Ultrasound Ultraprobe 10,000 Instructions

Steam Trap Inspection

Rev. 1.0

3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES

2-5 min

UE Systems Training

Needed: Ultraprobe 10,000 and CF card;


Stethoscope Module; Rubber Focusing Probe;
Headphones

Move Cursor to Function Bar by Pushing


Sens.Knob 2X and Press Enter for Setup Menu

Page 1 of 6

TQC

Work

Extension Rods for Stethoscope Module when


needed for Hard To Reach areas. NOTE: Never
Use Magnet for Steam Trap Inspection

If Using a Route, move Route to Flash Card, then


load to Ultraprobe by using Data Transer. Refer to
Instructions "Loading Route."

*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*

Verify

Ultrasound Ultraprobe 10,000 Instructions


Steam Trap Inspection
Rev. 1.0

3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES

2-5 min

UE Systems Training

TQC

Work

Be sure Instrument Setup is in Automatic, if


Baselining Manual. Load Group (Route) to
Ultraprobe.

If you are PERFORMING the BASELINE Readings,


proceed to NEXT INSTRUCTION, if not READ the
Picture Below.

Rotate for Application Select, Push Sens.Knob.


Rotate to Steam, push Sens.Knob (Enable Steam).
Press Enter to Exit.

A clean memory is a must to keep files organized.


It is Normal to Delete Records when changing
apps.

Page 2 of 6

*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*

Verify

Ultrasound Ultraprobe 10,000 Instructions


Steam Trap Inspection
Rev. 1.0

3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES

2-5 min

UE Systems Training

TQC

Work

Rotate to Module Type Select, push Sens.Knob.


Rotate to Stethoscope (STM); Push Sens to
Select/Exit

10

Rotate to Instrument Setup, Push Sens.Knob.


Rotate for "Manual", Push Sens.Knob to Select/Exit

11

Rotate to Enter Inspector ID [under "More" in Setup


Menu], push Sens.Knob. Enter 3 Alpha-Numeric
Characters use Sens.Knob.

12

Press Enter 2X to Save/Exit and return to Main


Display.

Page 3 of 6

*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*

Verify

Ultrasound Ultraprobe 10,000 Instructions


Steam Trap Inspection
Rev. 1.0

3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES

2-5 min

UE Systems Training

TQC

Work

13

IF Route has been baselined and established refer


to NOTE. Note does not hold true for
troubleshooting WITHOUT having a route loaded.

14

Push Sens.Knob 1X to move cursor to Frequency.


Rotate to 25 kHz.

15

Push Sens. 1X more [if cursor at S= push 2X]


moving Cursor to Function Bar. Rotate Sens until
Record WAV in prep for readings.

16

Push Sens.Knob 1X more to return cursor to S=


(Operational Home Spot for Adjusting Sensitivity).

Page 4 of 6

*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*

Verify

Ultrasound Ultraprobe 10,000 Instructions


Steam Trap Inspection

17

19

Rev. 1.0

3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES

2-5 min

UE Systems Training

Choose Test Point Near Discharge Orifice


(DownStream) of Steam Trap

Press Enter To Begin Recording Process of Steam


Trap, Record for at least 15 secs or until several
cycles of Trap occur.

Page 5 of 6

TQC

Work

18

Never Leave Stethoscope Module on Steam Trap


For Lengthy Periods. Heat will Transfer to Module
over Time.

20

Note BOTH screens confirmed before moving to


next point. After confirmation screens, unit returns
to Main Display, advanced to next point.

*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*

Verify

Ultrasound Ultraprobe 10,000 Instructions


Steam Trap Inspection

21

Rev. 1.0

3/2012 SAFETY: Wrist Strap to be Worn at ALL TIMES

2-5 min

UE Systems Training

Visual Instructions exist for Storing a WAV. A


sound recording is stored for Analysis of Steam
Trap in UE Spectralyzer Software.

Page 6 of 6

TQC

ALWAYS WEAR the HEADPHONES.


22

*Compact Flash Card & Module Inserted BEFORE Turning Ultraprobe ON*