Professional Documents
Culture Documents
Daily
Monthly
Annually
Procedure
Dosimetry
X-ray output constancy
e- output constancy
Mechanical
Localizing lasers
Collimator Size Indicator
Safety
Door Closing Safely
Door Interlock
Audiovisual monitor
Dosimetry
X-ray output constancy
e- output constancy
Backup monitor constancy
X-ray central axis dosimetry parameter (PDD, TAR) constancy
e- central axis dosimetry parameter constancy (PDD)
X-ray beam flatness constancy
e- beam flatness constancy
X-ray and e- symmetry
Safety Interlocks
Emergency off switches
Wedge, e-cone interlocks
Mechanical checks
Light/radiation field coincidence
Gantry/collimator angle indicators
Wedge position
Tray position
Applicator position
Field size indicators
Cross-hair centering
Treatment couch position indicators
Latching of wedges, blocking tray
Jaw symmetry
Localizing lasers
Field light intensity
Dosimetry
X-ray/e- output calibration constancy
Field size dependence of x-ray output constancy
Output fator constancy for e- applicators
Annually
Central axis parameter constancy (PDD, TAR)
Off-axis factor constancy
Transmission factor constancy for all treatment accessories
Wedge transmission factor constancy
Monitor chamber linearity
X-ray output constancy versus gantry angle
e- output constancy versus gantry angle
Off-axis factor constancy versus gantry angle
Arc mode
Safety interlocks
Follow manufactuer's test procedures
Mechanical checks
Collimator rotation isocenter
Gantry rotation isocenter
Couch rotation isocenter
Coincidence of collimator, gantry, and couch axes with isocenter
Coincidence of radiation and mechanical isocenter
Vertical travel of table
Respiratory Gating
Beam energy constancy
Temporal accuracy of phase/amplitude
Calibration of surrogate phase/amplitude
Interlock testing
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010
2. Klein EE Hanley J, Bayouth J, et al. Task Group 142 report: Quality Assurance of Medical Accel
2009;36(9): 4197-4212. http://dx.doi.org/10.1118/1.3190392
Non IMRT
Tolerance (+/-)
IMRT
SRS/SBRT
3%
3%
2mm
2mm
1.5 mm
2mm
1 mm
1mm
Functional
Functional
Functional
(+/-) 2mm
2mm or 1% on side
1 degree
2mm
2mm
2mm
2mm
2mm diameter
2mm/1 degree
Functional
2mm
(+/-) 1mm
>1mm
Functional
2% 5%/2%
5%/2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
Functional
2mm diameter
2mm diameter
2mm diameter
2mm diameter
2mm diameter
2mm diameter
2mm
2mm
2%
100 ms
100 ms
Functioning
Wilkins; 2010
Assurance of Medical Accelerators. Med Phys.
1mm diameter
Frequency
Daily
Quality Metric
Safety
System operation and accuracy
Geometric
Image quality
Image qaulity
Dose
Imaging system performance
Geometric
System operation
Asterick (*) indicates the option to perform the quality check activity. For example, phantom localization and reposi
(*) signifies the activity substition; not an eliminiation.
Italized font indicates that test preformance can be completed semiannually after 6-8 months after commissioning (on
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010.
Quality Check
Imaging dose
X-ray generator
Performance (kV systems only)
Tube potential, mA, ms accuracy, linearity
Anteroposterior, mediolateral, craniocaudal orientations are maintained
Long and short term planning of resources (disk space, manpower, etc.)
vity. For example, phantom localization and repositioning with couch shifts can be preformed OR geometric calibration maps. T
iannually after 6-8 months after commissioning (only if resulting data proves stable)
Tolerance (+/-)
Functional
Functional
2 mm
2 mm
Replace/refresh
1 mm
1 mm
Baseline
Baseline
< 2 mm (or <5 lp/cm)
Baseline
Baseline
Baseline
Baseline
Accurate
Support clinical use and
current imaging policies and
procedures
CT Simulator QA
Performance Parameter
Alignment of gantry lasers with center of imaging plane
Frequency
Daily
Monthly
Annually
Annually
Annually
Annually
Annually
Semi-annual
Post replacement of major generator component
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010.
Tolerances (+/-)
2mm
2mm over length of laser projection
2mm
2mm over length of laser projection
2mm over length and width of tabletops
1mm over the range of table motion
1 mm over the scan range
Brachytherapy Sources
Source
Long half-life (description)
Performance Test
Physical/chemical form
Source encapsulation
Radionuclide distribution and
source uniformity
Location of radionuclide
Mean of batch
Deviation from mean
Calibration Vertificiation
Physical/chemical form
Source encapsulation
Mean of batch
Deviation from mean
Radionuclide distribution and
source uniformity
Frequency
Initial purchase
Initial purchase
Initial purchase
Initial purchase
Initial purchase
Initial purchase
Every use
Initial purchase
Initial purchase
Every use
Every use
Every use
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010.
Tolerance
Recorded
Recorded
Recorded
1mm
3%
5%
Visual Check
Recorded
Recorded
3%
5%
Visual Check
Tolerance (+/-)
Functional
Present
Functional
Functional
Accurate
Accurate
Functional
Functional
Functional
1 mm
Functional
Functional
Functional/Accurate
Functional/Accurate
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010
Monthly
Annual
Performance Tested
Understand algorithm
Single field/source isodose distribution
MU calculation
Test cases
I/O system
Checksum
Subset of reference QA
I/O system
MU calculation
Reference QA test
I/O system
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010
Tolerance (+/-)
Functional
2% or 2mm
2%
2% or 2mm
1mm
No charge
2% or 2mm
1mm
2%
2% or 2mm
1 mm
TPS QA
Parameters Tested
Entry of Axial Contours
Creation of PTV Axial Contours, Given a CTV
Collimator Setting
Aperture Definition
Traditional Planning
0.3cm
3D Planning
0.1cm
0.5-10cm
0.3cm
1.0-2.0cm
0.2-0.5cm
0.5cm
<0.1cm
0.5cm
0.1cm
0.3cm or more
0.1cm
0.1cm
1.00o
N/A
<1.00o
<1.00o
1.00%
1.00%
>10.00%
1.00%
2.00-5.00cm
10.00%
1.00-5.00mm
2.00% is Achievable
>100.00%
2.00%
>1.00cm
N/A
1.00mm
Depedant on Many Factors*
N/A
Reasoning
Traditional contour typically obtained mechanically. 3-D contour typically obtained
from CT.
Traditional system uses a 2-D PTV drawn by hand around the CTV expansion onto
other 2-D contours is quite inaccurate, as it is totally manual. In3-D system, PTV
can be created by 3-D expansion around the CTV through the software.
Traditional system involves totally manual registration and countour transfer. 3-D
system registration has at best about 2.00mm reproducibility, plus additional
distortions, plus transfer of MRI contours to CT dataset.
Traditiona; system may force beam center to be on axial calculation plane or CT
slice. 3-D system allows any specified isocenter coordinates
Resolution of jaw positions typically 1.00mm, although traditional system will
usually specify field width and length with resolution of 0.5cm at best
Block shape not modeled in prototype traditional system, but may be entered with
digitizer for some types of systems. 3-D system may use computer-generated
aperture.
Traditional system may not display aperture shape and may not display divergence
effects
Resolution of gantry angle typically 0.1 degrees in 3-D system.
Table and/or collimator angles often not allowed or displayed in traditional system
Traditional beam models reproduce measured data. 3-D models may do no better
since they are not directly based on measurements of this situation.
Traditional beam models do not handle non-axial behavior. 3-D models are just as
accurate in non-acial directions as axial directions.
Depends on grid effects, models.
Traditional beam normalization depends only on central axis of beam on axial slice
for the open rectangular field in a water phantom. 3-D normalization includes all
effects, including scatter under blocks and inhomogeneity effects.
Traditional system cannot handle blocks, so it can make large errors under blocks.
3-D model accurately handles dose under blocks, perhaps with an accuracy of 1.002.00%.