MERCURY CALIBRATOR
DESCRIPTION

SHEET

EID/130/DS/02
ITEM CODE
81I-BNNNCA
SPECIFICATION

1 of 1

MANUFACTURER

THERMO FISHER SCIENTIFIC, USA

MODEL NO

81i-BNNNCA

DISPLAY

LCD

MASS FLOW MEASUREMENT ACCURACY

REPEATABILITY OF MASS FLOW CONTROL

LINEARITY OF MASS FLOW MEASUREMENTS

+/- 2% SET POINT OR +/- 1% FS, WHICHEVER IS LESS FROM


-20 TO 100% FS
+/- 2% OF READING OR 1% OF FULL SCALE, WHICHEVER IS LESS
-20 TO 100% FS
+/- 0.5% FS

ZERO AIR FLOW CONTROLLER

20 SLM

SPAN GAS FLOW CONTROLLER

50 SCCM

HG SOURCE CONTROLLER

0 - 15 C

10

HG CONCENTRATION STD OUTPUT

3 - 50 g/dscm

11

LINEARITY

+/- 2% FS

12

OPERATING TEMPERATURE

5 - 40 C

13

SIGNAL OUTPUT

ANALOG OUT (4-20mA)

14

POWER REQUIREMENT

220 VAC 50HZ @ 250W

15

MOUNTING TYPE

ON 19 RACK

16

PHYSICAL DIMENSIONS

16.75(W) X 8.62(H) X 23(D)

17

APPROXIMATE WEIGHT

21.8 KG

18

MODEL DE-CODIFICATION
81I
B
N
N
N
C
A

MERCURY CALIBRATOR
= 220 VAC 50/60 Hz
= HYDRATOR (STANDARD)
= NOT AVAILABLE YET (FUTURE OPTION)
= NOT AVAILABLE YET (FUTURE OPTION)
= 6 CHANNEL 4-20mA CURRENT OUTPUT, (STANDARD)
= BENCH MOUNTING (STANDARD)

NOTE :

10/20/2011

ISSUED FOR APPROVAL

DATE

REV

DESCRIPTION

SS

MNS

PREPARED CHECKED

SIEMENS
APPROVED

Product Specifications

Mercury Calibrator, Model 81i


Precision source of mercury for automated calibrations and checks

Key Features

Performs standard calibration upstream of the


inertial filter

Capability to perform diagnostic calibrations


directly to analyzer and critical orifice

Long lasting Hg source

Uses Peltier cooler vapor pressure control


and mass flow control

Wide concentration range for post-dilution


calibration

Ideal for periodic converter efficiency and


linearity testing

The Thermo Scientific Mercury Calibrator,


Model 81i Mercury Calibrator is one of
the four major components of the Mercury
Freedom System. The Model 81i utilizes
a vapor generator that allows standard
calibration and dynamic spiking into the
extraction probe. A wide calibration range
lets you directly calibrate the analyzer at
post-dilution concentrations. The calibrator
is ideally suited for daily zero/span checks
and routine converter efficiency and
linearity testing.

The Model 81i also offers the benefits of


the iSeries platform features flexible
communications, increased serviceability,
and an easier to use interface.

Designed to meet the provisions of 40CFR


Parts 60 and 75, the Mercury Freedom
System provides a complete mercury
monitoring solution that reports elemental,
ionic, and total mercury in exhaust stacks
from both coal-fired boilers and waste
incinerators.

You can easily program short-cut keys to


allow you to jump directly to frequently
accessed functions, menus or screens.
The larger interface screen can display up
to five lines of measurement information
while the primary reporting screen remains
visible.

Ethernet connectivity provides efficient


remote access, allowing the user to
download measurement information directly
from the instrument without having to be
on site. Captive hardware and slide-apart
modules allows easy access for service and
periodic maintenance.

Hg Calibrator, Model 81i


To maintain optimal product performance, you need immediate access to experts worldwide, as well as priority status when your air quality equipment
needs repair or replacement. We offer comprehensive, exible support solutions for all phases of the product life cycle. Through predictable, xed-cost
pricing, our services help protect the return on investment and total cost of ownership of your Thermo Scientic air quality products.

Product Specifications
Gas Dilution:
Accuracy of each mass flow measurement
Linearity of mass flow measurement
Repeatability of mass flow measurement
Mass Flow Controller:
Zero Air Flow Controller
Span Gas Flow Controller
Hg Source Control
Hg Concentration Standard Output
Linearity
Operating Temperature
Power Requirements
Size and Weight
Outputs
Inputs

+2% of reading or 1% of full scale, whichever is less - 20 to 100% full scale


0.5% of full scale
+2% of reading or 1% of full scale, whichever is less - 20 to 100% full scale
20 slm
50 sccm
0O - 15O C
3 - 50 g/dscm
+/- 2 % full scale
5O - 40O C
250 watts
16.75"(W) x 8.62"(H) x 23"(D), 48 lbs.
425 mm (W) x 219 mm (H) x 584 mm (D), 21.8 kg
Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail
Indication (standard). 0-20 or 4-20 mA Isolated Current Output (optional)
16 Digital Inputs (standard), 8 0-10 VAC, Analog Inputs (optional)

Model 81i Hg Calibrator


V4

MFC 2
0-20 lpm

NO

NC

Zero air
to analyzer

NO
C

V3

Cal gas to probe


Zero air to probe

NC

Temp
control
CDA
(20-40 psig)

Bypass
solenoid
C

NC
V1

Regulator
MFC 1
Set to 25 psig 0-50 sccm

Hg source

V2

NO
NC

V5

NO

NC

Hg scrubber

Calibration gas
to analyzer
Exhaust

This specification sheet is for informational purposes only and is subject to change without notice. Thermo Fisher Scientific makes no warranties, expressed or implied, in this product summary.
2009 Thermo Fisher Scientific, Inc. All rights reserved Thermo Fisher Scientific, Inc.
This product is manufactured in a plant whose quality management system is ISO 9001 certified.
Lit_Model#AQI_01/09

Environmental
Instruments Division
Air Quality Instruments

You might also like