CARD OPTION

AX

Reference Image Storage


with Display 2000
Installation and Setting Instruction

S iem ens AG 1998


The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

Print No.: RA43-030.031.09.02.02


Replaces: RA43-030.031.09.01.02

English
Doc. Gen. Date: 05.98

0-2

Revision

Chapter

Page

all

all

01

all

all

02

CARD OPTION

Revision

RA43-030.031.09
Rev.02
05.98

Page 2 of 4
TDPS 21

Siemens AG
Medical Engineering

Contents

0-3
Page

1 _______Requirements__________________________________________________ 1 - 1
System Requirements . . . . . . . . . . . . . . . .
Required Documents . . . . . . . . . . . . . . . .
Required Test Equipment and Meters . . . . . . . .
Parts List . . . . . . . . . . . . . . . . . . . . . .
Safety Information . . . . . . . . . . . . . . . . . .
Protection of Electrostatically-sensitive Components
Startup Information . . . . . . . . . . . . . . . . .

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.1-1
.1- 1
.1-1
.1- 1
.1-2
.1-2
.1-2

2 _______Installation ____________________________________________________ 2 - 1
Programming of DISPLAY 2000 . .
Installation of Reference Monitors .
M100 Modifications . . . . . . . .
KOORDINAT U Modifications . . .
KOORDINAT M Modifications . . .
Installation of Handswitch . . . . .
Overview Cabling . . . . . . . . .

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.
.
.
.
.
.
.

.2-1
.2- 3
.2-3
.2- 4
.2- 5
.2- 5
.2-6

3 _______Reference Image with Quantcor Option ____________________________ 3 - 1


Modifications Quantcor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Overview Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2

4 _______Startup _______________________________________________________ 4 - 1
HICOR Configuration . . . .
SIMOMED H Configuration .
SIMOMED HM Configuration
SIMOMED H/HM Adjustment
Function Test . . . . . . . .

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.
.
.
.
.

.4- 1
.4- 2
.4- 2
.4- 3
.4 - 3

5 _______Changes to previous version _____________________________________ 5 - 1

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 3 of 4
TDPS 21

CARD OPTION

0-4

Contents

This page intentionally left blank.

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 4 of 4
TDPS 21

Siemens AG
Medical Engineering

Requirements

1-1

System Requirements

This procedure is only valid for COROSKOP / BICOR TOP systems.

Required Documents

Description, Terminal Emulator Program: R43-030.114.01...


Setting Instructions, HICOR: RA43-030.032.06...
Adjustment Instruction, Simomed H: RA52-040.071.01...
Installation Instructions for Monitor Suspension (MTS): R53-040.031.01...

Required Test Equipment and Meters

Standard tool kit


AMP Tools
Ground wire test meter (see ARTD 3)
Service PC with Terminal Emulation Program (e.g. ST 320)
Serial cable (9-pole ... 9-pole) for connecting the PC to the HICOR / SIMOMED

Parts List

Reference Option Kit

47 96 939

Display 2000

27 98 671

Monitor Simomed HM 44 cm or

37 92 079

Monitor Simomed HM 54 cm or

37 92 087

Monitor Simomed H 44 cm or

11 63 000

Monitor Simomed H 54 cm

11 63 018

For installation of the Simomed HM monitor on a table:


- Rotating tilt console, Part No. 47 86 856 X2077 (44 cm)
- Rotating tilt console, Part No. 27 98 051 X2078 (54 cm)
- Power connector, Part No. 10 61 829 B1948
For installation of the Simomed HM monitor in the MTS:
- Adapter Kit 44, Part No. 37 92 350 X2077
- Adapter Kit 54, Part No. 27 98 069 X2078
- Power connector, Part No. 10 61 829 B1948

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 1 of 2
TDPS 21

CARD OPTION

1-2

Requirements
Safety Information

The "Safety Information" in the system manual and in the particular unit instructions must
be observed.

Protection of Electrostatically-sensitive Components

Integrated circuits and circuit boards on which MOS components are installed call for particularly careful handling because of their highly sensitive electrical structures and
because of their extremely high input resistance.
The ESD regulations that apply for electrostatically-sensitive components must be
observed.

Startup Information

If any work steps are carried out on the system, power must first be switched off.
The I.I. TV channel setting must be optimal.
The SIMOMED HM monitor(s) (ambient light sensor, standard changing time) must be
adjusted according to the adjustment instruction on service floppy

The HICOR has been fully tested and adjusted.

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 2 of 2
TDPS 21

Siemens AG
Medical Engineering

Installation

2-1

Programming of DISPLAY 2000

Note: The Display 2000 is composed of two identical, separate channels and contains two
identical circuit boards, one for each plane.

Remove the housing cover and install the jumpers as shown in the table (Fig. 1).
Close the housing.
Jumper

Position

Remark

J1

50/60 Hz

depends on line frequency

J2

OPEN

LIH

J3

INT

internal synchronization

J5

CLOSED

NORM

50Hz
60Hz

J3
Ext/Int

J1
50/60Hz

Component side

J5

J2

Fig. 1

75 Ohm switches on the front of the D2000: The three 75 Ohm switches belong to the
two outputs X12 and X22 and to input X30. If a socket is used, the corresponding switch
must be in the OFF position, if a socket is not used, the switch must be in the ON
position.

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 1 of 6
TDPS 21

CARD OPTION

2-2

Installation
Installation of Display 2000
The Display 2000 has to be installed in the M2 of the MTS.

Remove the two brackets from the Display 2000.


Take off the cover of the M2.
Slide the Display 2000 on the right hand side into the M2 and fix it with the screw (Fig. 2).
Connect the power cable to the Display 2000 and M2 X8 (if there are no empty terminals,
add one to X8.L1/N/Ground each and label them; the connecting bar needs to be cut to
the correct lenght).

Connect triax cable from Reference Monitor 1 to X13 at the Display 2000 using the
angled BNC connector (page 2-6).

For biplane: connect triax cable from Reference Monitor 2 to X23 at the Display 2000
using the angled BNC connector.

Pull a triax cable from the Hicor to the M2; unplug existing triax cable at J9 A and
establish new connection J9 A ---- X11 Display 2000.

For biplane: pull a triax cable from the Hicor to the M2; unplug existing triax cable at J9 B
and establish new connection J9 B ---- X21 Display 2000 (or: pull the triax cable from the
M2 to the control room and connect it to Display 2000 X14 and the 2nd Ref Monitor; Ref 1
images can then also be seen in the control room).

Pull a triax cable from the control room M101 to the M2 and connect it to Display 2000
X12; unplug the existing Hardcopy triax cable in the M101coming from J9 Hicor and
connect the new triax cable instead.

Pull the control signal cable from the M100 to the M2 and connect it at Display 2000
Remote.

M2

M2
X8,9

Display 2000

M4x8

Fig. 2

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 2 of 6
TDPS 21

Siemens AG
Medical Engineering

Installation

2-3

Installation of Reference Monitors

Install the monitor(s) in the examination room according to the Installation Instructions
for the monitor suspension system.

M100 Modifications

Install the sockets K21/1 and K21/2 in the frame of the M100 and connect the attached
wires to X109 and X110 (Fig. 3).

Connect cable D2000 Remote ---- K21/2.


Run control signal cable from M100 to the table and connect at M100 K21/1.

Diamentor
X109
X110

Z96
D94
Backplane

K21/1

Table

Display 2000
D95

K21/2

M100
Fig. 3

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 3 of 6
TDPS 21

CARD OPTION

2-4

Installation
KOORDINAT U Modifications

A
K23

Table
Control

Cover

K23A

Fig. 4

Drive table to maximum height and remove the upper and lower cover from the rear side
of the table.

Remove control console plug, unscrew K23 from panel and pull it out carefully.
Eject existing pins BB, CC and DD of the tableside K23 (not K23 connected to N10)
and move them to "X", "Y" and "Z" (page 2.6).

Take three spare wires of the cable harness between point A and B (on the lower part,
the cable harness is behind the connector plate, to have better access, the plate might
be unscrewed and the table rotated).

Fasten AMP connectors (female) to three spare wires at A and plug them into K23
pins BB, CC and DD; mark the corresponding spare wires at B:
K23.BB ---- K23A.BB
K23.CC ---- K23A.HH
K23.DD ---- K23A.KK

Eject existing pins "BB", "HH" and "KK" at both sides of K23A, mark and insulate them.
Fasten AMP connectors (male) to the spare wires at B and plug them into K23A BB,
HH and KK.

Run the control signal cable from M100 to the table and plug pins into K23A BB, HH
and KK.

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 4 of 6
TDPS 21

Siemens AG
Medical Engineering

Installation

2-5

KOORDINAT M Modifications

Eject existing pins BB, CC and DD of the tableside K23 (not K23 connected to N10)
and move them to "X", "Y" and "Z".

Run the control signal cable from M100 to the table and plug pins into K23 BB, CC
and DD.

Installation of Handswitch

Fig. 5

The handswitch has to be installed at the Control Console.


Open the Control Console of the stand and install the handswitch holder on the left side.
Remove the rubber cap of the bottom of the Control Console and connect the hand
switch plug.

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 5 of 6
TDPS 21

CARD OPTION

2-6

Installation
Overview Cabling

Handswitch
Control cable

Sto.1
Sto.2
1

Video cable

Optional video cable


X2

BB

DD CC

N10
K23

Hardcopy

OTV

Koordinat M/U

M101
M56/M501
BB
K23
M56/M501
BB
K23

DD CC

* *

DD CC

BB

change location of
existing wires:

KK HH

BB
CC
DD

X
Y
Z

Ref. 1

M503.K23A
Koordinat U only

M100
K21/1

(0V)

Ref. 2
(Biplane only)

D94.X110

(+24V)
D94.X109
K21/2

M2 (MTS)
Remote

X12

7
8
4
2

X13
5HI&RQWUROURRP

J9 (A)
Hardcopy

X11

X14

J9 (B)
Hardcopy

X21

X23

Power
Connector

X24

D2000/1

(either instead of Ref. 2, or


add. videocable and
Monitor necessary!)

D2000/2

Hicor
I/O panel

Display 2000

CARD OPTION

5HI&RQWUROURRP

(Option: add. videocable


and Monitor necessary)

X8 (M2)

RA43-030.031.09
Rev. 02
05.98

Page 6 of 6
TDPS 21

Siemens AG
Medical Engineering

Reference Image with Quantcor Option


Modifications Quantcor

3-1
3

The video signal for the Quantcor will be taken from the loop-through output X12 at the
Display 2000; during image transfer to the Quantcor, the image is put to the Hardcopy outputs 9(A,B) in N norm, therefore the reference monitor(s) will be dark for about 10s and
interferences might be visible.

Disconnect video cables


- I/O Hardcopy 9(A) --- M101 Video Print In
- Quantcor IF X3 --- M101 Video Print In

Establish video cable connections as seen in see Fig. 1.

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 1 of 2
TDPS 21

CARD OPTION

3-2

Reference Image with Quantcor Option


Overview Cabling

VSC 36

*1

x9

Video switcher

x1

x2

Quantcor IF
X2

x12

Logic &

*4

x13

PC Monitor

VSC 36.X8

Video Print
In
*5
*5

RS 232

HICOR

OTV

Frame grabber

x5

Customer PC

x6

x4

Dongle

Graphics card

x7

x3
*3

x8

RGB

x11

HC-Trigger J8

*3

BAS

OTV

Video Print
Remote

Hardcopy J9 A
Hardcopy J9 B

90 H / HM

Separation

Relay board
MON 1 J1

SIMOMED

M101

*2

Angiomatic
(Z11)
x3

Ex.RoomMon
Plane A

QUANTCOR Interface
OTV

M 101
OTV
Analog HC

Ctrl.Rm.Mon
SIMOMED HM
Option

* 1, only if VSC 36 is configured


* 2, only if VSC 36 is not configured
* 3, remove connection
* 4, connection already exists if
if VSC 36 is configured
* 5, remove connections

+DQGVZLWFK
0

7
X12
8 Remote
4
2
X13

5HIHUHQFH0RQLWRU

X11

X21

X23
'LVSOD\

5HIHUHQFH0RQLWRU

Fig. 1

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 2 of 2
TDPS 21

Siemens AG
Medical Engineering

Startup

4-1

HICOR Configuration

HICOR site configuration menu - 4


Copyright (c) 1990-1995 by Siemens Medical Systems Inc. - All Rights Reserved

NEXT MENU
PREVIOUS MENU
EXIT MENU
BEDSIDE CONTROLLER 1: ENABLED
BEDSIDE CONTROLLER 2: DISABLED
BICOR COMMUNICATION: ACS
AUTO GANTRY DOWNLOAD: ENABLED W/ACS
MICOR COMMUNICATION: DISABLED
HARDCOPY DEVICE: VIDEOMED H
TAPE DEVICE: VIDEOMED H
COLLIMATOR CONTROL: ENABLED
ARCHIVE DEVICE: T0
AUTO ARCHIVE DEFAULT: DISABLED
ACTIVE KEYS: <up arrow>, <down arrow>, <space>,

<return>

The Menu item: TAPE DEVICE and HARDCOPY DEVICE must be set to VIDEOMED H
(see Hicor Setting Instructions for entering the Configuration Menu).
Remark: If a Hardcopy N is used, an additional Scan Converter VSC36 has to be connected or the N-output of an existing Videorec can be used.

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 1 of 4
TDPS 21

CARD OPTION

4-2

Startup
SIMOMED H Configuration

1:1
1:1

The image format for 62,42KHz/100Hz (50Hz systems) and 61,38KHz/120Hz (60Hz
systems) must be set to 1:1.

SIMOMED HM Configuration

open Service Tool for Simomed HM.


click on <Configuration>.
click on <Disable scan>.
click on norm <12> and <8> , the norms are then marked, the image format 4:3 is
disabled and 1:1 is enabled.

click on <OK>.

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 2 of 4
TDPS 21

Siemens AG
Medical Engineering

Startup

4-3

SIMOMED H/HM Adjustment

Select the Hicor test image at the ECC.


Using the test image, adjust the same image geometry and sharpness as at the existing
monitors.

Adjust the maximum brightness and the maximum contrast according to the IQAP.
Adjust the brightness and the contrast according to the IQAP.
Measure the luminance with the dynamic test in field 2L and 5R and adjust the same
values with Hicor off (Bypass mode).

Function Test

Display a patient image on the live monitor(s).


Press >1 (and >2 if biplane) at the handswitch to display the image(s) on the reference
monitor(s).

Verify that image geometry, sharpness, brightness and contrast are the same on the live
and the reference monitors.

with Quantcor only:


- Select a patient image, press the <Evaluate> key at the ECC (available only if the
internal HICOR evaluation program is configured), press the <Image Trans Eval> key
-> The patient image will be transferred to the QUANTCOR PC and displayed.
. in the QUANTCOR control bar to switch the PC
- Click on the monitor symbol
image to the examination room monitor;
Note: If QUANTCOR.QCA and QUANTCOR.LVA run simultaneously, this can be
performed presently only from the program started first.
Briefly trigger radiation: the actual radiographic image must appear without delay on
the Live monitor.
Repeat procedure with DCM and Cine.
Remark: While a image transfer from Hicor to Quantcor PC is in progress, the reference monitor(s) will be dark for about 10s.

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 3 of 4
TDPS 21

CARD OPTION

4-4

Startup
This page intentionally left blank.

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 4 of 4
TDPS 21

Siemens AG
Medical Engineering

Changes to previous version

5-1

Chapter 1:
The chapter was completely revised.
Chapter 2:
The chapter was completely revised.
Chapter 3:
New content, original information moved to chapter 2.
Chapter 4:
The chapter was completely revised.

Siemens AG
Medical Engineering

RA43-030.031.09
Rev. 02
05.98

Page 1 of 2
TDPS 21

CARD OPTION

5-2

Changes to previous version


This page intentionally left blank.

CARD OPTION

RA43-030.031.09
Rev. 02
05.98

Page 2 of 2
TDPS 21

Siemens AG
Medical Engineering