Professional Documents
Culture Documents
Learner Objectives
Understand the importance of monitoring and interpreting ventilator graphics. Learn the basic shapes of waveforms. Identify the different types of waveforms and loops available on most ventilators. Learn to use the graphics to identify patient/ventilator problems, and make the appropriate adjustments.
Ventilator graphics are waveforms that reflect the patient-ventilator system and their interaction.
Learning to analyze and interpret ventilator graphics can be a very useful tool for respiratory therapists. The graphics can assist therapists in making recommendations and necessary adjustments to the ventilator.
Practitioners can use ventilator graphics to assess the condition of a patients lungs in the same way a cardiologist uses the graphics of an EKG to view the condition of the heart.
Hamilton Galileo
The graphics display will have different appearances and configurations, depending on the make and model of ventilator.
Scalars
Time
Loops
Generally, The ascending and descending ramps can be considered the same as exponential ramps, So you really only need to remember three shapes: Square, Ramp, and Sine waves.
Squares
Squares represent fixed, constant, or set parameters. For example, pressure setting in Pressure Control mode. Ramps represent variables. Will vary with changes in lung characteristics. Ramps can be accelerating or decelerating. Sine waves are seen with spontaneous, unassisted breathing.
Ramps
Sine
Types of Waveforms
Each time the ventilator delivers a breath, there are waveforms that are displayed on the graphics screen: Pressure, Flow, and Volume.
Types of Waveforms
(20)
The pressure scalar displays the amount of pressure generated with each breath. The scale is located on the left side in cmH20. The peak pressure of these breaths are about 22 cmH20.
Types of Waveforms
The flow scalar displays the flowrate, or speed associated with each breath. In this example, the breath begins at a fast flowrate and then decreases during inspiration.
Types of Waveforms
(375 ml)
(250 ml)
Volume in
Volume out
The volume scalar displays the amount of volume inhaled and exhaled with each breath. The scale is located on the left side in mls. These breaths are about 370 ml.
Types of Waveforms
Question:
What 3 waveforms are displayed on the graphics screen? Answer:
Pressure
Flow
Volume
Types of Waveforms
Volume Modes
Pressure Pressure
Pressure Modes
square
Flow
Flow
square
Volume
Volume
Types of Waveforms
Volume Modes Pressure Modes
Pressure
Pressure
Flow
Flow
Note: Some ventilators allow you select the desired flow pattern for different modes.
Types of Waveforms
SIMV (Vol. Control) + Pressure Support
Volume Breath
Pressure Breath
Pressure Breath
Volume Breath
For combination modes, such as SIMV/PS, the graphics can show both volume breaths and pressure breaths.
Pressure Waveform
Volume Modes
Pressure Pressure
Pressure Modes
square
Flow
Flow
square
Volume
Volume
Pressure Waveform
The pressure waveform can be used to assess:
Breath Type (Pressure vs. Volume) Air trapping (auto-PEEP) Airway Obstruction Bronchodilator Response Respiratory Mechanics (Compliance/Raw) Active Exhalation PIP, Pplat CPAP, PEEP Asynchrony Triggering Effort
Pressure Waveform
Volume Modes Pressure Modes
If the ventilator delivers a volume breath, the shape of the pressure waveform will be a ramp.
If the ventilator delivers a pressure breath, the shape of the pressure waveform will have a square shape.
square
Ramp = variable
Square = constant
Pressure Waveform
Volume Modes Pressure Modes
This means that pressure will vary, depending on lung characteristics; (compliance, airway resistance, etc.)
This means that pressure will remain the same during inspiration.
Ramp = variable
Square = constant
Pressure Waveform
Pressure waveform Pressure waveform with inspiratory pause
PIP Pplat Raw
Setting an inspiratory pause time or performing an insp. hold maneuver, will create a plateau on the waveform. The plateau allows for easy visualization of PIP, Pplat, and Raw. Adding an inspiratory pause time may improve distribution of ventilation.
Pressure Waveform
I-time Peak Inspiratory Pressure (PIP) Plateau Pressure (Pplat) Alveolar Distending Pressure
Elastic Recoil Pressure
Pressure Waveform
10 5
PEEP
No patient effort
PEEP
Patient triggered breath
PEEP
The baseline for the pressure waveform will be higher, when Positive EndExpiratory Pressure (PEEP), is added. PEEP is also a component of mean airway pressure (MAP). With patient triggered breaths, there will be a negative deflection just before the waveform.
Pressure Waveform
Peak Inspiratory Pressure (PIP)
I-time
Pressure Waveform
Increased Airway Resistance (Raw)
PIP
(Increased Raw) Pplat
Decreased Compliance
PIP
(Normal Raw) Pplat
Increased airway resistance (Raw) will cause the PIP to increase. The Pplat pressure remains normal.
Decreased lung compliance will cause the entire waveform to increase in size. The difference between PIP and Pplat will remain normal.
Pressure Waveform
Air-Trapping (auto-PEEP)
PEEP
+9 +5 +14
Total-PEEP
While performing an expiratory hold maneuver, trapped air will cause the waveform to rise above the baseline. An acceptable amount of auto-PEEP should be < 5cm H2O.
Pressure Waveform
Label the parts:
A
B C =G
Pressure Waveform
Question: What does this pressure waveform show?
Flow Waveform
Volume Modes
Pressure Pressure
Pressure Modes
square
Flow
Flow
square
Volume
Volume
Flow Waveform
If a volume breath is delivered, the shape of the waveform will be a square. This means that the flowrate stays the same during inspiration. If a pressure breath is delivered, the shape of the waveform will be a ramp. This means that the flowrate starts out high and then decreases during inspiration.
Square = constant
Ramp = variable
Flow Waveform
The flow waveform can be used to assess:
Air trapping (auto-PEEP) Airway Obstruction I-Time adjustment Bronchodilator Response Active Exhalation Breath Type (Pressure vs. Volume) Inspiratory Flow Asynchrony Triggering Effort
Flow Waveform
Volume Modes Pressure Modes
I-time
Set Flowrate
I-time
PIF
E-time
E-time
PEF
PEF
Flow Waveform
The decelerating flow pattern may be preferred over the constant flow pattern. The same tidal volume can be delivered, but with a lower peak pressure.
Flow Waveform
Airway Obstruction
PIF
= Normal
PEF
In patients with severe airway obstruction, the flow waveform can become a plateau. This can become a problem in flow-cycled modes, such as Pressure Support. (asynchrony, W.O.B.)
Flow Waveform
Auto-Peep (air trapping)
= Normal
If the expiratory portion of the waveform doesnt return to baseline before the start of the next breath starts, there could be air trapping. (emphysema, improperly set I:E ratio).
Flow Waveform
Bronchodilator Response
Pre-Bronchodilator I-time Post-Bronchodilator
Prolonged E-time
Normal E-time
To assess response to bronchodilator therapy, you should see an increase in peak expiratory flow rate. Also, the expiratory portion of the curve should return to baseline sooner.
Flow Waveform
Adjusting I-Time
Pressure Support, Volume Support
(Flow-cycled)
Pressure
Flow
Volume
In Pressure modes that are time-cycled (Pressure Control), the flow waveform should return to baseline. In Pressure modes that are flow-cycled (Pressure Support), the flow waveform does not return to baseline. (Adjustment is made with Insp. Cycle Off %)
Flow Waveform
Adjusting I-Time
I-time
(Too short)
I-time
(Too long)
Not here
Not here
Remember, this applies to Control modes with a decelerating flow (ramp). This includes Pressure Control, PRVC, SIMV (Press. Control), and SIMV (PRVC). Volume Control and SIMV (Vol. Control) have a constant flow.
Flow Waveform
Question: What does this flow pattern show?
Flow Waveform
Question: To assess improvement after a breathing treatment you should see what?
Pre-Bronchodilator I-time Post-Bronchodilator
Normal E-time
Improved PEF
Volume Waveform
Volume Modes
Pressure Pressure
Pressure Modes
square
Flow
Flow
square
Volume
Volume
Volume Waveform
Volume Modes Pressure Modes
The Volume waveform will generally have a mountain peak shape regardless of what mode, of ventilation.
There may also be a small plateau at the top of the volume waveform
Volume Waveform
The volume waveform can be used to assess:
Tidal Volume Active Exhalation Asynchrony Airway Resistance Air trapping (auto-PEEP) Leaks
Volume Waveform
Volume Waveform
Air-Trapping or Leak
Volume Loss
If the exhalation side of the waveform doesnt return to baseline, it could be from air-trapping (improperly set I-time, emphysema), or there could be a leak (ET tube, vent circuit, chest tube, etc.)
Volume Waveform
Increased Airway Resistance
= Normal
If there is an increase in airway resistance, it will take longer for the exhaled volume to return to baseline. This is more commonly seen with a damp or blocked expiratory filter/valve.
Volume Waveform
Question: The volume waveform is most commonly used to assess which two situations?
Is it Volume or Pressure?
Volume Modes Pressure Modes
Pressure
Pressure
square
Flow
square
Flow
Volume
Volume
Here is an easy way to tell what type of mode (or type of breath) this is, just by looking at the graphics. Look at the pressure waveform and remember the letter P. With Pressure breathsThe Pressure waveformwill have a Plateau.
Is it Volume or Pressure?
The pressure waveform has a plateau
Is it a Volume or Pressure mode? (Hint: look at the pressure waveform) Is it a Control mode or Support mode? (Hint: look at the flow waveform)
Volume Modes
Pressure Modes
Ventilator Graphics
PressureVolume Loops
Pressure-Volume Loops
Volume is plotted on the y-axis, Pressure on the x-axis. Inspiratory curve is upward, Expiratory curve is downward. Spontaneous breaths go clockwise and positive pressure breaths go counterclockwise. The bottom of the loop will be at the set PEEP level. It will be at 0 if theres no PEEP set. If an imaginary line is drawn down the middle of the loop, the area to the right represents inspiratory resistance and the area to the left represents expiratory resistance.
Pressure-Volume Loops
The Pressure/Volume Loop can be used to assess: Lung Overdistention Airway Obstruction Bronchodilator Response Respiratory Mechanics (C/Raw) Flow Starvation Leaks WOB Triggering Effort
Pressure-Volume Loops
Volume Modes
Pressure Modes
In Volume modes, the P/V loop will normally have a football shape.
Pressure-Volume Loops
500
250
10
15
20
If PEEP is added, the loop will begin at the set PEEP level. The top part of the P/V loop represents Dynamic compliance (Cdyn). Cdyn = volume/pressure
Pressure-Volume Loops
500
Plateau
250
10
20
Pressure modes deliver a set constant pressure during inspiration, creating a plateau on the pressure waveform. This will also create a plateau on the P/V loop.
Pressure-Volume Loops
Overdistention
500
beaking
250
10
20
Pressure continues to increase with little or no change in volume, creating a bird beak. Fix by reducing amount of tidal volume delivered.
Pressure-Volume Loops
Triggering Effort
WOB
tail
If the patient is triggering the breath, you will see a crossover or tail, at the beginning of the loop. As WOB increases, the tail will become larger.
Pressure-Volume Loops
Airway Resistance
500
ETT size too small, tube kinked, patient biting tube, etc.
As airway resistance increases, the loop will become wider. An increase in expiratory resistance is more commonly seen.
Pressure-Volume Loops
Increased Compliance
Cdyn
500
Decreased Compliance
500
Cdyn
250 250
10
20
10
20
Cdyn = Dynamic Compliance Examples: Examples: (volume/pressure) Emphysema, ARDS, CHF, Post Surfactant Therapy Atelectasis, Pleural Effusions
Pressure-Volume Loops
Increased Compliance
600 600/12 Cdyn = 50 ml/cmH20
Decreased Compliance
500
500
Increased volume
300
250
Decreased volume
250
12 Decreased pressure
22
10
20
10
Increased pressure
20
VT (Tubing compliance) Examples: Examples: Cdyn = PIP - PEEP Emphysema, ARDS, CHF, Post Surfactant Therapy Atelectasis, Pleural Effusions
Pressure-Volume Loops
A Leak
500
250
10
20
The expiratory portion of the loop does not return back to baseline. This indicates that there is a leak.
Pressure-Volume Loops
point of alveolar collapse
500
250
Inflection Points
Some lung protection strategies for treating ARDS, suggest setting PEEP just above the lower inflection point, to hold the alveoli open.
Pressure-Volume Loops
500
250
10
20
Question: What does this loop show? Answer: Decreased lung compliance. (ARDS, CHF, Atelectasis)
Pressure-Volume Loops
500
250
15
30
Question: What is happening when there is a bird beak appearance on the P/V loop? Answer: Lung overdistention. Pressure continues to increase, with no increase in volume.
Pressure-Volume Loops
Question: What does this P/V loop show? Answer: Increased expiratory airway resistance (Raw).
Ventilator Graphics
40 20 200 0 20 600
FlowVolume Loops
40
Flow is plotted on the y axis and volume on the x axis Flow volume loops used for ventilator graphics are the same as ones used for Pulmonary Function Testing, (usually upside down). Inspiration is above the horizontal line and expiration is below. The shape of the inspiratory portion of the curve will match the flow waveform. The shape of the exp flow curve represents passive exhalation. Can be used to determine the PIF, PEF, and Vt Looks circular with spontaneous breaths
Flow-Volume Loops
The Flow/Volume Loop can be used to assess: Air trapping Airway Obstruction Airway Resistance Bronchodilator Response Insp/Exp Flow Leaks Water or Secretion accumulation Flow Starvation Asynchrony
Flow-Volume Loops
60 40
20
Begin Inspiration
200 400 600
Begin Expiration
-20
-40
-60
Flow-Volume Loops
Constant Flow Variable Flow
The shape of the inspiratory portion of the curve will match the flow waveform.
Flow-Volume Loops
PEF
Airway Obstruction
Reduced PEF
scooping
PFT view
With conditions that cause airway obstruction (asthma), you will see a lower peak expiratory flow (PEF). You should also see scooping on the expiratory portion of the loop. *Scooping means that the volume is being exhaled at a slower rate because of obstruction.
Flow-Volume Loops
Airway Obstruction
scooping
Flow-Volume Loops
60
-20
loss of volume
-40
= Normal
-60
If there is air-trapping, or a leak, the loop will not meet at the starting point where inhalation starts and exhalation ends.
Flow-Volume Loops
Water or Secretions
If there is a collection of water in the ventilator circuit or a build up of secretions, you will see a jagged, sawtoothed pattern.
Flow-Volume Loops
20
Begin Inspiration
200 400 600
Begin Expiration
-60
Flow-Volume Loops
60 40
20
-20
-40
-60
Question: What does this F/V loop show? Answer: There is air-trapping or a leak. (ETT cuff, vent circuit)
Flow-Volume Loops
Question: What is the term used to describe the part of the loop indicated by the arrow? What causes it? Answer: This is known as scooping. Its caused by small airway obstruction.
Asynchrony
(out of sync)
Flow Starvation
The inspiratory portion of the pressure waveform shows a dip. Because of an inadequate flowrate , the patient is sucking in attempting to get more air. This causes a drop in pressure. A main disadvantage of constant flow modes (e.g. Volume Control), is that it may not meet the patients inspiratory demands.
Asynchrony
(out of sync)
Dip
Flow Starvation
Ventilator detects inadequate flow
Some ventilators have an adaptive flow system that will automatically increase flow to try and meet the patients demand. Notice that on the next breath, the flowrate has been increased.
Asynchrony
(out of sync)
F/V Loop
P/V Loop
Air-Trapping (auto-PEEP)
Causes:
Insufficient expiratory time Early collapse of unstable alveoli/airways during exhalation
How to Fix:
Give a bronchodilator treatment, adjust I-time, increase flow, adjust PEEP.
How to fix:
Give a bronchodilator treatment, suction patient, drain water, change HME, change ETT, add a bite block, reduce flowrate, change exp filter.
Compliance Changes
Decreased compliance
Causes
ARDS Atelectasis Abdominal distension CHF Consolidation Fibrosis Overdistention Pneumothorax Pleural effusion
Increased compliance
Causes
Emphysema Surfactant Therapy
Leaks
Causes:
Expiratory leak: ETT cuff leak , chest tube leak, BP fistula, NG tube in trachea Inspiratory leak: Loose connections, ventilator malfunction, faulty flow sensor
Asynchrony
Causes: (Flow, Rate, or Triggering)
Air hunger (flow starvation) Neurological Injury Improperly set sensitivity Pressure waveform: Patient tries to inhale/exhale in the middle of the waveform, causing a dip in the pressure Flow waveform: Patient tries to inhale/exhale in the middle of the waveform, causing erratic flows/dips in the waveform Pressure/Volume loop: Patient makes effort to breath causing dips on either inspiratory or expiratory side. Flow/Volume loop: Patient makes effort to breath causing dips in loop on either inspiratory or expiratory side.
Try increasing the flow rate, decreasing the I-time, or increasing the set rate to capture the patient to better meet the patients needs. Change the mode - sometimes changing from partial to full support will solve the problem. If neurological, may need paralytic or sedative. Adjust sensitivity
The pressure waveform does have a plateau, but its from an insp. hold or set insp. pause.
square
Ventilator Graphics
Rise Time
Rise Time
The inspiratory rise time is the time it takes to reach full inspiratory flow, or pressure, at the start of each breath. The rise time can be expressed as a percentage of the breath cycle time (%) or in seconds (s).
Rise Time
The Bart Simpson spike
Rise time overshoots desired pressure
Too fast
If rise time is set too fast, you can get an overshoot in the pressure wave, creating a pressure spike. If this occurs, you need to increase the rise time. This makes the flow valve open more slowly.
Rise Time
Rise Time
Rise Time
Too slow
If rise time is too slow, the pressure waveform will become more slanted, when it should look more square. This may affect Vt delivery and may not meet the patients inspiratory demands. If this occurs, you will need to decrease the rise time to open the valve faster.
In flow-cycled modes, like Pressure Support & Volume Support, the inspiratory cycle off % determines when the ventilator cycles from inspiration to expiration. Also know as: Inspiratory cycle threshold Inspiratory flow termination, Expiratory flow sensitivity, Inspiratory flow cycle %, E-cycle, etc
pressure
Inspiration ends
flow
The inspiration ends when inspiratory flow has dropped to a specific flow value.
Flow
50% 25%
30%
In the above example, the ventilator is set to stop inspiration and begin expiration, at 30% of the patients peak inspiratory flow (PIF).
100%
100%
60% 10%
The cycle off percentage is too high (60%), cycling inspiration off too soon.
This makes the breath too small. (not enough Vt.)
The cycle off percentage is too low (10%), This makes the breath too long. This forces the patient to actively exhale against the breath, creating a pressure spike. (increases WOB),
Rise Time
Question: What does this waveform show?
Question: This pressure support breath is set to insp. cycle off at 80%, 50%, 30%, or 10%?
100% 75%
Flow
50% 25%
Answer: 30%
Sources:
Rapid Interpretation of Ventilator Waveforms,
Waugh, Harwood, and Deshpande
Ventilator Waveform Analysis, Anatomy of Servo-i Graphics, Maquet, inc. Golden Moments in Mechanical Ventilation,
Pearson
Maquet, inc.
Ventilator Graphics
Thank You!