Professional Documents
Culture Documents
1.What is Validation?
At sites where products are developed and manufactured, a variety of equipment is used to create products through a complex series of work processes. To state that such products meet the expected quality standards, it is important to verify that the plant, equipment, and operational procedures are free of problems. Verification procedures must be established to ensure that the details of the verification are consistent, regardless of who performs it. The series of processes from performing the verification according to a defined procedure to documenting the verification results is generally known as "validation". Validation can be performed on a diverse range of objects, from tangible items such as plant and equipment to intangible work procedures and processes. This UV TALK LETTER describes details about the instrument validation of UV-VIS spectrophotometers.
Performance Items A spectrophotometer shines light at various wavelengths onto the sample and investigates the degree of absorption, reflection, and transmission of the light to perform qualitative or quantitative analysis of the sample. So, what sort of performance does a spectrophotometer offer? JIS K0115 "General rules for molecular absorptiometric analysis" prescribes the performance items that should be displayed by the instrument, as shown in Table 1. It can be seen from Table 1 that Table 1 Performance Items Listed in JIS K0115 Resolution Wavelength accuracy Wavelength repeatability Photometric accuracy Photometric repeatability Stray light Baseline stability Baseline flatness Noise level
the generic term "performance" encompasses various conditions depending on the point of focus. Instrument validation of a spectrophotometer involves selecting the items required to manage and determine the status of the instrument from among these performance items, andverifying them.
Wavelength Accuracy
Fig. 1 Emission Lines of a Deuterium Lamp The emission lines of a deuterium or low-pressure mercury lamp or the absorption peaks of an optical filterfor wavelength calibration are generally used to verify the wavelength accuracy. Fig. 1 shows the energy spectrum of a deuterium lamp. A deuterium lamp is known to exhibit sharp energy peaks (emission lines) at 656.1 nm and 486.0 nm wavelengths. Consequently, the instruments wavelength accuracy can be verified by measuring the energy spectrum of a deuterium lamp, investigating the wavelength of the peak near 656.1 nm, and then comparing its wavelength value to 656.1 nm. For example, if the detected peak wavelength is 656.2 nm, the error from the true 656.1 nm value is 0.1 nm, and this becomes the wavelength accuracy of the instrument.
How many multiple emission lines (or absorption peaks) should be used to confirm the wavelength accuracy and what error is permitted depends on the level of instrument performance required for the development and manufacture of the product. In cases where a sample spectrum is measured and the peaks must be specified within an error of 1 nm, a wavelength accuracy of 0.1 nm is probably adequate.
Stray Light
Stray light is light outside the specified wavelength that shines onto the sample. For example, when measuring the absorbance using light at 220 nm, accurate measurements are not possible if a lot of light at wavelengths other than 220 nm hit the sample. Let us consider the case of 0.01 % stray light outside the specified wavelength. Due to the effects of the stray light, a sample with 1 % transmittance (2 Abs) appears to have a transmittance of 1.01 % (1.9957 Abs). Alternatively, a sample with 0.01 % transmittance (4 Abs) appears to have a transmittance of 0.02 % (3.6990 Abs). That is, errors of 0.0043 Abs and 0.3010 Abs, respectively, occur. This case reveals that the effects of stray light increase as the sample absorbance increases.
Fig. 2 Stray Light Measurements Using Nal Aqueous Solution The presence of stray light causes distortion in thecalibration curve. Consequently, an instrument with low stray light is required for the quantitation of high-concentration unknown samples (with high absorbance) using a calibration curve created with standard samples.
An aqueous solution that is known to not transmit light at a specific wavelength, such as sodium iodide (Nal), is used to evaluate the amount of stray light. For example, an aqueous solution of Nal does not transmit light at 220 nm. Initially, the transmittance is measured with a shutter block that
completely shuts out all light mounted in the sample compartment (transmittance X). Next, the transmittance is measured with the aqueous solution of Nal in the sample compartment (transmittance Y). The instruments stray light is defined as (Y X). This value is used to evaluate the level of stray light. Fig. 2 shows a schematic diagram of this situation.
The aim of instrument validation is to confirm that the instrument offers adequate performance for the inspectionand manufacture of products. When performing an actual validation, it is important to select the appropriate inspection items and to set the evaluation criteria based on a sound understanding of the required instrument performance.
Fig. 3 Noise Level Measurement Results A spectrophotometer is made up of many components, some of which are consumables that deteriorate according to the time and frequency of operation. Instrument validation is also useful for determining the state of the constituent components of the instrument. Let's consider the noise level as an example. The noise level is one indicator of the condition of the light source (lamp) in the spectrophotometer. The noise level is defined as the maximum deviation (maximum distance between peak and trough) of the absorbance measured over one minute at a specific wavelength near 0 Abs. Fig.3 shows the graph of a measured noise level. As the relative noise increases when the emitted light intensity of the lamp drops off over time, the noise level becomes higher. A higher noise level means that the data reproducibility is lower. This has a negative effect when accurate photometric values are required. When it is necessary to detect extremely small absorption peaks, it may not be possible to detect the peaks correctly if the noise level is so high that they become buried in the noise.
The optical system of a spectrophotometer contains various mirrors for focusing and forming a spectrum. The surfaces of these mirrors can deteriorate over time. It is also possible for dust and dirt from the atmosphere to adhere to the mirror surfaces due to the environment where the instrument is installed. For example, the deterioration of a mirror that forms the spectrum can be one cause of the increase in stray light described above. Therefore, instrument validation can provide valuable information to diagnose the condition of an instrument. It is recommended to perform periodic instrument validation to continuously understand and manage the state of the instrument. It is also important to validate an instrument after consumables are replaced or the instrument is moved to a different place.
6. Conclusions
This UV TALK LETTER introduced the following points:
Instrument validation is important to determine the status of instruments used for the development and manufacture of products. To perform an actual instrument evaluation, select the required performance items and set the appropriate evaluation criteria for the inspection. The effective application of a validation program permits accurate and efficient instrument validation.
Validation of hardware
IRPrestige-21
This time, I will talk about how the validation of FTIR. There is a meaning "(legally) to enable" and to validate in English. After you shop at department stores, is called the (are pushed to be Valid in the U.S.) with whom to press the stamp of "free" to the parking ticket at the cash register also validate in English. To prove the "validity, validate to authenticate - since it is to ensure that you have the inspection required in equipment, the equipment is operating correctly," I can use "and that" Enable " It is used in the sense that "the noun validation is now also used to mean "verification, validation, validation, and validation." So, I will talk about software validation and hardware validation of FTIR. You said that validation of hardware of FTIR, and as the method to examine and verify that the FTIR is operating correctly. In distributed IR, by measuring the spectrum of polystyrene film, I was checking by reading the peak wave number and also serves as a calibration of the wave number. That the state of the device appears in the power spectrum because there are many in FTIR, how to determine the size and shape of the power spectrum is used as a simple way. As daily inspections in our FTIR instrument, and has adopted the confirmation of the size and shape of the power spectrum. As how to check the FTIR finer, public institutions as well as a number of issues a standard. Japanese Industrial Standards (JIS), as a standard industrial-based American Society etc. (JIS ASTM, the American version) for Testing and Materials, there is a provision Japanese Pharmacopoeia, the European Pharmacopoeia (EP) in the pharmacopoeia. Let's take a look at each one defines what kind of standard.
specified variations in the measured value of the transmittance and the wave number.
PAGE TOP
Pharmacopoeia first amendment. It is evaluated to be acceptable and to determine the difference in transmittance of each if both exceed the standard value
2870cm absorption spectra were measured absorption spectra of polystyrene film of about 0.04mm thick, the -1 -1 resulting 2850 cm and minima near difference transmittance (%) at the maximum in the vicinity of 18% or -1 -1 more, and, 1589 cm and 1583Cm minimum of around and the difference of transmittance (%) at the maximum in the vicinity is 12% or more
Wavenumber accuracy and is described as the right to EP2000 and fourteenth amendment Japanese Pharmacopoeia first supplement on performance tests and on the performance of the wave number accuracy. The calculated at the wave number used to determine the peak wavenumber spectrum of polystyrene film was measured to determine whether it contains the allowable range in this program. It is judged acceptable peak wavenumber of all specified if within the allowable range.
(Cm wave number scale, characteristic absorption wave number of the following convention, a polystyrene film 1 of), corrected using some. In addition, Figures in brackets represent the acceptable range of values of these. -1 3,060.0 ( 1.5) cm -1 1,583.0 ( 1.0) cm -1 2,849.5 ( 1.5) cm -1 1154.5 ( 1.0) cm -1 1,942.9 ( 1.5) cm -1 1,028.3 ( 1.0) cm -1 1601.2 ( 1.0) cm
Wavenumber reproducibility reproducibility of the wave number, and is described as the right to the fourteenth amendment Japanese Pharmacopoeia first amendment. Specify three points the peak wave number in this program, is obtained in two measurements of polystyrene film the peak wave number of the actual at each point, and then determined by whether or not within the allowable range error between them. It is judged acceptable peak wave number of the wave number if the given within the allowable range all.
Transmission and 3000 ~ 1000cm polystyrene film reproducibility of the wave number, when measuring was -1 repeated twice and the absorption at several points in, 3000cm , the difference in wave number 5 cm near 1 -1 -1 or less, 1000 cm 1 cm in the vicinity to and within.
-1
It should be noted that the wavenumber reproducibility is not listed in the EP2000.
PAGE TOP
Reproducibility of the transmittance for the reproducibility of the transmittance, in the EP2000 and fourteenth amendment Japanese Pharmacopoeia first amendment, and are described as below.
Transmission and 3000 ~ 1000cm polystyrene film reproducibility of the wave number, when measuring was repeated twice and the absorption at several points in the difference in the transmittance was 0.5% or less, ... and so.
-1
Specify three points a peak wavenumber This program is found in the two measurements the transmittance at each point is determined by whether or not within the allowable range error between them. It is judged acceptable difference in transmittance If it contains all of the allowable range. Behavior of the program validation program of Japanese Pharmacopoeia compatible, comes standard in IRsolution software that controls the FTIR-8400S/IRPrestige-21. Just follow the message you can do so simply by selecting from a menu of IRsolution, and displayed on the screen, you can do. You can check in all automatically when using the kit takes out external beam polystyrene film is mounted. When the measurement is completed, the report summarizing the test results are printed automatically. Validation execution result enlarged view power / polystyrene spectrum enlarged view
It is recommended to NIST SRM 1921a as of polystyrene film, but it is a situation where it is not available there is no stock in the NIST 27 June 2005. So I also recommend NIST traceable Polystyrene Test film for the secondary standard NIST-traceable International Crystal Laboratories of (ICL) made. This article is Shimadzu GL Sea and we deal with (list price 38,100 yen).International Crystal Laboratories of (ICL) made NIST traceable I also recommend Polystyrene Test film. This article is
Item
Decomposition
Wavenumber accuracy
3,027.1 ( 0.3) 2924 ( 2) 2,850.7 ( 0.3) 1944 ( 0.3) 1,871.0 ( 0.3) 1801.6 ( 0.3) 1,601.4 ( 0.3)
1583.1 ( 0.3) 1,181.4 ( 0.3) 1154.3 ( 0.3) 1,069.1 ( 0.3) 1,028.0 ( 0.3) 906.7 ( 0.3) 698.9 ( 0.5)
(Cm wave number scale, characteristic absorption wave number of the following -1 convention, a polystyrene film of), corrected using some. The values in parentheses indicate the range of acceptable values for these values . 3,060.0 ( 1.5) cm -1 2,849.5 ( 1.5) cm -1 1,942.9 ( 1.5) cm -1 1601.2 ( 1.0) cm
-1
-1
Validation of software