CALIBRATION REPORT, VICAT DEVICE

Instrument name: ID: Referent norm: Date of calibration: Calibration tools: Timer device: Length device: Weight device: ID*: ID*: ID*: EN196-3

R011/1

VICAMATIC, Automatic setting time device


Final report:

APPROVED

Table 1. Timer inspection: Units: Time (Vicat): Time device: Difference: Range test: s 30 1 15 60 minute 120 180 240 Test passed

30
0 OK

1
0 OK

15
0 OK

60
0 OK

120
0 OK

181
1 OK

240
0 OK

OK

Table 2. Needle inspection:


Measured diameter (mm) Needle diameter (mm): max Required (mm) min Range test:

d1

d2

d3

1.11

1.11
1.09 1.21

1.11
Test pas

OK

OK

OK

OK

Table 3. Needle inspection:


Measured length (mm) Needle length (45mm) Required (mm) Range test: min

47
45 OK

Test pass

OK

Table 4. Needle inspection:


Needle position (central longitudinal) visual description

YES

Yes , needle is in longitudinal center. No , needle is bended. Test passed:

OK

Table 5. Needle inspection:


Measured weight (g) Weight (needle + holder)(g): max Required (g) min Range test: OK 305 OK OK

299.2

297.1
295

298.3
Test pass

OK

Table 6. Mold inspection: Dimension (mm) Mold Dimension (mm) Required (mm) max min

R011/1

L1

L2

L3

69.0
65 85 OK

70.0
65 85 OK

40.0
39.5 40.5 OK Test pass

Range test: Table 7. Software inspection: Needle ZERO position, auto calibration Table 8. Software inspection: Software parameters Check in the last version of LI001 Table 9. Base plate dimension: All base plates used for setting time test has to be same thickness Table 10 Visual inspection:

OK
Test passed:

Yes , needle ZERO position is set

YES
No , needle ZERO position is not set

OK
Test passed:

Yes , parameters are set good

YES
No , parameters are not set good

OK
Test passed:

YES

Yes , all base plates are same thickens No , If not, remove all which are different

OK

YES NO YES YES

Device is clean Any visible damage (if YES, give description in annex) Does the room condition suitable for this device Electricity connector is proper and safe

Final Conclusion:

APPROVED

This device is properly calibrated and is approved to use in laboratory testing.

OK

If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this document)

Team for calibration:


1)

2)

Approved by: Signature:

CALIBRATION REPORT, VICAT DEVICE


Instrument name: ID: Referent norm: Date of calibration: Calibration tools: Timer device: Length device: Weight device: ID*: ID*: ID*: EN196-3

R011/1

VICAMATIC, Automatic setting time device


Final report:

--

Table 1. Timer inspection: Units: Time (Vicat): Time device: Difference: Range test: --------------s 30 1 15 60 minute 120 180 240 Test passed

--

Table 2. Needle inspection:


Measured diameter (mm) Needle diameter (mm): max Required (mm) min Range test:

d1

d2

d3

1.09 1.21 ----

Test pas

--

Table 3. Needle inspection:


Measured length (mm) Needle length (45mm) Test pass Required (mm) Range test: min 45 --

--

Table 4. Needle inspection:


Needle position (central longitudinal) visual description Yes , needle is in longitudinal center. No , needle is bended. Test passed:

--

Table 5. Needle inspection:


Measured weight (g) Weight (needle + holder)(g): max Required (g) min Range test: -305 --295 Test pass

--

Table 6. Mold inspection: Dimension (mm) Mold Dimension (mm) Required (mm) max min

R011/1

L1

L2

L3
Test pass

65 85 --

65 85 --

39.5 40.5 --

Range test: Table 7. Software inspection: Needle ZERO position, auto calibration Table 8. Software inspection: Software parameters Check in the last version of LI001 Table 9. Base plate dimension: All base plates used for setting time test has to be same thickness Table 10 Visual inspection: Device is clean

OK
Test passed:

Yes , needle ZERO position is set No , needle ZERO position is not set

-Test passed:

Yes , parameters are set good No , parameters are not set good

-Test passed:

Yes , all base plates are same thickens No , If not, remove all which are different

--

Any visible damage (if YES, give description in annex) Does the room condition suitable for this device Electricity connector is proper and safe

Final Conclusion:

--------

This device is properly calibrated and is approved to use in laboratory testing.

--

If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this document)

Team for calibration:

1)

2)

Approved by: Signature:

VICAMATIC, Automatic setting time device


Calibration description (EN 196-3)

R011/1

Table 5 Weight Needle and holder together 295-305(g)

Use L key for needle ZERO calibration

Use L key for needle replacement only

Table 7

Table 9
Table 2

Glass base plate

d1 d2 d3

Table 3 L = 45mm

Table 4

Table 6 L1 = 65-85mm

L3 = 40.0 2mm

L2 = 65-85mm NOT OK