Fumehood Filed Test Report

Customer Name
FH Location

Contact Person
Contact Number
Certification Date

Square Pharmaceuticals Ltd.


GPB QC Lab (Analytical
Lab)
3
4
Mohammad Razib Ahamed
+8801772109977

Type of Test
Face Velocity Test
Smoke Pattern Test
N
o

Manufacturer
Model No

N/A
N/A

Serial No
Type of the FH
Certification Due Date

70-04-14-FH01
Ducted Fumehood

Test Equipments Details


Description
Serial
Anemometer (TSI)
T95450813004
Smoke Generator
N/A

Test Name

Acceptance Range

Results

Calibration Due Date


Sept, 2014
N/A
Remarks

Tested to

ASHRAE 110:1995
Standard
ASHRAE 110:1995
Standard
ASHRAE 110:1995
Standard

01

At 100% Sash Opening

Face Velocity Test


0.50 m/s
0.54 m/s

Pass

02

At 50% Sash Opening

0.83 m/s

0.87 m/s

Pass

03

At 25% Sash Opening

1.15 m/s

1.55 m/s

Pass

***
01
02

01

All readings should be with a deviation 20 % at all point of sash opening


Local Visualization
Challenge
Gross Visualization
Challenge

Flow visualization Test


Mentioned at 2.3
Fair

Pass

Mentioned at 2.3

Pass

Fair

Airflow Mointor or Indication Display


Airflow Mointor or Indication Display working
Please Circle
YES
NO
Good?

ASHRAE 110:1995
Standard
ASHRAE 110:1995
Standard
Remark :

1.0 Face Velocity test

During re-certification, face velocity measurements are required, but this value does not represent the
containment performance of fume hood. Test will be done according to ASHRAE 110:1995 Standard.
Have to take the reading for 3 positions. 100 % sash opening, 50 % sash opening and 25 % sash
opening.
1) Sash 100% Open

Diamed
205 Shahid Sayed Nazrul Islam Swarani, Dhaka -1000.
Email : info@diamedbd.com

1
2Filed Test3Report
Fumehood
Square Pharmaceuticals Ltd. Manufacturer
Model No
GPB QC Lab (Analytical
Lab)
Serial No
Mohammad Razib Ahamed
Type of the3FH
+8801772109977
1
2
Certification Due Date

Customer Name
FH Location
Contact Person
Contact Number
Certification Date

4
N/A
N/A
70-04-14-FH01
4Ducted Fumehood

300mm
2) Sash 50% Open

300mm
3) Sash 25% Open

300mm

1.1

Face Velocity Test Results Table

Sash 100% open Test Data


Point
1

Point
2

Point
3

Point
4

Point
5

Point
6

Point
7

Point
8

0.52

0.54

0.55

0.52

0.56

0.53

0.52

0.54

Average

Acceptance
Met (Yes/No)

0.54 m/s

Yes

Sash 50% open Test Data


Point
1

Point
2

Point
3

Point
4

0.88

0.85

0.86

0.85

Average

0.86 m/s

Acceptance
Met (Yes/No)

Sash 50% open Test Data

Diamed
205 Shahid Sayed Nazrul Islam Swarani, Dhaka -1000.
Email : info@diamedbd.com

Yes

Fumehood Filed Test Report


Customer Name
FH Location
Contact Person
Contact Number
Certification Date

Square Pharmaceuticals Ltd.


GPB QC Lab (Analytical
Lab)
Mohammad Razib Ahamed
+8801772109977

Manufacturer
Model No

N/A
N/A

Serial No
Type of the FH
Certification Due Date

70-04-14-FH01
Ducted Fumehood

Point
1

Point
2

Point
3

Point
4

1.54

1.57

1.59

1.56

1.55 m/s

Average

Acceptance
Met (Yes/No)

Yes

2.0 Flow Visualization Test


2.1 Purpose
To visualize a hoods ability to contain vapors and any leak in the hood. This test consist of a
small local challenge and gross challenge to the Fume Hood.

2.2 Test Method


Local Visualization Challenge
1) A stream of smoke is discharged from the stick
Along the left and right walls.
Along the joints between the walls and the work surface.
Behind the airfoil
The work surface of the hood in a line parallel to the hood face
6 in. (150mm) behind the face of the hood.
Along top of the face opening
In an 8-in.(200mm) diameter circle on the back of the hood.
2) The smoke shall be carried to back of the hood and exhausted smoothly and not be
entrained in vortex at the top of the hood.
Gross Visualization Challenge
1) Place smoke Generator inside the cage, to release a large volume in the centre of the work
surface width and 6 in.(150mm) behind the sash. Start up the smoke generator and
observe the direction of the smoke flow.
2) Observe carefully the flow pattern and watch out for reverse flow direction of smoke and
escape from the Fume Hood. Grade the Fume Hood performance based on the table
above.

Diamed
205 Shahid Sayed Nazrul Islam Swarani, Dhaka -1000.
Email : info@diamedbd.com

Fumehood Filed Test Report


Customer Name
FH Location
Contact Person
Contact Number
Certification Date

Square Pharmaceuticals Ltd.


GPB QC Lab (Analytical
Lab)
Mohammad Razib Ahamed
+8801772109977

2.3 Acceptance Criteria


Grading
Fail
Poor
Fair
Good

Manufacturer
Model No

N/A
N/A

Serial No
Type of the FH
Certification Due Date

70-04-14-FH01
Ducted Fumehood

Observations

Smoke observed escaping from the hood.

Reverse flow of smoke near opening.


Lazy flow into opening along boundary.
Observe potential for escape.
Some reverse flow, not necessarily at opening.
No Visible escape.

No reverse flows. Active flow steam into hood around boundary.

2.4 Flow visualization Test Results Table


Test Name

Local Visualization
Challenge

Observation
Some

Gross Visualization
Challenge

reverse flow,
necessarily at opening

Grade

Acceptance
Met (Yes/No)

Fair

Yes

Fair

Yes

not

No Visible escape

Tested by :

Jinnatur Rahman Bipul


Service Engineer
Diamed

Diamed
205 Shahid Sayed Nazrul Islam Swarani, Dhaka -1000.
Email : info@diamedbd.com