CARD OPTION
Revision
RA43-030.031.09
Rev.02
05.98
Page 2 of 4
TDPS 21
Siemens AG
Medical Engineering
Contents
0-3
Page
1 _______Requirements__________________________________________________ 1 - 1
System Requirements . . . . . . . . . . . . . . . .
Required Documents . . . . . . . . . . . . . . . .
Required Test Equipment and Meters . . . . . . . .
Parts List . . . . . . . . . . . . . . . . . . . . . .
Safety Information . . . . . . . . . . . . . . . . . .
Protection of Electrostatically-sensitive Components
Startup Information . . . . . . . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.1-1
.1- 1
.1-1
.1- 1
.1-2
.1-2
.1-2
2 _______Installation ____________________________________________________ 2 - 1
Programming of DISPLAY 2000 . .
Installation of Reference Monitors .
M100 Modifications . . . . . . . .
KOORDINAT U Modifications . . .
KOORDINAT M Modifications . . .
Installation of Handswitch . . . . .
Overview Cabling . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.2-1
.2- 3
.2-3
.2- 4
.2- 5
.2- 5
.2-6
4 _______Startup _______________________________________________________ 4 - 1
HICOR Configuration . . . .
SIMOMED H Configuration .
SIMOMED HM Configuration
SIMOMED H/HM Adjustment
Function Test . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.4- 1
.4- 2
.4- 2
.4- 3
.4 - 3
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 3 of 4
TDPS 21
CARD OPTION
0-4
Contents
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 4 of 4
TDPS 21
Siemens AG
Medical Engineering
Requirements
1-1
System Requirements
Required Documents
Parts List
47 96 939
Display 2000
27 98 671
Monitor Simomed HM 44 cm or
37 92 079
Monitor Simomed HM 54 cm or
37 92 087
Monitor Simomed H 44 cm or
11 63 000
Monitor Simomed H 54 cm
11 63 018
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 1 of 2
TDPS 21
CARD OPTION
1-2
Requirements
Safety Information
The "Safety Information" in the system manual and in the particular unit instructions must
be observed.
Integrated circuits and circuit boards on which MOS components are installed call for particularly careful handling because of their highly sensitive electrical structures and
because of their extremely high input resistance.
The ESD regulations that apply for electrostatically-sensitive components must be
observed.
Startup Information
If any work steps are carried out on the system, power must first be switched off.
The I.I. TV channel setting must be optimal.
The SIMOMED HM monitor(s) (ambient light sensor, standard changing time) must be
adjusted according to the adjustment instruction on service floppy
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 2 of 2
TDPS 21
Siemens AG
Medical Engineering
Installation
2-1
Note: The Display 2000 is composed of two identical, separate channels and contains two
identical circuit boards, one for each plane.
Remove the housing cover and install the jumpers as shown in the table (Fig. 1).
Close the housing.
Jumper
Position
Remark
J1
50/60 Hz
J2
OPEN
LIH
J3
INT
internal synchronization
J5
CLOSED
NORM
50Hz
60Hz
J3
Ext/Int
J1
50/60Hz
Component side
J5
J2
Fig. 1
75 Ohm switches on the front of the D2000: The three 75 Ohm switches belong to the
two outputs X12 and X22 and to input X30. If a socket is used, the corresponding switch
must be in the OFF position, if a socket is not used, the switch must be in the ON
position.
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 1 of 6
TDPS 21
CARD OPTION
2-2
Installation
Installation of Display 2000
The Display 2000 has to be installed in the M2 of the MTS.
Connect triax cable from Reference Monitor 1 to X13 at the Display 2000 using the
angled BNC connector (page 2-6).
For biplane: connect triax cable from Reference Monitor 2 to X23 at the Display 2000
using the angled BNC connector.
Pull a triax cable from the Hicor to the M2; unplug existing triax cable at J9 A and
establish new connection J9 A ---- X11 Display 2000.
For biplane: pull a triax cable from the Hicor to the M2; unplug existing triax cable at J9 B
and establish new connection J9 B ---- X21 Display 2000 (or: pull the triax cable from the
M2 to the control room and connect it to Display 2000 X14 and the 2nd Ref Monitor; Ref 1
images can then also be seen in the control room).
Pull a triax cable from the control room M101 to the M2 and connect it to Display 2000
X12; unplug the existing Hardcopy triax cable in the M101coming from J9 Hicor and
connect the new triax cable instead.
Pull the control signal cable from the M100 to the M2 and connect it at Display 2000
Remote.
M2
M2
X8,9
Display 2000
M4x8
Fig. 2
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 2 of 6
TDPS 21
Siemens AG
Medical Engineering
Installation
2-3
Install the monitor(s) in the examination room according to the Installation Instructions
for the monitor suspension system.
M100 Modifications
Install the sockets K21/1 and K21/2 in the frame of the M100 and connect the attached
wires to X109 and X110 (Fig. 3).
Diamentor
X109
X110
Z96
D94
Backplane
K21/1
Table
Display 2000
D95
K21/2
M100
Fig. 3
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 3 of 6
TDPS 21
CARD OPTION
2-4
Installation
KOORDINAT U Modifications
A
K23
Table
Control
Cover
K23A
Fig. 4
Drive table to maximum height and remove the upper and lower cover from the rear side
of the table.
Remove control console plug, unscrew K23 from panel and pull it out carefully.
Eject existing pins BB, CC and DD of the tableside K23 (not K23 connected to N10)
and move them to "X", "Y" and "Z" (page 2.6).
Take three spare wires of the cable harness between point A and B (on the lower part,
the cable harness is behind the connector plate, to have better access, the plate might
be unscrewed and the table rotated).
Fasten AMP connectors (female) to three spare wires at A and plug them into K23
pins BB, CC and DD; mark the corresponding spare wires at B:
K23.BB ---- K23A.BB
K23.CC ---- K23A.HH
K23.DD ---- K23A.KK
Eject existing pins "BB", "HH" and "KK" at both sides of K23A, mark and insulate them.
Fasten AMP connectors (male) to the spare wires at B and plug them into K23A BB,
HH and KK.
Run the control signal cable from M100 to the table and plug pins into K23A BB, HH
and KK.
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 4 of 6
TDPS 21
Siemens AG
Medical Engineering
Installation
2-5
KOORDINAT M Modifications
Eject existing pins BB, CC and DD of the tableside K23 (not K23 connected to N10)
and move them to "X", "Y" and "Z".
Run the control signal cable from M100 to the table and plug pins into K23 BB, CC
and DD.
Installation of Handswitch
Fig. 5
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 5 of 6
TDPS 21
CARD OPTION
2-6
Installation
Overview Cabling
Handswitch
Control cable
Sto.1
Sto.2
1
Video cable
BB
DD CC
N10
K23
Hardcopy
OTV
Koordinat M/U
M101
M56/M501
BB
K23
M56/M501
BB
K23
DD CC
* *
DD CC
BB
change location of
existing wires:
KK HH
BB
CC
DD
X
Y
Z
Ref. 1
M503.K23A
Koordinat U only
M100
K21/1
(0V)
Ref. 2
(Biplane only)
D94.X110
(+24V)
D94.X109
K21/2
M2 (MTS)
Remote
X12
7
8
4
2
X13
5HI&RQWUROURRP
J9 (A)
Hardcopy
X11
X14
J9 (B)
Hardcopy
X21
X23
Power
Connector
X24
D2000/1
D2000/2
Hicor
I/O panel
Display 2000
CARD OPTION
5HI&RQWUROURRP
X8 (M2)
RA43-030.031.09
Rev. 02
05.98
Page 6 of 6
TDPS 21
Siemens AG
Medical Engineering
3-1
3
The video signal for the Quantcor will be taken from the loop-through output X12 at the
Display 2000; during image transfer to the Quantcor, the image is put to the Hardcopy outputs 9(A,B) in N norm, therefore the reference monitor(s) will be dark for about 10s and
interferences might be visible.
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 1 of 2
TDPS 21
CARD OPTION
3-2
VSC 36
*1
x9
Video switcher
x1
x2
Quantcor IF
X2
x12
Logic &
*4
x13
PC Monitor
VSC 36.X8
Video Print
In
*5
*5
RS 232
HICOR
OTV
Frame grabber
x5
Customer PC
x6
x4
Dongle
Graphics card
x7
x3
*3
x8
RGB
x11
HC-Trigger J8
*3
BAS
OTV
Video Print
Remote
Hardcopy J9 A
Hardcopy J9 B
90 H / HM
Separation
Relay board
MON 1 J1
SIMOMED
M101
*2
Angiomatic
(Z11)
x3
Ex.RoomMon
Plane A
QUANTCOR Interface
OTV
M 101
OTV
Analog HC
Ctrl.Rm.Mon
SIMOMED HM
Option
+DQGVZLWFK
0
7
X12
8 Remote
4
2
X13
5HIHUHQFH0RQLWRU
X11
X21
X23
'LVSOD\
5HIHUHQFH0RQLWRU
Fig. 1
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 2 of 2
TDPS 21
Siemens AG
Medical Engineering
Startup
4-1
HICOR Configuration
NEXT MENU
PREVIOUS MENU
EXIT MENU
BEDSIDE CONTROLLER 1: ENABLED
BEDSIDE CONTROLLER 2: DISABLED
BICOR COMMUNICATION: ACS
AUTO GANTRY DOWNLOAD: ENABLED W/ACS
MICOR COMMUNICATION: DISABLED
HARDCOPY DEVICE: VIDEOMED H
TAPE DEVICE: VIDEOMED H
COLLIMATOR CONTROL: ENABLED
ARCHIVE DEVICE: T0
AUTO ARCHIVE DEFAULT: DISABLED
ACTIVE KEYS: <up arrow>, <down arrow>, <space>,
<return>
The Menu item: TAPE DEVICE and HARDCOPY DEVICE must be set to VIDEOMED H
(see Hicor Setting Instructions for entering the Configuration Menu).
Remark: If a Hardcopy N is used, an additional Scan Converter VSC36 has to be connected or the N-output of an existing Videorec can be used.
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 1 of 4
TDPS 21
CARD OPTION
4-2
Startup
SIMOMED H Configuration
1:1
1:1
The image format for 62,42KHz/100Hz (50Hz systems) and 61,38KHz/120Hz (60Hz
systems) must be set to 1:1.
SIMOMED HM Configuration
click on <OK>.
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 2 of 4
TDPS 21
Siemens AG
Medical Engineering
Startup
4-3
Adjust the maximum brightness and the maximum contrast according to the IQAP.
Adjust the brightness and the contrast according to the IQAP.
Measure the luminance with the dynamic test in field 2L and 5R and adjust the same
values with Hicor off (Bypass mode).
Function Test
Verify that image geometry, sharpness, brightness and contrast are the same on the live
and the reference monitors.
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 3 of 4
TDPS 21
CARD OPTION
4-4
Startup
This page intentionally left blank.
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 4 of 4
TDPS 21
Siemens AG
Medical Engineering
5-1
Chapter 1:
The chapter was completely revised.
Chapter 2:
The chapter was completely revised.
Chapter 3:
New content, original information moved to chapter 2.
Chapter 4:
The chapter was completely revised.
Siemens AG
Medical Engineering
RA43-030.031.09
Rev. 02
05.98
Page 1 of 2
TDPS 21
CARD OPTION
5-2
CARD OPTION
RA43-030.031.09
Rev. 02
05.98
Page 2 of 2
TDPS 21
Siemens AG
Medical Engineering