Introduction
Over the past 10 years, integrated circuit chips (IC), different electronic devices and several other technology
products have become smaller and more efficient in
terms of performance. The major limiting factor for the
size of an integrated circuit chip and its speed is the ultra
trace (pg/g level) metal contamination of the silicon used
in their manufacturing. For example, high purity quartz is
used as a filler material in the encapsulation of discrete
and integrated circuit devices. The metal impurities in
ultrapure materials lead to higher dark currents that influence the chips ability to carry currents (Close and Yarwood 1971). Therefore, high purity materials and other
chemicals that are used in the IC manufacturing processes must be analysed for ultratrace metal impurities
and controlled (Luck 1998). While 10 years ago semiconductor equipment manufacturers institute (SEMI)
deemed that 10 ng/g purity levels were adequate for many
chemicals, today 100 pg/g levels are typical and for some
of the more critical materials even 10 pg/g levels are being proposed. Hence, there is a great need for the analysis
of high pure materials for trace element impurities at extremely low concentrations to evaluate the intrinsic properties of different materials required for the semiconductor
industry including process chemicals (high-purity water,
mineral acids, organic solvents, etc).
balaramv1951@yahoo.co.in
V Balaram
346
Table 1. Comparison of certified values and values obtained by HRICPMS using continuous
nebulization (Patterg and Matschat 1999).
Element
Ag
BCR074
Certified value uncertainty
(95% c.i.) (g/g)
1280 070
Measured isotope
107
109
As
Bi
Cr
Fe
Ni
078 014
010 003
< 010
114 006
104 011
Pb
097 005
Ag
Ag
75
As
Bi
52
Cr
56
Fe
58
Ni
60
Ni
209
207
208
Sb
058 003
121
123
Sn
< 007
118
120
HRICPMS
average of n values uncertainty
(95% c.i.) (g/g)
1285 030
1287 030
078 004
0098 0004
005 0016
114 011
098 004
114 004
Pb
Pb
096 0024
096 0021
Sb
Sb
060 0020
060 0018
Sn
Sn
0033 0007
0036 0012
347
Table 2. Trace element concentrations (ng/g) in high pure tellurium by ICPMS and GDMS in
comparison with purity requirements.
Element
Li
Be
B
F
Na
Al
P
K
Ca
Cr
ICPMS*
<1
<1
< 10
< 10
< 10
<1
< 10
< 10
<5
GDMS@
<3
<4
<4
< 50
<3
<2
<3
< 45
< 50
<4
7N
<5
<2
<5
< 10
15
20
2
< 30
< 30
<1
Element
ICPMS*
GDMS@
7N
Ni
Cu
Zn
As
Rb
Ag
Ba
Mo
Sr
Pb
< 10
<5
< 10
< 10
< 10
< 10
< 10
< 10
< 10
< 10
<2
<5
<9
<3
<2
<3
<2
<1
< 07
<2
<2
<2
<8
<1
<1
<1
<7
< 05
<1
<1
, not determined; 7N, purity requirements; *, at NGRI, Hyderabad (ELAN DRC II); @at Institute for
National Measurement Standards, National Research Council, Canada.
Table 3. Concentrations of some trace elements in sub-boil distilled water, sub-boil distilled hydrochloric
and nitric acids (Balaram et al 2000).
Sub-boil distilled
hydrochloric acid
Element
Be
B
V
Cr
Mn
Fe
Co
Ni
Cu
Zn
As
Se
Rb
Sr
Mo
Ag
Cd
Sb
Ba
Tl
Pb
Bi
Sub-boil distilled
water (ng/g)
NGRI1
< 001
288
010
< 001
< 001
< 001
< 002
< 002
< 002
025
001
< 001
056
049
< 001
< 001
< 001
004
004
< 001
NGRI1
(ng/g)
< 001
265
830
< 001
< 001
< 001
< 002
1315
< 002
231#
< 001
< 001
< 002
< 001
088
< 001
< 001
007
004
005
013
SEMI2
(g/g)
01
1
1
02
01
01
01
1
01
1
Sub-boil distilled
nitric acid (ng/g)
NGRI1
(ng/g)
< 001
1068
496
5455
2722
6601
064
< 002
7560
060
738
661
005
< 001
827
018
011
011
< 001
003
055
139
SEMI2
(g/g)
01
05
02
05
005
005
1
05
05
V Balaram
348
Conclusions
To keep pace with the increasingly stringent requirements of the analysis of ultra pure materials, analytical
instrumentation must undergo continual improvement.
The detection limits offered by different mass spectro-