760

Ventilator System
Exceptional ValueEasy to operate and service, the affordable 760 Ventilator is
ideal for both metropolitan hospitals and remote medical facilities.
Technological SophisticationHigh performance gas delivery system, combined with
frictionless piston and flow triggering, make the ventilator extremely responsive to
patients breathing efforts without the need for a compressed air source.
Powerful SystemThe 760 Ventilator delivers mandatory breaths in both volume
control ventilation and pressure control ventilation.
Simplicity Outside,
Sophistication Inside
The Puritan Bennett

760

Ventilator System provides


comprehensive, affordable
respiratory care in the Intensive
Care Unit. This easy-to-operate
ventilator offers all the
capabilities of the 740

Ventilator, plus added features


such as pressure control
ventilation to support pediatric
and adult patients with a wide
variety of clinical conditions.
Puritan Bennett is
one of the worlds
premier producers
of respiratory-related
medical products
serving patients
from the hospital
to the home.
Standard Accessories Catalog Number
Collector Vial Assembly G-061668-00
D/X7

Expiratory Bacteria Filter (case of 12) G-060526-00


Re/X 700

Expiratory Bacteria Filter, reusable, 22-mm ISO


connectors; one comes standard with every ventilator G-060525-00
D/Flex

Inspiratory Bacteria Filter, single-patient use,


22-mm ISO connectors; one comes standard with every
ventilator (carton of 12) 4-074601-00
Test Lung 4-000612-00
Flex Arm, brushed aluminum 4-032006-00
Selection Required Catalog Number
Hose Assembly, oxygen
DISS female x female (United States) 4-001474-00
DISS female x male (Canada) G-061269-00
NIST (United Kingdom) G-061201-00
Air Liquide (France) G-061190-00
Australia G-061198-00
Drger (Germany) 5-029059-00
Power Cord
North America and Japan G-061241-00
Australia G-061242-00
Continental Europe G-061243-00
Denmark G-061244-00
India/South Africa (old British-style plug with round prongs) G-061247-00
Israel G-061248-00
Italy G-061245-00
Switzerland G-061246-00
United Kingdom G-060135-00
Power Supply Kit
English 110 V G-062390-00
English 220 V G-062391-00
French Canadian 110 V G-062386-00
French 220 V G-062387-00
Spanish 110 V G-062394-00
Spanish 220 V G-062395-00
Portuguese 110 V G-062397-00
Portuguese 220 V G-062398-00
Japan 110 V G-062401-00
Japan 220 V G-062402-00
German 220 V G-062405-00
Italian 220 V G-062407-00
Russian 220 V G-062409-00
Poland 220 V G-062411-00
Manual, Operators
Czech G-061988-00
English G-061874-00
French G-061981-00
German G-061980-00
Italian G-061982-00
Japanese G-061987-00
Polish G-061985-00
Portuguese G-061984-00
Russian G-061986-00
Spanish G-061983-00
760 Ventilator System
Modes of Ventilation
Assist/Control (A/C)
Synchronous intermittent manda-
tory ventilation (SIMV)
Spontaneous (SPONT)
Breath Types
Volume controlled ventilation
(VCV), available in A/C or
SIMV modes
Pressure control ventilation (PCV),
available in A/C or SIMV modes
Pressure support ventilation (PSV),
available in SPONT or SIMV
modes
Mandatory (VCV) Settings
Respiratory rate: 1 to 70 breaths/min
Tidal volume: 40 to 2,000 mL,
compliance-compensated and
corrected to body temperature
and pressure, saturated (BTPS)
Peak flow: 3 to 150 L/min in A/C
and SIMV modes (mandatory
breaths), 300 L/min for
spontaneous breaths
Plateau: 0.0 to 2.0 seconds
Flow pattern: Square or
decelerating ramp
Mandatory (PCV) Settings
Respiratory rate: 3 to 70 breaths/min
Inspiratory pressure: 5 to 80 cmH
2
O
Inspiratory time: 0.2 to 8 seconds
I/E Ratio: 1:99 to 4:1
Rise time factor: 5 to 100
Spontaneous (PSV) Settings
Support pressure: 0 to 70 cmH
2
O
(0 to 70 hPa)
Rise time factor: 5 to 100
Exhalation sensitivity: 1% to 80%
Common Settings
PEEP/CPAP: 0 to 35 cmH
2
O
(0 to 35 hPa)
Trigger sensitivity: 1 to 20 L/min
% O
2
: 21% to 100%
Operator-Set Alarm Limits
Alarms designed to ISO and
European standards, and are
classified as high or medium
priority, with distinct audible and
visual presentations.
High rate: 3 to 100 breaths/min
Low insp pressure: 3 to 60 cmH
2
O
(3 to 60 hPa)
High pressure: 10 to 90 cmH
2
O
(10 to 90 hPa)
Low minute volume: 0 to 50 L
High tidal volume: 20 to 6,000 mL
Low tidal volume: 0 to 2,000 mL
Ventilator Status Indicators
ALARM
CAUTION
NORMAL
VENT INOP
SAFETY VALVE OPEN
ON AC/BATTERY CHARGING
ON INTERNAL BATTERY
ON EXTERNAL BATTERY
INTERNAL BATTERY LEVEL
Other Keys and Indicators
MESSAGE WINDOW: Shows up to
four lines, 20 characters each.
100% O
2
: Delivers 100% O
2
to the
patient for 2 minutes, then returns
to the current %O
2
setting.
MENU: View active and reset
alarms; run self tests; enter stand-
by mode; adjust settings including
alarm volume, apnea interval, VCV
flow type, ET tube size, humidifier
type and date/time, PCV timing
setting, and volume LED bar);
access oxygen sensor functions
(calibrate, enable/disable %O
2
alarm limits, enable/disable display
of sensor readings); display bat-
tery, software revision, and service
information.
SPEAKING VALVE MODE:
Allows clinician to ventilate
patients on speaking valve while
monitoring delivered volume.
APNEA PARAMS: Allows clinician
to select apnea ventilation
settings (respiratory rate, tidal
volume, peak flow). Apnea
interval is programmable from
10 to 60 seconds.
MANUAL INSP: Delivers one
mandatory breath.
SIMPLICITY OUTSIDE, SOPHISTICATION INSIDE
For more information, call 1-800-635-5267 or visit www.puritanbennett.com.
KNOB: Adjusts values or selects menu
options. A setting or alarm pulses in its
display window to indicate that the knob is
linked to the setting.
CURRENT: Lights to indicate that displayed
settings are in effect.
PROPOSED: Lights to indicate that displayed
mode, breath type, and settings are being
proposed.
ALARM SILENCE: Silences audible alarm
for 2 minutes.
ALARM RESET: Clears all alarm indicators.
Alarm reactivates if alarm condition persists.
CLEAR: Cancels proposed settings.
ACCEPT: Makes proposed settings effective.
INSPIRATORY PAUSE: Measures inspiratory
resistance and static compliance.
EXPIRATORY PAUSE: Measures intrinsic and
total PEEP.
Pneumatic Gas Sources
Air: Gas delivery system based on frictionless
piston design eliminates the need for a
compressor or hospital-grade wall air.
Oxygen: Must be supplied at 40 to 90 psi
(275 to 620 kPa) and at least 160 L/min
at standard temperature and pressure,
dry (STPD).
Environmental Specifications
Temperature
Operating: 41 to 113F (5 to 45C) at 10% to
95% relative humidity
Storage: -40 to 140F (-40 to 60C) at 10% to
95% relative humidity
Atmospheric pressure
Operating: 8.7 to 16.0 psi (600 to 1,100 hPa)
Storage: 7.3 to 16.0 psi (500 to 1,100 hPa)
Altitude
Operating: 4,570 m (15,000 ft) maximum
Storage: 15,240 m (50,000 ft) maximum
Power
Power supply input range: 100 to 120 V AC
nominal, 220 to 240 V AC nominal; 50/60 Hz;
2.9 A (100 to 120 V units) or 1.3 A (220 to
240 V units); 290 VA.
Earth leakage current (including ventilator,
power cord, and internal and external
batteries): Less than 300 A at 100-120 V AC
operation; less than 500 A at 220-240 V AC
operation.
Enclosure leakage current: Less than 50 A in
Normal Condition (all units); less than 300 A
in Single Fault Condition at 100-120 V AC
operation; less than 500 A in Single Fault
Condition at 220-240 V AC operation.
Internal battery
Operating time: Approximately 2.5 hours under
nominal conditions. Approximately 2 hours
under extreme conditions.*
Recharge time: 2.5 to 3 hours in ventilator.
Charges automatically while ventilator is
connected to AC power and power switch
is ON (including standby mode).
Charge level: Indicated on keyboard.
External battery
Operating time: Approximately 7 hours under
nominal conditions. Approximately 5.5 hours
under extreme conditions.*
Recharge time: 7.5 to 8 hours in ventilator, 3 to
4 hours using optional battery charger.
Charges automatically while ventilator is con-
nected to AC power and the power switch is
ON (including standby mode). Charge level
available under menu function.
* Nominal conditions: tidal volume 0.6 L, respiratory
rate 15/min, PEEP/CPAP 5 cmH
2
O, peak flow
60 L/min, plateau 0 seconds, peak pressure 30
cmH
2
O, mean pressure 8 cmH
2
O.
Extreme conditions: tidal volume 1.2 L, respiratory
rate 15/min, PEEP/CPAP 15 cmH
2
O, peak flow
60 L/min, plateau 0 seconds, peak pressure 64
cmH
2
O, mean pressure 24 cmH2O. Battery life
specifications are approximate and are for new,
fully charged batteries at 68F (20C) and sea level.
To ensure maximum battery life, maintain full charge
and minimize the number of complete discharges.
Compliance and Approvals
CE marked, indicating compliance to the
Medical Device Directive (93/42/EEC).
IEC 601-1 classification: Protection class I,
Type B, internally powered, drip-proof
equipment, continuous operation.
Complies with these International and
European standards:
IEC 601-1, EN 60601-1, IEC 601-1-2,
EN 60601-1-2, EN 794-1.
Approved to the type test requirements of
Annex III of the Medical Device Directive by
the notified body, TUV Rheinland.
Certified by these test agencies:
TUV Rheinland: EC Type Test Certificate to
the requirements of Annex III of the Medical
Device Directive.
CSA: C22.2 No. 601-1, CSA C22. 2 No. 601-2-
12, UL 2601-1, and NRTL certification.
Physical Characteristics
Weight
Ventilator only: 66 lb (30 kg)
Cart only: 45.5 lb (20.7 kg)
External battery: 28 lb (12.75 kg)
Shipping weight
Ventilator only: 118 lb (53.5 kg)
Cart only: 57 lb (23.6 kg)
Dimensions
Ventilator only:
14.9'' H x 20.3'' W x 14.6'' D
(378 mm H x 515 mm W x 370 mm D)
Ventilator and cart:
47.0'' H x 26.0'' W x 23.5'' D
(1,194 mm H x 660 mm W x 597 mm D)
Shipping dimensions
Ventilator only:
26.8'' H x 27.6'' W x 31.5'' D
(680 mm H x 700 mm W x 800 mm D)
Cart only:
36.0'' H x 26.0'' W x 12.0'' D
(915 mm H x 660 mm W x 305 mm D)
Connectors
Inspiratory limb connector: ISO 22-mm
conical male
Expiratory limb connector: ISO 22-mm
conical male to collector vial
Oxygen inlet: DISS male, DISS female, NIST,
Air Liquide, Drger, or Australian fitting
Gas exhaust port: ISO 30-mm conical
Mounting
Cart or shelf mount allows maximum flexibility
Shelf requirements: Able to hold 220 lb
(100 kg) (to accommodate ventilator,
humidifier, collector vial, external battery,
and brackets); 12'' (300 mm) deep minimum
for access to fan filter; 40'' (1,000 mm)
clearance above shelf; 20'' (500 mm)
clearance below shelf; 36'' (900 mm)
width (room for external battery)
Oxygen sensor life
2 years, nominal (actual life depends on
operating environment)
Alarm volume
74 to 88 dBA at 1 m (user-adjustable)
Communications
An optional communications panel allows
the 760 Ventilator to interface with a number
of external devices including a stand-alone
remote alarm, nurse call systems, central
stations and information systems, such
as CliniVision

Clinical Informatin System.


4280 Hacienda Drive
Pleasanton, CA 94588
Tel 925.463.4000
Toll Free 1.800.635.5267
www.puritanbennett.com
Tyco Healthcare UK LTD.
154 Fareham Road
Gosport, UK PO13 0AS
Tel +44.1329.224000
Specifications subject to
change without notice.
2003 Nellcor Puritan
Bennett Inc. All rights
reserved.
O.f 1213-0603
VE05100
Optional Accessories Catalog Number
Ventilator Breathing Circuits
Reusable, with water traps, adult G-061208-00
Reusable, with water traps, pediatric G-061223-00
Reusable, with heated wire, for Fisher & Paykel Humidifiers, adult G-061235-00
Reusable, with heated wire, for Fisher & Paykel Humidifiers, pediatric G-061237-00
Water Trap, in-line G-061213-00
D/X7

Expiratory Bacteria Filter, single-patient use (carton of 12) G-060526-00


Re/Flex

Inspiratory Bacteria Filter, reusable, 22-mm ISO connectors (each) 4-074600-00


Humidifier Bracket, adaptable G-062612-00
Shelf-Mount Kit, adaptable to most rail systems; includes brackets
for mounting collector vial and humidifier for use with Fisher & Paykel Humidifiers G-061279-00
for use with Hudson RCI ConchaTherm Humidifiers G-061601-00
External Battery Kit, designed for easy mounting on ventilator cart but can be used
without cart; includes battery and cover to protect battery terminals G-061140-00
External Battery Charger
North America and Japan G-061260-00
Australia G-061261-00
Continental Europe G-061500-00
Denmark G-061501-00
India/South Africa (old British-style plug with round prongs) G-061504-00
Israel G-061505-00
Italy G-061502-00
Switzerland G-061503-00
United Kingdom G-061499-00
Filter, air intake, package of 6, to be replaced every 1,000 hours of operation;
one comes standard with every ventilator G-061262-00
Filter, main fan, package of 6, to be replaced every 250 hours or as needed;
one comes standard with every ventilator G-061263-00
Standard Cart G-062611-00
Standard Cart Kit designed for in-house transport; includes dual oxygen cylinder
mounting kit and adaptable humidifier bracket. Order humidifier mount separately. G-062616-00
Dual E Cylinder Holder Bracket G-062613-00
Linkage Arm Kit, used to connect non-standard wheelchair to EasyCart G-061845-00
Humidifier Mounts
For use with Fisher & Paykel Humidifiers G-061227-00
For use with Hudson RCI ConchaTherm Humidifiers G-061228-00
Manual, Service
English G-061875-00
Preventive Maintenance Kits
15,000-hour G-061166-00
30,000-hour G-061167-00
Keyboard cleaning spray G-061576-00
O2 Regulator, for E cylinder (2 x DISS & 0-25 L/min) 6-900003-00
760 Ventilator System
UNIVERSAL RESERVOIR
Standard Cart