General Specifications
GS 6B8B2-02E

Model YT200 Temperature Transmitter

The YT200 is a two-wire temperature transmitter that accepts an input from a thermocouple, resistance thermometer detector or DC voltage (mV) and converts it to a 4 to 20 mA uniform signal for transmission. The intelligent YT200 supports a broad choice of input signals. When combined with the BT200 BRAIN TERMINAL, it permits programming input signals, spans, and parameters through two-way communication.

s FEATURES
Sensor-and range-selectable The sensor type and measuring ranges are programmable to suit specific applications. Communication function The YT200 permits remote setting and monitoring the sensor type, ranges, etc. from the BT200 BRAIN TERMINAL. Easy wiring The YT200 saves wiring cost because it requires an extension cable for the thermocouple and a 3-core cable for the resistance thermometer detector between the detector and the converter. I/O isolation It accepts the YT200 to work with the grounded thermocouple.

s STANDARD SPECIFICATIONS
q I/O Specifications
Thermocouple Input Type Minimum and maximum spans Measuring range Zero Elevation JIS, IEC Type R, S, B, K, E, J, T, N Refer to Table 1 Refer to Table 1 300 % of span or 25 V DC whichever smaller 10 % of span 10 % of span 1 M (3 k at power off) or higher 1 M or lower 10 per a wire or lower -0.5 to 4.0 V DC at the input terminal R50 (E-13) [] See Fig.1 4 to 20 mA DC (output signal range: 3.6 to 21.6 mA) Upscale, Downscale or OFF
T01E.EPS

Resistance Thermometer Detector JIS Pt100, JPt100 IEC Pt100

DC Voltage (mV) DC (mV)

300 % of span

300 % of span or 25 mV DC whichever smaller

Zero point adjustment range Span adjustment range Input resistance Input signal source resistance Input leadwire resistance Input tolerance Permissible load resistance Output Burnout

1 M or lower

(Continued)

Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180 Japan Tel.: 81-422-52-5690 Fax : 81-422-52-2018

GS 6B8B2-02E Copyright Apr. 1995(YK) 4th Edition Mar. 1998(YK)

<<Contents>> <<Index>>

q Basic Performance Specifications


Thermocouple Accuracy Repeatability Reference junction precision Voltage fluctuation effect Ambient temperature effect Ambient temperature limit Ambient humidity limit RFI effect Response time Burnout Refer to Table 1 0.05 % of span 0.5 C (0.9 F) 0.005 % of span/V 0.05 % of span/10 C (18 F) 40 to 85 C (40 to 185 F) 5 to 100 % R.H. (40 C) Tested per SAMA PMC 33.1C (IEC 801-3), frequency from 20 to 1000 MHz and field intensity up to 10 V/m 1 second or shorter (0 to 63 % response of the full scale) Upscale, downscale or off Burnout current : 0.1 A Burnout time : 60 seconds or shorter 100 M (500 V DC) or higher Between input and output, and between input and ground: 1000 V AC, rms Between output and ground: 500 V AC, rms 12 to 42 V DC (17.9 to 42 V DC for communication) Refer to Table 2
T02E.EPS

Resistance Thermometer Detector

DC Voltage (mV)

N.A.

Insulation resistance Withstand voltage Supply voltage EMC Conformity Standards

Table 1: Range and span limit, and accuracy

Type (TC) J K T R S B E N
*1

Input Ranges C (F) 200 (328) to 1200 (2192) 200 (328) to 1350 (2462) 200 (328) to 400 (752) 0 (32) to 1760 (3200) 0 (32) to 1740 (3164) 100 (212) to 1820 (3308) 200 (328) to 1000 (1832) 200 (328) to 1300 (2372) C (F) 200 (328) to 650 (1202) 200 (328) to 500 (932) 20 mV to 100 mV

Min / Max Span

Accuracy A/D C 0.25 0.35 0.25 0.60 0.50 0.77 *3 0.20 0.40 C 0.17 0.16 12 V D/A % of span 0.03 0.03 0.03 0.03 0.03 0.03 0.03 0.03 % of span 0.03 0.03 0.03 %
T03E.EPS

2 mV / 62 mV

(RTD) *2 Pt100 JPt100 (DC mV)


*1: *2: *3:

C (F) 15 (36) / 650 (1170) 15 (36) / 500 (900) 2 mV / 100 mV

JIS C 1602/IEC 584 Thermocouple Specification Pt100: IEC 751/JIS C 1604, R100/R0=1.3850 JPt100: JIS C 1604, R100/R0=1.3916 Accuracy guarantee of Type B thermocouple: 400 C or higher.

All Rights Reserved. Copyright 1995, Yokogawa Electric Corporation

GS 6B8B2-02E

4th Edition Mar.10,1998-00

<<Contents>> <<Index>>
Table 2. EMC Confirmity Stnadards
Frequency Range 30 - 1000 MHz Basic Standard
EN55011 Class A Group 1
T04E.EPS

a) For EMI (Emission)-EN55011: 1991


Test Item

Applicable Electromagnetic Radiation Disturbance

b) For EMS (Immunity)-EN50082-2: 1995


No. 1 2 Test Item Electrostatic discharge Radio - frequency electromagnetic field Amplitude modulated Radio - frequency electromagnetic field Pulse modulated Fast transients common mode Radio - frequency common mode Amplitude modulated Test Specification 4 kV (contact) 8 kV (air) 80 MHz - 1 GHz 10 V/m (unmodulated) 80 % AM 900 MHz 10 V/m (unmodulated) Duty 50 %, 200 Hz REP. 2 kV, 5/50 (Tr/Th) ns 5 kHz REP. 150 kHz - 80 MHz 10 V (unmodulated) 80 % AM (1 kHz) Source Impedance 150 Basic Standard IEC 1000 - 4 - 2 : 1995 Level 3 Performance Criteria B A IEC 1000 - 4 - 3 : 1995 Level 3 A IEC 1000 - 4 - 4 : 1995 Level 3 Draft IEC 1000 - 4 - 6 : 1995 Level 3 B

3 4

A
T05E.EPS

Note: Definition of performance criterion A This apparatus continues to operate with its measurement accuracy within 1 % of full scale on 20 mV on 200 span during the text. Note: Definition of performance criterion B This apparatus continues to operate without hang-up or falling into uncontrollable conditions during the test. No change of actual operating state or stored or date is allowed. Note: Wiring shall be made by shielded wire or metal conduits.

s STRUCTURE
Integral Indicator: LCD digital indicator 4-digit with unit C, F, K, R, % or mV Enclosure Material: Aluminum alloy casting Coating: Polyurethane resin backed finish Color case: Frosty white cover: Deep sea moss green Classification: JIS C 0920 Watertight NEMA 4X Mounting: Mounted on 2-inch pipe Weight: 2.4 kg(5.3 lb) Add 0.3 kg(0.7 lb) for integral indicator

s COMMUNICATION LINE CONDITIONS


Power supply voltage: 17.9 to 42 V DC (Refer to Figure 1) Load resistance: 250 to 600 V (including cable resistance) (Refer to Figure 1) Maximum line length: 2 km when CEV cable is used. Load capacitance: 0.22 F or less Load inductance: 3.3 mH or less Spacing to power line: 15 cm or more
Figure 1. Relationship between power supply voltage and load resistance

All Rights Reserved. Copyright 1995, Yokogawa Electric Corporation

GS 6B8B2-02E

4th Edition Mar.10,1998-00

<<Contents>> <<Index>>

s MODEL AND SUFFIX CODES


[Style : S2] Model YT200 Output Signal Suffix codes Description Temperature transmitter 4 to 20 mA DC BRAIN Communication JIS G1/2 female ANSI 1/2 NPT female DIN Pg13.5 female M20 femal Digital indicator None 2-inch pipe mounting(SECC) 2-inch pipe mounting(SUS304) None Lightning arrester Stainless steel tag plate Epoxy resin-baked coating Optional color FM explosionproof type CENELEC explosionproof type JIS explosionproof type CSA explosionproof type SAA explosionproof type FM intrinsically safe type CENELEC intrinsically safe type CSA intrinsically safe type SAA intrinsically safe type
T06E.EPS

............

-D . . . . . . . . . . Electrical 0........ Connection 2........ 3........ 4........ Integral Indicator D..... N..... Mounting Bracket A... B... N...
Optional Codes

/A /SCT /X1 /P /FMF /EF1 /JF1 /CF1 /SF1 /FS1 /ES1 /CS1 /SS1

All Rights Reserved. Copyright 1995, Yokogawa Electric Corporation

GS 6B8B2-02E

4th Edition Mar.10,1998-00

<<Contents>> <<Index>>

s OPTIONAL SPECIFICATIONS (For Explosion Protected Typed)


Item Description JIS Flameproof Approval, Ex ds IIB+H2Ty Amb. Temp. : -20 to 60 C, Process Temp. : -20 to 120 C FM Explosionproof Approval Explosionproof for Class I, Division 1, Groups B, C and D. Dust-ignitionproof for Class II, Division 1, Groups E, F and G. Hazardous (classified) locations, indoors and outdoors (NEMA 4) Division 2, 'SEALS NOT REQUIRED', Temperature class : T6 Amb. Temp. : -40 to 60 C (-40 to 140 F), Electrical connection : 1/2 NPT female CENELEC (KEMA) Flameproof Approval EExd IIC T6, Amb. Temp. : -40 to 80 C (-40 to 176 F) Electrical connection : 1/2 NPT, Pg13.5, M20 female CSA Explosionproof Approval Explosionproof for Class I, Groups C and D Dustignitionproof for Class II/III, Division I, Groups E, F, and G Class I, Division 2 'SEALS NOT REQUIRED', Temp. Class : T6 Encl Type 4 Amb. Temp. : -50 to 60 C Electrical connection : 1/2 NPT female SAA Explosionproof Approval AUS Ex 3337X. Ex d IIC T6, IP65 AMB. TEMP. -40 to 80 C connection : M20 female, Pg13.5 female FM Intrinsically safe Approval Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1, Groups E, F, & G and Class III, Division 1 Hazardous Locations. Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division 2, Groups E, F & G and Class III, Division 1 Hazardous Locations. Enclosure : ''NEMA 4X'' Temp. Class : T4, Amb. Temp. : -40 to 60 C (-40 to 140 F) Intrinsically Safe Apparatus Parameters Vmax=30 V, Imax=200 mA, Pmax=0.9 W, Ci=4 nF, Li=730 H Intrinsically safe type CENELEC (KEMA) Intrinsically safe Approval EEx ia IIC T4, Amb. Temp. : -40 to 60 C (-40 to 140 F) Vi=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 H CSA Intrinsically safe Approval Intrinsically safe for Class I, Groups A, B, C and Class II and III, Groups E, F and G. Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division 2 Groups F & G, and Class III Encl Type 4, Temp. Class : T4, Amb. Temp. : -40 to 60 C (-40 to 140 F) Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=2 nF, Li=730 H SAA Intrinsically safe Approval AUS Ex 3336X. Ex ia IIC T4, IP66 AMB. TEMP. -40 to 80 C Attached flameproof packing adapter *1 Electrical connection : G1/2 female Amplifier cover only : 8.5 to 11 mm 2 pcs. Code JF1

FMF

Explosionproof type

EF1

CF1

SF1

FS1

ES1

CS1

SS1

G2
T07E.EPS

*1:

If cable wiring is to be used to a JIS flameproof type transmitter, do not fail to add the YOKOGAWA-assured flameproof packing adapter.

All Rights Reserved. Copyright 1995, Yokogawa Electric Corporation

GS 6B8B2-02E

4th Edition Mar.10,1998-00

<<Contents>> <<Index>>

6
Unit : mm (approx. inch)

s EXTERNAL DIMENSIONS

All Rights Reserved. Copyright 1995, Yokogawa Electric Corporation

GS 6B8B2-02E

4th Edition Mar.10,1998-00