Professional Documents
Culture Documents
GROUP: 5 Thursday
Aims:
To use a pipette and check the accuracy and reproducibility of the method
Results
Part B: Pipetting accuracy and reproducibility
Complete the following table with the data from Ex1 Table 1 from the laboratory
manual (p. 41), and calculate the mean and percentage error for each
measurement.
ER1 Table 1: Measurements of dH20 containing 5 g mL-1 methyl red volumes
to assess pipetting accuracy and reproducibility
Pipet
te
setti
ng
(L)
Reading 1
(mg)
Reading 2
(mg)
Reading 3
(mg)
Mean
(mg)
% Error*
(%)
A20
5
0.00
16
16
15
17
16
0.06
55
57
59
56
57.33
0.04
180
181
179
180
180
0.00
A200
GROUP: 5 Thursday
A1000
245
244
247
247
246
0.004
750
751
752
751
751.33
0.002
*The difference between the mean and the expected value should be expressed as a percentage of
the expected value (e.g. if the mean = 48 mg and the expected value = 50 mg, the percentage
error is 2/50 100 = 4%
1/1
1/3
Absorbance
Wavelength
From Fig. 1 and Fig. 2, identify the wavelength where the Absorbance value
(Ordinate value) is at a maximum (i.e. what is max?)
a. Potassium
chromate
max
b. DNA
0.5889
Discussion
Part B
Comment on the reproducibility and accuracy of your pipetting technique, and whether it
differed for the three pipettes tested.
The reproducibility and accuracy of pipetting technique was within approximately 5%
error throughout the three pipettes. Generally, pipettes became more accurate from A20
to A1000.
What steps could you take in future to improve the reproducibility and accuracy (if
necessary)?
Hint: Refer to Fig. 3.
The most obvious area that could be improved on is using the recommended immersion
depth to improve reproducibility and accuracy. Through practice, a rough idea on how
deep to immerse the tip can improve (e.g. 1-2mm for a small pipette).
Part C & D
Explain the differences in parameters and method of the wavelength scans
performed for the two different solutions (i.e. potassium chromate versus DNA).
The parameters used for the wavelength scans performed for potassium
chromate and DNA were almost identical except for the range of wavelengths
used. A visible light spectrophotometry was performed for potassium chromate
from 300 nm - 500 nm. DNA was scanned using UV spectrophotometry and
parameters were set lower from 200 nm - 350 nm.
Conclusion
Week 1's practical served as an introduction to techniques of molecular sciences
and laboratory work. Pipettes, scales, spectrophotometers, centrifuges were all
demonstrated during the practical followed by hands on practical work in diluting
solutions with pipettes and analysing them using spectrophotometry. Not
everything went according to plan, the values for Part C are wrong because the
cuvette was placed the wrong way in the spectrophotometer. However, mistakes
such as this as well as self-feedback from analysing accuracy of results will
become a learning experience and can only be improved upon.