You are on page 1of 44

User manual

User manual
• Viofor JPS Classic
• Viofor JPS Standard
• Magnetic & Light Therapy

Med & Life 2005


PART A

Prior to use of Viofor JPS System apparatus, please read the present manual.
The User Manual for Viofor JPS System Apparatus is not meant as the main source
of medical information. Any information necessary for optimal use of the apparatus
can be found in professional publications and at Symposiums organized by Med & Life.
It is also advisable to seek medical consultations available on Med & Life Medical
Infoline, share information and experience with other MD’s and to visit the site:
www.medandlife.com

Consult instructions for use

NOTICE:
please read the enclosed documentation!

Do not use during pregnancy


(see: contraindications)!

Do not use if wearing electronic implants


(see: contraindications)!

Beam of intensive red or/and infrared radiation! Protect your eyes!


Do not look at the light!
Direct eye contact with the applicator can cause risk
as in case of class 3A laser.

Fragments marked in italics present viofortherapy theory elements and are mainly
for doctor’s use. You do not need to know these parts of the manual in order to be able
to operate the apparatus.
Application of Viofor JPS System

Application of Viofor JPS System


INTRODUCTION
User Manual for Viofor JPS System has of physical medicine. Thanks to modern
been designed for a wide range of private technologies applied in medical devices,
users as well as health care professionals, the application and effectiveness of variable
i.e. medical doctors of various specializations magnetic fields in medicine has also been
(including dentists and veterinarians) as well expanded.
as physiotherapists and rehabilitation special- Viofor JPS System (further referred to as
ists, who use or are interested in application Viofor JPS) is a set consisting of: a control-
of variable magnetic fields in medicine. ler with basic and/or additional accessories.
In recent years, medical world has been Unique design solutions enable a wide ap-
intensively seeking new solutions in diagnos- plication of therapeutic properties of variable
tics and therapy, especially of noninvasive magnetic fields. Viofor JPS provides a selec-
character, as well as treatment methods not tion of physical parameters of variable mag-
sufficiently implemented until now. netic fields used in:
Dynamic development observed in math- 1. magnetostimulation in JPS System
ematics, physics and technology (especially 2. magnetostimulation in JPS system paired
electronics) has resulted in a new quality with light energy (optional)

1 CHARACTERISTICS OF VIOFOR JPS


1.1 Usage
Viofor JPS System can be used:
Viofor JPS System is used - in medical and physiotherapy clinics
for the following types of therapy: - in dentistry
- Magnetostimulation of the whole body - in rehabilitation
or its parts - in sport medicine
- Magnetostimulation paired - in veterinary medicine
with light energy - at home as recommended by a physician
of selected body parts - at home prophylaxis

2 TYPES OF THERAPY

2.1 Magnetostimulation and series 0,08 – 0,3 Hz, whereas median


induction of generated fields are significantly
2.1.1 Characteristics
lower than those used in magnetotherapy.
of magnetostimulation
Viofor JPS generates variable magnetic
Magnetic fields in magnetostimulation fields, which cause changes in the organism
generated by Viofor JPS system are impulse similar to those caused by physical activity
fields of complex impulse shapes and signal (physical strain, physical activity); therefore
structure of multi-peak frequency spectrum. they support biological regeneration. Both
Basic frequencies are within 180 – 195 Hz, mental and physical regeneration is tightly
frequencies of impulse packets are within connected with exercise. Variable magnetic
12,5 to 29 Hz, packet groups within 2,8 – 7,6 Hz, field plays a similar beneficial role. Magneto-

3
Application of Viofor JPS System
stimulation stimulates in the organism electro- - stabilizing cell membranes
motor forces of such properties, which enable - anti-inflammatory
homeostasis, and in case of its disturbance - relaxative
stimulate return to a normal state. - anti-spastic
Biological effect of variable magnetic field - analgesic
generated by Viofor JPS is the result of: Analgesic action takes place mainly
1. electrodynamic effect of such field on ion due to an increase in the secretion of endo-
currents in the organism genous opiates from the ß-endorphin group,
2. magnetomechanic effect of the magnetic a substance responsible for the increase in
field on particles with uncompensated pain threshold. The effect of analgesic action
magnetic spin occurs not only during exposure to magnetic
3. ionic cyclotron resonance of cations and fields, but is also observed after exposure is
anions in the body fluids. discontinued. This determines the biological
Body fluids i.e. blood, the lymph and inter- hysteresis of magnetic field effect. Precisely
cellular fluid constitute so-called second mode such a mechanism of analgesic action on
conductors of the ionic character of electrical the internal opioid system confirms an an-
conductivity. tagonist of opiates blocking this effect
The essence of the Viofor JPS system is (Naloxone).
the optimal choice of two different electrical The basis for regeneration effect is
(and thus magnetic) signals as well as their mainly the intensification of the processes
mutual combination, in order to achieve the of oxygen utilization and tissue respira-
aforementioned three fundamental mecha- tion as a result of intensified diffusion as
nisms of variable magnetic field absorption well as oxygen capture by hemoglobin and
by the living matter. ELF (Extremely Low Fre- cytochrome. The increased capture of oxy-
quency) electromagnetic radiation may affect gen is associated with a stimulation of the
an organism in various ways. In magnetostim- processes of tissue respiration, DNA syn-
ulation, due to extremely low value of magnetic thesis and acceleration of the mitotic cycle.
induction, the most important are bio-electric, Thanks to this, magnetic fields treat trophic
biochemical and bioenergetic effect. lesions of the shank, accelerate the heal-
1. The bioelectric effect causes normaliza- ing of ulcerous niches, improve the cellular
tion in the potential of the cell membrane. metabolism of the central nervous system
In pathological cases, the potential drops and accelerate regeneration of damaged
due to penetration through the cell mem- peripheral nerves.
brane to the cell interior of positive ions, Very weak currents generated due to in-
e.g. Na+. In order to reverse the process, teraction on piezoelectric substances e.g.
the cell needs energy, which can be ob- collagen, stimulate the activity of osteoblasts.
tained from ATP hydrolysis. It is aided by intensification of vasodilatatory
2. The biochemical effect is based on an effects. It has also been observed that a mag-
increased enzymatic activity as well as netic field inhibits demineralization processes
reduced processes related with ATP. of the bones and, at the same time, increases
3. The bioenergetic effect is a factor that their mineralization – which is measured by
stimulates the nutrition and growth of cells densitometry.
and regulates intercellular processes lead- Anti-inflammatory effect is connected
ing to regeneration of the organism. with the stimulation of formation of c-AMP
The main biological effects of magneto- and E prostaglandin. Prostaglandin influ-
stimulation mechanisms are actions: ences the accumulation of c-AMP, which
- regenerative decreases the secretion of inflammation
- vasodilatatory mediators from neutrophiles, basophiles,
- angiogenetic mastocytes and lymphocytes.

4
Application of Viofor JPS System
In case of the nervous system, magnetic - conditions after bone fracture
fields also improve the interneuron conductivity and breakage
and modulation of neuron activity as well as the - conditions after operations
pineal gland’s twenty-four hour rhythms of mela- on the osseous system
tonin secretion, which is called” the scavenger - conditions after operations in which
of free radicals” responsible, among others, for anastomosis has been carried
the aging process. out with screws or nails
The variable magnetic fields generated - pseudo joints
by the Viofor JPS clearly modify the ampli- 2. soft tissue:
tude value of alpha and theta wave rhythms - conditions after injury with or without
in proper directions, characteristic for the damage to skin continuity
physiological state of relaxation - objectively - burns
measurable and indicated by value of the de- - bedsores
mobilization of the autonomous system. The
results of psychophysical testing, as a part of III. Action improving peripheral circulation
psychotherapy of emotionally immature sub- 1. improper local blood flow in the limbs:
jects point to magnetostimulation as a method - in diabetic angiopathy
of choice for these people all the more due to - in atherosclerotic angiopathy
the fact that this is a non-invasive method ap- - in crural ulceration
plied equally in medicine as in psychology. 2. accelerated absorption of haematomas
3. decreased lymphatic edema/swelling
2.1.2 Indications for magnetostimulation
Indications of magnetostimulation as IV. Anti-spastic action
a supplementary therapy (therapy of choice) - conditions after apoplexy
include: - degenerative diseases of the nervous
system, accompanied by spasticity
I. Analgesic action
1. Osseous-joint system: V. Relaxative action
- degeneration of the spine - neuroses
and osseous-joint system - sleep disorders
of the upper and lower limbs - prophylactic calming action,
- rheumatoid inflammation of the joints to counteract everyday stress
- stiffening inflammation (rigor) - stress in the common meaning of the
of the spinal joints word, and particularly after long-term
- joint injuries psychic/mental tension
2. soft tissue: - concentration disorders
- extrajoint rheumatism - so-called
fibromyalgia VI. Action decreasing oxygen deficit through
- traumas of soft tissue effect on the processes of:
- conditions after injury (cutting) of tissue - aerobic tissue respiration
continuity (including post-surgical) - anaerobic tissue respiration
3. migraines
VII.Magnetostimulation in dentistry:
II. Regeneration action - Dental treatment, e.g.: biological pulp
1. osseous system: treatment, pulpitis treatment, complica-
- osteoporosis, particularly post-meno- tions related to pulp gangrene or para-
pausal (concerns prophylaxis as well dontitis treatment, periodontitis, compli-
as therapy) cations following endodontic treatment
- Sudeck syndrome - Paradontitis treatment

5
Application of Viofor JPS System
(anti-inflammatory treatment, fighting lack of information on the effect of magnetic
pain, tissue regeneration, treatment fields on the fetus, the application of variable
of post-surgical scarring) magnetic fields in pregnancy should be con-
- Diseases of mucous membrane sidered inadvisable.
of oral cavity, diseases of small salivary Neoplastic disease in the majority of
glands, gingivitis cases has been of a vague etiology, and the
- After surgery (24 hours), e.g.: after natural history of neoplastic diseases is also
tooth extraction, after gouging following known only in some cases. Experimental data
apicectomy, in implant therapy, regarding the effect of magnetic fields on neo-
- Complications after dental surgery plastic cell cultures and animals with an active
(e.g. infected dry socket) neoplastic disease are contradictory. In accor-
- Complications after anesthesia, nerve dance with the principle primum non nocere,
damage variable magnetic fields should not be applied
- Orthodontic treatment (during treatment in a neoplastic disease, although many data
with permanent devices) indicate a positive effect of magnetic fields on
- Prosthetic treatment (after prosthesis neoplasms.
application) Active tuberculosis of the lungs is
a contraindication to the application of mag-
2.1.3 Contraindications netic fields, due to their interactive effect on
to magnetostimulation tissue. The influence of magnetic fields on
Contraindications apply to magnetostim- the utilization of oxygen, as well as on the
ulation as a method, and not the blood flow, is beneficial. Koch’s bacilli are
Viofor JPS apparatus. Emitted fields are aerobes, therefore, based on knowledge of
a contraindication in case of an existing the biological effects of variable magnetic
disease – whereas they are not a disease- fields, in case of TBC, they should not be
developing factor. The diseases mentioned applied.
in contraindications are caused by different Bleeding from the alimentary tract
factors. At present, there is no information should be divided into massive hemorrhag-
in literature regarding contraindications ing – from the varices of the gullet and
to magnetostimulation. However, in view peptic ulcers – and bleeding accompanying
of the comparatively short period gastritis type diseases. In case of intensive
of existence of magnetostimulation in rela- bleeding, the therapy of choice is an endo-
tion to magnetotherapy, the contraindica- scopic or surgical intervention. In case of
tions should be the same as in case bleeding observed with gastritis erosiva and
of magnetotherapy, i.e.: gastritis haemorrhagica, caution is also rec-
- pregnancy ommended in the application of magnetic
- active neoplastic disease fields. In case of gastric ulcer the relaxative
- active tuberculosis of the lungs and anti-spastic action of variable magnetic
- bleeding from the alimentary tract fields could be considered advisable, nev-
- serious infections of a viral, bacterial ertheless the decision of their application
and mycotic origin in this disease should be consulted with
- the presence of electronic implants a doctor, since it is frequently accompanied
- conditions after organ transplants by gastritis
Pregnancy is a state in which the dynam- Serious infections of a viral, bacterial
ics of forming new tissues is the greatest. and mycotic origin are a contraindication
As a rule, apart from necessary measures, to the application of magnetic fields. Each
everything else should be avoided during preg- type of these diseases should be treated
nancy. Until now, the influence of magnetic with antiviral, antibiotic and antimycotic phar-
fields on the fetus has been unknown. Due to maceuticals.

6
Application of Viofor JPS System
Electronic implants applied in mod- posure and affects people with significant
ern medicine, in most cases have good degenerative changes of the osseous-joint
screening from external interference. From system.
physical point of view, the fields applied in Summing up, having taken into consider-
magnetic field therapy, particularly in mag- ation the above observations, it should
netostimulation, have little influence on the be stated that the application of variable mag-
operation of electronic implants. The person netic fields is safe and the progress of ex-
ordering the therapy rarely has knowledge perimental and clinical knowledge will in all
of the electronic implant’s screening qual- likelihood result in reduction of the number
ity that is why it is safer for the patient to of contraindications.
refrain from exposure to variable magnetic
fields. 2.2 Magnetostimulation paired
with light energy
2.1.4 Side effects, which may occur due
to magnetostimulation
Until now, no serious side effects of mag- 2.2.1 Characteristics
netostimulation have been reported. The fol- of magnetostimulation paired
lowing effects might occur temporarily: with light energy
- tingling sensation The method of magnetostimulation paired
- numbing sensation with light energy utilizes the magnetic field
- warmth sensation effect described in paragraph 2.1 of the
- sleep disorders, particularly in elderly Manual.
people Light energy has a mainly topical effect
- irritation on tissues. Its body penetration depends
- concentration disorders on the light wavelength. Tissue reaction
- temporary intensification of pain. depends on energy absorption in each tis-
Tingling, cramped and warmth sensa- sue layer. The absorption is mainly affected
tion may occur particularly in the first three by: tissue thickness, blood supply and flow,
or four days of exposure. This applies in par- water content and presence of pigment.
ticular to post-traumatic areas of the osseous Infrared reactions start at the level of cell
system. membrane, and reactions to red light – in
Sleep disorders are observed only in ap- mitochondria.
prox. 3% of elderly patients. This could be At the tissue level, mechanism of biologi-
easily eliminated by exposure to increasing cal effect of variable magnetic fields and light
level of induction. energy are of similar nature. When applying
Irritation and/or concentration dis- magnetostimulation together with light, a syn-
orders concern particularly women during ergy effect could be expected, depending on
menopause as well as people with increased individual characteristics.
tension of the neurovegetative system. The Main biological effects of magnetostimu-
above symptoms are usually observed in lation paired with light energy should be as
the first days after exposure. The frequency follows:
of their occurrence may decrease through - improved oxygen assimilation by its car-
the application of an increasing value of in- riers, which improves ATP synthesis in
tensity. It is also advisable to inform patients oxidation-reduction systems aerobic and
about possibility of side effects. The above anaerobic respiration path; increased
symptoms affect approx. 4% of women and ATP concentration is directly related to
1% of men. the increased activity of ATP-dependent
Temporary intensification of pain enzymes, responsible, among oth-
usually occurs during the first period of ex- ers, for protein and collagen synthesis,

7
Application of Viofor JPS System
which better stimulates tissue regenera- - joint strain and injuries
tion process. - rheumatoid inflammation of the joints
- Strong stimulation of DNA synthesis and - stiffening inflammation (rigor) of the spi-
cell proliferation nal joints
- Vasodilatation effect due to direct relax- 2. soft tissue:
ation of vessel smooth muscular coat, - extrajoint rheumatism – so-called
related with accelerated processes of fibromyalgia
angiogenesis and tissue perfusion, which - traumas of soft tissue
improves regeneration - zoster
- Direct effect on liquid crystal mem- - neuralgia
brane structure and on modification of
membrane enzyme activity related with II. Regenerative effect
increased ATP-dependent ion pumps - skin and soft tissue inflammation
through increased ATP synthesis in mi- - skin and soft tissue post-inflammation
tochondria; each of the above processes condition
leads to changed membrane permeability, - psoriasis
thus to change in electrolyte and water - burns
distribution between cells or their organelle - bedsores
and surrounding areas; both processes - chronic and subacute arthritis
have a strong anti-inflammatory and anti- - initial treatment preceding massage,
swelling effect kinesitherapy and iontophoresis
- Restored ATP-dependent membrane
pump activity, which has been lost due III. Improvement of peripheral circulation
to disease-related changes – sodium-po- 1. improper local blood flow in the limbs:
tassium and calcium pumps in neurons, - in diabetic angiopathy
which leads to membrane hyperpolariza- - in atherosclerotic angiopathy
tion and stimulus blocking, even those of - in crural ulceration
high amplitude; as a result, pain stimuli 2. accelerated absorption of haematomas
conduction in afferent fibers is decreased, 4. decreased lymphatic edema
therefore a clear analgesic effect can be
observed IV. Anti-spastic effect
- Changed activity of serotoninergic - degenerative diseases of the nervous
synapses and stimulation of beta-endor- system, accompanied by spasticity
phin release and stimulation
of pituitary gland-adrenal gland axis V. Decreased oxygen deficit
accompanied by increased release and debt through affecting
of clycocorticoids adds to a strong the processes of:
analgesic effect - aerobic tissue respiration
- Regulation of immune system and rheo- - anaerobic tissue respiration
logical blood properties
2.2.3 Contraindications
2.2.2 Indications for magnetostimulation to magnetostimulation and light energy
with light energy treatment
Although there is no proof for negative
I. Analgesic effect effect of magnetostimulation on each meta-
1. osseous-joint system: bolic function, contraindications should be
- degeneration of the spine and osse- the same as in case of variable magnetic
ous-joint system of the upper and lower field (see paragraph 2.1.3.). Special
limbs attention should be paid to local neoplastic

8
Application of Viofor JPS System
changes and changes that are difficult 2.2.4 Side effects which may occur due
to identify, predisposition to bleeding to magnetostimulation with light energy
as well as acute inflammation of skin treatment
and soft tissue in the area exposed No significant side effects should be ex-
to therapy. Patients who are administered pected. The following effects might occur
drugs, which cause photosensitization temporarily:
or show symptoms of photosensitization - tingling sensation,
should not be treated – it is easy - numbing sensation,
to verify by checking patient’s - warmth sensation,
history. - skin erythema in the area of treatment.

3 THERAPY SELECTION
Viofor JPS apparatus is a unique and 5. Due to the possibility of sleep disorders, par-
original design that provides the user with the ticularly in elderly patients, treatment should
selection of one of three options of variable be avoided late in the evening. It particularly
magnetic field therapy: concerns treatment of the head and neck
1. magnetostimulation in JPS System area with the use of a pad, elliptic and point
2. magnetostimulation in JPS System paired applicator as well as ring applicators.
with light energy 6. Therapy should be preceded by applica-
tion of dEF1 or dEF2 program (see below),
In order to establish the application param- since it allows to determine individual sen-
eters and therapy options for a given patient, sibility to magnetic field therapy. The lists
the following factors should be considered: of programs are available for Classic version
- age
- general health Prophylaxis – based on default parameters
- course of the main illness preprogrammed by manufacturer
- co-existing illnesses (dEF1 and dEF2 lists are used only
- administered pharmaceuticals for magnetostimulation)
- the number of treatments and their duration
- other types of physical therapy.

3.1 Magnetostimulation
1. Therapy should be conducted upon
a diagnosis and doctor’s consultation
2. Magnetostimulation, like other forms of
physical therapy, usually does not bring
immediate effects. They depend on the
nature of the illness, its length and coexist-
ing illnesses. The individual reaction to this
type of treatment is also important.
3. Magnetostimulation complements primary
treatment. Changes in treatment (par-
ticularly pharmacological) should be intro-
duced only after doctor’s consultation.
4. When lying down on a large applicator
(mat, frame bed), the head should always
be positioned near the cable outlet.

9
Application of Viofor JPS System
A table containing applicator combinations
is on page 11. The applicators, when used
simultaneously on a patient, intensify the
effect of variable magnetic field, and, there-
fore, accelerate the therapeutic effect. Two
elliptic applicators of the same color cannot
be used on one patient in respect to the
same body part. It is, however, possible to
use any two applicators on two patients.
Other combinations require consultations
with the distributor.

3.1.1 Application parameters components

3.1.1.1 Programs (P)


1. P11) – JPS system with one impulse type
of simplified structure
2. P2 – JPS system with two impulse types
3. P31) – JPS system with two impulse types
of different impulse order and structure as
compared with P2
only (displayed by remote control). Stan-
dard version users can set the below-listed 3.1.1.2 Application modes (M)
parameters (M and I) manually – except for 1. M1 – application with a constant intensity;
program (P), which is a default program and the selected field intensity is constant
equivalent of P2 program in Classic version. throughout the whole time of application;
7. Each applicator can be plugged in any recommended in persons of good health
socket of Viofor JPS Classic controller. If
a default program has been selected, i.e.
dEF1 and dEF2, applicators assigned to
these programs should be plugged in ac-
cordance with markings on the controller
and tables indicated in item 6.
8. Magnetostimulation with the use of a large
applicator (cot, mat, short mat) can be ap-
plied in intervals of minimum six hours, two 2. M2 – application with increasing intensity;
or three times within 24 hours. the degree of field intensity increases
9. In case of long-term therapy in chronic ill- every 10 or 12 seconds from 0,5 to the se-
nesses, a week interval is recommended lected intensity, cyclically during the time
after several weeks of application. of application; recommended in persons
10. During therapy it is not possible to interfere of lowered general condition
with the application’s parameters – all keys,
apart from START/STOP and SOUND, are
inactive at this time.
11. Possible subjective side effects are usually
of temporary character.
12. Usage of an adapter enables to use two ap-
plicators at the same time without possibility
of applying different application parameters.
1)
for version Classic

10
Application of Viofor JPS System
3. M3 – application with an increasing-de- 0,5. For older people and children, chang-
creasing intensity; the degree of field inten- es may be introduced every 5-7 days, for
sity increases in 2 min. from 0,5 to the se- others every 1-2 days.
lected value, during the next 8 min. remains Proposed magnetic field intensity for
at the selected level, during the last 2 min. adults:
decreases to a level of 0,5; apply from an - level 0,5 do 121) (102)), depending on age
intensity of 3 or higher. and individual sensitivity to magnetic fields.
Proposed intensity for children:
- 0-2 years level 0,5-1
- 2-5 years level 1-2
- 5-11 years level 1-4
- 11-15 years level 1-6

3.1.1.4 Selection of application


parameters
Application parameters are selected
3.1.1.3 Magnetic field intensity ( I ) through a combination of: program (P)1),
Intensity from 0,5 to 12 corresponds mode (M) and intensity (I)
with proportional values of magnetic field 1. P11) + M1 – application with constant in-
emitted by the applicators. It is recom- tensity (8 min.)
mended to start therapy from magnetic 2. P11) + M2 – application with increasing
field intensity set at 1 and successively set intensity (12 min.)
to higher levels. The speed of setting to 3. P11) + M3 – application with increasing-de-
higher levels depends on the age and gen- creasing intensity (12 min.)
eral health of the user. For persons of high
sensitivity to magnetic fields it is recom- 1)
for version Classic
mended to start application from a level of 2)
for versioni Standard

Applicator combinations

11
Application of Viofor JPS System
4. P21) 2) + M1 i P31) + M1 – JPS system with pole direction. It is important in case of using
constant intensity (10 min.) of two applicators in one patient.
5. P21) 2) + M2 i P31) + M2 – JPS system with
increasing intensity (12 min.) NOTICE!
6. P21) 2) + M3 i P3 + M3 – JPS system with
increasing-decreasing intensity (12 min.) Never use two applicators with the same
The recommended modes of applica- color sides facing each other.
tion are based on wide clinical experience
of the teams of practicing physicians as well When a patient’s body or a part of the body
as multicenter research that evaluate the ef- is placed between two applicators generating
fectiveness of magnetostimulation with the heterogeneous magnetic field, the effect is
Viofor JPS apparatus. close to the homogenous magnetic field.
Each of the application types can be com- 1 m connection between two elements of
bined with selected field intensity according to the flat applicator enables simultaneous ex-
the above-mentioned instructions. posure of two body parts.
In order to establish the application param- Two or three large applicators can be used
eters for a given patient, the following factors at the same time (connected with one or two
should be considered: adapters) as well as different combinations of
- age applicators (e.g. large and point applicator, large
- general fitness and small, large and point). Point applicators
- course of the main illness work with large or small applicators facing the
- co-existing illnesses white side up. In case of using applicators in
- pharmaceuticals administered two or three patients at the same time, therapy
- time of the day, when therapy parameters are identical for all patients. There is
will be carried out no possibility of setting separate parameters for
- duration of magnetostimulation each applicator when used simultaneously.
- other types of physical therapy.
3.2 Magnetostimulation
3.1.2 Using more than one applicator with light energy
Additional accessories (adapter, extra ap-
plicators) enable optional usage of several CAUTION!
applicators in one or more patients in order to
increase the therapy effectiveness or to use STRONG RED AND/OR
the apparatus more economically. INFRARED RADIATION!
Example: PROTECT YOUR EYES!
– patient with rheumatoid inflammation of DO NOT LOOK AT THE LIGHT!
joints and significant degenerative chang-
es of hip joints – simultaneous application 1. Magnetostimulation paired with light en-
of a mat (whole body) and a pad (hips) ergy is used locally during the series of
– several large applicators used for several repeated treatments with the use of one
athletes, e.g. after training. of three magnetic-light applicators.
Two-color coding of applicators is used to 2. Each applicator generates a heteroge-
identify the direction of magnetic field poles neous variable magnetic field in the JPS
generated by an applicator. In case of the system. The light area of 48 diodes is lo-
same impulse direction, each element of the cated in within the coil and equals
applicator when, for example, facing white 19,6 cm2.
side up, will have the same magnetic field 3. Light energy penetrating the tissue de-
pends on the radiation properties (wave-
1)
for version Classic
lengths, angle of incidence), and on tissue
2)
for version Standard

12
Application of Viofor JPS System
properties (rate of hydration, perfusion, tion with constant intensity, where selected
pigment content). intensity of variable magnetic field does not
4. Tissue penetration will be the highest in change during the application.
case of infrared magnetic-light applica-
tor (48 diodes, wavelength 855 ± 5 nm 3.2.1.3 Magnetic field intensity ( I )
and one diode, wavelength 630 ± 5 mn), Intensity from 0,5 to 121) (102))
intermediate in case of combined mag- corresponds with proportionate values of
netic-light applicator (24 diodes, wave- variable magnetic field emitted by elliptic
length 855 ± 5 nm, 24 diodes, wavelength applicators. The highest mean impulse
630 ± 5 nm), and the lowest in case of value of variable heterogeneous magnetic
red magnetic-light applicator (47 diodes, field (in microteslas) on the surface of mag-
wavelength 630 ± 5 nm). netic-light applicators is 8.9 x 1 at P2M1
5. Energy power of each light type will in- (intensity level selected on the Viofor JPS
crease to intensities set on the controller. controller).
Between 0.5 – 6 it increases proportionally It is recommended to start the therapy
to intensity, and between 6 – 12 – it re- from the lowest intensity and increase it
mains on the same level. gradually.
6. Treatment with magnetic-light applicators
should be started from low induction val- 3.2.1.4 Selection of application parameters
ues of variable magnetic field, due to the Application parameters are selected
expected synergy of magnetostimulation through a combination of one of three pro-
and light energy. grams (P11), P22) or P31)), application mode
7. If the magnetostimulation has started with M1 (recommended) and magnetic field inten-
the use of elliptic applicator, then it is rec- sity (I).
ommended to lower down the induction of Treatment duration is set automatically
magnetic-light applicator by 50% as com- depending on P and M combination.
pared to the previous one in respect to the When the induction in Viofor JPS con-
same body area. troller increases from 0.5 to 6, light energy
8. Due to the applicator contact with the skin, density increases as well, whereas above
it is recommended to wash and dry the these values, i.e. from 6 to 12, only variable
skin area to be exposed. magnetic field induction will increase. Due to
9. When using magnetic-light applicators, the expected synergy of both types of radia-
never direct the light in the eye area! tion, the therapy should be started from low
induction values of variable magnetic field,
3.2.1 Application parameter components and, therefore, from low light energy density
level.
3.2.1.1 Programs ( P ) Selection of one of three magnetic-light
When using magnetic-light applicators, all 3 applicators should depend on the type of
programs are recommended, i.e. P11) P22) and light required for a given form of therapy.
P31). Energy light parameters are indicated in
technical data tables, pp. 32-34. 1. Red magnetic-light applicator R
(wavelength 630 ± 5 nm) is used mainly
3.2.1.2 Application modes ( M ) for skin disease treatment, such as
Viofor JPS apparatus enables three unhealing wounds and ulcerations
different application modes (M) for vari- (including bedsores), common acne and
able magnetic field described in paragraph zoster. It is recommended in dermatology
3.1.1.2. and cosmetic surgery.
When using magnetic-light applicators, 1)
for version Classic
M1 mode is recommended. It is the applica- 2)
for version Standard

13
Application of Viofor JPS System
2. Infrared magnetic-light applicator IR
(wavelength 855 ± 5 nm) in motor sys-
tem diseases (arthrosis; pain syndromes
in discopathy; syndromes due to strain
of muscles and periarticular soft tissue;
inflammation of tendons, fascia, tendon
sheaths and joint bursa; periarticular in-
flammation; bone fractures and delayed
union of fractured bones) as well as in
neurological diseases (neuralgia of pe-
ripheral nerves – intercostal neuralgia,
trigeminal neuralgia, post-zoster neural-
gia, sciatica; diabetic neuropathy).

3. Combined magnetic-light applicator R+IR


(wavelength 630/855 ± 5 nm) in chronic
skin inflammations; keloid, psoriasis, psori-
atic joint inflammation, ankylosing spondy-
litis, rheumatoid inflammation of the joints,
post-thrombotic syndrome.

Energy doses as well as type of light and


the number of treatments should be selected
individually; their parameters should be simi-
lar to those used in laser biostimulation, de-
pending on experience and preferences
of the therapist.

14
Usage of Viofor JPS System

PART B

15
Usage of Viofor JPS System

Usage of Viofor JPS System


4 VIOFOR JPS CLASSIC

4.1 General Information


The front panel of Viofor JPS Classic contains: membrane keyboard control unit, applica-
tion time display and LED diode indicating selected settings and passage of current through
an applicator. These elements are in the main control panel of the apparatus.

4.2 Viofor JPS Classic back panel

Drawing 1: Viofor JPS Classic back panel

1 – power switch 5– socket A3


2 – voltage switch (default setting 200/230V) 6– socket A2
3 – fuse base with a 0,25 A (250 mA) delay 7– socket A1
switch 8– service socket
4 – power cable – DO NOT USE!

4.3 Viofor JPS Classic front panel

Drawing 2: Viofor JPS Classic front panel

16
Usage of Viofor JPS System
1 – START/STOP button – starts or stops as well as programmed and programmable
an application lists.
2. Light-emitting diode / LED (yellow), which 11. LED diodes (green) signal current
flashes when the apparatus is operating polarization of the field impulses
3. Button to increase the magnetic field in- 12. A key, for applicator socket selection
tensity (>) 13. Light-emitting diode / LED, which signals
4. Button to decrease the magnetic field in- selection of A1 socket – mat: lights green
tensity (<) if the applicator circuit is closed, red – if
5. A series of light-emitting diodes / LED there is a break in the applicator’s circuit
(7 green, 3 yellow, 3 orange), which signal 14. Light-emitting diode / LED, which signals
the set and used magnetic field intensity selection of the A2 socket – pad, color
6. SOUND button – switches on / off beeper lighting as in point 13
signaling of apparatus operation 15. Light-emitting diode / LED, which signals
7. Light-emitting diode / LED (green), selection of the A3 socket, color lighting
which lights up whenever the beeper as in point 13
is switched on 16. M button (MODE) – mode of magnetic
8. ON button, which activates the apparatus field application
from STAND-BY mode and restores indi- 17. Light-emitting diodes / LED (green), which
cator lighting signal the chosen application mode: M1,
9. Light-emitting diode / LED (green), which M2 or M3
signals switching on of the apparatus 18. P button (program) – selection of the ap-
found in STAND-BY mode, when all indi- propriate application program
cators are unlit 19. Orange light-emitting diodes / LED, which
10. Four-digit display, which indicates applica- signal the selected program: P1, P2 or P3
tion time (before pushing START/STOP 20. Infrared receiver window
button), or remaining application time (for remote control)

4.4 Remote control unit

Remote control enables fast programming and setting.

4.4.1 Remote control details


1– START/STOP button switches
the application on and off
2. ON button - activation
of the apparatus from STAND-BY
mode and restoration of indicator
lighting
3. A1, A2, A3 applicator selector
buttons
4. M1, M2, M3 application mode
selector buttons
5. P1, P2, P3 program selector Drawing 3: Remote control unit
buttons
6. 0,5 to 12 intensity selection buttons 8. SEL button selects successive
7. B, C, D, E active set application application parameter sets
parameter selector buttons LI1, LI2, dEF1, dEF2

17
Usage of Viofor JPS System
9. OK button confirms set selection (pre- 4.6 Apparatus operational use
ceded by the SEL button) or monitoring of
the active set on the apparatus front panel 1. Press the power switch (1; see fig. 1,
display (not preceded by the SEL button) p. 16) on the back panel of the appara-
10.SOUND button - switches the beeper on tus. The parameters will be set after
and off, to indicate whether the unit is in a short test:
operation Applicator A1 – mat
11. CODE button – in standard version used (mat socket)
as one of the elements enabling counter Program P1
reading; in special version conditions the Mode M1
switching on of the apparatus (application with constant intensity)
Intensity – 1
4.4.2 Battery installation Time – 8:00 min
1. Remove the battery cover on the back of 2. Press the A button (on the control unit) or
the remote control unit in the direction indi- A1, A2, A3 (on the remote control) to select
cated by an arrow the applicator – the green diode correspond-
2. Place in the correct position (in accor- ing to this applicator will light up. If the LED
dance with the diagram inside the contain- lights red, this signals a circuit break or too
er) two AAA alkaline batteries (or equiva- large resistance in the applicator circuit
lent) (e.g. a loose pin).
3. Replace the battery cover 3. Press the P button to select the appro-
ADDITIONAL INFORMATION priate program (P1, P2, P3). An orange
- use only alkaline AAA batteries LED will show the program currently se-
- batteries will last depending on remote lected.
control usage, however they should be 4. Press the M button to choose the appropri-
changed at least once a year ate mode of application (M1, M2, M3).
- if the remote control unit is not used for A green LED will show the mode currently
a long time, remove the batteries selected.
- batteries should be removed during 5. Press the „>”(3) or „<”(4) buttons on the ap-
transportation paratus front panel or ”0,5” to ”12” buttons
- do not mix old batteries with new or on the remote control unit to increase or
batteries of different types decrease the intensity to the selected level.
- do not connect the batteries to an elec- A series of lit LED shows the selected level
tric current, take them apart, heat or of intensity.
put into fire. CAUTION! Each time the intensity button
Change of batteries is not covered by the on the front panel is pressed, the intensity
warranty and is within user’s discretion. level changes by one point - it is signaled
by a beep. With a change of program (P)
4.5 Apparatus installation and application mode (M), the application
time shown on the display also changes
Load batteries in the control unit. automatically (8 min, 10 min, 12 min).
Set the control unit on a stable base 6. After setting the appropriate application
(e.g. on a table) and connect the applicators. parameters press the START/STOP but-
Plug in the cable. If a mat is used, it should ton – this is accompanied by a prolonged
be laid on a flat, hard surface at a minimum beep. The clock is then activated to
distance of 50 cm from metal objects count down the application time in sec-
(e.g. radiators). onds from the set value to zero. During
For best effects, select places free from the application, a yellow LED pulsates
the interference of other magnetic fields. next to the START/STOP button. The

18
Usage of Viofor JPS System
application stops automatically after the By repeating the procedure described
set time has elapsed and is signaled by in points 2-6, 8 different sets of application
3 beeps. parameters can be programmed (4 for set
7. When START/STOP button is pressed LI1 and 4 for set LI2). The parameters will be
all buttons apart from START/STOP and stored in the apparatus’s memory until re-
SOUND will become inactive – stop re- placed by new ones.
sponding.
CAUTION! The START/STOP button will 4.8 Retrieving preprogrammed
not work if the circuit of the chosen applica- application parameters:
tor is broken (the applicator’s LED will light
red). If a break in the circuit occurs during 1. Press the ON button to activate the ap-
the application, it will be interrupted, the paratus
display will show “Err” and then the initial 2. Press the OK button on the remote con-
time of the application. trol unit to check which set is active – if
8. Depending on requirements, press the the required set is displayed, proceed
SOUND button to switch on or off the to the action described in pt.5, and then
sound signaling operation of the apparatus. onto pt. 6; if not:
9. If the apparatus is inactive for longer than 3. Press the SEL button on the remote
2 minutes, it switches to as STAND-BY control unit repeatedly until the required
mode. The green LED then lights up. To set is displayed
re-activate the apparatus press ON button. 4. Press the OK button on the remote con-
trol unit to confirm the set selection
4.7 Programming of application 5. Press one of the B, C, D or E buttons
parameters on the remote control unit to retrieve
selected preprogrammed application
Programming applies to sets LI1 parameters
and LI2. Sets dEF1 and dEF2 6. Press the START/ STOP button to start
(see item 6 p. 9) are permanently programmed the application.
by the manufacturer.
To program application parameters follow 4.9 PLACEBO mode
the instructions:
1. press the power switch (1; see fig. 1, p. 16) When the apparatus is off, press and hold
on the back panel of the control unit or, if it the intensity ”>” (3) button and press the
is in STAND-BY mode, activate it with the power switch. After the message ”8:00 min”
ON button. appears on the display release the ”>” (3)
2. set the required applicator (A) button. The apparatus will simulate all func-
3. select the application parameters i.e. pro- tions, but impulses will be sent to a subsid-
gram (P), mode (M) and intensity iary circuit. All applicators will remain inactive
4. press the SEL button on the remote control at this time.
unit to retrieve set LI1 or LI2
5. press the ON button on the front panel of 4.10 Counter
the control unit – the message SAVE will
appear on the display When ON-CODE-ON buttons on the
6. on the remote control unit press one of the remote unit are pressed in sequence within
B, C, D or E buttons, to store the applica- 1 second, the counter will appear on the
tion parameters display. Each digit is separated by a pe-
7. a prolonged beep informs us riod. The treatment is measured if longer
that the application parameters than 30 seconds. The user cannot reset
have been stored. the counter.

19
Usage of Viofor JPS System
5 VIOFOR JPS STANDARD
5.1 General Information dicating selected settings and passage of
The front panel of Viofor JPS Standard current through an applicator. These ele-
contains: membrane keyboard control unit, ments are in the main control panel of the
application time display and LED diode in- apparatus.

5.2 Viofor JPS Standard back panel

Drawing 4: Viofor JPS Standard back panel

1 – power switch 4 – power cable


2 – voltage switch (default setting 200/230V) 5 – applicator socket
3 – fuse base with a 0,25 A (250 mA) 6 – service socket
delay switch – DO NOT USE!

5.3 Viofor JPS Standard front panel

Drawing 5:
Viofor JPS Standard front panel

1 – START/STOP button – starts or stops 3. Button to increase the magnetic field


an application intensity (>)
2. Light-emitting diode / LED (yellow), 4. Button to decrease
which flashes when the apparatus the magnetic field
is operating intensity (<)

20
Usage of Viofor JPS System
5. A series of light-emitting diodes / LED If an applicator is plugged in, a green di-
(7 green, 3 yellow, 3 orange), which signal ode will light up. If the diode turns red, this
the set and used magnetic field intensity signals a circuit break or too large resistance
6. SOUND button – switches on / off beeper in the applicator circuit (e.g. a loose pin).
signaling of apparatus operation 2. Press the M button to choose the appropriate
7. Light-emitting diode / LED (yellow), which mode of application (M1, M2, M3). A green
lights up whenever the beeper is switched on LED will show the mode currently selected.
8. ON button, which activates the apparatus 3. Press the „>”(3) or „<”(4) buttons on the
from STAND-BY mode and restores indi- apparatus front panel or ”0,5” to ”12” but-
cator lighting tons on the remote control unit to increase
9. Light-emitting diode / LED (green), which or decrease the intensity to the selected
signals switching on of the apparatus level. A series of lit LED shows the select-
found in STAND-BY mode, when all indi- ed level of intensity.
cators are unlit CAUTION! Each time the intensity button
10.Four-digit display, which indicates applica- on the front panel is pressed, the intensity
tion time (before pushing START/STOP but- level changes by one point - it is signaled by
ton), or remaining application time as well as a beep. With a change of program (P) and
programmed and programmable lists. application mode (M), the application time
11. LED diodes (green) signal current polar- shown on the display also changes automati-
ization of the field impulses cally (8 min, 10 min, 12 min).
12.A LED diode, lights green if the applicator 4. After setting the appropriate application
circuit is closed, red – if there is a break in parameters press the START/STOP button
the applicator’s circuit – this is accompanied by a prolonged beep.
13.M button (MODE) – mode of magnetic The clock is then activated to count down
field application the application time in seconds from the set
14.Light-emitting diodes / LED (green), which value to zero. During the application, a yellow
signal the chosen application mode: M1, LED pulsates next to the START/STOP but-
M2 or M3 ton. The application stops automatically after
the set time has elapsed and is signaled by 3
5.4 Apparatus installation beeps.
5. When START/STOP button is pressed all
Set the control unit on a stable base (e.g. buttons apart from START/STOP and SOUND
on a table) and connect the applicators. Plug will become inactive – stop responding.
in the cable. CAUTION! The START/STOP button will not
If a mat is used, it should be laid on a flat, work if the circuit of the chosen applicator is
hard surface at a minimum distance of 50 cm broken (the applicator’s LED will light red).
from metal objects (e.g. radiators). If a break in the circuit occurs during the ap-
For best effects, select places free from plication, it will be interrupted, the display
the interference of other magnetic fields. will show “Err” and then the initial time of the
application.
5.5 Apparatus operational use 6. Depending on the needs the button SOUND
can switch on / off beeper signaling of appa-
1. Press the power switch (1; see fig. 1, p. 16) ratus operation.
on the back panel of the apparatus. The pa- 7. After 2 minutes of idleness the apparatus
rameters will be set after a short test: switches in the STAND-BY state, which is ac-
Mode M1 companied by turning off whole signalization.
(application with constant intensity) Then the green diode lights up. Reactivation
Intensity – 1 of the apparatus can be done by pressing
Time – 10:00 min the ON button.

21
Usage of Viofor JPS System

6 MAGNETIC & LIGHT THERAPY


6.1 General Information 6.2 ZAMS 01/2 power supply unit
Magnetic & Light Therapy set is used ZAMS 01/2 power supply unit enables
for magnetostimulation paired with light the users of Viofor JPS Classic or Viofor JPS
energy treatment. Viofor JPS controller Standard to apply magnetic and light therapy. It
generates control signal for all applicators adjusts the output controller signal to the input
manufactured by the company. ZAMS 01/2 signal of the magnetic-light applicator and syn-
power supply adjusts the output signal in the chronizes with light impulses.
controller to the input signal in magnetic-light The power supply unit is equipped with 6-pin
applicator. Magnetic-light applicator gener- socket to connect the applicator and the cable
ates heterogeneous variable magnetic field with a large jack, connected to a socket in
of a specific signal structure, defined as JPS Viofor JPS System controller. Large jack has
system, and emits light impulses of selected a blue label wit the power supply unit symbol. A
parameters synchronized with magnetic field diode (ON) in the middle of the front panel lights
impulses. in blue when the power supply unit is plugged in.

6.2.1 ZAMS 01/2 front panel


1 – signal diode
2 – cable with
a large jack
3 – 6 –pin socket

Drawing 6. ZAMS 01/2 front panel

6.2.2 ZAMS 01/2 back panel

1 – power on
2 – fuse base with
a 100 mA
delay switch
3 – power cable
4 – power voltage
switch
5 – power unit serial
number

Drawing 7. ZAMS 01/2 back panel

22
Usage of Viofor JPS System
6.3 Magnetic-light applicators and controller) are set at the proper volt-
age. If not – switch them accordingly and
The applicator casing contains a coil for exchange the fuses according to the de-
magnetic field therapy and LED diodes set scription on the back panels of the power
for light therapy. On the top of the casing supply unit and controller under the fuse
there is a triangular warning label, signal socket (electric cable should be discon-
diode in color depending on the applicator nected).
type (see Table below) and company logo. 2. Connect ZAMS 01/2 power supply unit
On the bottom there is a transparent glass, to the Viofor JPS controller socket with
through which the light is emitted. the large jack. Connect the plug of mag-
Each applicator is equipped netic-light applicator to the 6-pin power
with a cable with a 6-pin jack. supply unit socket and turn it right to
The cable is labeled with different color block it.
labels indicating: 3. Turn the power on in controller and power
- applicator serial number supply unit.
(from the jack’s side), 4. Place the applicator over the treated
- applicator type identifier area. Set the controller. Press „START”.
(from the casing side – R, IR, RIR) NOTICE:Due to direct contact of the ap-
Applicator is attached to the patient with plicator with the patient’s skin it is recom-
a regulated belt with a buckle. mended to clean the applicator with a disin-
fectant (Notice, do not allow the applicator to
6.4 Magnetic & Light Therapy get wet!).
set usage In case of therapy of open wounds and
bacterial or mycotic skin deformations, do
1. Make sure that the power voltage switches not touch the applicator to the skin. Keep it
(on the back panel of power supply unit 0,5-1 cm above the skin.

Drawing 8: LED diode layout in applicators: a) red light R, b) infrared light IR, c) mixed light RIR
(infrared applicator is equipped with one red LED diode that signals emission of invisible infrared radiation IR
during treatment)

a) b) c)

23
Usage of Viofor JPS System

7 SAFE USAGE OF VIOFOR JPS


1. The apparatus transported or stored in 10.For hygienic purposes, it is recommended
temperatures below the temperature in the to cover the applicators with disposable
room, where it will be installed, it can be foil or paper covers.
plugged in after at least 4 hours from deliv- 11. When the device is working:
ery. The equipment cannot be operated in - The personnel must keep at least
humid places 1 meter away from an active applicator
2. When installing Viofor JPS, check if the Caution! When using magnetic-light applica-
voltage is set correctly. tors, do not direct the light in the eye area!
NOTICE! Devices distributed in Europe - The device can be plugged in and out
are set by default to 203 V. When switch- only when the power switch is off.
ing the voltage to 115 V (when using - Plug the applicator in and out only
in other countries), fuse should be ex- when the power is off.
changed as indicated on a label. 12.After work:
3. Install and connect the apparatus accord- - Do not roll the mat,
ing to the instruction contained in the pres- but fold it twice
ent manual. (see drawing)
4. Do not plug the device if the cable or plug - Do not bend the pad
is damaged. Have the device serviced - Before cleaning the device, read appro-
5. Protect the device out of reach of small priate section of the instruction
children and mentally disabled persons. - Do not open the zipper around the mat.
6. Keep away from pets. 13.It is recommended to have the device
7. If the device is flooded or gets wet, do not checked every 1 to 2 years (service is free of
plug it in – have it serviced immediately. charge, transportation costs are covered
8. Before plugging the device in, put away by the client).
magnetic cards and electronic watches, 14.The device transported to the Med & Life
since they might be damaged. It applies service should be well protected against
both to the patients and the operators. damages (e.g. in original package)
9. Turn off all other electronic devices 15. Use only AAA alkaline batteries
(e.g. hearing aids, pain stimulators, etc.) for the remote control.
8 CLEANING, MAINTENANCE, REPAIRS
8.1 General information - In case of broken fuse in controller
- Clean the controller with the plastic clean- or ZAMS, exchange it
er, do not use water (when the controller and power supply
- Applicators can be cleaned with a cloth unit are plugged off!). Make sure
with a mild detergent and, if necessary, to use the proper fuse,
with disinfectant i.e. according to the label at the fuse socket.
- Wash ZAMS 01/2 only with detergent for - Do not short-circuit 6-pin sockets
plastics in ZAMS 01/2!
- Device does not require any maintenance - In case of damaged power cable
- Repairs can be done only by authorized or plug, do not use the set
service of Med & Life, as described in the and have it serviced.
warranty certificate. Any unauthorized re- - Magnetic-light applicators are not resistant
pairs and/or breaking of guarantee seals to high temperatures. Do not use auto-
will result in warranty termination! clave for disinfection!
- When changing the fuse, unplug the con- - If disposal is necessary, have it done
troller and power supply unit by a specialized company

24
Usage of Viofor JPS System
8.2 Service pack 8.2.2 Exchange of applicator plug
and adapter
8.2.1 Fuse exchange 1) Cut the damaged plug from the applicator
1) Unplug the device from the wall socket. cable.
2) Remove the fuse handle and the fuse 2) Unscrew the new plug casing. Install
with a proper screwdriver. The fuse handle the new plug on the applicator cable.
is removed by one-quarter turn 3) Remove external insulation from applicator
counter-clockwise. cable – approx. 1.5 cm.
3) Remove the fuse from the handle. 4) Remove approx. 0.5 cm of insulation
Insert a new fuse in accordance from the brown (or black) wire.
with the instruction on the back panel. 5) Remove approx. 1.5 cm of insulation
4) Place the handle in the socket from the blue (or white) wire.
with a proper screwdriver 6) Cover the uncovered wire ends
and make one-quarter with soldering agent.
turn clockwise. 7) Solder the brown (or black) wire
5) Plug the device in the wall socket to the middle plug electrode.
and turn on the power. 8) Solder the blue (or white) wire
If the device is not operating, to the side plug electrode.
it should be serviced. 9) Screw the plug casing together.
NOTICE: do not use fuses 10) Test the applicator with MFI indicator.
with parameters other If the applicator is not working, it should
than indicated. be serviced.

9 WARRANTY TERMS
Warranty cover free repair or exchange • pollution
of the device or its elements in case of manu- • flooding
facture and material defects. Exchange will • damages due to
apply to the damaged element only. - insufficient knowledge of the user
- incorrect installation or usage of the
The warranty does not cover: device
• mechanical damages - exploiting of the device in conditions
- due to device being dropped or hit others then specified by the manufacturer
- abrasions - incorrect parameters of power supply
- scratches - incorrect setting of the voltage switch
- dents - incorrect fuse
- tears • broken warranty seals
- discoloration due to usage of improper • damages during the transport of the
detergents and cleaning agents device to Med & Life authorized service
- detachment of magnetic-light
applicator glass due to improper In case of repair or structural changes
cleaning method by unauthorized persons, Med & Life shall
- torn or broken power supply cable, ap- have the right to revoke warranty obligations.
plicator cable or plug The manufacturer shall not be held responsi-
- improper disinfection method ble for any bodily damages or injuries due to in-
(e.g. in autoclave) correct usage of the device, its usage for purpo-
• regular wear and tear of applicators ses other than those indicated in the instruction
• applicator deformation manual, or any changes introduced to the devi-
• effects of atmospheric discharge ce without prior approval of the manufacturer.

25
Technical data and accessories

PART C

26
Technical data and accessories

Technical data and accessories


10 TECHNICAL DATA

10.1 Controllers
Controller Microprocessor

Trade name Viofor JPS System Classic/Standard

Voltage rating and frequency 220/230 V – 110/115 V (±10%, 50-60 Hz)

Power input in ready mode <8 VA

Maximum power input <15 VA

Output voltage in controller socket Constant voltage, max 20 V

Fuse in primary circuit T 0,25 A / with delay (for 110/115 V / T 0,5A)

Working mode Continuous running

Storage and transport temperature From -15 to +70oC

Relative working humidity Max 80%

Admissible storage and transportation humidity Max 90%

Output controller socket Jack

Safety class II

Protection type IP40 (water unprotected)

Protection B (from electric shock)

Magnetic induction intensity regulation


in Classic controller Gradual 0,5; 1; 2.................; 12

Magnetic induction intensity regulation


in Standard controller Gradual 0,5; 1; 2.................; 10

Application time 81), 10, 12 min

Automatic field direction change Depends on program and application

Continuation on the next page

1)
for Classic controller

27
Technical data and accessories
Continuation from page 27

Amplitude of current running through applicator


coils for intensity 0,5 0,05 ÷ 0,1 A

Amplitude of current running through applicator


coils for intensity 12 (Classic) 1,5 ÷ 2,0 A

Amplitude of current running through applicator


coils for intensity 10 (Standard) 1,4 ÷ 1,7 A

10.1.1 Compatibility 10.1.2 Casing


All Viofor JPS controllers A special ABS casing has been used,
are compatible with all Viofor JPS of maximal dimensions:
applicators and can work with width 330mm
magnetic-light applicators through depth 280mm
ZAMS 01/2 power supply unit. height 280mm
Viofor JPS Classic also works
with infrared remote 10.1.3 Mass
control unit. Mass of the device without applicators - < 4kg

10.2 Power supply unit ZAMS 01/2

Power 230V/115V (±10% -10%) – T100/200 mA

Input signal from Viofor JPS controller

Impulse output signal synchronized


with controller signal: 21 V ± 0,3 V

Max. power input 12 VA ± 2 VA

Working temperature 10°C – 40°C

Relative working humidity max 80 %

Storage and transport temperature -15°C - +70°C

Admissible humidity of storage and transport max 90%

Safety class II

Type of protection IP40

Protection B (from electric shock)

Working mode Continuous running

28
Technical data and accessories

10.3 Magnetic-light applicators


Power and input signal from ZAMS 01/2 power supply unit

Output signal Light impulses synchronized with magnetic field impulses:


- applicator light parameters: see table p. 32-34
- magnetic field parameters: peak value 100 x I[µT],
mean value: 8,9 x I [µT] where:
I – intensity set on the controller

Max power input 0,63 A

Working temperature 10°C – 40°C

Relative working humidity max 80 %

Storage and transport


temperature -15°C - +70°C

Admissible humidity
of storage and transport max 90%

Safety class for light energy 3A

Protection B (from electric shock)

Working mode Continuous running

11 APPLICATORS
11.1 Applicators surface, away from any household appliances
for magnetostimulation and TV or radio sets. Plug the applicator to
any controller socket*. Lay down on the mat,
11.1.1 Large applicator - mat with your head near the output cable. Set the
Used for the whole body desired application parameters and press
treatment. Contains 3 pairs START/STOP button.
of coils placed along
the mat and generating 11.1.2 Large applicator – short mat
heterogeneous variable Used for the whole body
magnetic field of different intensity: coil located treatment. Generates het-
in the feet area are the strongest, and in the erogeneous variable mag-
shoulder area – the weakest. Head should netic field. Used in cases,
always be located near the output cable. Mat when it is recommended to
is a portable applicator, in two colors, can be use two mats on one patient or if a regular
folded in 3. Size 180 x 65, folded 60 x 65 cm. mat is not handy (e.g. after surgery). Two
Installation: place the applicator on a hard colors. Size 157 x 65 cm, side cable output.

* In case of using default application parameters, mat, pad and point applicators should be connected in accordance
with symbols located over the controller sockets.

29
Technical data and accessories
Installation: plug the applicator into any con- is required. Can be used together
troller socket*. Cover the patient with the mat. with two-color applicator facing
Set the desired application parameters and white side up.
press START/STOP button. Installation: Plug the applicator into any
controller socket*. Point the plastic
11.1.3 Large applicator - frame bed applicator end to the place exposed
Metal frame bed, to treatment. Set the desired application
generates heterogeneous parameters and press START/STOP button.
variable magnetic field.
Used instead of large 11.1.6 Elliptic applicator (light and green)
applicator – mat For local application,
in doctor’s office. fastened with the belt on limbs
Equipped with applicator roll guides or with the tape in other places
for the large ring applicator. of application. Generates
Size: length 200 x width 60 x height 55 cm. heterogeneous variable
Installation: place the applicator away magnetic field, whose area
from household appliances and TV or radio of activity is within approx. 6 cm
sets. Plug the applicator into any controller circle. Used mostly for treatment
socket*. Set the desired application param- of small and medium joints,
eters and press START/STOP button. in dentistry and laryngology.
White and greed applicators, when used
11.1.4 Small applicator - pad separately, work identically. If two elliptic
Used for local treatment, applicators are used on the same spot,
can be placed over or un- e.g. wrist in the same patient, only different
der selected body part. color applicators can be connected.
Contains one pair of coils Installation: plug the applicator
generating heterogeneous into any controller socket*. Place
variable magnetic field. The strongest effect the applicator on the treated area,
of magnetic field is located near each coil. fasten with the strap, with an adhesive tape
Do not bend the pad. Two color. or hold it in one position. Set the desired
Size 50 x 30 cm. application parameters and press
Installation: Plug the applicator into any START/STOP button.
controller socket*. Place the applicator
on or under the treated body part. Set the 11.1.7 Ring applicator – small,
desired application parameters and press 30 cm diameter
START/STOP button. For local application on limbs.
Generates homogenous
11.1.5 Point applicator (S and Z) variable magnetic field of line
For local application. of force parallel to the coil
Generates heterogeneous axis, therefore parallel
variable magnetic field to patient’s body, which increases
of point-like characteristics. application effectiveness. Equipped
Used in dentistry, on selected with two cable sockets, marked H and L.
acupuncture points and in cases When applicator is connected to socket H,
when a narrow beam the induction of magnetic field generated
of variable magnetic field by the applicator is three times stronger
of significant induction than in socket L.

* In case of using default application parameters, mat, pad and point applicators should be connected in accordance
with symbols located over the controller sockets.

30
Technical data and accessories
Installation: select therapy type: H or L - Infrared applicator IR - 47 infrared LED
and plug in the cable in the selected diodes, one LED red diode
applicator socket, in accordance - Combined applicator R+IR - 24 infrared
with symbols on the cable. Plug the other and 24 red LED diodes
end to any controller socket. Place the limb Light area of magnetic-light applicators
inside the applicator (applicator should is 19,6 cm2. LED diode terminals touch
be place exactly above the treated area) polycarbonate plate 0,75 mm thick.
on a flat base. Set the desired parameters Plate is place directly on patient’s body.
and press START/STOP button. Applicators are fastened with the belt
on limbs and with tape on other parts
11.1.8 Ring applicator – large, of the body. Synergy of both types
65 cm diameter of electromagnetic radiation
Used together with the large is expected.
applicator – frame bed.
Equipped with rolls that can be 11.2.1 Red magnetic-light applicator (R)
moved in the frame bed (wavelength 630 ±5 nm – shal-
guides. In the area of applica- low penetration) is used mainly
tion, generates homogenous for skin disease treatment, such
variable magnetic field, as unhealing wounds and
of line of force parallel ulcerations (including bedsores),
to the coil axis, therefore common acne and zoster. It is recommended
parallel to patient’s body, in dermatology and cosmetic surgery.
which increases application effectiveness.
Used as local applicator for selected 11.2.2 Infrared magnetic-light applicator (IR)
body parts. (wavelength 855 ±5 nm
Installation: Plug properly – deep penetration)
(see assembly instruction) ring applicator in motor system diseases
installed on the guide rails of the frame-bed (arthrosis; pain syndromes
applicator to any controller socket. in discopathy; syndromes
Lie down on the applicator and set the ring due to strain of muscles and periarticular
applicator. Set the desired application soft tissue; inflammation of tendons,
parameters and press START/STOP button. fascia, tendon sheaths and joint bursa;
If frame-bed applicator and large ring periarticular inflammation; bone fractures
applicators have to be used and delayed union of fractured bones)
simultaneously, they should be connected as well as in neurological diseases
through an adapter. (neuralgia of peripheral nerves – intercos-
tal neuralgia, trigeminal neuralgia,
post-zoster neuralgia, sciatica;
11.2 Magnetic-light applicators diabetic neuropathy).

Three magnetic-light applicators for lo- 11.2.3 Mixed magnetic-light


cal use (red (R), infrared (IR) and combined applicator (RIR)
red/infrared (R+IR)). Generate heteroge- (wavelength 630/855 ±5 nm
neous variable magnetic field of physical – medium penetration)
parameters of magnetostimulation in JPS in chronic skin inflammations;
System, whose area of activity is within ap- keloid, psoriasis, psoriatic
prox. 6 cm circle. The applicator coil con- joint inflammation, ankylosing
tains: spondylitis, rheumatoid inflammation
- Red applicator R - 48 red LED diodes of the joints, post-thrombotic syndrome.

31
Technical data and accessories
11.3 Maximal induction of magnetic field generated
by Viofor JPS System coils

B[µT] B[µT]
Type of applicator Mean value Impulse peak value Comments
(at setting M1 P2)

Large applicator
Mat/frame bed 2.5 x I 28 x I Heterogeneous field*

Small applicator - pad 5xI 56 x I Heterogeneous field

Point applicator S 10 x I 112 x I Heterogeneous field

Point applicator Z 100 x I 1120 x I Heterogeneous field

Elliptic applicator 8.9 x I 100 x I Heterogeneous field


(light and green)

Small ring applicator 0.9 x I 10 x I for L Homogenous field**


Ø 30 cm 2.7 x I 30 x I for H

Large ring applicator 0.9 x I 10 x I Homogenous field


Ø 65 cm

Magnetic-light 8.9 x I 100 x I Heterogeneous field


applicator

Accuracy ±10%.
I – is the intensity set on Viofor JPS System controller
* Heterogeneous field measured on the applicator surface in the area of maximal device readings
** Homogenous field** measured in the geometric center of the applicator. Homogeneity with accuracy ± 5 % obtained
in approx. 70 % applicator internal volume.

11.4 Radiation value parameters of magnetic-light applicators


connected to Viofor JPS controller
Light area 19,6 cm2.

In red applicators R: 48 red LED diodes, ø5 mm, 2Θ1/2 = 15°

In infrared applicators IR: 47 infrared LED diodes, ø5 mm, 2Θ1/2 = 30°;


1 red diode, ø5 mm, 2Θ1/2 = 15°

In mixed applicators (RIR): 24 red LED diodes, ø5 mm, 2Θ1/2 = 15° and
24 infrared LED diodes, ø5 mm, 2Θ1/2 = 30°

Red and infrared light parameters constitute 50 % of below listed parameter values

32
Technical data and accessories
RED APPLICATOR R

Value Unit 0,5 1 2 3 4 5 6 ÷ 12

Impulse power mW 210 210 210 210 210 210 210

Mean power for P11) mW 10,9 25,2 30,2 31,5 32,3 32,8 33,6

Mean power for P2 mW 9,7 30,7 35,4 36,7 37,5 38,0 39,0

Mean power for P31) mW 9,9 31,8 36,7 38,0 38,8 39,3 40,3

Impulse power density mW/cm2 10,7 10,7 10,7 10,7 10,7 10,7 10,7

K1 impulse energy mJ 0,27 0,36 0,76 0,79 0,81 0,82 0,84

K2 impulse energy mJ 0,10 0,71 0,78 0,79 0,81 0,82 0,84

K1 impulse energy density µJ/ cm2 13,9 32,1 38,6 40,2 41,2 41,8 42,9

K2 impulse energy density µJ/ cm2 5,4 36,4 40,0 40,7 41,2 41,8 42,9

Energy provided during


treatment for P11) J 5,2 12,1 14,5 15,1 15,5 15,7 16,1

Energy provided during


treatment for P2 J 5,8 18,4 21,3 22,0 22,5 22,8 23,4

Energy provided during


treatment for P31) J 5,9 19,1 22,0 22,8 23,3 23,6 24,2

Energy density for P11) J/cm2 0,27 0,62 0,74 0,77 0,79 0,80 0,82

Energy density for P2 J/cm2 0,30 0,94 1,09 1,12 1,14 1,17 1,19

Energy density for P31) J/cm2 0,30 0,98 1,12 1,17 1,19 1,20 1,23

1) for Viofor JPS Classic controller

The table shows light parameters of the red light applicator depending on the intensity
set on the Viofor JPS controller.

INFRARED APPLICATOR IR

Value Unit 0,5 1 2 3 4 5 6 ÷ 12

Impulse power mW 1440,00 1440,00 1440,00 1440,00 1440,00 1440,00 1440,00

Mean power for P11) mW 74.88 172.80 207.36 216.00 221.76 224.64 230.40
Continuation on the next page

33
Technical data and accessories
Continuation from page 33

Mean power for P2 mW 66.59 210.59 243.07 251.83 257.24 260.58 267.26

Mean power for P31) mW 67.74 218.42 251.60 260.58 266.11 269.57 276.48

Impulse power
density mW/cm2 73.47 73.47 73.47 73.47 73.47 73.47 73.47

K1 impulse energy mJ 1.87 4.32 5.18 5.40 5.54 5.62 5.76

K2 impulse energy mJ 0.72 4.90 5.33 5.47 5.54 5.62 5.76

K1 impulse energy
density µJ/ cm2 95.51 220.41 264.49 275.51 282.86 286.53 293.88

K2 impulse energy
density µJ/ cm2 36.73 249.80 271.84 279.18 282.86 286.53 293.88

Energy provided
during treatment
for P11) J 35.94 82.94 99.53 103.68 106.44 107.83 110.59

Energy provided
during treatment
for P2 J 39.95 126.35 145.84 151.10 154.34 156.35 160.36

Energy provided
during treatment
for P31) J 40.64 131.05 150.96 156.35 159.67 161.74 165.89

Energy density for P11) J/cm2 1.83 4.23 5.08 5.29 5.43 5.50 5.64

Energy density for P2 J/cm2 2.04 6.45 7.44 7.71 7.87 7.98 8.18

Energy density for P31) J/cm2 2.07 6.69 7.70 7.98 8.15 8.25 8.46
1) for Viofor JPS Classic controller

The table shows light parameters of infrared light applicator depending on the intensity
set on the Viofor JPS controller.

Wavelengths of emitted light (λP) are as follows:


• 630±5 nm for red light,
• 855±5 nm for infrared light.

CAUTION!

STRONG RED AND/OR INFRARED RADIATION!


PROTECT YOUR EYES!
DO NOT LOOK AT THE LIGHT!

34
Table of contents

TABLE OF CONTENTS
1 CHARACTERISTICS OF VIOFOR JPS ................................................................................ 3
1.1 Usage ............................................................................................................................ 3
2 TYPES OF THERAPY .......................................................................................................... 3
2.1 Magnetostimulation ....................................................................................................... 3
2.1.1 Characteristics of magnetostimulation .................................................................. 3
2.1.2 Indications for magnetostimulation ........................................................................ 5
2.1.3 Contraindications to magnetostimulation .............................................................. 6
2.1.4 Side effects, which may occur due to magnetostimulation.................................... 7
2.2 Magnetostimulation paired with light energy ................................................................. 7
2.2.1 Characteristics of magnetostimulation paired with light energy ............................ 7
2.2.2 Indications for magnetostimulation with light energy ............................................. 8
2.2.3 Contraindications to magnetostimulation and light energy treatment ................... 9
2.2.4 Side effects which may occur due to magnetostimulation
with light energy treatment .................................................................................... 9
3 THERAPY SELECTION ........................................................................................................ 9
3.1 Magnetostimulation ...................................................................................................... 9
3.1.1 Application parameters components ................................................................... 10
3.1.1.1 Programs (P) ............................................................................................... 10
3.1.1.2 Application modes (M)..................................................................................11
3.1.1.3 Magnetic field intensity ( I ) ..........................................................................11
3.1.1.4 Selection of application parameters ............................................................ 12
3.1.2 Using more than one applicator .......................................................................... 12
3.2 Magnetostimulation with light energy ......................................................................... 13
3.2.1 Application parameters components ................................................................... 13
3.2.1.1 Programs ( P ) ............................................................................................. 13
3.2.1.2 Application modes ( M )............................................................................... 13
3.2.1.3 Magnetic field intensity ( I ) ......................................................................... 14
3.2.1.4 Selection of application parameters ............................................................ 14
4 VIOFOR JPS CLASSIC ....................................................................................................... 16
4.1 General information .................................................................................................... 16
4.2 Viofor JPS Classic back panel ................................................................................... 16
4.3 Viofor JPS Classic front panel .................................................................................... 16
4.4 Remote control unit .................................................................................................... 17
4.4.1 Remote control details ........................................................................................ 17
4.4.2 Battery installation ............................................................................................... 18
4.5 Apparatus installation ................................................................................................. 18
4.6 Apparatus operational use.......................................................................................... 18
4.7 Programming of application parameters .................................................................... 19
4.8 Retrieving preprogrammed application parameters ................................................... 19
4.9 PLACEBO mode ........................................................................................................ 19
4.10 Counter ..................................................................................................................... 20
5 VIOFOR JPS STANDARD................................................................................................... 20
5.1 General information................................................................................................... 20
5.2 Viofor JPS Standard back panel ............................................................................... 20

35
Table of contents
5.3 Viofor JPS Standard front panel ................................................................................ 20
5.4 Apparatus installation ................................................................................................ 21
5.5 Apparatus operational use ........................................................................................ 21
6 MAGNETIC & LIGHT THERAPY ......................................................................................... 22
6.1 General information .................................................................................................... 22
6.2 ZAMS 01/2 power supply unit .................................................................................... 22
6.2.1 ZAMS 01/2 front panel ........................................................................................ 22
6.2.2 ZAMS 01/2 back panel ........................................................................................ 22
6.3 Magnetic-light applicators........................................................................................... 23
6.4 Magnetic & Light Therapy set usage .......................................................................... 23
7 SAFE USAGE OF VIOFOR JPS ........................................................................................ 24
8 CLEANING, MAINTENANCE, REPAIRS ............................................................................ 24
8.1 General information ....................................................................................................... 24
8.2 Service pack .................................................................................................................. 25
8.2.1 Fuse exchange ...................................................................................................... 25
8.2.2 Exchange of applicator plug and adapter .............................................................. 25
9 WARRANTY TERMS .......................................................................................................... 25
10 TECHNICAL DATA ............................................................................................................ 27
10.1 Controllers ................................................................................................................ 27
10.1.1 Compatibility ...................................................................................................... 28
10.1.2 Casing ............................................................................................................... 28
10.1.3 Mass .................................................................................................................. 28
10.2 Power supply unit ZAMS 01/2 .................................................................................. 28
10.3 Magnetic-light applicators......................................................................................... 29
11 APPLICATORS .................................................................................................................. 29
11.1 Applicators for magnetostimulation .............................................................................. 29
11.1.1 Large applicator - mat........................................................................................ 29
11.1.2 Large applicator – short mat .............................................................................. 29
11.1.3 Large applicator - frame bed ............................................................................. 30
11.1.4 Small applicator - pad ........................................................................................ 30
11.1.5 Point applicator (S and Z ) ................................................................................. 30
11.1.6 Elliptic applicator (white and green ).................................................................. 30
11.1.7 Ring applicator – small, 30 cm diameter ........................................................... 30
11.1.8 Ring applicator – large, 65 cm diameter ............................................................ 31
11.2 Magnetic-light applicators ......................................................................................... 31
11.2.1 Red magnetic-light applicator (R) ...................................................................... 31
11.2.2 Infrared magnetic-light applicator (IR) ............................................................... 31
11.2.3 Combined magnetic-light applicator (RIR)......................................................... 31
11.3 Maximal induction of magnetic field generated by Viofor JPS System coils ............. 32
11.4 Radiation value parameters of magnetic-light applicators
connected to Viofor JPS controller ............................................................................ 32

AAVI-EN-I-04/2005

36
Technical data
Awards
and accessories
and honors
Viofor JPS has won many prestigious national and international awards, such as:
- award of “Polish Product of the Future 2000”
- gold medal at the Health Spa and Rehabilitation Fair “Zdroj 2001”
- awards at 50th World Exhibition of Development, Research and New Technologies
“Brussels Eureka” in Brussels:
• gold medal and distinction
• special award of the Italian Delegation
• special award of Croatian Society of Inventors INOVA
- gold medal at „Concours Lepine” Fairs 2002 Paris
- gold medal at IENA 2002 Fairs in Nuremberg
- gold medal at the International Fair “Invest Technology 2002” in Warsaw
- metal at Intarg Fair 2002 in Katowice
and many others.

37
Awards and honors
Med & Life, the manufacturer of Viofor JPS system has a certified Quality System according to
ISO norms: ISO 9001:2000 and ISO 13485:2003.

Med & Life is also a recipient of many prestigious awards:


- Leader of Entrepreneurship 2000
- Cup and Diploma of the Minister of Education
and President of KBN for innovative activity
- Leader of Innovation 2002
- Award at “Huzar Europy 2003” competition.
For more information on the manufacturer see: www.medandlife.com

38
Additional informations

EC CONFORMITY DECLARATION
Med & Life herewith declares, that:
Systems for therapy and rehabilitation
• Viofor JPS System Standard with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)
• Viofor JPS System Classic with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)
• Viofor JPS System Clinic with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)
• Viofor JPS System Delux with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)
conform the essential requirements and the guidelines of the Council Directive
93/42/EEC - Medical Devices and with harmonized standards:
– EN IEC 60601-1 „Medical electrical equipment” with the supplements
and linked standards.
– PN-EN 60825:2000/A2:2002 (U) Safety of laser products. Equipment classification,
requirements and user`s guide.
– is a subject of the procedure mentioned in the Annex II of the Directive 93/42/EEC
Med & Life has implemented the Quality Management System according
to ISO 9001:2000/ISO 13485:2003 under the surveillance of the Notified Body
nr 0120, SGS United Kingdom Ltd Systems & Services Certification.

39
Notes

40