(1) EC TYPE-EXAMINATION CERTIFICATE
(2) Equipment ar protective system intended for use in potentially explosive
atmospheres - Directive 94/9/EC
(3) _EC-Type Examination Certificate Number
TUV 01 ATEX 1766 X
(4) Equipment: Flow meter type FLOWSIC 600
(5) Manufacturer: SICK Engineering GmbH
(6) Address: 10-01458 Ottendorf-Okrilla, Bergener Ring 43
(7) _ This equipment or protective system and any acceptable variation thereto is specified in the
schedule to this certificate and the documents therein referred to.
(8) The TUV. HannoveriSachsen-Anhalt eV. TUV CERT-Certification Body, notified body
number N° 0032 in accordance with Article 9 of the Council Directive of the EC of March 23,
1994 (94/9/EC), certifies that this equipment or protective system has been found to comply
with the Essential Health and Safety Requirements relating to the design and construction of
equipment and protective systems intended for use in potentially explosive atmospheres
given in Annex II to the Directive.
‘The examination and test results are recorded in the confidential report N° 01 YEX 133086
(9) Compliance with the Essential Health and Safety Requirements has been assured by
compliance with:
EN 50 014:1997 EN 50 018:1994 EN 50 019:2000 EN 50 020:1994
(10) If the sign "X" is placed after the certificate number, it indicates that the equipment or
protective system is subject to special conditions for safe use specified in the schedule to
this certificate.
(11) This EC-type examination certificate relates only to the design and construction of the
Specified equipment or protective system according to Directive 94/9/EC, Further
Fequirements of this Directive apply to the manufacture and placing on the market of this
‘equipment or protective system.
(12) The marking of the equipment or protective system must include the following:
® 26 eexdeiiaic 4
‘TOV Hannover/Sachsen-Anhalt eV. Hanover, 2001-11-14
‘TOV CERT-Zertifsierungsstelle
am TOV 4
1D-20519 Hannover
Yk vot
Headot ne
Centeton Body
‘This certificate may only be reproduced without any change, schedule included.
Excorpts or changes shall be allowed by the TUV Hannover/Sacheen-Anhalt eV page 14(13)
ay,
SCHEDULE
(14) EC-TYPE EXAMINATION CERTIFICATE N° TUV 01 ATEX 1766 X
(15) Description of equipment
The flow meter type FLOWSIC 600 is intended for the contactless measurement of the
volumetric flow of gas.
The maximum permissible ambient temperature is 60°C. The maximum permissible media
temperature in dependence on the temperature class has to be taken from the following table:
Electrical Data
[Temperature class temperature |
aa
13.
i 12
Allnon intrinsically circuits are realised in the type of protection Increased Safety".
Supply circuit
(terminals 1+, 2-)
Field connections
Current output
(terminals 31,32)
Switched outputs
(terminals 41, 42; 51, 52; 81, 82)
Data interface
(terminals 33, 34)
Intrinsically safe version
Supply circuit
(terminals 1+, 2-)
12...82 V dc. resp, 85...230V acc.
253
Un = 253V
active passive
18V, Ip = 35 mA 30 V, Is = 35 mA
Us = 30 V, I = 100 mA
5V, Ip = 175 mA
in type of protection ,Intrinsic Safety" EEx ia IIC/EEx ib IIA
only for the connection to certified intrinsically safe
circuits
Maximum values: U;= 16 V
k =200 mA
B= 26w
effective internal capacitance 4 nF
The effective internal inductance is negligibly small.
age 24“oy,
Schedule EC-Type Examination Certificate N° TUV 01 ATEX 1766 X
Field connections
Current output in type of protection ,Intrinsic Safety” EEx ia IIC/EEx ib IIA
(terminals 31, 32) The maximum values in dependences on the type of
protection have to be taken from the following table:
EEx ib iA
U 22.4V 224V
i 194 mA 155 mA
P. 1072 mW 857 mw
[C. | 4100 nF 163 nF
bo 7 mH. mH
Characteristic line: linear
or in type of protection ,Intrinsic Safety’ EEx ia 1IC/EEx ib IIA
for the connection to certified intrinsically safe circuits
Maximum values: U;= 30V
k= 100 ma
P, = 750 mw
effective internal capacitance 4 nF
The effective internal inductance is negligibly small
‘Switched outputs in type of protection ,Intrinsic Safety" EEx ia IIC/EEx ib IA
(terminals 41,42; §1, 52; 81,82) _for the connection to certified intrinsically safe circuits
Maximum values: U;= 30 V
= 100 ma
, = 750 mw
effective internal capacitance 4 nF
The effective intemal inductance is negligibly small
Data interface in type of protection .Intrinsic Safety" EEx ia IICIEEx ib 11A
(terminals 33, 34) The maximum values in dependences on the type of
Protection have to be taken from the following table:
EEx ib lA EEx ia iC
US 5,88 V 5,88 V
a 313 mA 313 mA
Po 460 mW 460 mW.
[c,. 41000 uF 43 uF
Le 1,5 mH 0,2mH
Characteristic line: linear
page 34Ty,
Schedule EC-Type Examination Certificate N° TUV 01 ATEX 1766 X
or in type of protection , intrinsic Safety” EEx ia IIC/EEx ib WA
for the connection to certified intrinsically safe circuits
Maximum values: U,= 10 V
P, =1420 mw.
The effective intemal inductance and capacitance are
negligibly small.
Probe circuits in type of protection ,Intrinsic Safety" EEx ia IIC/EEx ib IIA
(probe connector 1...8) The maximum values in dependence on the type of
Protection have to be taken from the following table:
LC EEx ib HA EExialiC |
Us 60,8V 38,9V |
| 95 mA ‘60 mA
Py] 1444 mw 584 mW
Cy 300nF | __34nF
Ca mH mH
Characteristic line: linear
Routine test
In the case of the use in the extended ambient temperature range of -40°C to + 60°C the
overpressure test has to be carried out at every enclosure with 18.5 bar according to EN 50018,
Paragraph 15.1.3.1, as routine test.
(16) Test documents are listed in the test report No.: 01 YEX 133086.
(17) Special conditions for safe use
The combination of intrinsically safe and non intrinsically safe cirouits at the field connections is
not permitted.
(18) Essential Health and Safety Requirements
no additional ones
page 4/4(16)
(17)
(18)
a
Translation
2. SUPPLEMENT to
EC TYPE-EXAMINATION CERTIFICATE No. TUV 01 ATEX 1766 X
of the company: SICK Engineering GmbH
Bergener Ring 43,
1D-01458 Ottendorf-Okrilla
In the future, the flow meter type FLOWSIC 600 may also be manufactured and operated
according to the test documents listed in the test report.
‘The amendments concern the marking. The marking will be in the future.
11/2 G EEx de ib fia] lc Ts
The ultrasonic sensors of the flow meter must only be operated in area that require apparatus
of category when atmospheric conditions are present (temperature from -20°C to +60°C.
pressure from 0.8 bar to 1.1 bar),
|n case that the explosive atmosphere does not require category-1-apparatus, the maximum
permissible media temperature in dependence on the temperature class has to be taken from
the table:
(Temperature class_| Media temperature
Ta 103"
_ 3 i= 168°C.
72 200°C
The electrical data, the special condition for safe use and all other data apply unchanged for
this supplement.
Test documents are jisted in the test report N° 03 YEX 550772.
Special conditions for safe use
The combination of intrinsically safe and non-intrinsically safe circuits at the field connections is
not permitted,
Essential Health and Safety Requirements
no additional ones
‘9von cant cniteco ne Hermon 2084724
eee eran
wor
Se tmen
Si vabh
Sei aay
page 1Ty,
Translation
3. SUPPLEMENT to
EC TYPE-EXAMINATION CERTIFICATE No. TUV 03 ATEX 1766 X
of the company: SICK Engineering GmbH
Bergener Ring 43
D-01458 Ottendorf-Okrilla
In the future, the flow meter type FLOWSIC 600 may also be manufactured and operated
according to the test documents listed in the test report.
‘The amendments concem the internal design and the electrical data.
The marking is: 1/2G EEx de ib [ia] HC T4
The ultrasonic sensors of the flow meter must only be operated in areas that require apparatus
of category 1 when atmospheric conditions are present (temperature from -20°C to +60°C,
pressure from 0.8 bar to 1.1 bar).
In case that the explosive atmosphere does not require category-1-apparatus, the maximum
Permissible media temperature in dependence on the temperature class has to be taken from
the table:
Temperature class_| Media temperature
14 403°C
73 168°C
12 200°C
Electrical data
All non-intrinsically safe connections are realised in the type of protection Increased Safety.
Supply circuit U_ = 12.82 V dc, resp. 85...230V a.c.
(terminaist+, 2-) Un = 253V
Field connections Un = 253V
Current output active passive
(terminals 31,32) Us = 18, lp = 35. mA Us = 30 V, Ip = 35 mA
‘Switched outputs Us = 30 V, la = 100 mA
(terminals 41, 42; 51, 52; 81, 82)
Data interface Us=5V, le = 175 mA
(terminals 33, 34)tM
3. Supplement to EC Type-Exai ion Certificate No. TUV 01 ATEX 1766 X
intrinsically safe version
Supply circuit in type of protection Intrinsic Safety EEx ia IIC/EEx ib IIA
{terminals 1+, 2-) only for the connection to certified intrinsically safe
circuits
Maximum values: U;= 16V
Pe 26W
effective intemal capacitance 4 nF
The effective internal inductance is negligibly small
Field connections
Voltage-/current output in type of protection Intrinsic Safety EEx ia IIC/EEx ib 1A
(terminals 31, 32) ‘The maximum values in dependence on the type of
protection have to be taken from the following table:
EEx ib ilA | EEx ia lic
["U,[ aa4Vv— | 22a
| 155mA_| 155 mA
P, | 857 mW | 857 mW
C. | 4100 nF | 163 nF
ei 7 mH. mH
Characteristic line: linear
or in type of protection Intrinsic Safety EEx ia IIC/EEx ib 1A
only for the connection to certified intrinsically safe
circuits
Maximum values: U;= 30V
= 100mA
P= 750 mW
effective internal capacitance 4 nF
The effective internal inductance is negligibly small
‘Switched outputs in type of protection Intrinsic Safety EEx ia IIC/EEx ib IA.
(terminals 41, 42; 51, 52; 81,82) for the connection to certified intrinsically safe circuits
Maximum values: 30V
100 mA
750 mW
effective intemal capacitance 4 nF
‘The effective internal inductance is negligibly small.3. Supplement to EC Type-Examination Certificate No. TUV 01 ATEX 1766 X
Data interface
(terminals 33, 34)
or
Probe circuits
(probe connector 1...8)
Routine test
in type of protection Intrinsic Safety EEx ia IIC/EEx ib 1A
‘The maximum values have to be taken from the following
table:
EEx ib A | Ex ia tic
588V | 588V
[313 mA_ | 313mA.
460 mW.
43 uF
0.2mH
Characteristic line: linear
in type of protection Intrinsic Safety EEx ia IIC/EEx ib 1A
for the connection to certified intrinsically safe circuits
Maximum values U;= 10V
k= 275 mA
P= 1420 mw
The effective internal inductance and capacitance are
negligibly small
in type of protection intrinsic Safety Ex ia lIC/IIB/IIA
The maximum values in dependence on the type of
protection have to be taken from the table:
Ex ia WA | EExia WB | EEx ia lic
6o8v | 512V | 389V
95mA_ | 80mA_| 60 mA
1444 mW _| 1024 mwW_| “584 mW.
300nF | 187nF | 34nF
PELE
Characteristic line: linear
In the case of the use in the extended ambient temperature range of -40°C to + 60°C the
overpressure test has to be carried out at every enclosure with 18.5 bar according to EN 50018,
paragraph 15.1.3.1, as routine test.(16)
(17)
(18)
ny
3. Supplement to EC Type-Examination Certificate No. TUV 01 ATEX 1766 X
Test documents are listed in the test report N° 03 YEX 551035,
Special conditions for safe use
The combination of intrinsically safe and non-intrinsically safe circuits at the field cones
not permitted.
sis
Essential Health and Safety Requirements
no additional ones
‘TOV NORD CERT GmbH & Co. KG Hanover, 2003-11-21
‘TOV CERT-Cortifcstion Body
Am TOV
30519 Hannover
(0591 986-1470
(0511 986-2555
fet
Head of the
Certification Body>
TOVNORD
Translation
4. SUPPLEMENT to
EC-TYPE EXAMINATION CERTIFICATE No. TUV 01 ATEX 1766 X
Equipment: Flow meter type Flowsic 600
Manufacturer: Sick Engineering GmbH
Address: Bergener Ring 43
01458 Ottendorf-Okrilla
In the future, the flow meter type Flowsic 600 may also be manufactured and operated according to
the documents listed in the test report.
The amendments concern the optional use of the buffer battery type BR2032 and the mechanical
design.
The buffer battery is intrinsically safe and may be replaced inside of hazardous explosive area.
For the battery replacement the stated waiting time before the enclosure may be opened must be
observed.
All further data including the special conditions for safe use apply unchanged for this supplement.
The test documents are listed in the test report N° 05 YEX 552124.
‘TOV NORD CERT GmbH & Co. KG Hannover, 2008-04-20
‘Am TOV
30519 Hannover
Tel: #49 511 088-1470
Fax: 449 941 086-4
Head of the
Certification Body
page 114TEV NORD
Transiation
5.SUPPLEMENT
to
EC-TYPE EXAMINATION CERTIFICATE No. TI
NV 01 ATEX 1766 X
Equipment: Flow meter type FLOWSIC 600
Manufacturer: Sick Engineering GmbH
Address: Bergener Ring 27
01458 Ottendorf-Okrilla
Amendments:
In the future, the flow meter type FLOWSIC 600 may also be manufactured and operated according
to the documents listed in the test report.
The amendments concer the printed circuit boards SPU Board, IO Board, Power Board and
Backplane. In addition an profibus interface can be used.
Electrical Data
All non-intrinsically safe connections are realised in the type of protection .Increased Safety’
‘Supply circuit U_ = 12.42 Vd. resp. 85...230V a.c.
(terminals 1+, 2-) Un = 263V
Field connections Un = 253V
Current-/switched output active passive
(terminals 31, 32) Up=18V, Ip=35MA Up = 30V, Ip = 35 mA
Switched outputs Up=30V, le = 100 mA
(terminals 41, 42; 51, 52; 81, 82)
Data interface Us=5V, Ip=175 mA
(terminals 33, 34; 81, 82)
Intrinsically safe version
Supply circuit in type of protection , Intrinsic Safety” EEx ia IIC/EEx ib IA
(terminals 1+, 2-) only for the connection to certified intrinsically safe circuits
Maximum values: U, 16 Vv
h 200 mA
5 26 W
Effective intemal capacitance: 4nF
‘The effective internal inductance is negligibly small.
page 114TOV NORD
5. Supplement to EC-Type Examination Cerificate No. TUV 01 ATEX 1766 X
Field connections
Current-/switched output
(terminals 31, 32)
or
Switched outputs
(terminals 41, 42; 51, 52; 81, 82)
Data interface RS485
(terminals 33, 34; 81, 82)
in type of protection ,Intrinsic Safety” EEx ia HIC/EEx ib IIA
‘The maximum values in dependences on the type of
Protection have to be taken from the following table:
EEx ib A | EExiali¢
Up | 221v | 221v
ly | 155mA_| 155 mA
Py | 857 mw_| 857 mw
Co | 4100 nF | 163 nF
by Z mH ‘1 mH
Characteristic line: linear
in type of protection ,Intrinsic Safety” EEx ia IIC/EEX ib 1A
only for the connection to certified intrinsically safe circuits
Maximum values: U, = 30 V
i 100 mA
P, = 750 mw
Effective internal capacitance: 4nF
The effective internal inductance is negligibly small.
in type of protection .Intrinsic Safety” EEx ia IIC/EEx ib 1A
only for the connection to certified intrinsically safe circuits
Maximum values: U; = 30 V
4 100 mA
P, = 750 mw
Effective internal capacitance: 4nF
The effective internal inductance is negligibly small
in type of protection ,Intrinsic Safety” BEx ia IIC/EEx ib A
‘The maximum values in dependences on the type of
Protection have to be taken from the following table:
BEx Ib WA | EEx ia lic
UW | 58ev | 5.66V
I) | 313 mA_| 313 mA
Py | 460 mwW_| “460 mw.
Co
bo
41000 ,F [43 uF |
1.5mH | 0.2mH_|
Characteristic line: linear
age 2a\
TuV NORD
5. Supplement to EC-Type Examination Cortficate No. TUV 01 ATEX 1766 X
or
Data interface Profibus PA
(terminals 33, 34)
Probe circuits
(probe connector 1...8)
in type of protection ,Intrinsic Safety" EEx ia IC/EEx ib 1A
only for the connection to certified intrinsically safe circuits
Maximum values: U, = 10 V
k= 275 mA
P, = 1420 mw
The effective internal inductance and capacitance are
negligibly small.
in type of protection , Intrinsic Safety” EEx ia IIC/EEx ib A
only for the connection to certified intrinsically safe circuits
Maximum values: U, 30 V
4 100 mA
P, = 750 mw
The effective internal inductance and capacitance are
negiigibly small.
in type of protection ,Intrinsic Safety” EEx ia IICIBVIIA
The maximum values in dependences on the lype of
protection have to be taken from the following table:
EEx ia li | BExia iB [ EEx ia lic
Up 60,8 V 51,2 V 38,9V
ly 95 mA 80 mA 60 mA
Po | 1444 mW | 1024 mW | 584 mW
Co. 300 nF 187 nF 34 nF
Lo eB l emma
Characteristic line: linear
All other data apply unchanged for this supplement.
The equipment incl. of this amendments meets the requirements of
EN 50 014:1997 +A1+A2
EN 50 020:2002
EN 50 018:2000+A4 EN 50 019:2000
(18) The test documents are listed in the test report N° 06 YEX 552611.
page 34~
TOV NORD
5. Supplement to EC-Type Examination Cerfificate No. TUV 01 ATEX 1766 X
(17) Special conditions for safe use
All data apply unchanged for this supplement.
(18) Essential Health and Safety Requirements
All data apply unchanged for this supplement.
‘TOV NORD CERT Gmbi48 Co. KG
am TV
1D-30519 Hannover
‘+49 (0) 511 986-1455,
Hoad of the
Certification Body
Hannover, 2006-01-31
eye 444