Set Point
60F
(15.6C)
-2F
(-18.9C)
140F
(60C)
Tolerance
.9F
0.5C
2F
1.1C
2F
1.1C
Time
90 minutes
90 minutes
90 minutes
Sensitech Inc. 800 Cummings Center Suite 258X Beverly, MA 01915-6197 USA: 1-800-843-8367 Intl: +1-978-927-7033 Fax: 978-921-2112
The chamber is allowed to stabilize at each set point and it remains at that temperature for the
specified time. The monitors record data throughout the cycle. Also recording data during the
temperature cycle is a 3rd party certified NIST traceable precision RTD temperature probe and data
logger. This NIST traceable probe is certified to be 10x greater in accuracy than the monitors being
validated. The data recorded by the monitor is compared to the data recorded by the data logger to
determine the compliance of the monitor. These tests are performed to validate that the monitor's
circuitry performs as designed.
Proprietary testing software is used to validate that the monitor operated properly throughout the
functional test and environmental validation. The software test includes a verification of memory read
and write, microprocessor functionality, temperature measurement, and time clock functionality. Each
monitor is shipped with a Certificate of Validation.
Additionally Sensitech Inc. warrants the validation period for TempTale electronic temperature and
humidity monitors against defects in materials or workmanship for a period of one (1) year from the