DigitalMultimeterCalibration
Purpose
Thepurposeofthisprocedureistoprovideinstructionforthecalibrationofdigital
multimeters(DMM).

DownloadCalibrationControl
Email:*
Name:*

Scope

Company:*

Allmultimeters(multirangeinstrumentsusedtomeasurevoltage,current,and

DayPhone:

resistance)aretobeverifiedincalibrationnolessthanonceperyear.

SignIn&DownloadSoftware

Procedure
PreliminaryInstructionsandNotes
1. Readthisentireprocedurebeforebeginningthecalibration.
2. CalibrationshallbeperformedinanenvironmentthatconformstoManufacturer
Specifications.

http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

1/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

3. ThedigitalmultimeterwillhereafterbereferredtoastheInstrumentUnderTest
(IUT).
4. VerifythattheIUTisclean.
5. VisuallyexaminetheIUTforanyconditionthatcouldcauseerrorsinthecalibration.
6. Ifanyoftherequirementscannotbemet,refertotheapplicablemanufacturer
manual.
7. Ifamalfunctionoccursoradefectisobservedwhilecalibrationisinprogress,the
calibrationshallbediscontinuedandnecessarycorrectiveactiontakenifcorrective
actionaffectsameasurementfunctionpreviouslycalibrated,thefunctionshallbe
recalibratedbeforetheremainderoftheprocedureimplemented.

ReferenceMaterial
ApplicableManufacturersManualorBrochures.

Specifications
ThespecificationsoftheIUTaredeterminedbytheapplicablemanufacturer's
documentation.Ifthemanufacturer'sdocumentationisnotavailable,thenthe
specificationsidentifiedinthisprocedureareused.

EquipmentRequired
TheStandardslistedbelowshouldbeselectedonthebasisoftheirhigheraccuracylevel
whencomparedtotheunitundertest.EquivalentStandardsmustbeequaltoorbetter

thantheMinimumUseSpecification.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

2/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

MinimumUseSpecificationsforStandardslistedare1/4theaccuracyrequiredbythe
IUT.
RotekModel2500Calibratororequivalent
DiodeTestFixture
TestLeads
ContinuityTestFixture
CapacitanceTestFixture
Thermometer
Hydrometer

SetUp
Turnallpoweron(calibratorandtheunittobecalibrated).Allowtheinstrumentsto
stabilizeforapproximately5minutes.Conductthetestsinanambienttemperatureof25
5Candarelativehumidityoflessthan80%.
Note:IftheIUThasanautomaticrangingfeature,thenthefunctionbeingchecked(i.e.
voltage,current,orresistance)mustbeverifiedinthevariablerangemodeaswell.

BatteryReplacement
ReplacethebatteryintheIUTandmakesurethebatterycontactsarenotdirty.

DetailedProcedure
DISPLAYANDSWITCHTEST
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

3/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

TurntheIUTonandverifythatallLCDsegmentsareworkingandnotdim.Verifythatthe
selectorswitch(s)is/areworkingproperly.MakesuretheLowBatteryindicatorisnot
showing.
RESISTANCECHECK

ConnectthecalibratorbetweentheV/ohmandcommoninputterminalsoftheIUT.Use
thefollowingcharttoverifytheresistancevalues.
Step

Range

Input

Display

200ohm

Short

00.0to00.5

2kohm

Short

0.000to0.001

2kohm

1kohm

.998to1.002

20kohm

10kohm

9,98to10.02

200kohm

100kohm

99.8to100.2

2000kohm

1Mohm

998to1002

2000Mohm

Open

0.10to00.0

CONTINUITYTEST

1. Selectthecontinuitycheckfunctiononthecalibrator.
2. ConnectthetestleadstotheV/ohmandcommonterminalsoftheIUT.
3. Momentarilyshortthetestleadstogetherandobservethatthetonesounds.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

4/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

4. Connectthetestleadstothecontinuitytestfixturebetweenthe100ohmtest
points.Notoneshouldbeheardindicatingnonterminatingconnection.
5. Connectthetestleadstothecontinuitytestfixturebetweenthe50ohmsetup.The
testtoneshouldindicatecontinuitybetweenthetestpoints.
DCVOLTAGETEST

1. UsethecalibratortosupplythecorrectDCvoltagelistedbelow.
2. Setthecalibratorforazerovoltinput.
3. ConnectthecalibratoroutputtotheV/ohmandcommoninputterminalsoftheIUT.
Note:Connecttheground/common/lowsideoftheRotek2500tothecommonon
theIUT.

4. Withreferencetothetablebelow,selecttheIUTvoltagerangeandsetthe
calibratoroutputtothecorrespondingIUTinputvoltage.Testperthefollowing
chartandverifythatthedisplayfallswithinthelimitsgiven.
Step

Range

Input

Display

200mV

190mV

189.7to190.3

200mV 190mV 189.7to190.3

2V

0.0V

0.001to0.001

2V

1.9V

1.897to1.903

20V

19V

18.97to19.03

http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

5/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

200V

100V

99.84to100.16

1000V

400V

399.3to400.6

ACVOLTAGETEST

UsethecalibratortosupplythecorrectACvoltageinthefollowingtest:
1. Setthecalibratorto60Hz.
2. ConnectthecalibratoroutputtotheV/ohmandcommoninputterminalsoftheIUT.
Connecttheground/common/lowsideofthecalibratortocommonontheIUT.
3. Withreferencetothetablebelow,selecttheIUTvoltagerangegiveninstep1and
setthecalibratoroutputtothecorrespondingIUTinputvoltage.Verifythatthe
displayreadingiswithinthelimitsshown.
Step

Range

Input

200mV 100mV

Display
99.0to101.0

2V

1V

0.990to1.010

20V

10V

9.90

200V

100V

99.0to100.1

750V

400V

294to406

DCCURRENTTEST

UsethecalibratortosupplythecorrectDCcurrentlistedinthetablebelow:
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

6/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

1. SettheoutputofthecalibratortozeromA.
2. Connecttheoutputofthecalibratortothelowestcurrentterminal(typically200
mA)andcommoninputterminalontheIUT.
Note:Donotexceedthemaximumcurrentratingofthemeter.Onlytestfor
valuesthemeteriscapableofreading.
3. Withreferencetothetablebelow,selecttheIUTcurrentrangeandsetthe
calibrationoutputtoprovidethecorrespondingIUTinputcurrent.Verifythatthe
displayreadingiswithinthelimitsshown.
Step

Range

Input

Display

200?A

190?A

189.7to192.3

2mA

1.9mA

1.897to1.923

20mA

19mA

18.97to19.23

200mA 190mA 189.7to192.3

4. DisconnectthecalibratorandchangethecurrentterminalontheIUTtothehighest
setting(typically10A)leavingthecommoninputterminalconnected.
5. Usethefollowingtabletoverifythatthemeterisfunctioningwithinthespecified
parameters.
Step Range Input

Display

10A

1A

0.98to1.02

10A

5A

4.90to5.10

http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

7/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

10A

9A

8.82to9.18

ACCURRENTTEST

UsethecalibratortosupplythecorrectACcurrentlistedinthetablebelow:
1. Usethefollowingtabletoverifythatthemeterisfunctioningwithinthespecified
parameters.
2. SettheoutputofthecalibratortozeromA.
3. Connecttheoutputofthecalibratortothelowcurrentinputterminal(typically200
mA)andthecommonterminaloftheunitundertest.
Note:Donotexceedthemaximumcurrentratingofthemeter.Onlytestfor
valuesthemeteriscapableofreading.
4. Withreferencetothetablebelow,selecttheIUTcurrentrangeandsetthe
calibratoroutputtoprovidethecorrespondingIUTinputcurrent.Verifythatthe
displayreadingiswithinthelimitsshown.
Step

Range

Input

Display

200?A

190?A

18.27to19.28

2mA

1.9mA

1.872to1.928

20mA

19mA

18.72to19.28

200mA 190mA 187.2to192.8

5. DisconnectthecalibratorandchangethecurrentterminalontheIUTtothehighest
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

8/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

setting(typically10A)leavingthecommoninputterminalconnected.
6. Usethefollowingtabletoverifythatthemeterisfunctioningwithinthespecified
parameters.
Step Range Input

Display

10A

1A

0.97to1.03

10A

5A

4.87to5.13

10A

9A

8.77to9.23

LOGICTEST

Usethisproceduretotestweatherthelogicfunctionisindicatingthecorrectstate.
1. SettheIUTtothelogictestmodeandinsertthetestleadsintotheV/ohmand
commonterminals.
2. Shorttheleadstogetherandverifythatthemeterindicatesalowlogicstate.
3. Setthecalibratorto0.400VDCandconnectthemetertotheoutputterminals.
Verifythatthemeterindicatesalowlogicstate.
4. Setthecalibratorto3.500VDCandverifythattheIUTindicatesahighlogicstate.
FREQUENCYTEST

1. ConnectanoscilloscopetoafunctiongeneratortomeasureACsinewaves.
2. Setthefunctiongeneratortooutput10VACandusetheoscilloscopetosetthe
frequenciesfromthefollowingtable.
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

9/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

3. ConnecttheIUTleadstotheV/ohmandcommonterminalsontheIUTandtothe
groundandoutputofthefunctiongenerator.
4. Verifythemeterisdisplayingproperresultsfromthetable.
Step Range

Input

Display

Hz

60Hz@10V

59.4to60.6

Hz

1kHz@10V

.990to1.010k

Hz

1MHz@10V .990to1.010M

CAPACITANCETEST

1. SettheIUTtomeasurecapacitance,insertthetestleadsintotheV/ohmand
commontestterminals.
2. Connectthemetertothecapacitancetestfixtureandusethefollowingtableto
determinetheabilityoftheIUT.
Step

Range

Input

Display

Capacitance

10microF

9.85to10.15

Capacitance

1microF

.985to1.015

Capacitance 100microF 98.5to101.5

DIODETEST

1. SettheIUTtothediodecheckfunctionandconnectthetestleadstotheV/ohm
http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

10/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure

andcommonterminals.
2. Connectthetestleadstothediodetestfixtureandfollowthestepsinthe
proceedingtabletoverifydiodetestoperation.
Step

Range

Setup

Polarity

Display

Tone

DiodeCheck

Short

N/A

.000to.001

Yes

DiodeCheck

Si

Forward .800to.600

Yes

DiodeCheck

Si

Reverse

No

DiodeCheck

Ge

Forward .300to.500

Yes

DiodeCheck

Ge

Reverse

No

1.

1.

Note:Somemetersarenotdesignedtoemitatone.

ReasonforReissue
RevisionAFirstRelease

Software

Software

GetHelp

Forum

About

ContactUs

CalibrationControl

Printers

HelpPages&Videos

ForumHome

AboutUs

ContactUs

Download

Scanner

RemoteSupport

CalibrationControl

OurCustomers

EmailSales

VersionHistory

MinimumRequirements

OnlineMeeting

Feedback

ImportantLinks

EmailSupport

FeatureList

PDFBrochure

OnSiteServices

Twitter

SiteMap

DesktopSupport

http://www.apesoftware.com/calibrationprocedures/digitalmultimeter.aspx

11/12

23/01/2016

FreeDigitalMultimeterCalibrationProcedure