Professional Documents
Culture Documents
skin distance (SSD), a 10cm x10cm2 open field, and a water phantom aligned to the isocenter.
The collimator rotated from 90 degrees to 270 degrees with gantry in the initial position. An ion
chamber is placed in the water phantom 5 cm deep to the isocenter. 100 monitor unit (MU) was
delivered at 400 MU/min dose rate.
Wedge Factor Measurement: The calculation point is at the ion chamber reading point. The
wedge factor was calculated by using equation below.1
WF=
100 cGy dose was delivered three times at each of three conditions: without wedge at 90 degrees
of collimator, with wedge at 90 degrees and 270 degrees of collimators. Dose outputs were
measured at the same point of the central axis of the beam at the same depth. Three sets of data
from each condition collected from the ion chamber reader outside of the treatment room.
Average of the readings was used in calculation of wedge transmission factor.
Table 1: Beam output with and without wedge for 6 MV beam at 5 cm depth, 10x10cm2 field
size
Reading
#1
Reading
18.360
12.841
12.891
#2
Reading
18.359
12.843
12.889
#3
Average
(18.362+18.360+18.359
(12.843+12.841+12.843)/3 (12.888+12.891+12.889)/3
readings
)/3=18.360
=12.842
=12.889
The dose rate with wedge was taken from an average reading with 15-degree physical wedge at
90-degree and 270-degreee of the collimator.
(12.842+12.889)nC/2=12.866 (nC)
Applying measurement to the transmission factor calculation formula:
WF=
Discussion: The steel wedge factor for a 6 MV photon to a 10 x 10 cm2 and a fixed 100 cm SSD
treatment field at 5 cm depth is 0.701. Because the physical wedge has not been used for so long
time at our department, the physicist I worked with could not find the wedge factor reference
table for this Varian 21 EX treatment machine. However, we checked two other similar treatment
machines: Varian Clinac 600CD and Varian Clinac 2300 CD, with reference wedge factors 0.701
and 0.787 respectively. By comparing, we think the wedge factor we measured was consistent to
the wedge factors measured on the other same type of treatment machines. We were satisfied
with our experiment.
In my experiment, the wedge was aligned to perpendicular to the ion chamber and was
fixed centrally at the central axis of the beam. Wedge factors were measured at two positions
with 90-degree apart, which further confirmed the central alignment and right measurement point
of the wedge. Therefore, the reading for the measurement is maintained consistently. The
numbers used in final calculation were taken from average readings in each measurement. All
aimed to improve the accuracy of the experiment with more precise results.
My experiment agreed with the attenuation principle of a physical wedge that the beam
output was found to decrease with wedge in the path in comparing with beam output in an open
field.2 Table 1 represents this finding with all the measurement. Therefore, when a wedge is used
in radiotherapy, the dose calculation must be corrected with wedge factor.
Clinical application: In treatment of neck, head, or breast cancers with uneven body thickness,
wedges can be used to avoid too high dose section or hot spot in the thinner part of the body.
With the help of my preceptor Matt, we selected an example of thoracic spine treatment with an
AP beam. Due to the different thickness of the patients chest and the anatomy curve of the spine,
the dose will be unnecessary hotter near to the neck part. By adding a 15-degree wedge with
the heel portion towards the neck perpendicular with the beam, the hot spot was pulled off the
margin of the body, which creased a desirable even dose distribution in the body with the
isocenter at the target. As the treatment machine was calibrated at 1 cGy/MU, the output factor
was calculated by looking up in the effective field size table and percentage depth dose for 6 MV
beam table. With all other factors remained consistently, we then calculated the MU of each
treatment plan with and without a wedge. Figure 1 attached below lists the detail of the hand
calculation.
From the chart, MU without wedge was 348 Mu and 496 MU with a wedge. I listed the
formal calculation here.1
MU no wedge =
prescribed dose
300 cGy
=
=348 MU
cGy
1 cGy
%DDoutput factor
0.86730.9952
Mu
1 MU
MU with wedge =
prescribed dose
cGy
%DDoutput factorWF
Mu
300 cGy
1 cGy
0.86730.99520.701
1 MU
=496 MU
The deference between open field and with a wedge can be calculated as below.
%diff =
If the wedge transmission factor was missed in the dose calculation with wedge in place, the MU
setting would be 29.8% less than desired dose. The target would receive less dose, and the cancer
will be undertreated.
Conclusion: My experiment represents an individual physical wedge has a transmission
factor in the path of treatment beam, and it decreased the beam output by comparing with the
MU settings without a wedge. Along with the slop of the wedge, beam intensity decreased
progressively across the field. We created an ideal treatment field with the isocenter at target
volume by adding a 15-degree wedge in my clinical example. However, one more wedges and/or
greater ankles of wedges could be used in a treatment depending on the treatment sites in order
to produce desirable dose distribution. We tried to keep all other variable factors the same in my
experiment to get a consistent measurement of wedge factor. We calculated the 15-dgree wedge
affected the monitor units of the beam by 29.8% in our example. This is a significant impact of
radiation dose. This finding from my experiment is also applied to all other angles of wedges
with associated wedge factors.2
References
1. Bentel, GC. Radiation Therapy Planning. 2nd ed. Colombia: McGraw-Hill; 1996: 49-53,
137.
2. Khan FM. The Physics of Radiation Therapy. 4th ed. Philadelphia, PA: Lippincott
Williams and Wilkins; 2010: 181-185.