If PalmSens is completely blocked, the next procedure is used to load the instrument with the firmware and to restart the instrument. Necessary tools: Size 14 spanner Phillips or TORX T-10 screwdriver Flat screwdriver Pointy tool
Step 1
Untighten the bolt very carefully with the spanner, without scratching the surface of the housing. Screw it off using your fingers.
Step 2
If present remove the two protective caps using the pointy tool by sticking it between the cap and housing and levering it out.
Step 3
Remove the sensor side of the housing.
The window now tells you to press Enter + Esc + < + > . Since PalmSens does not respond to the keys, please omit this message and continue with step 5.
Step 5
IMPORTANT: Shortcircuit the two pins next to the blue tiny resetbutton and press this button while keeping the two pins shortcircuited.
Note: in case it is hard to press the blue reset button behind the two short-circuited pins, it is also possible to remove the ac-adapter as well as batteryconnector for a few seconds and replace the batteryconnector, while keeping the two pins short-circuited. This requires that the ac-adapter is NOT connected and the battery has a reasonable charge left.
. In case the procedure is not successful, please repeat the procedure again from step 4.