Manual number: GP0001i0 (vers. 0.4)
CONTENTS
1. CHARACTERISTICS ....................................................................................................................3
1 .1 . Mechanical features: ............................................................................................................ 3
1 .2 . Electrical features:................................................................................................................ 3
1 .3 . Electronic features: .............................................................................................................. 3
1 .4 . Indications on display .......................................................................................................... 3
1 .5 . Operating features: ............................................................................................................... 4
2. DESCRIPTION OF OPERATING FEATURES..................................................................4
2 .1 . KEYS FUNCTIONS ............................................................................................................ 4
3. USERS FUNCTIONS .............................................................................................................5
4. USERS SETUP........................................................................................................................5
5. ADVANCED CONFIGURATION (FOR TECHNICIANS ONLY) .............................................8
6. SERIAL CONNECTION (OPTIONAL)..................................................................................10
7. EXTERNAL LOAD CELL (OPTIONAL) ..............................................................................10
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 2 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
1. Characteristics
Gpp 01 Palm-size instrument purposely studied to radio receive the weight from a
transmitting board or process the weight of a load cell lodged within the container.
This instrument allows the weight to be calibrated and sent along a serial line.
Plastic housing
Alloy steel load cell
Mechanical resistance of the cell to overload without alterations, higher than 200% the end-
scale value.
Safety level greater than 5
Stable weight: it is signalled by the indication of measurement unit lighting up: if the
indication is off, the weight is unstable, if it is on, the weight is stable.
Battery: the state of the battery is signalled by the indicator in the form of a battery. The
notches on the indicator mark the state
Net: the net weight is indicated by the message NET.
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 3 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
KEY ON/OFF (PN) it will switch on the display backlighting for 10 seconds (on request).
KEY ON/OFF (PT) it allows the instrument to be switched on and off.
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 4 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
3. USERS FUNCTIONS
Digital filter: if the function is enabled, an oversized weight-stabilizing filter is inserted.
Automatic zero on switching on the unit: if the function is enabled, any time the
instrument is switched on, it reaches the zero state by cancelling small zero differences.
This function can be used to zeroize chains, accessories, etc.
Be careful. It will cancel all the weight that is present on switching on the unit, we suggest
you do not use it in the presence of automatic switch-off.
Instrumental zero: it allows the dynamometer zero to be performed at any time without
damaging the weight calibration. Unlike the zero performed by using the Tare key or the
remote control, this key will remain stored even on switching off the unit.
Switch off: if the function is enabled and the dynamometer remains unused, for the set time
and with stable weight, the automatic switching off will take place.
Rate: Setting of the display update frequency (rate) (This setting can vary the batterys life).
Codes: In this section, advanced settings can be made.
4. USERS SETUP
To gain access tot he instruments settings, simultaneously press the keys MODE and
TARE until the text Filt is displayed.
OPERATING PROCEDURE
1 FILt The display unit will show FILt, which corresponds to the
selectable digital filter
MODE Press MODE to move to the following parameter.(2)
MODE Press MODE for a long time to go back to the weighing state
TARE Press TARE to gain access to the digital filter selection, then
MODE Press MODE to change over between F no and F SI where:
F no standard filter
F SI oversized filter.
TARE Press TARE to gain access to the selection
TARE Set with TARE a filter value 1 to 9
MODE Press MODE for a long time to store the selected value and go
back to the settings selection mode.
2 AUt 0 The display unit will show Aut 0, which corresponds to the
automatic zero on switching on the unit.
MODE Press MODE to move to the following parameter.(3)
MODE Press MODE for a long time to go back to the weighing state
TARE Press TARE to gain access to the automatic zero selection, then
MODE Press MODE to change over between A no and A SI where:
A no automatic zero inhibited
A SI automatic zero enabled.
TARE Press TARE to enter the selection
MODE Set with MODE and TARE a weight value within which the
automatic zero will be performed.
MODE Press MODE for a long time to store the selected value and go
back to the settings selection mode.
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 5 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
3 ZEro The display unit will show -ZEro-, which corresponds to the zero
setting.
4 SPEGn The display will show SPEGn which corresponds to the time-
based switch-off
MODE Press MODE to move to the following parameter.(5)
MODE Press MODE for a long time to go back to the weighing state.
6 RATE The display unit will show RATE, which corresponds to the
selection of the display unit updating rate.
MODE Press MODE to move to the following parameter.(7)
MODE Press MODE for a long time to go back to the weighing state
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 6 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
Press TARE to store the selected value and go back to the settings
TARE selection menu.
7 CODE The display unit will show CODE which corresponds to the
advanced configuration section.
MODE Press MODE to go back to the first parameter.(1)
MODE Press MODE for a long time to go back to the weighing state
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 7 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 8 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 9 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 10 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
N.B.C. Elettronica
GPP 01
STATEMENT OF COMPLIANCE
This unit is in compliance with the Electromagnetic Compatibility requirements set forth in
the following Directives:
89/336/EEC dated 3 May 1989 and later modifications (92/31/EEC dated 28 April 1992
and 93/68/EEC dated 22 July 1993)
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 11 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it
Manual Number: GP0001i0 Vers. 0.4
INFORMAZIONI AGLI
UTENTI
NBC elettronica srl Via Tavani 2/C 23014 Delebio (So) page 12 of 12
Phone 0342682149 Fax 0342682150 E-mail: nbc@nbc-el.it http://www.nbc-el.it