You are on page 1of 10

Vibration Measuring

Instruments

By:
Dr. B. K. Sridhara,
NIE, Mysore

eNotes for the session on EDUSAT


Programme

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 1


Chapter 5
Vibration Measuring Instruments

5.1 Introduction
In practice the measurement of vibrations becomes necessary due to following reasons.
1. To determine natural frequencies, modal shapes and damping ratios. The
measurement of frequencies of vibration and forces developed is necessary to
design active vibration isolation systems.
2. The theoretically computed vibration characteristics of a machine or structure may
be different from the actual values due to the assumptions made in the analysis. (To
verify the analytical models).
3. Periodic measurement of vibration characteristics of machines and structures
becomes essential to ensure adequate safety margins. (Preventive maintenance).
4. Measurement of input and resulting output vibration characteristics of a system
helps in identifying the system in terms of its mass, stiffness and damping.

5.2 Vibration Measurement Scheme


Figure 5.1 shows the basic features of a vibration measurement scheme.
1. Vibrating machine or structure.
2. Vibration transducer or pick up.
3. Signal conversion instrument.
4. Display / recording.
5. Data analysis.

2 3 4

1 5

Figure 5.1 Vibration Measurement Scheme

The motion of a vibrating body is converted in to an electrical signal by the vibration


transducer or pick up. The transducer transforms changes in mechanical quantities such as
displacement velocity, acceleration in to changes in electrical quantities such as voltage or
current. (Electrodynamic pick up, electromagnetic pick up, piezo electric pick up,
inductive displacement pick up, LVDT pick up, capacitive pick up). Since the output
signal of a transducer is too small to be recorded directly, a signal conversion instrument is
used to amplify the signal to the required value (Amplifier). The output from the signal
conversion instrument can be displayed on a display unit or stored in a computer for later
use (Oscilloscope, A to D converters, milli voltmeters, computers etc., the data can then be

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 2


analyzed to determine the desired vibration characteristics of the machine. Depending upon
the quantity measured the vibration measuring instrument is called a vibrometer, a velocity
meter, an accelerometer, a phase meter or a frequency meter. To summarise, following are
the guidelines.
1. Displacement measurements may be useful for studying low frequency vibrations,
where corresponding velocity and acceleration measurements are too small for
practical purposes.
2. Velocity measurements may be useful at intermediate frequencies where
displacement measurements are likely to be small to measure conveniently.
3. Acceleration measurements may be useful at high frequencies.

Instead of the above, vibration analyzers can also be used. Several commercial vibration
analyzers are available today. They consist of a vibration pick up and an FFT (Fast Fourier
Transformation) analyser, a balancing kit for phase measurement and an inbuilt computer.
The pick up essentially a piezo electric type with a natural frequency of 25 kcps. (KHz).
Built in double integration is also available for displacement plots. FFT converts time
domain signal to a signal in frequency domain to identify the frequencies of concern.

5.3 Vibration pick ups: Seismic Instruments


The commonly used vibration pick ups are called seismic instruments. The basic element
of many vibration measuring instrument is a seismic unit which is basically a spring mass-
damper system mounted on a vibrating body on which measurements are to be made as
shown in Figure 5.2.

x
Casing

C
y

Figure 5.2Seismic Unit

Depending on the frequency range utilized displacement, velocity or acceleration is


indicated, by the relative motion of the suspended mass with respect to the case.

Behaviour of Seismic unit


Consider the equation of motion of spring-mass-damper system, subjected to base
excitation, as shown in Figure 5.3.

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 3


ωt-φ
x = X Sin (ω φ)

y = Y Sin ωt
Base

Figure 5.3
.. . . – K (x-y)
mx = - C (x-y)

if..
Z = x-y; relative displacement the equation of motion becomes
. .
mZ + CZ + KZ = mω2Y Sin ωt

from this
Z r2
= [1- r ] + [2 r]2
2 2
Y
φ = tan-1 [2 r/1-r2]

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 4


The parameters that influence Z/Y and φ are: (1) frequency ratio r = ω/ωn. (2) Damping
factor , as shown in the Figure 5.4.

Range for Range for Vibrometer


Accelerometer
=0

= 0.1
Z/Y = 0.15
= 0.25
= 0.5

= 1.0

r = ω/ω
ωn

Figure 5.4. Frequency response curves


Figure 5.4 also shows the range of frequencies corresponding to which a seismic
instrument act as a vibrometer or an accelerometer. Type of instrument is determined by
the useful range of frequencies with respect to the natural frequency (ωn) of the instrument.
The relative displacement Z, may represent the displacement or acceleration depending
upon ωn of the seismic unit and frequency of vibrating body, ω.

5.4 Vibrometer or Seismometer


It is an instrument with low natural frequency. Therefore,

ω >>>>> ωn

r >>>> 1, r is very large.

Z/Y ≈ 1, in particular when r > 3

Z/Y ≈ 1, (independent of )

∴Z=Y

Relative displacement of the seismic mass = displacement of base.

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 5


∴Z = X-Y, X = 0, ∴Z=Y

Hence the seismic mass remains stationary. It remains undisturbed in space. The
supporting casing moves the vibrating body. Thus the relative displacement between the
casing and the mass is the true displacement of the casing. Like wise, the relative velocity
between the casing and the mass is the true velocity of casing. Usually, the relative motion
Z is converted into electric voltage. The seismic mass is a magnet moving relative to the
coils fixed to the case, as shown in Figure 5.6.

0 0
x 0 0
0 0 Seismic mass

Figure 5.6

The voltage generated is proportional to the rate of cutting of magnetic field. Therefore the
output of the instrument is proportional to the velocity of the vibrating body. Such
instruements are called velometers. A typical instrument of this kind may have a natural
frequency of 1 Hz to 5 Hz and a useful range of 10 Hz to 2000 Hz. The sensitivity of such
instruments may be in the range of 20 mV/cm to 350 mV/cm. Both the displacement and
acceleration are available from the velocity type transducer by means of the integrator or
the differeniator provided in most signal conditioner units.

Limitation of Vibrometers
In order to have r >>>1, ωn should be very small. This means that, the mass must be very
large and the spring must have a very low stiffness. Therefore, a vibrometer is a spring-
mass-damper system with a very large mass and a flexible spring. This results in bulky
instrument, which is not desirable in many applications.

In practice, a vibrometer may not have a large value of r, and hence the value of Z, may not
be exactly equal to Y. In such cases the true value of Y, can be computed from:
Z = r2
Y [1- r2]2 + [2 r]2

5.5 Accelerometer
It is an instrument with high natural frequency. When the natural frequency of the
instrument is high compared to that of the vibrations to be measured, the instrument
indicates acceleration.

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 6


Then
ω <<<< ωn,
r <<<<< 1,
the factor [r – (ω/ωn)2]2 + (2 r)2 approaches unity.

∴Z (ω/ωn)2.Y

(1/ωn2). ω2Y

Hence, Z α ω2 Y, which implies that Z is proportional to the acceleration of the vibrating


body. Thus in order to make r <<<< 1, ωn should be very large. Hence K should be very
large and m should be small. This means that, the instrument needs a small mass and
spring of large stiffness. Therefore, the instrument will be very small in size and compact.

Due to their small size and high sensitivity accelerometers are preferred in vibration
measurements. The acceleration measured can be integrated once or twice with the help of
modern electrical circuits to obtain velocity and displacement of the system.

Thus the difference between a vibrometer and an accelerometer is in its natural frequency.
In vibrometer it is very small where as in accelerometer it is very high. The principle of
construction remains same.

5.6 Useful Frequency Range


The useful range of accelerometer can be seen from the following graph for different
amounts of damping . Useful frequency range is that range of r between which the
maximum error is less than 0.01 %. The useful frequency range for un damped
accelerometer is very much limited. However, with = 0.7 the useful frequency range is
quite large, that is, between 0 ≤ ω/ωn ≤ 0.20, as shown in Figure 5.7.

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 7


Figure 5.7. Useful frequency range

Thus an instrument with a natural frequency of 100 Hz has a useful frequency range of 0 to
20 Hz with negligible error. (Up to 20 Hz the error is less than 0.01%). Figure 5.8 shows
accelerometers.

Figure 5.8. Accelerometers

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 8


Numerical Examples on Vibration Measuring Instruments

1. A vibrometer having a natural frequency of 4 rad/sec and = 0.2 is attached to a


structure that executes harmonic motion. If the difference between the maximum and
minimum recorded value is 8 mm, find the amplitude of vibration of structure when its
frequency is 40 rad/sec.

ωn = 4 rad/sec, = 0.2
Z = Relative amplitude Z 4 mm
= 8/2 = 4 mm 8 mm
Mean
ω = 40 rad/sec
r = ω/ωn = 40/4 = 10
Z = r2
= 1.0093
Y [1- r2]2 + [2 r]2

Z/Y = 1.0093 ∴ Y = 3.9631 mm

2. A vibrometer has a natural frequency of 10 cps and has a damping ratio of 0.7. It is used,
by mistake, to measure vibrations of a fan base at an exciting frequency of 180 rpm. The
measured vibration velocity of the fan base is 3 mm/s. What is the actual velocity of the
fan base?

For a vibrometer,
Z = r2
Y [1- r2]2 + [2 r]2
In the present case, ωn = 10 cps = 62.8 rad/sec.
Exciting frequency ω = 180 rpm = 18.84 rad/sec.
Hence
r = ω/ωn = 0.3
(Z/Y) = 0.09/(0.8281 + 0.1764) = 0.089

Hence Y = Z/0.089 = 33.6 mm/s.

It may be noted that the actual velocity is beyond permissible limits, whereas what is
read is well below the permissible limit. Hence one should be very careful in selecting
the proper instrument.

3. A seismic instrument is fitted to measure the vibration characteristics of a machine


running at 120 rpm. If the natural frequency of the instrument is 5 Hz and if it shows 0.004
cm determine the displacement, velocity and acceleration assuming no damping.

ƒn = 5 Hz ; ωn = 2πƒn = 10π rad/sec = 31.4 rad/sec


N = 120 rpm ω = 2πN/60 = 12.56 rad/sec
r = ω/ωn = 12.56/31.4 = 0.4

Z = 0.004 cm = 0.0004 mm

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 9


For seismic instruments

Z = r2 , =0
Y [1- r2]2 + [2 r]2
Z r2 r2
Y = [1- r2]2
=
[1- r2]

∴Displacement Y = Z(1-r2) / r2 = 0.021 cm

Velocity V = ω Y = 2πN/60* 0.021 = 0.26 cm/sec

Acceleration a = ω2Y = ω(ωY) = 3.265 cm/sec2

4. A vibrometer indicates 2 percent error in measurement and its natural frequency is 5 Hz.
If the lowest frequency that can be measured is 40 Hz, find the value of damping factor.

Solution:
Data: ωn = 5Hz, ω = 40 Hz, error = 2%

r = ω/ωn = 40/5 = 8

Z/Y = 1.02 (since the error is 2%)


Z r2
=
Y [1- r2]2 + [2 r]2

(1.02)2 = 82/(1-64)2 + (16 )2

= 0.35

Dr. B. K. Sridhara, Asst. Professor, Mechanical Engineering, NIE, Mysore-8 10

You might also like