You are on page 1of 9

Electronic

Calibration
technology
pressure
measurement

Flush pressure transmitter


For viscous and particle-laden media
Model S-11
WIKA data sheet PE 81.02

for further approvals


see page 6

Applications
General industrial applications
Food and beverage industry

Filling and packing machinery


Dosing technology

Level measurement

Special features
High-quality product

Many configurations possible


Flush process connection

Large stocks for short delivery times


Vacuum-tight

Fig. left:

Fig. right:

Pressure transmitter model S-11

Pressure transmitter model S-11 with


cooling element

Description
Expert for viscous and particle-laden media
The model S-11 pressure transmitter with flush diaphragm
has been specifically designed for the measurement of
viscous, paste-like, adhesive, crystallising, particle-laden and
contaminated media, which would clog the pressure channel
of conventional process connections.
Through its optimised design, the flush process connection
enables cleanability with the wetted diaphragm integrated
into the process. Low-maintenance and trouble-free pressure
measurement is thus also guaranteed in critical applications
with frequently changing media.
High accuracy, a robust design, high-quality workmanship
and the high flexibility of configuration are key features of the
model S-11.

Flush process connection


All process connections of the flush pressure transmitter
are made of stainless steel and are fully welded and isolate
the medium from the pressure measuring instrument via a
positive seal. A reliable, dead-space free sealing between
the process connection and the measuring medium is thus
assured.
For high medium temperatures of up to 150 C (302 F),
the pressure transmitter is also available with an integrated
cooling element.
Specifically for the food and beverage industry, a version
with internal system fill fluid in accordance with FDA 21 CFR
178.3750 can be chosen.

WIKA data sheet PE 81.02 06/2016


Data sheets showing similar products:
High-quality pressure transmitter for general industrial applications; model S-20; see data sheet PE 81.61
Pressure transmitter for applications in hazardous areas; model IS-3, see data sheet PE 81.58
Pressure transmitter for highest pressure applications; model HP-2; see data sheet PE 81.53

Page 1 of 9

Measuring ranges
Gauge pressure
bar

psi

Measuring range

0 ... 0.6

0 ... 1

0 ... 1.6

10

Measuring range

0 ... 2.5

0 ... 4

0 ... 6

0 ... 10

0 ... 16

0 ... 25

0 ... 40

Overpressure limit

10

17

35

35

80

50

Measuring range

0 ... 60

0 ... 100

0 ... 160

0 ... 250

0 ... 400

0 ... 600

Overpressure limit

120

200

320

500

800

1,200

Measuring range

0 ... 15

0 ... 20

0 ... 30

0 ... 50

0 ... 60

0 ... 100

80

0 ... 150

Overpressure limit

145

145

145

240

240

500

500

Measuring range

0 ... 160

0 ... 200

0 ... 250

0 ... 300

0 ... 400

0 ... 500

0 ... 600

Overpressure limit

1,160

1,160

1,160

1,160

1,160

1,160

1,160

Measuring range

0 ... 750

0 ... 1,000

0 ... 1,500

0 ... 2,000

0 ... 3,000

0 ... 5,000

0 ... 6,000

Measuring range

Overpressure limit

1,740

0 ... 0.25

0 ... 6

1,740

0 ... 0.4

0 ... 10

2,900

0 ... 0.6

35

80

Measuring range

0 ... 15

0 ... 25

0 ... 50

72.5

145

Measuring range

-0.6 ... 0

0 ... 1

7,200

0 ... 1.6

10

0 ... 100

0 ... 250

240

500

-0.4 ... 0

-0.25 ... 0

-1 ... +5

1.5

-1 ... +0.6

-1 ... +1.5

-1 ... +3

-1 ... +24

Measuring range

-1 ... +9

-1 ... +15

Overpressure limit

Measuring range

35

-30 inHg ... 0

80

-30 inHg ... +30

Overpressure limit

72.5

240

Measuring range

-30 inHg ... +200

-30 inHg ... +300

Overpressure limit

1,160

17

-0.1 ... 0

-1 ... 0

10

0 ... 4

10

-0.16 ... 0

Overpressure limit

0 ... 2.5

11,600

1,160

Measuring range

Overpressure limit

11,600

0 ... 16

35

Overpressure limit

4,600

Overpressure limit

Vacuum and +/- measuring range

psi

0 ... 0.4

Measuring range

bar

0 ... 0.25

1.5

Absolute pressure

psi

0 ... 0.16

Overpressure limit

bar

0 ... 0.1

Overpressure limit

10

17

35

50

-30 inHg ... +60

240

-30 inHg ... +100


500

-30 inHg ... +160

1,160

1,160

The given measuring ranges are also available in mbar, MPa and further units.
Vacuum tightness
Yes

Page 2 of 9

WIKA data sheet PE 81.02 06/2016

Output signals

Accuracy specifications

Output signal

Accuracy at reference conditions

Selectable versions
Current (2-wire)

4 ... 20 mA

Current (3-wire)

0 ... 20 mA

Voltage (3-wire)

DC 0 ... 10 V
DC 0 ... 5 V

Other output signals on request.


Load
Depending on the signal type the following loads apply:
Current (2-wire): (power supply - 10 V) / 0.02 A
Current (3-wire): (power supply - 3 V) / 0.02 A
Voltage (3-wire): > max. output signal / 1 mA

Voltage supply
Power supply
The power supply depends on the selected output signal
4 ... 20 mA (2-wire): DC 10 ...30 V
0 ... 20 mA (3-wire): DC 10 ...30 V
DC 0 ... 10 V:
DC 14 ... 30 V
DC 0 ... 5 V:
DC 10 ...30 V

Reference conditions (per IEC 61298-1)


Temperature
15 ... 25 C (59 ... 77 F)
Atmospheric pressure
860 ... 1,060 mbar (12.47 ... 15.37 psi)
Humidity
45 ... 75 % r. h.
Power supply
DC 24 V
Mounting position
Calibrated in vertical mounting position with process
connection facing downwards.

WIKA data sheet PE 81.02 06/2016

Selectable versions
Standard

0.5 % of span

Option

0.25 % of span 1)

1) Only for measuring ranges 0.25 bar

Including non-linearity, hysteresis, zero offset and end value


deviation (corresponds to measured error per IEC 61298-2).
Calibrated in vertical mounting position with process connection facing downwards.
Non-linearity (per IEC 61298-2)
0.2 % of span BFSL
Non-repeatability
0.1 % of span
Temperature error in rated temperature range
Rated temperature range:
0 ... 80 C (32 ... 176 F)
Mean temperature coefficient of zero point:
Measuring range > 0.25 bar: 0.2 % of span/10 K
Measuring range 0.25 bar: < 0.4 % of span/10 K
Mean temperature coefficient of span:
0.2 % of span/10 K
Long-term drift
0.2 % of span/year
Adjustability of zero point and span
Adjustment is made using potentiometers inside the
instrument.
Not possible for IP68 cable outlet.
Zero point: 5 %
Span:
5%

Time response
Settling time
2 ms

Page 3 of 9

Operating conditions

Electrical connections

Ingress protection (per IEC 60529)


The ingress protection depends on the type of electrical
connection.

Short-circuit resistance
S+ vs. U-

Electrical connection

Ingress protection

Circular connector M12 x 1 (4-pin)

IP67

Cable outlet IP67

IP67

Cable outlet IP68

IP68 1)

Angular connector DIN 175301-803 A

1) Adjustability of zero point and span not possible.

IP65

The stated ingress protection only applies when plugged in


using mating connectors that have the appropriate ingress
protection.
Vibration resistance
Process connections without cooling element:
20 g (IEC 60068-2-6, under resonance)
Process connections with cooling element:
10 g (IEC 60068-2-6, under resonance)
Shock resistance
Process connections without cooling element:
1,000 g (IEC 60068-2-27, mechanical)
Process connections with cooling element:
400 g (IEC 60068-2-27, mechanical)
Permissible temperature ranges
Also meets EN 50178, tab. 7, operation (C) 4K4H,
storage (D) 1K4, transport (E) 2K3.
Process connections without cooling element
Medium

Standard

-30 ... +100 C

-22 ... +212 F

Option

-30 ... +125 C

-22 ... +257 F

Ambient

-20 ... +80 C

-4 ... +176 F

Storage

-40 ... +100 C

-40 ... +212 F

Process connections with cooling element


Medium

-20 ... +150 C

-4 ... +302 F

Ambient

-20 ... +80 C

-4 ... +176 F

Storage

-40 ... +100 C

-40 ... +212 F

Page 4 of 9

Reverse polarity protection


U+ vs. UOvervoltage protection
DC 36 V
Insulation voltage
DC 500 V with NEC class 02 voltage supply (low voltage and
low current max. 100 VA even under fault conditions)
Connection diagrams
Angular connector DIN 175301-803 A

1
3
2

U+

2-wire

3-wire

U-

S+

Wire cross-section max. 1.5 mm2 (AWG 16)


Cable diameter 6 ... 8 mm (0.24 ... 0.31")

Circular connector M12 x 1 (4-pin)

U+

2-wire

3-wire

U-

S+

Cable outlet

U+

2-wire

brown

3-wire

U-

green

green

S+

white

Shield

grey

grey

brown

Wire cross-section 6 x 0.5 mm (AWG 20)


Cable diameter 6.8 mm (0.27")
Cable lengths 1.5 m, 3 m, 5 m, 10 m, 15 m
(4.9 ft, 9.8 ft, 16.4 ft, 32.8 ft, 49.2 ft)

Other connections on request.

WIKA data sheet PE 81.02 06/2016

Process connections
Selectable versions
Thread

Available measuring ranges

G B flush 1)
G 1 B flush

1)

Hygienic G 1 B flush
(in accordance with 3-A Sanitary Standards)

0 ... 2.5 to 0 ... 600 bar

0... 50 to 0 ... 6,000 psi

0 ... 0.1 to 0 ... 1.6 bar

0 ... 15 psi

0 ... 0.1 to 0 ... 25 bar

0 ... 15 to 0 ... 300 psi

1) Process connection also available with cooling element.

Cooling elements
For higher medium temperatures, process connections with a cooling element are available (see Operating conditions).
Sealing
Process connection

Max. medium
temperature

Sealing material and max. pressure limitation

without cooling element

up to 100 C (212 F)

NBR
up to 600 bar (8,700 psi)

FKM/FPM
up to 600 bar (8,700 psi)

EPDM
up to 200 bar (2,900 psi)

up to 125 C (257 F)

NBR
up to 600 bar (8,700 psi)

FKM/FPM 1)
up to 400 bar (5,800 psi)

EPDM
up to 200 bar (2,900 psi)

with cooling element

up to 150 C (302 F)

FKM/FPM 1)
up to 300 bar (4,350 psi)

EPDM
up to 200 bar (2,900 psi)

Hygienic

up to 150 C (302 F)

EPDM
up to 200 bar (2,900 psi)

1) Minimum permissible medium and ambient temperature -20 C / -4 F

Standard

Option 1

1)

Option 2

O-ring for G B flush process connection optionally available with FFKM seal material (pressure limitation 600 bar (8,700 psi)
for all medium temperatures)
The sealings listed under Standard are included in the delivery.

Materials
Wetted parts

G B and G 1 B flush: 316TI


G 1 B flush, hygienic: 316L

For sealing materials see Process connections

Internal system fill fluid


Selectable versions
Standard

Synthetic oil

Option

Food-compatible system fill fluid per FDA 21 CFR


178.3750

WIKA data sheet PE 81.02 06/2016

Page 5 of 9

Approvals
Logo

Description

Country

European Community

EU declaration of conformity

EMC directive 1)

EN 61326 emission (group 1, class B) and interference immunity (industrial application)

Pressure equipment directive

CSA

Canada

EAC

Eurasian Economic Community

GOST
Metrology, measurement technology

Russia

KazInMetr
Metrology, measurement technology

Kazakhstan

MTSCHS
Permission for commissioning

Kazakhstan

BelGIM
Metrology, measurement technology

Belarus

Uzstandard
Metrology, measurement technology

Uzbekistan

3-A
Sanitary Standard

USA

Safety (e.g. electr. safety, overpressure, ...)


Hazardous areas
Electromagnetic compatibility

only for instruments with process connection hygienic G 1 B


This instrument is 3-A marked, based on a third party verification for conformance to the 3-A standard.
1) The existence of strong electromagnetic fields in a frequency range of < 2.7 GHz can result in increased measuring errors up to 1 %. Do not install the instruments in the vicinity of strong
electromagnetic sources of interference (e.g. transmitting devices, radio equipment), or use sheath current filters where applicable.

Approvals and certificates, see website

Page 6 of 9

WIKA data sheet PE 81.02 06/2016

Dimensions in mm (inch)
Pressure transmitter
with angular connector DIN 175301-803 A

with M12 x 1 circular connector

with cable outlet IP67

Weight approx. 0.2 kg / 7 oz

Weight approx. 0.2 kg / 7 oz

Weight approx. 0.2 kg / 7 oz

with cable outlet IP68

with cooling element for medium


temperatures up to 150 C

with cooling element for medium


temperatures up to 150 C and hygienic
process connection

Weight approx. 0.2 kg / 7 oz

Weight approx. 0.3 kg / 10 oz

Weight approx. 0.5 kg / 18 oz

Electrical connection

Angular connector
DIN 175301-803 A
Circular connector M12 x 1

Dimension Process connection


G B, G 1 B

without cooling element

G 1 hygienic

with cooling element

with cooling element


109 (4.29)

H1

54 (2.13)

69 (2.72)

H2

83.5 (3.29)

98 (3.56)

80 (3.15)

H1

54 (2.13)

69 (2.72)

80 (3.15)

H2

67 (2.64)

82 (3.23)

93 (3.66)

Cable outlet IP67

H1

54 (2.13)

69 (2.72)

80 (3.15)

Cable outlet IP68

H1

58 (2.28)

73 (2.87)

84 (3.31)

WIKA data sheet PE 81.02 06/2016

Page 7 of 9

Process connections

D1

L1

L2

L3

G B 18
23
20.5 10
(0.71) (0.91) (0.81) (0.4)

L4

15.5
(0.61)

D1

L1

L2

L3

G 1 B 30
23
20.5 10
(1.19) (0.91) (0.81) (0.4)

L4

15.5
(0.61)

D1

G1B
29.5
hygienic (1.17)

L1

25
(0.99)

L2

9
(0.36)

L3

15.5
(0.61)

Surface roughness of wetted metallic parts, Ra 0.76 m

For information on tapped holes and welding sockets, see Technical information IN 00.14 at www.wika.com.

Accessories
Welding socket
Designation

Order number

Welding socket for G B flush

1192299

Welding socket for G 1 B flush

1192264

Welding socket for G 1 B hygienic flush

14145179

Welding socket for G 1 B hygienic flush with leak-control channel


Fulfils the 3-A Sanitary Standards

14145183

For further adapter systems for measuring instruments with G 1 B hygienic process connection, see data sheet AC 09.20.
Mating connector
Designation

Order number
without cable

2 m cable,
shielded

5 m cable,
shielded

10 m cable,
shielded

with cable gland, metric

11427567

14100465

14100466

with cable gland, conduit

11022485

straight

14086880

14086883

14086884

angled

14086889

14086891

14086892

Angular connector DIN 175301-803 A

Circular connector M12 x 1 (4-pin)

Page 8 of 9

WIKA data sheet PE 81.02 06/2016

Sealings for mating connectors


Mating connector

Order number

Angular connector DIN EN 175301-803 A

1576240

Blue (WIKA)

Brown (neutral)

11437902

Sealings for process connection


Thread size and sealing

Order number
NBR

FPM/FKM

EPDM

FKKM

O-ring

14072275

14072276

14072277

14073739

Profile sealing

1039067

1039075

1538306

O-ring

1108247

1099094

1535056

Profile sealing

1100386

1145967

11522381

O-ring

2225859

Profile sealing

11522381

GB

G1B

G 1 B hygienic

Ordering information
Model / Measuring range / Output signal / Accuracy / Electrical connection / Medium temperature / Process connection /
Sealing / System fill fluid

06/2016 EN

2012 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.

WIKA data sheet PE 81.02 06/2016

Page 9 of 9

WIKA Alexander Wiegand SE & Co. KG


Alexander-Wiegand-Strae 30
63911 Klingenberg/Germany
Tel. +49 9372 132-0
Fax +49 9372 132-406
info@wika.de
www.wika.de