You are on page 1of 4

FOP-MIV

MEDICAL

APPLICATION
 Cardiovascular: LV pressure, arterial BP
 Pharmacology: Drug and Fluid injecon
 Neurosciences: Intracranial pressure
 Spine - Intradiscal pressure
 Bone - Intramedullary pressure
 Urology - Bladder/Ureter pressure
 MRI RFI Gang - Arterial blood pressure or
LV pressure for image gang
 Respiratory / Pulmonology
 Otorhinolaryngology: Inner ear pressure
 Ophthalmology: Intraocular pressure
 Gastro intesnal

FOP-MIV

BARE SHEATHED SHEATHED FILLED WITH GEL


FOP-MIV-10: 300mmHg FOP-MIV-20: 300mmHg FOP-MIV-21: 300mmHg
FOP-MIV-30: 0-150PSI FOP-MIV-40: 0-150PSI FOP-MIV-41: 0-150PSI

SPECIFICATIONS
Pressure Range1 -300mmHg to 300mmHg -300mmHg to 7500mmHg
2
Resoluon <0.3mmHg 2.6mmHg
3
System Accuracy 3mmHg 50mmHg

Zero thermal eect4 -0.4mmHg / C 3 mmHg / C


Proof pressure >4000mmHg >22 500mmHg
Cable Sheathing Nylon Sheathing, OD: 0.9 mm
Tip Terminaon Bare / Sheathed / Sheathed with gel / Custom design
Standard Sensor Length 2 Meters
Connector SCAI, SCAI is a SCA connector with smart chip communicang
calibrang data to the signal reading module
1. Relave to atmospheric pressure
2. With a FPI-HS reading module and a lter of 30Hz applied.
3. M260 sensor and FPI-HR module : includes reproducibility (sensor/module exchange), repeatability and hysteresis,
non-linearity, scale error, oset error, condioner temperature compensaon error.
4. Determined between 10C and 70C at atmospheric pressure (760mmHg)

Preliminary
FPI-HR / FPI-HS module
MEDICAL

The FPI-HR and FPI-HS signal condi oner are


designed for laboratories and OEM applica ons

Description

The FPI-HR and FPI-HS like all FPI-Modules are com-


pables with evoluon chassis and with the evolu-
on soware1.

The FPI-HR is suitable for pressure measurements


and for temperature measurement.

Compatible EVOLUTION chassis


 EVO-SD-2 (up to 2 modules)
 EVO-SD-5 (up to 5 modules)
 EVO-RM-8 (up to 8 modules)

Dimensions Specifications

FPI-HR FPI-HS
Number of channel (s) 1 or 2 1
Resoluon2 <0.3mmHg <0.1mmHg
Sampling rate up to 250Hz (1 channel) 15kHz
125Hz (2 channels)
Atmospheric No Yes
self-compensaon
Analog output 0 to 5V
16 bits resoluon
Analog output delay 8ms (1 channel) 130s
response3 16ms (2 channels)
Power consumpon 5 Was 12 Was
Power consumpon 24VDC
Operang temperature 10C to 50C
Storage temperature -30C to 80C
Communicaon USB via EVO chassis,
TS 35 DIN RAIL

1. The evoluon so
ware is included in the evoluon chassis which is sold separately.
2. Using a low pass lter of 30Hz.
3. Delay between the physical phenomenon and the analog output change.

Preliminary
EVOLUTION chassis
MEDICAL

The evolution chassis are the easiest way to configure and use evolution modules.

Description Key Features


evoluon chassis footprint, communicaon capabilies and i-evo module for communicaon and for
speed make it the ideal tool for laboratory and in site test power supply distribuon
environments.
USB communicaon
The evoluon chassis can house dierent module types with
dierent channel capabilies to combine results from a single Evoluon soware for sensors and module
conguraon and for data acquision up to
acquision source. 5 k samples/sec. total.
evoluon chassis have a dierent number of module slots, External data acquision system required for
depending on the model: higher acquision rate > 5k samples/sec.
Module capacies, communicaon ports, and overall width Full bandwidth via analog output connectors
specicaons dier from one model to the other.
USB communicaon interface is available on all chassis.
The SD-2, SD-5 and RM evoluon chassis package includes the
following components:
evoluon chassis unit,
i-evo module, Applications
Power supply adaptor and cord,
Laboratory measurements with evoluon modules
USB interface cable,
Module removal tool, Easy set-up of evoluon modules before migrang
modules in your own equipment
User guide,
CD containing so"ware driver and manual (pdf).

Specifications
Model EVO-SD-2 EVO-SD-5 EVO-RM
Communicaon USB USB USB
Data logging Via computer Via computer Via computer
Number of modules Up to 2 Up to 5 Up to 8
Power supply 24 VDC 70 W 24 VDC 70 W 24 VDC 150 W
Evoluon soware Included Included Included
Maximum rate of acquision1 5 k samples/sec. total 5 k samples/sec. total 5 k samples/sec. total
Dimensions W:133 x H:177 x D:156mm W:269 x H:177 x D:156mm W:483 x H:132 x D:175mm
1. With the evolution software and chassis. Analog output is available directly on the reading modules, offering full acquisition rate. Ex. FPI-HS plugged on analog is at 15Ksamples/sec.

Preliminary
EVOLUTION system
MEDICAL

EVOLUTION Software and Solution Summary


Configure and control Simple monitoring and
the reading instrument real-time graphing
The most common set-up Users may choose reading the
users will be congure the 0- actual measurement, or plot
5V analog output level to the in real-me with user specied
pressure range of interest, but screen refresh rates.
end user will also enjoy the
visually conrm proper Export data
communicaon between
catheters and instrument. While users may generally
prefer to use 250Hz/125Hz
analog output on the FPI-HR,
data may also be recorded and
saved in mulple les formats

Other Accessories: Extensions cable


Be sure to purchase this 3 meter extension
cable when a longer working distance is THR-10 and a FOP-M260-10 on an US dime.
required, but also can be removed when
working close to the subject.

FISO Technologies inc Phone +1.418.688.8065 Email info@so.com


500 St-Jean-Baste Ave, Suite 195 Fax +1.418.688.8067 Web www.so.com
Qubec (Quebec) Canada G2E 5R9
DOC: MC-00217 R4 FISO Technologies Inc. reserves the right to make any changes in the specicaons without prior noce.