ELECTROCHEMICAL
SCIENCE
www.electrochemsci.org
In this work, two new loperamide potentiometric electrodes were prepared and used for
pharmaceutical analysis. In case of liquid membrane electrode, the best results were obtained from the
electrode with the membrane composition of 5% ion-pair, 1% NaTPB, 30% PVC, and 64% DBP. The
proposed sensor shows a Nernstian response (slope of 57.2 mV per decade) in the range of 3.010-51.010-2 mol L-1. In case of carbon paste electrode modified by MWCNTs, calibration graph slope is
about 58.5 mV per decade in the concentration range of loperamide hydrochloride 1.010-5-1.010-2
mol L-1. Both sensors work well in the laboratory conditions and successfully applied in analysis of
loperamide in some pharmaceutical formulations.
1. INTRODUCTION
Loperamide,
Fig.
1,
(4-(p-chlorophenyl)-4-hydroxy-N,N,-dimethyl-,-diphenyl-1piperidinebutyr amide hydrochloride), an opiate-receptor agonist, is broadly used as an efficient drug
for control and symptomatic relief of acute non-specific diarrhea and chronic diarrhea associated with
inflammatory bowel diseases [1]. Recently, it has been also reported that loperamide could have some
attention as an antihyperalgesic agent reducing pain without any central nervous system side effects
[2].
Many methods have been reported for determination of loperamide in pharmaceutical or
biological samples, including gas chromatography-mass spectrometry (GC-MS) [3] colorimetry [4],
7644
N
Cl
O
H3C
HO
N
CH3
Here, lopramide hydrochloride and sodium tetraphenyl borate were used in synthesis of the
ion-pair complex which acts as a sensing element in the electrodes. The synthesized ion-pair was used
in making the both kinds of electrodes. Two kinds of ion selective electrode, PVC membrane electrode
and carbon paste one are made and characterized and then used for pharmaceutical analysis of
loperamide in some formulations.
2. EXPERIMENTAL PART
2.1. Apparatus
The glass cell in which the potentiometry was carried out into contained an Ag/AgCl electrode
(Azar electrode, Iran) as a reference electrode and loperamide selective electrode as an indicator
electrode. Both electrodes were connected to a mili-voltmeter (0.1).
The following cell was assembled for the conduction of the EMF (electromotive force)
measurements:
7645
7646
was dipped into the mixture for about 10 s so a thick transparent membrane was formed. The tube
was then pulled out from the mixture and kept at room temperature for about 5 h. Afterwards,
the tube was filled with an internal filling solution of loperamide hydrochloride (1.010-3 mol L-1 ).
The electrode was finally conditioned for 20 h by soaking in the same solution [20-22].
7647
1
2
3
4
5
6
7
8
Composition (%)
PVC
Plasticizer
Ion-pair
30
30
30
30
30
30
30
30
DBP, 67
DBP, 65
DBP, 63
DBP, 64
DBP, 63
NB, 64
BA, 64
DBP, 69
3
5
7
5
5
5
5
0
Additive
(NaTPB)
0
0
0
1
2
1
1
1
Slope
(mV decade-1)
LR (mol L-1)
29.60.6
52.40.4
51.30.5
57.20.4
54.50.6
21.80.6
27.70.5
3.70.6
1.010-4-5.010-3
5.010-5-5.010-3
5.010-5-5.010-3
3.010-5-1.010-2
3.010-5-1.010-2
5.010-5-1.010-2
1.010-4-1.010-2
5.010-4-1.010-3
R2
0.908
0.957
0.942
0.997
0.990
0.905
0.894
-
Response
time
57 s
46 s
50 s
26 s
33 s
1 min
53 s
4 min
7648
response time of the carbon paste electrode. Addition of 5% MWCNT to the composition increased the
response to a Nernstian slope of about 58.50.3 mV per decade (no. 7). From Table 2, it was obvious
that in the absence of ion-pair and presence of other components (no. 9), the response of the modified
CPE was very low (slope of 3.20.8 mV per decade).
Composition (%)
Graphite
Paraffin
Ion-pair
MWCNTs
80
70
60
55
55
57
55
53
55
20
20
20
20
25
20
20
20
20
10
20
25
20
20
20
20
-
3
5
7
5
1
2
3
4
5
6
7
8
9
Slope
(mV decade-1)
LR (mol L-1)
Response
time
R2
36.80.7
45.50.5
46.30.6
30.20.8
53.60.3
58.50.3
56.60.5
3.20.8
5.010-4-5.010-3
1.010-5-1.010-2
1.010-5-1.010-2
5.010-5-5.010-3
1.010-5-1.010-2
1.010-5-1.010-2
1.010-5-1.010-2
5.010-4-1.010-3
1.1 min
50 s
54 s
1.5 min
31 s
20 s
44 s
3 min
0.899
0.915
0.912
0.927
0.9
0.997
0.945
0.811
250
160
140
Nano-composite electrod
y = -58.5x + 318.4
R = 0.997
200
120
E/mV
E/mV
150
100
80
100
60
40
50
20
0
8
-log[Loperamide.HCl]
-log[Loperamide.HCl]
Figure 2. Calibration curves of loperamide nano-composite and PVC membrane electrode. The
results are based on 5 replicate measurements.
The measuring range of a potentiometric sensor is the linear part of the calibration graph as
shown in Figure 2.
7649
For many ion-selective electrodes the measuring range can extend from 1 molar to 10-6 or
even 10-7 molar concentrations [41-44]. Slope of the calibration curve of PVC membrane electrode is
57.2 mV per decade of the loperamide hydrochloride concentration and a standard deviation of 0.4
mV after five replicate measurements. A linear response towards the loperamide hydrochloride
concentration was from 3.010-5-1.010-2 mol L-1. Calibration graph slope for CPEs is 58.5 mV per
decade of loperamide hydrochloride concentration in the range of 1.010-5-1.010-2 mol L-1 with a
standard deviation of 0.3 mV after eight replicate measurements. Detection limit was calculated
from the intersection of two extrapolated segments of the calibration graph. In this work,
detection limit of the PVC membrane sensor was 1.010-5 mol L-1 and in case of nano-composite
carbon paste electrode was 9.010-6 mol L-1 which was calculated by extrapolating the two segments
of the calibration curves.
50
E/mV
40
30
20
10
0
0
10
pH
Figure 3. Applicable pH of the electrodes in the test solution of 1.010-4 mol L-1
To examine the effect of pH on the electrode responses, the potential was measured at specific
7650
concentration of the loperamide hydrochloride solution (1.010-4 mol L-1) from the pH value of 1.0 up
to 9.0 (concentrated NaOH or HCl solutions were employed for the pH adjustment) by PVC
membrane electrode. The results showed that the potential remained constant despite the pH change
in the range of 3.5 to 6.0, which indicates the applicability of this electrode in the specified pH range.
Relatively noteworthy fluctuations in the potential vs. pH behavior took place below and above
the formerly stated pH limits. In detail, the fluctuations above the pH value of 6.0 might be justified
by removing the positive charge on the drug molecule. Fluctuations below the pH value of 3.5 were
caused by removal of the membrane ingredients or analyte in the solution. In both electrodes the same
trend were observed.
PVC membrane
Slope
(mV per decade)
DL (mol L-1 )
57.2
1.010-5
Second
57.0
-5
Third
56.8
Fourth
56.4
Fifth
56.1
Sixth
52.5
Seventh
51.3
3.510
Eighth
49.2
7.510
Ninth
42.0
1.510
Tenth
33.7
First
1.510
-5
3.010
-5
5.510
-5
7.010
-4
1.010
-4
-4
-3
-3
3.510
Nano-composite
CPE
Slope
(mV per decade)
58.5
58.4
58.2
58.0
57.7
57.4
DL (mol L-1)
9.010-6
-6
9.510
-5
1.010
-6
2.510
-5
4.010
-5
5.010
-5
55.5
9.010
53.2
2.510
50.0
7.010
46.0
-4
-4
-4
9.510
For this estimation, three electrodes were employed extensively (1 hour per day) for 10
weeks. After 5 weeks utilization of PVC membrane electrode, two changes were observed: a slight
gradual decrease in the slope and an increase in the detection limit. As it can be seen from Table
3, this time in case of carbon paste was 6 weeks which shows the long-term stability of this kind of
sensor in comparison with PVC membrane electrodes.
In PVC membrane electrodes after several time of usage, the membrane ingredients
leak from the organic layer and affect the membrane response. While in CPEs the surface of the
7651
Labeled amount
Sample 1
Sample 2
Sample 3
Sample 4
2 mg/CAP
2 mg/CAP
2 mg/Tab
2 mg/Tab
2.170.13 mg/CAP
2.150.08 mg/CAP
2.090.11 mg/Tab
1.930.07 mg/Tab
2.100.10 mg/CAP
2.140.09 mg/CAP
2.100.07 mg/Tab
1.920.10 mg/Tab
3.7.2. Selectivity
Table 5. Selectivity coefficients of various interfering compounds for loperamide hydrochloride
sensors
Interfering ion
Na+
K+
NH4+
Ca2+
Mg2+
ClNO3Lactose
Glucose
Log KMPM
(PVC membrane electrode)
-3.2
-3.4
-3.0
-4.1
-4.0
-3.9
-4.0
-4.4
-4.1
Log KMPM
(Nano-composite CPE)
-3.5
-3.6
-3.1
-4.1
-4.2
-3.7
-4.2
-4.5
-4.0
7652
Selectivity, which describes an ion-selective electrodes specificity toward the target ion in the
presence of interfering ions, is the most important characteristic of these devices. The potentiometric
selectivity coefficients of the loperamide hydrochloride electrode were evaluated by the matched
potential method (MPM) [55-60]. The resulting values of the selectivity coefficients are shown in
Table 5. Note that all selectivity coefficients are about 10-3, suggesting were interferences negligible
in the performance of the electrode assembly.
3.7.4. Ruggedness
For ruggedness of the methods a comparison was performed between the intra- and interday assay results for loperamide hydrochloride obtained by two analysts.
The RSD values for the intra- and inter-day assays in the cited formulations performed in the
same laboratory by the two analysts did not exceed 4.1%. On the other hand, the robustness was
examined while the parameter values (pH of the solution and the laboratory temperature) changed
slightly. Loperamide hydrochloride recovery percentages were good under most conditions, and not
showing any significant change when the critical parameters were modified.
4. CONCLUSIONS
In the present work, two types of potentiometric electrodes were constructed for
determination of loperamide hydrochloride. The sensors demonstrated advanced performances with a
fast response time, a lower detection limit of 1.010-5 mol L-1 for PVC membrane electrodes and
9.010-6 mol L-1 potential responses across the range of 3.010-5-1.010-2 mol L-1 and 1.010-51.010-2 mol L-1. The sensors enabled the loperamide hydrochloride determination in pharmaceutical
formulations. Both sensors respond based on ion-exchange mechanism. The best PVC membrane
electrode performance was achieved by a membrane composition of 30% PVC, 64% DBP, 1%
NaTPB and 5% ion-pair. Then, a carbon paste electrode was designed to improve the analytical
responses. The best electrode was composed of 20% ion-pair, 20% paraffin, 5% MWCNTs and 55%
graphite.
References
1. K. Lavrijsen, D. Van Dyck, J. Van Houdt, J. Hendrickx, J. Monbaliu, R. Woestenborghs, W.
Meuldermans, and J. Heykants, J. Pharmacol. Exp. Ther., 23 (1995) 354
7653
7654