Professional Documents
Culture Documents
ITEM 10020153
Products mentioned in this manual are mentioned for identification purposes only. Product names appearing in this manual may or may not be registered trademarks or copyrights of their respective companies.
24 April, 2007
Zetec, Inc. 8226 Bracken Place SE, Ste. 100 Snoqualmie, WA 98065 USA Phone 425-974-2700 or 800-643-1771 (US only) Fax 425-974-2701 www.zetec.com
CONTENTS
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . .1-1
About This Operating Guide Warranty & Quality Information Electronic and Mechanical Equipment Probes, Cables and Spare Parts General Terms O.E.M. Equipment Customer-Paid Equipment Repairs Return Policy 1-1 1-1 1-1 1-2 1-2 1-2 1-2 1-2
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Configuration . . . . . . . . . . . . . . . . . . . . . . . .4-1
Naming the Configuration Selecting Configuration Options Config Type External Encoder Adjusting the Sample Rate Sample Rates in Multiplexed Mode Sample Rates in Simultaneous Mode Setting Test Frequency Values Selecting a Drive Setting Selecting a Channel Storing a Configuration 4-4 4-5 4-5 4-7 4-8 4-8 4-8 4-9 4-10 4-11 4-12
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Table of Contents
Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Selecting the Language 9-3
CONTENTS
Table of Contents
Setting the Speaker On/Off Setting the Time/Date Setting the IP Address Selecting the Graphics Dump Destination Setting the Remote Inputs
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Table of Contents
MRPC Rate Calculations RFT RFT Hardware Setup Configuration Setup Single Frequency HVAC Hardware Setup HVAC System Setup Steps Test Plan Setup
List of Figures
CONTENTS
Figure 2-1 Figure 2-2 Figure 3-1 Figure 3-2 Figure 3-3 Figure 3-4 Figure 3-5 Figure 3-6 Figure 3-7 Figure 3-8 Figure 3-9 Figure 3-10 Figure 3-11 Figure 3-12 Figure 3-13 Figure 3-14 Figure 3-15 Figure 3-16 Figure 3-17 Figure 3-18 Figure 3-19 Figure 3-20 Figure 3-21 Figure 3-22 Figure 4-1 Figure 4-2 Figure 4-3 Figure 4-4 Figure 4-5 Figure 4-6 Figure 4-7 Figure 4-8 Figure 4-9 Figure 4-10 Figure 4-11 Figure 4-12 Figure 5-1 Figure 5-2 Figure 5-3 Figure 5-4 Figure 5-5 Figure 5-6 Figure 5-7 Figure 5-8 Figure 5-9 Figure 5-10 Figure 5-11 Figure 6-1 Figure 6-2 Figure 6-3 Figure 6-4 Figure 6-5
MIZ-28 Front and Back Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1 WEEE Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2 On/Off Switch & Coil Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2 Bobbin Coil Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3 Front Panel Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4 Navigation Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5 MIZ-28 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5 Primary & Secondary Function Buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6 MIZ-28 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7 Plugging in the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9 Keyboard Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9 Select Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11 Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12 Bobbin Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13 Bobbin Probe Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14 Bobbin Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14 MRPC Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15 MRPC Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15 MRPC Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16 RFT Hardware Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17 RFT Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17 HVAC Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19 AC3 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19 HVAC Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20 Open Configuration Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1 Configuration Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2 Configuration Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3 Config Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4 Config Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5 Config Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6 External Encoder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7 Sample Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8 Frequency Select. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9 Drive Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10 Channel Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11 Config Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12 Open Config List Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1 Config List Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2 Config List Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2 List Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3 Recall Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4 Copy to Mass Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5 Change Directory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6 Copy to Internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6 Remove Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7 Change Directory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8 Config Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9 Open Test/Review Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1 Test/Review Menu 1 StructureTest Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2 Test/Review Menu 2 StructureTest Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3 Testing Mode Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4 Encoder Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
List of Figures
CONTENTS
Figure 6-6 Figure 6-7 Figure 6-8 Figure 6-9 Figure 6-10 Figure 6-11 Figure 6-12 Figure 6-13 Figure 6-14 Figure 6-15 Figure 6-16 Figure 6-17 Figure 6-18 Figure 6-19 Figure 7-1 Figure 7-2 Figure 7-3 Figure 7-4 Figure 7-5 Figure 7-6 Figure 7-7 Figure 7-8 Figure 7-9 Figure 8-1 Figure 8-2 Figure 8-3 Figure 8-4 Figure 8-5 Figure 8-6 Figure 9-1 Figure 9-2 Figure 9-3 Figure 9-4 Figure 9-5 Figure 9-6 Figure 9-7 Figure 9-8 Figure 10-1 Figure 10-2 Figure 11-1 Figure 11-2 Figure 11-3 Figure 11-4 Figure 11-5 Figure 12-1 Figure 12-2 Figure 12-3 Figure 12-4 Figure 12-5 Figure 12-6 Figure 12-7
Manual Tube Info Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6 Message Summary Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7 User Message Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8 Summary Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9 Tube End Detection Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-10 Adjust Tolerance Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-12 Test/Review Menu 1 StructureReview Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-14 Test/Review Menu 2 StructureReview Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15 Data Cursor Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-16 Adjust Data Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-17 Measure Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-18 Speed Check Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-19 Tube End Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-20 Filter Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21 Open Analysis Setup Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1 Analysis Setup Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2 Curve Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3 Select Measurement Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4 Adjust Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5 Curve Table Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6 Data Mixing Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7 Perform Mix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8 Setup File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9 Open Toolbox Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1 Toolbox Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2 Toolbox Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2 Frequency Plot Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3 Drive Plot Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5 Material List Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-6 Open Utilities Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1 Utilities Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-2 Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-3 Speaker On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4 Time/Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5 IP Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6 Graphics Dump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7 Remote Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8 Open Diagnostics Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1 Diagnostics Menu Structure - Two Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-2 Open Test Plans Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-1 Test Plans Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-2 Current Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-3 Select Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-4 Import Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-5 Open Mass Storage Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-1 Mass Storage Setup Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-2 Recording Directory Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-3 File Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-4 Record To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-5 Spool To . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-6 Network IP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-7
List of Figures
CONTENTS
Figure 12-8 Figure 12-9 Figure 12-10 Figure 13-1 Figure 13-2 Figure 13-3 Figure 13-4 Figure 14-1 Figure 14-2 Figure 14-3 Figure 14-4 Figure 14-5 Figure 14-6 Figure 14-7 Figure 14-8 Figure 14-9 Figure 14-10 Figure 14-11 Figure 14-12 Figure 14-13 Figure 14-14 Figure 14-15 Figure 14-16 Figure 14-17 Figure 14-18 Figure 14-19 Figure 14-20 Figure 14-21 Figure 14-22 Figure 14-23 Figure 14-24 Figure 14-25 Figure 14-26 Figure 14-27 Figure 14-28 Figure 14-29 Figure 14-30 Figure 14-31 Figure 14-32 Figure 14-33
Mount Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-8 Mount NetDRV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-9 Time Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-10 Mass Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-1 Mass Storage Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-2 Data Storage Directory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-2 List Directories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-3 Bobbin Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-1 Select Mass Storage Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-2 Mass Storage Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-3 Bobbin Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-4 Pull Calibration Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-6 Measure Max Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-7 Enter & Save Summary Form Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-8 MRPC Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-9 Select Mass Storage Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-10 Mass Storage Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-11 Select Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-11 Configuration Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-12 Test/Review Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-14 Enter & Save Summary Form Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-15 RFT Hardware Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-17 Single Frequency Configuration Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-19 HVAC Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-20 Select Mass Storage Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-21 Mass Storage Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-22 Select Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-22 Configuration Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-23 Test/Review Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-24 Data Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-25 Enter & Save Summary Form Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-26 Select Test Plans. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-27 Import Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-28 Change Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-28 Change Dest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-29 Import Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-30 Select Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-30 Current Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-31 Run Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-32 Run Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-32
CONTENTS
1.0 Introduction
The MIZ-28 is used by NDT inspectors to inspect tubing found in a wide range of heat exchanger designs. This testing can be performed using either the eddy current testing (ECT) technique for non-ferromagnetic tubing materials or the remote field testing (RFT) technique for inspection of ferromagnetic tubing.
1: GENERAL DESCRIPTION
Information contained in this document is subject to change without notice. Neither Zetec, Inc. nor any person acting on behalf of Zetec, Inc. makes any warranty, expressed or implied, with respect to the use of any information, apparatus, method or process disclosed in this document. Zetec, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. This document contains proprietary information which is protected by copyright. All rights are reserved. No part of this document may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language or computer language, in any form or by any means, electronic, mechanical, magnetic, optical, chemical, manual or otherwise, without the express prior written permission of:
1: GENERAL DESCRIPTION
www.zetec.com customerservice@zetec.com Phone 425-974-2700 or 800-643-1771 (US only) Copyright Zetec, Inc. 2005
1: GENERAL DESCRIPTION
1.2.1
All Zetec manufactured electronic and mechanical products are warranted, to the original purchaser only, against defects in material and workmanship for a period one year from the date of shipment. Zetec does not warrant against defects caused by misuse, abnormal operating conditions, alterations or damage such as fire, flood, wind and lightning. If installation/commissioning is performed by Zetec personnel as a condition of the purchase order, the warranty for mechanical systems will go into effect on the date of installation not to exceed 30 days beyond the original ship date.
1: GENERAL DESCRIPTION
Page 1-1
1.2.2
All Zetec manufactured probes, cables, and spare parts are considered consumables and require periodic replacement due to normal wear from use. Consumables have Zetecs quality guarantee that promises to deliver goods that are manufactured to specification, functional, and free from defects, as per the agreed conditions of sale.
1.2.3
General Terms
Equipment warranty does not include periodic calibration or minor maintenance as outlined in Zetec operating manuals. All returned merchandise must be sent freight prepaid by buyer, properly boxed to prevent damage in transit. The foregoing warranties of all Zetec products are in lieu of all other warranties, expressed or implied, including but not limited to the implied warranties or merchantability and fitness for a particular purpose. Limitation of Liability: Sellers total liability for any and all losses and damages arising out of any and all causes whatsoever including, without limitation, defects in the goods, services, software, documentation, or sitework supplied under this Agreement, shall in no event exceed the purchase price of the applicable item(s).
1.2.4
O.E.M. Equipment
For products not manufactured by Zetec, the manufacturers warranty will apply.
1.2.5
All customer paid repairs are warranted for 90 days (parts and labor). This warranty is limited to failures in areas related to the previous repair. After a warranty repair, the current calibration is validated; however, a new calibration is not provided free of charge.
1.2.6
Return Policy
Before returning items (uncontaminated only), please call Zetec to obtain a Return Merchandise Authorization (RMA) number. This RMA number will help us adjust your account as quickly as possible. To obtain an RMA number, you must provide the following information: reason for return, date of purchase, your purchase order number (on the packing list), and Zetecs reference number (on the packing list). If you suspect damage by the carrier, request that the carriers agent be present when the shipment is unpacked. If concealed damage is found, file a claim with the carrier immediately.
Page 1-2
Product Description
CONTENTS
1: GENERAL DESCRIPTION
The MIZ-28 instrument is used for eddy current inspections of heat exchanger tubing in balance-of-plant applications. It works seamlessly with all existing Zetec software and hardware products. Zetecs MIZ-28 offers a color display, selectable data presentation, menu-driven user interface, high digital sampling rates, bandpass and differential filters, and two independent balance controls. This versatile, fully digital tester provides maximum stability, reliability, and sensitivity.Testing capabilities include heat exchanger and air conditioner tube testing, weld inspection, and a broad variety of other eddy current applications. Packaged in a rugged portable carrying case, the MIZ-28 is ideal for industrial environments. An active matrix color display shows test information from almost every angle. The use of color provides quick channel identification without having to read the setup screen. The real-time display constantly indicates the coil impedance point shift and closely emulates a storage oscilloscope display. The MIZ-28 presents the phase and amplitude vectors of the eddy current signal as a two-dimensional Lissajous figure. Simultaneous display of up to four Lissajous figures and two strip charts enables you to efficiently monitor multiple data channels. The MIZ-28 internal hard drive adds 40 GB of data storage capacity to the MIZ-28 for data acquisition and tube list storage.
CA
UT
Figure 2-1
1: GENERAL DESCRIPTION
Page 2-1
2.1 Cleaning
W AR
When cleaning the display, care must be taken to avoid damage. For best results, use a lint free cloth with minimal water or a mild detergent. Use only detergents or cleansers which are recommended for use on plastics.
2.2 Maintenance
The MIZ-28 is designed to be maintenance free. However, if it does require repair, do not attempt to repair it yourself. Contact Zetec for more information. Do not open the MIZ-28 box. Any maintenance to be performed on internal components must be carried out by trained maintenance personnel at Zetec only.
Figure 2-2
CA
WEEE Label
The next section of this operating guide explains general operation of the MIZ-28.
Page 2-2
UT
IO N
NI NG
CONTENTS
1: GENERAL DESCRIPTION
If anything is missing, please contact Zetecs Customer Service department at 425-974-2700 immediately.
Page 3-1
Figure 3-1
Page 3-2
CONTENTS
Figure 3-2
3.
Proceed to Section 3.8 for quick tutorials or Section 14.0 for detailed tutorials.
Page 3-3
Figure 3-3
4 9 6 7 5
the Select Button to choose it. 5 Help Button. Press the Help button to display information about the current page selected. Press the Help button again to exit the Help display. 6 Print Button. Press the Print button to send the current screen image to the local USB printer or to a mass storage device in .bmp file format. Print button operation is setup on the Utilities page. 7 Menu Display/Hide Button. Any time during operation, you can press the Menu Display button to display the Main Menu. Press the Menu Display button again to hide the Main Menu. The Menu organizes the MIZ-28s functions into related tasks for instrument setup and testing. 8 Function Keys. Twelve function keys are available for oper-
ating the MIZ-28 features. Function key actions and labels change according to the selected mode. The horizontal (primary) function keys select a feature for adjustment. The vertical (secondary) function keys select a parameter value to change for the selected feature. 9 Menu Button Labels. The display labels identify the function of each menu button and change according to the selected mode.
Page 3-4
CONTENTS
1: GENERAL DESCRIPTION
Figure 3-4
Navigation Controls
1: GENERAL DESCRIPTION
1. 2.
Click on the Menu Display button. The Main Menu (figure 3-5) opens. Use the Control Knob to highlight the menu item you want to open.
Figure 3-5
3.
Page 3-5
4.
The button labels along the bottom of the screen (activated via primary function buttons F1 through F7) change depending on the original Main Menu item you selected. The button labels along the right hand side of the screen (activated via secondary function buttons F8 through F12) change depending upon which primary function button, F1 through F7, has been selected.
Figure 3-6
Page 3-6
CONTENTS
1: GENERAL DESCRIPTION
Figure 3-7
1: GENERAL DESCRIPTION
Configuration Screen. A configuration is a set of MIZ-28 test parameters that are used for NDT inspection. These parameters include sample rate, test frequencies, drive and channel. Use the Configuration screen to create and maintain these configurations. See Section 4.0 for details. Config List Screen. The Configuration List screen provides the ability to manage test configurations. It also lists all saved configurations and allows you to select a configuration to edit. Configurations can be stored internally or on an external storage device. The number of stored configurations is limited only by the size of the storage medium. See Section 5.0 for details. Test/Review Screen. Use the Test/Review screen to acquire and review data. See Section 6.0 for details. Analysis Setup Screen. Use the Analysis Setup screen to set up curve parameters and mix data channels, and to adjust the volt scale. See Section 7.0 for details. Toolbox Screen. The Toolbox screen allows you to plot graphs of amplitude versus frequency (freq plot) and amplitude versus drive (drive plot), as well as to view a list of materials. There are also prime frequencies for a given wall thickness and standard depth of penetration at a given frequency. See Section 8.0 for details. Utilities Screen. The Utilities screen provides access to Language Selection (English, German, Italian, Spanish, and French), Speaker Control, Time/Date Input, IP Address Setting, Graphics Dump, and Remote Inputs. The MIZ-28 has built-in USB and TCP/IP support. See Section 9.0 for details. Diagnostics Screen. The Diagnostics screen contains diagnostic tools that can be activated to assure that the instrument is within calibration specifications, as well as to troubleshoot inputs and outputs from other devices. The fea-
Page 3-7
tures in the Diagnostics menu are intended for use by Zetec calibration and repair or other personnel who are trained in MIZ-28 maintenance, and are not intended for typical field use applications.See Section 10.0 for details. Test Plans Screen. Use the Test Plans screen to import and utilize test plans. See Section 11.0 for details. Mass Storage Setup Screen. The Mass Storage Setup screen allows you to specify how to record the data. Use it to set the recording directory, file format, recording destination, network IP address, mount point, and time zone. Data can only be recorded when the Acquire function is active. See Section 12.0 for details. Mass Storage Screen. The Mass Storage screen enables you to open, delete and move data files between internal and mass storage directories. See Section 13.0 for details.
Page 3-8
CONTENTS
Figure 3-8
1: GENERAL DESCRIPTION
USB Ports
1: GENERAL DESCRIPTION
When a keyboard is plugged in, an icon displays in the upper right corner of the screen.
Figure 3-9
Keyboard Icon
Keyboard icon
When the main MIZ-28 menu is displayed, use the Enter key to select the current highlighted selection. When using the data cursor, the Up and Down arrow keys scroll the data, and the left and right arrow keys zoom the charts in and out, respectively.
Page 3-9
The Home key on the keyboard toggles between coarse and fine adjustment. When adjusting the configuration, press the Enter key to perform the selected function (i.e., highlight a drive, press the Enter key to make an adjustment and press Enter again to select the next one). In Test/Review, when the SELECT QUADRANT/CHANNEL (F8) button is selected, use the left and right arrow keys to change the quadrant, and use the up and down arrow keys to change the channel. When vertical or horizontal position is selected, the up and down keys will adjust the vertical position, and the right and left arrow keys will adjust the horizontal position. In testing mode (acquiring data) with a Model III Controller connected, the following keys can be used. Up arrow key runs the pusher forward at the fwd run speed. Down arrow key runs the pusher reverse at the reverse speed. Right arrow key runs the pusher forward at fwd jog speed. Left arrow key runs the pusher reverse at the rev jog speed. Enter key starts recording and starts the pusher running reverse. Enter key again stops recording and stops the pusher.
Tube ID fields in testing mode (acquiring data). + key increments the current tube encode field. key decrements the current tube encode field. * key increments the field to adjust. / key decrements the field to adjust.
When using a keyboard, the following CNTRL functions will take you directly to a screen. Ctrl C Configuration screen Ctrl L Configuration list screen Ctrl T Test / review screen Ctrl A Analysis Setup Ctrl B Toolbox screen Ctrl U Utilities screen Ctrl D Diagnostics screen Ctrl P Test plans screen Ctrl S Mass storage setup Ctrl R Mass storage dir listing
Page 3-10
CONTENTS
1: GENERAL DESCRIPTION
3.
1: GENERAL DESCRIPTION
Page 3-11
4.
Figure 3-11
Select Language
5.
Press the corresponding language (F8 to F12) button to select one of the following languages: English (F8) Deutsch (F9) Italiano (F10) Espaol (F11) Franais (F12)
Page 3-12
CONTENTS
3.8.1
1.
1: GENERAL DESCRIPTION
Connect bobbin hardware per the bobbin system drawing (see Drawing Number 10020312 in Appendix A for a larger illustration that includes the parts list.)
2.
In some testers a mode of operation (differential, absolute, driver/pick-up, etc.) might be selected in the test configuration process. In the MIZ-28 this step is not required. The specific probe adaptor selected will define whether the displayed information is differential or absolute once the coil functionality is selected (On/Off). For the bobbin probe application, those channels activated under Coil 1 will be operating in the self-comparison differential mode. Channels activated under Coil 2 will be operating in the external-reference differential mode.
NO
TE
1: GENERAL DESCRIPTION
Page 3-13
3. 4. 5. 6.
Verify system sensitivity and signal rotation. Select data storage mode or device: USB, internal hard drive, or network. Store all system parameters. Record summary, calibration runs, and data.
Note: For a step by step bobbin application process refer to Section 14.1.
Page 3-14
CONTENTS
3.8.2
1.
MRPC Exam
Connect MRPC hardware per the MRPC system drawing (see Drawing Number 10020311 in Appendix A for a larger illustration that includes the parts list.)
2.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 3-15
3. 4. 5. 6.
Verify system sensitivity and signal rotation. Select data storage mode or device: USB, internal hard drive, or network. Store all system parameters. Record summary, calibration runs, and data.
Note: For a step by step MRPC application process refer to Section 14.2.
Page 3-16
CONTENTS
3.8.3
1.
Connect RFT hardware per the RFT system drawing (see Drawing Number 10020313 in Appendix A for a larger illustration that includes the parts list.)
2.
Page 3-17
3. 4. 5. 6. 7.
Verify system sensitivity and signal rotation. Select data storage mode or device: USB, internal hard drive, or network. Store all system parameters. Record summary and calibration runs for the settings used in the calibration standard. Based on responses from tubing in the heat exchanger it will be necessary to modify the set-up parameters (predominantly frequency) to achieve maximum signal/noise ratio and flaw sensitivity in the tubes to be inspected. This process may have to be performed more than once in a given heat exchanger.
Page 3-18
CONTENTS
3.8.4
1.
HVAC
Connect HVAC hardware per the HVAC system drawing (see Drawing Number 10020340 in Appendix A for a larger illustration that includes the parts list.)
2.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 3-19
3. 4. 5. 6.
Verify system sensitivity and signal rotation. Select data storage mode or device: USB, internal hard drive, or network. Store all system parameters. Record summary, calibration runs, and data.
Note: For a step by step AC3 application process refer to Section 14.4
The next section of this operating guide describes the Configuration screen.
Configuration
CONTENTS
4.0 Configuration
A configuration is a set of MIZ-28 test parameters, including sample rate, test frequencies, drive and channel, used for NDT inspection. Use the Configuration screen to create and maintain these configurations. In addition to any custom configurations you might create, the MIZ-28 also includes four canned configurations: Bobbin Coil, MRPC, RFT and HVAC. These are accessed via the Config List screen (Section 5.0).
1: GENERAL DESCRIPTION
NG
1. 2. 3.
To open the Configuration screen, follow these steps: Press the Menu Display button. Use the Control Knob to select (highlight) CONFIGURATION. Press the Select button.
W AR
NI
The standard configurations are not write-protected. Use care when working with them.
1: GENERAL DESCRIPTION
Figure 4-1
Page 4-1
Figure 4-2
Configuration Screen
The default Configuration screen has the following primary function buttons across the bottom of the screen:
Config Name (F1) Config Options (F2) Sample Rate (F3) Frequency Select (F4) Drive Select (F5) Channel Select (F6) Config Storage (F7)
When you press these buttons, the F8 F12 secondary function buttons change, as illustrated in Figure 4-3.
Page 4-2
Configuration
CONTENTS
Figure 4-3
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
COIL NUMBER (F8) FIELD (F9) TOGGLE ON/OFF (F10) SET ALL (F11)
1: GENERAL DESCRIPTION
The rest of this section describes each menu function in the same order as shown above.
1: GENERAL DESCRIPTION
Page 4-3
4. 5. 6. 7.
Figure 4-4
Config Name
Page 4-4
CONTENTS
Figure 4-5
Config Options
4.2.1
Config Type
The MIZ-28 supports both multiplexed and simultaneous injection testing. Multiplexed It is recommended you use the multiplexed mode for the following probes:
Bobbin (differential and absolute with external reference) MRPC Array probes (that do not require external multiplexing hardware) AC3 probes
1: GENERAL DESCRIPTION
The MIZ-28 supports four test frequencies in multiplexed mode. Simultaneous Injection In simultaneous injection mode, the number of available frequencies depends on the number of coils being used and the number of boards present in the unit. With the two analog board MIZ-28, you can run a two coil test at four frequencies, a four coil test at two frequencies, or an eight coil test at one frequency.
1: GENERAL DESCRIPTION
Page 4-5
Figure 4-6
Config Type
To select a config type: 1. 2. 3. 4. Press the CONFIG OPTIONS (F2) button. Use the Control Knob to scroll to CONFIG TYPE. Press the ADJUST (F9) button and use the Control Knob to select MULTIPLEXED or SIMULTANEOUS. Press the SELECT (F8) button.
Page 4-6
CONTENTS
4.2.2
External Encoder
Selecting External Encoder allows you to set the number of encoders. If you enable the encoder (up to two), you must also configure the encoder.
Figure 4-7
External Encoder
To configure an encoder: 1. 2. 3. 4. Use the Control Knob to scroll through the encoder configuration options. Press the ADJUST (F9) button to activate the highlighted option.
1: GENERAL DESCRIPTION
Use the Control Knob to scroll through the selected option configurations. Press the SELECT (F8) button to confirm the new option configuration.
1: GENERAL DESCRIPTION
Page 4-7
Figure 4-8
Sample Rate
4.3.1
Four different filters 4K, 8K, 12K and 16K are available for use in the signal processing circuitry when operating in multiplexed mode. The 4K filter produces the best signal-to-noise, though at the cost of sample rate. The 16K filter allows faster sampling rates at the cost of signal-to-noise. Switching from the 4K filter to the 8K filter, you can expect to lose 3dB of signal-to-noise. Switching from 8K to 12K or 12K to 16K filter, you can expect to lose 1.5dB of signal-to-noise. The instrument automatically selects the best possible filter for the configured parameters and displays it next to the maximum sample readout. As you increase or decrease the sample rate, the filter setting will change.
4.3.2
Four filters are available in simultaneous mode: 240Hz pass band, 400Hz pass band, 800Hz pass band and 1600Hz pass band. The trade-off when using these filters is signal-to-noise versus pull speed. Given the general rule of 10Hz of pass band per inch pulling speed, the 240Hz filter has about 24ips max pull speed and the 400Hz has about 40ips max pull speed. As you start decreasing test frequencies, the max pull speed value displayed to the right of the sample rate decreases. You should expect the 240Hz filter to have a 2.5dB gain in signal to noise.
Page 4-8
CONTENTS
1: GENERAL DESCRIPTION
A special filter design is implemented in the MIZ-28 to improve performance at low test frequencies. This is to improve RTF testing performance and signal noise. This filter will automatically be selected if the configuration allows for it. When the filter is applied, you will see an increase in the max pull speed (ips) value on the configuration screen and an LF (low frequency filter) will appear. This filter is only applicable in the SI mode. You can select up to four instrument test frequencies, depending on the test configuration. To select a test frequency: 1. 2. 3. Press the FREQUENCY SELECT(F4) button. Use the Control Knob to select the frequency you want to change. Press the ADJUST (F9) button and use the Control Knob to adjust the frequency value.
1: GENERAL DESCRIPTION
TE
Figure 4-9
For help on frequency selection, see Section 8.3, Viewing the Material List .
NO
Frequency Select
Page 4-9
Page 4-10
CONTENTS
Figure 4-11
Channel Select
Page 4-11
The next section of this operating guide describes the Config List screen.
Page 4-12
CONTENTS
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 4-13
Config List
CONTENTS
1: GENERAL DESCRIPTION
Figure 5-1
Page 5-1
Figure 5-2
The default Config List screen has the following primary function buttons across the bottom of the screen:
List Options (F1) Change Directory (F2) Config Name (F3)
When you press these buttons, the F8 F12 secondary function buttons change, as illustrated in Figure 5-3.
Figure 5-3
CHANGE DIRECTORY (F2) NOTE: Only available with a mass storage device connected CONFIG NAME (F3)
INTERNAL (F8) MASS STORAGE(F9) Right arrow(F8) Left arrow (F9) SPACE (F10) BACK SPACE (F11) CLEAR (F12) RETURN (F7)
The rest of this section describes each menu function in the same order as shown above.
Page 5-2
CONTENTS
1: GENERAL DESCRIPTION
Figure 5-4
List Options
5.1.1
Press the SCROLL (F8) button to scroll through the listed configurations.
5.1.2
Saving a Configuration
Press the SAVE CONFIG (F9) button to store the current configuration to the list. If a configuration of the same name already exists, you will be given the option to overwrite the existing configuration.
1: GENERAL DESCRIPTION
Page 5-3
5.1.3
1. 2.
Recalling a Configuration
To recall a configuration: Press the RECALL CONFIG (F10) button to enable configuration selection. Rotate the Control Knob to select (highlight) the desired Configuration.
Figure 5-5
Recall Configuration
3.
Press the Select button to recall the highlighted configuration. A message displays confirming that the configuration is recalled.
Page 5-4
NO
If the current Configuration has any unsaved changes in it, a message dialog will appear, asking if you want to continue with the recall operation without saving the current modified configuration.
TE
CONTENTS
5.1.4
Configurations can be copied back and forth between internal storage and mass storage. If a configuration of the same name exists in the location that you are copying the configuration to, you will be prompted as to whether or not you want to overwrite the existing configuration. Copying from Internal to Mass Storage If a USB mass storage device is attached to the instrument, the COPY TO MASS STORAGE (F11) button will be active. (If there is no USB mass storage device attached, this button label will be blank.) 1. 2. Press the COPY TO MASS STORAGE (F11) button. Highlight the configuration you want to copy.
1: GENERAL DESCRIPTION
Figure 5-6
3. 4.
Press the Select button. The message Configuration copied to mass storage will be displayed at the bottom of the screen.
1: GENERAL DESCRIPTION
Page 5-5
Copying from Mass to Internal Storage 1. Press the CHANGE DIRECTORY (F2) button.
Figure 5-7
Change Directory
2. 3. 4.
Press the MASS STORAGE (F9) button. Press the LIST OPTIONS (F1) button. All the Mass Storage configurations will be listed. Press the COPY TO INTERNAL (F11) button.
Figure 5-8
Copy to Internal
Page 5-6
CONTENTS
5. 6. 7.
Highlight the configuration you want to copy. Press the Select button. The message Configuration copied to internal storage will be displayed at the bottom of the screen.
5.1.5
1. 2.
Removing a Configuration
To remove a configuration: Press the REMOVE CONFIG (F12) button to enable configuration selection. Rotate the Control Knob to select (highlight in green) the Configuration in the list.
1: GENERAL DESCRIPTION
Figure 5-9
Remove Configuration
3.
IO N CA UT
The configuration is deleted immediately, so be certain you want to remove it before pressing the Select button.
1: GENERAL DESCRIPTION
Page 5-7
Page 5-8
NO
When the current directory is Mass Storage, the F11 key of the List Options menu changes to Copy to Internal.
TE
CONTENTS
1: GENERAL DESCRIPTION
4. 5. 6. 7.
Figure 5-11
Config Name
NO
TE
1: GENERAL DESCRIPTION
The next section of this operating guide describes the Test/Review screen.
Page 5-9
Test / Review
CONTENTS
1: GENERAL DESCRIPTION
The menu functions change based on whether you are in Testing Mode (Section 6.1) or Review Mode (Section 6.2).
Figure 6-1
Page 6-1
Figure 6-2
ENCODER (F3) If you have only one encoder, this button displays ZERO ENCODER, and no secondary function buttons are displayed. Depending on how many encoders you have, there could be less secondary functions. MANUAL TUBE INFO (F4)
SELECT ID FIELD (F8) ADJUST (F9) TOGGLE FIELD(F10) ENABLE STD RUN (F11) DISPLAY NULL (F12)
MESSAGE SUMMARY (F5) When you press this button, the function keys to the right appear and the overall menu changes. Use the keys to the right to work with the message window.
START RECORDING (F6) When you press this button, recording starts and the button label changes to Stop Recording. STOP ACQUIRE (F7) When you press this button, Acquisition stops and the menu changes to that shown in Figure 6-12
Page 6-2
CONTENTS
Figure 6-3
This button will only be present if a Model III Controller is connected to the RS-232 connector on the back panel of the MIZ-28. If a Model III is connected, a pusher icon will display in the top right corner of the display. ADJUST TOLLERANCE(F3)
1: GENERAL DESCRIPTION
SELECT QUADRANT (F8) TOGGLE (F9) VERTICAL ADJUST (F10) HORIZ ADJUST (F11) HORZ (F12)
ADJUST DATA(F4)
1: GENERAL DESCRIPTION
CHANNEL ROTATION (F10) VERTICAL ADJUST (F11) HORIZ ADJUST (F12) RUN TEST PLAN (F5) SCROLL (F8) GO TO TOP (F9) FIND NEXT UNTESTED (F10) MARK UNMARK (F11) START RECORDING (F6) When you press this button, recording starts and the button label changes to Stop Recording. STOP ACQUIRE (F7) When you press this button, Acquisition stops and the menu changes to that shown in Figure 6-12 .
Page 6-3
6.1.1
Data Control
Press the DATA CONTROL (F2) button to change the channel and quadrant, chart rate, and persist time. From this submenu, you can also manually clear and balance the data display.
Figure 6-4
Select Quadrant / Channel 1. Press the SELECT QUADRANT CHANNEL (F8) button. When the word QUADRANT is capitalized, you can use the Control Knob to scroll through the four X-Y displays and strip charts. 2. Press F8 again and CHANNEL is capitalized. Now the Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display.
Chart Rate Press the CHART RATE (F9) button, and then use the Control Knob to adjust the rate at which the strip charts scroll. The number that displays on the Chart Rate button indicates how many slices of data will be plotted to the chart per second. Clear Press the CLEAR (F10) button to clear the data off the display. Persist Time Press the PERSIST TIME (F11) button to adjust the amount of time that will elapse between automatic screen clears. Adjustment ranges from OFF to 10 seconds in 100 millisecond internals.
Page 6-4
CONTENTS
Balance Press the BALANCE (F12) button to electronically balance the eddy current signals. This function will take the current data values and shift them to a value of zero. Two balance functions are available if more than one balance is selected in the configuration.
6.1.2
Encoder
1: GENERAL DESCRIPTION
The ENCODER (F2) button will only be present if two encoders have been configured.
Figure 6-5
NO TE
If there is no encoder is configured, the ENCODER button will not display. If only one encoder is configured, the ENCODER button will display ZERO ENCODER.
Encoder Screen
Zeroing an Encoder 1. 2. Press the ENCODER (F3) button. Press the ZERO ENCODER (F8 or F9) for the encoder you would like to zero.
1: GENERAL DESCRIPTION
Page 6-5
6.1.3
If test plans are not being used, it is necessary to manually encode the data files. Press the MANUAL TUBE INFO (F3) button to access the tube info control buttons.
Figure 6-6
Selecting an ID Field Press the SELECT ID FIELD (F8) button to enable the Control Knob to scroll through the tube identification fields. The selected field is highlighted in blue. Adjusting an ID Field Press the ADJUST (F9) button to adjust the selected field. The selected field is highlighted in green when the adjustment is enabled. Pressing the Select button in this mode will toggle the selection between select and adjust. After making an adjustment, pressing the Adjust (F9) button again will change the row or column encode to increment or decrement.
Toggling an ID Field Press the TOGGLE FIELD (F10) button to toggle the ID field between alpha numeric and numeric. Enable/Disable Standard Run Press the ENABLE STD RUN (F11) button to change the row and column values to the standard run values (typically 999). The values are operator selectable using the SELECT and ADJUST buttons above. When standard run is
Page 6-6
NO
TE
CONTENTS
enabled, the button label changes to DISABLE STD RUN. Pressing the DISABLE STD RUN (F11) button changes the row and column values back to what they were previously. Display Null Press the DISPLAY NULL (F12) button to center the data position on the screen. This function does not change the data values.
1: GENERAL DESCRIPTION
6.1.4
Message Summary
Press the MESSAGE SUMMARY (F5) button to access the Message Summary screen. The first screen you see is a user defined list of 25 messages. At the bottom of this screen are two primary function buttons, USER MESSAGE and SUMMARY FORM (these buttons are described in Sections 6.1.5 and 6.1.6, respectively). The Message Summary secondary function buttons are described below.
Figure 6-7
Scroll Press the SCROLL F8) button to enable the Control Knob to scroll up and down through the list of messages. Write Message Press the WRITE MESSAGE (F9) button to write the selected message to the data directory. Change Message Press the CHANGE MESSAGE (F10) button to change the text in the selected message. This action displays the same character naming screen as that used for naming configurations (Section 4.1). Alternatively, use the keyboard to change the text.
1: GENERAL DESCRIPTION
Page 6-7
6.1.5
User Message
In the Message Summary screen (Section 6.1.4), press the USER MESSAGE (F4) button to enable a blank text screen. This screen is used for message input to the data directory for later use when analyzing data.
Figure 6-8
Scroll Message Press the SCROLL (F8) button to enable the Control Knob to scroll the cursor through the message. Write Message Press the WRITE MESSAGE (F9) button to write the message to the data directory. Clear Message Press the CLEAR MESSAGE (F10) button to clear the current message. Save Message Press the SAVE THIS MESSAGE (F11) button to save the current message form. Recall Message Press the RECALL MESSAGE (F12) button to recall the saved message form.
Page 6-8
CONTENTS
6.1.6
Summary Form
In the Message Summary screen (Section 6.1.4), press the SUMMARY FORM (F5) button to access the test summary form. This default form includes parameters such as owner, plant, unit, and a configuration summary of the current parameters. To fill out the summary form, you must attach a USB keyboard. Press the Tab button to move the cursor to the next valid entry location.
1: GENERAL DESCRIPTION
Press the Enter button to move the cursor to the beginning of the next line.
Figure 6-9
Summary Form
Scroll Press the SCROLL (F8) button to enable the Control Knob to scroll the cursor through the Summary Form. Write Summary Press the WRITE SUMMARY (F9) button to write the current form.
Default Summary To restore the summary form default parameters, press the DEFAULT SUMMARY (F10) button. Save Summary Form Press the SAVE THIS FORM (F11) button to save any changes made to the summary form.
NO TE
When the file format is set to Eddynet in the mass storage setup, writing a summary form will automatically create a new data directory (cal group) for recording.
1: GENERAL DESCRIPTION
Page 6-9
Recall Summary Form Press the RECALL FORM (F12) button to recall the summary form that was last saved. This function can be useful for clearing the specific operator data out of the form.
6.1.7
Press the START RECORDING (F6) button to start recording a data file to the current data directory. The current recording path and file name is shown under the tube encode. When the START RECORDING (F6) button is pressed, the F6 button label changes to STOP RECORDING and a list of the last 5 files recorded displays in the lower right corner of the screen.
6.1.8
Stop Acquisition
Press the STOP ACQUISITION (F8) button to stop the data acquisition and switch to data review mode (Section 6.2).
6.1.9
The TUBE END DETECTION (F2) button on the second page of buttons, will only be present if a Model III Controller is connected to the RS-232 connector on the back panel of the MIZ-28. When a Model III is connected, a pusher icon is displayed in the top right corner of the display. The tube end detection submenu allows you to enable or disable tube end detection, set a tube end threshold delay (which allows for entry into the tube without detecting the near side tube end), enable or disable the auto retract function and, when in review mode, set the tube end threshold.
Setting the Tube End Threshold 1. Acquire the data from a tube end and press the STOP ACQUIRE button.
Page 6-10
CONTENTS
2.
Scroll the tube end signal to the cursor and press the SET THRESHOLD button. This will take the peak measurement of the tube end and save it to NVRAM. A message at the bottom of the page will display saying, Threshold has been set.
Tube End Detection Press the TUBE END DETECTION (F2) button to enable or disable the function.
1: GENERAL DESCRIPTION
When tube end detection is enabled, every time the pusher starts running the data will be monitored for a signal that is greater than 80% of the set threshold limit (tube end signal). When a signal of great enough amplitude is detected, the instrument will turn off the pusher. The exception is when Standard Run is enabled. In this case, the instrument will not monitor for the threshold signal. Threshold Delay Press the THRESHOLD DELAY (F10) button, and then use the Control Knob to change the delay time in milliseconds (mS). The Threshold Delay is the amount of time that the instrument will wait to begin processing data after a pusher move has been initiated. This delay allows the probe to enter the tube without triggering a threshold event. Auto Retract Press the AUTO RETRACT (F11) button to enable or disable the function. When auto retract is enabled and a threshold event is encountered in the forward direction, recording will be started and the pusher will automatically start running in the reverse direction. The next threshold event will stop the pusher and stop recording. In addition, when auto retract is enabled, the pusher will start running in the reverse direction whenever the START RECORDING button is pressed. Any time the pusher is running and the STOP RECORDING button is pressed, the pusher will be stopped.
Page 6-11
Figure 6-11
NO
Adjust Tolerance Screen
Select Quadrant Channel Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now you can use the Control Knob to select a channel for the highlighted X-Y or strip chart display. Toggle the alarm on or off Press the TOGGLE button (F9) to turn the alarm on or off. Adjust the height of the alarm area 1. Press the VERTICAL ADJUST (F10) button to activate the height of the alarm area. 2. Use the control knob to adjust the height of the alarm area. Adjust the width of the alarm area 1. Press the HORIZ ADJUST (F11) button to activate the width of the alarm area. 2. User the control knob to adjust the width of the alarm area.
Page 6-12
TE
Adjusting the scaling (HORIZ/#) does not effect the analysis calculations.
CONTENTS
Adjust the scale of the data 1. Press the HORIZ (F12) to scale the flaw data. 2. Use the control knob to adjust the scale of the flaw data.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 6-13
SELECT QUADRANT CHANNEL (F8) TOGGEL (F9) VERTICAL ADJUST (F10) HORIZ ADJUST (F11) HORIZ (F12)
SELECT QUADRANT CHANNEL (F8) SPAN (F9) CHANNEL ROTATION (F10) VERTICAL ADJUST(F11) HORIZ ADJUST (F12)
MEASURE (F5)
SELECT QUADRANT CHANNEL (F8) CHANNEL ROTATION (F9) PEAK TO PEAK (F10) VERTICAL MAX (F11) MAX RATE (F12)
START ACQUIRE (F7) When you press this button, Acquisition begins, and the menu structure changes. See Figure 6-2.
Page 6-14
CONTENTS
1: GENERAL DESCRIPTION
DATA POSITION (F10) CURSOR SIZE (F11) ADJUST DISTANCE (F12) FILTER SETUP(F4) SELECT QUADRANT CHANNEL (F8) FILTER TYPE OFF (F9) HIGH FREQ CUTOFF (F10) DIFF DELTA (F11) SET ALL (F12) START ACQUIRE (F7) When you press this button, Acquisition begins, and the menu structure changes. See Figure 6-2.
Page 6-15
6.2.1
Press the CURSOR CONTROL (F2) button to access the cursor control options.
Select Quadrant / Channel Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now you can use the Control Knob to select a channel for the highlighted X-Y or strip chart display. Chart Zoom Press the CHART ZOOM (F2) button to expand the data in the strip chart. The chart will zoom from the cursor location. At a chart zoom of 100%, there is one data point per strip chart line. If you cannot zoom in as far as you would like, reduce your cursor size. Data Position Press the DATA POSITION (F10) button to scroll the data through the cursor. Cursor Size Press the CURSOR SIZE (F11) button to change the amount of data that is displayed in the four X-Y displays. Display Null Press the DISPLAY NULL (F12) button to center the data back in the X-Y displays.
Page 6-16
CONTENTS
6.2.2
Adjust Tolerance
6.2.3
Adjust Data
Press the ADJUST DATA (F4) button to access the data adjustment control menu. Here you have the ability to select displayed channels, adjust the size of the displayed signal, rotate the channels data, and adjust the datas vertical and horizontal position. For quick adjustment of data, zoom the strip chart and only cursor the data of interest.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Select Quadrant / Channel Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now the Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display. Span Press the SPAN (F9) button and rotate the Control Knob to adjust the size of the displayed signal for the selected XY display or strip chart. Channel Rotation Press the CHANNEL ROTATION (F10) button, and then use the Control Knob to rotate the selected channels rotation from 0 to 360 degrees. Vertical Adjustment Press the VERTICAL ADJUST (F11) button, and then rotate the Control Knob to adjust the vertical data position of the current X-Y display. The vertical position has no effect when a strip chart is highlighted.
Page 6-17
Horizontal Adjustment Press the HORIZONTAL ADJUST (F11) button, and then rotate the Control Knob to adjust the horizontal data position of the current X-Y display. The horizontal position has no effect when a strip chart is highlighted.
6.2.4
Measure
Press the MEASURE (F5) button to access the measure controls. There are three measurement types in the MIZ-28: peak to peak, vertical max, and max rate. For quick adjustment and measurement of data, zoom the strip chart and only cursor the data of interest.
Select Quadrant / Channel Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now the Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display. Channel Rotation Press the CHANNEL ROTATION (F9) button to rotate the current channels data to achieve the desired angle of the flaw response. Peak to Peak Press the PEAK TO PEAK (F10) button to show all the channels displayed peak to peak measurement. This measurement determines the maximum signal amplitude in any phase. Measurement displays in A to D counts and angle. Vertical Maximum Press the VERTICAL MAX (F11) to display all the channels vertical maximum measurement. Vertical maximum measurement determines the maximum signal amplitude in the vertical direction only, and displays in A to D counts.
Page 6-18
CONTENTS
Maximum Rate Press the MAX RATE (F12) button to display the maximum rate measurement. The maximum rate measurement determines the greatest rate of change in the data signal. This measurement displays the peak to peak A to D measurement and the maximum rate angle.
6.2.5
1: GENERAL DESCRIPTION
The probe speed check controls are the same as the cursor controls with the addition of a button labelled ADJUST SPEED. To check the speed of the probe when the data was collected, follow the steps below. 1. 2. 3. 4. Press the SPEED CHECK (F2) button on the second button page. Position a known landmark (typically a tube support) in the center of the data cursor in the strip chart. Adjust the cursor size to place one of the outer cursor lines on another known landmark. The distance and speed are displayed at the bottom of the screen. Press the ADJUST DISTANCE (F12) button, and then use the Control Knob to adjust the distance value to the known distance between the two landmarks of the cursor location. The speed value will update to reflect the rate of travel at the time of acquisition.
1: GENERAL DESCRIPTION
Page 6-19
6.2.6
The TUBE END DETECTION (F3) button on the second page of buttons will be present only if a Model III Controller is connected to the RS-232 connector on the back panel of the MIZ-28. When a Model III is connected, a pusher icon is displayed in the top right corner of the screen. This menu is the same as previously discussed in the test mode section with one exception in review mode a new button, SET THRESHOLD, is available.
Setting the Tube End Threshold Acquire the data from a tube end, and then press the STOP ACQUIRE button. Scroll the tube end signal to the cursor and press the SET THRESHOLD (F8) button. This will take the peak measurement of the tube end and save it to NVRAM. A message at the bottom of the page will appear saying, Threshold has been set.
CONTENTS
6.2.7
Filter Setup
Filter adjustments do not affect the analog probe signal. They are digital filters that optimize the presentation and aid in signal analysis. This allows you to experimentally change the filters in review mode without the need to reacquire the data. There are two types of filters available with the MIZ-28: a first order differential filter and a low pass filter. The differential filter is adjustable from 2 to 40 points; the low pass filter is adjustable from 0 Hz to about half of the current sample rate. When a channel has a filter associated with it, you will see the words DIFF, LP, or LP+DIFF on the end of the X-Y channel information.
1: GENERAL DESCRIPTION
Select Quadrant / Channel Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now the Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display. Filter Type Press the FILTER TYPE (F9) button and rotate the Control Knob to change the filter type associated with the selected channel. The options are off, low pass, differential, or a combination of low pass and differential. The low pass filter is a 41 coefficient digital filter with sharp cutoff characteristics. Its useful in suppressing a variety of unwanted signals including electrical interference. The first order differential filter displays the difference between the present data point and a previous data point. The closer the data points are together in time, the more highpass action the filter creates. The process is sometimes known as continuous balance, because the displayed signal has been compared to some recent value, rather than a fixed point. This filter is useful when the probe is scanned across dissimilar materials, as it tends to keep the signal on screen. The differential filter is sensitive to the direction of data change. Rising edges will show as positive signals, and falling edges will show as negative signals. For this reason, a single unfiltered pulse, which deflects upwards and then returns to null, will appear filtered as a positive and then a negative signal.
1: GENERAL DESCRIPTION
Page 6-21
High Frequency Cutoff Press the HIGH FREQ CUTOFF (F10) button and rotate the Control Knob to adjust the cutoff frequency of the low pass filter for the selected channel. The range is 0 Hz to about half of the current sample rate. Differential Delta Press the DIFF DELTA (F11) button to adjust the number of data points that are applied to the differential filter. Set All Press the SET ALL (F12) button to set all channels configured to the current channels filter settings.
6.2.8
Start Acquire
When you press the START ACQUIRE (F7) button, Acquisition begins and the screen changes to Test Mode (Section 6.1).
The next section of this operating guide describes the Analysis Setup screen.
Page 6-22
Analysis Setup
CONTENTS
1: GENERAL DESCRIPTION
Figure 7-1
Page 7-1
Figure 7-2
SELECT QUADRANT CHANNEL (F8) CHART ZOOM (F9) DATA POSITION (F10) CURSOR SIZE (F11) DISPLAY NULL(F12)
MEASURE (F3)
SELECT CHANNEL (F8) CHANNEL ROTATION (F9) PEAK TO PEAK (F10) VOLTAGE MULT (F11) MAX RATE (F12)
SELECT POINT (F8) ADJUST PERCENT (F9) DATA POSITION (F10) SET THIS CHANNEL (F11) SET ALL FREQS (F12)
SELECT CHANNEL (F8) SELECT NUM MIXES (F8) MIX 1 CHAN 1 (F9) MIX 1 CHAN 2 (F10) PERFORM MIX (F12)
Page 7-2
CONTENTS
Figure 7-3
Selecting the Curve Type 1. Press the SELECT CHANNEL (F8) button. 2. 3. 4. Rotate the Control Knob to select a channel. Toggle the CURVE TYPE (F9) button to select ANGLE, VOLT, or NO CURVE for the selected channel.
1: GENERAL DESCRIPTION
If you selected ANGLE or VOLT, press NUM ID POINTS (F10) to set the number of ID points, then press NUM OD POINTS (F11) to set the number of OD points. When you select an angle curve, you can set from 1 to 4 ID points and 3 to 5 OD points. When you select a volt curve, you can set from 3 to 7 points to describe the curve.
1: GENERAL DESCRIPTION
Page 7-3
7.3 Measure
Press the MEASURE (F3) button to select the measurement type of peak to peak or max rate. The measurement results will be displayed at the top of the curve display.
Figure 7-4
Voltage Mult Use the VOLTAGE MULT (F11) button to normalize all channels. Channel Rotation Use the CHANNEL ROTATION (F9) button to rotate the signals.
Page 7-4
CONTENTS
1: GENERAL DESCRIPTION
Figure 7-5
Adjust Curve
1: GENERAL DESCRIPTION
1. 2. 3. 4. 5. 6. 7.
Press the ADJUST CURVE (F4) button. Press the SELECT POINT (F8) button and select the first point to setup. Press the DATA POSITION (F10) button and position the data to the appropriate flaw signal. To set only the current channel selected, press SET THIS CHANNEL (F11). To set all the channels for this coil, press SET ALL FREQS (F12). The selected point will automatically increment to the next point. Adjust the data position to the next flaw on the standard and repeat for all the curve points.
Page 7-5
Figure 7-6
1. 2.
Press the SELECT CHANNEL (F8) button. Rotate the Control Knob to scroll through the channels.
Page 7-6
CONTENTS
Figure 7-7
1. 2. 3. 4. 5.
Collect a tube support signal in test mode. In the Analysis Setup screen, press the DATA MIXING (F6) button. Adjust SELECT NUM MIXES (F8) to 1. Press the MIX CHAN 1 (F9) button and select the first channel of the mix. Press the MIX CHAN 2 (F10) button and select the second channel of the mix.
Page 7-7
6.
Press the PERFORM MIX (F12) button to calculate and apply the mix coefficients.
Figure 7-8
Perform Mix
7.
To add another mix channel, press SELECT NUM MIXES (F8) and use the Control Knob to select 2. Repeat the previous channel selection and perform mix process as above.
Page 7-8
CONTENTS
Figure 7-9
Setup File
Save Setup Press the SAVE SETUP (F8) button to save the current analysis setup to the current recording directory. A directory will be created in the current recording directory (cal group) named MIZ28_SETUP, and within that directory the setup file will be saved as analysis.stp. Recall Setup Press the RECALL SETUP (F9) button to recall the setup file from the current directory.
1: GENERAL DESCRIPTION
The next section of this operating guide describes the Toolbox screen
1: GENERAL DESCRIPTION
Page 7-9
CONTENTS
1: GENERAL DESCRIPTION
Figure 8-1
Page 81
The Toolbox screen initially displays the following system information: the instruments serial number, the firmware revision number and the revision date.
Figure 8-2
Toolbox Screen
Serial number
Figure 8-3
SCROLL (F8) SET FREQUENCY (F9) SET WALL THICKNESS (F10) mm Ms/m (F11) Mils IACS
Page 8-2
CONTENTS
1: GENERAL DESCRIPTION
Figure 8-4
8.1.1
Select Coil
Press the SELECT COIL (F8) button, and then use the Control Knob to select which coil data you are plotting.
8.1.2
Adjust Position
Press the ADJUST POSITION (F9) button, and then use the Control Knob to move through the plot frequencies.
8.1.3
Adjust Range
Press the ADJUST RANGE (F10) button, and then use the Control Knob to zoom out (show a broader frequency range) or zoom in on the frequency of interest.
8.1.4
Adjust Scale
1: GENERAL DESCRIPTION
Press the ADJUST SCALE (F11) button, and then use the Control Knob to adjust the amplitude to bring big signals down into the screen or raise small signals up into the screen.
Page 83
8.1.5
Adjust Drive
Press the ADJUST DRIVE (F12) button, and then use the Control Knob to change the amount of drive driven out to the test coil.
Page 8-4
NO TE
When the frequency readout is RED, it means that the front end is saturated at that frequency and drive combination.
CONTENTS
1: GENERAL DESCRIPTION
Figure 8-5
1: GENERAL DESCRIPTION
8.2.1
Select Coil
1: GENERAL DESCRIPTION
Press the SELECT COIL (F8) button, and then use the Control Knob to select which coils data is being plotted.
8.2.2
Adjust Frequency
Press the ADJUST FREQUENCY (F9) button, and then use the Control Knob to change the test frequency applied to the coils.
1: GENERAL DESCRIPTION
Page 85
8.3
The MIZ-28 provides a list of materials with a cross-reference table to %IACS or MS/m. It also has two control fields that can help determine the best test frequency or depth of penetration for a specific material.
Figure 8-6
8.3.1
Press the SCROLL (F8) button, and then use the Control Knob to scroll through the list of materials.
8.3.2
Press the SET FREQUENCY (F9) button, and then use the Control Knob to change the test frequency. The standard depth of penetration values change to reflect the standard depth of penetration in the material at the set test frequency.
8.3.3
Press the SET WALL THICKNESS (F10) button, and then use the Control Knob to select the wall thickness of the tube to be tested. The Prime Frequency values will change to reflect the prime frequency for the particular material with the new wall thickness.
8.3.4
Press the
mm Ms/m Mils IACS mm Ms/m Mils IACS (F11) button to toggle the table between MS/m or %IACS.
The next section of this operating guide describes the Utilities screen.
Page 8-6
Utilities
CONTENTS
9.0 Utilities
The Utilities screen provides access to Language Selection (English, German, Italian, Spanish, and French), Speaker Control, Time/Date Input, IP Address Setting, Graphics Dump, and Remote Inputs. The MIZ-28 has built-in USB and TCP/IP support. To open the Utilities screen: 1. Press the Menu Display button. 2. 3. Use the Control Knob to select (highlight) UTILITIES. Press the Select button.
1: GENERAL DESCRIPTION
Figure 9-1
Page 9-1
Figure 9-2
IP ADDRESS (F4)
GRAPHICS DUMP(F5)
Page 9-2
CONTENTS
Figure 9-3
Select Language
2.
Press the corresponding language (F8 to F12) button to select one of the following languages: English (F8) Deutsch (F9) Italiano (F10) Espaol (F11) Franais (F12)
Page 9-3
Figure 9-4
Speaker On/Off
Page 9-4
CONTENTS
1: GENERAL DESCRIPTION
IO N
Figure 9-5
CA
Time/Date
UT
Setting the Time/Date will affect the acquire time of the recorded data.
Page 9-5
Figure 9-6
IP Address
Page 9-6
CONTENTS
1: GENERAL DESCRIPTION
Figure 9-7
Graphics Dump
Page 9-7
Figure 9-8
Remote Inputs
The next section of this operating guide describes the Diagnostics screen.
Page 9-8
Diagnostics
CONTENTS
10.0 Diagnostics
The MIZ-28 has built-in diagnostic tools, which can be activated to assure that the instrument is within calibration specifications.
1: GENERAL DESCRIPTION
N CA UT
2. 3.
The features in the Diagnostics menu are intended for use by Zetec calibration and repair or other personnel who are trained in MIZ-28 maintenance. The functionality of this menu is not intended for typical field use applications. For information about MIZ-28 diagnostics beyond that documented here, contact Zetec at 425-9742700.
To open the Diagnostics screen: 1. Press the Menu Display button. Use the Control Knob to select (highlight) DIAGNOSTICS. Press the Select button.
IO
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 10-1
LINEARITY (F2)
BANDWIDTH (F3)
MORE 1 OF 2 (F6) Changes menu to Screen 2 as illustrated below RETURN (F7) Moves you out of current screen
SCREEN 2
Primary Function Buttons
SIGNAL PATH (F1)
NOISE (F2)
FULL CAL(F4)
MORE 1 OF 2 (F6) Changes menu to Screen 1 as illustrated above RETURN (F7) Moves you out of current screen
The next section of this operating guide describes the Test Plans screen.
Page 10-2
Test Plans
CONTENTS
1: GENERAL DESCRIPTION
Figure 11-1
This section describes the menu functions of the Test Plan. For information on using test plans, see Test Plan Setup in Section 14.5.
1: GENERAL DESCRIPTION
Page 11-1
Figure 11-2
Page 11-2
CONTENTS
Figure 11-3
11.1.1 Scroll
Press the SCROLL (F8) button, then use the Control Knob to scroll through the outputs.
11.1.2 Go To Top
Press the GOTO TOP (F9) button to go to the top of the list.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 11-3
Figure 11-4
11.2.1 Scroll
Press the SCROLL (F8) button, and then use the Control Knob to scroll through the available test plans.
11.2.2 Open
When you have highlighted a test plan, press the OPEN (F9) button to open it.
Page 11-4
CONTENTS
Figure 11-5
11.3.2 Scroll
Press the SCROLL (F8) button to scroll through the list of test plans available for importing.
1: GENERAL DESCRIPTION
The next section of this operating guide describes the Mass Storage Setup screen.
1: GENERAL DESCRIPTION
Page 11-5
CONTENTS
1: GENERAL DESCRIPTION
1. 2. 3.
Press the Menu Display button. Use the Control Knob to select (highlight) MASS STORAGE SETUP. Press the Select button.
Page 12-1
NETWORK IP (F4)
Page 12-2
CONTENTS
1: GENERAL DESCRIPTION
2. 3.
Enter the name in the same way as entering a configuration name (see Naming the Configuration on page 4), or use the keyboard. When you go back to Test/Review and press the START RECORDING button, the new directory is automatically created and becomes the current recording directory.
1: GENERAL DESCRIPTION
Page 12-3
There are two file formats: PC and Eddynet. The PC format is used when Zetecs PC analysis software application is used to analyze the data. In the PC file format, the first three characters are the entry number, the next two characters are the ID identifier, the following three characters are the row (ROW) identifier, and the last three characters are the column (COL) identifier. For example: ###IDROW.COL. Summaries and messages are identified by an entry number followed by a message file extension. For example: ###.MSG. The Eddynet format is used when Zetecs Eddynet software application is used to analyze the data. When the Eddynet file format is selected, the Recording Dir selection is not displayed, as the recording selection is automatically selected when a new summary is written. When a new summary is written, it is assumed that you are starting a new cal group. The instrument looks at the recording directory, finds the highest cal group number, and creates a new directory for the next number. The format of the cal group name is the ID name, the ID identifier, the leg identifier (I or O), the word cal and cal group number. For example: ID##ICAL00001.
Page 12-4
CONTENTS
The Eddynet file naming convention is a letter indicating the file type (D-data, M-message, or S-summary), the leg identifier (I or O), the first letter of the row name, the row identifier, the first letter of the column name, the column identifier, the letter I (Index), and the file index number. For example: DOR###C###I###.
1: GENERAL DESCRIPTION
Page 12-5
Page 12-6
CONTENTS
1: GENERAL DESCRIPTION
Page 12-7
To set the mount point: 1. 2. Press the MOUNT POINT (F5) button. Press the ADJUST (F8) button, and then use the Control Knob to change the selection.
Page 12-8
NO
TE
The /raw ###/ and /mnt/rod###/raw mount points are for recording to UNIX computers.
CONTENTS
1: GENERAL DESCRIPTION
Page 12-9
The next section of this operating guide describes the Mass Storage screen.
Mass Storage
CONTENTS
1: GENERAL DESCRIPTION
Page 13-1
SCROLL (F8) OPEN DIRECTORY (F9) DELETE DIRECTORY (F10) COPY TO MASS STORAGE (F11) NOTE: Available only when a mass storage device is connected. COPY TO NETWORK (F10) NOTE: Available only when a network drive is mounted.
13.1.1 Scroll
Press the SCROLL (F8) button to scroll through the list of data files.
Page 13-2
CONTENTS
13.1.2 Open
Press the OPEN (F9) button to open the selected data file.
13.1.3 Delete
Press the DELETE (F10) button to delete a data file. At this point, you can still scroll through the files. Press the Select button to delete a data file.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
13.2.1 Scroll
Press the SCROLL (F8) button to scroll through the outputs.
NO
TE
1: GENERAL DESCRIPTION
Page 13-3
Page 13-4
CONTENTS
1: GENERAL DESCRIPTION
IO N CA UT
It is recommended that you are familiar with Eddy Current theory and application before using the MIZ-28, and that you use it only for the functions for which it has been designed. Zetec offers a wide variety of courses dealing with electromagnetic inspection theory and applications. This includes classes that are specific to either eddy current testing (ECT) or remote field testing (RFT). Some formal courses can be used to help qualify inspectors to meet specific certification criteria for either data acquisition or data analysis. Other informal courses can be designed with variable content and duration to meet specific application needs. Contact the Training Department at 425-974-2700 for more details.
14.1Bobbin
14.1.1 Bobbin Hardware Setup
Figure 14-1 illustrates the hardware setup. See Drawing Number 10020312 in Appendix A for a larger illustration that includes the parts list.
1: GENERAL DESCRIPTION
Page 14-1
2.
Page 14-2
CONTENTS
3. 4.
As required for your application, press the RECORDING DIR (F1) button to change the name of the Recording Directory (see Section 4.1 for information on changing names). If needed, press the RECORD TO (F3) button to change where you record data to. This is the location where your setup, messages, summaries and data will be saved. Press ADJUST (F8) and use the Control Knob to change the Record To destination. Press RETURN (F7) when you are finished. Press the Menu Display button. Rotate the Control Knob until Configuration is highlighted. Press the Select button.
5. 6.
Page 14-3
7. 8. 9.
Press the CONFIG NAME (F1) button. Name the configuration, and press the RETURN (F7) button. Press the CONFIG OPTIONS (F2) button. If CONFIG TYPE is not already highlighted, use the Control Knob to select it.
10. Press the ADJUST (F9) button. 11. Use the Control Knob to select MULTIPLEXED. 12. Press the SELECT (F8) button. 13. Press the SAMPLE RATE (F3) button. 14. Use the Coarse/Fine Button in conjunction with the Control Knob to set the desired Samples per Second value. The Coarse/Fine button turns the red LED light off and on. When its on, the button is in coarse control, which provides larger adjustment increments.
15. Press FREQUENCY SELECT (F4), then press ADJUST (F9). Use the Coarse/Fine Button in conjunction with the Control Knob to set to the proper frequency. 16. Press the Select button. 17. Press DRIVE SELECT (F5). 18. Rotate the Control Knob until the desired voltage has been highlighted blue. Push the ADJUST (F9) button. This will turn the highlighted voltage green. Adjust by rotating the Control Knob to the correct drive voltage. You
Page 14-4
NO
TE
When setting the sample rate, use a minimum of 400 samples per 12 inches per second.
CONTENTS
can adjust each voltage drive separately, or adjust one and press the SET ALL (F10) button to change all the drive voltages settings at once. Use of lower drive voltages in specific frequency time slots may be required to prevent signal saturation of large volume flaw responses. A flow chart that defines the functional data analysis logic of the output from each channel should be used to establish the proper drive voltages in each frequency time slot. The largest amplitude response from a calibration standard that is representative of a known rejectable flaw type or condition will help define these limits.
NO
TE
1: GENERAL DESCRIPTION
19. Press the CHANNEL SELECT (F6) button. 20. Select COIL NUMBER (F8). This enables you to move the Control Knob horizontally across the coil selection boxes. After youve selected a coil, press the FIELD (F9) button and use the Control Knob to move vertically through the selected coils frequencies. (In summary, when COIL NUMBER (F8) is highlighted, the Control Knob selects horizontally between coils. When FIELD (F9) is highlighted, the Control Knobs selects vertically through the selected coils frequency fields.) Press the TOGGLE (F10) button to activate channels for data collection. Press the SET ALL (F11) button to turn on frequencies for that selected coil. In some testers a mode of operation (differential, absolute, driver/pick-up, etc.) might be selected in the test configuration process. In the MIZ-28 this step is not required. The specific probe adaptor selected will define whether the displayed information is differential or absolute once the coil functionality is selected (On/Off). For the bobbin probe application, those channels activated under Coil 1 will be operating in the self-comparison differential mode. Channels activated under Coil 2 will be operating in the external-reference differential mode.
NO TE
1: GENERAL DESCRIPTION
21. Press the CONFIG STORAGE (F7) button to store the configuration or bring up the config list. Press the SAVE CONFIG (F8) button to save the modified config. The instrument will display the message, Configuration saved, at the bottom of the screen. 22. Press the Menu Display button, highlight TEST/REVIEW and press the Select button. 23. Place the Reference probe in the reference standard, and the Test probe in the calibration standard. Press the SELECT QUADRANT Channel (F8) button. Use the Control Knob to scroll through the channel fields. 24. Press SELECT QUADRANT Channel (F8) again. Now CHANNEL will be capitalized and the selected field can be adjusted with the Control Knob. 25. Press SELECT quadrant CHANNEL (F8) to enable you to move to the upper left area of the lissajous. Press SELECT QUADRANT Channel (F8) and use the Control Knob to change the channel to C1 400 KHz. Move to the upper right area of the lissajous and change that channel to C2 400 KHz. 26. Move to the top of the first strip chart and adjust the channel to display C1F1 V. Move to the top of the second strip chart and adjust to C2F3 V. 27. Press the START ACQUIRE (F7) button and, with the test probe in good metal, press the DISPLAY NULL (F12) button to hardware null the instrument. 28. Pull the probe through the calibration standard at the same speed as the actual exam speed. Then press the STOP ACQUIRE (F7) button.
Page 14-5
29. Press the DATA POSITION (F10) button, and use the Control Knob to move to the appropriate point in the data pull. 30. Press CHART ZOOM (F9) and use the Control Knob to zoom the strip chart to full screen. 31. Press DATA POSITION (F10) to move the cursor window over the 100% TWH. Pressing the CURSOR SIZE (F11) button and adjusting with the Control Knob will enlarge or shrink the data window. 32. Press the ADJUST DATA (F4) button, and then press the SPAN (F9) button to allow adjustment of the span. Turn the Control Knob to adjust signal to appropriate display size. Spans on strip charts are not controlled by spans changed in the lissajous, and must be changed as separate items. 33. Press the MEASURE (F6) button, and then the MAX RATE(F12) button to give angles for the signal inside the data window.
Page 14-6
CONTENTS
34. Press the CHANNEL ROTATION (F9) button and use the Control Knob to adjust channels to the appropriate rotation. 35. Press the Menu Display button, highlight CONFIGURATION, and press the Select button. 36. Press the CONFIG STORAGE (F7) to store the configuration, including rotations and spans. 37. Press SAVE CONFIG (F8) to save the modified configuration. The message Configuration saved, will be displayed at the bottom of the screen. 38. Via the Menu Display button and Control Knob, return to the TEST/REVIEW screen. 39. Press START ACQUIRE (F7). 40. Press the MESSAGE SUMMARY (F5) button to access the message menu. Press F5 again (now labeled SUMMARY FORM) to access the summary form. 41. Enter all appropriate information in the form and press SAVE THIS FORM (F11). 42. Press the WRITE SUMMARY (F9) button. Important: at this point a new cal group has been generated.
Page 14-7
43. Press the MANUAL TUBE INFO (F3) button. 44. Press ENABLE STD RUN (F11). This will change your encodes to the 999/999 for recorded calibration pulls. 45. Press START RECORDING (F6), pull a calibration standard run, and then press STOP RECORDING (F6). 46. Repeat 2 more times. 47. Press DISABLE STD RUN (F11) and update the tube encodes. 48. Start Testing.
Page 14-8
CONTENTS
14.2MRPC
14.2.1 MRPC Hardware Setup
Figure 14-8 illustrates the hardware setup. See Drawing Number 10020311 in Appendix A for a larger illustration that includes the parts list.
1: GENERAL DESCRIPTION
Page 14-9
2.
Page 14-10
CONTENTS
3. 4.
As required for your application, press the RECORDING DIR (F1) button to change the name of the Recording Directory (see Section 4.1 for information on changing names). If needed, press the RECORD TO (F3) button to change where you record data to. This is the location where your setup, messages, summaries and data will be saved. Press ADJUST (F8) and use the Control Knob to change the Record To destination. Press RETURN (F7) when you are finished. Press the Menu Display button. Rotate the Control Knob until Configuration is highlighted.
5.
Page 14-11
6.
7. 8. 9.
Press the CONFIG NAME (F1) button. Name the configuration, and press the RETURN (F7) button. Press the CONFIG OPTIONS (F2) button. If CONFIG TYPE is not already highlighted, use the Control Knob to select it.
10. Press the ADJUST (F9) button. 11. Use the Control Knob to select MULTIPLEXED. 12. Press the SELECT (F8) button. 13. Rotate the Control Knob to highlight EXTERNAL TRIGGER. 14. Press the ADJUST (F9) button. 15. Using the Control Knob, toggle the external trigger ON. 16. Press the SELECT (F8) button. 17. Press the SAMPLE RATE (F3) button. 18. Use the Coarse/Fine Button in conjunction with the Control Knob to set the desired Samples per Second value. The Coarse/Fine button turns the red LED light off and on. When its on, the button is in coarse control, which provides larger adjustment increments. The sample rate for MRPC applications is selected based on the rotation speed (RPM) of the probe head and careful control of the axial translation speed of the probe.
Page 14-12
NO
TE
CONTENTS
19. Press FREQUENCY SELECT (F4), then press ADJUST (F9). Use the Coarse/Fine Button in conjunction with the Control Knob to set to the proper frequency. 20. Press the Select button. 21. Press DRIVE SELECT (F5). 22. Rotate the Control Knob until the desired voltage has been highlighted blue. Push the ADJUST (F9) button. This will turn the highlighted voltage green. Adjust by rotating the Control Knob to the correct drive voltage. You can adjust each voltage drive separately, or adjust one and press the SET ALL (F10) button to change all the drive voltages settings at once. 23. Press the CHANNEL SELECT (F6) button. 24. Select COIL NUMBER (F8). This enables you to move the Control Knob horizontally across the coil selection boxes. After youve selected a coil, press the FIELD (F9) button and use the Control Knob to move vertically through the selected coils frequencies. (In summary, when COIL NUMBER (F8) is highlighted, the Control Knob selects horizontally between coils. When FIELD (F9) is highlighted, the Control Knobs selects vertically through the selected coils frequency fields.) Press the TOGGLE (F10) button to activate channels for data collection. Press the SET ALL (F11) button to turn on frequencies for that selected coil.
1: GENERAL DESCRIPTION
There are many possible options for the channel selections. The channel and coil number selections shown here are based on the number and position of the actual coils on the probe and the specific probe adaptor used. The MRPC probe used in this example has three different coils: a standard pancake coil (Coil 1), a coil that is wound such that it is sensitive to axially oriented flaws (Coil 2), and a third coil that is wound and mounted such that it is sensitive to circumferentially oriented flaws (Coil 5). The use of Coils 1, 2 and 5 on the Configuration Screen is based on the wiring of the probe adaptor used in this example.
NO TE
1: GENERAL DESCRIPTION
25. Press the CONFIG STORAGE (F7) button to allow you to store the config or bring up the config list. Press the SAVE CONFIG (F8) button to save the modified config. The instrument will display the message, Configuration saved, at the bottom of the screen. 26. Press the Menu Display button, highlight TEST/REVIEW and press the Select button. 27. Place the Reference probe in the reference standard, and the Test probe in the calibration standard. Press the SELECT QUADRANT Channel (F8) button. Use the Control Knob to scroll through the channel fields. 28. Press SELECT QUADRANT Channel (F8) again. Now CHANNEL will be capitalized and the selected field can be adjusted with the Control Knob. 29. Press SELECT quadrant CHANNEL (F8) to enable you to move to the upper left area of the lissajous. Press SELECT QUADRANT Channel (F8) and use the Control Knob to change the channel to C1 400KHz. Move to the upper right area of the lissajous and change the channel to C2 400KHz. Move to the lower right area of the lissajous and change the channel to C5 400KHz. 30. Move to the top of the first strip chart and adjust the channel to display C1F1 V. Move to the top of the second strip chart and adjust to C2F3 V. 31. Press the START ACQUIRE (F7) button and, with the test probe in good metal, press the DISPLAY NULL (F12) button to hardware null the instrument.
Page 14-13
32. Pull the probe through the calibration standard at the same speed as the actual exam speed. Then press the STOP ACQUIRE (F7) button. 33. Press the DATA POSITION (F10) button, and use the Control Knob to move to the appropriate point in the data pull. 34. Press CHART ZOOM (F9) and use the Control Knob to zoom the strip chart to full screen. 35. Press DATA POSITION (F10) to move the cursor window such that Coils 1 and 2 show a maximum response from the selected (100%) axial EDM notch.
A calibration standard drawing may be required to verify flaw location and orientation within each standard. Each cal standard could have a range, from 20% to 100%, of ID and/or OD EDM notches. There might also be multiple flaws in the same cross-sectional area. Since the three coils on the probe used in this example are oriented at 120 degree intervals around the probe body, each coil could have a maximum flaw response at slightly different time locations within the data stream.
36. For coil 5, repeat with the 100% Circ indication. Pressing the CURSOR SIZE (F11) button and adjusting with the Control Knob will enlarge or shrink the data window. 37. Press the ADJUST DATA (F4) button, and then press the SPAN (F9) button to allow adjustment of the span. Turn the Control Knob to adjust the signal to the appropriate display size.
NO TE
Page 14-14
38. Press MEASURE (F6), and then PEAK TO PEAK (F10) to give angles for the signal inside the data window. 39. Press the CHANNEL ROTATION (F9) button and use the Control Knob to adjust channel responses to the appropriate rotation (noise horizontal, flaws up to the left). The response for Coil 1 (the pancake coil) and Coil 2 (the axially sensitive coil) should be adjusted from a known (100%) Axial notch indication.
NO
Spans on strip charts are not controlled by spans changed in the lissajous, and must be changed as separate items.
TE
CONTENTS
40. Repeat the phase rotation adjustment for Coil 5 (the circumferential flaw sensitive coil) from a 100% circumferential notch in the standard. 41. Navigate back to the Configuration screen. 42. Press the CONFIG STORAGE (F7) to store the configuration, including rotations and spans. 43. Press SAVE CONFIG (F8) to save the modified configuration. The message Configuration saved, will be displayed at the bottom of the screen. 44. Via the Menu Display button and Control Knob, return to the TEST/REVIEW screen. 45. Press START ACQUIRE (F7). 46. Press the MESSAGE SUMMARY (F5) button to access the message menu. Press F5 again (now labeled SUMMARY FORM) to access the summary form. 47. Enter all appropriate information in the form and press SAVE THIS FORM (F11). 48. Press the WRITE SUMMARY (F9) button. Important: at this point a new cal group has been generated.
1: GENERAL DESCRIPTION
49. Press the MANUAL TUBE INFO (F3) button and then press ENABLE STD RUN (F11). This will change your encodes to the 999/999 for recorded calibration pulls. 50. Press START RECORDING (F6), pull a calibration standard run, and then press STOP RECORDING (F6). 51. Repeat 2 more times. 52. Press DISABLE STD RUN (F11) and update the tube encodes. 53. Start Testing.
1: GENERAL DESCRIPTION
Page 14-15
Example: 300 RPM = 5 RPS (Rotations Per Second) 60 400 Samples per Second = 80 data points per scan 5 RPS 400 Samples per Second = 0.1666 inches per second pull speed (30 axial) (80 dps) Circumferential Digitizing rate = (Tester Sample rate) Circumferential scan (in inches per second) Circumferential scan rate (in inches per second) = (Rotations Per Second) x (Circ. in inches) (Rotations Per Second) = RPM 60 (Circumference in inches) = (3.1416) x (dia.) (use I.D.) = (circumferential samples per inch))
0.75 OD tubing with 0.050 wall has an ID of 0.650 Example: 300 RPM = 5 RPS (Rotations Per Second) 60 0.650 ID x 3.1416 = 2.377 Circumference in inches 5 RPS x 2.377 = 11.88 Circumferential scan rate (in inches per second) 400 Samples per Second = 33.7 Circumferential Samples per inch 11.88
Page 14-16
CONTENTS
14.3RFT
This section describes the RFT setup for ferromagnetic tubing. Required equipment: Miz-28 Item # 10020198 RFT Amplifier Item # 10003344-1 Cable-1/2-12P/2X18P-3FT-MIZ-28 RFT AMP DIFF/ABS Cable Item # 10019715-3 RFT Probe-Dual Exciter 12 Ft. Ground Cable Assembly Item # 2-8350-1 (also available in lengths of 15, 25 and 50 feet) RFT Dual Exciter Extension Cable Item # 10011614-* Keyboard with USB connector.
1: GENERAL DESCRIPTION
1.
1: GENERAL DESCRIPTION
TE NO
2. 3.
Connect 18-Pin (DIFF) to Coils 1 & 2 Connect 18-Pin (ABS) to Coils 5 & 6
Connect 12-Pin and RFT Dual Exciter Extension Cable to RFT Amplifier Tester Coil 1. Attach the ground cable assembly to the amplifier and to the component to be inspected.
Page 14-17
4. 5.
On the amplifier, select Dual Mode. Connect the power cord to the amplifier. Turn the MIZ-28 on.
10. Use the Control Knob to select SIMULTANEOUS. 11. Press the Select button. 12. Use the Control Knob to highlight NUM FREQS. 13. Press the ADJUST (F9) button. 14. Use the Control Knob to change the number of frequencies to 1. 15. Press the Select button. 16. Press FREQUENCY SELECT (F4), then press ADJUST (F9). Use the Coarse/Fine Button in conjunction with Control Knob to set to the desired value. 17. Press the Select button. 18. Press DRIVE SELECT (F5), then ADJUST, (F9). Use the Coarse/Fine Button in conjunction with the Control Knob to set the correct drive voltage. 19. Press the Select button.
TE
Page 14-18
If multiple frequencies are desired, a maximum of 20 volts is available for all timeslots.
20. Press the CHANNEL SELECT (F6) button. 21. Select COIL NUMBER (F8), then use the Control Knob to select Channel 1. Press the TOGGLE (F10) button to turn it on. 22. Use the Control Knob to select Channel 5, then the TOGGLE (F10) button to turn it on. 23. Toggle all other channels off. 24. Press the SAMPLE RATE (F3) button. Use the Coarse/Fine Button in conjunction with the Control Knob to set the desired Samplers per Second value. 25. Press the CONFIG STORAGE (F7) button. 26. Press the SAVE CONFIG (F8) button. Your setup should look like that in Figure 14-16.
NO
CONTENTS
27. Change to the TEST/REVIEW menu (press the Menu Display Button, rotate the Control Knob to highlight TEST/ REVIEW and press the Select button). 28. Press the START ACQUIRE (F7) button. 29. Toggle the RFT power switch to the On position. 30. Insert the probe into a tube in the interior of the tube bundle and watch the signal formation as the probe is removed (on-the-pull) from the tube bundle.
1: GENERAL DESCRIPTION
Nominal probe pull speed will typically be less than 12/second to maintain high quality signal responses. In most inspections this will be accomplished manually instead of using automated pusher/puller equipment.
31. Note the responses from the support plates, background noise (PV variations), and the tube end exit signal. 32. Repeat this process in several other tubes in the same area of the bundle, noting any variation in the expected or nominal responses to the signals of interest. 33. Adjust the frequency (as necessary) to correct the phase angle of the responses to these known signal sources to be as close as possible to those responses seen in the calibration sample. At the same time keep in mind that it is important to try to maintain a favorable S/N ratio.
NO TE
TE
NO
It may be necessary to adjust frequency more than once during the examination to compensate for tube wall condition variations found across the entire bundle.
1: GENERAL DESCRIPTION
Page 14-19
14.4
HVAC
Page 14-20
CONTENTS
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
2.
Page 14-21
3. 4.
As required for your application, press the RECORDING DIR (F1) button to change the name of the Recording Directory (see Section 4.1 for information on changing names). If needed, press the RECORD TO (F3) button to change where you record data to. This is the location where your setup, messages, summaries and data will be saved. Press ADJUST (F8) and use the Control Knob to change the Record To destination. Press RETURN (F7) when you are finished. Press the Menu Display button. Rotate the Control Knob until Configuration is highlighted.
5.
Page 14-22
CONTENTS
6.
7. 8. 9.
Press the CONFIG NAME (F1) button. Name the configuration, and press RETURN (F7). Press CONFIG OPTIONS (F2). If CONFIG TYPE is not already highlighted, use the Control Knob to select it.
10. Press the ADJUST (F9) button. 11. Use the Control Knob to select MULTIPLEXED. 12. Press the SELECT (F8) button.
1: GENERAL DESCRIPTION
13. Press the SAMPLE RATE (F3) button. 14. Use the Coarse/Fine Button in conjunction with the Control Knob to set the desired Samples per Second value. The Coarse/Fine button turns the red LED light off and on. When its on, the button is in coarse control, which provides larger adjustment increments.
15. Press FREQUENCY SELECT (F4), then press ADJUST (F9). Use the Coarse/Fine Button in conjunction with the Control Knob to set to the proper frequency. 16. Press the Select button. 17. Press DRIVE SELECT (F5). 18. Rotate the Control Knob until the desired voltage has been highlighted blue. Push the ADJUST (F9) button. This will turn the highlighted voltage green. Adjust by rotating the Control Knob to the correct drive voltage. You can adjust each voltage drive separately, or adjust one and press the SET ALL (F10) button to change all the drive voltages settings at once. 19. Press the CHANNEL SELECT (F6) button.
NO
TE
When setting the sample rate, use a minimum of 400 samples per 12 inches per second.
1: GENERAL DESCRIPTION
Page 14-23
20. Select COIL NUMBER (F8). This enables you to move the Control Knob horizontally across the coil selection boxes. After youve selected a coil, press the FIELD (F9) button and use the Control Knob to move vertically through the selected coils frequencies. (In summary, when COIL NUMBER (F8) is highlighted, the Control Knob selects horizontally between coils. When FIELD (F9) is highlighted, the Control Knobs selects vertically through the selected coils frequency fields.) Press the TOGGLE (F10) button to activate channels for data collection. Press the SET ALL (F11) button to turn on frequencies for that selected coil. 21. Press the CONFIG STORAGE (F7) button to allow you to store the config or bring up the config list. Press the SAVE CONFIG (F8) button to save the modified config. The instrument will display the message, Configuration saved, at the bottom of the screen. 22. Press the Menu Display button, highlight TEST/REVIEW and press the Select button.
The examples shown are for a single AC3 probe application. This fulfills most inspection requirements. Channel 1 (9.0kHz bobbin) provides good sensitivity to pit-like indications in the finned area. Channel 2 (3.0 kHz bobbin) provides good resolution of wear damage in the saddle area. Channel 3 (3.0 kHz axially-wound coils) provides detection of circumferential cracking. Other probe input port and channel selection options are possible and, in rare situations, the use of two AC3 probes can also be configured to provide additional detection and sizing.
23. Press the SELECT QUADRANT Channel (F8) button. Use the Control Knob to scroll through the channel fields. 24. Press SELECT QUADRANT Channel (F8) again. Now CHANNEL will be capitalized and the selected field can be adjusted with the Control Knob. 25. Press SELECT quadrant CHANNEL (F8) to enable you to move to the upper left area of the lissajous. Press SELECT QUADRANT Channel (F8) and use the Control Knob to change the channel to C1 9KHz. Move to the upper right area of the lissajous and change the channel to C1 3 KHz. Move to the lower right area of the lissajous and change the channel to C3 3KHz. 26. Move to the top of the first strip chart and adjust the channel to display C1F1 V. Move to the top of the second strip chart and adjust to C2F3 V.
Page 14-24
NO TE
CONTENTS
27. Place the probe in the reference standard. 28. Press the START ACQUIRE (F7) button and, with the test probe in good metal, press the DISPLAY NULL (F12) button to hardware null the instrument. 29. Pull the probe through the calibration standard at the same speed as the actual exam speed. Then press the STOP ACQUIRE (F7) button. 30. Press the DATA POSITION (F10) button, and use the Control Knob to move to the appropriate point in the data pull.
1: GENERAL DESCRIPTION
31. Press CHART ZOOM (F9) and use the Control Knob to zoom the strip chart to full screen. 32. Press DATA POSITION (F10) to move the cursor window over the 100% TWH. Pressing the CURSOR SIZE (F11) button and adjusting with the Control Knob will enlarge or shrink the data window. 33. Press the ADJUST DATA (F4) button, and then press the SPAN (F9) button to allow adjustment of the span. Turn the Control Knob to adjust signal to appropriate display size. Spans on strip charts are not controlled by spans changed in the lissajous, and must be changed as separate items. 34. Press the MEASURE (F6) button, and then the PEAK TO PEAK (F10) button to give angles for the signal inside the data window. 35. Press the CHANNEL ROTATION (F9) button and use the Control Knob to adjust channels to the appropriate rotation. Typically, Channels 1 and 2 are adjusted to put the 100% hole response at 40 degrees, or the dent response seen into and out of the land area can be adjusted to the horizon.
1: GENERAL DESCRIPTION
Page 14-25
36. Repeat for Channel 3 (Coil 3 / 3 KHz) on the circumferential notch in the standard. Rotate this response to 90 degrees. Since axially-wound coils are at different position within the probe body, this circumferential notch indication will be shifted in time from any bobbin coil responses.
Depending on the circumferential notch orientation with respect to the axially-wound coils, the notch response may be either positive or negative from the nominal balance point. These coils will only respond to circumferential indications.
37. Navigate back to the Configuration screen. 38. Press the CONFIG STORAGE (F7) to store the configuration, including rotations and spans. 39. Press SAVE CONFIG (F8) to save the modified configuration. The message Configuration saved, will be displayed at the bottom of the screen. Note: the remaining steps apply if you are recording data for later use. 40. Via the Menu Display button and Control Knob, return to the TEST/REVIEW screen. 41. Press START ACQUIRE (F7). 42. Press the MESSAGE SUMMARY (F5) button to access the message menu. Press F5 again (now labeled SUMMARY FORM) to access the summary form. 43. Enter all appropriate information in the form and press SAVE THIS FORM (F11). 44. Press the WRITE SUMMARY (F9) button. Important: at this point a new cal group has been generated.
45. Press the MANUAL TUBE INFO (F3) button. 46. Press ENABLE STD RUN (F11). This will change your encodes to the 999/999 for recorded calibration pulls. 47. Press START RECORDING (F6), pull a calibration standard run, and then press STOP RECORDING (F6). 48. Repeat 2 more times.
Page 14-26
NO
TE
CONTENTS
49. Press DISABLE STD RUN (F11) and update the tube encodes. 50. Start Testing.
1: GENERAL DESCRIPTION
4.
From the Import Test Plans screen, select the IMPORT TEST PLAN (F5) button to list all the available plans from the source directory.
1: GENERAL DESCRIPTION
Page 14-27
5.
For the source directory, select either MASS STORAGE or NETWORK. The Mass Storage option will only display when a USB device is attached; the Network option will only display when a network drive is mounted.
6.
If the current destination is not correct, press the CHANGE DEST (F2) button. Select the destination directory (INTERNAL, MASS STORAGE, or NETWORK).
Page 14-28
CONTENTS
7.
Press the INTERNAL (F6) button to import the test plan to the internal hard drive.
As an example of the test plan possibilities, you can import a test plan to the internal hard drive, remove the USB mass storage, and record to a network drive. The test plan can also be run directly from the USB device, or the test plan can be maintained on a network drive.
8.
Press the SCROLL (F6) button to select the desired test plan, and then press the IMPORT TEST PLAN button. For example, this screen shows that the water box 2 test plan.csv file will be imported to the internal drive. You can import as many test plans as you like.
NO
TE
Page 14-29
9.
Press the SELECT TEST PLAN (F4) button to display a list of all the test plans that have been imported.
10. Press the SCROLL (F8) button to select the desired plan, and then press the OPEN button. This will open the test plan and automatically put you in the Current Test Plan screen. Here you have the option of scrolling through the plan to select an entry to start testing. 11. Press the GOTO TOP (F9) button to place the cursor at the top of the list.
Page 14-30
CONTENTS
12. Press the FIND NEXT UNTESTED (F10) button to find the next untested entry from the current cursor location. 13. Press the MARK/UNMARK (F11) button to manually unmark and mark entries.
14. From here, press the Main Menu button and select TEST/REVIEW. 15. Select the START ACQUIRE (F7) button. 16. Select the RUN TEST PLAN (F4) button. The selected entry from the test plan shows up bold in the test window. 17. Press the START RECORDING (F6) button and the test is automatically marked off the list as tested (first character = *). The next entry to be tested shows up in bold.
Page 14-31
18. To skip entries in the list, go to the Test Plan screen (select TEST PLAN from the Main Menu) and scroll down past the entries you want to skip over. When you go back to the Test/Review screen, the current entry will be where you left the cursor, and testing will continue down the list. Alternatively, from the Run Test Plan menu, select the SCROLL (F8) button to scroll down to the desired entry. 19. To re-test a tube, scroll back to the desired tube in the Run Test Plan menu or from the Test Plan screen. Press the START RECORDING button to test the next untested entry in the list.
The next section of this operating guide contains MIZ-28 Specifications and features.
Page 14-32
CONTENTS
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
Page 14-33
CONTENTS
NO
TE
Multiple USB devices can be operated together but only one at a time of each type, i.e., only 1 printer, 1 keyboard, or 1 mass storage USB device at a time.
1: GENERAL DESCRIPTION
Page 15-1
Description
Pin
Function
1 1
44
1 2 3 4 5 6 7 8 9
The Ethernet interface requires CAT 5 cable that contains separate transmit and receive pairs for proper operation with 100BaseTX. For operation with 10Base-T, a CAT 3 cable or better is required. 1 2 3 4 5 6 7 8
Page 15-2
CONTENTS
Description
Encoder 2Q Encoder 2A External Trigger Analog Out 1 Analog Out 2
Pin
10 1 16 8 17
Function
1: GENERAL DESCRIPTION
Monitor signals are available at the rear panel on a standard 15-pin high-density female D-Sub.
1: GENERAL DESCRIPTION
DRIVER 2B RETURN COIL 2B COIL 2B RETURN COIL 2A RETURN COIL 2A GROUND DRIVER 1B RETURN COIL 1B COIL 1B RETURN COIL 1A RETURN COIL 1A DRIVER 1A DRIVER 2B DRIVER 2A RETURN DRIVER 2A DRIVER 1B DRIVER 1A RETURN NOT USED
Probe Connectors A B C D E F G H J K L M N P R S T U
Page 15-3
15.2Instrument Control
The instrument can be controlled with the MIZ-28 front panel or USB keyboard. For detailed information, see Section 3.0, General Operation.
Page 15-4
CONTENTS
15.3MIZ-28 Features
The MIZ-28 features are described below.
1: GENERAL DESCRIPTION
15.3.3 Memory
RAM The MIZ-28 instrument allocates 8 MB of RAM memory as a buffer for eddy current data. This buffer provides for data setup, review, and signal measurement. NVSRAM The instrument has NVSRAM for the current configuration and user variables, so that when power is cycled off and on the instrument the settings will not be lost. Hard Drive The MIZ-28 instrument has a 40 GB internal hard drive where configurations and data can be stored and recalled. Test plans can be stored and run from the internal hard drive. USB Mass Storage Configurations and data may also be stored on an external USB mass storage device connected to the instrument. Test plans can be imported to the internal hard drive from USB devices or stored and run from the USB device itself.
1: GENERAL DESCRIPTION
1: GENERAL DESCRIPTION
TE NO
Only one USB device of each type may be connected to the MIZ-28 at a time.
1: GENERAL DESCRIPTION
Page 15-5
15.3.9 Filters
Three filter types are available: bandpass, 1st-order differential high-pass, and a combination differential bandpass filter. The bandpass filter is adjustable from 0 Hz to half the sample. The high end of the bandpass filter limit is relative to the sample rate setting. The differential high-pass filter is a continuous balance type that is particularly useful for keeping the signal on screen when scanning across dissimilar materials. The differential filter can be adjusted from 2 to 40 data points.
Page 15-6
CONTENTS
Function
Determines the maximum signal amplitude in any phase. Determines the maximum signal amplitude in the vertical direction only. Determines the greatest rate of change.
Display
Amplitude and angle.
Vert Max
Amplitude.
Max Rate
1: GENERAL DESCRIPTION
Page 15-7
Appendix A
CONTENTS
Appendix A
N IO
The MIZ-28 should only be installed and run by persons who have been adequately trained by authorized Zetec personnel.
This section contains the following configuration drawings for the Bobbin Coil, MRPC, RFT and HVAC setups.
1: GENERAL DESCRIPTION
10020311 System Config IG MIZ-28 MRPC 3D Probe Pusher 10020312 System Config IG MIZ-28 Diff/Abs Bobbin 10020313 System Config IG MIZ-28 RFT 10020339 System Config IG MIZ-28 AC-3 with Reference Probe 10020340 System Config IG MIZ-28 AC-3 without Reference Probe
CA
UT
1: GENERAL DESCRIPTION
10020341 System Config IG MIZ-28 AC-3 w/Diff/Diff ADP & no Reference Probe The last three drawings are for HVAC applications.
Page A-1
Index
Index
Numerics
10/100Base-T connector 15-2 4K filter 4-8
C
cables 1-2 calibration 1-2 canned configurations 4-1, 5-1 Change Dest button 11-5, 14-28 Change Message button 6-7 Change Source button 11-5 changing directory 5-8 channel 3-7, 4-1, 4-11 selecting 4-11, 7-3, 7-6 Channel Rotation button 6-17, 6-18, 7-4 Channel Select button 4-2, 4-11 character table 4-4, 5-9 Chart Rate button 6-4 Chart Zoom button 6-16 checklist 3-1 cleaning 2-2 Clear button 4-4, 5-9, 6-4 Clear Message button 6-8 clearing specific operator data 6-10 Coarse/Fine adjustment button 3-4, 14-4 LED indicator 3-4, 14-4 Coil Number button 4-11 coils 3-13, 3-15, 3-17, 3-19 select 8-3, 8-5 Config List button 5-1 Config List screen 3-7 Config Name button 4-4, 5-9 Config Options button 4-5, 4-6 Config Storage button 4-2, 5-1
A
AC3 probe 4-5 acquisition, start 6-22 Adjust button 4-10, 12-4, 12-7 Adjust Curve button 7-5 Adjust Data button 6-17 Adjust Distance button 6-19 Adjust Frequency button 8-5 adjusting drive 8-4 frequency 8-5 position 8-3 range 8-3 scale 8-3 Analysis Setup screen 7-1 array probe 4-5 Auto Retract button 6-11
B
back panel 2-1 Back Space (F11) button 4-4, 5-9 Balance button 6-5 Balance Number button 4-11 bandpass filter 2-1 bobbin probe 3-2, 3-13, 4-5, A-1 configuration 4-1 bobbin probe adapter 3-1
Page I-1
configuration Bobbin Coil 4-1 bobbin probe 4-1 canned 4-1, 5-1 display 15-4 HVAC 4-1 list 5-3 list screen 5-1 MRPC 4-1 name 4-2, 4-4, 5-9 number of 15-5 recalling 5-4 removing 5-7 RFT 4-1 saving 5-3 storage 15-5 storing 4-12 type 4-2, 4-5 Configuration screen 3-7, 4-1 Control Knob 3-4, 3-5 Copy to Internal button 5-6 Copy to Mass Storage button 5-5, 13-4 Copy to Network button 13-4 copyright 1-1 Current Test Plan button 11-3 Cursor Control button 6-16, 7-3 Cursor Size button 6-16 Curve Setup button 7-3 Curve Table button 7-6 Curve Type button 7-3
D
data display 15-4 file 3-8 opening 13-3 setting 13-2 storage capacity 2-1 Data Control button 6-4 Data Files button 13-2 Data Mixing button 7-7 Data Position button 6-16, 7-5
date input 3-7, 9-1, 9-5 Default Summary button 6-9 Delete button 13-3 deleting data files 3-8, 13-3 depth of penetration 8-6 destination, change 11-5 Diagnostics screen 3-7, 10-1 Diff Delta button 6-22 differential filter 2-1 digital recording 15-5 internal hard drive 15-5 shared network drives 15-5 USB mass storage 15-5 directory changing 5-8 copying to mass storage 13-4 copying to the network drive 13-4 deleting 13-4 listing 13-1 opening 13-3 display positioning 15-4 rotation 15-4 scaling see also ZOOM Display Null button 6-7, 6-16 drive 3-7, 3-13, 3-15, 3-17, 3-19, 4-1 adjusting 8-4 selecting 4-2 Drive button 4-10 drive plot 8-1, 8-5, 15-5 graph 3-7 setting 8-5 Drive Plot button 8-5 Drive Select button 4-10 drive setting 4-10 selecting 4-10
E
eddy current testing (ECT) 1-1
Page I-2
Index
Eddynet 15-5 file format 12-4 software 12-4 electronic equipment 1-1, 2-2 Enable Std Run button 6-6 English language, select 3-11 Enter key 3-9 ET Analysis for Windows 15-5
G
general operation 3-13-20 German language, select 3-11 Go To Top button 11-3, 14-30 Graphics Dump button 9-7 graphics dump destination printer options 9-7 selecting 9-7 Greenwich Mean Time (GMT) 12-10
F
ferromagnetic tubing 1-1, 14-17 Field button 4-11 file format 3-8 Eddynet 12-4 PC 12-4 selecting 12-4 File Format button 12-4 Filter Type button 6-21 filters 15-6 12K 4-8 16K 4-8 1st-order differential high-pass 15-6 4K 4-8 8K 4-8 bandpass 2-1, 15-6 combination differential bandpass 15-6 differential 2-1 Find Next Untested button 11-3, 14-31 firmware upgrades 15-6 French language, select 3-11 freq plot graph 3-7, 8-1 frequency 3-13, 3-15, 3-17, 3-19 adjusting 8-5 plot 8-3, 15-5 select 4-2 set 8-6 test 8-6 Frequency Plot (F1) button 8-3 Frequency Select button 4-9 front panel 2-1, 3-4 control 15-4 function keys 3-4
H
Help button 3-4 High Freq Cutoff button 6-22 Home key 3-10 Horizontal Adjust button 6-18 HVAC 3-19, A-1 probe configuration 4-1
I
ID points 7-3 Import Plan button 11-5 Import Test Plan button 11-5, 14-27, 14-29 initial setup 3-2 instrument control front panel control 15-4 Internal button 14-29 internal hard drive 3-14, 3-16, 3-18, 3-20 internal storage 5-6 internal to mass storage, copying 5-5 IP address 3-7, 3-8 set 9-1, 9-6 IP address, network 12-7 IP Field button 12-7 ISO 9001 1-1 Italian language, select 3-11
Page I-3
K
keyboard 3-1 control 15-4 operation 3-93-10 shortcuts 3-10
L
language, selecting 3-7, 3-11, 9-1, 9-3 Left Arrow button 4-4, 5-9 limitation of liability 1-2 Lissajous 2-1 List Dirs button 13-3 list of materials 3-7, 8-1 List Options button 5-3, 5-6
M
Main Menu button 14-31 maintenance 2-2 Manual Tube Info button 6-6 Mark/Unmark button 11-3, 14-31 mass storage 5-5, 13-1, 14-28 setup 12-1 Mass Storage button 5-6 mass storage device 9-7 Mass Storage screen 3-8, 13-1 Mass Storage Setup screen 3-8, 12-1 mass to internal storage, copying 5-6 material list 8-6 Max Rate button 6-19 Max Rate measurement 15-7 Measure button 6-18, 7-4 measurement, signal 15-7 mechanical equipment 1-1 memory 15-5 hard drive 15-5 NVSRAM 15-5 RAM 15-5 USB mass storage 15-5 menu button labels 3-4
Menu Display button 3-4, 3-5, 11-1 message recall 6-8 save 6-8 Message Summary button 6-7 Mix Chan 1 button 7-7 Mix Chan 2 button 7-7 MIZ-28 features 15-5 test system 1-1 Model III Controller 3-10, 6-2, 6-3, 6-10, 6-20 Mount NETDRV button 12-9 mount point 3-8 setting 12-8 Mount Point button 12-8 MRPC 3-15 MRPC probe 3-15, 4-5, A-1 configuration 4-1 multiplexed mode 4-5, 4-8, 4-10, 4-11
N
name, configuration 4-2, 4-4, 5-9 network 3-14, 3-16, 3-18, 3-20, 9-7 drive 12-9 icon 12-9 IP address 12-7 source directory 14-28 Network IP button 12-7 network, copy to 13-2, 13-4 new cal group 14-7, 14-15, 14-26 non-ferromagnetic tubing 1-1 NVRAM 6-20
O
O.E.M. equipment 1-2 OD points 7-3 opening data file 13-3 directory 13-3 test plan 11-4, 14-30
Page I-4
Index
owner 6-9
P
Pacific time 12-10 PC file format 12-4 Peak To Peak button 6-18 Perform Mix button 7-8 Persist Time button 6-4 plant 6-9 position adjust 8-3 data 6-16 power switch 3-4 primary function buttons 3-6 prime frequency 3-7, 8-6 Print button 3-4 printing 3-7, 9-1 probe adaptor 3-2 effect on configuration 3-13, 14-5 probe drive adjustment 15-6 probes 1-2 product description 2-1 pusher icon 6-20
removing configuration 5-7 test plan 11-4 Return button 4-4, 5-9 return policy 1-2 rev jog speed 3-10 reverse speed 3-10 revision date 8-2 RFT 3-17, A-1 RFT probe configuration 4-1 Right Arrow button 4-4, 5-9 RMA number 1-2 root directory 12-8 rotation 3-14 RS-232 connector 15-2 Run Test Plan button 6-13, 14-31
S
sample rate 3-7, 4-1, 4-2, 15-6 adjusting 4-8 Sample Rate button 4-8 Save button 7-9 Save This Form button 6-9 Save This Message button 6-8 scale, adjust 8-3 Scroll button 8-6, 9-7, 11-3, 11-5, 13-2, 13-3, 14-29, 14-30, 14-32 secondary function buttons 3-6 Select button 3-4, 3-5, 4-4, 5-9, 9-7, 11-1, 134 Select Channel button 7-3, 7-6 Select Coil button 8-3, 8-5 Select Filter button 4-8 Select ID Field button 6-6 Select Num Mixes button 7-7, 7-8 Select Point button 7-5 Select Quadrant Channel button 3-10, 6-4, 6-12, 6-16, 6-17, 6-18, 6-21 Select Test Plan button 11-4, 14-30
R
range, adjusting 8-3 raw directory 12-8 Recall Form button 6-10 recall message 6-8 Recall Setup button 7-9 Record To button 12-5 record to destination selecting 12-5 Recording Dir button 12-3 recording directory 3-8 red frequency readout 8-4 remote field testing (RFT) 1-1 remote inputs, setting 9-8 Remove Config button 5-7
Page I-5
selecting coils 8-3, 8-5 drive 4-2 frequency 4-2 record to destination 12-5 serial number 8-2 Set All button 4-10, 6-22 Set All Freqs button 7-5 Set Frequency (F9) button 8-6 Set This Channel button 7-5 Set Threshold button 6-11, 6-20 Set Wall Thickness button 8-6 setting frequency 8-6 IP address 9-1, 9-6 mount point 12-8 network IP address 12-7 remote inputs 9-8 speaker on/off 9-4 wall thickness 8-6 setup initial 3-2 mass storage 12-1 Setup File button 7-9 signal measurement capability 15-7 signal processing circuitry 4-8 signal/noise ratio 3-18 simultaneous injection mode 4-5 testing 4-5 simultaneous mode 4-8, 4-10, 4-11 source, change 11-5 Space button 4-4, 5-9 Span button 6-17 Spanish language, select 3-11 spare parts 1-2 speaker control 3-7, 9-1 speaker on/off 9-4
specifications cleaning 2-2 dimensions 15-1 display 15-1 frequency range 15-1 input/outputs 15-1 language support 15-1 memory 15-1 operating temperature range 15-1 power 15-1 storage temperature range 15-1 test method 15-1 weight 15-1 Speed Check button 6-19 Spool To button 12-6 standard depth of penetration 3-7, 8-1, 8-6 Start Acquire button 6-2, 6-22, 14-31 Start Recording button 6-10, 6-13, 14-31, 14-32 Stop Acquire button 6-10 Stop Acquisition button 6-10, 6-13 Stop Recording button 6-10, 6-13 storing configuration 4-12 strip charts 15-4 Summary Form button 6-9 summary form, recall 6-10 system setup A-1
T
test configurations see configuration test frequencies 3-7, 4-1, 4-5, 4-9 test plan importing 11-5 opening 11-4, 14-30 removing 11-4 selecting 11-4 setup 14-27 test plan file (.csv) 14-27 Test Plans screen 3-8, 11-1 Test/Review screen 3-7 testing mode 3-10
Page I-6
Index
Threshold Delay button 6-11 threshold, tube end 6-20 time input 3-7, 9-1, 9-5 time zone 3-8 setting 12-10 Toggle Field button 6-6 Toggle On or Off button 4-11 Toolbox screen 3-7, 8-1 tube encode field 3-10 Tube End Detection button 6-10, 6-11, 6-20 tube end threshold 6-20 tutorial 14-1 type, configuration 4-2, 4-5
W
wall thickness, set 8-6 warranty 1-1 WEEE 2-2 Write Message button 6-7, 6-8 Write Summary button 6-9
Z
zoom chart 6-16
U
unit 6-9 Unmount NETDR button 12-9 USB 3-14, 3-16, 3-18, 3-20, 9-7 devices hub 15-5 keyboard 6-9, 15-5 mass storage 5-5, 13-1, 15-5 port 15-4, 15-5 printers 15-5 support, see TCP/IP USB key 3-1 USB port 3-9 User Message button 6-8 Utilities screen 3-7, 9-1
V
Vert Max measurement 15-7 Vertical Adjust button 6-17 Vertical Max button 6-18 VGA monitor output 15-4 Voltage Mult button 7-4 Volts P-P measurement 15-7
Page I-7