You are on page 1of 76

No.

X730-324EH

INSTALLATION

MANUAL

FOR
DIAGNOSTIC

ULTRASOUND

SSA-340A
(ZC730=324E*H)

TOSHIBA

@ TOSHIBA

CORPORATION

CORPORATION

ALL RIGHTS RESERVED

1994

SYSTEM

No. 2C730-324E*D

IMPORTANT!
1.

No part of this manual may be copied or reprinted,


without written permission.

2.

The contents of this manual are subject to change without prior notice
and without our legal obligation.

C-l

in whole or in part,

No. X730-324E*H

REVISION RECORD
REV.

DATE

REASON
/AUTHOR

(MM/yy)

INI.

021'94

Mr. Nakajima, Kikuchi

*A

121'94

Mr. Nakajima, Kikuchi

*B

021'95

Mr. Nakajima

*C

091'95

Mr. Mita

*D

101'96

Mr. Okumoto

*El

051'97

Mr. Okumoto

*F

061'98

Mr. Okumoto

*d

081'98

Mr. Sato

*H

031'99

Mr. Okumoto

R-l

PAGE
CHANGED
-------

SER

No-

DOC.
PRODUCT.

TM-WW

No. 2C730-324E*E

CONTENTS
Page
1.

SCOPE ____________________~~~~~~~----~- -___________________~~~~~~~~~~ 1

2.

INSTALLATION CONDITIONS ____________________~~~~~~~~~~~~~~~~~~~~~~~~~

3.

OPERATING CONDITIONS _______~~~~~~-~~~~~~~~_~~~~~~~~~~~~~~~~~________2

3.1

Power

3.2

Grounding______---__--------

3.3

Operating Conditions ------------ ____________________~~~~~~~~~~~ 3

3.4

Storage and Transportation Conditions ____________________------ 3

Supply

__________~~~~~~~~~~~~~~~~~~~~~____~~~~_____~_~~___~

____________________~~~~~~~~~~~~~~~

4.

UNPACKING ___________________~~~~~~~~~~~~___~~_~~~~~~~~~~___~_~~~_~~~3

5.

PREPARATION FOR INSTALLATION ___________~________~~~~~~~~~~~~~~~~~~~~

6.

ASSEMBLY ____________________~~~~~~~~~-~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~

7.

6.1

Mounting

6.2

Mounting the Probe ____________________~~~~~~~~~~~~~~~~~~~~~~~~~

the

Cable

~~~_~~~~~~_~~~~~_________~~_~~~~___~_~_~~~~~~

CONNECTING AND STORING OPTIONAL DEVICES -----------------------------7


7.1

Setting the Ultrasound System ____________________~~~~~~~~~~~~~~

7.2

Mounting

7.3

the

Peripherals

Rack

Connecting the Camera Unit


(same for both multi-exposure

~~~~~~~~_~~~_~~~~~~~~~_~~~~~~~~~~~~

type and black/white type) -------20

7.4

Connecting the VCR ____________________~~~~~~~~~~~~~~~~~~~~~~~~~

22

7.5

Connecting the Imager ____________________~~~~~~~~~~~~~~~~~~~~~~

27

7.6

Connecting the Black/White Printer ____________________~~~~~~~~~ 31

7.7

Connecting the Footswitch ______________________________________33

7.8

Connecting the UIPS-342A ____________________~~~~~~~~~~~~~~~~~~~

7.9

Connecting the UJUR341A/UJUR342A -___________________~~~~~~~~~~~ 37

35

7.10 Replacing and Connecting the Monitor ____________________------- 38


7.11

8.

Connecting

the

UICF340A

-------------~~--~~~~~------~-----------3~~

7.12 Connecting the SDS-340A ____________________~~~~~~~~~~~~~~~~~~~~

40

FUNCTIONAL CHECK ____________________~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~

43

8.1

Settings at Power-On ____________________~~~~~~~~~~~~~~~~~~~~~~~

43

8.2

Operational Check __________-_-_______~~~~~~~~~~~~~~~~~~~~~~~~~~

44

-a-

No. 2C730-324E*G

CONTENTS - continued
Page
9.

LOCALIZED LABELS ----------- ______~_____________~~~~~~~~~~~~~~~~~~~~~

-b-

60

No. X730-324E*A

1.

SCOPE
This installation manual describes:
(1) Installation procedures for the SSA-340A
(2) Procedures for connecting optional devices
printer, VCR, etc.) to the SSA-340A

2.

(black/white or color

INSTALLATION CONDITIONS
The SSA-340A must not be used:
(1) near heat sources or in damp locations
(2) in environments containing flammable gases
(3) in intense electric or magnetic fields (such as locations near
transformers)
(4) near devices generating high frequencies

l-

No. X730-324E*A

3.
3.1

OPERATING CONDITIONS
Power Supply
100 to 127 VAC f 10% (50160 Hz) or 220 to 240 VAC f 10E (50/60 Hz)
1.5 kVA or less for the standard configuration of the main unit
Capacity of the AC outlet

3.2

(1) 100 V regions .. . Operating panel (1) :


(2)
..
VCR panel

Total 100 V Max. 4.1 A

(2) 110 V regions .. . Operating panel (1) :


(2)
:
VCR panel

Total 110 V Max. 3.7 A

(3) 120 V regions .*. Operating panel (1) :


(2)
:
VCR panel

Total 120 V Max. 3.23 A

(4) 127 V regions . .. Operating panel (1) :


(2)
:
VCR panel

Total 127 V Max. 3.2 A

(5) 220 V regions ... Operating panel (1) :


(2)
:
VCR panel

Total 220 V Max. 1.86 A

(6) 230 V regions ... Operating panel (1) :


(2)
:
VCR panel

Total 230 V Max. 1.78 A

(7) 240 V regions ... Operating panel (1) :


(2)
:
VCR panel

Total 240 V Max. 1.7 A

Total 100 V Max. 0.75 A

Total 110 V Max. 0.68 A

Total 120 V Max. 0.6 A

Total 127 V Max. 0.59 A

Total 220 V Max. 0.34 A

Total 230 V Max. 0.32 A

Total 240 V Max. 0.31 A

Grounding
The SSA-340A must be grounded using the grounding wire of the three-wire
power cable. The grounding resistance must be 100 $2 or less.
When the SSA-340A is used with units that have power supplied from
outlets other than the AC outlet provided with the SSA-340A, use the
equipotential terminal to ensure that each unit is at the same
potential.

-2-

No. 2C730-324E*F

3.3

Operating Conditions
Temperature
:
Relative humidity
:
Atmospheric pressure:

3.4

Storage and Transportation Conditions


Temperature
:
Relative humidity
:
Atmospheric pressure:

4.

5OC to 35OC
35% to 85X (no condensation)
700 hPa to 1060 hPa

-10C to 60C
30% to 95X (no condensation)

700 hPa to 1060 hPa

UNPACKING
Confirm that the following items are provided when unpacking the SSA340A.
Quantity

(2) 3.75 MHz probe (PVF-375MT) ____________________~~~~~~~~~~~~~~

(3) Accessories ____________________~~~~~~~~~~~~~~~~~~~~~~~~~~~~~

1 set

(a) Acoustic coupler


(b) Fuse:
:

5.

(1)

For 100 V to 127 V, 5


1
For 220 V to 240 V, 2
0.5

A
A
A
A

(for
(for
(for
(for

external
external
external
external

AC
AC
AC
AC

output) (1)
output) (1)
output)
output)

(c) Probe hanger

(1)

(d) Operation Manual

(3)

(e) Operation Note

(1)

(f) Installation Manual

(1)

(g) Installation Quality Check Sheet

(1)

PREPARATION FOR INSTALLATION


(1) Personnel
Tools

2 persons
Phillips screwdriver (large and small)
Metric Allen wrench (M6)
Bladed screwdriver (small)
Tweezers

-3

one each
one
one
one pair

No. 2C730-324E*A

6.
6.1

ASSEMBLY
Mounting the Cable Arm
(1) Orientate the tip of the cable arm toward the rear of the main unit,
insert it into the arm mounting hole on the probe holder, and turn
the tip of the arm toward the front of the main unit by 180' to
mount the cable arm.

cated
.

Figure 6.1-l

No. 2C730-324E*A

6.2

Mounting the Probe


(1) Hang the probe cable on the hook of the cable arm to
being caught by the casters.

Figure 6.2-l

-5-

prevent

it

from

No. 2C73Q-324E*D

(2) Connect the connector of the probe to the connector on the main
unit. When connecting the probe, insert the probe connector into
the connector on the main unit and turn the lock handle clockwise
(NLOCK1fdirection) to secure the connector.

Probe connector
on the
main u

Figure 6.2-2

-6

No. 2C730-324E*F

7.
7.1

CONNECTING AND STORING OPTIONAL DEVICES


Setting the Ultrasound System
Perform setting for the functions of PRINT switches A and B and setting
for allowing the magneto-optical disk unit to be controlled by the
diagnostic ultrasound system.
The settings on pages "SERVICE" and "OTHER" of the preset menu of the
diagnostic ultrasound system are described below:
(1) SERVICE menu
(a> Perform the setting for the functions of PRINT switches A and B.
There are four possible function settings each for PRINT
switches A and B as shown below.
Set the functions according to the application.
(Function selection)
.. . The B/W printer (or camera) operates in black/white
mode and the color printer (or camera) operates in
color mode.
(B/W)
... The B/W printer (or camera) always operates
regardless of the mode.
(COLOR) . . . The color printer (or camera) always operates
regardless of the mode.
(MO)
. .. Allows data to be recorded onto the magneto-optical
disk.
(AUTO)

(b) Perform setting for the control of the magneto-optical


from the diagnostic ultrasound

disk unit

system.

START OF TEXT 1 NONE

COMMUNICATING 1 DISP. ON

END OF TEXT

TERMCHAR

CR t LF

-7-

11000

No. 2C730-324E*F

OTHER menu
Set the controls of the magneto-optical

disk unit as follows.

The table below shows the specifications for various settings of MO


CAPA SEL.

CAPA SEL
3

High quality

1 Intermediate

1
0

7.2

Average compression ratio

Compression mode

No. of images that


can be recorded
100 approx.

111
quality

I Standard
I High capacity

l/l0

1000 approx.

l/20

2000 approx.

l/30

3000 approx.

Mounting the Peripherals Rack


The procedures for mounting the peripherals

rack are described below.

(1) Before mounting the peripherals rack, connect the cables (signal
cable, control cable, power supply cable) for all the optional
devices to be mounted to the panel of the main unit as shown below.

-8

No. 2C730-324E*A

(a) Connecting the color printer cables

I
Cable (C):

Power cable
Signal and control cables
Figure 7.2-l

a. Connect the "37pin-Dsub" connector of the signal and control


cables (cable (C)) to the [COLOR PRINTER] connector of the
diagnostic ultrasound system and tighten the screw to prevent
the connector from being disconnected.
b. Connect the straight-type connector of the power cable to the
AC outlet connector of the diagnostic ultrasound system.
Caution:

Overload will occur if either the color printer or


black/white printer draws its power from the VCR
panel. Refer to subsection 3.1 on page 2 to
connect the power.

9 -

No. 2C730-324E*A

(b) Connecting the black/white printer cables

or

Cable (D):

I T

Power cable
Signal and control cables

Figure 7.2-2
a. Connect the "15pin-Dsub" connector of the signal and control
cables (cable (D)) to the [B/W PRINTER] connector of the
diagnostic ultrasound system and tighten the screw to prevent
the connector from being disconnected.
b. Connect the straight-type connector of the power cable to the
AC outlet connector of the diagnostic ultrasound system.
Caution:

(2)

Overload will occur if either the color printer or


black/white printer draws its power from the VCR
panel. Refer to subsection 3.1 on page 2 to
connect the power.

Remove the probe holder.

(3) Mount the frame to the support below the operating panel of the main
unit on the right and left with 2 retaining screws (A) and 2 spring
.
washers. At the same time, pass the cables already connected
through the notch on the far side of the rack.
(4) Place rubber caps into the two holes for the retaining screws on the
left side.

10

No. X730-324E*D

(5) Slide plate (A) to the middle rack of the frame and secure it with
the 4 retaining screws (B) from below.

Probe holder

Be P articularly careful
rega rding the engagement
proc edures for the
circ led sections.
Frame
\1

-rLPlate

(A)

-Retaining

screws (B) (M4, 4 total)

Figure 7.2-3
(6)

Mount plate (C) onto plate (B) with 2 nuts and washers and cover it
with the blank plate.
(Figure 7.2-4)

(7) Mount plate (B), onto which plate (C) has been mounted, to the upper
rack of the frame with 2 retaining screws (C). (Figure 7.2-5) 5
(8) Place rubber caps into the 2 holes for retaining screws.

11

No. X730-324E*D

(9) Remount the probe holder in its original position.

Plate (B)

Figure 7.2-4

MV-200 (magneto-optical
disk unit)

aining screw (C) (M3, 2 total)


Rubber cap (2)

Figure 7.2-5

- 12 -

No. X730-324E*A

(IO) Connect the supplied cable (C) for the color printer.
details, refer to the figure and table below.)

(For

Cable from the upper panel


To each OUTPUT (R, G, B, and SYNC)

each INPUT (R, G, B, and SYNC)


Power cable

o-

Cable (C) (For the color printer)

B SYNC (For the TCP-7070, do not connect


the BNC of SYNC OUT.)

e control jack (pin jack type) for the UP-1 850 (Sony)
Remote control jack for the TCP-7070
(Do not connect this connector.)

Figure 7.2-6

On supplied cable (C)


for the color printer

On the TCP-7070

On the UP-1850
(SOW

R-IN

IN

INPUT

G-IN

IN

G/G+SYNC

INPUT

B-IN

IN

INPUT

SYNC-IN

H/HtV-SYNC

INPUT

SYNC

R-OUT

OUT

OUTPUT

G-OUT

OUT

G/GtSYNC

OUTPUT

B-OUT

OUT

OUTPUT

SYNC-OUT

Do not connect.

OUTPUT

SYNC

TCP

REMOTE

Do not connect.

UP

Do not connect.

REMOTE 2

VY

Do not connect.

Do not connect.

Connectors other than those listed above do not need to be


connected.
Caution:

Be extremely careful because the remote control jack is


easily disconnected.

13

No. X730-324E*A

(11) Set the switch and DIP switch on the rear of the printer as shown
below.
(a) UP1850
DIP switch
1

ON

OFF

OFF

OFF

OFF

OFF

OFF

OFF

Switch
75 sz
ON
OFF

(b) TCP7070
cl> Switch (IMPEDANCE)
75 Sz m-1

HIGH

The DIP switch for RS-232C DATA IN does not need to be set.
<2> Turn ON the main power supply SW at the upper left on the
rear.
(12) Connect the power cable to the power input of the color printer.
(13) Turn ON the power and check the functions of the printer. If
horizontal misalignment is seen on printout from the TCP7070, remove
the top cover of the printer to perform centering adjustment (refer
to the next page).
If vertical misalignment is seen on the printout, use the hand SW
supplied with the TCP7070 to perform adjustment (see below).
When the UP1850 is used, it is not necessary to adjust the display
position.
[Adjustment procedures for display position of the TCP7070]
(a) Adjusting the vertical display position
<l> Use the hand SW to perform the following settings.
PRINT AREA
TOP
c -4: 16> -2
BOTTOM x-16: -2> -2

- 14 -

No. X730-324E*A

(b) Adjusting the horizontal display position


<I> Use the hand SW to perform the following settings.
PRINT AREA
LEFT
RIGHT

<0:8>0
K-8: O> 0

Remove the 6 screws in the

(FRONT side)

<3>
TOP VIEW (FRONT side)

Remove the 2 screws in the


figure on the left and lift
up the [VIDEO] PWB towards
the front side together with
the rear panel to which the
S-VIDEO and VIDEO connectors
are connected.

(FRONT side)

Turn VR830 on the [RGB] PWB


mounted below the [VIDEO] PWB
counterclockwise (when viewed
from the REAR side) until it
is near maximum.

(REAR side)

15

No. X730-324E*D

<5> Return the PWB lifted up in step <3> to its original


position and perform print operation to check the printed
horizontal display position.
If the printed horizontal
display position is misaligned to the right, perform step
<3> again and slowly turn VR830 on the [RGBJ PWB clockwise
until the position is correct.
<6> When the horizontal display position has been adjusted,
reverse steps x2> and <3> to secure the PWB to the top cover
with screws.

Color printer
TCP-7070 (Mitsubishi)
UP-1850
(Sony)

Plate

Frame

For the
UP-18
(2)

.e CD)

Figure 7.2-7

- 16 -

No. 2C730-324E*D

(14)

Mount the color printer on the base, being careful with regard to
the connection cables.
l

For the TCP-7070 (Mitsubishi), mount 4 plates (D) onto the base
with retaining screws (D) (2 each, 8 total) and then adjust the
feet of the printer to mount it.
Caution:

When mounting the TCP-7070 (Mitsubishi), confirm that


the feet of the TCP-7070 are one step each. If they
are two steps each, remove one step from each foot and
tighten the remaining steps with the supplied short
screws.
(Be sure to use the supplied screws. The
original screws used for the two-step feet are too long
and will come into contact with the circuits in the
printer, causing a short circuit, if they are used to
mount the one-step feet.)

For the UP-1850 (Sony), mount 2 plates (E) onto the base with
retaining screws (D) (2 each, 4 total) and then adjust the feet of
the printer to mount it.

(15) Slide the base onto which the printer has been mounted into the
frame.

-Retaining
screw (E)

Figure 7.2-8

- 17 -

(M4)

No. X730-324E*A

(16) After sliding the base in half way, temporarily secure it to the
retaining hole in the middle with 2 retaining screws (E) to bundle
the cables.
Caution:

Be sure to perform this temporary securing.


the printer may fall.

Otherwise,

(17) After the cables are bundled, remove the temporary retaining screws,
slide the base all the way in, and secure it with 6 retaining screws
(EL

To REMOTE
Cable from the
upper panel

)--Power

Cable (D) (For the


black/white
printer)

Figure 7.2-9
(18)

Connect supplied cable (D) as shown in the figure above.

(19)

Connect the power cable to the power input of the black/white


printer.

(20)

Turn ON the power to check the operation of the printer.

- 18 -

No. X730-324E*D

ion of the
ratin g name
ze
attac El

TP-8700/8010

screw (F) (M4 flat-head screw)

-Retaining
-Retaining

screw (G) (M3 flat-head screw)

Figure 7.2-10
(21) Mount plate (F) to the bottom of the base
PI.

with 4 retaining screws

(22) Insert the TP-8700, being careful with regard to the cables
connected to plate (F), and mount it with 4 retaining screws (G).
the side of the UZRI340A.
(23) Attach the supplied rating nameplate to
(The rating nameplate may have already been attached in some cases.)

19

No. 2C730-324E*A

7.3

Connecting the Camera Unit (same for both multi-exposure


black/white type)

type and

(I) Connect each of the supplied cables as shown in the figure below.

For the
multi-exposure

For the black/white type

Figure 7.3-l

- 20 -

No. X730-324E*A

(2) Mount plate (A) on the support section at the lower portion of the
operating console of the main unit with retaining screws (A) (four).
* However, one of screws (A) must be tightened together with the
grounding wire connected to the camera.
(3) Slide the camera unit on which plate (B) has been mounted into plate
(A) mounted on the support section of the main unit and secure it
with resin-head screws (two).
* Be careful not to subject the cable to excessive stress.

screw

&

Plate (A)
Retaining
screw (A)

Multi-exposure camera unit or


black/white camera unit

Figure 7.3-2

21

No. X730-324E*A

7.4

Connecting the VCR

m
OI:

VCR

CONTROL

-S-VHS

(TO)

VW
l-AUDIO

OUT
REMOTE

\\-I I I

To the AC LINE on the front panel


of the main unit
Figure 7.4-1

- 22 -

No. X730-324E*B

(1) Connecting the VCR to the main unit


VCR side

Main-unit

AUDIO IN

AUDIO OUT

AUDIO OUT

AUDIO IN

VIDEO IN

VHS TO

VIDEO OUT

VHS FROM

side I

Connection cable
Cable (A) (VCR t AUDIO cable)

S VIDEO IN

I S-VHS TO

(Cable (B) (cable for S-VHS)

S VIDEO OUT

1 S-VHS FROM

REMOTE

I VCR CONTROL
AC LINE
I_

1 Cable

Power CNN

(C) (control cable)

cable
I Power
~
~~

Connect the VCR according to the table above.


impedance of the video signal to 75 0.

Also set the

It is not recommended that the VHS input/output signals be used.


Use the S-VHS input/output signals. The PAL version does not support
and cannot use the VHS input signal.

- 23 -

No. X730-324E*A

Resin-head
screw
Spring washer
Flat washer

Plate (A)

Figure 7.4-Z

Remove the screws (four) retaining the base cover and secure plate
(A) together with the base cover with the resin-head screws, spring
washers, and flat washers (four each).
* For the AG-7350
figure above.

(Panasonic), mount plate (A) as shown in the

* For the SVO-95OOMD (Sony), mount plate (A) so that the angled
section is orientated inward.

- 24 -

No. X730-324E*A

(3) After completing cable connection, mount the VCR on the base of the
main unit, pass the fixing bands through plate (A) as shown in the
figure below, and secure the VCR using the bands.
*Be careful with regard to the routing of the cable.
(4) Attach the rating plate supplied on the rear or side of the main
unit.

Fixing bands

Plat

Figure 7.4-3

- 25 -

No. 2C730-324E*A

Figure 7.4-4

- 26 -

No. 2C730-324E*B

7.5

Connecting the Imager


(1) Connect the cable as shown in the figure below.

VHS

(TO)

front panel of the main unit


Figure 7.5-l
The PAL version does not support and cannot use the VHS input/output
signals.

- 27 -

No. X730-324E*A

Figure 7.5-Z

Remove the screws retaining the base cover (four) and resecure plate
(A) together with the base cover with retaining screws (A) (four).
* The only imager which can be mounted is the MP-40 manufactured
IIE.

28

by

No. X730-324E*A

(3) After cable connection is completed, mount the imager on the base of
the main unit, pass the fixing bands through plate (A) as shown in
the figure below, and fix the imager using the bands.
* Be careful with regard to the routing of the cable.

Fixing bands

Figure 7.5-3

29

No. X730-324E*A

* To mount the imager and the peripherals rack at the same time,
replace the supplied feet with the rubber feet of the imager
(four).
(4) Attach the rating plate supplied on the rear or side of the main
unit.

Figure 7.5-4

- 30 -

No. X730-324E*A

7.6

Connecting the Black/White Printer


(1) Connect supplied cable (A) as shown in the figure below.
(2) Connect the power cable to an AC outlet.
either of the right or left outlets.)

1 B/W PRINTER

1 or

(It can be connected to

To REMOTE
To

Cable (A)

Figure 7.6-2

(3) Attach the rating plate supplied on the UZRI341A.


rating plate is already attached.)

- 31 -

(In some cases, a

No. 26730-324E*D

(4) Mount plate (A) on the support section at the lower of the operating
console of the main unit with retaining screws (A) (four).
* Plate (A) can be mounted at either the center or the left edge,
(5) Mount the TP-8700/8010 on plate (A) secured at the support of the
main unit.
(6) Fix the TP-8700/8010 to plate (A) with retaining screws (B) (four).

Plate (A)

TP-8700/8010

taining screw (A)


Retaining screw (B) -I

Figure 7.6-Z

- 32 -

No. 2C730-324E*A

7.7

Connecting the Footswitch


(1) Connect the footswitch to the connector for the footswitch at the
front of the power supply.
(The settings at the time of shipment from the factory are "A: PRINT
A" and "B: FREEZE".)

Front panel of the


power SUPPlY

-Footswitch

Figure 7.7-l

- 33 -

No. X730-324E*A

(2) To switch the settings of the footswitch, remove the panel cover and
main panel and the knobs on the sub-panel (four) and switch the
rotary SW on the panel PWB.
For each footswitch, A and B, the settings are "1. FREEZE", "2.
PRINT A", "3. PRINT B", and "4. RECIPAUSE".

Figure 7.7-2

(3) For the panel cover and main panel, remove the 15 screws from the
bottom as shown in the figure below.
(4) For the knobs on the sub-panel, remove the five setscrews.

- 34 -

No. X730-324E*E

7.8

Connecting the UIPS-342A

Shield plate
-__

for the

screw (C)

r-

Rear coyer

Retaining
screw (E)
>

Knob for hanging


the cable

Rubber
cap

Figure 7.8-l

(1) Remove the knobs for hanging the cable (two), rubber cap, and
retaining screws (A) (two each) and remove the rear cover from the
main unit.
(2) Remove retaining screws (B) (four) to remove the fan plate from the
main unit.
* Be sure to disconnect the power CNN of the fan.
(3) Remove retaining screws (C) (11) to remove the shield plate for the
T&R rack from the rack.

- 35 -

No. 2C730-324E*E

(4) Remove retaining screws (D) to remove plate (A) which connects the
PROBE-SEL-PWB and PULSER-PWB in the T&R rack.
(5) Remove the rubber cap and retaining screws (E) (eight) on the side
of the system.
(6).Pull out the PROBE-SEL-PWB on the left interior of the T&R rack and
insert the UIPS-342A.
(7) Reverse steps (1) to (5) to remount the covers.
(8) Attach the rating plate supplied on the rear of the equipment.

- 36 -

No. X730-324E*E

7.9

Connecting the UJUR341A/UJUR342A

Xi,oht

Retaining

side cover (upper)

wer):
screw

(B)

screw (C)

Figure 7.9-l

(1) Remove the rubber caps (five) and retaining screws (A) (seven in
total) to remove the lower right side cover from the main unit.

(2) Remove retaining screws (B) (six) to remove the shield plate.
(3) Remove retaining screws (C) (eight) to remove the shield plate for
the D&D rack.
(4) Insert the ECG/NONFADE PWB into the third slot from the left of the
D&D rack (ECGINONFADE).
(5) After removing the cable arm, remove the two retaining screws (D) to
remove the BOSS.
(6) Remove the four retaining screws (E) and remove the upper right side
cover while lifting it up.
- 37 -

No. 2C730-324E*D

(7) Remove the four retaining screws (F) to remove the decorative cover.

cover

rcrew

Figure 7.9-2

- 37A -

CT)

No. 2C730-324E*D

(8) Mount the ECG panel to the frame with the four retaining screws (F).

Clamp

Shield the cables for


ECG and DC IN and
clamp them.

rew
.ning scre

Clamp (for the UJUR341A only)

Clamp

-Frame
Shield

Figure 7.9-3
(9) Extend the ECG cables from the notch on the left side, bundle them
using the clamp attached to the frame, and pass them through the
opening to the D&D rack.
*

For the UJUR341A (The ECG cables are mesh-shielded,)


Fix each cable with the supplied clamp so that the end of the
shield of each cable (up to approximately 5 mm) is pinched between
(Use the two retaining screws (D).)
the clamp and the frame.

(10) Connect CNNs ECGlJE and ECGZJE to the CNN on the free edge side of
the ECG/NONFADE PWB.
(11) Reverse steps (1) to (6) to remount the shield plate and covers.
(12) Attach the supplied rating nameplate on the rear of the system.
- 37B -

No. 2C730-324E*D

(13) Be sure to use the attached cable for connection to the external
recording meter (only TMSE). Pin assignments of the connection
cable are given below:

Pin No. I Color of core wire-Ipin


1

2
3

White
I

4
5

Red

Black
Yellow

Blue

I
I
I

N.C.
DC
GND
N.C.

I N.C.

Figure 7.9-2

- 37c -

No. 2C730-324E*C

7.10

Replacing and Connecting the Monitor

Monitor rear cover

b
Retaining screw (A)

Figure 7.10-l
(1) Remove the retaining screws (A) and (B) to remove the monitor rear
cover.

BNC for the VIDEO signal


t

GND terminal
Monitor set screw

Figure 7.10-2
(2) Remove the power CNN and the BNC for the VIDEO signal
(3) Remove the wing nut to remove the GND terminal.
(4) Remove the monitor setscrew.
(5) Lift the monitor unit straight up to remove the monitor.

- 38 -

No. 2C730-324E*C

(6) Remove the rubber caps, screws, and knobs to remove the left side
cover and the rear cover.

Retaining screw
I

Rectangular rubber cap


Rear

coverI-

Round rubber cap

<

Left side cover

Knob

Retaining screw

screw

Round rubber cap

Figure 7.10-3

- 38A -

No. 2C730-324E*D

(7) Remove the retaining screw to remove the right side cover (lower).
(8) Remove the lower retaining screw and pull out the boss.
(9) Remove the retaining screw and remove the right side cover (upper)
while lifting it.

BOSS

Right side cover (upper)

r cap

cover (lower)

Figure 7.10-4

- 38B -

No. 2C730-324E*C

(10) Remove the left side cover shield plate and the rear shield plate.
* Remove the fan power cable.
(11) Remove the retaining screws of the top cover and remove the cover
by lifting it.

Retaining
screw

screw
Rear
shield

,plafe/

Left side cover


shield plate

*.

Retaining plate

Figure 7.10-5

- 38C -

No. 2C730-324E*F

(12)

Remove the retaining screw to remove the clamp. When the 15"
monitor is to be mounted, also remove the monitor mounting boss.

(13) Remove the grounding terminal secured with the T&R rack retaining
screw. Retighten the rack retaining screw.
(14) Remove the retaining screw to remove
plate.
(15) Remove the relay connectors

the brace and the power shield

(BWZPN and BW2JN).

(16) Remove the signal cable connector(M25JE)


monitor cable.

from D&D and remove the

-Clamp

_A
P

Monitor mounting boss

Figure 7.10-6

- 38D -

No. 2C730-324E*F

(17) Attach the connector (Ml8JE) of the color monitor


the D&D motherboard connector (M18PE).

signal cable to

(18) Attach the connector (POlOJN) of the color monitor power cable to
the power supply unit connector (POlOPN).
19) After routing the color monitor cables connected in (17) and (18)
into the top plate as illustrated below, mount the clamp with the
retaining screw. When the 15" monitor is to be mounted, mount the
dedicated monitor mounting boss to the top plate with the retaining
screw.

Color monitor

Monitor
boss (d.edicated fo
the 15" monitor)

Figure 7.10-7

- 38E -

No. X730-324E*F

(20) Install the power shield plate and the brace in that order by
tightening the retaining screws.
(21)

Install the shield plates and the outer cover by reversing steps
(6) to (II). When the 15" monitor is to be mounted, the dedicated
top cover must be used.

(22)

Pull out the R, G, B, VD, and HD signal BNC and power cables from
the inside of the main unit.

(23)

Insert the monitor into the boss of the neck, taking care that the
cables are not disconnected.

(24) Connect the R, G, B, VD, and HD signal BNC connectors to the


corresponding IN ports of the monitor.
(25) Connect the power cable to the monitor.
(26) Install the monitor setscrew.
dedicated monitor setscrew.

For the 15" monitor, mount the

(27) Mount the monitor rear cover by reversing step (1). For the 15"
monitor, mount the rear cover with screws at four locations on the
rear. ,
* Do not use RGB. OUT at the rear of the 10" color monitor,

since
the design makes it physically impossible.
For the 15" monitor,
mount the rear cover with screws at four locations on the rear.

R; G, B, VD, and HD
signal BNC
I

Monitor setscrew
(for the 15 monitor)
[Monitor setscrew
(for the IO monitor)]
Figure 7.10-8

- 38F -

No. X730-324E*H

(28) If the 15" color monitor is mounted, the display size (normally
12") can be switched from 12" to 15". Remove the cap of the
monitor rear cover by removing its screw to expose the monitor
rear panel, and set the slide switch on the panel. to "15" (refer
to figure 7.10-g).
Slide switch
IN

Figure 7.10-9

HS/CS

vs

Monitor rear panel

- 38G -

No. 2C730-324E*A

(4) Insert each PWB in the following slots inside the D&D rack.
(5 types)

PWB name

Slot position

ADC/LB/CAL

3rd from the right

FIL/CORR

4th from the right

(5) Reverse steps (1) to (3) to reattach each cover.


(6) Attach the rating plate supplied on the rear of the equipment.

39

No. X730-324E*D

7.12

Connecting the SDS-340A

Shield plate for the T&R rack


-Retaining

screw (C)

late

.
rky\ .

Rear cover

\.

>

Retaining

screw (B)

Knob for hanging


the cable

Rubber cap

Figure 7.12-l

(1) Remove cable hanging knobs (Z), rubber caps, and retaining screws
(A) (2 each) to remove the rear cover from the main unit.
(2) Remove retaining screws (B) (four) to remove the fan plate from the
main unit.
* Be sure to disconnect the power CNN of the fan.
(3) Remove retaining screws (C) (11) to remove the shield plate for the
T&R rack.
(4) Insert the PHASE DETECTOR PWB into the first slot from the right
(PHASE DETECTOR) inside the T&R rack.

- 40 -

No. X730-324E*D

Figure 7.12-Z

(5) Remove rubber caps (5) and retaining screws (D) (7 in total) to
remove the right side cover from the main unit.
(6) Remove retaining screws (E) (six) to remove the shield plate.
(7) Remove retaining screws (F) (eight) to remove the shield plate for
the D&D rack.

- 41 -

No. 2C730-324E*A

(8) Insert each PWB into the following slots inside the D&D rack.
(2 types)

PWB name

Slot position

FFT I/O

I 2nd from the right

FFT/CONT/AUDIO

1 1st from the right

(9) Connect PEASE DETECTOR to FFT/CONT/AUDIO


supplied.

using the BNC cable

(10) Reverse steps (1) to (3) and (5) to (7) to reattach the covers.
(11) Attach the rating plate supplied on the rear of the equipment.

- 42 -

No. 2C730-324E*C

8.
8.1

FUNCTIONAL CHECK
Settings at Power-On
Turn ON the power of the equipment.
that:
l

The power cable is connected.

One or more probes are connected.

Before turning ON the power, confirm

(1) Power-on status


After turning the power on, connect the PVF-375MT to probe connector
A and confirm that the correct PROBE ID has been entered.
Also, perform the following checks:
(a) Indicator lamps
a. Check that the power pilot lamp lights.
(The available
b. Check that various switch lamps light.
switches are lit in green and the switch currently selected
is lit in orange.)
(b) Monitor display
a. Make sure that after power-on, the required images are
displayed on the TV monitor.

Note:

No characters must be outside the display frame of the


monitor.

(2) Brightness of lamps


Check that turning the brightness adjustment control changes the
brightness level.
Relationship between the control and the brightness level
<Control>
Leftmost position
Clockwise
Counterclockwise

<Brightness level>
Off
High
Low

(3) Buzzer
Check that turning the volume adjustment control changes the sound
level of the operating sounds.
Relationship between the control and the volume
<Control>
Leftmost position
Clockwise
Counterclockwise

<Volume>
Off
High
Low

- 43 -

No. 2C730-324E*A

8.2

Operational Check
Check that the desired operation of the system is correctly performed.
(I) Image display functions

No.

Procedure

Item

lisplay mode

1.1

B-mode switch
selection

Remarks

s Display

Connect the specified probes


to probes A, and carry out
the operational checks
indicated below.

Requirement

Ultrasound
image display

(1) POWER ON
Real time

Same as above

(2)

(3) Push
H

ml
B

R;?al time

(4) Push
UEI

Freeze

(5) Push

Real time

Real time

&
Ezll

- 44 -

Switch lamp
which lights

'udgement Note

No. 2C730-324E*A

No.

Item

1.2

M-mode switch
selection

Procedure

Requirement

(1) Push

. Display

Remarks

Ultrasound
image display

B
llIlIl

Switch lamp
which lights

\\I
.

/I\\

Real time

FJJ
III
M

~
Scroll

(3)

* Scroll

Real timl

Scroll

(5) Push

l!Il

-/ cl
c
Scroll

(2) Push

I/

&
cl

- 45 -

Tudgement

Note

No. 2C730-324E*A

No.

1.3

Procedure

Item

BDF mode
switch
selection

Requirement

. Display

Remarks

Ultrasound
image display

(1) -POWER ON

(2)

Push

Real time

IL3
a

(3)

Real time

(4) Push

Freeze

(5) Push

(6) Push

Freeze

a
Cl

@
El

(7)

- 46 -

Real time

Switch lamp
which lights

Tudgement Note

No. X730-324E*A

No.

Procedure

Item

- Display

Remarks

Set PRESET MtB SIMULT of


COLOR to ON.
1.4

MDF mode
switch
selection
the
trackballs
referred to
here are all
those of the
palm
controller

Requirement
(1)

MENU (B)
[D/M POS]
ON

Ultrasound
image display

Switch lamp
which lights

Real time

G
J,
\

Generaiipurpose
marker

(3) Push

(4) Trackball
operations

Rotate the trackball to the


left and right, and make
sure that marker M on the
BDF mode image moves to the
left and the right and that
the appropriate MDF mode
images are displayed
according to the particular
new positions If marker M.

(5 )

Scroll

(6) Push

ZF

@
u

Scroll
Real time
\II,

El:
.

Et3

twice

II\\

Scroll

- 47 -

Tudgement Note

No. X730-324E*A

No.

Item

Procedure

- Display

1 Scroll
(9) During the mode where
freeze BDF images remain
on the display shown
above, press the [B
FRESH] (MENU (PW)) switch
and check that one BDF

- 48 -

Remarks

Judgement

Note

No. 2C730-324E*A

No.

Item

1.5

PW mode switch
selection

Procedure

Requirement

- Display

Ultrasound
image display

Remarks

Switch lamp
which lights

(1) POWER ON

(2)

M.ENU (B)
[D/M POS]
ON
General-burpose
marker

(3) Push

ICI
MEti (PW)
[D EXP]
OFF
Doppler range
gate marker

(4)

(5) Push

. El
Cl-/
\ \I/,

Scpll

El
4%
LA

49

\\

D
/

El
I

rudgement Note

No. 2C730-324E*A

Item

No.
GAIN

Procedure

- Display

B GAIN
Rotate the GAIN control
clockwise from the left end
to the right end and make
sure that the gain level and
gain display value of the
resulting ultrasound echo
image increase smoothly.
62 - 100 dB 2 dB step
M GAIN
Rotate the GAIN control from
the left end to the right end
in M-mode display to confirm
that the gain and its
indication for the ultrasound
echo image increase smoothly.
52 - 110 dB 1 dB step

STC

Activate those sliding


controls of the STC that
correspond to individual
depths and closely observe the
resulting displays made on the
monitor.
During these operations, check
that moving each sliding
control of the STC to the
right increases the
corresponding gain and that
moving each sliding control of
the STC to the left reduces
the corresponding gain.
Check that the range of gain
adjustment with those STC
controls changes according to
the selected probe frequency.

- 50 -

Remarks

Judgement

Note

No. 2C730-324E*A

(2) Display scale


No.
1

Procedure

Item
PAN/EXPAND
function check

. Display

Connect the specified probes to


probes A, then turn power on, and
set the
(1)

mode.

Turn on the

switch.

B-mode image PAN/EXPAND


Check that rotation of the
rotary switch causes the
displayed image to be
upscaled/downscaled with its
center as reference.
GIQ

Check that upward, downward,


rightward, and leftward
movement of the trackball
moves the center of the image
correspondingly.

- 51 -

Remarks

Judgement

Note

No. 2C730-324E*A

(3) Measurement
No.
1

functions

Procedure

Item
Function
panel

/+
w

- Display

Check that turning on the


4:

trackball function panel

displays measurement
keys as follows:
2

Measurement
related to Bmode images

2.1

Distance
measurement

Remarks

function entry

Carry out the operational checks


described below.

Requirement

(1) Push

,,;"
I-J

(2) Trackball
operations

Ultrasound image
display
Check that a CHl
starting point
marker is displayed
in the middle of the
monitor.
D: Distance (mm)
Check that the CHl
starting point
marker moves.

Switch lamp
which lights

(3) Push

0
EI

(4) Trackball
operations

(5) Push

Q
III

Check that the


starting point of
channel 1 is fixed
and that a CHl
ending point marker
is displayed above
that point.
Check that the CHl
ending point marker
moves. Also, check
that the two points
are connected by a
succession of dots
at 1 cm pitch and
that the distance
between those points
is displayed.
Check that the
ending point of
channel 1 is fixed
and that channel
identification
character "1" is
displayed above the
CHl starting point
marker.

- 52 -

D O.OMM

I
,

&**,

*MM

D 100 MM

JudgeNote
ment

No. X730-324E*A

No.

Item

Procedure

Requirement

(6) Operate
fQ@.

(7) Measurement menu


[Z] set

- Display

Ultrasound image
display
Check that the
channel identification character
moves clockwise or
counterclockwise
around the starting
point marker.
Note: Whether the
channel
identification
character is
displayed
depends on the
setting (ON or
OFF) of CH
DISPLAY in the
PRESET menu.
Check that a CH2
starting point
marker is displayed
above the starting
point of channel 1.

(8) Check that measurement of CH2 is


possible with the same procedure
as that of CHl and that the
results are displa.yed on the
right of the CH1 measurement
results.
(9) Check that setting [3] of the
measurement menu allows
measurement of CH3 and that the
results are displayed on the
right of the CH2 measurement
results.
(10) Check that setting [43 of the
measurement menu allows
measurement of CH4 and that the
results are displayed on the
right of the CH3 measurement
results.

- 53 -

Remarks

Switch lamp
which lights
+I

I/

s0100

JudgeNote
ment

No. 2C730-324E*A

(4) Body mark display


No.

Item

Body mark
display

1.1

Body mark
display

Procedure

Remarks

- Display

Perform the procedures below to


check the operation of body mark
display.
on the main panel to
display body marks on the
screen.
(2) Continue to press

the

desired body mark appears.


Alternatively, press the body
mark key on the full keyboard of
the main panel.

to select the desired


application.
1.2

Probe mark
movement,
angle
adjustment

(1) Use the left trackball to move


the probe mark to the desired
position.
n Q
(2) Use the
knob to rotate the

probe mark.

1.3

Posture
message
display

(1) Press

to display the body

mark menu.
(2) Press POSITION in the menu
several times and select the
desired posture comment.

- 54 -

JudgeNote
ment

No. X730-324E*A

No.

1.4

Item
Movement of
the display
positions of
body marks
and posture
messages

Procedure

- Display

(1) Select MARK MOVE in the menu


(The menu disappears).
(2) Use the trackball to move body
marks, probe marks, and posture
messages.

Only probe marks


may be moved.
1.5

Fixing body
mark
positions

to

fix the position.of

the body mark.


1.6

Deletion of
body marks

is pressed twice with the


body mark position fixed, the mark
will disappear.

- 55 -

Remarks

JudgeNote
ment

No. 2C730-324E*F

(5) Keyboard entry functions


No.
1

Procedure

Item
Character
display in
the ID area

Remarks

- Display

JudgeNote
ment

Enter the information for the


patient to be examined into the
fixed character area. While another
function which uses the keyboard
(measurement, registration, etc.) is
not active, keyboard entry to this
function can be performed.
The character display format is
shown below (There is no menu to fix
this function).
(1) Entry information
l

Up to 38 characters

11111222228s8884:~44466656
.:
02~4880248802468024080i#48802487
0
1
2

Area where the


cursor cannot
move (Note)

8
4

Free character
area

Fixed character area

---_-__-______-_____~~~~-~~~~~~~------

-------_-

81

Note:

Does not include the


fixed character area.

- 56 -

Area where the


cursor cannot
move

No. 2C730-324E*E

No.

Item
Keyboard
entry

Procedure

- Display

Remarks

<Using the full keyboard>

Advanced by one
character after
deleting the character
Returned by one
character after
deleting the character
Press to enter the
upper symbol on the
keys with upper/lower
symbols.

cl

(4)Gf

Moves the cursor to the


left, right, up or
down.

(5) DEL
a

Deletes all characters


in the area where the
cursor is displayed
(image or ID area).

El

6)

IDll

The cursor moves to the


X-coordinate of the
position at which
character entry starts,
and the Y coordinate
one column below.
(Example)

A
n

- 57 -

C [CR]

Judge.
Note
ment

No. 2C730-324E*A

(6) Initial setting functions


No.
1

Procedure

Item

Remarks

- Display

Judge
ment

Note

NEW PATIENT
Check that pressing the

'

switch

ii3ta

during any mode enters the system


into the preset status existing at
that time.

(7) Printing functions


Printing functions are described for the hardware units listed
below. Of all these units, only those which are to be connected
during installation must be checked for normal operation.
Camera unit (Sonoprinter)
. VCR
l

No.

1
1.1

Procedure

Item
Camera unit
(Sonoprinter)
Shutter

- Display

Remarks

Use the camera unit assembly check


sheet to carry out the checks listed
below.
(I) Change over the EXPOSURE INDEX
value to 1, 2, 3, etc. up to 9,
in that order, and then press
the shutter switch each time.
Visually check that the shutter
release time correspondingly
increases (removal of the film
pack allows you to see the
shutter release/lock mechanism).

Visual check

(2) Check that the shutter satisfies


the following requirements:
Mode
FREEZE

Requirement
The shutter must
correctly be
released/locked
according to the
selected EXPOSURE
INDEX (shutter
speed).
At the same time, a
buzzer must begin
sounding.

50

Shutter SW
Press the
PRINT A
switch or
the PRINT B
switch.

JudgeNote
ment

No. 2C730-324E*A

No.

Item

VCR
2
2.1 Remote
control

Procedure

Display

To record using the VCR, operate


the VCR directly to place it
into a status in which recording
is possible.

to start recording.

When &
IOII
recording
that each
recording

is pressed again,
is paused. Confirm
time it is pressed,
and pause alternate.

Perform the following procedures


to confirm that VCR playback
images-are displayed on the TV
monitor.

panel.
(b) Select [VCR] in the
displayed menu.
(c) Press the PLAY switch of the
VCR for playback.

59

Remarks

JudgeNote
ment

No. 2C730-324E*G

9.

LOCALIZED LABELS
The system is supplied with the following labels in 12 local languages,
Attach the labels in your local
language
at the appropriate positions by
referring to the corresponding English labels.
(The local language
labels should be attached above or beside the corresponding English
labels.)
Note:

When local language labels are attached above English labels


containing illustrations, care should be taken to avoid masking
the illustrations.

Label contents and drawing Nos.


(1) Warning label indicating that the system must not be used
in an atmosphere containing flammable gases

YNF5939

(2) Caution label describing precautions for transducer handling

YNF5940

(3) Caution label indicating that the operator must not lean
on the operating panel

YNF5942

(4) Label indicating the location of the circuit breaker


in the power supply section

YNF5950

(5) Label describing the procedures for resetting the circuit


breaker in the power supply section

YNF5950

(6) Label indicating the location of the AC power input section

YNF5950

(7) Label indicating the location of the power switch section

YNF5950

- 60 -

TOSHIBA

Y=OSHIBA

CQRPORATION

1385,SHIMOIS~IGAMI,OTAWARA-SHLTOCHIGI-KEN

324-8550,JAPAN