Professional Documents
Culture Documents
01-999121-00 K 0606
Overview
AutoSwitchable probes and 4 Nucleus probes are (listed in Table 1) combine simplified 4-nucleus operation with the versatility of multinuclear broadband nuclear observation. Table 1. AutoSwitchable and 4 Nucleus Probes
Magnet (MHz) 200 200 200 PFG 200 PFG 300 300 300 PFG 300 PFG 300 300 PFG 300 300 PFG 400 400 400 PFG 400 PFG 400 PFG 400 400 Bore (mm) 54 89 54 89 54 89 54 89 54 54 89 89 54 89 54 54 89 54 89 Sample Range (mm) 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5
1 1
Probe Part No. 01-910569-00 01-910570-00 01-907876-00 01-907875-00 01-910572-00 01-910573-00 01-910575-00 01-91197-00 01-903109-05 01-904083-05 01-903110-05 01-904084-05 01-910550-00 01-910581-00 01-910583-00 01-904089-06 01-904090-06 01-910585-00 01-910586-00
1H/19 F/13C/31P 1 1H/19 F/13C/31P 1H/19 F/13C/31P 1H/19 F/13C/31P 1H/19 F/13C/31P 1H/19 F/13C/31P, 1H/19F/{15N31P} 1H/19 F/13C/31P, 1H/19F/{15N31P} 1H/19 F/13C/31P, 1 H/19F/{15N31P} 1H/19 F/13C/31P, 1H/19F/{15N31P} 1H/19 F/13C/31P 1H/19 F/13C/31P 1H/19 F/13C/31P 1H/19 F/13C/31P, 1H/19F/{15N31P} 1H/19 F/13C/31P, 1H/19F/{15N31P} 1H/19 F/13C/31P, 1H/19F/{15N31P} 1H/19 F/13C/31P, 1H/19F/{15N31P}
ATB (Automation Triple Resonance Broadband) probes (listed in Table 2) support a wide possible array of liquids NMR techniques in an automation environment where performance is critical. The inner coil of this probe is broadband tunable and is constructed with circuitry to completely eliminate the need to tune this coil on a sample to sample basis. The outer coil is simultaneously tuned to 19F and 1H making X-{1H, 19F} indirect detection experiments and 1H or 19F direct observe NMR experiments convenient. Table 2. Automation Triple Resonance Broadband Probes
Magnet (MHz) 300 ATB PFG 300 ATB PFG 400 ATB PFG 400 ATB PFG Bore (mm) 54 89 54 89 Sample Range (mm) 5 5 5 5
1H/19 F/{15N31P} 1H/19F/{15N31P} 1H/19 F/{15N31P} 1H/19 F/{15N31P}
01-999121-00 K 0606
Safety Precautions
Safety Precautions
CAUTION: CAUTION:
Tighten a capacitor stick to finger-tight. Overtightening can damage the probe. Do not attempt to tune the AutoSwitchable, 4 Nucleus, or ATB probe before thoroughly reading the instructions in this manual. Any slight altering of the tuning on the AutoSwitchable can completely throw off the delicate balance between the two nuclei resonant on a port.
01-999121-00 K 0606
a. b. 3.
Connect the appropriate filters (low-pass and lock reject) to the lowband preamplifier. Connect the lowband cable between the filter and the OBS connector on the probe. Connect a notch filter (2H bandpass notch) to the lock preamplifier. Connect the lock cable between the filter and the LOCK connector on the probe.
Tuning Controls
The tuning controls for the AutoSwitchable and ATB probes extend from the bottom of the probe base. Shown in Figure 1 is the base of the AutoSwitchable. The base of the ATB is similar.
Guide bore for 1/4-wave and capacitor sticks (shown with 1/4-wave stick) High-band tuning controls (shown in UP positions)
High-band match
AutoSwitchable Probes
The tuning controls of this probe are operated with the coaxial, brass-colored tuning knobs. Both the high-band channel (labeled 1H/19F in red) and the low-band channel (labeled X-CHANNEL in green) are tuned by independent controls that control two resonances for each channel. The match function is controlled by the smooth knob on the bottom. The knurled knob controls two tune functions: DOWN position the word DOWN is visible and the gear system for the counter is engaged (for X-channel tuning), UP position with the word UP is visible and the gear system engaged.
01-999121-00 K 0606
Tuning Characteristics
ATB Probes
Tuning the 1H/19F channel This probe tunes much like the AutoSwitchable probe for 1H/19F tuning. Follow the instructions appropriate for the system: Tuning the X channel The X-Channel tuning knob on the Automation Triple Resonance Broadband probe does not have an UP and DOWN position. X-Channel tuning procedures for the ATB probe are the same as the Autoswitchable when it is operated with the X-channel tuned to a single nucleus. This procedure is provided in Tuning the Low-Band Channel, page 8. The next sections describe the tuning characteristics of the tune control.
Tuning Characteristics
High-Band
Figure 2 shows the tuning characteristics of the high-band tune controls for AutoSwitchable, 4 Nucleus, and ATB probes. Both resonances are moved simultaneously with the knob in the DOWN position. Clockwise rotation of the knob increases the frequency. The 1H resonance is moved with the knob in the UP position. Clockwise rotation of the knob increases the separation and the resonance frequency.
DOWN position UP position
19F
1H
19F
1H
01-999121-00 K 0606
DOWN position
UP position
31P
Tuning Ranges
Use the appropriate table as a guide for choosing capacitor or inductor sticks. Probe ranges are approximate and valid for empty probe only. Ranges are slightly lower with a sample in a probe.The ATB probe solvent tolerance is optimized when it is tuned to 13C.
400-MHz AutoSwitchable
Capacitor Stick (pF) none G 0 (blue knob) G 18 (blue knob) G 36 (blue knob) 1/4-wave (gold knob) Range (MHz) 17093 (31P, 13C) 9545.8 5945.5 46.239.5 (15N) 4-nucleus (13C/31P)
4-nucleus (13C/31P)
400-MHz ATB
Capacitor Stick (pF) none 6.8 (red knob) G 0 (blue knob) G 18 (blue knob) G 36 (blue knob) G 51 (blue knob) G 68 (blue knob) Range (MHz) 170116 (31P) 11891 (13C) 9270 7155.3 55.547.3 48.343.6 42.338.9 (15N)
01-999121-00 K 0606
CAUTION:
1.
Remove the capacitor stick from the protective shipping container. The stick value is labeled at the bottom end of the stick and color coded for easy identification: Red knobs regular capacitor sticks (not required for all probes) Blue knobs grounded capacitor sticks Gold knobs 1/4-wavelength sticks (not used on ATB probes) The top of the stick has a flat region that serves as a key for proper orientation of the capacitor stick in the probe.
2. 3. 4. 5.
Insert the stick into the probe bottom until the tip of the stick hit a barrier. Hold the plastic stick end (not the colored knob) and slightly rotate the stick until the key engage and the stick move inward slightly. Continue inserting the stick the rest of the way. Use finger tight. Do not overtighten. When the threads of the stick knob touch the bottom of the probe, twist the knob to affix the stick to the probe.
2.
3. 4. 5. 6. 7. 8.
01-999121-00 K 0606
01-999121-00 K 0606
easier return. Preset the counter to factory value, add match (clockwise) to find resonance, use DOWN position and match to optimize. 300-MHz probes AutoSwitchableuse the G 51pF capacitor stick (blue knob) ATBuse the G 75pF capacitor stick (blue knob) 400-MHz probes AutoSwitchableuse the G 36pF capacitor stick (blue knob) ATBuse the G 68pF capacitor stick (blue knob) 3. Return cables to the same position as connected prior to tuning.
Tuning 300-MHz AutoSwitchable Probe to 63130 MHz or the ATB Probe to 70125 MHz
The following steps tune a 300-MHz AutoSwitchable probe to nuclei in the frequency range of 63130 MHz or ATB probe to nuclei in the frequency range of 70125 MHz. 1. Choose the appropriate capacitor stick: AutoSwitchable Probe: 79130 MHz, no stick 6379 MHz, 6.8 pF stick ATB probe: 90125 MHz, no stick 7091 MHx 7.5 pF stick 2. 3. Preset the Broadband match knob to fully counterclockwise and then turn it several turns clockwise. Preset the counter by turning the Broadband tune knob. AutoSwitchable probe only engage the DOWN position and turn the tune knob to a value appropriate for the nucleus of interest. Find an approximate value by interpolating between 0 and the setting for 31P for frequencies above 79 MHz or interpolating between 0 and 80 for frequencies in the range of 6379 MHz. 4. 5. 6. Turn the Broadband tune knob (the UP position for tuning an AutoSwitchable probe) and the match knob to find the reflection minimum. Repeat with the tune knob (the UP position for tuning an AutoSwitchable probe) and the match knob to minimize the reflection meter reading. Record the counter reading for future reference.
Tuning 300-MHz AutoSwitchable Probe to 3063 MHz or the ATB Probe to 2971 MHz
The following steps tune a 300-MHz AutoSwitchable probe to nuclei in the frequency range of 3063 MHz. 1. 2. Select the appropriate capacitor stick, refer to the tables in Tuning Ranges, page 6. Turn the Broadband tune knob (the DOWN position for tuning an AutoSwitchable probe) position to find the resonance. If the probe has been operated at this frequency before, refer to the notes made at that time for the appropriate counter setting.
01-999121-00 K 0606
AutoSwitchable probe only the UP position of the tune knob has no effect with grounded capacitor sticks (blue knob). Leave the knob in the 4-nuc position for easier return to 4-nucleus operation. 3. 4. Repeat step 2 with the tune knob (the DOWN position for tuning an AutoSwitchable probe) and the match knob to minimize the reflection meter reading. Record the counter reading for future reference.
Tuning 400-MHz AutoSwitchable Probe to 94170 MHz or the ATB Probe to 116170 MHz
The following steps tune a 400-MHz AutoSwitchable probe to nuclei in the frequency range of 94170 MHz or 400-MHz ATB probe to nuclei in the frequency range of 116170 MHz. Use no capacitor stick. 1. 2. Preset the Broadband match knob to fully counterclockwise and then turn it several turns clockwise. Preset the counter by turning the Broadband tune knob (the DOWN position for tuning an AutoSwitchable probe) to a value appropriate for the nucleus of interest. If the probe has been operated at this frequency before, refer to the notes made at that time for the appropriate counter setting. Otherwise find an approximate value by interpolating between the settings for 13C and 31P. Turn the Broadband tune knob (the UP position for tuning an AutoSwitchable probe) and the match knob to find the reflection minimum. Repeat step 2 with the tune knob (the UP position for tuning an AutoSwitchable probe) and the match knob to minimize the reflection meter reading. Record the counter reading for future reference.
3. 4. 5.
Tuning 400-MHz AutoSwitchable Probe to 3994 MHz or the ATB Probe to 38.9118 MHz
The following steps tune a 400-MHz AutoSwitchable probe to nuclei in the frequency range of 3994 MHz or 400-MHz ATB probe to nuclei in the frequency range of 38.9118 MHz. 1. 2. Select the appropriate capacitor stick according to the tables in Tuning Ranges, page 6. Turn the Broadband tune knob (the DOWN position for are tuning an AutoSwitchable probe) to find the resonance. If the probe has been operated at this frequency before, refer to the notes made at that time for the appropriate counter setting. AutoSwitchable probe only note: note that the UP position of the tune knob has no effect with grounded capacitor sticks (blue knob), leave it in 4-nuc position for easier return to 4-nucleus operation. 3. 4. Repeat step 2 with the tune knob (the DOWN position for tuning an AutoSwitchable probe) and the match knob to minimize the reflection meter reading. Record the counter reading for future reference.
10
01-999121-00 K 0606
P C N
1. Counter settings are based on an empty probe. 2. Tuning in UP position does not affect the circuit when using grounded capacitor sticks.
1. Counter settings are based on an empty probe. 2. Tuning in UP position does not affect the circuit when using grounded capacitor sticks.
01-999121-00 K 0606
11
UP Tuner DOWN Tuner (Number of Turns) (Counter Number) 29 2 44 2 53 2 no effect counter # 3 counter # 3 counter # 8 counter # 4
31P 13 15
UP Tuner DOWN Tuner (Number of Turns) (Counter Number) 45 2 45 2 49 10 no effect counter # 1 counter # 1 counter # 2 counter # 2
(4-nuc)
12
01-999121-00 K 0606
Sample H Observe 90 Pulse Width, page 14 H Spinning Resolution & Lineshape, page 14 400 MHz systems: 200 and 300 MHz systems: 1% chloroform in acetoned6 5% chloroform in acetoned6 1% chloroform in acetoned6 5% chloroform in acetoned6 4Hz 1% H2O/99% D2O 0.1% ethylbenzene, 0.01% TMS 99.89% deuterochloroform (CDCl3) 0.05% trifluorotoluene in benzened6 0.05% trifluorotoluene in benzened6 0.0485 M triphenylphosphate in CDCl3 0.0485 M triphenylphosphate in CDCl3 40% p-dioxane in benzened6 (ASTM) 40% p-dioxane in benzened6 (ASTM) 40% p-dioxane in benzened6 (ASTM) 40% p-dioxane in benzened6 (ASTM) 0.1% ethylbenzene, 0.01% TMS 99.89% deuterochloroform (CDCl3)
00-968120-89 00-968120-99 00-968120-89 00-968120-99 00-901855-01 00-968120-70 00-968120-82 00-968120-82 00-968120-87 00-968120-87 00-968120-69 00-968120-69 00-968120-69 00-968120-69
1H
Spinning Sidebands, page 15 400 MHz systems: 200 and 300 MHz systems:
Sensitivity, page 17 Observe 90 Pulse Width, page 17 Sensitivity, page 17 Observe 90 Pulse Width, page 18 Sensitivity, page 18 Observe 90 Pulse Width, page 18 Spinning Resolution & Lineshape, page 19 Spinning Sidebands, page 20 Sensitivity, page 20
01-999121-00 K 0606
13
Do not repeat this calibration if the automated version from the VnmrJ Calibrate Probe window was run. 1. 2. 3. 4. 5. 6. Enter rtp('/vnmr/tests/') su Tune the probe on the ethylbenzene sample. Enter pw=5 at=1 lb=1 spin='n' Obtain the spectrum, and then adjust the phase for a positive signal. Place the cursor on the quartet and type movetof. Determine the correct tpwr as follows: a. b. c. Enter tpwr=50 pw=7 ga. After the acquisition is complete, enter aph to phase the spectrum. Set the pw to twice the pw90 specification for the probe and array tpwr from 48 to 58 (up to 62, if necessary). Observe the spectra as they are being collected. When the peak of interest passes through the null and goes negative, the acquisition can be halted by entering aa. The tpwr value corresponding to the first negative spectrum should then be used for pw90 calibrations. 7. 8. 9. Expand the spectrum so that the quartet is centered on the screen with wp=250. Enter ai. Set vs so that the signal occupies about half the vertical size of the screen.
10. Enter vp=50. 11. Set up the following array for determining the pw90 for a probe: a. b. Enter r1=xx/5, where xx is the specified pw90 specification for the probe. Enter array('pw',20,r1,r1) to create a pw array covering approximately 360 degrees of nutation.
12. Enter ga. 13. Enter dssh after the last spectrum is obtained. 14. The pw for the first minimum is the 180 pulse and this value is 90 pulse. 15. Write the results in the forms provided in Probe Test Results, page 21.
1H
This test measures the 50%, 0.55%, and 0.11% levels of the CHCl3 sample. This test and the 1H spinning sidebands test (the next test) must be passed simultaneously. 1. 2. 3. Put the appropriate CHCl3 sample in the probe and spin it. Enter rtp('/vnmr/tests/H1lshp') nt=1 vs=1000 su to set up the system hardware. Set the 1H 90 pulse width to the value determined during the calibration.
14
01-999121-00 K 0606
4.
Enter ga to acquire the spectrum. Phase the spectrum, then plot it using wp=250. Enter vs=10k and plot the expanded spectrum. Make sure the spectrum is normalized. If floor vibration results in excessive noise around the base of the peak, nt can be changed to a larger value (e.g., nt=4 or nt=16); however, if extreme vibrations are present, it may be impossible to measure the lineshape accurately.
CHCl3 peak
13
13
C satellite
5.
Measure lineshape as the linewidth of the CHCl3 peak at 50%, 0.55%, and 0.11% of the main peak amplitude. Refer to Figure 4 for the measurement points. Write the results in the forms provided in Probe Test Results, page 21.
60
40
20
6.
1H
Spinning Sidebands
Using the appropriate CHCl3 sample, measure the 1H spinning sidebands on the same spectrum as 1H lineshape. Use the spectra and parameter set from the 1H lineshape test. Plot the spectrum again using a large enough value of wp to show all the spinning sidebands. Measure spinning sideband amplitudes as a percentage of the main peak. Spinning sidebands occur at frequency intervals on either side of the central peak equal to the spinning rate. The sidebands may not be split. The standard test requires nt=1 or nt=4. If sidebands meet specifications at nt=1, repeating the test at nt=4 is unnecessary.
This test and the 1H lineshape test (the previous test) must be passed simultaneously. 1. 2. 3.
4.
Write the results in the forms provided in Probe Test Results, page 21.
01-999121-00 K 0606
15
8.
Enter setgcal to convert this width (delta) and the gradient strength (gzlvl1) to a gradient strength per DAC value (gcal). The setgcal macro needs the physical extent of the rf sensitivity region (1.6 cm), as well as the gradient strength (gzlvl1) and the resulting plateau width (delta). When the system asks Set this value? respond with y.
delta
Hz
The value of gcal times the appropriate value below gives the maximum achievable gradient strength. Performa I: gcal 2048. Performa II: gcal 32767. 9.
Repeat the experiment with gzlvl1 set to one of the following. The delta should be equal to that previously measured. Performa I: gzlvl1=250 Performa II: gzlvl1=4000
PFG Recovery 1. 2. 3. 4. Enter grecovery. For Performa I, enter at=1 in='n'. Set tpwr and pw for the 90 pulse width. Enter the appropriate value of gzlvl1: Figure 6. Gradient Recovery Pattern Profile Performa II: gzlvl1=30/gcal to set gzlvl1 to 30 G/cm. 5. 6. 7. 8. 9. Enter ga to start the acquisition. Enter wft ds(arraydim). Enter ga to start the acquisition. Enter wft ds(arraydim). Phase the final spectrum, adjust its vertical scale, and enter dssh. The resulting profile should show a smooth and monotonic recovery pattern, similar to Figure 6. Performa I: gzlvl1=10/gcal to set gzlvl1 to 10 G/cm.
16
01-999121-00 K 0606
H Sensitivity
1. 2. 3. 4. 5. 6. Insert the ethylbenzene sample. Enter rtp('/vnmr/tests/H1sn') su. Tune the probe.
1
Quartet
Set the PW90 value that was determined from the H PW90 calibration.
3800 3600 3400 3200 3000 2800 2600 2400 2200 2000
Hz
Enter ga to acquire the spectrum. Phase the spectrum and display the quartet using the region from the right of the aromatic peak to the right of the quartet. Place the cursor to the left (downfield) of the quartet (see Figure 7). Enter dsnmax(200). The computer calculates the S/N ratio. Answer 'y' to plot the results. Write the results in the forms provided in Probe Test Results, page 21
Aromatic peak
7. 8. 9.
19F
1. 2. 3. 4.
5.
Setup the following array to determine the pw90 for a probe: a. b. Enter r1=xx/5, where xx is the specified pw90 specification for the probe. Enter array('pw',20,r1,r1) to create a pw array covering approximately 360 degrees of nutation.
6. 7.
19
When the last spectrum is obtained, enter dssh. The first null spectra is the 180 pulse and the 90 pulse width is half the 180 pulse. Write the results in the forms provided in Probe Test Results, page 21.
F Sensitivity
Enter rtp('/vnmr/tests/F19sn') su. Tune the probe with the appropriate sample inserted. Set the PW90 value that was determined from the 19F PW90 calibration. Enter ga to acquire the spectrum. Enter dsnmax(200).
1. 2. 3. 4. 5.
01-999121-00 K 0606
17
The computer calculates the S/N ratio. Answer 'y' to plot the results. 6.
31
Write the results in the forms provided in Probe Test Results, page 21.
1. 2. 3. 4. 5. 6. 7.
8.
Setup the following array to determine the pw90 for a probe: a. b. Enter r1=xx/5, where xx is the specified pw90 specification for the probe. Enter array('pw',20,r1,r1) to create a pw array covering approximately 360 degrees of nutation.
9.
When the last spectrum is obtained, enter dssh . The first null spectra is the 180 pulse; therefore, the 90 pulse width is one-half the 180 pulse.
10. Write the results in the forms provided in Probe Test Results, page 21.
31P
Sensitivity
Insert the triphenylphosphate sample. Enter rtp('/vnmr/tests/P31sn') su. Tune the probe with the appropriate sample inserted. Set the PW90 value that was determined from the 31P PW90 calibration. Enter ga. Enter dsnmax(2000). The computer calculates the S/N ratio. Answer 'y' to plot the results. Write the results in the forms provided in Probe Test Results, page 21.
1. 2. 3. 4. 5. 6. 7.
13C
1. 2. 3.
18
01-999121-00 K 0606
4. 5. 6. 7. 8. 9.
Tune the probe. Enter a dpwr 3 or 4 dB less than dpwr of H2, dm='nny' d1=10 su. Enter ga. Place the cursor in the center of the dioxane triplet and type movetof. Enter pw=5 at=2 lb=3.5 fn=4096. Obtain the spectrum and adjust the phase for a positive signal. Array dof to set the decoupler frequency on the dioxane 1H resonance. Optimize the position of the decoupler (dof) to within 0.4 Hz of the correct frequency to obtain the best possible resolution.
10. Expand the spectrum so that the signal is centered on the screen with wp=100. 11. Enter ai. 12. Change vs so that the signal occupies about half the vertical size of the screen and set vp=50. 13. Determine the correct tpwr as follows: a. b. c. d. Enter tpwr=44 pw=7 ga. Enter aph to phase the spectrum after the acquisition completes. Set the pw to twice the pw90 specification for the probe. Array tpwr from 48 to 58. Observe the spectra as they are being collected. When the peak of interest passes through the null and goes negative, the acquisition can be halted by entering aa. The tpwr value corresponding to the first negative spectrum should then be used for pw90 calibrations. 14. Setup the following array to determine the pw90 for a probe: a. b. Enter r1=xx/5, where xx is the specified pw90 specification for the probe. Enter array('pw',20,r1,r1) to create a pw array covering approximately 360 degrees of nutation.
15. Enter d1=30 ga. 16. Enter dssh after the last spectrum is obtained. Change vs if necessary. 17. Determine pw for the first minimum. This is the 180 pulse. The 90 pulse is one-half the 180 pulse. 18. Write the results in the forms provided in Probe Test Results, page 21.
13C
This test measures the 50%, 0.55%, and 0.11% levels of the ASTM sample and is not required for ATB probes. This test and the 13C spinning sidebands test (the next test) must be passed simultaneously. 1. 2. 3. 4. 5. Insert the appropriate ASTM sample in the probe and spin it. Enter rtp('/vnmr/tests/C13res') su. Tune the probe. Enter nt=1. Change the decoupler power dpwr to 3 or 4 dB below the dpwr of H2. This will be the dpwr for 1H in the probe file. Array dof to set the decoupler frequency on the dioxane 1H resonance.
01-999121-00 K 0606
19
To obtain the best possible resolution, it may be necessary to optimize the position of the decoupler (dof) to within 0.4 Hz of the correct frequency. 6. 7. 8. Enter ga to acquire the spectrum. Save the spectrum use this spectrum for 13C spinning sideband test. Measure the linewidth and plot the spectrum as follows: a. b. 9.
13C
Set wp=50. Enter res to determine the linewidth at 50% of the decoupled peak and plot the spectrum. Increase vs by 100 times, plot the expanded spectrum, and measure the linewidth at 0.55% and 0.11%.
Write the results in the forms provided in Probe Test Results, page 21.
Spinning Sidebands
This test (not required for ATB probes) and the 13C spinning resolution and lineshape test (the previous test) must be passed simultaneously. 1. 2. Use the spectrum obtained for the 13C lineshape and resolution and test and measure the 13C spinning sidebands. Measure spinning sideband amplitudes as a percentage of the peak as follows. a. Plot the spectrum again using a large enough value of wp to show all the spinning sidebands, including fourth order. Spinning sidebands occur at integer multiples of the spinning rate and appear symmetrically about of the central peak. b. 3.
13C
The standard test uses nt=4. If sidebands meet specifications at nt=4, repeating the test at nt=16 is unnecessary.
Write the results in the forms provided in Probe Test Results, page 21.
Sensitivity
The specification for the 13C lineshape and resolution test must be met before performing this test. 1. 2. 3. 4. 5. Enter rtp('/vnmr/tests/C13sn') su. Tune the probe with the appropriate sample inserted. Set the 90 pulse width to the value determined in the calibration.
13C
Benzene-d6
Enter ga. Wait at least 8 minutes between acquisitions to ensure that the C6D6 lines are fully relaxed. The signal of interest is the highest line of the C6D6 triplet (see Figure 8).
p-Dioxane
9000
8000
7000
Hz
Figure 8.
13C
Sensitivity
6. 7. 8. 9.
Use the cursors to select the best 1400 Hz noise region between the dioxane and benzene peaks. Enter dsnmax(1400). Plot the results and record the value on the spectrum. Write the results the forms provided in Probe Test Results, page 21.
20
01-999121-00 K 0606
01-999121-00 K 0606
21
Probe Test Results for 4-nuc and ASW Probe Serial No. Date Installer Signature 90 Pulse Width
Nucleus
1
Contracted PW90
Notes, tpwr/dpwr
H F P C
19 31 13
H2
1H Spinning Sidebands
Nucleus
1H 13C
Contracted %
Results %
Notes
Results
1H
Nonspinning (CHCl3)
Results
Sensitivity
Nucleus
1H 19F 31P 13C
Contracted S:N
Results
Notes
22
01-999121-00 K 0606
Probe Test Results for ATB Probe Serial No. Date Installer Signature 90 Pulse Width
Nucleus
1
Contracted PW90
Notes, tpwr/dpwr
H F P C
19 31 13
1H Spinning Sidebands
Nucleus
1H 13C
Contracted %
Results %
Notes
Spinning (CHCl3)
Results
1H
Nonspinning (CHCl3)
Results
(S:N)
Results
Notes
01-999121-00 K 0606
23
AutoSwitchable, 4 Nucleus, and ATB NMR Probes Pub. No. 01-999121-00, Rev. K 0606 Technical contributor: Knut Mehr, Rob Rice, John Sandoval, Carlos Gil, Bao Nguyen, Susan Klein Technical writer and editor: Everett Schreiber Revision history: A0599 Initial release, ER 2750 B1099 ECO 8703; updated S/N specs, added tuning chart section C1099 ECO 8719; changed MERCURY S/N 1H and 19F specs (10/29/99). D0501 Updated probe part numbers per ECO 9108. E0501 300- and 400-MHz ATB probes; Updated S/N specs for MERCURYplus ECO 9140 F0901 Added specs to 300 ATB probe for MERCURY 300 4-nuc per ECO 9207 G0902 Updated specifications per ECO 9466 H1002 Added Magnet Interface Box updates I0104 Updated specifications J0505 General updates K0606 Added probes and updates per ECO 13019, updated parameter files and macros for VnmrJ2.1B Copyright 2006, Varian, Inc. 3120 Hansen Way, Palo Alto, California 94304 1-800-356-4437 http://www.varianinc.com All rights reserved. Printed in the United States
The information in this document has been carefully checked and is believed to be entirely reliable. However, no responsibility is assumed for inaccuracies. Statements in this document are not intended to create any warranty, expressed or implied. Specifications and performance characteristics of software and hardware described in this manual may be changed at any time without notice. Varian reserves the right to make changes in any products herein to improve reliability, function, or design. Varian does not assume any liability arising out of the application or use of any product or circuit described herein; neither does it convey any license under its patent rights nor the rights of others. Inclusion in this document does not imply that any particular feature is standard on the instrument. AutoSwitchable, ATB, VnmrS, UNITYINOVA, and MERCURYplus registered trademarks or trademarks of Varian, Inc. Other product names are trademarks of their respective holders.
24
01-999121-00 K 0606