Professional Documents
Culture Documents
REVIEW
1. The __________ is responsible for examination equipment calibration records.
Manufacturer. T-130
2. At least ______ copy (ies) of an NDE procedure must be available to the
manufacturers NDE personnel at the worksite.
One. T-150
3. The difference between an inspection and examination per section V of the ASME
code is:
Examinations are QC functions performed by the Manufactures employees.
Inspections are performed by the authorized Inspector. T-170
4. How must all NDE personnel be qualified?
Per the requirements of the referencing code or standard. T-140 (a)
5. What is not part of the minimum content of a written radiographic procedure?
Type of emulsion used. T-221.1
6. How can compliance with a written radiographic procedure be demonstrated?
By verifying the proper density and demonstrating the IQI image
requirements. T-221.2
7. What finished surface is required of butt welds?
They must not have any irregularities or contours that will mask defects or
interfere with interpretation. T-222.2
8. One reason for a system of radiographic film identification is:
So the location of defects will be traceable to the weldment. T-224
9. Intensifying screens may be used _______________.
When performing RT in accordance with ASME V, Article 2. T-232
10. IQIs may be what types?
Wire & Hole types. T-233
11. Viewing facilities for radiographs shall ______________.
Have subdued background lighting that will not cause glare on the film. T-234
12. The direction of the central beam of radiation should be ____________ the area of
interest whenever practical.
Centered on. T-273
13. Where are hole type IQIs place on the radiographing welds?
Adjacent to or on the weld but not in the area of interest. T-277
14. Where are location markers placed if they are to appear as a radiographic image?
On the part to be radiographed. T-275
15. The IQI is normally placed on which side of a part?
The source side except when inaccessibility prevents hand placement on that
side. T-277.1 (a)
16. What designation is used to indicate the IQI is on the film side?
A lead letter F placed next to or on the IQI. T-277.1 (b)
17. How many IQIs should appear on each radiograph?
It depends on the configuration used to set up the shot (panoramic multi-film,
single shots, etc.). T-277.2
ASME V
REVIEW
18. Shims may be placed under the IQIs to stimulate weld reinforcement to assure the
density in the area of interest is no more than ______ lighter than the density through
the IQI.
15%. T-277.3
19. What blemishes are permitted on film as long as they do not interfere with
interpretation and do not mask or become confused with discontinuities in the area of
interest.
Fogging & false indications from defective screens, scratches, crimps, static
marks & dirtiness, processing defects such as streaks & water marks. T-281
20. The minimum density requirements for single film viewing are:
2.0 for film made with gamma ray and 1.8 for film made with an X-ray
machine. T-282
21. When using only one IQI per film density, measurements may vary through the area
of interest by no more than ________ or additional IQIs will be required.
Minus 15% or plus 30%. T-282
22. What is an essential indication of sensitivity for image quality of a radiograph?
Display of the specified hole or wire of the IQI used. T-283
23. Excessive backscatter is indicated by:
A light image of the lead letter B on a dark background. T-284
24. A tank is built with plate under 2 thick. The geometric unsharpness of the
radiographs shall not exceed:
0.020, T-285
25. Information that may not be included in the documentation accompanying the
radiographs:
Development time & exact shim material specifications. T-291
26. When radiographs are presented to the authorized inspector:
The manufacturer will have interpreted all the film and will have indicated the
disposition of each on the report and will have also included all the other
information required on the report. T-292
27. Written UT procedures do not include?
Search unit configuration or density attenuation. T-522
28. Section V, article 5 specifically states __________ unless equal or better sensitivity
can be demonstrated using other frequencies:
Examinations shall be conducted with a pulse-echo UT instrument capable of
generating frequencies over a range of at least 1 MHz to 5MHz. T-531
29. As a minimum UT equipment must be calibrated:
At the beginning and end of each examination, anytime that malfunctioning is
suspected and when examination personnel are changed. T-534
30. For thickness measurements the frequency required to be used ____________.
May be any frequency range capable of resolving the thickness range to be
measured. T-544.1 (a)
31. UT examination report shall include:
The locations of the thickness readings taken and the identification of the
operator who carried out the examination. T-591
32. UT equipment calibration records shall include:
ASME V
REVIEW
ASME V
REVIEW