
delivered to such sensitive facilities be of a higher equipment that is important to the well
quality [1]. Power quality problems may be caused being and safety of patients.
by generation of harmonics by equipment itself,
external sources like lightning or internal sources
like interaction of loads.
Common sources of power quality problems found
in hospitals include : inadequate wiring and

 978-1-4244-8546-8/10/$26.00 2010 IEEE 34


operate (termed as the fundamental frequency;
usually 50 Hz or 60 Hz).They can cause overheating
of rotating equipment, transformers, and current-
carrying conductors, premature failure or operation
of protective devices (such as fuses), harmonic
resonance, blinking of incandescent lights,
Transformer Saturation, Capacitor Fai lure
Harmonic Resonance, Circuit Breakers Tripping
Inductive Heating and Overload, Computer
Malfunction or Lockup - Voltage Distortion .
2. POWER QUALITY (D. Voltage Fluctuations
The definition of power quality given in the IEEE
dictionary [5] originates in IEEE Std 1100 (better Voltage fluctuations are systematic variations of the
known as the Emerald Book) [78]: Power quality is voltage envelope or a series of random voltage
the concept of powering and grounding sensitive changes.
equipment in a matter that is suitable to the This may cause Flickering of incandescent lamps,
operation of that equipment. High power qualities System halts.
implies safety, reliable service and lower initial and
operational costs. Power quality is concerned with (E). Frequency Fluctuations
deviations of the voltage from its ideal waveform
(voltage quality) and deviations of the current from The power system frequency is directly related to
its ideal waveform (current quality). Such a the rotational speed of the generators on the system.
deviation is called a "power quality phenomenon" or At any instant, the frequency depends on the balance
a "power quality disturbance." Some of the power between the load and the capacity of the available
quality disturbance are generation. When this dynamic balance changes,
small changes in frequency occur. This may cause
(A)Transients synchronous equipment failure.
Transients or surges are defined as a transient wave 3. MATERIALS AND METHODOLOGY
of current, voltage or power in an electric circuit.
Voltage transients may cause a microprocessor to The existing standards do not consider the voltage
read voltage levels incorrectly, resulting in incorrect harmonic and sags in the medical electrical
data processing (ones being read like zeros) or equipment.
altered stored data/setting. This can also cause loss The methodology applied consisted of taking
(or corruption) of data , Physical damage of measurements of parameters regarding of electrical
Electronic equipments like PCB cards, SMPS and power quality in health care facilities. Particular
system halts . attention was paid to the voltage harmonic content
and to momentary voltage transients, especially
(B) Interruption sags. The electrical current flow needed to operate
three types of medical equipment usually found at
Interruption is complete loss of supply voltage or typical health care facilities are measured during
load current, depending on its duration. This may their operations. The purpose is to show the
cause equipment damage, ruination of product, the polluting characteristics of X-ray, Computerized
cost associated with downtime, cleanup, and restart. Axial Tomography and Magnetic Resonance
Imaging equipments.
(C) Sag/Swell
4. CASE STUDY
Sag is a reduction of AC voltage at a given
frequency for the duration of 0.5 cycles to 1 Study was carried out on power quality of few
minutes time. Swell is the reverse form of sag, instruments commonly used in nursing home and
having an increase in AC voltage for duration of 0.5 hospital.
cycles to 1 minutes time. They can cause equipment X-ray equipment current consumption
damage, data corruption, errors in industrial measurements
processing, data errors, flickering of lights, The X-ray equipment in the operating theatres at the
degradation of electrical contacts, semiconductor hospital is supposed to be normally run from the
damage in electronics, Insulation degradation. UPS system. However, due to light flickering while
being supplied by the UPS, the X-ray machine was
(D) Harmonics supplied by an essential power supply instead.
Hence in a blackout situation the X-ray equipment
Harmonics are sinusoidal voltages or currents would loose power until the start-up of the diesel
having frequencies that are integer multiples of the generator. During the two minute change over the
frequency at which the supply system is designed to operating theatre staffs are unable to monitor

35
2010 International Conference on Industrial Electronics, Control and Robotics

patients condition. The neutral current on the


essential supply is found to be large and have very
high total harmonic distortions. The X-ray machine
operates on two modes:
- Screening: This mode is a continuous mode of
operation for patient monitoring purposes. This
mode produces relatively small amount of X-raying
energy compared to the next mode.
- Snapshots: This mode is monitoring and produces
high levels of X-raying energy but for a period of up
to 20-30 seconds so there is a high demand for
electrical power and temporary voltage sags can
occur, compromising the performance of sensitive
equipment connected to the power supply. Longer
periods could cause excessive heating and damage Fig. 3: Current peak caused by magnetic resonance
the X-ray tube. imaging during its operation.
For both modes, the X-raying energy is adjustable.
This is an important observation as it is crucial to
maintain similar conditions between different Heart monitor test results: The heart monitor is
used for continual monitoring of a patient in surgical
measurements for valid comparisons. 
centres and intensive care units producing
electrocardiograms as well as for monitoring
The use of X-ray fluoroscopy has increased temperature, breathing and other patient data. In
dramatically in recent years and is spreading beyond addition to the waveform in the electrocardiogram, it
the radiology department where users traditionally can indicate the heart rate and provide important
have extensive fluoroscopy training. Figure 2 shows information like oxygen saturation levels in arterial
potential for excessive radiation exposure to blood (SpO2) and arterial pressure. The equipment
personnel and patients. used in the tests performs those measurements in a
non-invasive manner. A patient simulator was used
to generate voltage pulses on a steady and regular
basis, similar to those generated by a heart. The
equipment tested has an internal battery for
operation in cases of energy outages. In all tests, an
110 V voltage powered it. The maximum value
obtained for the third harmonic voltage during
measurements in several hospitals was 3%, thus the
values used during tests were far superior, with the
objective to verify the behaviour of the equipment
under conditions that can be far more critical. This
equipment keeps a steady operation when subjected
to voltages with high harmonic content and critical
momentary voltage sags.
Pulmonary ventilator test results: The ventilator
Fig.2: Peaks caused by X-ray equipment during its type tested is completely microprocessor based and
operation. designed to be employed in respiratory insufficiency
in pediatric and adult patients, with a body mass
Computerized axial tomography current between 6 and 150kg. It is very practical equipment,
consumption measurements with direct digital controls for the main ventilator
parameters. The ventilation cycle starts by means of
It is one of the most reliable and safe methods of pressure and since it incorporates a ventilation
examination in modern electrical signals are emitted monitor, it allows for an extensive monitoring of
by the atomic nuclei of body tissues. A circular patients conditions, increasing safety. Besides, an
antenna surrounding the patient intercepts these audiovisual alarm system with written messages on
signs. The intensity of the signal varies according to the control screen gives quick alarm identification.
the kind of tissue. A computer assigns each signal to The presence of lungs was simulated through the use
a point corresponding to the body area being of a rubber balloon, provided by the manufacturer
examined and turns them into an image on the for this purpose. alteration in current waveform
screen. The current peak during its employment is between sine voltage and the supply with completely
shown in figure distorted voltage, no alterations in the regularity of
the equipments ventilation cycles could be noticed.

36
amplitude of current is almost twice in this case, but
the distortion is very different. The current
magnitude of the third harmonic is slightly reduced
as compared to idle operation, and its percentage in
the spectrum is significantly reduced since the total
current magnitude has doubled.
Total harmonic distortion THD is the most
frequently used power quality index. Most of this
equipment produces large third harmonic current
and hence the violation of the third harmonic limit.
This also explains why the process of x-raying did
not have much effect on the THD of the neutral
current[4]. The THDIs
on the ES and UPS supply were measured and found
to be 4.1% and 3.3% respectively. The r.m.s
voltage of the ES supply is also 4 to 5 V higher than
the UPS supply. Figure 5 shows the neutral current
and its spectrum. The current waveform is severely
distorted with a THD of 38.16% without x-raying
from the ES supply.
Another important outcome of the power quality
Fig 4: Voltage sag sequence generation to simulate audit is in regard to the condition of the electrical
test conditions of the Table 4 system. Installation of new equipment did not go in-
hand with improvements in infrastructure. As in
5. OBSERVATIONS most facilities changes occur at a fast pace, new
equipment is added almost daily, e.g., office
There is a lack of standard related to test cases equipment, diagnosis equipment, and patient
considering the effects of high harmonic and voltage treatment equipment. In the past a single power
sags. The quality of electrical installation in health outlet per bed in patient rooms was more than
care facilities needs to be improved to reduce risks enough. Today two or three outlets are insufficient.
to the patients and to provide better health treatment. Unfortunately, services are added most of the time
Figure 5 shows the lamp r.m.s voltage over a 4 without improving feeders, panel board capacity,
minute interval. In this period two x-raying circuit protection and infrastructure in general.
snapshots were taken as shown in the Figure. The Therefore, a system that was once stiff is now
variation of peak voltage was measured to be equal operating near its limits. This change in operating
to 40 V between two successive period. Figure 5 point has already caused malfunctions in equipment
clearly indicates severe voltage fluctuations while x- such as patients intravenous pumps (used to
raying using the UPS system. However, with the ES administer serum and other medicines to patients).
supply no voltage fluctuations were found. Overall system efficiency may be improved,
resulting on less repairs, reduced loss-of-life and
better maintenance records. This kind of study
would give an ability to identify and prevent power
quality events. This study will also serve as a guide
for installations of new equipment and future
expansions in the hospital. The study is especially
important for the BVH since its electrical installation
is not suited to deal with many power quality events.
Further analytical investigation of harmonic
propagation and unbalance conditions is required.
As a first step, the grounding bonding in the
facilities is being studied so that existing and future
problems are mitigated.
Surgery suits equipment includes the surgery In most hospitals, the latest equipment is controlled
luminaries. These luminaries were experiencing by computers, which require uninterrupted power
frequent burn out. This problem was identified supply for proper functioning. Even though hospitals
thorough the audit. The solution was to use a circuit rely on the power supplied by the government, they
with less harmonic distortion. The current distortion do have generators as back up during power-cuts.
resembles a step square wave. The steps follow a The generators use diesel, which is costlier than
sinusoidal wave. This behaviour agree with the way electricity. Hence UPS have to be used to their
this machine operates, were continuous x-ray maximum capacity to control wastage. With the
"shoots" draw high current in steps. As with idle conventional resources depleting day-by-day, the
operation, there is minimal voltage distortion. The need to bank on alternate energy sources is
unavoidable. The common resource of all is solar

37
2010 International Conference on Industrial Electronics, Control and Robotics

energy, which is in abundance in a tropical country


like India.

6. RECOMENDATIONS
Design of an additional low cost filter unit and use
of intelligent UPS system with its implementation at
various power supply outlets will ensures power
quality enhancements, RMS/power current level
reduction, efficient power/energy utilization and
effective demand side management. Source current
harmonic will also be reduced by the dynamic
controller power quality enhancement. This
translates into improved power factor and energy
saving. The quality of electrical installation in health
care facilities needs to be improved to reduce risks
to the patients and to provide better health treatment.
REFERENCES
[1] D. Stymiest, W. Henry, Power Prognosis: Surgery
Required!, Consulting-Specifying Engineer, no. 3, September
1998, pp. 71 76.
[2] Power Quality Analyzer-Applications Guide, Rev. 2, Fluke
Corporation, Everett, WA, June 1998.
[3] G. T. Heydt, Electric Power Quality, 2nd ed., Stars in a Circle
Publications, Scottsdale, 1994.
[4] Recommended Practices and Requirements for Harmonic
Control in Electrical Power Systems, IEEE Standard 519, NY,
1992.
[5] Dougherty, J.G. and W.L. Stebbins, 1997. Power quality: A
utility and industry perspective. Proceeding of the Conference on
Textile, Fiber Film Industry Technical, May 6-8, IEEE Computer
Society, Washington DC., USA., pp: 1-10. DOI:
10.1109/TEXCON.1997.598528
[6] Sersen, E. and J. Vorsic, 2008. Quality of Electricity Supply
as a Service. The Energy Agency of the Republic of Slovenia.
IEC Std. 60601, 2008. Electrical safety testing for medical
equipment.
[7] Hoevenaars , T. et al., 2003. Interpreting IEEE Std
519 and meeting its harmonic limits in VFD applications.

38

You might also like