Professional Documents
Culture Documents
Abstract
Sand screen selection relies on accurate particle size information for the sands that need to be controlled. Within the oil
industry there is awareness of the differences between dry sieve analysis and laser particle size analysis (LPSA), but there can
also be wide variations between LPSA performed in different laboratories even when such variables as sample preparation
are taken into consideration. Recent work by the authors has shown that such variables as sonication time and amplitude can
have a significant effect on certain sands, and there is evidence to indicate that the sensitivity to sonication is dependent on
the mineralogy of the sand. Other factors have also been examined to address the widespread view within the oil industry that
LPSA exaggerates the fines content, whereas the opposite view is held in sedimentology and soil science in that the coarse
fraction is believed to dominate. Work has been performed to examine which view might be most relevant for reservoir
sands, and guidelines for performing LPSA are proposed.
Introduction
The particle size of reservoir rocks is routinely measured for a number of reasons, including sand control selection, and as
such is probably considered a straightforward process. Most of the literature concerning grainsize analysis of geological
materials comes from the fields of sedimentology and soil science. Historically, particle size was measured by a mixture of
dry sieving (for grain sizes above 45 microns), and sedimentation (for the less the 45 micron fraction). Sedimentation
measures the settling velocity of particles through fluids to infer particle size. However, the oil industry dry sieving without
any further analysis for the less than 45 micron fraction was most often utilised. More recently laser light scattering (LPSA)
has become the preferred method of particle sizing, since it is quick to perform and covers the entire size range in a single
measurement.
All methods of particle sizing have their disadvantages and the two methods used in the oil industry do give different results,
in part due to particle shape and also degree of dispersion. Geological materials have wide grain size distributions and are
typically composed of irregular shaped particles, and so the particle size obtained during analysis depends on which aspect of
the particle is measured. Different methods of analysis measure different aspects of size and so would not be expected to give
the same results. Dry sieving generally only measures down to 45 micron, and has a lower resolution than LPSA. There is
also a common perception within the oil industry that laser light scattering exaggerates the fines in a distribution. Sampling
may also be an issue, especially for LPSA which typically requires less than 1 gram of sand for each measurement.
This paper explains the inherent difficulties of particle sizing and summarises the expected differences between dry sieving
and LPSA. However, the authors have been aware for some time that there can be a variation in the laser particle size data
measured on the same sand in different laboratories which cannot be totally attributed to sampling issues. The work described
here concerns a reservoir sand which gave very variable LPSA results. Laser particle size measurements under various
conditions on two different instruments provided insights into the causes. Further detailed work on other reservoir sands has
identified analysis procedures commonly used in the field of soil science which are not appropriate to the measurement of
reservoir rocks for the purposes of identifying optimum sand control options. This paper describes commonly-used LPSA
procedures and their potential effect on the measured grain size distributions. In the light of these results a method believed
more appropriate to sand control work is suggested.
SPE 165119
Sphere
of
same
maximum
diameter
Sphere
of
same
Sedimentation
rate
Sphere
of
same
minimum
diameter
Sphere
of
Same
volume
(LPSA)
Sphere
of
passing
same
sieve
opening
Figure 1. Illustration of different equivalent spheres for one irregular shaped particle (from Malvern4)
Measurement Techniques: Dry sieving
Dry sieving was traditionally used to measure particle size in the oil industry before the widespread use of LPSA. Dry sieving
works well for coarse sands with little fine material.
The sample is placed on the top sieve of a stack of sieves arranged in order of decreasing aperture size. Usually the sieve
stack is placed on a mechanical sieve shaker for a set time. The contents of each sieve are then weighed. However, samples
with high fines contents often blind off the finer sieves and require manual brushing to force them through. This aspect
introduces an operator-dependent feature to the process. Dry sieving is a dusty, rather time-consuming, process so the results
often depend on the diligence of the operator.
An alternative to a mechanical sieve shaker is the sonic sifter is an instrument which potentially provides a more efficient
method of dry sieving. The particles are lifted and forcibly dropped in a column of oscillating air at a frequency of thousands
of cycles per minute, which encourages better dispersion and helps prevent the sieves blinding off. Sonic sifters are able to
handle much finer dry powders than standard sieving, and a smaller sample is required (around 10g). But with samples
containing significant fines it is noticeable that fines will coat the sides of the sieves after use.
Measurement techniques: Laser Particle Sizing
Laser diffraction is more properly called laser light scattering and works on the principle that particles scatter light in all
directions with an intensity pattern that is dependent on the particle size. The angle of scatter is inversely proportional to
particle size, and the intensity of the diffracted beam at any angle is a measure of the number of particles with a specific cross
sectional area in the beams path2.
A sample dispersed in a suitable fluid, usually water, is passed through a beam from a monochromatic light source, usually a
laser. The light scattered by the particles at various angles is focused by the Fourier lens onto a specific spot irrespective of
the particles position or velocity. Therefore a single, composite diffraction pattern is formed containing a contribution from
all the particles in the measurement cell. The diffraction pattern is measured by multi element detectors placed in appropriate
positions. Numerical values relating to the scattering pattern are recorded and then transformed using an appropriate optical
model and mathematical procedure into a volumetric size distribution3 related to grain size by assuming the particles to be
spheres.
SPE 165119
Sonic
probe
sample
reservoir
detector
Projection
lens
Fourier
lens
laser
measurement
cell
pump
SPE 165119
Optical model: The scattering model that is generally used is Mie theory. Mie theory predicts both the intensity of light
scattered from the particle surface, as well as the behaviour of light which is transmitted and refracted through the particle. In
order to use this model correctly the complex refractive index of the particle is required. This complex refractive index
comprises a real and imaginary part. The real part (commonly called the refractive index of the material) predicts the
diffraction from the particle surface, and can be easily measured. The imaginary refractive index accounts for the attenuation
of light as it passes through the particle, and is commonly referred to as the absorption of the material4. For the purposes of
laser diffraction, the absorption of a material only has to be specified to an accuracy of an order of magnitude; instruments
give options from .001 to 1. Absorption depends on the transparency of the particles and also surface roughness, and can be
estimated by comparing the volume of particles calculated by the instrument to the real volume of particles using different
absorption values. The absorption value giving the closet correlation will be the correct value (Malvern application note ref
4). Generally for geological materials absorption values of 0.1 are used. The scattering pattern can be re-analyzed with
different optical models without having to re-run the sample, making it straightforward to examine the sensitivity of the
distribution to various optical parameters.
Commonly used LPSA procedures for measuring geological materials
The procedures used for LPSA measurement of geological samples have often originated in soil science or core analysis,
since that is generally the background of the commercial labs offering such a service. Particle size is used to characterise and
classify soils and requires that clays are dispersed as much as possible. So the methods employed tend to be fairly aggressive.
A normal soil science technique could comprise the following steps:
However, the authors laboratory came to PSD analysis purely for sand control purposes, and adopted the following method
for LPSA measurements
This technique gave 4 measurements for each sand, and was intended to identify any non-reproducibility in the measurement.
SPE 165119
30
25
75-125um ballotini
volume %
20
75-125um sand
212-300um ballotini
15
212-300um sand
500-710um ballotini
500-710um sand
10
10
100
Diameter (microns)
1000
Another difference between sieve analysis and LPSA is related to sample dispersion. Where sands contain significant fines
the small particles become attached to the coarser grains by electrostatic forces, and therefore do not pass properly down
through a sieve stack. However, in an LPSA measurement the sample is dispersed in water, which breaks down these forces
allowing the fines to be detected and resulting in LPSA measuring more fines than dry sieving. These two effects (sample
dispersion and grain shape) mean that one would expect a bigger D10 and smaller D90 from an LPSA measurement
compared to the dry sieve analysis of the same sand; Figure 4 gives typical examples.
100
90
80
Cumulative %
70
60
depth A: LPSA
50
depth A: sieve
40
depth B: LPSA
30
depth B: sieve
20
10
0
1000
100
diameter (microns)
10
SPE 165119
Blott and Pye6 reasoned that the results are explained by greater noise in the diffraction data collected at low angles (that is
from coarse particles) compared with high angles (fine particles). Therefore there is potential for over-estimation of the
proportion of coarse particles.
In practice, these extreme bimodal distributions are not very representative of real geological materials, and so some other
mixtures were prepared to give a wider, continuous distribution more similar to a poorly sorted sand sample. Figure 6 shows
the comparison between sieve analysis and LPSA for four mixes with different amounts of fines present. The results show
good agreement between sieve and LPSA for low fines contents. With high fines the LPSA measurement becomes more
variable, but still with reasonable agreement with the sieve measurement.
100
90
80
Cumulative %
70
60
50
40
45um-63um + 250um-355um
30
45um-63um + 710um-840um
20
45um-63um + 1000um-1180um
10
0
1000
100
Diameter (microns)
10
Mix 5 LPSA
cumulative %
60
Mix 5 sieve
50
Mix 1 LPSA
40
Mix 1 sieve
30
Mix 4 LPSA
Mix 4 sieve
20
Mix 6 LPSA
10
0
Mix 6 sieve
1000
100
diameter (microns)
10
SPE 165119
measurement. Bi-modal distributions are difficult to measure and the authors were less concerned about the apparent
variation between Lab A and Lab B and more concerned that our measurement was not bi-modal as well. Varying the
instrument settings did not reproduce the commercial labs bi-modal distributions; modifying the refractive indices in the Mie
optical model did not greatly affect the shape of the distribution, and reducing the pump speed simply reduced the coarse end
of the distribution (Figure 8). Also increasing the sonication strength and time did not create the bimodal distributions seen in
the other lab measurements .
The instrument used by the authors is from a different manufacturer to the instruments used by both Lab A and Lab B. We
will call the instrument used by ourselves instrument C and that used by Labs A and B instrument AB. Since changing the
settings in instrument C did not reproduce the measurements made in Labs A and B some tests were done on an instrument
made by the same manufacturer as instrument AB. The same basic principles and algorithms are used by all manufacturers,
and so the purpose was to examine whether different instrument components (pump, sonicator, etc) could affect the
distribution obtained or whether some aspect of sample preparation influenced results. This alternative instrument was an old
model which only measured up to 300 microns, and so the sample was passed through a 250 micron sieve and the <250
micron fraction was analysed. The sonication on instrument AB caused the sample to become bimodal, this sample was
collected from the instrument and re-measured in instrument C and again gave a bi-modal distribution, Figure 9.
Measurements with no sonication gave good agreement between both instruments (Figure 9). These results showed that
there is little difference in PSDs measured on a sample in the same state, and so it was inferred that the sonication strength is
greater on instrument AB. To confirm this, in a subsequent test a bi-modal distribution was also obtained by placing the
sample in an ultrasonic bath in our own lab (rather than using the sonicator built into the instrument) and then measuring it in
instrument C; this measurement is also shown in Figure 9. Further work with maximum sonication strength of this sieve
fraction on our own instrument (C) was still not powerful enough to create a bi-modal distribution (Figure 10)
100
90
80
Cumulative %
70
60
Lab A
50
40
Lab B
30
Lab C
20
10
0
1000
100
Diameter (microns)
10
Cumulative %
70
60
Lab A
50
Lab B
40
30
20
10
1000
100
Diameter (microns)
10
Figure 8. Effect of instrument pump speed on size distribution measured for sand X
SPE 165119
100
90
80
Cumulative %
70
60
50
instrument AB - no sonication
40
instrument C - no sonication
30
20
10
0
1000
100
Diameter (microns)
10
Figure 9. Size distribution measured for <250um fraction of sand X on instrument AB and instrument C
100
90
80
Cumulative %
70
60
50
instrument AB - no sonication
40
instrument C - no sonication
30
20
10
0
100
Diameter (microns)
10
Figure 10. Effect of instrument sonication on size distribution measured for sand X
Obtaining a bi-modal distribution suggests that grains of a particular size are affected by the sonication, since a narrow range
of grain sizes (between approximately 70 micron and 140 micron) are absent from the measurement. To further examine the
effect of sonication the sample was divided into sieve fractions by wet sieving (to minimise fines retention). Each sieve
fraction was sonicated in the ultrasonic bath and then examined under a microscope. Figure 11 shows the results for the 6390 micron sieve fraction for this sand. In the non-sonicated sieve fraction there are many brown grains which when sonicated
have been broken down. Further sonication turns these grains into a fine powder. Therefore sonication was able to break
down specific grains in the sand, but the sonicator in instrument C was not powerful enough to produce the effect.
No sonication
1 min sonication
Figure 11. Effect of increasing sonication time on sand X 63-90 micron fraction
30 min sonication
SPE 165119
In addition to sonication Lab A also used SHMP during the measurement process. As mentioned previously, this is
commonly used as a dispersant and works by exchanging divalent ions such as calcium for sodium on mineral surfaces,
specifically clays. Figure 13 shows that after treatment with SHMP the distribution is bimodal. XRD analysis indicated this
sample contains a relatively high proportion of mixed layer clays which will be very sensitive to this type of treatment. As in
the sonication tests individual sieve fraction were treated with SHMP, and it was found by XRD that most of the grains in the
63 90 micron size range are brown clay aggregates which are dispersed by the SHMP. Figure 12 shows the 63 90 micron
sieve fraction after treatment with SHMP. Overall, the delicate grains (which were believed to be mixed-layer clays)
dispersed by SHMP and broken down by sonication were very prevalent from 45 micron to 180 micron in size. The effective
removal of these grains by both sonication (Lab A uses sonication until there is no change in the distribution) and SHMP
creates the bi-modal distributions.
It was possible that the clay aggregates could have been created as a result of the cleaning and drying process during the
sample preparation which baked the clay into aggregates. If this were the case then the aggregates would be expected to
separate in other aqueous fluids. Some measurements after soaking in synthetic formation water were performed; the result is
shown in Figure 13 and indicates that dispersion in pore water (as would occur in the field) has a less pronounced effect than
SHMP.
Cumulative %
70
60
50
40
Lab C - no sonication
30
LAB A
20
10
0
100
10
Diameter (microns)
10
SPE 165119
therefore the effect of sonication and SHMP was investigated on a range of reservoir sands to see how significant their
influence is on more typical sands.
The authors standard measurement routine was to pre-wet the sand sample with methanol before adding it to the sample
reservoir. Sonication was used during the loading of the sample to speed up dispersion, and also for one minute before
measuring the sample. For the study described below the samples were pre-soaked in demineralised water overnight to aid
dispersion without the use of sonication. To do this the amount of sample required to give the correct obscuration for the
instrument is required so that the whole soaked sample can be used in the measurement.
1.
The effect of sonication
Figures 14 to 17 show the effect of progressive sonication on four sands. The sample was introduced into the instrument and
sonicated for 1 minute before each measurement. A total of 20 measurements were made on the same sample. As a control, a
duplicate sample was measured without sonication but including 1 minute wait between each measurement to give both
samples the same time in the instrument to take into account any effect of degradation by the pump.
1000
1000
100
100
d10: no sonication
d10: no sonication
d50: no sonication
d50: no sonication
d90: no sonication
d90 no sonication
d10: sonication
d10: sonicated
d50: sonicated
d50: sonication
d90: sonication
10
10
15
20
25
30
35
40
45
d90: sonicated
10
10
15
20
25
30
35
40
45
1000
100
100
d10: no sonication
d10: no sonication
d50: no sonication
d50: no sonication
d90: no sonication
d90: no sonication
d10: sonicated
d10: sonicated
d50: sonicated
d50: sonicated
d90: sonicated
10
10
15
20
25
30
35
40
45
d90: sonicated
10
10
15
20
25
30
35
40
Figures 14 to 17 show that soley circulating the sample through the system tends to increase the fines content and sonication
further increases this effect. Generally the d50 and D10 are relatively unaffected; only in one instance was the whole
distribution made finer. Because the finer fractions are preferentially affected compared to the coarser grains this can lead to
significant changes in the uniformity with relatively small changes to the D90 of the sand. Table 1 shows how the changes
alter the uniformity coefficient for the North Sea sand S990. Based on Weatherfords screen selection criteria the appropriate
sand control changes from wire wrap screens to metal mesh screens over the measurement time. Tables 2 to 4 show the
results for the other sands tested. The effect of sonication may well be more pronounced with other instruments if they have
more powerful sonic probes.
SPE 165119
11
Table 1. The effect of circulation and sonication on the calculated PSD parameters for S990
Time
(mins)
D90
D50
76
232
10
68
20
62
30
40
No sonication
D10
Uc
<45 m
D90
514
3.55
6.4
58
219
529
4.44
8.2
232
566
3.94
7.4
44
207
571
5.51
10.1
229
566
4.34
8.2
38
205
578
6.37
11
59
226
556
4.52
8.6
35
205
581
7.03
11.5
55
225
543
4.79
9.0
32
204
580
7.62
12
<45 m
Table 2. The effect of circulation and sonication on the calculated PSD parameters for S655
No sonication
Time
(mins)
d90:
d50:
d10:
Uc
<45 m
d90:
d50:
d10:
Uc
45 m
90
383
848
4.8
8.6
61
350
791
6.5
8.6
10
63
382
873
6.9
10.3
41
349
849
9.6
10.3
20
40
377
855
10.6
11.6
28
346
825
14.0
11.6
30
31
378
872
13.9
13.1
19
340
778
20.5
13.1
40
26
378
881
16.5
13.7
16
339
771
24.5
13.7
Table 3. The effect of circulation and sonication on the calculated PSD parameters for S77
No sonication
Time
(mins)
d90:
d50:
d10:
Uc
<45 m
d90:
d50:
d10:
Uc
45 m
12
125
282
12.6
25
15
134
333
10.9
21.2
10
12
127
327
13.1
25.1
12
113
288
12.1
26.4
20
12
125
316
13.4
25.8
10
98
246
12.3
29.9
30
11
122
309
13.8
26.7
10
93
238
12.4
31.5
40
11
120
304
14.1
27.2
89
231
12.6
32.8
Table 4. The effect of circulation and sonication on the calculated PSD parameters for S1472
No sonication
Time
(mins)
d90:
d50:
d10:
Uc
<45 m
d90:
d50:
d10:
Uc
45 m
27
279
681
12.1
11.3
42
286
771
7.9
10.2
10
29
290
888
11.9
11
38
294
847
9.1
10.4
20
22
293
911
15.7
11.5
33
296
890
10.5
10.7
30
19
292
845
18.4
11.9
34
299
838
10.3
10.6
40
17
291
810
20.2
12
29
293
757
11.7
11
Both sonication and circulation though the system reduce the sand D90, apart from S1472, where it increases. This is an
artefact of the shape of the distribution, Figure 18 shows that the effect of increasing the amount of very fine material shifts
the D90 upwards in this sample.
12
SPE 165119
100
90
80
cumulative %
70
60
50
40
no sonication
30
20
sonicated
10
0
1000
100
diameter (microns)
10
Figure 18. The effect of circulation and sonication on the S1472 sand
100
100
90
90
80
80
70
70
60
cumulative %
cumulative %
To investigate whether the effect of sonication is due to dispersion or breaking up of grains it is useful to examine sieve
fractions. Figures 19 to 22 show the LPSA analysis on sieve fractions obtained by a process of wet and dry sieving on sand
S1472 (The effect of sonication on this sand is shown in Figure 16). The material was wet sieved through a 45 micron sieve
to remove all fines then dry sieved into fractions. Figures 19 to 22 show a progressive increase in fines in the 45 and 63
micron sieve fractions with sonication, this is less apparent for the 90 micron sieve fraction and the 180 micron fraction does
not change.
no sonication
no sonication
50
1 min sonication
40
2 min sonication
60
no sonication
no sonication
50
1 min sonication
40
2 min sonication
30
3 min sonication
30
22 min sonication
20
23 min sonication
20
32 min sonication
33 min sonication
10
0
1000
100
10
42 min sonication
10
0
1000
100
100
100
90
90
80
80
70
70
60
no sonication
50
no sonication
40
1 min sonication
2 min sonication
30
20
10
diameter (microns)
no sonication
50
no sonication
40
1 min sonication
2 min sonication
12 min sonication
20
32 min sonication
10
60
30
22 min sonication
100
cumulative %
cumulative %
0
1000
10
diameter (microns)
diameter (microns)
22 min sonication
10
1
0
1000
100
10
diameter (microns)
To ensure these changes are attributable solely to sonication a control was performed in a similar way as the sonication tests
SPE 165119
13
shown in Figures 14-17. A sieve fraction was measured 20 times without sonication then the same sample measured a further
20 times with 1 minute sonication before each measurement. The final measurement would therefore have had a total of 20
minutes sonication. Figures 23, 25 and 27 show the cumulative particle size distribution from these measurements. In all
cases sonication increases the fines compared to merely pumping the sample around the measurement system.
To find out if the sonication is aiding dispersion or breaking the grains the sonication probe was removed from the instrument
and set up in a system to mimic the sample vessel of the instrument, and so give equivalent sonication. Figures 24, 26 and 28
show sieve fractions of each sand before sonication and after sonication for 1 minute and 5 minutes. It is apparent that there
are more fines with the sonicated samples and grains have broken. The degree of fines production in the photomicrographs
reflects the LPSA measurements, in that for example in sample S1472 there are less fines and this sample shows least affect
of sonication.
100
90
80
cumulative %
70
60
no sonication
50
40
sonicated
30
20
10
0
1000
100
10
diameter (microns)
Figure 23. S1472 63micron sieve fraction measured 20 times without sonication, then the same sample sonicated for 1 min
before 20 further measurements.
no sonication
1 min sonication
2 min sonication
Figure 24. Photomicrographs of the 63-90 micron sieve fraction of sand S1472, with different amounts of sonication.
14
SPE 165119
100
90
80
cumulative %
70
60
50
no sonication
40
30
sonicated
20
10
0
1000
100
10
diameter (microns)
Figure 25. S655 125-180 micron sieve fraction, measured 20 times without sonication, then the same sample sonicated for 1
min before each of 20 further measurements.
no sonication
1 min sonication
5 min sonication
Figure 26. Photomicrographs of the 125-180 micron sieve fraction of sand S655, with different amounts of sonication.
100
90
80
cumulative %
70
60
no sonication
50
40
30
sonicated
20
10
0
1000
100
10
diameter (microns)
Figure 27. S77 90-125 micron sieve fraction measured 20 times without sonication, then the same sample sonicated for 1
min before each of 20 further measurements.
SPE 165119
15
no sonication
1 min sonication
5 min sonication
Figure 28. Photomicrographs of the 90-125 sieve fraction of sand S77, with different amounts of sonication.
It should be noted that the shape of the particle size distribution curves means that differences that look small when the data
is plotted can become quite significant when translated into parameters such as the uniformity coeffient or sorting coefficient.
Where the distribution intersects the 90% gridline at a shallow angle, small changes in the visual appearance of comparison
plots can actually be quite large numerically. Where samples are bi-modal (or become bi-modal during measurement) this
effect can be very significant if the data is used to calculate parameters used in criteria for sand screen selection.
2.
Effect of SHMP
The same samples used to examine the effect of sonication were also treated with SHMP to determine whether it affects all
sands. The samples were soaked in SHMP overnight then measured without sonication. Table 5 summarises the results, they
indicate that for the sands tested SHMP has very little impact on the particle size.
Table 5. Summary of measurements made after the sample was soaked in SHMP and demin water
SHMP
B77
Demineralised water
d90:
11.7
d50:
118
d10:
247
<45 m
26.6
d90:
13.5
d50:
125
d10:
294
45 m
24.4
S655
72
359
797
8.4
69
373
813
8.5
S1472
36
259
620
7.8
27
279
681
11.3
S990
64
236
529
7.8
76
232
514
6.4
16
SPE 165119
100
90
80
Cumulative %
70
60
50
40
30
20
external lab
10
0
1000
100
Diameter (microns)
10
Figure 29. S77 measured on different instruments with an additional measurement from an external lab with a different
procedure.
100
90
80
Cumulative %
70
60
50
40
Instrument AB - demin soak, no sonication
30
20
10
0
1000
100
Diameter (microns)
10
100
90
80
Cumulative %
70
60
instrument AB - demin soak, no sonication
50
40
30
external lab measurement (30 min soniation)
20
10
1000
100
Diameter (microns)
10
SPE 165119
17
100
90
80
Cumulative %
70
60
Instrument C - no sonication
50
Insturment AB - no sonication
40
30
External Lab
20
10
1000
100
Diameter (microns)
10
This approach may, impact on the sample dispersion achieved, and so some steps can be taken to aid measurement. The
sample can be pre-wetted with methanol; this is useful especially with reservoir sands which may be intermediately wetting
even after cleaning. Samples for measurement could also be pre-soaked (for 2 to 3 hours or overnight in water) to ensure they
are fully wetted. Gentle agitation in a shaker bath could also assist wetting/dispersion. The difficulty with this approach is
that the amount of sample prepared must be correct for the measurement since all the sample needs to be used to ensure
reproducibility. This problem can be addressed by preparing multiple smaller samples and adding them until the correct
sample loading is achieved. For the actual measurement, it is beneficial to measure the sample initially without sonication,
18
SPE 165119
and then repeat the measurement with sonication. This allows the sonication effects to be observed for instance an increase
in grain size resulting from agglomeration.
Even with careful sample preparation and measurement there will still be some natural variation in the particle size
distribution obtained, and different laboratories will still have some differences in results. These variations tend to occur at
the fine end of the distribution, which can significantly impact on sand screen selection when it is based purely on some
criteria derived from size parameters. Differences in size distributions which look small when plotted can be significant in
terms of selection criteria. For example, a change in D90 from 10 micron to 5 micron may not look large in a plot of data,
but would double the uniformity coefficient of the distribution and could increase the measured fines content. Either of these
effects could change the preferred sand control from stand-alone screens to gravel packing. Since it is difficult to accurately
measure the complete size distribution, and in any case the exact state of the sand downhole is generally unknown, it may be
beneficial to move away from screen selection based only on criteria that rely on the fine end of a distribution, which is the
hardest to measure.
Conclusions
Based on the work described here, the following conclusions can be drawn:
In general, different instruments give similar results if ultrasonic dispersion is not used.
In itself LPSA does not exaggerate fines content (<45 mcron fraction), however, some of the procedures associated
with laser particle sizing can increase the fines content.
o Sonication can break up grains, and
o The use of water softeners such as SHMP can have a dramatic affect on samples containing significant
amounts of mixed layer clays.
F. Storti and F. Balsamo - Particle size distributions by laser diffraction: sensitivity of granular matter strength to
analytical operating procedures. Solid Earth, 1, 2548, 2010 19 April 2010
Eshel, G; Levy, G, J; Mingelgrin, U and Singer, M,J. - Critical Evaluation of the Use of Laser Diffraction for
Particle-Size Distribution Analysis. Soil Sci. Soc. Am. J 68 (736-743) 2004
ISO 13320
Malvern Mastersizer 2000 application note: Determination of the particle absorption for laser diffraction size
calculations. MRK1308-01
Blott, Simon J; Croft, Debra, J; Pye, Kenneth; Saye, Samantha, E and Wilson, Helen, E. Forensic Geoscience:
Principles, Techniques and Applications, ed Pye K and Croft D J, Geological Society of London , Special
SPE 165119
6.
7.
19