Professional Documents
Culture Documents
16881696
An Overall Stem Effect, including Stem Leakage and Stem Scatter, for a
TM30013 Farmer-type Chamber
Dae Cheol Kweon
Department of Radiologic Science, Shin Heung College University, Uijeongbu 480-701, Korea
Jae-Seung Lee
Department of Radiation Oncology, Good Samaritan Hospital, Pohang 791-704, Korea and
Department of Physics, Soonchunhyang University, Asan 336-745, Korea
Eun-Hoe Goo
Department of Radiology, Seoul National University Hospital, Seoul 110-744, Korea and
Department of Physics, Soonchunhyang University, Asan 336-745, Korea
Moon-Jib Kim
Department of Physics, Soonchunhyang University, Asan 336-745, Korea
Jae-Eun Jung
Department of Biomedical Engineering, Sahm Yook Seoul Medical Center, Seoul 130-711, Korea
Kyung-Rae Dong
Department of Radiological Technology, Gwangju Health College University, Gwangju 501-701, Korea and
Department of Nuclear Engineering, Chosun University, Gwangju 501-759, Korea
Woon-Kwan Chung
Department of Nuclear Engineering, Chosun University, Gwangju 501-759, Korea
-1688-
An Overall Stem Effect, including Stem Leakage and Stem Scatter Dae Cheol Kweon et al.
-1689-
Table 1. Physical characteristics of the Farmer-type ionization chamber used in this study.
Chamber
modela)
TM30013
Length
(mm)
26.0
Volumne
(cm3 )
0.600
Wall thickness
PMMA
(g/cm3 )
1.19
Graphite
(g/cm3 )
1.85
Stem materials
(Purity of Al, %)
Chamber
Dummy
stem
stem
99.98
99.28
Length of stem
(mm)b)
107.7
a)
b)
I. INTRODUCTION
the calibration [14]. In this case, due to radiation interaction, electrons ejected from air near the chamber end,
from the dielectric in the metal stem or from the cable
can reach the central electrode, generating the stem effect, which reduces charges [1416]. Such a stem effect
depends on the length of the chamber stem in the field
border. Thus, a correction will be necessary whenever
the length of the exposed stem differs from that at the
time of the chamber calibration [15].
Therefore, this study intends to calculate the stem effect correction factor (Psem ). This factor is the result,
which is shown in different forms, of a leakage current
that is frequently inevitable in the measurement of exposure doses of high-energy photon beams. We measured the dependences of stem leakage (ksem.leak ) and
the stem scatter (ksem.scatter ) on the length of the chamber stem exposure when measuring the exposure dose of
high-energy X-rays generated by using a linear accelerator (LINAC), and we calculated the overall stem effect
correction factor that was caused by the metal stem of
the ionization chamber. In addition, we calculated and
compared the ratios of the stem leakage and the stem
scatter to the overall stem effect.
-1690-
2. Measurement Methods
Journal of the Korean Physical Society, Vol. 58, No. 6, June 2011
The calibration of the stem leakage was done by turning around the ionization chamber or the collimator. In
ter
A. Measurement of the stem leakage
An Overall Stem Effect, including Stem Leakage and Stem Scatter Dae Cheol Kweon et al.
-1691-
-1692-
Journal of the Korean Physical Society, Vol. 58, No. 6, June 2011
kstem.scatter =
(2)
Table 2. Chamber orientation and stem material dependencies, with their estimated relative uncertainties, for the
TM 30013 ionization chamber measured for 6- and 10-MV
X-rays.
Values
(3)
(4)
kstem.scatter
100(%).
Pstem
(5)
Rstem.leak =
kstem.scatter =
III. RESULTS
We used the PTW model TM 30013 (vented sensitive
volume of 0.6 cm3 ) Farmer-type chamber for high-energy
X-rays emitted from a LINAC with a view to calculating
the dependence of the overall stem effect correction factor of the ionization chamber on the length of the stem
exposed in the exposure measurement. The results are
as follows:
TM30013
6 MV
10 MV
0.0101
0.0179
0.0244
0.0207
0.0271
0.0397
0.0248
0.0298
a)
An Overall Stem Effect, including Stem Leakage and Stem Scatter Dae Cheol Kweon et al.
-1693-
Table 3. Stem leakage correction factors (kstem.leak ) determined for various lengths of the stem exposed to 6- and
10-MV X-ray in the PTW model TM30013 (vented sensitive
volume of 0.6 cm3 ) Farmer-type ionization chamber.
Table 4. Stem scatter correction factors (kstem.scatter ) determined for various lengths of the stem exposed to 6- and
10-MV X-ray in the PTW model TM30013 (vented sensitive
volume of 0.6 cm3 ) Farmer-type ionization chamber.
6 MV
Length of stem
b)
a) Relative
exposed (cm) kstem.leak
uncertainty
1
1.0004
0.0335
2
0.9998
0.0306
3
0.9979
0.0300
4
0.9965
0.0221
5
0.9967
0.0327
6
0.9980
0.0314
7
0.9991
0.0291
8
0.9995
0.0335
9
0.9989
0.0258
10
0.9995
0.0298
11
0.9991
0.0277
12
0.9993
0.0249
13
0.9997
0.0233
14
0.9993
0.0180
15
0.9994
0.0340
Length of
6 MV
10 MV
stem
exposed
Relativeb)
Relative
kstem.scatter
kstem.scatter a)
(cm)
uncertainty
uncertainty
1
0.9752
0.0260
0.9712
0.0340
2
0.9734
0.0394
0.9689
0.0300
3
0.9724
0.0298
0.9684
0.0267
4
0.9695
0.0267
0.9660
0.0394
5
0.9684
0.0327
0.9655
0.0260
6
0.9664
0.0327
0.9630
0.0260
7
0.9680
0.0298
0.9637
0.0340
8
0.9688
0.0340
0.9662
0.0416
9
0.9690
0.0307
0.9677
0.0333
10
0.9689
0.0153
0.9684
0.0277
11
0.9696
0.0300
0.9690
0.0267
12
0.9723
0.0300
0.9690
0.0267
13
0.9723
0.0300
0.9719
0.0249
14
0.9739
0.0300
0.9737
0.0277
15
0.9742
0.0269
0.9736
0.0359
10 MV
Relative
kstem.leak
uncertainty
1.0003
0.0269
0.9997
0.0371
0.9982
0.0291
0.9964
0.0267
0.9968
0.0214
0.9980
0.0482
0.9989
0.0277
0.9992
0.0249
0.9989
0.0277
0.9992
0.0258
0.9989
0.0359
0.9995
0.0249
0.9995
0.0291
0.9992
0.0348
0.9993
0.0298
a)
average uncertainty was 2.55 102 % for the measurement repeated for ten times with the same measurement
method.
In a square field where the length of the stem exposure was increased by 1 cm, we measured the ionization
current after placing the dummy stem in a direction opposite to the sensitive volume of the ionization chamber
and to the ionization current after removing the dummy
stem. Based on measured values of the ionization current
and Eq. (2), we calculated the uncertainty for the stem
scatter correction factor (kstem.scatter ) and the repeated
measurements. Table 4 show the relation of the stem
scatter correction factor (kstem.scatter ) to the length of
stem exposure when 6-MV and 10-MV X-rays were used
for the Farmer-type chamber of TM 30013 in a PTW
model. Unlike the case with stem leakage, as soon as the
beam reached the chamber stem, 2.48% and 2.88% of
the stem scatter took place in the metal stem for 6- and
10-MV X-rays, respectively. The stem scatter increased
gradually with increasing length of stem exposure before
reaching maximum values of 3.36% for 6-MV X-rays and
a)
-1694-
Journal of the Korean Physical Society, Vol. 58, No. 6, June 2011
IV. DISCUSSION
overall stem effect. Figure 4 shows the overall stem effect correction factor (Pstem ), which includes both the
stem leakage and the stem scatter that may take place
in the metal stem of the ionization chamber. The result shows that the stem effect contributed 2.44 3.56%
and 2.85 3.89% to exposure dose in all regions of the
metal stem for 6- and 10-MV X-rays, respectively. In
particular, the stem effect contributed to the exposure
dose at the maximum field size (which included the calibrated field size of the ionization chamber at 10 10
cm2 ) mostly used in clinical trials. Figure 5 shows the ratios at which the stem leakage (Rstem.leak ) and the stem
scatter (Rstem.scatter ) contributed to the overall stem effect. Excluding the middle part of the chamber stem (the
length of the stem exposure is 4 6 cm), the stem effect
caused by the stem leakage was approximately less than
5%, which is very small, and around 95% of the stem
effect took place in stem scatter. Therefore, as shown in
An Overall Stem Effect, including Stem Leakage and Stem Scatter Dae Cheol Kweon et al.
ward. Such stem leakage showed similar correction factors for 6- and 10-MV X-rays without any dependency
on the photon energy (Table 3).
Furthermore, the measurement result for stem scatter
showed that unlike the case with the stem leakage, as
soon as the beam reached the chamber stem, the scattered photons generated in the metal stem made a contribution to the exposure dose. As the length of the stem
exposure increased, the stem scatter increased gradually
and reached maximum values of 3.36% for 6 MV X-rays
and 3.70% for 10 MV X-rays in the middle part of the
chamber stem (the length of the stem exposure is 6 cm)
before decreasing on an irregular basis afterward. Such
stem scatter increased with increasing photon energy in
the chamber stem and showed similar correction factors
regardless of the photon energy after the end part of the
stem (Table 4).
Based on the results mentioned above, it can be said
that the stem effect, including stem leakage and the stem
scatter, was attributable to (1) the ions within the stem
that could be measured and (2) the ions that were generated between the end part of the stem and the cable. The
electrons emitted from the air in the space from the middle part of the stem to the end part of the stem have an
increasing probability of ion recombination as they move
farther away from the collection electric field, which hinders a normal measurement. Therefore, the stem effect
decreased gradually after the middle part of the chamber stem. However, as the beam reached the metal stem,
the electrons emitted from the metal stem arrived at the
center electrode, making a clear contribution to the exposure dose. Such a stem effect was found to increase
linearly up to the middle part of the metal stem. In
conclusion, the Farmer-type ionization chamber that is
frequently used to measure the exposure dose, with a
sensitive volume of 0.6 cm3 , showed a stem effect of less
than around 4% that increased linearly up to the middle
part of the metal stem, but afterward decreased linearly
because the distance from the collection electric field became large, even though the length of the stem exposure
increased, which eventually increased the recombination
of ions. In such a region of ion recombination, the stem
effect showed similar correction factors regardless of the
photon energy because the effect had nothing to do with
the number of the electrons emitted from the metal stem.
In this study on the stem effect, we suggest two points.
First, when the ionization chamber has a small sensitive
volume and the a metal stem with a operating voltage of
400 V (for example, the ionization chamber has a sensitive volume of 0.125 cm3 with a stem length of 4.25 cm),
the ion recombination region will decrease, and the number of the electrons collected in the center electrode will
increase because the distance between the center electrode and the chamber stem decreases. Therefore, the
stem effect will increase linearly throughout the metal
stem, which requires consideration of more correction
factors to the stem effect. Second, it should be noted
that the stem effect makes it, maximum contribution to
-1695-
V. CONCLUSION
We measured the stem effect of a Farmer-type ionization chamber that had recently been in use for exposure
dose measurements and present some conclusions based
on the measurement results. We repeated the measurements for five weeks in the same method to observe the
stem effect for various changes in the measurement environment. The measurement results showed that the stem
effect increased slightly with decreasing temperature or
increasing pressure. Since the stem effect correction factor showed a slight change, not absolutely but relatively,
according to the given measurement environment, an individual measurement of the stem effect of ionization
chambers to be applied in clinical trials in the division of
radiation oncology, are required. Furthermore, radiation
therapy focuses on verification of prescription dose at
a certain depth of tissue rather than on measurement of
the exposure dose, so that the absorbed dose is measured
in most of the cases. Also there exist various absorbers
with the different depths for measurement. For these
reasons, we believe that a study should be conducted on
a new stem correction method for the absorbed dose.
REFERENCES
[1] G. F. Knoll, Radiation Detection and Measurement, 3rd
ed. (John Wiley & Sons, Michigan, 2000), p. 140.
[2] B. Johansson and G. Wickman, Phys. Med. Biol. 42, 133
(1997).
[3] International Commission on Radiation Units and Measurements, ICRU Report No. 24, 1972.
[4] International Commission on Radiation Units and Measurements, Use of Computers in External Beam Radiotherapy with High-energy Photons and Electrons, ICRU
Report (ICRU, Bethesda, MD, 1987), No. 42.
Journal of the Korean Physical Society, Vol. 58, No. 6, June 2011
791 (1966).
[17] American Association of Physicists in Medicine, RTC
Task Group 51, Med. Phys. 26, 1847 (1999).
[18] International Atomic Energy Agency, Absorbed Dose Determination in External Beam Radiotherapy, Technical
Reports Series (IAEA, Vienna, 2000), No. 398.
[19] H. Tolli and K. A. Johansson, Phys. Med. Biol. 43, 3171
(1998).
[20] J. S. Tasi, M. J. Rivard and M. J. Engler, Med. Phys.
27, 2215 (2000).
[21] C. M. Ma and A. E. Nahum, Phys. Med. Biol. 40, 63
(1995).
[22] C. M. Ma, C. W. Coffey, L. A. DeWerd, C. Liu, R. Nath,
S. M. Seltzer and J. P. Seuntjens, Med. Phys. 28, 868
(2001).
[23] G. A. Ezzel, J. M. Galvin, D. Low, J. R. Palta, I. Rosen,
M. B. Sharpe, P. Xia, Y. Xiao, L. Xing and C. X. Yu,
Med. Phys. 30, 2089 (2003).
[24] E. H. Goo, J. S. Lee, M. J. Kim, K. R. Dong, D. C.
Kweon and W. K. Chung, J. Korean Phys. Soc. 57, 506
(2010).
[25] P. R. Almond, Med. Phys. 37, 3011 (2010).