INTRODUCTION
Microorganisms such as bacteria are responsible for decomposing organic waste. When
organic matter such as dead plants, leaves, grass clippings, manure, sewage, or even food waste
is present in a water supply, the bacteria will begin the process of breaking down this waste.
When this happens, much of the available dissolved oxygen is consumed by aerobic bacteria,
robbing other aquatic organisms of the oxygen they need to live. Biochemical oxygen demand or
BOD is a procedure to determine the amount of oxygen consumed by the microorganisms in the
waste water chemically. It is commonly used as the indicator to show the cleanliness of the waste
water.
2.0
OBJECTIVE
To measure the amount of dissolved oxygen in the waste water for a specified period of
THEORY
BOD is defined as the amount of oxygen required by living organisms in the stabilization
of the organic matter of water. If the oxidation of an organic compound is carried out by
microorganisms using the organic matter as a food source, the oxygen is consumed. The greater
the amount of organic matter present, the greater the amount of oxygen utilized. The BOD test is
indirect measurement of organic matter. Since the test is performed over a five day period, it is
often refferred as five day BOD (BOD5).
It is generally assumed that the rate at which oxygen is consumed directly proportional to
the concentration of degradable organic matter remaining at any time. Different result would be
obtained at different temperatures because biochemical reaction rates are temperature-dependent.
The kinetics of the BOD reaction are, for particle purposes, formulated in accordance with firstorder reaction kinetics and may be expressed as
dL / dt = - kLt
where, Lt is the amount of the first stage BOD remaining in the water at time t and k is the
reaction rate constant. This equation can be integrated as
ln Lt / L = - kt or The amount of BOD remaining at time t equals
Lt = L e-kt
and y the amount of BOD that has been exerted at any time t, equals
BODt = L( 1 e-kt )
For five day BOD (BOD5), the sample is kept in incubator which remains around 20 oc. Most of
the biological process sprrd up as the temperature increase and slow down as the temperature
drop. The temperature of 20oc is standard temperature in lab and it is suitable for the bacteria
active in breaking down the waste.
4.0 APPARATUS AND MATERIAL
a.
b.
c.
d.
e.
f.
g.
h.
Beaker
5 unit of Bottles
Lake UTHM water
Measuring cylinder
Deionised water
pH Meter
DO meter
BOD incubator
5.0 PROCEDURES
BOD ID
Volume Of
Volume Of
Initial Do
Samples (ML)
Dilution Water
(Mg/L)
4
4
4
0
0
(ML)
296
296
296
300
300
8.21
8.20
8.25
8.20
8.21
Sample 1
Sample 2
Sample 3
Blank 1
Blank 2
Show all the calculation and state if any of the data needs to thrown out.
1. DO3 Blank (1) = (8.20 - 7.80) mg/L = 0.4 mg/L is more than 0.2 mg/L
2. DO3 Blank (2) = (8.21 - 7.87) mg/L = 0.34 mg/L is more than 0.2 mg/L
Final Do (Mg/L)
7.52
7.61
7.55
7.80
7.87
3. DO3 Sample (1) of 4ml wastewater = (8.21- 7.52) = 0.69 mg/L is less than 2 mg/L
4. DO3 Sample (2) of 4ml wastewater = (8.20- 7.61) = 0.59 mg/L is less than 2 mg/L
5. DO3 Sample (3) of 4ml wastewater = (8.25- 7.55) = 0.70 mg/L is less than 2 mg/L
7.0
DISCUSSION
7.0
APENDIX
Figure 2 DO Meter