Professional Documents
Culture Documents
www.elsevier.com/locate/ces
Received 10 August 2004; received in revised form 2 June 2005; accepted 19 June 2005
Available online 1 August 2005
Abstract
An inexpensive method was developed to measure local gas holdup based on withdrawal of the airliquid dispersion in bioreactors
under non-isokinetic conditions. The method was tested with several suction probe designs and withdrawal pressures using coalescing
and non-coalescing model media as well as during yeast fermentations. Simultaneously, gas holdup was measured manometrically. With
a straight end probe and vacuum pressure of 3 kPa, local gas holdup measurements presented a relative error < 6% compared to the
manometric method, independently of the media composition. This method was used to characterize local gas holdup in a bubble column.
Two new parameters are proposed to characterize non-gassed volume and gas holdup homogeneity.
2005 Elsevier Ltd. All rights reserved.
Keywords: Bioreactor; Bubble column; Fermentation; Hydrodynamics; Coalescing; Non-coalescing media
1. Introduction
Gas holdup is dened as the fraction of gas volume in
gasliquid dispersion and is commonly used to characterize
oxygen mass transfer and mixing of aerated vessels. Compared to global gas holdup, local gas holdup measurements
allow a ner analysis of hydrodynamics in aerated vessels.
Several methods have been developed for this purpose, based
on conductivity (Bin et al., 2001), optical (Schweitzer et al.,
2001), and gamma ray tomography (Kemoun et al., 2001)
detectors. These methods are time consuming and require
complex calibration or expensive equipment. Greaves and
Kobbacy (1984), Tabera (1990), and Yang and Wang (1991)
have proposed another approach, based on volumetric measurement of gas and liquid phases after they are withdrawn
from the gasliquid dispersion. Their method is simple and
Corresponding author. Tel.: +52 55 50 61 33 20;
fax: +52 55 50 61 33 13.
E-mail address: thalasso@mail.cinvestav.mx (F. Thalasso).
0009-2509/$ - see front matter 2005 Elsevier Ltd. All rights reserved.
doi:10.1016/j.ces.2005.06.010
inexpensive in as much as no specially designed equipment is required. However, Greaves and Kobbacy (1984) and
Hofmeester (1988) have stated that the aerated media must
be withdrawn under isokinetic conditions, that is, with the
same withdrawal speed as the speed of bubbles arriving at
the suction probe. In stirred media, Yang and Wang (1991)
recommended a withdrawal velocity in the sampling probe
equal to 50% of the stirrer tip speed, which is similar to
Hofmeesters (1988) recommendation of maintaining withdrawal conditions proportional to the kinetic conditions of
the aerated media. The purpose of isokinetic conditions is
to guarantee that the gas/liquid ratio E extracted is equal to
the gas/liquid ratio of the aerated media. Due to the difference of gravitational effect on the gas and the liquid phases
entering the probe, the ratio withdrawn would be different
from the true ratio if isokinetic conditions are not ensured
(Nevers, 1991).
The need to maintain either isokinetic conditions or
withdrawal conditions proportional to the conditions prevailing in the aerated media is a signicant limitation. This
6930
2. Experimental
2.1. Bubble column and operating conditions
Fig. 1(A) shows the glass bubble column (internal diameter: 0.12 m; height: 1.6 m; working volume 14 L) tted with
a glass porous plate (0.02, 0.04, 0.06 or 0.12 m diameter).
The bubble column was aerated at supercial gas velocities
vg ranging from 0.25 102 to 3.5 102 m/s. The air
supply was controlled by a needle valve and measured with
a variable area owmeter (Cole Parmer, USA).
During the experiment, distilled water was used as coalescing medium and 0.13 M KCl aqueous solution as noncoalescing medium. The total height Hd of the gasliquid
dispersion was kept constant at Hd /(2R) = 10.4, where R
is the column radius.
Measurements were also carried out in the same bubble
column during an aerobic batch fermentation with Candida
utilis NRRL-Y900. The medium contained initially (g/L):
glucose 13.0; yeast extract 1.0; (NH4 )2 SO4 1.37; NaH2 PO4
1.26, dissolved in tap water. Except glucose, all reagents
were of analytical grade. The operating conditions were as
follows: operating liquid volume 10 L; temperature 33 C.
The air ow rate was changed from 1.7 to 10.2 L/ min (vg
from 0.26102 to 1.50102 m/s) as explained in Section
4.5. The pH was automatically maintained at 4.5 using a
2 M NaOH solution and an automatic pH controller (B&C
(1)
R
0
2E(r)r dr.
(2)
6931
i
Vj
(3)
j =1
(4)
i
i
j =1 j Vj
Vtot
(5)
for 0 v Vdead ,
for Vdead v Vtot ,
(6)
where Vdead and the exponent 1 are solved for numerically in such a way as to make (6) best t the experimental
data. This eliminates the need to decide subjectively how
small a value of i is to be considered insignicant. A value
of = 1 describes a mixture completely homogeneous in
the complement of the dead region, while large values of
correspond to a strongly non-homogeneous mixture. Finally,
the relative dead volume is given by
0
Vdead
=
Vdead
,
Vtot
(7)
0i =
i
.
0
(8)
6932
0
The three dimensionless parameters 0 , and Vdead
give a
concise description of the state of the reactor. Fig. 2 presents
the four possible cases of gas holdup homogeneity (i.e., =1
0
0 > 0).
or > 1 combined with Vdead
= 0 or Vdead
(9)
i=1
When the straight and the inclined orices were used, E was
linearly correlated to the average manometric gas holdup
(r 2 > 0.99). The straight end probe presented a higher slope
and an ordinate to origin closer to zero. These results are in
accordance with Greaves and Kobbacy (1984), who reported
that straight capillary orices induce higher ow eld at
the suction point and therefore higher air withdrawal. The
straight end probe was selected for follow-up experiments.
4.3. Withdrawal pressure
4. Results
4.1. Overall operative conditions
A preliminary experiment was run in order to evaluate
the overall operative conditions required by the withdrawal
method. When the apparatus was operated at vacuum pressure > 35 kPa signicant differences between repetitions
were observed. At vacuum pressure < 1 kPa, the sampling
time was greater than 5 min and was considered too long.
Several probe orientations were also tested. The error between repetitions was signicantly reduced when the withdrawal probe was positioned horizontally instead of vertical
or inclined position. Therefore, it was decided to use the
probe in horizontal position and with vacuum pressure
within the 135 kPa range for follow-up experiments.
4.2. Probe design
Three probe designs were compared using distilled water
at vacuum pressure of 3 kPa. The average extraction parameter E was calculated by means of the discrete version of
= 0.55E 0.0094.
(10)
6933
Fig. 7. Local gas holdups obtained from the withdrawal method ( ) and
the local gas holdups measured with the manometric method (), as well
as the average gas holdups obtained from the withdrawal method ( - - - ), the manometric method (- - - - - - ), and the expansion method
( ). Operating conditions: vg = 3.42 cm/s, h/(2R) = 6.67 and
sparger diameter of 12 cm, using distilled water. Bars are 1 standard
deviation of three replicates. (-): dimensionless.
6934
Table 1
Global gas holdup obtained by withdrawal method and global gas holdup measured by expansion method
Sparger dia (m)
vg (cm/s)
Distilled water
KCL 0.13 M
0 (%)
0E (%)
Error (%)
0 (%)
0E (%)
Error (%)
0.12
0.54
1.71
3.42
2.32
8.54
16.14
2.13
7.85
14.74
8.1
8.1
8.6
2.32
10.90
25.79
2.27
11.33
24.11
2.1
3.9
6.5
0.06
0.54
1.71
2.17
6.05
2.01
6.54
7.3
8.1
2.33
7.79
2.19
8.06
6.0
3.4
0.04
0.35
1.71
1.17
4.80
1.07
4.52
8.5
5.8
1.55
5.27
1.44
5.36
7.0
1.9
0.02
0.20
0.54
0.44
1.39
0.40
1.49
9.0
7.1
0.71
1.57
0.65
1.45
8.4
7.6
vg ; supercial gas velocity, 0E : global gas holdup measured by expansion method, 0 : global gas holdup obtained by withdrawal method.
Fig. 8. Power law model ts to the local gas holdups obtained from
the withdrawal method at several supercial gas velocities vg . Sparger
diameter of 12 cm at h/(2R) = 6.75 using distilled water. vg = 0.034 m/s,
( ), p = 2.1, error = 0.9%; vg = 0.017 m/s (), p = 6.2, error = 1.2%;
vg = 0.005 m/s (), p = 7.0, error = 3.3%. Power law model (). Bars
are 1 standard deviation of three replicates. (-): dimensionless.
6935
Fig. 9. Radial and axial gas holdup distributions at vg = 0.54 and 1.71 cm/s for distilled water and KCl 0.13 M aqueous solution, with sparger diameters
0
and parameters. x axis values: r/R = 1, 0.75, 0.5,
of 2, 4, 6 and 12 cm. The homogeneity of the gasliquid dispersion are represented by Vdead
0.25, 0, 0.25, 0.5, 0.75, 1 (dimensionless), y axis values: h/(2R) = 0.16, 1.83, 3.45, 5.16, 6.75, 8.45, 10.10 (dimensionless), (local gas holdup/average
gas holdup): < 0.30 (), 0.310.80 ( ), 0.811.20 ( ), 1.211.40 ( ), > 1.40 ().
5. Discussion
Taking into account the total air and liquid withdrawn
volumes as well as the time needed to complete E measure ments, the withdrawn ow rate was estimated as
88 103 L/ min at 3 kPa vacuum pressure. This ow rate
corresponds to a suction velocity into the probe of 2.93 m/s.
According to Yu and Kim (1991), under similar operating
conditions to this work, a maximum radial distribution of
local bubble velocity from 0.40 up to 0.75 m/s is expected.
There is therefore no doubt that the suction speed was
clearly greater than the bubble speed and that the withdrawal
method presented here was operated under non-isokinetic
conditions.
The results obtained indicate that under non-isokinetic
conditions, the withdrawn air/liquid ratio is greater than the
6936
0
Fig. 11. Relative dead volume Vdead
for various sparger diameters, at
supercial gas velocity of 0.54 cm/s: distilled water ( ), solution of KCl
0.13 M (); and at supercial gas velocity of 1.71 cm/s: distilled water
(), solution of KCl 0.13 M (). (-): dimensionless.
6. Conclusion
The withdrawal method for local gas holdup measurement
is simple, versatile and inexpensive. This method can be
Notation
E
E0
E
f (v)
h
Hd
kL a
N
p
r
R
vg
V
Vdead
0
Vdead
VG
Vi
Vio
VL
Vtot
Greek letters
h
hw
i
oi
0
0E
h
6937
Acknowledgements
The authors acknowledge the nancial support of the Instituto Mexicano del Petrleo (IMP) through a FIES project
(FIES-98-107-VI) and CONACyT (Mxico). The authors
also thank C. I. Hernndez Gonzlez for his technical assistance and L. Dendooven and L. B. Flores-Cotera for their
critical reading of the manuscript. S. Garca-Salas was holder
of a grant from COFAA-I.P.N. and SUPERA and gratefully
acknowledges the COTEPABE-I.P.N., Mxico.
References
Akita, K., Yoshida, F., 1974. Bubble size, interfacial area, and liquid-phase
mass transfer coefcient in bubble columns. Industrial and Engineering
Chemistry, Process Design and Development 13 (1), 8491.
Bin, A., Duczmal, B., Machniewski, P., 2001. Hydrodynamics and ozone
mass transfer in a tall bubble column. Chemical Engineering Science
56, 62336240.
Chisti, M.Y., 1989. Airlift Bioreactors. Elsevier, England. pp. 294297.
Deckwer, W.D., 1992. Bubble Column Reactors. Wiley, New York. pp.
190191.
Frhlich, S., Lotz, M., Larson, B., Lbbert, A., Schgerl, K., Seekamp,
M., 1991. Characterization of a pilot plant airlift tower loop reactor: III.
Evaluation of local properties of the dispersed gas phase during yeast
cultivation and in model media. Biotechnology and Bioengineering 38,
5664.
Greaves, M., Kobbacy, K., 1984. Measurement of bubble size distribution
in turbulent gasliquid dispersions. Chemical Engineering Research
and Design 62, 312.
Hofmeester, J.J.M., 1988. Gas holdup measurements in bioreactors. Trends
in Biotechnology 6, 1922.
Joshi, J.B., 2001. Computational ow modeling and design of bubble
column reactors. Chemical Engineering Science 56, 58935933.
Kemoun, A., Rados, N., Li, F., Al-Dahhan, M., Dudukovic, M., Mills,
P., Leib, T., Lerou, J., 2001. Gas holdup in a trayed cold-ow bubble
column. Chemical Engineering Science 56, 11971205.
Kumar, S.B., Moslemian, D., Dudukovic, M.P., 1997. Gas-holdup
measurements in bubble columns using computed tomography.
A.I.Ch.E. Journal 43 (6), 14141425.
Lbbert, A., 1991. Characterization of Bioreactors. In: Schgerl, K., (vol.
Ed.), Measuring, Modeling and Control. In: Rehm, H.J., Reed, G.,
(Eds.), Biotechnology. Weinheim, New York, pp. 119120.
Miller, G.L., 1959. Use of dinitrisalicilic acid reagent for determination
of reducing sugar. Analytical Chemistry 31, 426428.
Nevers, N., 1991. Fluid Mechanics for Chemical Engineers. McGraw-Hill,
New York. pp. 449457.
Parasu, U.V., Joshi, J.B., 1999. Measurement of gas holdup proles in
bubble column by gamma ray tomography: effect of sparger design
and height of dispersion in bubble columns. Transactions of Institution
of Chemical Engineers 77 (A), 303317.
6938