COMPARISON OF THE RESPIRABLE FRACTION

FROM THREE DIFERENT DPI DEVICES

Miriam Sanz Cermeo and Helena Maria Cabral Marques


UCTF, Faculdade de Farmcia, Universidade de Lisboa, PORTUGAL

1. Introduction

Inhalation is a method of delivery which has been known and used for decades, mainly in
respiratory diseases. There are two ways of treating asthma: a) to dilate the bronchi, i.e. to treat
an asthma attack as it occurs and the drugs used are bronchodilators such as salbutamol; b) to
inhibit bronchoconstriction, by attacking the inflammation with steroidal anti-inflammatories.
Dry powder inhalers (DPIs) offer a unique opportunity for the delivery of drugs to the lung as
aerosols.
These devices combine powder technology with device design in order to disperse dry particles
as an aerosol in the patients inspiratory airflow [1].
The deposition of particles is largely controlled by their impaction on pulmonary surfaces. The
larger particles (> 20 m) impact in the oropharynx; the particle velocity increases as the
airways become narrower at deeper levels, and successively smaller particles impact in narrower
vessels. Thus the terminal bronchioles are only reached by particles smaller than 2-3 m.
Unfortunately these smallest particles may not impact at these deep levels, since they do not
diffuse rapidly enough to encounter an epithelial surface before exhalation removes them from
the respiratory tract [2].
To introduce drug particles into the lung they must be < 5 m in aerodynamic diameter [1]. This
is generally achieved by milling the powder prior to formulation. Small particles are notoriously
difficult to disperse. The forces governing dispersion are well documented and consist mainly of
electrostatic, Van der Waals, and capillary forces [1].
One approach that has been taken to improve the dispersion of dry powders is the inclusion of an
excipient, notably lactose. The lactose particles are intended to act as carrier particles for the
drug and as such are in a much large size range, 60-80 m [1]. The drug particles are dispersed
and can traverse the upper respiratory tract while the excipient particles do not pass beyond the
mouth piece of the device or the mouth and throat of the patient [1].
In order to study better these excepient / drug interations, as dry powders, three devices were
compared these work three devices were compared: the Microhaler, the Rotahaler and the
FlowCaps.

2. Materials and Methods

2.1. Devices
The three devices used are capsule-based inhalers:
MicrohalerTM - the capsule is pierced manually at both ends, using the pin mechanism existing
inside the device. The powder is then released through the capsule holes and inhaled through a
screened tube by means of the circulating air which generates a rotating motion in capsule (i.e.
spinning of the capsule).
RotahalerTM - the capsule is inserted into the device, wherein, by rotating the device it is opened
(or broken) in two halves: the capsule body containing the powder falls into the device, while the
capsule cap is retained in the entry port (hole) and the powder is inhaled through a screened tube
[3]. The capsule body containing the powder experiences an erratic motion in the air stream,
causing dislodge particles to be entrained and inhaled [1].
Flowcaps - is based on a new concept in capsule-based inhalers: the dancing cloud which
causes the emission of the powder contents of a motionless capsule [4]. The capsule is cut at
both ends by blades. As the admission of air into the capsule is severely restricted through very
small cuts, i.e. narrow slits, a low pressure is created in that area of the capsule, causing the
powder to rush towards it (against the direction of the airflow), and it gradually becomes
entrained towards the mouthpiece [4].
The figures represent the 3 devices used:
1) MicrohalerTM

2) Rotahaler

3) FlowCaps

2.2. Deposition of salbutamol sulphate


The in vitro deposition of pulmonary aerosol formulation was tested using the Twin-stage
liquid impinger (Twin Impinger, Copley, U.K.) [5], a two-stage separation device for assessing
the drug delivery from inhalation delivery devices. The discharged aerosol is fractionated by
firing through a simulated oropharynx and then through an impinger stage of defined
aerodynamic particle size cut-off characteristics [6]. The fine (pulmonary) fraction which
penetrates is collected by the lower impinger.

This glass impinger, in which the aerosol particles impinge on to the liquid and surfaces due to
their inertia in a deflected airstream, has proved valuable for routine quality assessment of
aerosols during product development, stability testing and for quality assurance and comparison
of commercial products [6]. To enable correct functioning of most powder inhalers, a airflow of
60 L min-1 is recommended [5, 6].
The blend tested, salbutamol sulphate and lactose, batch n F6-9617, was prepared by using a
special mixer. The blend particle size distribution obtained by dry dispersion (Malvern
Mastersizer) was as follows: 90 % < 98.72 m; 50 % < 62.28 m and 10 % < 22.62 m.
Hard gelatine capsules n. 4 (for the FlowCaps) and n. 3 (for the Microhaler and
Rotahaler) were filled with 22 mg (Balance: Mettler AG 204. Delta Range) of the mixture
(corresponding to 200 g salbutamol per capsule). These capsules were then placed in the
Microhaler, Rotahaler or FlowCaps Dry Powder Inhaler devices and tested using the Twinstage liquid impinger [5]: 7 and 30 ml of Hydrochloric acid 0,1 M were introduced into the upper
and lower impingement chambers, respectively. The vacuum pump operated for 5 seconds at 60
L/min [6, 5], air flow rate and the capsule content was discharged by the turbulent air stream.
Five capsules were fired in succession for each determination.
At the end of this operation, the apparatus was disassembled and the inner surfaces were washed
separately with Hydrochloric acid 0,1 M. The amount of active substance collected in each of the
stages or portions (capsule + DPI device, throat, upper and lower impingement chambers) was
assayed by UV spectrophotometry (UV / VIS Spectrophotometer, Hitachi U 2000) at 276 nm,
and the salbutamol was quantified.

3. Results and Discussion

Besides the results for the respirable fraction (Lower impingement chamber), results obtained for
all other compartments of the Twin Impinger: Throat, Upper impingement chamber and also the
remaining dose in the capsule and device with the three different devices are shown in the
following table:

Microhaler

Capsule+device

Throat

Upper

Lower

Total

Determ. 1

11,4

6,5

77,9

12,0

107,8

Determ. 2

10,2

7,1

73,6

9,5

100,4

Determ. 3

11,4

8,3

79,7

8,3

107,7

Average

11,0

7,3

77,1

9,9

105,3

SD

0,7

0,9

3,1

1,9

4,2

rsd

6,3

12,6

4,1

19,0

4,0

Rotahaler

Capsule+device

Throat

Upper

Lower

Total

Determ. 1

28,6

13,2

56,3

7,1

105,2

Determ. 2

29,2

12,0

48,3

7,1

96,6

Determ. 3

25,5

13,8

57,6

7,7

104,6

Average

27,8

13,0

54,1

7,3

102,1

SD

2,0

0,9

5,0

0,3

4,8

rsd

7,2

7,1

9,3

4,8

4,7

FlowCaps

Capsule+device

Throat

Upper

Lower

Total

Determ. 1

8,9

17,5

58,8

15,7

100,9

Determ. 2

9,5

18,8

52,6

14,5

95,4

Determ. 3

9,5

18,8

55,7

14,5

98,5

Average

9,3

18,4

55,7

14,9

98,3

SD

0,3

0,8

3,1

0,7

2,8

rsd

3,7

4,1

5,6

4,7

2,8

These results are expressed as the percentage of drug dose filled into one capsule (200 g).
As the DPI devices combine powder technology with device design, different devices may give
different results even with the same formulation and in the same conditions. This is one of the
reasons for the observed differences between the three devices tested.
All these systems have a mechanism for aerosolising the powder. Reproducible dose metering
and dispersion characteristics are affected by particle size, rugosity, shape, moisture content,
surface chemical composition and charge [1]. As the powder blend tested was from the same
batch for all experiments performed in this work (uniformity of content in 20 samples tested had
a rsd of less than 4 %), and consequently those characteristics were the same for all
determinations, there were small differences intra-device (relative standard deviations are
reasonable for most of the compartments: between 3.7 and 5.6% for the FlowCaps, between for
4.8 and 9.3% for the Rotahaler, and between 4.1 and 19.0% for the Microhaler).
The data show that FlowCaps had the best shot-to-shot reproducibility.
The induction of turbulent flow in narrow tubes can be associated with an enhanced
deaggregation of the powder agglomerates. For this reason internal geometry of the device is of
great importance, for example the dimensions of channels through which the inspired airflow
passes and the release mechanism of the powder from the capsule. This may cause the
differences between the three devices as the Microhaler delivers the dose through a capsule
hole made by a pin, the Rotahaler delivers the dose through the open capsule and the
FlowCaps delivers the dose through a cut made by a blade.

For a more illustrative information the same results are listed out also as the emitted fine particle
fraction in the following tables:

Microhaler
Cap. & Dev.(
g) Stage I (
g) Stage II (
g)

Emitted dose
(
g)

FPF (% < 6.4 g)


% emitted
% nominal
dose
dose
12,4
12,0

Total recovery
% nominal
dose
107,8

Determ. 1

22,8

168,8

24,0

192,8

Determ. 2

20,4

161,4

19,0

180,4

10,5

9,5

100,4

Determ. 3

22,8

176,0

16,6

192,6

8,6

8,3

107,7

Average

22,0

168,7

19,9

188,6

10,5

9,9

105,3

SD

1,4

7,3

3,8

7,1

1,9

1,9

4,2

rsd

6,3

4,3

19,0

3,8

18,2

19,0

4,0

Rotahaler
Cap. & Dev.(
g) Stage I (
g) Stage II (
g) Emitted dose (
g)

FPF (% < 6.4 g)

Total recovery

% emitted dose % nominal dose % nominal dose

Determ. 1

57,2

139,0

14,2

153,2

9,3

7,1

105,2

Determ. 2

58,4

120,6

14,2

134,8

10,5

7,1

96,6

Determ. 3

51,0

142,8

15,4

158,2

9,7

7,7

104,6

Average

55,5

134,1

14,6

148,7

9,8

7,3

102,1

SD

4,0

11,9

0,7

12,3

0,6

0,3

4,8

rsd

7,2

8,9

4,7

8,3

6,5

4,7

4,7

FlowCaps
Cap. & Dev.(
g) Stage I (
g) Stage II (
g) Emitted dose (
g)

FPF (% < 6.4 g)

Total recovery

% emitted dose % nominal dose % nominal dose

Determ. 1

17,8

152,6

31,4

184,0

17,1

15,7

100,9

Determ. 2

19,0

142,8

29,0

171,8

16,9

14,5

95,4

Determ. 3

19,0

149,0

29,0

178,0

16,3

14,5

98,5

Average

18,6

148,1

29,8

177,9

16,7

14,9

98,3

SD

0,7

5,0

1,4

6,1

0,4

0,7

2,8

rsd

3,7

3,3

4,6

3,4

2,4

4,6

2,8

where:
Emitted dose = Stage I + Stage II
Fine Particle Fraction (FPF) % emitted = ((Stage II)/(Stage I + Stage II))*100
Fine Particle Fraction (FPF) % nominal = (Stage II/nominal dose)*100
Total Recovery = Capsule and device retention + Stage I + Stage II
Nominal dose = capsule content i.e. 200 g

The RotahalerTM not only has poor emptying performance (showed by the high amount retained
in device and capsule), but also a poor aerosolisation efficiency (as a low % FPF was obtained).
Both MicrohalerTM and FlowCaps have good emptying efficiencies (as showed by the low
amount retained) and Flow Caps has the higher % FPF (is it significantly different, p < 0.05).
The data seems to support Hovione's claim that by designing a device where the capsule is
stationary so all possible energy from patient's inhalation is directed to aerosolising the powder
(rather than rotate and/or shake the capsule too), an improved performance can be obtained.
It should be noted that the FlowCaps device used in this work (code-named Dolphin) was not
yet in the market and at the moment Hovione has evolved the FlowCaps device (code-named
Midget) which resulted from an improvement / upgrading of the former device.

4. Conclusions

FlowCaps seems to be the best device for the salbutamol sulphate - lactose mixture used, under
our working conditions if compared to the MicrohalerTM and RotahalerTM. The results obtained
with the devices can be ranked in the following order: FlowCaps > MicrohalerTM >
RotahalerTM.

5. References

Hickey, A. J., "Inhalation Aerosols: Physical and Biological Basis of Therapy ". Marcel Dekker,
New York. 1996. P: 451-467.
Swarbrick, J., Boylan, J.C., "Encyclopaedia of Pharmaceutical Technology". 1995. Vol. 12. P:
160, 416-420.
Swarbrick, J., Boylan, J.C., "Encyclopaedia of Pharmaceutical Technology". 1994. Vol. 9. P:
287-292 and 321-325.
Villax, P., Brito, V., McDerment, I. "A Capsule-based dry powder inhaler". In: FlowCaps
Information Pack. DY002-rev. 5. June 1997. P:1-4.
Eur. Pharm. 1997. 3rd edition. P: 144.
Hallworth, G.W., Westmoreland, D.G. "The Twin-Impinger: a simple device for assessing the
delivery of drugs from metered dose pressurized aerosol inhalers". J. Pharm. Pharmacol. 1987,
39: 966-972.

6. Acknowledgements

We gratefully acknowledge Hovione - Produtos Farmacuticos S.A. for providing the


FlowCaps device.
This work was partially supported by the SOCRATES / ERASMUS European Programme.