Professional Documents
Culture Documents
MANUAL
Prevacuum Table - top Autoclave
Models 2340, 2540 EPV
TECHNICIAN
MANUAL
Prevacuum Table - top Autoclave
Models 2340, 2540 EPV
TABLE OF CONTENTS
PARAGRAPH
PAGE NO.
1.
INTRODUCTION
2.
SYMBOL DESCRIPTION
3.
INSTALLATION INSTRUCTIONS
3.1
4.
5.
6.
7.
Connections
TESTS
4.1.
Installation Tests
4.2
Periodical Tests
WATER QUALITY
5.1
5.2
Reverse Osmosis
6.1
6.2
10
6.3
AC - Board - AC-T1
11
12
7.1
Method of Calibration
12
7.2
13
7.3
15
8.
TEST POINTS
16
9.
CHANGING PARAMETERS
17
9.1
General
17
9.2
Changing Parameters
17
9.3
23
9.4
Test Mode
23
10.
REPLACEMENT PROCEDURES
24
10.1
24
10.2
25
10.3
26
10.4
26
10.5
28
10.6
29
10.7
30
10.8
31
10.9
32
11.
PAGE NO.
33
34
35
36
37
38
39
41
TROUBLESHOOTING
42
11.1
42
11.2
Pre-process malfunction
43
11.3
In Process Malfunction
49
11.4
Components Malfunction
54
11.5
Mechanical malfunction
67
12.
69
13.
VALVES NUMBERING
79
PAGE NO.
Rear View
DIG- T2 Board
ANL T2 Board
10
AC-T1 Board
11
Electronic Box
27
Water Reservoir
40
Front View
74
Vessel Assembly
75
76
77
78
Piping Drawing
81
Electrical Drawing
82
1.
INTRODUCTION
This manual, together with the operators manual, forms the complete edition
of the Operation and Maintenance instructions. This manual is intended for the
use of the technician. It is forbidden for unqualified and unauthorized
personnel to service the autoclave in accordance with the instructions in this
manual. Any unauthorized service may result in the invalidation of the
manufacturers guarantee.
The qualified technician shall be an authorized electrician with the right
qualifications in electronics and shall be familiar with the local
technical/electrical regulations.
2.
SYMBOL DESCRIPTION
Caution! Consult accompanying documents
Stand by
3.
INSTALLATION INSTRUCTIONS
Network and connection should comply with the device consumption. It must
comply with local installation and safety rules and regulations.
The voltage supplied to the device must comply with the label + 5%.
In order to avoid any injury by electrical hazard, it is mandatory for the
customer to have installed an earth leakage relay in the electrical board to
which the autoclave is connected.
This relay disconnects all the poles of the electric power line in case of
accidental contact with the instrument metal enclosure, by the operator or
another person, leading to a dangerous leakage current.
Set the hind legs of the autoclave to give the body an inclination of 2 towards
the front; this ensures that water is completely drained out of the chamber
through the opening on the bottom in front of the chamber. To check pour a
glass of water into the bottom of the autoclave ensuring water flows out of the
chamber.
Ensure while placing the autoclave, to leave space around the machine, to give
the technician access to service the machine also making sure the autoclave is
on a leveled and rigid surface.
Ensure the following connections to the utilities sources and drain (refer to the
drawing REAR VIEW below).
3.1
Connections
Connect the exhaust from pump to the drain by 5/16 tube.
Note: All the connections listed above will be executed with flexible
hoses, which are supplied with the autoclave.
Connect the power cord of the autoclave to a power outlet, 230V-16A.
REAR VIEW
4.
TESTS
4.1. Installation Tests
The service technician shall perform the following
preliminary checks before operating the autoclave:
a.
Leveling Check
Check that the autoclave is leveled.
b.
c.
Continuity Check
Check the continuity of the grounding connection.
d.
Safety Check
Check the safety elements; safety valve and the door locking
mechanisms.
e.
Programs Check
Run basic programs of the autoclave and check the operation
sequences, the sterilization parameters etc.
f.
Integrity Check
Perform a visual check to verify that there are no dents,
scratches, broken gauges, etc.
After the above steps are performed, the autoclave is ready for operation.
4.2
Periodical Tests
TEST
PERIOD
5.
WATER QUALITY
5.1
< 15 mg/l
Silica
< 2 mg/l
Iron
< 0.2mg/l
Cadmium
Lead
Chloride
< 3 mg/l
Phosphate
Conductivity
< 50 s/cm
pH
6.5 to 8
Appearance
Hardness
Attention:
The use of water that do not comply with the table above may have
severe impact on the working life of the sterilizer and can invalidate
the manufacturers guarantee.
5.2
Reverse Osmosis
A Reverse Osmosis system may be used to improve the quality of the
water used to generate steam in the autoclave chamber. The use of
mineral free will contribute to better performance and longer life of the
autoclave.
PLUG
SK T
20
19
JP1
19
JP1
P1
DIG-T2 BOARD
+ C2
37
20
37
U12
JP5
12
JP7
JP6
R3
6.
R2
C9
JP2
U8
U16
R1
C5
C7
U11
U14
U3
U18
C4
JP3
C3+
U21
RP1
JP9
1
Q1
Y1
JP10
1
BZ1
U15
U17
DIG
T3
97-05 C25 C24
1
JP4
U4 R4
C16
C17
JP3 C23
JP12
R9
U22 D1
C15
U19
U1
C6
C22
D S1
2
1
JP8
1
34
33
20
37
JP1
1
R8
19
R7
C1
C1
U1
C1
U1
+C2
R5
R6
R1
R1
C1 U1
JP
C1
C8
R3
RP
C9
U2
C2
1
C1
1 U2
U7
R2
C2
C1 R1
U5
U6
U8
U1
R1
C5
C7
U1
U1
U3
C3
U2
1
U4 R4
C1
C1 6
RP
Q1
Y1
BZ1
U1
U1
H2
U1
C4
D IG
C2
97-05
JP3
C2
U2
C2
C2
R9
D1 C1
U1
C6
U1
2
1
DS
34
33
20
19
JP1
P1
12 1
JP7
JP2
JP5
JP6
JP9
JP4
JP3
1
6.2
JP12
JP10
1
JP8
10
JP11
1
AC - Board - AC-T1
This board receives command signals from ANL-T2 board designated
to activate AC devices. It converts them to AC mode.
The AC-T1 board provides AC filtering for the power supply to the
control system and protection from sharp current fluctuations.
The AC-T1 board includes:
AC drivers for the AC devices (pump, heaters, etc.) up to 6A.
AC - input connector JP3.
Filter circuit based on L1C5, the power transfer from the filter to
power supply via JP2.
Spike protection - RV13.
Three - 230/115 VAC output channels, triac control circuit.
Q1 - 16Amp - JP4/1,2
Q2 - 8 Amp - JP7/1
Q3 - 8 Amp - JP7/2
AC-T1 BOARD
(JP3,JP4,JP7 ON THE BOTTOM SIDE OF THE BOARD)
JP5
JP1
R5
R9
RV3
1
R1
JP2
RV2
6.3
U3
R12
C3
R10
R11
U2
R6
R8 C2
R7
U1
R4
R2
C1 R3
JP6
JP7
Q3
JP4
L1
JP3
1
Q2
Q1
C5
ETYPEAC-97
CAT TECH.
11
RV1
Method of Calibration:
The calibration of temperature and pressure is performed digitally.
The temperature and pressure measuring circuits are designed with
components having 1% accuracy.
The temperature circuit is linear and has an output of 100mV2400mV
for a temperature range 20C150C.
The pressure circuit is also linear and has an output of
100mV2400mV for a pressure range 0400 kPa.
The measuring at the A/D is limited for values higher than 2400mV or
below 100mV.
It is obvious that although the accuracy of the circuit components is
1%, the sum deviation may reach 5%. Therefore calibration is
necessary.
The system has a non-volatile memory in which the offset and gain
data of the sensors are stored.
Calibration is performed by use of the keyboard (or the communication
port for computer use).
Following is a description of the calibration of temperature and
pressure using the keyboard. The calibration procedure is identical to
all autoclave models.
Definition of Gain and Offset:
Every straight line can be defined by the equation ax+b=y. a is the
deviation from y=0 and b is the slope of the line. In our case a=Gain
and b=Offset.
The system is based on calculation of offset and gain of two points as
in the following example
If the actual measured pressures are 100 kPa & 300 kPa and the
displayed temperatures are 90 kPa & 310 kPa respectively, the input
Vs pressure graph will be as follows.
READ
Y
310
300
ACTUAL
PRESURE
7.
100
90
X
V1
V2
INPUT FROM SENSOR
The calibration method enables to input this data into the system in
order to perform automatic correction of OFFSET and GAIN.
It is necessary to define the two points, which have to be input to the
system, prior to performing calibration.
12
7.2
key.
key
R 60.0
key.
The cursor on the upper row will move to R60.0, change to R68.0
by means of UP/DN
keys.
Press the PROGRAM
key.
key.
13
Remarks:
When cursor is set under any figure, pressing the START/STOP
key stops calibration.
When the cursor is set under any figure, it enables to feed this
value to the reading of temperature by pressing the
PARAMETERS key.
Example:
If the cursor is set under R68.0 on the upper row and the actual
temperature of the autoclave is now 66.0C, by pressing the
PARAMETERS key, the temperature reading will be corrected from
R68.0 to R 66.0.
This calibration method enables the technician to change a certain point
without affecting a second point.
Example:
When the working temperatures are 134C and 121C but the actual
temperature are 132C and 121C
The measured temperature for 134C must be corrected without
changing the temperature at 121C,
Considering the above example, the setting of data can be done as
follows:
A 121.0
R 121.0
A 132.0
R 134.0
The system will calibrate the new offset and gain and retain them into
the non-erasable memory.
As a result, while running a 134C program the correction is
automatically made, while at 121C no change is felt.
14
7.3
key.
key.
key again.
The system calculates the new values for offset and gain and stores
these values in the non-volatile memory.
This completes the calibration procedure for the pressure.
15
8.
TEST POINTS
The test point list provides testing points on the junctions on board to assist in
locating the malfunction.
NU TP
FUNCTION
VALVE
TP1
GND
TP2
+5V DC
TP3
+12V DC
TP4
OUT - HEATERS
ON-5V; OFF-0V
TP5
OUT-VAC. PUMP
ON-5V; OFF-0V
TP6
TP7
OUT-AIR-RELEASE VALVE
ON-5V; OFF-0V
TP8
OUT-EXH-VALVE
ON-5V; OFF-0V
TP9
OUT-WATER TO RES.
ON-5V; OFF-0V
TP10
OUT-VACUUM VALVE
ON-5V;OFF-0V
TP11
OUTDOOR LOCK
ON-5V; OFF-0V
TP12
OUT-AIR RELEASE
ON-5V; OFF-0V
TP13
TP14
TP15
IN FLOAT HIGH.
ON-5V; OFF-0V
TP16
IN DOOR SWITCH
ON-5V; OFF-0V
TP17
IN THERMOSTAT
ON-5V; OFF-0V
TP 18
TP19
TP20
TP21
IN TEMPERATURE
134C 1.97V
TP22
TP23
TP24
IN PRESSURE
2.0B 2.031V
TP25
TP26
IN ELECTRODE HIGH
0V-YES; 2.5V NO
16
9.
General
The software for the control system of the EPV autoclave is EpvEn320.
The software contains a table of parameters of which part of them
defines the autoclave, and part defines the processes in the autoclave.
This document describes parameters and their task in the software.
9.2
Changing Parameters
Access to the program is achieved by inserting the access code.
Access code 1
This access code enables change of the sterilization
time, sterilization temperature and the drying time.
Access code 11
changes that may be done by the supervisor.
Enables to change a few parameters
Access code 13
changes that may be done by the technician.
Enables to change all the parameters
To change the parameters listed below, proceed as follows:
a.
b.
c.
d.
e.
The pre-set parameters values will replace the last default values and
become the updated default values.
9.2.1
9.2.2
Cycle
105C
137C
8
0
Cycle
1
1 minute
3 minutes
59 minutes
Value 10
15
15
18
30
3.5
17
9.2.3.
9.2.4
9.2.5
1
1 minute
minutes
99 minutes
Cycle
Value
10
30
30
9.2.6
Cycle
Value
Heat T.O. - The maximum time allowed for the heating stage
(testing)
After this time, if the system does not enter the sterilization stage,
it will receive a Low Heat message and the process will abort.
Entry Code
11
Resolution
1min
Minimum Value 20min
Maximum Value 53 min
Set value
Cycle
18
8
0
9.2.7
9.2.8
Cycle
6
95
120 134
9.2.9
11
1
1
5
Cycle
Value
9.2.10
9.2.11
9.2.12
9.2.13
11
Ex Mode
Fast Ex
1
N.A.
2
Slow Ex (Waste)
3
Slow Ex (Liquids)
4
Set value
9.2.14
Cycle
Value
9.2.15
20
9.2.16
Req Prs+
This parameter defines the required addition to the sterilization
pressure in kpa.
For example, for a sterilization temperature of 121C the required
pressure is 204kpa. Since the system controls the sterilization
process according to pressure and temperature, if Req Prs+ equals
, the system will maintain the pressure at 204kpa. If the value
is at 5kpa, the system will be maintained at 209kpa, and so on.
Entry Code
11
Resolution
1
Minimum Value 0kpa
Maximum Value 20kpa
Set value
Cycle
Value
Cycle
Value
1`
21
22
9.3
9.4
Test Mode
Turn on the main power switch, while pressing the UP key
continuously, to enter the Test mode.
The output test will start (valves, heater, pump), each output will be
activated for one second and its name will be displayed. To switch
from one output to another, press each time the UP key.
Keep pressing the UP key, the inputs test is performed, pressing each
time the UP key the next input is selected.
The same way the temperature and pressure values are displayed.
23
10.
REPLACEMENT PROCEDURES
10.1
2.
3.
Dismantle the air filter from the service opening cover (3).
4.
5.
6.
3
1
2
24
10.2
2.
3.
Unscrew the safety valve and remove it from the safety valve base.
4.
5.
Perform one cycle and verify that the valve operates correctly.
Safety valve
25
10.3
10.4
1.
2.
Remove the screws that connected the plastic panel with the digital
electronic board to the autoclave (2).
3.
4.
5.
6.
7.
8.
2.
Remove two screws from the bottom side and from the top (1, 3).
3.
4.
5.
6.
Test any cycle and verify that the autoclave operates as required.
26
6
Electronic box
Control panel
Electronic box
27
10.5
2.
3.
4.
5.
6.
7.
8.
9.
28
10.6
2.
3.
29
10.7
2.
3.
30
10.8
Replacing
Caution
Before starting, disconnect the instrument from the power
source and ensure that there is no pressure in the
autoclave.
Allow the autoclave to cool before removing outer covers.
10.8.2
1.
2.
3.
4.
5.
6.
Cleaning
1.
2.
3.
31
10.9
2.
3.
Remove the nut (3) and the ring for drain valve (2).
3.
4.
5.
CMT240-0020
CMT240-0003
VLV170-0066
Item
Cat No.
SRV000-0224
SRV000-0232
32
2.
3.
Install the new pressure gauge using Teflon tape for sealing the
thread.
4.
5.
33
2.
3.
4.
5.
6.
Operate the autoclave and verify that the pin is retracted when the
cycle reaches stand-by stage.
34
3.
4.
Unscrew the two screws (2) fastening the printer to the panel.
5.
6.
7.
35
2.
Disconnect the wires (1), (2) from the door switch (3).
3.
4.
Reconnect the wires the microswitch. Verify that the wire is placed
on the isolating cover (4) and does not touch the chamber.
5.
6.
1
3
36
2.
3.
Remove the four screws connecting the circuit breaker to the panel
(1).
4.
5.
6.
7.
8.
37
38
description
The water reservoir is a one-compartment reservoir that
supplies water to the chamber to generate steam. The
reservoir is equipped with two outlets at the bottom. One
outlet is connected to the chamber to supply water to
generate steam. The second outlet is connected to the drain
vale to enable draining the reservoir. The reservoir must be
supplied with water conforming to requirements specified in
paragraph 5.1 (Water for Generating Steam).
A float switch is mounted on the wall of the reservoir. The
float stops the cycle when the water level drops below the
float, to avoid working without water.
Flexible tubing connects the aeration inlet of the vacuum
pump to the upper side of the pump reservoir.
10.16.2
Replacement
It is possible to replace damaged items on the reservoir or to
replace the complete assembled reservoir. Since the items on
the reservoir tend to deteriorate after a long period, it is
recommended to replace the complete assembly.
Caution!
Before starting, disconnect the instrument from the
power source.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
39
10.1
Water reservoir
40
3.
4.
1
2
ceases.
41
42
11.
TROUBLESHOOTING
11.1
Para.
11.3.1
Para.
11.3.2
Low Temp.
Low Heat
4, 5
High Temp.
Para.
11.3.3
12, 16
1
1
High Pres.
1
7
Para.
11.5
6, 7, 16,
1
19, 21
6, 7, 16,
19, 20, 3
21
11,16,
1
18, 19
3, 25
Door Unlock
Low Water
Para.
11.4
15, 20
Low Pres.
Add Water
Para.
11.3.4
24
1, 6, 7,
22, 25, 1, 2
26, 27
42
43
1.
Symptom
1.5
1.4
1.3
1.2
1.1
TROUBLESHOOTING
11.2 Pre-process malfunction
11.
1.4
Fasten contacts.
1.3
1.2
1.1
Corrections
44
1.
Symptom
1.6.1
1.6
TROUBLESHOOTING (cont.)
11.2
Pre-process malfunction (cont.)
(cont.)
11.
Fix connections.
Corrections
45
1.
Symptom
1.6.2
TROUBLESHOOTING (cont.)
11.2
Pre-process malfunction (cont.)
(cont.)
11.
Corrections
46
1.
Symptom
1.6.4
1.6.3
TROUBLESHOOTING (cont.)
11.2
Pre-process malfunction (cont.)
(cont.)
11.
1.6.4
Corrections
47
3.
4.
5.
(cont.)
Symptom
5.2
5.1
4,1
4.1
4.2
3.1
2.1
2.2
TROUBLESHOOTING (cont.)
11.2
Pre-process malfunction (cont.)
2.
1.
11.
5.2
Reconnect cable.
Replace the keyboard.
Replace the DIG-T2 board.
4.1
4.1
4.2
5.1
Reconnect cable.
Replace the DIG-T2 board
3.1
2.1
2.2
Corrections
48
7.
Symptom
7.1
6.1
TROUBLESHOOTING (cont.)
11.2 Pre-process malfunction (cont.)
6.
11.
7.1
6.1
Corrections
49
3.
4.
Symptom
4.1
4.2
3.1
2.2
2.1
1.
TROUBLESHOOTING (cont.)
11.3 In Process Malfunction
11.3.1 Water Inlet Stage
2.
1.
11.
Piping is blocked
The water inlet valve is faulty.
2.2
3.1
4.1
4.2
2.1
1.1
Corrections
50
Symptom
TROUBLESHOOTING (cont.)
11.3 In Process Malfunction (cont.)
11.3.1 Water Inlet Stage (cont.)
4.7
5.
Water does not enter the
If no water enters the chamber, there is no water
chamber during FILL WATER in the reservoir and display indicates that there
stage.
is water:
4.1
Float is stuck. Check if the cooling coil
has moved and prevents the float from
moving or if float is stuck due to dirt.
4.2
Float is faulty.
There is water in the reservoir and the display
indicates that there is water. Despite this, water
does not enter the chamber.
4.3
No voltage at solenoid input.
Verify command at Test Point TP9 from
electronic box.
4.4
solenoid faulty.
4.5
Water inlet pipe is clogged.
4.6
Water inlet valve (21) is faulty.
11.
Replace float.
Fix disconnection.
Replace plunger or complete solenoid.
Clean water inlet pipe
See symptom s 3 & 6 in para. 11.4.
Replace water inlet valve.
Replace plunger or complete solenoid.
4.2
4.3
4.4
4.5
4.6
4.7
4.1
Corrections
51
2.
3.
Symptom
2.1
3.1
3.4
3.3
3.2
1.1
TROUBLESHOOTING (cont.)
11.3 In Process Malfunction (cont.)
11.3.2 Heating Stage
1.
11.
3.1
3.2
3.3
3.4
2.1
1.1
Corrections
52
5.
4.
Symptom
4.2
4.1
TROUBLESHOOTING (cont.)
11.3 In Process Malfunction (cont.)
11.3.2 Heating Stage (cont.)
11.
5.1
5.2
5.3
4.2
Corrections
53
1.
2.
1.
Symptom
Symptom
1.1
1.2
2.1
5.4
5.2
5.2
5.1
TROUBLESHOOTING (cont.)
11.3 In Process Malfunction (cont.)
11.3.3 Sterilization stage
11.
5.2
5.3
1.1
1.2
Replace fuse.
increase temperature adjustment by
turning the potentiometer screw.
Corrections
5.1
2.1
5.1
Corrections
54
1.
Symptom
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
1.4
1.5
1.2
1.3
1.1
Corrections
55
3.
2.
Symptom
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
3.3
2.3
2.2
3.1
3.2
2.1
Corrections
56
(cont.)
4.
Symptom
Check the solenoid coil input voltage
with the solenoid connected. remove the
cover that protects the terminals of the
coil. If the valve is open despite the
voltage being less than 2VDC, there is a
mechanical problem.
If all items checked above conform to
requirements, ANL-T1 board is faulty.
3.4
3.3
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
3.
11.
4.2
4.3
4.4
3.4
4.1
3.3
Corrections
57
Symptom
6.
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
6.3
6.4
6.5
5.2
6.1
6.2
5.1
Corrections
58
8.
7.
Symptom
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
7.5
7.4
8.1
7.3
7.1
7.2
Corrections
59
(cont.)
9.
Symptom
Check the connection between connector
JP5/6 and the valve.
Check the solenoid coil input voltage
while the solenoid is connected to the
valve. If the valve is closed, despite the
voltage being higher than 10VDC, there
is a mechanical problem with the valve.
If all items checked above conform to
requirements, ANL-T1 board is faulty.
8.4
8.3
8.2
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
8.
11.
8.4
9.1.2
9.1.3
8.3
9.1.1
8.2
Corrections
60
(cont.)
Symptom
Check the connection between JP5/7 on
ANL-T1 board to the locking solenoid.
Check if the cable is not damaged.
Check the solenoid input voltage, without
intermitting the solenoid. if the voltage is
higher than 10VDC, the problem is the
solenoid or a mechanical problem,
preventing the locking pin from moving
freely inside the solenoid.
If all items checked above conform to
requirements, ANL-T1 board is faulty.
9.4
9.3
9.2
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
9.
11.
10.2
10.3
9.4
9.3
10.1
9.2
Corrections
61
12.4
12.3
12.2
12.1
Symptom
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
12.4
12.3
12.2
11.3
11.2
12.1
11.1
Corrections
62
14.1
15.3
15.4
15.2
15.1
14.2
13.1
13.2
Symptom
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
15.2
15.4
15.3
15.1
14.2
13.2
14.1
13.1
Corrections
63
Symptom
18.1
19.1
18.4
18.2
18.3
17.1
17.2
16.4
16.3
16.1
16.2
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
replace ANT-T1board.
16.4
18.4
18.2
18.3
19.1
18.1
17.2
16.3
17.1
16.1
16.2
Corrections
64
Symptom
23.1
23.2
23.3
22.1
22.2
22.3
21.3
21.4
21.1
21.2
20.4
20.2
20.3
20.1
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
20.1
20.2
20.3
20.4
21.1
21.2
21.3
21.4
22.1
22.2
22.3
23.1
23.2
23.3
Corrections
65
Symptom
26.3
26.2
26.1
25.4
25.3
25.1
25.2
24.2
24.3
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
25 3 R l
d
d i
26.1 Replace faulty thermostat.
25.1
24.3
24.2
24.1
Corrections
66
Symptom
28.1
27.2
27.1
TROUBLESHOOTING (cont.)
11.4 Components Malfunction
11.
Corrections
67
2.
3.
4.
5.
6.
7.
Symptom
2.1
2.2
3.1
7.1
Wear of components.
5.1
6.1
4.1
1.2
TROUBLESHOOTING (cont.)
11.5 Mechanical malfunction
1.
11.
7.1
6.1
5.1
4.1
3.1
2.2
2.1
1.2
1.1
Corrections
12.
DESCRIPTION
2340
2540
ARM067-0012
ARM067-0012
ARM067-0023
ARM067-0023
BOL190-0024
BOL190-0024
Sheet screw
BOL194-0340
BOL194-0340
CMT100-0002
CMT100-0002
Air jet
CMT100-0006
CMT100-0006
CMT240-0004
CMT240-0004
Nut CTOT84-5010-02
CMT240-0020
CMT240-0020
CMT240-0022
CMT240-0022
CMT240-0023
CMT240-0023
ELE039-0031
ELE039-0031
Sleeve 1/4
FIT100-0216
FIT100-0216
Sleeve 5/16
FIT100-0217
FIT100-0217
FIT100-0226
FIT100-0226
FIT100-0227
FIT100-0227
FIT100-0246
FIT100-0246
Fitting 5/16*1/4
FIT100-0290
FIT100-0290
FIT100-0308
FIT100-0308
GAS080-0010
GAS080-0010
NUT192-0234
NUT192-0234
U-nut 4.8*2.0-5.0
NUT192-0236
NUT192-0236
NUT193-0263
NUT193-0263
NUT193-0292
NUT193-0292
Teflon bushing
NUT193-0340
NUT193-0340
68
DESCRIPTION
2340
2540
SRV000-0251
SRV000-0251
SVL029-0014
SVL029-0014
THE007-0001
THE007-0001
CMT240-0003
CMT240-0003
Tightening nut
CMT201-0003
CMT201-0003
Door electromagnet
SOL027-0002
SOL027-0002
Screw
BOL191-0032
BOL191-0032
Door switch
ELE036 0009
ELE036-0009
Spring
SPR177-0012
SPR177-0012
Screw
BOL191-0033
BOL191-0033
Nut CTOT84-5010-02
CMT240-0020
CMT240-0020
Drain valve
VLV170-0066
VLV170-0066
O-Ring 10*2.5
SRV000-0224
SRV000-0224
O-Ring 6*2
SRV000-0232
SRV000-0232
CMT098-0001
CMT098-0001
CMT196-0001
CMT098-0001
COT204-0055
COT204-0055
ELE039-0048
ELE039-0048
Nut 5/32
NUT192-0217
NUT192-0217
NUT193-0291
NUT193-0291
GAS086-0001
GAS086-0001
THE003-0001
THE003-0001
THE006-0002
THE006-0002
HEA011-0001
HEA011-0002
SOL026-0004
SOL026-0004
CTP201-0004
CTP201-0004
69
DESCRIPTION
2340
2540
WIR040-0016
WIR040-0016
WIR040-0003
WIR040-0003
Keypad panel
CPN064-0007
CPN064-0007
ARM172-0007
ARM172-0007
Water Reservoir
RES075-0002
RES075-0002
Tightening nut
CMT201-0003
CMT201-0003
Locking pin
COT204-0065
COT204-0065
Outer cover
COV240-0006
COV240-0006
Rear cover
RCV240-0007
RCV240-0007
Air jet
CMT100-0006
CMT100-0006
AC-T2 board
CTP201-0060
CTP201-0060
ANL-T2 board
CTP201-0100
CTP201-0100
DIG-T2 board
CTP201-0102
CTP201-0102
CTP201-0065
CTP201-0065
CTP201-0082
CTP201-0082
CTP201-0083
CTP201-0083
CTP201-0084
CTP201-0084
COT204-0055
COT204-0055
ELE035-0001
ELE035-0001
ELE035-0012
ELE035-0012
ELE035-0031
ELE035-0031
ELE036-0001
ELE036-0001
FIL175-0011
FIL175-0011
GAS080-0002
GAS080-0003
Condenser
70
DESCRIPTION
2340
2540
GAS080-0007
GAS080-0007
GAU029-0009
GAU029-0009
LOK240-0001
LOK240-0001
LOK254-0001
LOK254-0001
Console base
POL062-0002
POL062-0002
POL063-0002
POL063-0002
POL065-0006
POL065-0006
POL067-0004
POL067-0004
PUM057-0007
PUM057-0007
SOL026-0015
SOL026-0015
SOL026-0016
SOL026-0016
SOL027-0002
SOL027-0002
SPR137-0004
SPR137-0004
Door Cover
SRV000-0231
SRV000-0231
SVL029-0004
SVL029-0004
THE002-0003
THE002-0003
Printer DPU 20
THE002-0005
THE002-0005
THE005-0001
THE005-0001
THE005-0003
THE005-0003
THE007-0001
THE007-0001
Tray
TRY240-0001
TRY240-0001
Front leg
WHE070-0012
WHE070-0012
WHE070-0014
WHE070-0014
WIR040-0003
WIR040-0003
SRV000-0252
SRV000-0252
Washer
71
DESCRIPTION
2340
2540
CTP201-0061
CTP201-0061
POL067-0003
POL067-0003
LOK240-0007
LOK240-0007
Locking base
LOK240-0012
LOK240-0012
LOK240-0014
LOK240-0014
LOK240-0031
LOK240-0031
Bakelite handle
HAN071-0003
HAN071-0003
NUT193-0339
NUT193-0339
Okolon disc
LOK240-0017
LOK240-0017
LOK240-0020
LOK240-0020
72
FRONT VIEW
Ref.
Description
Cat No.
VLV170-0066
CMT240-0003
LOK240-0001
Door switch
ELE036-0009
Autoclave cover
POL067-0004
SRV000-0252
Safety valve
SVL029-0004
CMT100-0006
10
Pressure gauge
GAU029-0009
11
POL065-0006
12
Main switch
ELE035-0012
13
Printer
THE002-0005
14
CTP201-0061
15
Completion to panel
POL063-0002
16
Panel base
POL062-0002
17
18
73
POL067-0003
VESSEL ASSEMBLY
74
L:\WORD-PLT\man237.doc
No.
Description
Cat. No.
Bushing
N/A
Locking base
LOK240-0012
LOK240-0014
LOK240-0031
Bakelite handle
HAN071-0003
N/A
NUT193-0339
10
Okolon disc
Bushing (1) + Locking screw housing (3) +
Closing bridge c clip (8)
LOK240-0017
LOK240-0007
N/A
75
LOK240-0020
76
Cap
Strainer
element
Gasket
Description
Strainer
Housing
Cat. No.
FIL175-0027
Strainer element
SRV000-0235
Teflon gasket 3 mm
GAS082-0007
FIL175-0026
77
13.
VALVES NUMBERING
The valves in the drawing and the manual are numbered according to their
function. The following list includes all the valve numbers that are in use in
Tuttnauer
0.
1.
2.
3.
4.
5.
FUNCTION
FEED WATER
MINERAL FREE
WATER
COMPRESSED
AIR
AIR
VACUUM
01.
02.
03.
11.
12.
13.
14.
15.
16.
17.
18.
21.
22.
23.
24.
25.
31.
32.
33.
34 1.
34 2.
38 1.
38 2.
39 1.
39 2.
Air inlet
Air inlet - to chamber
To splash cooling pipe
To door 1 seal
To door 2 seal
Open door 1
Open door 2
Close door 1
Close door 2
41.
42.
43.
44.
45.
51.
52.
53.1
53-2.
Vacuum - break
Vacuum - to pump
Vacuum - from door 1 seal
Vacuum - from door 2 seal
78
6.
7.
8.
9.
DRAIN
EXHAUST
GAS
STEAM
61.
62.
63.
64.
65.
66.
67.
68.
69.
70.
71.
72.
73.
74.
75.
76.
77.
78.
79.
81.
82.
83.
Inlet
Main inlet
Inlet through humidifier
90.
Steam from building source
91.
Steam to jacket (From outer source)
92.
Steam inlet
93.
Steam to chamber
94 1. Steam to door 1 seal
94 2. Steam to door 2 seal
95.
Steam to heat exchanger
96.
Steam to sanitary filter
97.
Steam from steam generator
98.
Steam to activate ejector
79
PIPING DRAWING
80
ELECTRICAL DRAWING
81