Professional Documents
Culture Documents
80225
Table of Contents
Introduction ......................................................................................................................................... 2
Alkalinity .............................................................................................................................................. 3
Bacteria, Acid-producing ............................................................................................................... 11
Bacteria, Heterotrophic Aerobic................................................................................................... 15
Bacteria, Slime-forming.................................................................................................................. 19
Bacteria, Sulfate-reducing ............................................................................................................. 23
Bacteria, Iron-related ...................................................................................................................... 27
Barium ................................................................................................................................................ 31
Boron .................................................................................................................................................. 37
Chloride .............................................................................................................................................. 43
Chloride, HR ...................................................................................................................................... 49
Conductivity ...................................................................................................................................... 55
Hardness, Calcium .......................................................................................................................... 61
Hardness, Total ................................................................................................................................ 69
Iron, Total ........................................................................................................................................... 77
pH ........................................................................................................................................................ 83
Sulfate................................................................................................................................................. 87
Sulfide................................................................................................................................................. 93
TPH (Total Petroleum Hydrocarbons) ......................................................................................... 99
Procedures Explained
Barium .................................................................................................................................................. 1
Hardness, Total and Calcium .......................................................................................................... 3
Iron ........................................................................................................................................................ 4
Conductivity and Total Dissolved Solids ..................................................................................... 6
Turbidity and Total Suspended Solids ....................................................................................... 10
Introduction
This manual is made up of test procedures and additional explanatory notes for testing of oil and gas field waters.
The first part of the manual contains the test procedure documents. Explanatory documents are found in the
second part of the manual.
Explanatory documents include information about the purpose of a test, recommended instrumentation,
interpreting test results, and information about test interferences and challenges.
Alkalinity
DOC316.53.01308
Method 10244
Digital Titrator
Scope and Application: For oil and gas field shale waters.
Test preparation
Before starting the test:
Four drops of Bromcresol Green-Methyl Red Indicator Solution1 can be substituted for the Bromcresol Green-Methyl Red
Indicator Powder Pillow.
Four drops of Phenolphthalein Indicator Solution1 can be substituted for the Phenolphthalein Indicator Powder Pillow.
For added convenience when stirring, use the TitraStir stirring apparatus1.
meq/L Alkalinity = mg/L as CaCO3 50
1
Quantity
each
Digital titrator
Graduated cylinder
Alkalinity
Page 3
Alkalinity
Test procedure
See
Table 1
1. Select a sample
volume and titration
cartridge from Table 1 on
page 5.
Note: Typical oil and gas
field water levels are 100
600 mg/L CaCO3.
Alkalinity
Page 4
4. Use a graduated
cylinder or pipet to
measure the sample
volume from Table 1 on
page 5.
digits x multiplier =
mg/L as CaCO3
P alkalinity
Example: 100 mL of
sample was titrated with
the 1.600 N cartridge and
250 digits were used to
reach the end point. The
concentration is 250 x 1.0
= 250 mg/L as CaCO3.
Alkalinity
Test procedure (continued)
See
Table 4
digits x multiplier =
mg/L as CaCO3
total alkalinity
Example: 100 mL of
sample was titrated with
the 1.600 N cartridge and
250 digits were used to
reach the end point. The
concentration is 250 x 1.0
= 250 mg/L as CaCO3.
Multiplier
1040
100
0.1600
0.1
40160
25
0.1600
0.4
100400
100
1.600
1.0
200800
50
1.600
2.0
5002000
20
1.600
5.0
10004000
10
1.600
10.0
Total alkalinity
Phenolphthalein alkalinity
pH 4.9
pH 8.3
pH 4.6
pH 8.3
pH 4.3
pH 8.3
pH 4.5
pH 8.3
pH 4.5
pH 8.3
pH 4.5
pH 8.3
Alkalinity
Page 5
Alkalinity
Interferences
Table 3 lists substances that can interfere with this test.
Interference level
Chlorine
Chlorine at levels above 3.5 mg/L may cause a yellow-brown color when the Bromcresol
Green-Methyl Red Powder Pillow is added. Add one drop of 0.1 N Sodium Thiosulfate to the
sample to remove chlorine before the test is started.
Color or turbidity
Do not filter, dilute or alter the sample. Color or turbidity can mask the color change of the
end point. Use a pH meter instead of the color indicators and titrate to a pH of 8.3 for
phenolphthalein acidity. For total alkalinity, refer to Table 2 on page 5 for the correct end point
pH.
Oils or solids may cover the pH probe and cause a slow or sluggish response. Clean the
probe immediately after use (refer to Maintenance of pH probes).
Maintenance of pH probes
To measure the complex oil and gas shale waters, probe maintenance is essential. Follow the
probe maintenance or cleaning procedures provided in the probe documentation.
Clean the probe when the following conditions occur:
Drifting/inaccurate readings
Calibration errors
Alkalinity
Collect samples in clean plastic or glass bottles. Fill completely and tighten the cap.
Prevent excessive agitation or prolonged exposure to air. Complete the test procedure as
soon as possible after collection for best accuracy.
The sample can be stored for 24 hours if cooled to 4 C (39 F) or below. If biological activity is
suspected, analyze the sample within 6 hours.
Alkalinity
Page 7
Alkalinity
The bicarbonate alkalinity equals 0 mg/L.
Check: (Refer to step 6.)
90 mg/L hydroxide alkalinity + 160 mg/L carbonate alkalinity + 0 mg/L bicarbonate alkalinity =
250 mg/L
The above answer is correct; the sum of the three alkalinity types equals the total alkalinity.
Sample result
Hydroxide alkalinity
equals:
Carbonate alkalinity
equals:
Bicarbonate alkalinity
equals:
Total Alkalinity
Phenolphthalein Alkalinity = 0
Total Alkalinity
Phenolphthalein Alkalinity
times 2
Total Alkalinity
Phenolphthalein Alkalinity
greater than one-half of Total
Alkalinity
2 times Phenolphthalein
Alkalinity minus Total
Alkalinity
Accuracy check
End point confirmation
Use a buffer pillow with the same pH as the end point with the indicator to make sure the end point
color is accurate.
Total alkalinityAdd 50 mL of deionized water to a flask. Add one pH 4.5 buffer powder
pillow and one Bromcresol Green-Methyl Red Indicator Powder Pillow and swirl to mix.
Use this solution for comparison during the titration with the sample.
Ampule breaker
Alkalinity
Page 8
Alkalinity
Summary of method
The sample is titrated with sulfuric acid to a colorimetric end point corresponding to a specific pH.
Phenolphthalein alkalinity is determined by titration to a pH of 8.3, as evidenced by the color
change of phenolphthalein indicator and indicates the total hydroxide and one half the carbonate
present. The M (methyl orange) or T (total) alkalinity is determined by titration to a pH between 4.3
and 4.9 and includes all carbonate, bicarbonate and hydroxide. Alternatively, to determine the total
alkalinity end points, use a pH meter and titrate to the specific pH required for the sample
composition.
Quantity/Test
Unit
Item no.
2271900
100/pkg
94399
100/pkg
94299
varies
each
1438801
varies
each
1438901
Quantity/Test
Unit
Item no.
each
1690001
each
50546
each
50838
each
50840
each
50841
each
50842
Unit
Item no.
16/pkg
1427810
Required apparatus
Description
Digital Titrator
Flask, Erlenmeyer, graduated, 250-mL
Graduated cylinderselect one or more based on range:
Recommended standards
Description
Alkalinity Standard Solution, Voluette Ampule 0.500 N Na2CO3, 10-mL
Alkalinity
Page 9
Alkalinity
Unit
Item no.
25/pkg
89568
25/pkg
89868
each
2095352
each
1970001
500 mL
27249
each
1451538
each
1451520
Water, deionized
each
1465100
each
2087079
100 mL MDB
2329232
100 mL MDB
16232
pH meter
each
each
1940000
each
1940010
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
06/2011, Edition 1
Bacteria, Acid-producing
DOC316.53.01328
Visual determination
APB-BART*
Semi-quantitative
Scope and Application: For the determination of acid-producing bacteria in brine solutions, produced waters and
hydraulic fracturing waters.
*APB-BART is a trademark of Droycon Bioconcepts Inc.
Test preparation
Procedure notes:
Do not touch or contaminate the inside of the tube or lid. Use aseptic technique.
To prevent contamination when the cap is removed from the inner tube, set the cap down directly on a clean surface. Do not
invert the cap.
Do not shake or swirl the tube after the sample is added. Let the ball float to the top at its own speed.
Test procedure
2. Collect at least 20 mL
of sample in the outer
tube.
Bacteria, Acid-producing
Page 11
Bacteria, Acid-producing
Interferences
If the original sample is acidic (pH < 6.0), neutralize the pH (pH 6.9 to 7.2) with sterile KOH. This
adjustment stresses the bacteria, so subtract 2 days from the Days to reaction in Table 1.
Water samples that contain more than 6% salt can give false negatives. Dilute all samples that
have more than 6% salt with sterile distilled water until the salt concentration is less than 6%.
Test results
Presence/Absence
When acid-producing bacteria are present, the color of the solution changes from a purple to a
yellow-orange color. The solution often becomes cloudy.
Aggressivity
800,000
High
70,000
High
9,000
High
1500
Moderate
500
Moderate
150
Moderate
<100
Low
<100
Low
Summary of method
When acid-producing bacteria are in the sample, the sample becomes acidic (pH 3.5 to 5.5) during
incubation. A pH indicator, bromocresol purple, in the APB-BART vial changes from a purple to a
orange or yellow color as the pH decreases. This change occurs at a pH of 5.2 to 5.8.
The acid-producing bacteria produce acids in very reductive (oxygen-free) environments. If
oxygen is present, then the APB do not generate acidity in the water, but can generate acidity at
the interface between the biofilm and the supporting material (e.g., concrete, steel).
Disposal
Sterilize the reacted sample before disposal. Refer to Figure 1.
Bacteria, Acid-producing
Page 12
Bacteria, Acid-producing
Figure 1 Disposal
Quantity/Test
Unit
Item no.
9/pkg
2831409
Bacteria, Acid-producing
Page 13
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
DOC316.53.01329
Visual determination
HAB-BART*
Semi-quantitative
Scope and Application: For the determination of total aerobic bacteria in brine solutions, produced waters and
hydraulic fracturing waters.
*HAB-BART is a trademark of Droycon Bioconcepts Inc.
Test preparation
Procedure notes:
Do not touch or contaminate the inside of the tube or lid. Use aseptic technique.
To prevent contamination when the cap is removed from the inner tube, set the cap down directly on a clean surface. Do not
invert the cap.
Do not shake or swirl the tube after the sample is added. Let the ball float to the top at its own speed.
Test procedure
2. Collect at least 20 mL
of sample in the outer
tube.
Test results
Presence/Absence
When heterotrophic aerobic bacteria are present, the color of the solution changes from a blue to a
light or medium yellow color. The solution often becomes cloudy.
Aggressivity
7,000,000
Very high
500,000
High
50,000
Moderate
7,000
Low
The color is
bleached from the
bottom to the top.
Aerobic bacteria
The color is
bleached from the
top to the bottom.
Summary of method
When heterotrophic aerobic bacteria (HAB) are in the sample, the bacteria consume oxygen
during incubation. When the oxygen is depleted, the bacteria react with the methylene blue dye in
the HAB-BART vial and reduce the dye to its colorless form. The faster the color change, the
higher the level of respiration and the larger or more aggressive the bacteria population.
Aerobic bacteria can cause several problems in water, including slime formations, turbidity, taste
and odor, corrosion, health risks and hygiene risks. When a problem is found, more tests are
recommended to give more information about the microbial problem. This method does not give
information about the particular groups of bacteria that can be present.
Disposal
Sterilize the reacted sample before disposal. Refer to Figure 2.
Quantity/Test
Unit
Item no.
9/pkg
2490409
27/pkg
2490427
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
Bacteria, Slime-forming
DOC316.53.01327
Visual determination
SLYM-BART*
Semi-quantitative
Scope and Application: For the determination of slime-forming bacteria in brine solutions, produced waters and
hydraulic fracturing waters.
*SLYM-BART is a trademark of Droycon Bioconcepts Inc.
Test preparation
Procedure notes:
Do not touch or contaminate the inside of the tube or lid. Use aseptic technique.
To prevent contamination when the cap is removed from the inner tube, set the cap down directly on a clean surface. Do not
invert the cap.
Do not shake or swirl the tube after the sample is added. Let the ball float to the top at its own speed.
Test procedure
2. Collect at least 20 mL
of sample in the outer
tube.
Bacteria, Slime-forming
Page 19
Bacteria, Slime-forming
Test results
Presence/Absence
When slime-forming bacteria are present, the solution becomes cloudy. Refer to Figure 1.
Figure 1 Negative versus positive test results
Negative (absent/non-aggressive)
Positive (present/aggressive)
Aggressivity
1,800,000
Very high
350,000
High
66,500
High
12,500
Moderate
2500
Moderate
500
Moderate
100
Low
10
Low
Bacteria, Slime-forming
Page 20
Cloudy growth or
layered platesslimeforming bacteria
Blackened liquid
pseudomonads
and enterics
Thread-like
strands
tight slime
bacteria
Bacteria, Slime-forming
Summary of method
When slime-forming bacteria are in the sample, one or more types of slime will grow in the SLYMBART vial during incubation. The slime is typically seen as a cloudy or gel-like growth, which can
be in one location or occur throughout the sample. These growths are usually white, grey, yellow
or beige in color and can darken over time. Slime-forming bacteria typically produce the thickest
slime under aerobic (oxidative) conditions, which can be seen around the floating ball.
Iron-related bacteria also produce slime, but it is typically thinner and involves the accumulation of
various forms of iron. Slime-forming bacteria can make large amounts of slime without iron.
Disposal
Sterilize the reacted sample before disposal. Refer to Figure 3.
Figure 3 Disposal
Quantity/Test
Unit
Item no.
9/pkg
2432509
27/pkg
2432527
Bacteria, Slime-forming
Page 21
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
Bacteria, Sulfate-reducing
DOC316.53.01326
Visual determination
SRB-BART*
Semi-quantitative
Scope and Application: For the determination of sulfate-reducing bacteria in brine solutions, produced waters
and hydraulic fracturing waters.
*SRB-BART is a trademark of Droycon Bioconcepts Inc.
Test preparation
Procedure notes:
Do not touch or contaminate the inside of the tube or lid. Use aseptic technique.
To prevent contamination when the cap is removed from the inner tube, set the cap down directly on a clean surface. Do not
invert the cap.
Do not shake or swirl the tube after the sample is added. Let the ball float to the top at its own speed.
SRB grow predominantly deep within biofilms and not directly in water. Make sure to get a representative sample.
Test procedure
2. Collect at least 20 mL
of sample in the outer
tube.
Bacteria, Sulfate-reducing
Interferences
If the sample contains more than 20 ppm hydrogen sulfide (H2S), the test can give a false positive.
To remove hydrogen gas from the sample, add 30 mL of sample to the outer tube, cap and shake
for 10 seconds. Let stand for 20 seconds.
Test results
Presence/Absence
When sulfate-reducing bacteria are present, a black slime forms in the tube. Refer to Figure 1.
Figure 1 Negative versus positive test results
A black slime
ring forms
around the ball
and/or there is
a black slime
growth at the
bottom of the
tube.
The solution
has no black
slime.
Negative (absent/non-aggressive)
Positive (present/aggressive)
Aggressivity
6,800,000
Very high
700,000
High
100,000
High
18,000
Moderate
5000
Moderate
1200
Moderate
500
Moderate
200
Low
Bacteria, Sulfate-reducing
Page 24
Cloudy solution
anaerobic bacteria
Bacteria, Sulfate-reducing
Summary of method
When sulfate-reducing bacteria are in the sample, sulfate is reduced to hydrogen sulfide (H2S) in
the SRB-BART vial during incubation. The H2S reacts with the ferrous iron in the test vial to form
black iron sulfides. This sulfide commonly forms either in the base as a black precipitate and/or
around the ball as an irregular black ring.
SRB tend to grow in anaerobic conditions deep within biofilms (slimes) as a part of a microbial
community. SRB may not be present in the free-flowing water over the site of the fouling. Sulfatereducing bacteria can cause problems such as strong odors, blackening of equipment, slime
formations and the start of corrosive processes.
Disposal
Sterilize the reacted sample before disposal. Refer to Figure 3.
Figure 3 Disposal
Quantity/Test
Unit
Item no.
9/pkg
2432509
27/pkg
2432527
Bacteria, Sulfate-reducing
Page 25
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
Bacteria, Iron-related
DOC316.53.01325
Visual determination
IRB-BART*
Semi-quantitative
Scope and Application: For the determination of iron-related bacteria in brine solutions, produced waters and
hydraulic fracturing waters.
*IRB-BART is a trademark of Droycon Bioconcepts Inc.
Test preparation
Procedure notes:
Do not touch or contaminate the inside of the tube or lid. Use aseptic technique.
To prevent contamination when the cap is removed from the inner tube, set the cap down directly on a clean surface. Do not
invert the cap.
Do not shake or swirl the tube after the sample is added. Let the ball float to the top at its own speed.
IRB grow predominantly on surfaces and not directly in water. Make sure to get a representative sample.
Test procedure
2. Collect at least 20 mL
of sample in the outer
tube.
Bacteria, Iron-related
Test results
Presence/Absence
When iron-related bacteria are present, a foam or a brown slime ring forms around the ball and/or
a brown slime forms at the bottom of the tube. Refer to Figure 1.
Figure 1 Negative versus positive test results
Foam or a brown
slime ring forms
around the ball
and/or there is a
brown slime growth
at the bottom of the
tube.
Negative (absent/non-aggressive)
Positive (present/aggressive)
Aggressivity
540,000
Very high
140,000
High
35,000
High
9,000
Moderate
2300
Moderate
500
Moderate
150
Moderate
25
Low
Foam around
ballanaerobic
bacteria
Bacteria, Iron-related
Page 28
Green cloudy
pseudomonads
Red
cloudy
enteric
bacteria
Cloudy
heterotrophic
bacteria
Black solution
pseudomonads
and enterics
Bacteria, Iron-related
Summary of method
When iron-related bacteria are in the sample, a series of reactions occur in the redox and nutrient
gradients that develop in the IRB-BART vial during incubation. The IRB use the nutrients and ferric
iron in the vial to grow and can be seen as foam, clouding, slime and/or color changes.
The bacteria that can be seen in this test include iron oxidizing and reducing bacteria, the
sheathed iron bacteria, Gallionella, pseudomonads and enteric bacteria. Positive results can be a
possible cause of biofouling problems such as plugging, corrosion, cloudiness and color.
Disposal
Sterilize the reacted sample before disposal. Refer to Figure 3.
Figure 3 Disposal
Quantity/Test
Unit
Item no.
9/pkg
2432309
27/pkg
2432327
Bacteria, Iron-related
Page 29
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
Barium
DOC316.53.01311
Turbidimetric Method1
Method 10251
Powder Pillows
Adapted from Snell and Snell, Colorimetric Methods of Analysis, Vol. II, 769 (1959).
Test procedure
1. Push PRGM.
The display shows
PRGM ?
Initial setup: go to
Instrument Setup on
page 33 to add the
program to the
instrument.
7. Push ZERO.
The cursor moves to the
right, then the display
shows:
0 mg/L Ba and LR, MR
or HR
Barium
Page 31
Barium
9. Push
TIMER ENTER
A 5-minute reaction
period starts.
Do not move the sample
cell during the reaction
period.
Interferences
Known interferences are shown in Table 1. The interference levels are applicable to an undiluted
10-mL sample. The interference levels increase proportionally as the sample is diluted.
Table 1 Interferences
Substance
Interference Level
Calcium
Sodium chloride
Magnesium
Silica
Strontium
The interference level is dependent on the sample matrix and the barium concentration. When the
barium concentration is zero, there is no interference from strontium. The best results occur when the
barium concentration is less than 20 mg/L and when the strontium concentration (as mg/L) is equal to
or less than the barium concentration.
Turbidity
Accuracy Check
Standard Additions Method
Use the standard additions method to validate the test procedure, reagents and instrument and to
find if there is an interference in the sample.
1. Fill three sample cells with sample as specified in steps 4 and 5 of the test procedure.
2. Use a TenSette Pipet to add 0.1, 0.2 and 0.3 mL of a 1000 mg/L Barium Standard Solution to
the sample cells. Mix fully.
3. Complete the test procedure for each sample.
Barium
Page 32
Barium
4. Review the results. The barium concentration should increase by 10 mg/L for Ba LR,
100 mg/L for Ba MR or 1000 mg/L for Ba HR for each 0.1 mL of standard that is added.
5. If the concentration does not increase by the correct amount, refer to Standard Additions in
Section 1 of the procedures manual.
Standard Solution Method
Use a 50 mg/L Barium Standard Solution to validate the test procedure, reagents and the
instrument. Select the Ba LR test range in step 3 and use 10 mL of the standard solution instead of
the sample in step 4. To adjust the result, refer to Standard Adjust.
To prepare this standard solution, add 5.0 mL of a 1000 mg/L Barium Standard Solution to a 100mL volumetric flask. Dilute to the mark with deionized water and mix fully.
Standard Adjust
The standard adjust option is recommended when program 125 is used.
1. Measure the concentration of a 50 mg/L Barium Standard Solution. Select the Ba LR range
and use 10 mL of the standard solution. Keep the sample cell in the instrument.
2. Push the SETUP key and use the arrow keys to scroll to the STD option.
3. Push ENTER.
4. Push the numbers 50 to make the instrument read the value of the standard solution
concentration.
5. Push ENTER to complete the adjustment.
Note: The MR and HR calibration curves are adjusted proportionally when the Ba LR calibration curve is
adjusted. Refer to Section 1, Standard Curve Adjustment of the procedures manual.
Method Performance
Precision
In a single laboratory, with a 70 mg/L barium standard solution, two representative lots of powder
pillows and the instrument, a single operator got a standard deviation of 1.2 mg/L barium.
Estimated Detection Limit (EDL)
The EDL for program 125 Ba LR is 2 mg/L Ba. For more information on derivation and use of the
estimated detection limit, refer to Section 1 of the procedures manual.
Instrument Setup
This procedure adds program 125 to a DR/820, DR/850 or DR/890 instrument.
1. Push the ON key to turn on the instrument.
2. Push the SETUP key.
3. Push the down arrow key until the prompt line shows USER.
4. Push the ENTER key.
5. Push the numbers 8138, then push ENTER.
6. Refer to Table 2. Find the number from the Enter column that corresponds to Line Number 1
on the display. Push these numbers on the keypad, then push ENTER. Continue to add the
numbers that correspond to each line number on the display.
Note: Use the arrow keys to scroll and review or change numbers at any time.
Barium
Page 33
Barium
Table 2 Instrument setup
Line Number
Enter
Line Number
Enter
125
29
82
24
30
66
72
31
97
32
32
33
72
34
82
35
65
36
32
37
10
38
11
39
66
12
66
40
200
13
169
41
14
125
42
15
112
43
16
44
88
17
45
18
46
19
47
30
20
66
48
21
97
49
44
22
32
50
23
76
51
24
82
52
25
66
53
26
97
54
25
27
32
55
28
77
56
255
Summary of Method
Barium ions in the sample react with the BariVer 4 Barium Reagent to make a barium sulfate
precipitate, which is held in suspension by a protective colloid. The amount of precipitate is
proportional to the barium concentration. The BariVer 4 Method is especially useful for brines and
produced waters where both barium and sulfate are in the sample and precipitation cannot be
started by the addition of more sulfate.
Barium
Page 34
Barium
Units
Item No.
1 pillow
100/pkg
1206499
6/pkg
2401906
OPTIONAL REAGENTS
Description
Barium Standard Solution, 1000 mg/L
Water, deionized
Units
Item No.
100 mL
1461142
4L
27256
OPTIONAL APPARATUS
Description
Units
Item No.
2088638
100/pkg
189457
1457442
Funnel, poly, 65 mm
108367
1970001
50/pkg
2185696
1970010
50/pkg
2199796
1451537
1465100
Barium
Page 35
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
09/2011, Edition 1
Boron
DOC316.53.01309
Carmine Method
Method 10252
0 to 50 mg/L B
Powder Pillows
Test procedure
1. Push: PRGM
The display shows:
PRGM ?
Initial setup: go to
Instrument Setup on
page 39 to add the
program to the
instrument.
4. Use a 100-mL
graduated cylinder to
measure 75 mL of
concentrated sulfuric acid.
Add the acid to a 250 mL
plastic Erlenmeyer flask.
Notice! Prepare the
solution in a wellventilated area or use a
fume hood to prevent
injury.
Boron
Page 37
Boron
Reagent preparation
Mix one BoroVer 3 Reagent Powder Pillow per 75 mL of concentrated sulfuric acid. Add the
powder pillow while stirring. The preparation of this solution generates gaseous HCl when the
indicator pillow is added to the sulfuric acid. Prepare the solution in a well-ventilated area or use a
fume hood to prevent injury. This solution is stable for up to 48 hours if stored in plastic containers.
Do not store in borosilicate glassware (Pyrex or Kimax) for more than one hour. The solution
may pull boron from these containers.
New vials can contain residual amounts of reactive boron from the glass manufacturing process.
For best results, fill the vials with 34 mL of the prepared BoroVer 3 Reagent for 30 minutes.
Discard the reacted solution. Flush and dry the vials for use in other analysis.
Boron
Page 38
Boron
The BoroVer 3/Sulfuric Acid solution is very acidic. Make sure that the solution is neutral (pH 69)
before disposal. Refer to the current material safety data sheets (MSDS) for safety protocols and
disposal information.
Accuracy Check
Standard Solution Method
Use a 30 mg/L Boron Standard Solution to validate the test procedure, reagents and the
instrument.
1. To prepare the standard solution, add 3.0 mL of a 1000 mg/L Boron Standard Solution to a
100-mL volumetric flask. Add deionized water to the mark, stopper and mix fully.
2. Use the 30 mg/L Boron Standard Solution as an alternative to the sample in step 7 of the
procedure.
Standard Adjust
Use the reading with the 30 mg/L Boron Standard Solution to adjust the calibration curve.
1. Push the SETUP key and use the arrow keys to scroll to the STD option.
2. Push ENTER.
3. Push the numbers 30 to measure the value of the standard solution concentration.
4. Push ENTER to complete the adjustment. Refer to Section 1 Standard Adjust of the
procedures manual for more information.
Method Performance
Precision
In a single laboratory, with a standard solution of 25 mg/L boron and two representative lots of
reagent with the instrument, a single operator obtained a standard deviation of 0.8 mg/L boron.
Estimated Detection Limit
The estimated detection limit for program 126 is 2.2 mg/L B. For more information on derivation
and use of the estimated detection limit, refer to Section 1 of the procedures manual.
Summary of Method
Boron reacts with a mixture of carminic acid and sulfuric acid and then shows a reddish to bluish
color. The amount of color is directly proportional to the boron concentration.
Instrument Setup
This procedure adds the program 126 to a DR/850 or DR/890 instrument.
1. Push ON to turn on the instrument.
2. Push SETUP.
3. Push the down arrow until the prompt line shows USER and push ENTER.
4. Push the numbers 8138, then push ENTER.
5. Refer to Table 1. Find the number from the Enter column that corresponds to Line Number 1
on the display. Push these numbers on the keypad, then push ENTER. Continue to add the
numbers that correspond to each line number on the display.
Note: Use the arrow keys to scroll and review or change numbers at any time.
Boron
Page 39
Boron
Table 1 Instrument setup
Boron
Page 40
Line Number
Enter
Line Number
Enter
1
2
126
29
42
30
72
31
32
33
34
35
64
36
117
37
10
40
38
11
140
39
12
65
40
13
211
41
14
79
42
15
222
43
16
44
55
17
45
128
18
46
19
47
20
20
66
48
21
32
49
22
72
50
23
82
51
24
52
25
53
26
54
143
27
55
28
56
255
Boron
Unit
Item No.
1pp/10 tests
100/pkg
1417099
75mL/20 tests
500 mL
97949
Water, deionized
100 mL
27242
6/pk
2275806
6/pk
2350206
2089846
108142
Pipet, 0.21.0 mL
BBP078
100/pk
BBP079
Pipet, 1.05.0 mL
BBP065
75/pk
BBP068
1970001
50/pk
2185696
OR
1970010
50/pk
2199796
OPTIONAL REAGENTS
Description
Unit
Item No.
100 mL
191442
Sodium Hydroxide, 6 N
1000 mL
2332453
Unit
Item No.
1000/pk
2185628
250/pk
2199725
1864100
LZP320
2927902
pair
2410105
6/pk
2930004
188/pk
293280
OPTIONAL APPARATUS
Description
Boron
Page 41
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
09/2011, Edition 1
Chloride
DOC316.53.01306
Method 10246
Digital Titrator
Scope and Application: For gas and oil field shale waters.
Test preparation
Before starting the test:
mg/L sodium chloride = mg/L chloride x 1.65
meq/L chloride = mg/L chloride / 35.45
For added convenience when stirring, use the TitraStir stirring apparatus.
Quantity
Digital titrator
Water, deionized
4 liters
250
0.1
125
0.2
50
0.5
Chloride
Page 43
Chloride
Table 1 Guidelines to estimate sample sizes (continued)
Number of digits
25
1.0
10
2.0
5.0
20
50
See
Table 1
1. Select a sample
volume from Table 2 on
page 45. Refer to
Determine sample sizes
on page 43.
4. Use a TenSette
graduated cylinder or pipet
to measure the sample
volume from Table 2 on
page 45 into a 250 mL
Erlenmeyer flask.
Chloride
Page 44
digits x multiplier =
mg/L Cl
Example: 1.0 mL of
sample was titrated with
the 1.128 N cartridge and
200 digits were used to
reach the end point. The
concentration is 200 x 50 =
10,000 mg/L Cl.
Chloride
Multiplier
100400
50
1.128
1.0
2501000
20
1.128
2.5
10004000
1.128
10.0
250010,000
1.128
25.0
500020,000
1.128
50
10,00040,000
0.5
1.128
100
25,000100,000
0.2
1.128
250
50,000200,000
0.1
1.128
500
Technique tips
Demineralized water or other sources of chloride-free water may be used in place of the
deionized water.
The TitraStir stirring apparatus can provide a more convenient or reproducible stirring
technique.
If the precipitate formed is red or orange, but the solution color is yellow, the test will give low
results. Greater agitation or swirling is required during the titration. The test should be
repeated. To eliminate the red or orange precipitate formation:
Interferences
Table 3 lists substances that can interfere with this test.
Table 3 Interfering substances
Interfering substance
Interference level1
Bromide
Cyanide
Iron
Iodide
Orthophosphate
pH
Chloride
Page 45
Chloride
Table 3 Interfering substances (continued)
Interfering substance
Interference level1
Sulfide
Sulfite
Concentrations above 10 mg/L interfere with this method. To eliminate sulfite interference,
add three drops of Hydrogen Peroxide, 30%, to the sample before the test is started.
Accuracy check
Use the standard additions method to determine whether the sample has an interference and
confirm the analytical technique.
Required items for accuracy check:
Ampule breaker
Chloride
5. Titrate the prepared solution from yellow to red-brown. The end point should be 250 digits
25 digits.
6. Calculate the actual standard concentration for a 1 mL sample size (refer to
Table 2 on page 45).
Digits required x digit multipliers = mg/L Cl(Example: 250 digits x 50 multiplier = 12,500 mg/L Cl-)
Summary of method
The sample is titrated with Silver Nitrate Standard Solution in the presence of potassium chromate
(from the Chloride 2 Indicator Powder). The silver nitrate reacts with the chloride present to
produce insoluble white silver chloride. After all the chloride has been precipitated, the silver ions
react with the excess chromate present to form a red-brown silver chromate precipitate, and marks
the end point of the titration.
Quantity/Test
Unit
Item no.
2288000
1 pillow
50/pkg
105766
varies
each
1439701
Quantity/Test
Unit
Item no.
Digital Titrator
each
1690001
each
50546
each
50838
each
50840
each
50841
Required apparatus
Description
each
50842
each
1720500
each
1970001
Pipet tips
50/pkg
2185696
Unit
Item no.
16/pkg
1425010
2196800
Recommended standards
Description
Chloride Standard Solution, Voluette Ampule, 12,500-mg/L Cl, 10-mL
Voluette breaker
Chloride
Page 47
Chloride
Unit
Item no.
100/pkg
69257
each
108367
473 mL
14411
100 mL MDB
245032
each
2095352
100/pkg
241899
100 mL MDB
244932
each
1940000
each
1940010
4L
27256
Water, deionized
pH Test Strip, 014 pH
100/pkg
2601300
Pipet tips
1000/pkg
2185628
500 mL
18349
Sampling bottle
250 mL
2087076
Dropper, glass
5/pkg
1419705
each
19700-10
each
4157800
50/pkg
2199796
250/pkg
2199725
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
05/2011, Edition 1
Chloride, HR
DOC316.53.01322
Method 10255
Powder Pillow ISA
Scope and Application: For the determination of high concentrations (1 M) of chloride in brine solutions,
produced waters and hydraulic fracturing waters.
Test preparation
Procedure notes:
The instructions in this procedure are specific to the HQd meters. The SensION+ meters can also be used but the navigation
and menus will be different.
Refer to the meter user manual for meter operation. Refer to the probe user manual for probe maintenance.
Set the date and time in the meter before the probe is attached. The probe must have the correct service-life time stamp.
If complete traceability is necessary, enter a sample ID and operator ID. Refer to the HQd meter manual
for more information.
Calibrate the probe regularly for the best measurement accuracy. Refer to Calibrate the probe.
Stir the standards at a slow and steady rate to prevent the formation of a vortex.
Air bubbles under the probe tip can cause slow response or measurement errors. Gently shake the probe to remove
bubbles.
Keep the temperature of the calibration standards within 2 C for best results.
Between measurements, rinse the probe with deionized water. Blot dry with a lint-free cloth.
Items to collect:
Description
Quantity
Deionized water
varies
1
11.55 g
Stir bars
Stirrer, electromagnetic
Chloride, HR
Page 49
Chloride, HR
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
3. Put the probe in a 3.55 g/L chloride standard solution for a minimum of 30 minutes. Refer to
Prepare the calibration standards.
Unitsg/L
5. Go to Calibration Options.
6. Set the calibration options for high-range measurements:
Std setcustom
Calibration unitsg/L
Laboratory balance
Deionized water
Chloride, HR
b. Dilute to the mark with deionized water. Mix fully.
1. Add 25 mL of the
3.55 g/L, 12.5 g/L and
35 g/L chloride standard
solution to three beakers.
4-7
5. Put the probe into the
3.55 g/L chloride standard
solution. Tap the probe to
remove any air bubbles.
Chloride, HR
Page 51
Chloride, HR
Measure samples
Dilute samples that are greater than 35 g/L
If the chloride concentration is greater than 35 g/L (1 M), dilute the sample to a lower
concentration. Complete the steps that follow to make a 1:10 (10-fold) dilution.
1. Measure 2.5 mL of the sample in a 25-mL graduated cylinder.
2. Add deionized water to the 25-mL line.
3. Pour the diluted sample into a beaker.
4. Measure the concentration with the HR chloride measurement procedure.
5. Multiply the result by 10 to get the concentration of the sample before dilution.
1. Add 25 mL of sample
to a 50-mL beaker.
1-6
5. Put the probe into the
sample. Tap the probe to
remove any air bubbles.
Chloride, HR
Page 52
7. Repeat steps 1
8. When done, rinse the
through 6 for each sample. probe.
Chloride, HR
Interferences
The sensing element can respond to other ions in addition to chloride and cause a positive error. If
Chloride ISA is added to the standards and samples, the effect of interfering ions is minimized.
Samples can contain oxidizing agents such as Copper (Cu2+), Iron (ferrous) (Fe2+) and
Permanganate (MnO4-). Refer to Table 1.
Interference
Mercury
Can deposit a layer of salt on the sensing surface and cause probe errors.
Quantity/Test
Unit
Item no.
100/pkg
2318069
108041
ISECL18101
Recommended standards
Description
Unit
Item no.
454 g
18201H
Unit
Item no.
62011
108140
4L
27256
ISECL18103
LZW9652C.97.002
1457445
Deionized water
IntelliCAL Chloride Ion Selective Electrode (ISE), 3 m cable
sensION+ 9652C Chloride Combination Ion Selective Electrode (ISE), 1 m
Volumetric flask, 200 mL, Class A, glass
1
cable1
Use with sensION+ meters or other meters that have a BNC connection.
Chloride, HR
Page 53
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
05/2012, Edition 1
Conductivity
DOC316.53.01324
Method 10256
Conductivity Meter
Test preparation
Meter
Standard probe
Rugged probe1
CDC40101, CDC40103
Items to collect:
Description
Quantity
Conductivity
Page 55
Conductivity
Items to collect: (continued)
Description
Quantity
500 mL
500 mL
500 mL
500 mL
500 mL
500 mL
500 mL
Conductivity
Conductivity
Page 56
Conductivity
Conversions
Table 2 shows the conversions to change the readings on the display to other conductivity units.
From
To
mS/cm
S/cm
mS/cm 1000
S/cm
mS/cm
S/cm 0.001
S/cm
mhos/cm
S/cm 1
mS/cm
mmhos/cm
mS/cm 1
S/cm
mg/L TDS
S/cm 0.641
g/L TDS
mg/L TDS
mS/cm
g/L TDS
mS/cm 0.64
mg/L TDS
g/L TDS
mg/L TDS
gpg TDS
g/L TDS
gpg TDS
S/cm
ohms cm
1,000,000 S/cm
mS/cm
ohms cm
1,000 mS/cm
TDS is an empirically-derived value from the conductivity measurement. A value of 0.64 is selected here for simplicity and suitability to oil and
gas field waters.
Table 3 shows typical temperature correction values for selected solutions from the linear
temperature correction option.
Percent per C
Ultrapure water
4.55
Salt (NaCl)
2.125
NaOH
1.72
Dilute ammonia
1.8810
10% HCl
1.325
5% sulfuric acid
0.9698
Interferences
To remove the conductivity from hydroxide ions, neutralize the pH of the sample:
a. Add 4 drops of phenolphthalein indicator solution to 50 mL of sample. The solution
becomes pink.
b. Add gallic acid solution by drops until the pink color completely goes away.
c. Measure the conductivity.
For more information on conductivity measurements in oil and gas field waters, refer to the
Conductivity and Total Dissolved Solids Procedures Explained section of the Hydraulic
Fracturing Water Analysis Handbook.
Conductivity
Page 57
Conductivity
Accuracy check
Measure the conductivity of a sodium chloride (NaCl) standard solution that has a similar value to
the conductivity of typical samples. The measured conductivity should be close to the known value
of the standard solution. If the value is outside of the accuracy limits on the standard solution label,
complete a calibration with this standard solution. Refer to the meter user manual.
Method performance
The accuracy of a conductivity measurement depends on many factors that are associated with
the overall conductivity system, which includes the meter, electrode and calibration solutions.
Refer to the documentation for the electrode or the meter for more information.
Summary of method
Electrolytic conductivity is the movement of ions in a solution, which makes an electrical current
and is the reciprocal of the solution resistivity. The ions come from inorganic dissolved solids (e.g.,
chloride, nitrate, sulfate and phosphate anions and sodium, calcium, magnesium, iron and
aluminum cations). Organic material such as oils, phenols, alcohols and sugars do not have
enough conductivity for a useful estimate of concentration.
Conductivity meters measure the resistance that occurs in an area of the solution that is defined
by the physical design of the probe. A voltage is applied between the electrodes, and the voltage
drop caused by the resistance of the solution is used to calculate the conductivity per centimeter.
The basic unit of measure for conductivity is the Siemen (or mho), which is the reciprocal of the
ohm. Other common units for aqueous solutions are milliSiemens/cm (103 S or mS/cm) and
microSiemens/cm (106 S or S/cm).
Unit
Item no.
HQ40d meter
HQ40d53000000
HQ30d meter
HQ30d53000000
HQ14d meter
HQ14d53000000
CDC40101
CDC40103
CDC40105
CDC40110
CDC40115
CDC40130
Unit
Item no.
100 mL
2307542
100 mL
1440042
100 mL
210542
Recommended standards
Description
NaCl conductivity standards:
Conductivity
Page 58
Conductivity
Recommended standards
Description
Unit
Item no.
100 mL
2307442
500 mL
C20C250
500 mL
C20C270
500 mL
C20C280
500 mL
S51M001
500 mL
S51M002
500 mL
S51M003
500 mL
S51M004
Unit
Item no.
108042
50 mL SCDB
1442326
500 mL
88449
15 mL SCDB
16236
62014
4L
27256
Conductivity
Page 59
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
Barium, 8014
Hardness, Calcium
DOC316.53.01318
Method 10253
Digital Titrator
Test preparation
Before starting the test:
Magnesium is not included in the results but must be present for a sharp endpoint. If magnesium is not present, add one to
two drops of Magnesium Standard Solution, 10-g/L as CaCO3 to the sample before the test is started.
mg/L Ca = Ca hardness in mg/L as CaCO3 x 0.40
A 0.1-g scoop of CalVer 2 Calcium Indicator Powder can be used in place of the CalVer 2 Calcium Indicator Powder Pillow.
If samples cannot be analyzed immediately, refer to Sample collection, preservation and storage. Adjust the pH of preserved
samples before analysis.
For added convenience, use the TitraStir stir plate1.
1
Quantity
1 pillow
1 bottle
1 cartridge
Digital titrator
Graduated cylinder
Water, deionized
Erlenmeyer flask, 250-mL
4L
1
Hardness, Calcium
Page 61
Hardness, Calcium
Test procedure
See
Table 1
or
Table 2
1. Select a sample
volume and titration
cartridge from Table 1.
4. Use a graduated
cylinder or pipet to
measure the sample
volume identified in step 1.
Put the sample in a clean,
250-mL Erlenmeyer flask.
If the sample volume is
less than 100 mL, dilute to
about 100 mL with
deionized water.
Hardness, Calcium
Page 62
Hardness, Calcium
Multiplier
100400
100
200800
50
5002000
20
10004000
10
10
20008000
20
500020,000
50
10,00040,000
100
20,00080,000
0.5
200
50,000200,000
0.2
500
200
0.2
100
0.5
50
25
10
10
20
Hardness, Calcium
Page 63
Hardness, Calcium
Samples can be used for up to 6 months if preserved and kept at room temperature. Preserve
samples only when immediate analysis is not possible.
To preserve the sample:
1. Add 1.5 mL of nitric acid per 1 liter (1 quart) of sample to the sample. Mix fully.
2. Measure the sample pH.
3. If the sample pH is greater than 2, add more nitric acid in 0.5-mL increments. Mix fully and
check the pH of the sample after each addition until the sample pH is 2 or less.
Before analysis:
1. Add Potassium Hydroxide Standard Solution to the preserved sample in increments. Mix fully
and check the pH of the sample after each addition until the sample pH is 7.
2. Make a volume correction for the nitric acid and hydroxide added.
Total volume = sample + acid + hydroxide
Volume correction = (total volume) divided by (sample volume)
Corrected test result = (test result) x (volume correction)
Interferences
WARNING
Chemical hazard. Potassium cyanide is toxic. Always add potassium cyanide after the
potassium hydroxide. Obey local hazardous waste regulations for disposal of all cyanidecontaining waste.
An interfering substance can prevent the color change at the titration endpoint. A dilution can often
reduce the interference to a level at which the substance does not interfere. If an interference is
suspected, decrease the sample volume, dilute to about 100 mL and do the test again. Table 3
shows substances that can interfere with this test.
Table 3 Interfering substances
Substance
Acidity
Alkalinity
Aluminum
Causes a slow endpoint but up to 200 mg/L aluminum can be tolerated if sufficient time is
given for the color change.
Barium is a by-product from the drilling process and can be present in very high
concentrations in process and flow back waters. The barium will be titrated directly with the
calcium titration.
Barium
If strontium and calcium are not present in the sample, the barium will precipitate at pH 13
similar to magnesium. However, most produced and flow back water samples have
strontium, calcium and barium present at high concentrations. If the barium concentration is
known, it can be subtracted from the calcium hardness by converting the mg/L Ba
concentration to mg/L as CaCO3 by multiplying the Ba concentration by 0.729.
Chloride
Saturated solutions do not give a distinct endpoint. The test can be run directly in sea water.
Cobalt
Interferes at all levels. 0.5 grams of potassium cyanide can be added after the potassium
hydroxide solution to remove interference from up to 20 mg/L cobalt.
Copper
Interferes at 0.1 mg/L copper. 0.5 grams of potassium cyanide can be added after the
potassium hydroxide solution to remove interference from up to 100 mg/L copper.
Hardness, Calcium
Page 64
Hardness, Calcium
Table 3 Interfering substances (continued)
Substance
Iron
However, since the hardness concentrations in these fracing fluids are very high, the use of a
smaller sample volume will decrease the iron interference. If when using a smaller sample
volume the iron concentration is still present at a concentration greater than 100 mg/L, add
the CDTA powder pillow to decrease this interference.
Magnesium
Manganese
Nickel
Interferes at 0.5 mg/L nickel. 0.5 grams of potassium cyanide can be added after the
potassium hydroxide solution to remove interference from up to 200 mg/L nickel.
Orthophosphate
Causes a slow endpoint but does not interfere if the calcium phosphate that forms is given
time to dissolve again during the titration.
Polyphosphates
Strontium
The strontium will be titrated directly with the Ca hardness titration. If the strontium
concentration of the sample is known, it can be subtracted from the result by converting the
mg/L Sr concentration to mg/L as CaCO3 by multiplying the Sr concentration by 1.142.
Temperature
Samples at 20 C (68 F) or colder should be titrated slowly near the endpoint to provide
sufficient time for the color change.
Zinc
Interferes at 5 mg/L zinc. 0.5 grams of potassium cyanide can be added after the potassium
hydroxide solution to remove interference from up to 100 mg/L zinc.
May be greater than the buffering capacity of the reagents. Adjust the pH before starting the
test (refer to Sample collection, preservation and storage).
Accuracy check
Use the standard additions method to identify if the sample has an interference and to make sure
that the test procedure was completed correctly.
Prerequisites:
Ampule breaker
Pipet tips
Hardness, Calcium
If more or less titrant was used, the problem can be due to user technique, an interference (refer to
Interferences) or a problem with reagents or apparatus.
Summary of method
The sample is made alkaline (pH 12 to 13) with potassium hydroxide to precipitate magnesium as
magnesium hydroxide. CalVer 2 Calcium Indicator is added and combines with any calcium to give
a red color. As the EDTA is added, it reacts with all the free calcium, barium (as long as both
strontium and calcium are present) and strontium present. At the endpoint of the titration, when no
free calcium ions are present, the EDTA removes the calcium complexed with the indicator. The
indicator then changes from red to blue.
Quantity/Test
Unit
Item no.
2447500
1 pillow
100/pkg
85299
12 mL
100 mL MDB
28232H
varies
1439901
Required apparatus
Description
Quantity/Test
Item no.
Digital titrator
1690001
50546
50838
50840
50841
50842
1970001
50/pkg
2185696
5/pkg
1720500
Unit
Item no.
16/pkg
218710
500 mL
2833349
Recommended standards
Description
Hardness, Calcium
Page 66
Hardness, Calcium
Unit
Item no.
113 g
28114H
29 mL
102233
100/pkg
1408099
29 mL
102233
100 mL
42532
1L
74053
113 g
28014
500 mL
15249
500 mL
254049
50 mL
245026
Potassium Cyanide
125 g
76714
2095352
1940000
1940010
4L
27256
500 mL
28249
1451538
1451520
1465100
2087079
100 mL MDB
2329232
100 mL MDB
16232
pH meter
Pipet tips
50/pkg
2185696
1970010
Spoon, measuring, 1 g
51000
90700
51100
Water, deionized
Potassium Hydroxide, 8 N
2196800
12/pkg
2087076
Hardness, Calcium
Page 67
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
09/2011, Edition 1
Hardness, Total
DOC316.53.01317
Method 10247
Digital Titrator
Test preparation
Before starting the test:
Four drops of Hardness 2 Indicator Solution or a 0.1-g scoop of ManVer 2 Hardness Indicator Powder can be added instead
of the ManVer 2 Hardness Indicator Powder Pillow.
Total Hardness as Ca = mg/L Total Hardness as CaCO3 x 0.40
mg/L Total Hardness as CaCO3 = mg/L Ca as CaCO3 + mg/L Mg as CaCO3.
If samples cannot be analyzed immediately, refer to Sample collection, preservation and storage. Adjust the pH of preserved
samples before analysis.
For added convenience, use the TitraStir stir plate1.
1
Quantity
1
2 mL
1 cartridge
Digital titrator
Graduated cylinder
Water, deionized
Erlenmeyer flask, 250-mL
4L
1
Hardness, Total
Page 69
Hardness, Total
Test procedure
See
Table 1
or
Table 2
1. Select a sample
volume from Table 1.
Refer to Identify the
sample volume if the
concentration ranges are
unknown.
4. Use a graduated
cylinder or pipet to
measure the sample
volume identified in step 1.
Put the sample in a clean,
250-mL Erlenmeyer flask.
If the sample volume is
less than 100 mL, dilute to
about 100 mL with
deionized water.
5. Add 2 mL of Hardness
1 Buffer Solution. Swirl to
mix.
Hardness, Total
Page 70
Hardness, Total
Multiplier
100400
100
200800
50
5002000
20
10004000
10
10
20008000
20
500020,000
50
10,00040,000
100
20,00080,000
0.5
200
50,000200,000
0.2
500
200
0.2
100
0.5
50
25
10
10
20
Hardness, Total
Samples can be used for up to 7 days if preserved and kept at or less than 4 C (39 F). Preserve
samples only when immediate analysis is not possible.
To preserve the sample:
1. Add 1.5 mL of nitric acid per 1 liter (1 quart) of sample to the sample. Mix fully.
2. Measure the sample pH.
3. If the sample pH is greater than 2, add more nitric acid in 0.5-mL increments. Mix fully and
check the pH of the sample after each addition until the sample pH is 2 or less.
Before analysis:
1. Let the sample temperature increase to room temperature.
2. Add 5.0 N sodium hydroxide to the preserved sample to change the pH to 7. Mix fully.
3. If a significant amount of nitric acid was added to the sample, make a volume correction for the
add nitric acid and hydroxide.
Total volume = sample + acid + hydroxide
Volume correction = (total volume) divided by (sample volume)
Corrected test result = (test result) x (volume correction)
Interference
WARNING
Chemical hazard. Potassium cyanide is toxic. Always add potassium cyanide after the
potassium hydroxide. Obey local hazardous waste regulations for disposal of all cyanidecontaining waste.
An interfering substance can prevent the color change at the titration endpoint. A dilution can often
reduce the interference to a level at which the substance does not interfere. If an interference is
suspected, decrease the sample volume, dilute to about 100 mL and do the test again. Table 3
shows substances that can interfere with this test.
Table 3 Interfering substances
Substance
Acidity
Alkalinity
This test can tolerate 10,000 mg/L alkalinity and can be done directly in sea water.
Aluminum
Aluminum interferes at levels greater than 0.20 mg/L aluminum. 0.5 grams of potassium
cyanide can be added after the buffer solution to remove interference of up to 1 mg/L
aluminum.
As an alternative, the interference can be removed by the addition of a CDTA powder pillow.
Refer to Table 4 and Table 5.
Barium
Cobalt
Hardness, Total
Page 72
Barium is a by-product from the drilling process and can be present in very high
concentrations in process and flow back waters.
The barium will be titrated directly with the total hardness titration. If the barium concentration
of the sample is known, it can be subtracted from the result. Refer to Table 5.
Interferes at all levels. 0.5 grams of potassium cyanide can be added after the potassium
hydroxide solution to remove interference from up to 20 mg/L cobalt.
As an alternative, the interference can be removed by the addition of a CDTA powder pillow.
Refer to Table 4 and Table 5.
Hardness, Total
Table 3 Interfering substances (continued)
Substance
Copper
As an alternative, the interference can be removed by the addition of a CDTA powder pillow.
Refer to Table 4 and Table 5.
Interferes at levels greater than 8 mg/L by causing an orange-red to green endpoint.
Accurate results can still be obtained at levels up to 20 mg/L iron with this endpoint. The iron
interference can be removed by the addition of a CDTA powder pillow when the
concentrations are greater than 100 mg/L. Refer to Table 4 and Table 5.
Iron
However, since the hardness concentrations in these fracing fluids are very high, the use of a
smaller sample volume will decrease the iron interference. If when using a smaller sample
volume the iron concentration is still present at a concentration greater than100 mg/L, add
the CDTA powder pillow to decrease this interference.
Interferes at levels greater than 5 mg/L. The interference can be removed by the addition of
a CDTA powder pillow. Refer to Table 4 and Table 5.
Manganese
Interferes at 0.5 mg/L nickel. 0.5 grams of potassium cyanide can be added after the
potassium hydroxide solution to remove interference from up to 200 mg/L nickel.
Nickel
As an alternative, the interference can be removed by the addition of a CDTA powder pillow.
Refer to Table 4 and Table 5.
Orthophosphate
Causes a slow endpoint but does not interfere if the calcium phosphate that forms is given
time to redissolve during the titration.
Polyphosphates
Although less common than calcium and magnesium, other polyvalent metal ions cause the
same hardness effects and will be included in the results.
Sodium chloride
Strontium
The strontium will be titrated directly with the total hardness titration. If the strontium
concentration of the sample is known, it can be subtracted from the result. Refer to Table 5.
Interferes at 5 mg/L zinc. 0.5 grams of potassium cyanide can be added after the potassium
hydroxide solution to remove interference from up to 100 mg/L zinc.
Zinc
As an alternative, the interference can be removed by the addition of a CDTA powder pillow.
Refer to Table 4 and Table 5.
May exceed the buffering capacity of the reagents. Adjust the pH before starting the test
(refer to Sample collection, preservation and storage).
The addition of one CDTA Magnesium Salt Powder Pillow will remove metals interferences at or
below the levels shown in Table 4. If more than one metal is present at or greater than the
concentrations in Table 4, the addition of another CDTA Magnesium Salt Powder Pillow may be
necessary.
Table 4 Interference level with CDTA pillow
Substance
Aluminum
50
Cobalt
200
Copper
100
Iron
100
Manganese
200
Nickel
400
Zinc
300
Hardness, Total
Page 73
Hardness, Total
The results given with CDTA Magnesium Salt include the hardness contributed by the metals. If
the concentration of each metal is known, a correction can be made to get the hardness
contributed by calcium and magnesium only. The hardness contributed per mg/L metal ion is
shown in Table 5.
Metal hardness = (mg/L of metal in the sample) x (hardness equivalence factor)
Calcium and magnesium hardness = (total hardness) minus (metal hardness)
Table 5 Hardness equivalence factors
Substance
Aluminum
3.710
Barium
0.729
Cobalt
1.698
Copper
1.575
Iron
1.792
Manganese
1.822
Nickel
1.705
Strontium
1.142
Zinc
1.531
Accuracy check
Use the standard additions method to identify if the sample has an interference and to make sure
that the analytical technique is correct.
Standard additions method (sample spike)
Prerequisites:
Ampule breaker
Pipet tips
Hardness, Total
Page 74
Hardness, Total
Standard solution method
Complete the following test to make sure that the reagents and user technique are accurate.
Prerequisites:
Summary of method
In the total hardness test procedure, the water sample is first buffered (using an organic amine and
one of its salts) to a pH of 10.1. An organic dye, calmagite, is added as the indicator for the test.
The organic dye reacts with calcium and magnesium ions to give a red-colored complex.
EDTA (ethylenediaminetetraacetic acid) is added as a titrant. The EDTA reacts with all free
calcium, magnesium, barium and strontium in the sample. At the endpoint of the titration, when
free magnesium ions are no longer available, EDTA removes magnesium ions from the indicator.
The indicator then changes from red to blue.
Quantity/Test
Unit
Item no.
2448100
100/pkg
85199
2 mL
100 mL MDB
42432
Description
Quantity/Test
Unit
Item no.
Digital titrator
1690001
50546
50838
50840
50841
50842
1970001
50/pkg
2185696
5/pkg
1720500
Description
Unit
Item no.
1L
12153
16/pkg
218710
Required apparatus
Recommended standards
Hardness, Total
Page 75
Hardness, Total
Optional reagents and apparatus
Description
CDTA Magnesium Salt Powder Pillow
Unit
Item no.
100/pkg
1408099
113 g
28014
29 mL
102233
100 mL
42532
1L
74053
50 mL
245026
500 mL
15249
500 mL
254049
125 g
76714
2095352
1940000
Potassium Cyanide
1940010
4L
27256
1451538
1451520
Water, deionized
1465100
1970010
Spoon, measuring, 1 g
51000
90700
51100
50/pkg
2199796
2196800
12/pkg
2087076
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
09/2011, Edition 1
Iron, Total
DOC316.53.01310
FerroVer Method
Method 10249
Powder Pillows
Scope and Application: For oil and gas field waters; digestion is required for determining total iron
USEPA approved for reporting wastewater analysis (digestion is required)
Test procedure
1. Push: PRGM
The display shows:
PRGM ?
Initial setup: go to
Instrument setup on
page 80 to add the
program to the
instrument.
Iron, Total
Page 77
Iron, Total
9. Push: ZERO
The cursor moves to the
right, then the display
shows:
0.0 mg/L Fe LR, Fe MR
or Fe HR
11. Push:
TIMER ENTER
A three-minute reaction
period starts.
An orange color will form
if iron is present.
If the sample contains
visible rust or if five drops
of EDTA were used in
step 7, the sample should
be allowed to react for at
least five minutes.
Accuracy Check
Standard Additions Method
1. Fill three sample cells with sample as specified in steps 4 and 5 of the test procedure.
Iron, Total
Page 78
Iron, Total
2. Open a 25 mg/L Iron Voluette Ampule Standard Solution.
3. Use the TenSette Pipet to add 0.1, 0.2, and 0.3 mL of the 25 mg/L standard, respectively, to
the samples and mix fully.
4. Complete the test procedure for each sample.
5. Examine the results. The iron concentration should increase by 0.25 mg/L for Fe LR,
2.5 mg/L for Fe MR and 25.0 mg/L for Fe HR for each 0.1 mL of standard that is added.
6. If these increases do not occur, refer to Section 1 Standard Additions of the procedures
manual for troubleshooting information.
Standard Solution Method
Prepare a 1.0-mg/L iron standard solution.
Dilute 1.00 mL of Iron Standard Solution, 100 mg/L Fe, to 100 mL with deionized water.
Use 10 mL of the standard in step 4 to complete the Fe LR procedure for powder pillows. Results
should be between 0.90 mg/L and 1.10 mg/L Fe.
Method Performance
Precision
In a single laboratory, with a standard solution of 2.0 mg/L Fe and two of powder pillows with the
instrument, a single operator got a standard deviation of 0.02 mg/L Fe.
Estimated Detection Limit (EDL)
The EDL for program 127 is 0.1 mg/L Fe. For more information on derivation and use of the
estimated detection limit, refer to Section 1 of the procedures manual.
Interferences
Interfering Substances and Suggested Treatments
Interfering Substance
Barium, Ba2+
The dilution of samples to measure iron lowers most barium concentrations below
interference levels. No effects are seen on analyzed solutions which contain less
than 50 mg/L of Ba. No effects are seen when 1.0 or 0.1 mL of sample is used in step
4. A turbidity may show at higher levels. Add 5 drops of EDTA Solution to the solution
in step 5 and do the analysis again. Allow the sample to react for 5 minutes.
Calcium, Ca2+
Cl-
Cu2+
Chloride,
Copper,
Prevents color development. Dilute the sample and test again to verify results.
Magnesium
Molybdate, Molybdenum
High Sulfide Levels,
S2-
Strontium, Sr2+
Strontium by itself does not interfere. Strontium in combination with Barium will cause
a precipitate to form. The dilution of samples to measure iron lowers most strontium
concentrations below interference levels. No effects are seen on analyzed solutions
which contain less than 50 mg/L of combined Ba and Sr. No effects are seen when
1.0 or 0.1 mL of sample is used in step 4. A turbidity may start at higher levels. Add 5
drops of EDTA Solution to the solution in Step 5 and do the analysis again. Allow the
sample to react for 5 minutes
Iron, Total
Page 79
Iron, Total
Interfering Substance
Sample pH (extreme)
Summary of Method
FerroVer Iron Reagent reacts with all soluble iron and most insoluble forms of iron in the sample to
produce soluble ferrous iron. Ferrous iron reacts with 1,10-phenanthroline indicator in the reagent
to form an orange color in proportion to the iron concentration.
Instrument setup
This procedure adds the program 127 to a DR/820, DR/850 or DR/890 instrument.
1. Push the ON key to turn on the instrument.
2. Push the SETUP key.
3. Push the down arrow key until the prompt line shows USER.
4. Push the ENTER key.
5. Push the numbers 8138, then push ENTER.
6. Refer to Table 1. Find the number from the Enter column that corresponds to Line Number 1
on the display. Push these numbers on the keypad, then push ENTER. Continue to add the
numbers that correspond to each line number on the display.
Note: Use the arrow keys to scroll and review or change numbers at any time.
Iron, Total
Page 80
Iron, Total
Enter
Line Number
Enter
127
29
82
24
30
70
73
31
101
32
32
33
72
34
82
35
65
36
32
37
10
38
11
39
66
12
64
40
200
13
24
41
14
81
42
15
235
43
16
44
33
17
45
128
18
46
19
47
15
20
70
48
21
101
49
180
22
32
50
23
76
51
24
82
52
25
70
53
26
101
54
132
27
32
55
28
77
56
255
Iron, Total
Page 81
Iron, Total
Unit
Item No.
1 pillow
100/pkg
2105769
6/pkg
2401906
2 drops
50 mL
2241926
OPTIONAL REAGENTS
Description
Unit
Item No.
500 mL
88449
100 mL
1417542
16/pkg
1425310
100 mL MDB
245032
4L
27256
OPTIONAL APPARATUS
Description
Unit
Item No.
2196800
108140
108142
50546
1457442
1465100
Pipet, serological, 2 mL
53236
1970001
50/pkg
2185696
1000/pkg
2185628
1451535
1970010
50/pk
2199796
250/pk
2199725
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
09/2011, Edition 1
pH
pH
DOC316.53.01323
Method 10255
pH meter
Based on Standard Method 4500-H+B, ASTM Method D1293-84(90)/(A or B) and USEPA Method 150.
Test preparation
Meter
Standard probe
Rugged probe1
Items to collect:
Description
Quantity
Beakers/sample containers
pH
Page 83
pH
Sample pH measurement (calibration required)
5. In three separate
beakers or containers,
prepare fresh buffers of
4.0, 7.0 and 10.0 pH.
pH
Page 84
pH
Collect samples in clean plastic or glass bottles. Fill completely and cap tightly.
For general electrode storage, use the Hach storage solution or a 3 M potassium chloride
(KCl) solution.
A contaminated glass bulb or fouled electrode can cause a slow response time.
Do not clean the bulb too often because the bulb life can decrease.
To clean an electrode that has contamination from oils and fats, put the electrode tip in a
detergent solution. Clean with a soft brush or ultrasonic bath. Do not scratch the glass bulb.
Interferences
The sodium error is low but increases at pH values that are higher than pH 11. The acid error is
negligible. For more information, refer to the documentation for the electrode or the meter.
Accuracy check
Electrode operation
The electrode operation is satisfactory when the calibration slope is within the specified range
(typically 58 (3) mV at 25 C).
Calibration accuracy
Measure the pH of a fresh buffer solution. A calibration is satisfactory when the measured pH
agrees with the known pH value of the buffer.
Method performance
The accuracy of a pH measurement depends on many factors that are associated with the overall
pH system, which includes the meter, electrode and calibration buffers. Refer to the
documentation for the electrode or the meter for more information.
Summary of method
A combination pH electrode develops an electrical potential at the glass/liquid interface. At a
constant temperature, this potential varies linearly with the pH of the solution.
The pH is a measure of the hydrogen ion activity in a solution and is defined as log10 aH+, where
aH+ is the activity of the hydrogen ion. The sample pH can change when carbon dioxide is
absorbed from the atmosphere. In water that has a high conductivity, the buffer capacity is typically
high and the pH does not change significantly.
pH
Page 85
pH
Unit
Item no.
HQ40d meter
HQ40d53000000
HQ30d meter
HQ30d53000000
HQ11d meter
HQ11d53000000
PHC10101
PHC10103
PHC30101
PHC30103
PHC10105
PHC10110
PHC10115
PHC10130
30 mL
2841700
500 mL
2756549
Recommended standards
Description
pH color-coded buffer solution kit (NIST), 500 mL, includes:
Unit
Item no.
2947600
500 mL
2283449
500 mL
2283549
500 mL
2283649
50/pkg
2226966
Powder
pillows1
50/pkg
2227066
50/pkg
2227166
500 mL
S11M001
500 mL
S11M002
pH 7.000 0.010 at 25 C
500 mL
S11M004
pH 10.012 0.010 at 25 C
500 mL
S11M007
500 mL
S11M009
500 mL
S11M010
500 mL
S11M011
Unit
Item no.
2758101
Sample bottle, cleaned and certified, HDPE, suitable for EPA reporting, 500-mL
2758201
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
08/2012, Edition 1
Sulfate
DOC316.53.01312
SulfaVer 4 Method1
Method 10248
Powder Pillows
Adapted from Standard Methods for the Examination of Water and Wastewater.
Test procedure
1. Push PRGM.
The display shows
PRGM ?
Initial setup: go to
Instrument Setup on
page 89 to add the
program to the
instrument.
7. Push ZERO.
The cursor moves to the
right, then the display
shows:
0 mg/L SO4 and L, M or
H
Sulfate
Page 87
Sulfate
9. Push
TIMER ENTER
A 5-minute reaction
period starts.
Do not move the sample
cell during the reaction
period.
Interferences
Known interferences are shown in Table 1. The interference levels are applicable to an undiluted
10-mL sample. The interference levels increase proportionally as the sample is diluted.
Table 1 Interferences
Substance
Interference Level
Barium
Interferes at all levels. The greater the barium concentration compared to the sulfate concentration,
the greater the error. Samples with high barium concentrations will generally give a result that is
20% lower than the actual sulfate concentration.
Calcium
Chloride
Magnesium
Silica
Turbidity
Accuracy Check
Standard Additions Method
Use the standard additions method to validate the test procedure, reagents and instrument and to
find if there is an interference in the sample.
1. Fill three sample cells with sample as specified in steps 4 and 5 of the test procedure.
2. Use a TenSette Pipet to add 0.1, 0.2 and 0.3 mL of a 1000 mg/L Sulfate Standard Solution to
the sample cells. Mix fully.
3. Complete the test procedure for each sample.
Sulfate
Page 88
Sulfate
4. Review the results. The sulfate concentration should increase by 10 mg/L for SO4 L,
100 mg/L for SO4 M or 1000 mg/L for SO4 H for each 0.1 mL of standard that is added.
5. If the concentration does not increase by the correct amount, refer to Standard Additions in
Section 1 of the procedures manual.
Standard Solution Method
Use a 50 mg/L Sulfate Standard Solution to validate the test procedure, reagents and the
instrument. Select the SO4 L test range in step 3 and use 10 mL of the standard solution instead of
the sample in step 4. To adjust the result, refer to Standard Adjust.
To prepare this standard solution, add 5.0 mL of a 1000 mg/L Sulfate Standard Solution to a 100mL volumetric flask. Dilute to the mark with deionized water and mix fully.
Standard Adjust
The standard adjust option is recommended when program 128 is used.
6. Measure the concentration of a 50 mg/L Sulfate Standard Solution. Select the SO4 L range
and use 10 mL of the standard solution. Keep the sample cell in the instrument.
7. Push the SETUP key and use the arrow keys to scroll to the STD option.
8. Push ENTER.
9. Push the numbers 50 to make the instrument read the value of the standard solution
concentration.
10. Push ENTER to complete the adjustment.
Note: The MR and HR calibration curves are adjusted proportionally when the SO4 L calibration curve is
adjusted. Refer to Section 1, Standard Curve Adjustment of the procedures manual.
Method Performance
Precision
In a single laboratory, with a 50 mg/L sulfate standard solution, two representative lots of powder
pillows and the instrument, a single operator got a standard deviation of 0.5 mg/L sulfate.
Estimated Detection Limit (EDL)
The EDL for program 128 SO4 L is 4.9 mg/L SO4. For more information on derivation and use of
the estimated detection limit, refer to Section 1 of the procedures manual.
Summary of Method
Sulfate ions in the sample react with barium in the SulfaVer 4 Sulfate Reagent to make an
insoluble barium sulfate precipitate. The amount of precipitate is proportional to the sulfate
concentration. The SulfaVer 4 also contains a stabilizing agent to hold the precipitate in
suspension.
Instrument Setup
This procedure adds program 128 to a DR/820, DR/850 or DR/890 instrument.
1. Push the ON key to turn on the instrument.
2. Push the SETUP key.
3. Push the down arrow key until the prompt line shows USER.
4. Push the ENTER key.
5. Push the numbers 8138, then push ENTER.
Sulfate
Page 89
Sulfate
6. Refer to Table 2. Find the number from the Enter column that corresponds to Line Number 1
on the display. Push these numbers on the keypad, then push ENTER. Continue to add the
numbers that correspond to each line number on the display.
Note: Use the arrow keys to scroll and review or change numbers at any time.
Sulfate
Page 90
Line Number
Enter
Line Number
Enter
128
29
77
24
30
83
72
31
79
32
52
33
32
34
72
35
65
65
36
32
198
37
10
169
38
11
251
39
66
12
66
40
200
13
25
41
14
10
42
15
61
43
16
44
80
17
45
18
46
19
47
30
20
83
48
21
48
49
44
22
52
50
23
32
51
24
76
52
25
83
53
26
79
54
37
27
52
55
28
32
56
255
Sulfate
Units
Item No.
1 pillow
100/pkg
2106769
6/pkg
2401906
Units
Item No.
OPTIONAL REAGENTS
Description
Sulfate Standard Solution, 50 mg/L
500 mL
257849
500 mL
2175749
4L
27256
Water, deionized
OPTIONAL APPARATUS
Description
Cylinder, graduated mixing, 10 mL
Units
Item No.
2088638
100/pkg
189457
1457442
Funnel, poly, 65 mm
108367
1970001
50/pkg
2185696
1970010
50/pkg
2199796
1451537
1465100
Sulfate
Page 91
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
06/2012, Edition 2
Sulfide
DOC316.53.01320
Method 10254
Adapted from Standard Methods for the Examination of Water and Wastewater.
USEPA accepted for wastewater analysis. Procedure is equivalent to USEPA method 376.2 or Standard Method
4500-S2- D for wastewater.
Test procedure
1. Push PRGM.
The display shows
PRGM ?
Initial setup: go to
Instrument setup on page
95 to add the program to
the instrument.
5. Sample
preparation: Add the
sample volume that is
specified for the test
range to a clean sample
cell.
4. Blank preparation:
Add 25 mL of deionized
water to a sample cell.
A 25-mL graduated
mixing cylinder can be
used in steps 4 and 5.
8. Swirl to mix.
S LR: 25 mL
S MR: 2.5 mL
S HR: 0.25 mL
Use a pipet to measure
0.25 mL or 2.5 mL.
Sulfide
Page 93
Sulfide
11. Push
TIMER, then ENTER.
A 5-minute reaction
period starts.
Soluble sulfides
Complete the steps that follow to measure the concentration of soluble sulfides in the sample.
1. Use a centrifuge to make a separation of the soluble and insoluble sample components. Make
sure that the centrifuge tubes are filled completely and have a cap.
2. Use the liquid portion as the sample in the test procedure to measure the concentration of
soluble sulfides.
To make an estimate of the insoluble sulfides, subtract the soluble sulfide concentration from the
total sulfide concentration.
Sulfide
Page 94
Sulfide
Interferences
The known interferences are shown in Table 1. The interference levels are applicable to an
undiluted 10-mL sample. The interference levels increase proportionally as the sample is diluted.
Table 1 Interferences
Substance
Interference
Barium
Concentrations greater than 20 mg/L react with the sulfuric acid in Sulfide 1 Reagent
and form a BaSO4 (barite) precipitate. To correct for this interference:
1. Use a 0.25-mL or 2.5-mL sample volume in the test procedure and add deionized
water to the 25-mL line.
2. Let the sample fully react with both reagents.
3. After the 5 minute reaction period, pour the sample into a 50-mL beaker.
4. Pull the sample into a 60 cc Luer-Lock syringe.
5. Put a 0.45-m filter disc on the Luer-Lock tip and filter the sample into a clean
sample cell for measurement. Use deionized water to prepare the blank.
Can decrease the blue color or prevent the full color development.
Can prevent the full color development. Make a dilution if the sample has a high sulfide concentration. Some sulfide loss can occur when the sample is diluted.
Turbidity
Accuracy check
Sulfide standard solutions are not stable and must be prepared by the user. Refer to Standard
Methods, 4500S for preparation and standardization instructions.
Method performance
Precision
In a single laboratory, with standard solutions of 0.73 mg/L sulfide and two representative lots of
reagent, a single operator got a standard deviation of 0.02 mg/L sulfide with the instrument.
Estimated detection limit (EDL)
The EDL for program 129 is 0.01 mg/L S2-.
Summary of method
Hydrogen sulfide and acid-soluble metal sulfides react with N, N-dimethyl-p-phenylenediamine to
form methylene blue. The intensity of the blue color is proportional to the sulfide concentration.
Instrument setup
This procedure adds program 129 to a DR/850 or DR/890 instrument.
1. Push SETUP.
2. Push the DOWN ARROW until the display shows USER.
3. Push ENTER.
4. Push the numbers 8138, then push ENTER. The display shows LINE 1?
Sulfide
Page 95
Sulfide
5. Refer to Table 2. Find the 1 in the Line Number column, then read across to find the numbers
in the Enter column. Push these numbers on the keypad, then push ENTER.
6. Continue to add the numbers that correspond to each line number on the display.
Note: Use the arrow keys to scroll and review or change numbers at any time.
Sulfide
Page 96
Line number
Enter
Line number
Enter
129
29
42
30
83
74
31
32
32
72
33
82
34
35
65
36
32
37
10
38
11
39
66
12
63
40
200
13
129
41
14
202
42
15
184
43
16
44
80
17
45
18
46
244
19
47
20
20
83
48
21
32
49
44
22
76
50
23
82
51
24
52
25
83
53
26
32
54
167
27
77
55
28
82
56
255
Sulfide
Quantity/Test
Unit
Item no.
2244500
1 mL
100 mL MDB
181632
1 mL
100 mL MDB
181732
10 mL
4L
27256
Quantity/Test
Unit
Item no.
BBP078
100/pkg
BBP079
BBP065
75/pkg
BBP068
Pipets, variable volume, one BBP078 and one BBP065 with tips
LZP320
6/pkg
2401906
Unit
Item no.
50041H
29 mL
221120
2088640
Flask, Erlenmeyer, 50 mL
50541
29 mL
211220
1451540
Water, deionized
Required apparatus
Description
1465100
2258700
50/pkg
2513603
Sulfide
Page 97
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
06/2012, Edition 2
TPH, 10050
Method 10050
This test is semi-quantitative. Results are expressed as greater or less than the threshold value used.
Test preparation
Cell orientation
Adapter
DR 6000
LZV902.99.00020
DR 5000
A23618
DR 3900
LZV846 (A)
LZV583 (A)
Quantity
Marker, laboratory
Wipes, disposable
Single Wavelength
OK
1. Press
Single Wavelength.
Press OPTIONS and the
button.
Enter 450 nm and press
OK.
2. Label an Antibody
Cuvette for each calibrator
and each sample to be
tested.
Insert an adapter if
required (Instrumentspecific information).
8. Immediately pipet
0.5 mL of TPH Enzyme
Conjugate into each
calibrator and sample
cuvette.
The same pipette tip can
be used repeatedly for this
step.
Zero
3. Submerge the
capillary tube below the
surface of the liquid to be
pipetted. Slowly and
smoothly draw the Wiretrol
plunger up until the bottom
of the plunger tips reaches
the appropriate volume
line.
Touch the end of the tube
to the side of the vessel to
release remaining drops
on the capillary tube tip.
Loading the RackThe cuvette rack is designed so that it may be inverted with the cuvettes in
place. Identify each cuvette with a sample or calibrator number and insert all the cuvettes in the
rack before beginning the procedure. Fit the cuvettes snugly into the rack, but do not force them or
they may be difficult to remove and their contents may spill. The cuvettes should remain in place
when the rack is inverted and tapped lightly.
MixingSet the rack on a hard, flat surface that is at least twice the length of the rack. Hold the
rack by one end and vigorously slide it back and forth along its long axis for 30 seconds. The rack
should move through a distance equal to its own length in each direction.
If sample reading < calibrator reading, TPH concentration in sample > calibrator reading
If sample reading > calibrator reading, TPH concentration in sample < calibrator reading
Example:
Readings
TPH Calibrator #1: 0.480 Abs
TPH Calibrator #2: 0.360 Abs
Sample #1: 0.200 Abs
Sample #2: 0.400 Abs
Sample #3: 0.550 Abs
Interpretation for a Soil Sample
Sample #1The sample reading (0.200 Abs) is less than the readings for both calibrators.
The concentration of TPH in the sample is greater than 50 ppm diesel fuel.
Sample #2The sample reading (0.400 Abs) is between the readings for the TPH calibrators.
The concentration of TPH in the sample is between 20 ppm and 50 ppm diesel fuel.
Sample #3The sample reading (0.550 Abs) is greater than the readings for both calibrators.
The concentration of TPH in the sample is less than 20 ppm diesel fuel.
Interpretation for a Water Sample
Sample #1The sample reading (0.200 Abs) is less than the readings for both calibrators.
The concentration of TPH in the sample is greater than 5 ppm diesel fuel.
Sample #2The sample reading (0.400 Abs) is between the readings for the TPH calibrators.
The concentration of TPH in the sample is between 2 ppm and 5 ppm diesel fuel.
Sample #3The sample reading (0.550 Abs) is greater than the readings for both calibrators.
The concentration of TPH in the sample is less than 2 ppm diesel fuel.
Sensitivity
The antibodies used in the TPH Test Kit react with a variety of compounds found in petroleum
fuels; however, each TPH calibrator has been formulated to represent a specific concentration of
diesel fuel. To use the calibrators for other TPH compounds, see Table 2 for soil or Table 3 for
water to select the proper TPH calibrator for the compound, sample and range you want to test.
Example:
To use the TPH calibrators for gasoline, find Gasoline in the first column of Table 2 or Table 3.
Read across the column to find the ppm represented by each calibrator. For gasoline,
calibrator #1 = 15 ppm, calibrator #2 = 35 ppm, etc.
TPH calibrator #1
(ppm)
TPH calibrator #2
(ppm)
TPH calibrator #3
(ppm)
TPH calibrator #4
(ppm)
Diesel fuel
20
50
100
200
Gasoline
15
35
70
140
Kerosene
35
75
140
250
Benzene
20
45
85
160
Toluene
15
30
50
90
Ethylbenzene
15
35
75
m-Xylene
20
35
70
o-Xylene
10
20
40
80
p-Xylene
16
BTEX
15
25
45
TPH calibrator #1
(ppm)
TPH calibrator #2
(ppm)
TPH calibrator #3
(ppm)
TPH calibrator #4
(ppm)
Diesel fuel
10
20
Gasoline
1.5
3.5
14
Kerosene
3.5
7.5
14
25
Benzene
4.5
8.5
16
Toluene
1.5
Ethylbenzene
0.5
1.5
3.5
7.5
m-Xylene
0.9
3.5
o-Xylene
p-Xylene
0.3
0.5
0.9
16
BTEX
0.5
1.5
2.5
4.5
Dilution multiplier
0.5
100
1.0
50
2.0
25
5.0
10
10.0
25.0
Interferences
Interfering substance
Interference level
Interferes above 2 ppm. Remove with 1 drop per 100 mL sodium thiosulfate (0.1 N).
To store the samples, collect them in glass or Teflon containers that have been washed with
soap and water and rinsed with methanol. The container should be capped with a Teflon-lined
cap. If a Teflon cap is not available, aluminum foil rinsed in methanol may be used as a
substitute cap liner.
When collecting water samples, fill the container completely (no head space) and cover the
container with a tightly-sealed lid immediately after collection.
WaterChill the sample in an ice bath or refrigerator to limit the loss of volatile compounds.
Store samples no longer than 24 hours.
Summary of method
This method provides semi-quantitative screening for TPH based on thresholds as diesel fuel in
the following concentrations:
Immunoassay tests use antigen/antibody reactions to test for specific organic compounds in water
and soil. Antibodies specific for TPH are attached to the walls of plastic cuvettes. They selectively
bind and remove TPH from complex sample matrices. A prepared sample and a reagent
containing enzyme-conjugate molecules (analyte molecules attached to molecules of an enzyme)
are added to the Antibody Cuvettes. During incubation, enzyme-conjugate molecules and TPH
compete for binding sites on the antibodies. Samples with higher levels of analyte will have more
antibody sites occupied by TPH and fewer antibody sites occupied by the enzyme-conjugate
molecules.
After incubation, the sample and unbound enzyme conjugate are washed from the cuvette and a
color-development reagent is added. The enzyme in the conjugate catalyzes the development of
color. Therefore, there is an inverse relationship between color intensity and the amount of TPH in
the sample. The resulting color is then compared with a calibrator to determine whether the TPH
concentration in the sample is greater or less than the threshold levels. The TPH concentration is
inversely proportional to the color development: the lighter the color, the higher the TPH
concentration. Test results are measured at 450 nm.
Unit
Catalog Number
20 cuvettes
2774300
500 mL
27248
Description
Unit
Catalog Number
Required apparatus
2/pkg
2581802
Marker, laboratory
each
2092000
each
1970001
1000/pkg
2185628
each
4879900
Wipes, disposable
box
2097000
Unit
each
2796900
medium1
each
2550502
50/pkg
2185696
20/pkg
2657205
each
2775200
Catalog Number
20/pkg
2124720
20/pkg
2567620
200 mL
2567729
20/pkg
2592920
20/pkg
2179020
Spatula, disposable
2/pkg
2569320
250 g
709929
Unit
Catalog number
each
2936701
500/pkg
1473800
Weighing papers
each
2550700
Graduated cylinder, 10 mL
each
108138
20/pkg
2852200
250/pkg
2568905
100 mL MDB
32332
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
09/2011 Edition 1
Procedures Explained
Barium
Introduction
Barium is a naturally occurring byproduct of the drilling process. Barium, along with
strontium, calcium and bicarbonate, can cause scale deposits in flow line pipes in the
presence of fracturing water with high sulfate concentrations. Barium and strontium react
with sulfate to form a white precipitate, which creates scaling inside the pipes.
Recommended Instrumentation
Matrix Challenges
Strontium will interfere with barium at concentration of 20 mg/L for both analytes. If the
sample concentration of barium and strontium are diluted to a concentration at or below
20 mg/L, the interference from strontium becomes negligible. When strontium is present in
the sample without barium, Sr is undetectable with the BariVer 4 reagent. A 100 mg Sr/L
spike in a sample without Ba produced a concentration of 2 mg Ba/L, which is the
method's lowest detectable concentration.
Figure 1 and Table 1 show the effect that strontium has on the barium concentration when
both Ba and Sr are present at the same concentrations.
Ba Spike recovery
% Difference
20
24
30
20
40
42
58
28
50
50
80
38
60
58
92
37
80
72
127
43
100
89
154
42
1 Both
Ba and Sr are spiked at the same concentration. For example, the first combined standard set was spiked with 20 mg/L Ba and Sr, the
second set was spiked with both 40 mg/L of Ba and Sr, and so on. As the concentration of both analytes in the combination spike increases,
the % difference between the Ba spike and the combination spike of Ba and Sr increases based on the nature of the Sr interference with the
BariVer 4 chemistry.
If the barium and the strontium concentrations are close to the same concentration, the
analyst can dilute them within the appropriate concentration range for the Ba method (2 to
100 mg/L). Dilute the Ba and Sr concentrations so they are both at or below 20 mg/L to
avoid the Sr interference.
Recommended Instrumentation
CDTA Magnesium Salt Powder Pillows (iron interference removal) (Item no. 1408099)
Matrix Challenges
Total Hardness
Produced and flowback water can have elevated levels of barium, strontium and iron.
The presence of iron interference causes a redorange to green endpoint. The iron
interference can be removed by the addition of a CDTA powder pillow when the
concentrations are above 100 mg/L. However, since the hardness concentrations in these
matrices are very high, the use of a smaller sample volume will reduce the iron
interference. After using the smaller sample volume, if iron concentration is still believed to
be present at a concentration above 100 mg/L, the addition of the CDTA powder pillow will
reduce this interference. CDTA does not affect barium or strontium interference and these
ions will be titrated directly along with calcium and magnesium. Therefore, the total
hardness value will include the divalent cations of barium, strontium, calcium and
magnesium. If the barium and strontium concentrations of the sample are known, these
interfering cations can be subtracted from the total hardness value by performing a
molecular weight conversion of both barium and strontium to CaCO3. The hardness
equivalence factor for barium and strontium is 0.729 and 1.142, respectively. Multiply the
barium and strontium concentrations by these equivalence factors to convert their
concentrations to CaCO3 and subtract these values from the total hardness concentration
to achieve a more accurate hardness value for just magnesium and calcium hardness.
See the Hardness Equivalence Factor Table in the total hardness procedure for other
metal hardness equivalence factors. To report the true total hardness of the sample, do
not subtract out the barium and strontium concentrations.
Calcium Hardness
For the calcium hardness titration, laboratory studies have shown that a sample containing
barium and strontium will not precipitate at pH 13, and both will be titrated along with the
calcium in the sample. If strontium and calcium are not present in the sample, the barium
will precipitate at pH 13. However, most produced and flowback water samples have
strontium, calcium and barium present at high concentrations. If the concentrations of
barium and strontium are known, these ions can be subtracted from the calcium hardness
titration in the same manner as the total hardness titration. Convert the barium and
strontium to CaCO3 using the equivalence factors listed above. After subtracting out the
barium and strontium, the resulting value will be the calcium hardness as CaCO3. To
convert calcium as CaCO3 to the elemental form of just calcium, multiply the Ca hardness
value by 0.4.
Iron
Introduction
Iron is a byproduct of the drilling process, with ferrous (Fe+2) being detrimental to the
fracing fluid. The fracing fluids have chemicals added to prevent the precipitation of metal
oxides, iron being one of the most important metals. When preparing the fracing fluid, the
ferrous iron concentration should be less than 10 mg/L to prevent fluid degradation. The
FerroVer chemistry will detect both ferrous and ferric (Fe+3) iron. Iron in ground water is
normally present in the ferrous form, which oxidizes quickly to ferric iron with exposure to
air.
Recommended Instrumentation
FerroVer Iron Reagent Powder Pillows (for 10 mL samples) (item number 2105769)
Determination of iron
Test results are dependent upon sample pretreatment steps and the iron reagent reaction
conditions. The FerroVer Iron Reagent contains strong reducing agents plus 110
phenanthroline indicator. The reducing agents reduce ferric iron present in the sample to
ferrous iron. The phenanthroline indicator then reacts with the ferrous iron to form an
orange color in proportional to the iron concentration. The strong reducing agents in
FerroVer will convert most insoluble iron forms to soluble ferrous iron and be included in
the test results. A total iron test to determine soluble and insoluble iron will require a mild
acid digestion followed by analysis with FerroVer Iron Reagent. Samples that have been
filtered before analysis to remove particulates or samples that are analyzed with iron
reagents containing ascorbic acid as the reducing agent will usually give much lower test
results. It therefore becomes important to document the sample pretreatment steps when
comparing test results with previous analysis or with other outside laboratories using other
colorimetric or instrumental methods of analysis.
Matrix Challenges
In fracing fluids, barium and strontium can be present at high concentrations, and
concentrations above 50 mg/L will interfere with the iron analysis by forming a precipitate.
Strontium by itself will not interfere with the iron analysis, however; when strontium and
barium are in the solution together (this combination is common in fracing fluids) the
strontium acts as a catalyst increasing the barium interference. This precipitate / turbidity
interference can be eliminated with the addition of a 1 M EDTA solution. The EDTA does
not chelate the iron, but will remove barium and strontium from the sample. The amount of
EDTA needed to overcome this interference depends on the concentration of the barium
and strontium in the sample. Laboratory studies have shown that 1 to 2 drops of a 1 M
EDTA solution added to a 10 mL sample volume is sufficient to remove the barium
interference of approximately 50 to 200 mg/L Ba/Sr. Concentrations of 200 to 1000 mg/L
Ba/Sr, use up to 5 drops of 1 M EDTA to eliminate the interference. If there is precipitate in
the sample after the addition of the FerroVer reagent powder pillow, the analyst can add
more EDTA or dilute the sample to reduce the barium/strontium interference. Avoid
adding to much EDTA, up to 10 drops, to the 10 mL sample. Over dosing the sample with
EDTA will impede the color development of the FerroVer reagent. It was observed in the
laboratory that 10 drops of EDTA in a 10 mL sample caused the colorimetric reaction to
take up to 30 minutes to develop. The addition of EDTA will tie up the barium, strontium
and the iron, since they are all cations, but the phenanthroline indicator in the FerroVer
reagent has a stronger affinity to the iron than the EDTA does and will eventually complete
the colorimetric reaction. It is recommended to use 5 drops of EDTA or less with the 10 mL
sample volume. At extremely high concentrations of both Ba and Sr in the sample (above
Recommended Instrumentation
Produced and flowback water have conductivity values that are in the mS/cm range, ~10
200+ mS/cm (~10 150 g/L as TDS). As Figure 1 shows, to accurately measure
conductivity values at this elevated level it is necessary to use a 4pole conductivity cell
with an enhancement from either graphite, stainless steel, or platinum*.
Matrix Challenges
Due to the high ranges of conductivity in these sample matrices, it is recommended to use
a metal enhanced 4pole cell. Users that don't have this type of cell can dilute the sample
to get it into the appropriate range for the cell's specifications. However, laboratory studies
have shown it's possible to produce a 30% increase in conductivity values with the diluted
samples compared to the undiluted samples. Hach recommends not diluting the samples
for conductivity measurements to avoid this error.
Figure 2 shows some of the effects of an increase in conductivity on diluted samples.
* The enhanced 4-poled cells have a layer of metal (graphite, stainless steel, or platinum) on the poles to minimize the effects
of polarization and increase the concentration range. (For more information on conductivity theory, download the Conductivity
Theory and Practice from the Hach website.)
Sample1
Undiluted
Diluted
%Diff
235
344
31.7
208
258.8
19.6
197.1
245.8
19.8
174.5
205.2
15.0
Samples were diluted 1:1 and the conductivity measured. The value was then multiplied by 2 to get the final diluted result.
Higher sample conductivities will produce larger positive errors when the sample is
diluted.
The conductivity cells were calibrated using a single point calibration using three different
standards, 1408 S/cm, 12.85 mS/cm, and 111.3 mS/cm standard; none of the three
calibrations improved the difference between the undiluted and diluted samples. However,
it is always recommended to calibrate using a calibration standard that approximates the
range of conductivity values expected in the sample to be tested. Figure 3 provides
guidance on choosing the appropriate standard concentration.
TDS factors
The different Hach Company meter platforms offer direct measurements for TDS; it is up
to the operator to select the type of conversion factor that best matches the true TDS
value. The TDS is typically used to estimate the amount of total dissolved solids in the
sample. The standard method to determine TDS is to filter and evaporate the sample to
dryness at 180C, then weigh the residue. Hach Method 8163 is available for determining
the total dissolved solids using the standard method. If required, Method 8163 can be
used to determine the conversion factor for a specific solution or sample matrix.
To determine the conversion factor for a specific solution of a known TDS value, measure
the solution's conductivity and divide the mg/L TDS value by the conductivity value
reported. For example, a solution of a known TDS value of 64 g/L and the measured
conductivity value of 100 mS/cm has a conversion factor of 64/100 or 0.64. It is important
to know the conversion factor being used, especially when comparing your TDS results
with another lab's results, another test site or when comparing results with previously
published or referenced data.
The different TDS concentration conversion options for the HQd meters are as sodium
chloride (NaCl), a generic default factor of 0.5, or a userentered custom value. The
operator can choose any factor within the custom field; a common factor for high salinity
samples is 0.64. For the MP6 meter, the TDS factor options are as NaCl, as potassium
chloride (KCl), 442 and userentered. The MP6's meter default is the KCl which is used for
conductivity, the NaCl is used for resistivity (mineral/salt), the 442 factor is an algorithm
that is used for estimating TDS in natural waters, and the userentered factor option.
Maintenance
Due to the nature of the produced and flowback water, the operator needs to be sure to
rinse the conductivity cell off with clean water. Do not allow the cell to soak, or store the
cell in these samples. Once the cell has been rinsed off, blot and store dry.
Ordering information
To order one of the 4pole conductivity cells, refer to Table 1. For more meter and probe
options, visit www.hach.com.
Table 1 CDC401 Graphite Conductivity Cell ordering information
Description
Item number
CDC40101
CDC40103
CDC40105
CDC40110
CDC40115
CDC40130
Recommended Instrumentation
Laboratory method, Gravimetric method (Hach method 8158) for nonfilterable total
suspended solids.
10
Matrix Challenges
The laboratory turbidimeters, A and AN, can measure above 1000 NTUs because of the
ratio measurement feature that corrects for color interference. The laboratory
turbidimeters are not portable for field analysis, but they can be used in an onsite mobile
lab. However, for ease of use and smaller footprint, the 2100Q is adequate for most
produced and flowback water samples. If the sample is over range, a simple 1:1 dilution
should lower the turbidity concentration within the 0 to 1000 NTU range. High levels of
color may cause high results.
The challenge for the gravimetric TSS procedure is that this application is a laboratory
test. It is possible to perform the analysis in a mobile lab, but there is an abundance of lab
equipment needed for this procedure, i.e. oven, analytical balance, vacuum pump,
desiccators, etc. The method requires moderate laboratory skill, and test results are not
available for 3 to 4 hours. Samples having a high TSS load or high oil residual levels may
require sample size adjustments.
The TSS portable probe is ideal for measuring suspended solids in the field. Its 10 m cable
allows the probe to be lowered into the storage container to spot check the suspended
solids or turbidity of the produced or flowback water. The probe is best used after
calibration or correlation to the gravimetric TSS procedure. The TSS probe also provides
immediate results for process control and reduces the need for the time consuming
suspended solids lab analysis. Samples having high levels of oil or hydrocarbon residuals
may coat the probe and require routine cleaning. Samples having variable color or
particulate size can cause variable test results.
11
Hach Company/Hach Lange GmbH, 2011-2012. All rights reserved. Printed in U.S.A.