SERVICE MANUAL

SINGLE PATIENT DIALYSIS MACHINE

PREFACE
Thank you for purchase of Single Patient Dialysis Machine, SURDIAL. The service manual is designed to give information required to check/maintain and repair SURDIAL a qualified service engineer and technician. Any operation or work contained in the manual should be informed by them. They should have knowledge and ability sufficient to replace microprocessor based electronic component with new one, and should have general knowledge concerning fluid lines of dialysis machine. Also they should read and full understand contents of the manual before executing component replacement or maintenance of SURDIAL. Use the manual by referring to the operation manual of SURDIAL. Generally, description common to the both manuals is minimal with exception that the description is contained in some chapters. Information required to operate SURDIAL is contained in the operation manual (and technical information is also found in some chapters) and information related to maintenance, component replacement and various settings of SURDIAL is contained in the service manual.

Description in the service manual is protected by copyright of the manufacturer. Copy, reproduction or preparation of version in any foreign language of the description partially or wholly should not be allowed without obtaining prior approval by the manufacturer.

NIPRO SURDIAL (SEM) ver. OR2102-9902

CAUTION -1

CAUTIONS FOR SAFETY


To use the machine (SURDIAL) safely, it shall be operated in accordance with specified procedures and shall be maintained regularly after reading and full understanding cautions included in this Operation Manual. Procedures for operation and cautions for safety contained in the manual are only applicable to the case when the machine is used for specified purpose. It shall be noted that any danger caused by use of the machine based on any procedures not specified in the manual shall be responsible for user. (1) Messages for safety are classified into the following three categories:

DANGER Cases critically causing death or severe injury of user by misuse of the machine WARNING Cases potentially causing death or severe injury of user by misuse of the machine CAUTION Cases potentially causing injury of user and/or loss of property by misuse of the machine (2) NOTE Reference to operation of the machine

NIPRO SURDIAL (SEM) ver. OR2102-9902

CAUTION -2

DANGER

Before use of the machine, read and understand this Operation Manual. The machine can be used safely and reliably when operated and maintained as specified in the manual. The machine shall be operated by well trained and qualified person only to avoid danger and/or the machine damage due to incorrect use. Be familiar with procedure how to stop the machine in emergency. Full understand procedure how to stop the machine to prevent incidence of serious problems. The machine shall be operated by well trained and qualified person only unless otherwise instructed formally. Be familiar with procedure how to take corrective actions if the machine is stopped by any reason. Take corrective actions specified in the manual, after checking the machine conditions. DO NOT touch board, wiring and terminal with hand in the case other than maintenance. DO NOT touch wiring and terminal not insulated. DO NOT touch board with wet hand. Turn OFF the power breaker and remove plug when touch terminal. Take the following care to avoid potential danger to patient when blood line connection and disconnection: DO NOT enter air bubbles into patient body. DO NOT connect blood line to the patient before completion of priming operation. DO NOT connect blood line to the patient during rinsing operation. DO NOT enter harmful virus or chemical into patient body. Check frequently that blood line is connected correctly to avoid danger to patient due to blood leak caused by disconnection of the line when incorrectly connected, and due to rupture of the line caused by excess pressure application.

NIPRO SURDIAL (SEM) ver. OR2102-9902

CAUTION -3

WARNING
DO NOT use electric wave generating electronic device in any room or building where single patient dialysis machine is operated. If the device such as portable telephone, transceiver or radio control toy is operate or used in any place in hospital, malfunction of instrument used for medical treatment may be caused. Manage and instruct any person not to bring or use any electric wave generating device in hospital to avoid any failure. DO NOT change parameter setting during dialysis excluding the case of emergency. Change parameter setting before connecting with patient. Set parameter as intended after full understanding its function. DO NOT touch power cable on the floor to prevent electric shock due to fluid beak caused by the machine trouble. Provide drain port near the machine. DO NOT touch switch unnecessary for the machine operation during dialysis. DO NOT incline the machine at an angle 5 (Maximum) under loading any object on container support or hunger to prevent falling down of the machine. DO NOT load any object during transport. DO NOT incline the machine at an angle of 10 (Maximum) under no load to prevent falling down of the machine.

NIPRO SURDIAL (SEM) ver. OR2102-9902

CONTENTS -1

CONTENTS
PREFACE ..................................................................................................................... PREFACE-1 CAUTIONS FOR SAFETY............................................................................................ CAUTION-1 DANGER........................................................................................................... CAUTION-2 WARNING......................................................................................................... CAUTION-3 CAUTION .......................................................................................................... CAUTION-4 1. APPLICATION .............................................................................................................. 1-1 2. CONTRAINDICATION ................................................................................................. 2-1 3. CAUTIONS.................................................................................................................... 3-1 4. SYMBOL FOR SAFETY................................................................................................ 4-1 5. FLOW CHART 5.1. Fluid and blood line ............................................................................................ 5-1 5.1.1. Flow chart ................................................................................................. 5-1 5.1.2. List of input components .......................................................................... 5-2 5.1.3. List of output components ........................................................................ 5-3 5.1.4. List of other components ......................................................................... 5-4 6. DISASSEMBLING AND REPLACEMENT 6.1. Input components................................................................................................ 6.1.1-1 6.1.1. Conductivity cell Assy (CD) ...................................................................... 6.1.1-1 6.1.2. Dialysate pressure sensor Assy (PD) ...................................................... 6.1.2-1 6.1.3. Blood leak sensor Assy (BLD) ................................................................. 6.1.3-1 6.1.4. Pressure SW: AW-3200-63-A (PS) .......................................................... 6.1.4.1 6.2. Output components............................................................................................. 6.2.1-1 6.2.1. Two way electromagnetic valve : AV-3200-318-SC (V2/V3/V12/V13/V14(V15/V17)) .......................................................................... 6.2.1-1 6.2.2. Two way electromagnetic valve: AV-3202-63-S (V4 / V11) ....................... 6.2.2-1 6.2.3. High vacuum electromagnetic valve: AV-3202-63-SH (V10) ..................... 6.2.3-1 6.2.4. Three way electromagnetic valve: AV-3302-63-S (V5 / V6 / V7 / V8)......................................................................... 6.2.4-1 6.2.5. Bellowphragm pump: VCP-60 (Vpa / VPb) ............................................... .6.2.5-1 6.2.6. Blood pump Assy (BP / BPC) .................................................................... .6.2.6-1 6.2.7. Micro gear pump Assy (P1/P2) .................................................................. .6.2.7-1 6.2.8. Venous clamp Assy (CLV) ........................................................................ .6.2.8-1

NIPRO SURDIAL (SEM) ver. OR2102-9902

CONTENTS -2
6.3. Other components ............................................................................................ . 6.3.1-1 6.3.1. Deaeration bath Assy (AS 1) ................................................................. . 6.3.1-1 6.3.2. Deaeration bath Assy (AS2) .................................................................. . 6.3.2-1 6.3.3. Buffer tank W Assy (BTW)...................................................................... . 6.3.3-1 6.3.4. Buffer tank D Assy (BTD1 / BTD2)........................................................ . 6.3.4-1 6.3.5. Check valve(C1/C2) ................................................................................ . 6.3.5-1 6.3.6. Supply water filter (F) ............................................................................. . 6.3.6-1 6.3.7. Flow meter Assy (FM / FS) .................................................................... . 6.3.7-1 6.3.8. Reducing valve Assy (PR) ..................................................................... . 6.3.8-1 6.3.9. Relief valve(RV1/RV2/RV3) ................................................................... . 6.3.9-1 6.3.10. Sampling port Assy (SL) ...................................................................... . 6.3.10-1 6.3.11. Chamber Assy (VCa / VCb) ................................................................. . 6.3.11-1 6.3.12. Concentrate tank connectors(for A /B/disinfect concentrates) ................. . 6.3.12-1 7. CONSTRUCTION OF ELECTRICAL PART 7.1. Components of electrical part ............................................................................ . 7-1 7.1.1. Components at upper part of electrical part............................................ . 7-1 7.1.2. Component at lower part of electrical part .............................................. . 7-1 7.1.3. Placement of electrical part at machine .................................................. . 7-2 7.1.4. Components at left side in electrical POWER BOX ............................... . 7-3 7.1.5. Components at right side in electrical POWER BOX ............................. . 7-3 8. CONNECTION, REPLACEMENT AND ADJUSTMENT 8.1. Operation panel sheet ........................................................................................ . 8.1-1 8.2. Operation panel board ........................................................................................ . 8.2-1 8.3. LCD module ......................................................................................................... . 8.3-1 8.4. Bubble sensor board ........................................................................................... . 8.4-1 8.5. Dialysate pressure sensor board....................................................................... . 8.5-1 8.6. Main board ........................................................................................................... . 8.6-1 8.7. Venous and arterial pressure sensor boards ................................................... . 8.7-1 8.8. DSUB board ......................................................................................................... . 8.8-1 8.9. Buzzer board........................................................................................................ . 8.9-1 8.10. Sub-board .......................................................................................................... . 8.10-1 9. MAINTENANCE MODES 9.1. Display .................................................................................................................. 9.1-1 9.2. Maintenance modes ............................................................................................ 9.2-1 9.3. Maintenance mode menu screen....................................................................... 9.3-1 9.4. Maintenance mode 1 screen .............................................................................. 9.4-1

NIPRO SURDIAL (SEM) ver. OR2102-9902

CONTENTS -3
9.5. Maintenance mode 2 screen .............................................................................. 9.5-1 9.6. Maintenance mode 3 screen .............................................................................. 9.6-1 9.7. Maintenance mode 4 screen .............................................................................. 9.7-1 9.8. Maintenance mode 5 screen .............................................................................. 9.8-1 9.9. Maintenance mode 6 screen .............................................................................. 9.9-1 9.10. Maintenance mode 7 screen ............................................................................ 9.10-1 9.11. Maintenance mode 8 screen ............................................................................ 9.11-1 9.12. Maintenance mode 9 screen ............................................................................ 9.12-1 9.13. Maintenance mode 10 screen.......................................................................... 9.13-1 10. ADJUSTMENT 10.1. Measurement of chamber capacity and adjustment of silicone volume ............... 10.1-1 10.2. Adjustment of venous and arterial pressure sensors........................................... 10.2-1 10.3. Adjustment dialysate pressure sensor ................................................................ 10.3-1 10.4. Adjustment of temperature sensor ...................................................................... 10.4-1 10.5. Adjustment of pressure SW ................................................................................. 10.5-1 10.6. Adjustment of blood leak sensor.......................................................................... 10.6-1 10.7. Adjustment of reducing valve............................................................................... 10.7-1 10.8. Adjustment of priming end pressure .................................................................... 10.8-1 10.9. Adjustment of closed circuit pressure .................................................................. 10.9-1 10.10.Adjustment of priming flow rate .......................................................................... 10.10-1 10.11.Adjustment of bypass flow rate .......................................................................... 10.11-1 10.12.Adjustment of gaps between housing and rotor rollers of blood pump............... 10.12-1 10.13. Adjustment of blood pump flow rate ................................................................. 10.13-1 10.14. Leak test and gap adjustment of clamp ............................................................ 10.14-1 10.15. Adjustment of bubble sensor ............................................................................. 10.15-1 10.16. Adjustment of conductivity and concentration ................................................... 10.16-1 10.17. Adjustment of contrast ...................................................................................... 10.17-1 11. DIP Switches (DIP SW's) 11.1. Operation of DIP SW's ........................................................................................ 11.1-1 11.2. Function's of DIP SW's ....................................................................................... 11.2-1 12. VARIOUS DATA 12.1. Loaded CPU and ROM........................................................................................ 12.1-1 12.2. Wiring in POWER box ......................................................................................... 12.2-1

NIPRO SURDIAL (SEM) ver. OR2102-9902

1-1

1. APPLICATION

(a)

Single patient dialysis machine, SURDIAL is applied to treatment of patients with acute or chronic renal failure by dialysis when the application is instructed by doctor.

(b)

SURDIAL can be used at both hospital and home.

NIPRO SURDIAL (SEM) ver. OR2102-9902

2-1

2. CONTRAINDICATION
Any application other than dialysis using SURDIAL is contraindicated.

NIPRO SURDIAL (SEM) ver. OR2102-9902

3-1

3. CAUTIONS
Take care of the followings in use of the machine: 1. The machine shall be operated by well trained technician only. 2. Cautions for installation
(1) Install the machine at place without fear of contact with water. (2) Install the machine at place without fear of adverse effects caused by abnormal

atmospheric pressure, temperature and/or humidity; insufficient ventilation, direct exposure to sunlight; and dust, salt and/or sulfur containing air.
(3) Handle the machine under avoiding slant placement, vibration and shock during

installation and transportation.


(4) DO NOT install the machine at chemical storing or gas generating place. (5) Frequency, voltage and amperage (power consumption) of power supply should be

as specified.
(6) Connect the machine with earth correctly.

3. Cautions prior to starting operation


(1) Check SW contacts, polarity, dial settings, meters and the like to assure correct

operation of the machine.


(2) Check correct connection of the machine with earth. (3) Check correct connections of all wires. (4) Take care that operation of the machine combined with other device may cause

adverse effects.
(5) Recheck integrity of lines connected to patient directly.

4. Cautions during operation


(1) Monitor always whether the machine and patient are maintained normally. (2) If trouble of the machine or patient is detected, take appropriate actions such as

the machine stop so that the patient is kept safe.


(3) Take care that there is no direct contact of the machine to patient.

NIPRO SURDIAL (SEM) ver. OR2102-9902

3-2 5. Cautions after operation


(1) Turn off power after returning operation SW's, dials and the like to original positions

based on prescribed procedures.


(2) DO NOT forcibly pull any cord and the like to disconnect.. (3) Take care the followings to store or transport the machine: 1) Store the machine at place without fear of contact with water. 2) Store the machine at place without fear of adverse effects caused by abnormal

atmospheric pressure, temperature and/or humidity; insufficient ventilation, direct exposure to sunlight; and dust, salt and/or sulfur containing air.
3) Handle the machine under avoiding slant placement, vibration and shock during

storage and transportation.


4) DO NOT store the machine at chemical storing or gas generating place. (4) Clean accessories, cords, 7 and the like in order and clean them to store. (5) Store the machine after cleaning it for the next use without any trouble.

6. If the machine is damaged, indicate the damaged conditions to request repair by specialist.

7. DO NOT remodel or modify the machine.

8. Check and maintenance


(1) Perform regular check and maintenance of the machine and its components. (2) When the machine is used after storage for some time, check and maintain so that

it can be operated normally and safely.

9. Read and understand other CAUTIONS contained in the operation manual.

NIPRO SURDIAL (SEM) ver. OR2102-9902

4-1

4. SYMBOL FOR SAFETY


a) The symbol shown below means that user of SURDIAL must read and understand this Service Manual before starting operation of the machine for safe and effective use of it.

b) The symbol shown below means that electric shock is caused by contact with high voltage applied live part.

NIPRO SURDIAL (SEM) ver. OR2102-9902

5-1

5. FLOW CHART
5.1. Fluid and blood lines
5.1.1. Flow chart

Fig. 2-1 Flow chart

NIPRO SURDIAL (SEM) ver. OR2102-9902

5-2 5.1.2. List of input components


Symbol T1 T2 T3 CD PD <PA> PV BLD LS1 LS2 LS3 LS5 PH1 PH2 PH3 PH4 PS FS FSW BPC BD Component Temperature sensor 1 Temperature sensor 2 Temperature sensor 3 Conductivity cell Dialysate pressure sensor Arterial pressure sensor Venous pressure sensor Blood leak sensor Temperature limit SW Coupler SW Syringe overload SW Clamp position sensor Vpa original point sensor Vpa rotation detection sensor VPb original point sensor Vpa rotation detection sensor Pressure SW Closed line flow sensor Float SW Blood pump cover SW Bubble sensor Function Supply water temp. measurement Heater outlet water temp. measurement Dialysate temp. measurement Dialysate conductivity measurement Dialysate line pressure measurement Arterial line pressure measurement Venous line pressure measurement Blood leak detection Heater surface temp. limiting Coupler detection Syringe overload limiting Clamp position detection Vpa original point detection Vpa rotation detection VPb original point detection VPb rotation detection Supply water pressure limiting Flow detection in closed line Level detection in AS2 BP cover open detection Bubble detection in venous line

NIPRO SURDIAL (SEM) ver. OR2102-9902

5-3 5.1.3 List of output components EMV: electromagnetic valve


Symbol V2 V3 V4 V5 V6 V7 V8 V9 V10 V11 V12 V13 V14 <V15> <V17> P1 P2 CLV VPa VPb H BP BZ1 BZ2 Lamp SP Component EMV for priming EMV for drainage EMV for AS2 EMV for fresh fluid in VCa EMV for drainage from VCa EMV for fresh fluid in VCb EMV for drainage from VCb EMV for bypassing dialysate EMV for gas discharge EMV for rinse EMV for disinfectant concentrate EMV for A concentrate EMV for B concentrate EMV for NaCI concentrate EMV for acetic acid concentrate Deaeration pump Dialysate pump Venous clamp Bellowphragm pump A Bellowphragm pump B Heater Blood pump Buzzer 1 Buzzer 2 Lamp Syringe pump Function Open/close of priming line Open/close of drainage line Open/close of upper line of AS2 Line changeover of fresh fluid side of A chamber Line changeover of drainage side of A chamber Line changeover of fresh fluid side of B chamber Line changeover of drainage side of B chamber Dialysate bypass line changeover Open/close of gas discharge line Open/close of rinse line Open/close of disinfectant concentrate line Open/close of A concentrate line Open/close of B concentrate line Open/close of NaCI concentrate line Open/close of acetic acid concentrate line Deaeration from supply water/fluid supply Dialysate supply Occlusion of venous line Transfer of viscous at A chamber side Transfer of viscous at B chamber side Heating of supply water Peristaltic blood supply Panel SW clicking Audible alarm Main lamp Syringe driving

NIPRO SURDIAL (SEM) ver. OR2102-9902

5-4 5.1.4. List of other components


Symbol AS1 AS2 BTW BTD1 BTD2 C1 C2 F FM NV1 NV2 PC1 PC2 PR R1 R2 RV1 RV2 RV3 SL VCa VCb <HEX> Component Deaeration bath1 Deaeration bath2 Buffer tank W Buffer tank D1 Buffer tank D2 Dialysate check valve Bellowphragm check valve Supply water filter Flow meter Priming flow rate control valve Dialysate flow rate control valve Bypass pinch cock AS2 pinch cock Supply water reducing valve Flow rate limiting connector 1 Flow rate limiting connector 2 Relief valve 1 Relief valve 2 Relief valve 3 Sampling port Viscous chamber A Viscous chamber B Heat exchanger Function Deaeration from supply water Deaeration from dialysate Deaeration acceleration Dialysate mixing 1 Dialysate mixing 2 Dialysate back flow prevention Negative pressure maintenance in Bellowphragm line Removal of foreign matters from supply water Dialysate flow rate measurement Priming flow rate control Dialysate flow rate control Bypass flow rate control Gas discharge pressure control Supply water pressure reduction Bypass flow rate limiting Gas discharge flow rate limiting Priming pressure control Dialysate pressure control Relief of excess supply water pressure Dialysate sampling VCS control VCS control Heat exchange between supply/drain water

NIPRO SURDIAL (SEM) ver. OR2102-9902

6-1

6. DISASSEMBLING AND REPLACEMENT


6.1. Input components
6.1.1. Conductivity cell Assy (CD)

Fig. 6.1.1-1 Conductivity cell Assy

No. 1 2 3 4 5 6

Component Sensor holder O-ring Pin Pan head Phillips screw (small) Hose port CD harness #9315 P3 #9315 M3 x 8 #9193 #9344

Requirement 38 11 104 1

n 1 3

Note

Silicone rubber

38 12 104

3 6 With spring/flat washers

42 12 002 71 32 100

3 0

1 1

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.1.2-1 6.1.2. Dialysate pressure sensor Assy (PD)

Sensor

Fig. 6.1.2-1 Dialysate pressure sensor Assy

No. Component 1 Sensor BKT 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cover Dialysate pressure sensor board Spacer Spring washer Small pan head Phillips screw Small pan head Phillips screw Small pan head Phillips screw Small pan head Phillips screw Manifold Holder Hose port Pressure sensor O-ring Small pan head Phillips screw

Requirement #9344 29 11 100 0 #9344 29 15 100 0 SBY-0159 BSB-310 For M3 M3 x 6 M3 x 6 M3 x 10 M3 x 6 #9344 29 12 100 0 #9344 29 13 100 0 #9193 42 12 002 3 P-8100-102G-10-SB P14 M3 x 10

n 1 1 1 4 4 4 2 3 2 1 1 2 1 1 2 Brass

Note

With spring/flat washers With spring/flat washers With spring/flat washers With spring/flat washers

Silicone rubber With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.1.3-1 6.1.3. Blood leak sensor Assy (BLD)

Fig. 6.1.3-1 Blood leak sensor Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Component Upper body cover Body casing Light receiver holder Light emitter holder Condensing lens Rear body cover O-ring (1) O-ring (2) Light emitter element Light receiver element Light receiver board Light emitter board Main board L-shaped hose port Knob ....... fight

Requirement TM-0005-02 TM-0006-02 TM-0007-02 TM-0008-02 TM-0009-01 TM-0010-02 TK-0016-01 TK-0017-01 TK-0018-01 TK-0019-01 TE-0002-01 TE-0003-01 TE-0004-01 #9328 39 12 001 4 TK-0021-01 TK-0022-01

n 1 1 1 1 2 1 1 1 1 1 1 1 1 2 2 6

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.1.4-1 6.1.4 Pressure SW: AW-3200-63-A (PS)

Fig. 6.1.4-1 Pressure SW

No. 1 2 3 4 5 6 7 8 9

Component Screw Bonnet Spring holder Spring Shaft Casing Diaphragm Valve body Screw

Requirement M3 x 35 D3C150 D3H210 D3K370 D3I1200 D3D140 D3A090 D3B1181 M3 x 30

n 2 1 1 1 1 1 1 1 2

Note With spring/flat washers

With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.1-1

6.2 Output components


6.2.1 Two way electromagnetic valve: AV-3202-318-SC (V2/V3/V12/V13/V14 (V15/V17))

Fig. 6.2.1-1 Two way electromagnetic valve

No. 1 2 3 4 5 6 7 8 9 10 Spring

Component Actuator Assy

Requirement D3K430

n 1 1 1 1

Note

Diaphragm holder Plunger Diaphragm Seat Spacer Valve body Screw Nut D3B1161 D3Y180

1 1 2 1 2 2

Plunger Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.2-1 6.2.2 Two way electromagnetic valve: AV-3202-63-S (V4/V11)

Fig- 6.2.2-1 Two way electromagnetic valve

No. 1 2 3 4 5 6 7

Component Actuator Assy Back-less pipe Spring Plunger Diaphragm Valve body Screw

Requirement D3Y080 D3I170 D3K340 D3I630 D3A090 D3B1181 M3 x 25

n 1 1 1 1 1 1 2

Note

With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.3-1 6.2.3 High vacuum electromagnetic valve: AV-3202-63-SH (V10)

Fig. 6.2.3-1 High vacuum electromagnetic valve

No. 1 2 3 4 5 6 7

Component Actuator Assy Back-less pipe Spring Plunger Diaphragm Valve body Screw

Requirement D3Y110 D3I990 D3K040 D3I1000 D3A090 D3B1181 M3 x 25

n 1 1 1 1 1 1 2

Note

With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.4-1 6.2.4. Three way electromagnetic valve: AV-3302-63-S (V5/V6/V7/V8)

Fig. 6.2.4-1 Three way electromagnetic valve

No. 1 2 3 4 5 6 7 8 9 10

Component Actuator Assy Push rod Diaphragm Valve body Stem Diaphragm Spring holder Spring Base Screw D3Y170 D3I660 D3A090 D3B990 D3I670 D3A090 D3H110 D3K150 D3F120 M3 x 50

Requirement

n 1 1 1 1 1 1 1 1 1 2

Note

With spring washer

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.5-1 6.2.5. Bellowphragm pump: VCP-60 (VPa/VPb)

Fig. 6.2.5-1 Bellowphragm pump Assy


No. 1 2 3 4 5 6 7 8 9 Component Upper part of cylinder Middle part of cylinder Lower part of cylinder BF diaphragm Piston BF diaphragm Sensor guide rod Rotation detection plate Drive shaft Requirement n 1 1 1 1 1 1 1 1 1 Applicable grease: Made by NOK. For ball and roller bearing (STABURAGS NBU 8 EP) 10 11 Photo sensor board Water proof cover SBY0176 1 1 Note

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.6-1 6.2.6. Blood pump Assy (BP/BPC)

Fig. 6.2.6-1 Blood pump Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.6-2

No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

Component Cover Hinge shaft Shaft A Shaft B BP rotor Roller Housing Rubber cap Packing Rubber sheet cover Magnet BKT (2) Pin Brush-less DC motor Bearing Limit SW Small pan head Phillips screw Small pan head Phillips screw Small pan head Phillips screw Small pan head Phillips screw Small pan head Phillips screw

Requirement #9314 61 23 114 0 #9193 61 16 010 0 #9314 61 17 185 0 #9314 61 19 102 0 #9314 61 16 168 1 #9314 61 18 188 0 #9331 61 11 002 2 #9314 61 27 100 1 #9314 61 28 197 0 #9314 61 12 001 0 #9314 61 29 219 0 #9314 61 13 237 0 FE6PF10G-314 687ZZ SS-01 M5 x 25 M5 x 12 M4 x 50 M2.3 x 10 M3 x 5

n 1 1 2 1 1 3 1 1 1 1 1 1 1 6 1 1 2 4 2 2 1 2 3 2 1 1 1 1 1 1

Note

Attached to harness side

Attached to motor With spring/flat washers With spring/flat washers

Set screw with hexagonal cavity M5 x 12 Set screw with hexagonal cavity M4 x 6 Set screw with hexagonal cavity M3 x 5 Guide Ring type magnet Magnet holder E-ring Seal Small pan head Phillips screw Receiver metal #9193 61 12 010 0 14 x 4 x 3 #9193 61 18 010 2 E5 #9193 61 21 010 0 M3 x 4 #9314 61 32 001 1

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.7-1 6.2.7. Micro gear pump Assy (P1/P2)

Fig. 6 2 7-1 Micro gear pump Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.7-2

No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32

Component Gear casing Mounting disk End disk Back casing Inner magnet Outer magnet Mounting plate Drive gear shaft Counter gear shaft Drive gear Counter gear Spring Teflon seal Gasket Motor flange Set screw with hexagonal cavity Small pan head Phillips screw Small pan head Phillips screw Small pan head Phillips screw Motor Hexagon headed bolt Spring washer Crown washer Small pan head Phillips screw Insulation collar Spring washer Capacitor cover Small pan head Phillips screw Capacitor Soundproof sheet Small pan head Phillips screw Hose port

Requirement

n 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

Note

M3 x 8 M2.5 x 4 M3 x 30

6 3 3 1 4 4 1 1 2 2 1 2 1 1 2 With spring/flat washers

#9193 42 12 002 3

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.8-1 6.2.8. Venous clamp Assy (CLV)

Fig. 6.2.8-1 Venous clamp Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.2.8-2

No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

Component Solenoid BKT Solenoid base Head guide Clamp guide Head Solenoid O-ring Spring Hexagon headed bolt

Requirement #9344 62 11 002 0 #9344 62 12 100 0 #9314 62 23 232 0 #9344 62 14 002 0 #9344 62 15 002 0 RGY-AC0 505DNP P6 No.1112 M3 x 8

n 1 1 1 1 1 1 2 1 3 7 4 1 1 1 1 1 1 1 1 2 2

Note

With spring/flat washers With spring/flat washers With spring washer

Small pan head Phillips screw M3 x 6 Small pan head Phillips screw M4 x 6 Hexagon nut Hexagon headed bolt Hexagon nut Sensor plate Hexagon headed bolt Hexagon nut Limit SW M5 M3 x 20 M3 #9314 62 16 001 1 M3 x 10 M3 SS-O1GL

Small pan head Phillips screw M3 x 6

With spring/flat washers

Attached to harness side With spring/flat washers

Small pan head Phillips screw M2.3 x 10 Small pan head Phillips screw M5 x 10

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.1-1

6.3. Other components


6.3.1. Deaeration bath Assy (AS1)

Fig. 6.3.1-1 Deaeration bath Assy

No. 1 2 3 4 5 6 7 8 9

Component Casing (1) Casing (2)-L Float Metal mesh O-ring Small pan head Phillips screw Hexagon nut Flat washer Small truss Philips screw

Requirement #9193 34 11 006 2 #9193 34 12 006 0 #9193 34 14 002 2 #9193 30 15 002 2 G50 M4 x 15 M4 For M4 M4 x 8

n 1 1 1 1 1 4 4 4 2

Note

Silicone rubber With spring/flat washers

With Nylon washer

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.2-1 6.3.2. Deaeration bath Assy (AS2)

Fig. 6.3.2-1 Deaeration bath Assy

No. 1 2 3 4 5 6 7 8 9 10 Casing Packinq

Component Casing (1)

Requirement #9193 34 11 006 2 #9331 24 12 134 0 #9191 10 19 001 2 #9193 30 15 002 2 #0193 97 00 069 1 G50 OLV-2P (special) M4 x 20 M4 M4 x 8

n 1 1 1 1 1 2 1 4 4 2

Note

Metal mesh Plate O-ring Float SW Small pan head Phillips screw Hexagon nut Small truss Phillips screw

Silicone rubber With spring/flat washers With Nylon washer

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.3-1 6.3.3. Buffer tank W Assy (BTW)

Fig. 6.3.3-1 Buffer tank W Assy

No. 1 2 3 4 5 6 7 8

Component Lower tank Upper tank (B) BT bracket O-ring Small pan head Phillips screw Hexagon nut Hexagon headed Phillips screw Hexagon headed Phillips screw

Requirement #9193 69 37 004 3 #9193 69 55 004 2 #9344 35 13 101 0 G70 M4 x 20 M4 M4 x 10 M4 x 10

n 1 1 1 1 4 4 2 2

Note

Silicone rubber With spring/flat washers With spring/flat washers With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.4-1 6.3.4. Buffer tank D Assy (BTD1/BTD2)

Fig. 6.3.4-1 Buffer tank D Assy

No. 1 2 3 4 5 6 7 8

Component Lower tank Upper tank (8) BT bracket O-ring Small pan head Phillips screw Hexagon nut Small truss Phillips screw

Requirement #9193 69 37 004 3 #9193 69 55 004 2 #9344 35 13 101 0 G70 M4 x 20 M4 M4 x 8

n 1 1 1 1 4 4 2 2

Note

Silicone rubber With spring/flat washers With spring/flat washers With Nylon washer

Hexagon headed Phillips screw M4 x 10

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.5-1 6.3.5. Check valve 6.3.5.1. Dialysate check valve: #919342300041(AC-2200-55-S) (C1)

Fig. 6.3.5-1 Dialysate check valve

No. 1 2 3 4 5 6

Component Body Seat holder Seat Spring Lock nut O-ring Sub-body

Requirement ACB060 ACH090 ACL101 ACK290 ACH080 P9DC ACB070

n 1 1 1 1 1 1 1

Note

Check valve seat Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.5-2 6.3.5.2. Bellowphragm check valve (C2)

Fig. 6.3.5-2 Bellowphragm check valve

No. 1 2 3 4 5 6 Casing

Component Diaphragm Spring receiver Spring Valve body Hose port

Requirement ACC050 D2A181 ACH130 ACK220 ACB140 #9193 42 12 002 3

n 1 1 1 1 1 2

Note

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.6-1 6.3.6. Supply water filter (F)

Fig. 6.3.6-1 Supply water filter

No. 1 2 3 4 5 6 Cap Mesh

Component

Requirement #9193 42 20 002 2 #9193 42 22 002 1 #9314 20 17 002 0 CF200 #9193 42 21 002 3 #9314 59 21 107 0

n 1 1 1 1 1 1

Note

Packing Filter Body Arrow mark seal

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.7-1 6.3.7. Flow meter Assy (FM/FS)

Fig. 6.3.7-1 Flow meter Assy

No. 1 2 3 4 5 6 7 8 Flow meter

Component L-shaped hose port Hose port Bracket Small pan head Phillips screw Photo-micro sensor Hexagon headed Phillips screw Hexagon headed Phillips screw

Requirement S1-1AB #9328 39 12 001 4 #9193 42 12 002 3 #9344 31 13 100 2 M3 x 6 EE-SPY-412 M3 x 16 M4 x 10

n 1 1 1 1 2 1 2 2 FM

Note

With spring/flat washers FS With spring/flat washers With pring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.9-2 6.3.9.2. Relief valve 3: AC-3200-63H-T (RV3)

Fig. 6.3.9-2 Relief valve 3

No. 1 2 3 4 5 6 7 8

Component Valve body Seat Packing Seat holder Spring Lock nut Casing Name plate

Requirement

n 1 1 1 1 1 1 1 1

Note

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.10-1 6.3.10. Sampling port Assy (SL)

Sampling port

Fig. 6.3.10-1 Sampling port Assy

No. 1 2 3 4 5 6 7 8

Component Packing Bracket Syringe holder Packing (B) Valve Spring Body

Requirement #9331 34 #9344 34 #9331 34 #9193 69 #9331 34 #9193 88 #9331 34 18 16 12 53 17 67 11 100 0 100 0 110 0 004 2 100 0 004 1 100 2

n 1 1 2 1 1 1 1 1

Note

Small pan head Phillips screw M3 x 8

With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.8-1 6.3.8. Reducing valve Assy (PR)

Fig. 6.3.8-1 Reducing valve Assy

No. 1 2 3 4 5

Component Reducing valve Flange PR hose port Hose port Small truss Phillips screw

Requirement 8.RW83 #9344 21 13 100 0 #9344 21 14 100 0 #9331 21 16 129 0 M4 x 8

n 1 1 1 1 2

Note

With Nylon washer

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.9-1 6.3.9. Relief valve 6.3.9.1. Relief valve 1/2: AC-2200-55-TA / AC-2200-55-P (RV1/RV2)

Fig. 6.3.9-1 Relief valve 1/2

No. 1 2 3 4 5 6 Body

Component

Requirement ACB060 ACH090 ACL010 ACH080 P9DC ACB070

n 1 1 1 1 1 1 1

Note

Seat holder Seat Spring Lock nut O-ring Sub-body

Check valve seat Assy

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.11-1 6.3.11. Chamber Assy (VCa / VCb) 6.3.11.1. Chamber Assy (overall)

Fig. 6.3.11-1 Chamber Assy (overall)

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.11-2

No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Component Chamber frame Rubber sheet Sheet Bellowphragm pump T-pipe Viscous plug Polyurethane tube Shapper Hexagon head Phillips bolt T-pipe Silicone tube Shapper Nylon string Hexagon head Phillips bolt Hexagon head Phillips bolt

Requirement #9344 10 11 100 0 #9314 10 36 001 0 #9331 10 28 002 0 VCP-60 #9328 29 11 001 3 #9193 11 11 010 0 US code 15-10 SNP-4 M4 x 15 198-19-21-02E 6 1x3 SNP1.510C2 T18R M4 X 10 M4 x 15

n 1 1 2 2 2 2 10 8 1 2 3 4 4

Note

with spring/flat washers

With spring/flat washers With spring/flat washers

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.11-3 6.3.11.2. Chamber Assy (Viscous chamber)

Fig. 6.3.11-2 Chamber Assy (Viscous chamber)

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.11-4

No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Diaphragm

Component Inner spacer Outer spacer Chamber Silicone oil Clamp Clamp fitting Shaft E-ring Bolt with thrust pad T-pipe Hose port Orifice Silicone tube Valve bracket Plate Hose port with flange O-ring 3 way electromagnetic valve Small pan head Phillips screw Hexagon headed bolt Spring washer Snapper

Requirement #9314 10 12 001 1 #9314 10 13 001 7 #9314 10 14 001 7 #9344 10 17 100 1 THF450-100* #9314 10 15 102 0 #9314 10 39 105 0 #9314 10 38 002 0 E-6 BJ746-08001 #9331 10 25 100 1 #9314 39 19 100 0 #9331 10 29 100 0 5 x 11 x 120L #9344 10 15 002 0 #9193 10 20 011 0 #9314 10 35 207 0 P10 AV-3302-63-S M4 x 6 M4 x 50 for M4 SNP-4

n 4 2 2 4 4 2 4 8 2 4 4 2 4 2 4 4 4 4 8 8 8 2

Note

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.12-1 6.3.12.2. Concentrate tank connector (for disinfectant concentrate)

Fig. 6.3.12-2 Concentrate tank connector (for disinfectant concentrate)

No. 1 2 3 4 5 6 7

Component Connector holder Packing Holder base Nut Silicone tube .......... FEP tube Color seal

Requirement #9193 69 40 004 4 #9193 69 44 004 3 #9193 69 41 004 3 #9193 33 11 010 1 5 x 11 x 50L TF185 #9193 59 31 010 2

n 1 1 1 1 1 1 1

Note

6 x 8 x 330L Green

NIPRO SURDIAL (SEM) ver. OR2102-9902

6.3.12-2 6.3.12. Concentrate tank connectors 6.3.12.1. Concentrate tank connector (for A and B concentrates)

Fig. 6.3.12-1 Concentrate tank connector (for A and B concentrates)

No. 1 2 3 4 Socket

Component

Requirement QJ-2P,NPT1/8 5 x 11 x 50L TF185 #9193 59 32 010 2 #9193 59 33 010 2

n 1 1 1 1 1

Note

Silicone tube ..FEP tube Color seal

6 x 8 x 330L Pink for A concentrate Orange for B concentrate

NIPRO SURDIAL (SEM) ver. OR2102-9902

7.1

7. CONSTRUCTION OF ELECTRICAL PART


7.1. Components of electrical part

Fig. 7.1-1 Electrical part

7.1.1. Components at upper part of electrical part Operation panel Operation panel board LCD module Bubble sensor board Electrical POWER BOX

7.1.2. Component at lower part of electrical part Diaysate pressure sensor board

NIPRO SURDIAL (SEM) ver. OR2102-9902

7.2 7.1.3. Placement of electrical part at machine

Fig. 7.1-2 Placement of electrical part

NIPRO SURDIAL (SEM) ver. OR2102-9902

7.3 7.1.4. Components at left side in electrical POWER BOX

Fig.7.1-3 Components al left side in electrical POWER BOX 7.1.5. Components at right side in electrical POWER BOX

Fig. 7.1-4 Components at right side in POWER BOX

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.1.1

8. CONNECTION, REPLACEMENT AND ADJUSTMENT


8.1. Operation panel sheet

Fig. 8.1-1 Operation panel

[General description and application]


Key SW's are encased in sheet of the operation panel. Adhesive tape placed at rear side of the panel sheet is applied to front side of upper part of machine.

[Connection]
Connect flexible cable to CN3 of [operation panel board].

[Adjustment]
Not required

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.1.2

[Replacement]
(1) Turn off power SW of machine. (2) Remove left side cover from upper part of machine (3) Disconnect flexible cable from connector by unlocking it.

Fig. 8.1-2 Disconnection of flexible cord


(4) Defoliate operation panel sheet from machine. (5) Prepare new panel sheet. (6) Clean place for the sheet to be applied. (7) Remove release paper from rear side of the sheet. (8) Apply the sheet while passing flexible cable through hole of machine. (9) Engage the cable with connector followed by locking it. (10) Turn on power SW of machine. (11) Check correct function of each SW of operation panel. (12) Return the removed cover to the original position.

CAUTION

Replace operation panel sheet after turning off power SW under care not to be charged with static electricity. Handle flexible cable carefully so that it is not folded and/or damaged. Otherwise malfunction may be caused. DO NOT apply excess external force to operation panel because key SW's and patterns encased in the panel may be folded and/or damaged, resulting in malfunctioning of them, in particular when the panel itself is stored or handled.

DO NOT reused removed panel sheet. Disconnect flexible cable from connector after unlocking it. If the cable is pulled out under unlocking the connector, electrodes placed at the connector surface may be separated, causing machine failure.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.2.1

8.2. Operation panel board

Fig. 8.2-1 Operation panel board

[General description and functions]


Operation panel board is positioned at rear side of operation panel placed at upper part of machine. Contrast of LCD is adjusted by VR1 of the board. Flow rate of blood pump is set by VR2 of the board. Clicking is caused by pressing each SW of operation panel (BZ1). Each LED of operation panel is ON or OFF according lo executed SW operation.

[Connection]
CN1: Connect to CN22 of [main board] through harness. CN2: Connect to CN10 of [main board] through harness. CN3: Connect to flexible cable of [operation panel). CN4: Connect to CN2 of [LCD module] through harness. CN5: Connect to CN1 of [LCD module].

[Adjustment]
See 10.17. Adjustment of contrast.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.2.2

[Replacement]
(1) Turn off power SW of machine. (2) Remove left cover from upper part of machine. (3) Disconnect all connectors. (4) Remove green cap from upper part of knob of blood pump flow rate setting volume placed at front side of machine and loosen exposed screw by screwdriver under pressing the knob by hand to remove it. (5) Remove board fixing nuts from machine inside to remove the board. (6) Prepare new board. (7) Remove seal, if provided, from buzzer (BZ1). (8) Attach the new board. (9) Attach knob of blood pump flow rate setting volume under care of correct 0 point positioning. (10) Connect the all disconnected connectors. (11) Turn on power SW of machine. (12) Try to adjust contrast as per the procedure specified in 10.17. Adjustment of contrast. (13) Check correct functioning of each LED. (14) Check correct functioning of each SW of operation panel. (15) Return the removed cover to the original position.

CAUTION

Replace operation panel board after turning off power SW under care not to be charged with static electricity. It is recommended to replace the board after cooling to room temperature because temperature of resistors placed on the board immediately tuning off power is high, potentially causing burn. Handle flexible cable carefully so that it is not folded and/or damaged. Otherwise malfunction may be caused. Disconnect flexible cable from connector after unlocking it. If the cable is pulled out under unlocking the connector, electrodes placed at the connector surface may be separated, causing machine failure. Unlock as per the procedure specified in 8.1 Operation panel. LCD may not function normally after the board replacement. If so, adjust contrast.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.3.1

8.3. LCD module

Fig. 8.3-1 LCD module

[General description and construction]


LCD module is constructed by LCD, back-light and board. LCD is STN type and can indicate 40 letters in each of 4 lines. Back-light is constructed by LED's. CN1 is connector for back-light and is soldered to board.

CN2 is connector to transmit signal to LCD.

[Connection]
CN1: Connect to CN5 of [operation panel board]. CN2: Connect to CN4 of [operation panel board] through harness.

[Adjustment]
See 10.17 Adjustment of contrast.

[Replacement]
(1) Turn off power SW of machine. (2) Remove left cover from upper part of machine. (3) Disconnect CN1 and CN2 from LCD module. (4) Remove 4 fixing screws or nuts to take out the module. (5) Prepare new LCD module. (6) Remove protective sheet from the new module surface. (7) Attach the module under care of correct positioning of UP mark on the board. (8) Connect CN1 and CN2 to LCD module. (9) Turn on power SW of machine. ( 1 0 ) Try to adjust contrast as per the procedure specified in 10.17 Adjustment of contrast.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.3.2

CAUTION

DO NOT apply excess mechanical impact to LCD because it is constructed by glass. If broken, take care not to be injured by the glass fragments. If liquid crystal is leaked from damaged LCD and your skin or cloth is contaminated with the crystal, rinse out using soap immediately. Also oral contamination with it should be avoided. DO NOT scratch polarizing plate covering LCD because the plate is composed of soft material. Replace the module after turning off power SW under care not to be charged with static electricity. DO NOT try to disassemble LCD module because it is integrated as one unit. After the replacement, LCD may not function correctly. If so, try to adjust contrast.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.4.1

8.4. Bubble sensor board

Fig. 8.4-1 Bubble sensor board

[General description and function)


Bubble sensor board is placed at upper part of machine. The board is used to detect bubble in blood line in cooperation with bubble sensor.

[Connection]
CN601: Connect to connector of [bubble sensor]. CN602: Connect to CN3 of [main board] through harness.

[Adjustment]
See 10.15 Adjustment of bubble sensor.

[Replacement]
(1) Turn off power SW of machine. (2) Remove left cover from upper part of machine. (3) Disconnect all connectors. (4) Remove bubble sensor board under pressing spacer lock used to fix it. (5) Prepare new bubble sensor board. (6) Attach the new board and check correct locking. (7) Connect all the disconnected connectors. (8) Turn on power SW of machine. (9) Adjust the sensor as per the procedure specified in 10.15 Adjustment of bubble sensor. (10) Check correct function of bubble sensor. ( 1 1 ) Return the removed cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.5.1

8.5. Dialysate pressure sensor board

Fig. 8.5-1 Dialysate pressure sensor board

[General description and function]


Dialysate pressure sensor board is placed at top of lower part of machine. The board is used to measure pressure in dialysate line in cooperation with dialysate pressure sensor.

[Connection]
CN701: Connect to connector of [dialysate pressure sensor]. CN702: Connect to CN14 of [main board] through harness.

[Adjustment]
See 10.3. Adjustment of dialysate pressure sensor.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.5.2

[Replacement]
(1) (2) (3) Turn off power SW of machine. Remove left cover from upper part of machine and left cover from lower part of it. Remove dialysate pressure sensor ASSY as per the procedure specified in 6.1.2 Dialysate pressure sensor ASSY. (4) Disconnect all connectors. (5) Loosen screws used to fix dialysate pressure sensor board to remove it. (6) Prepare new dialysate pressure sensor board. (7) Attach the new board. (8) Connect all the disconnected connectors. (9) Attach dialysate pressure sensor ASSY. (10) Turn on power SW of machine. (11) Adjust dialysate pressure sensor as per the procedure specified in 10.3. Adjustment of dialysate pressure sensor. (12) Check correct function of dialysate pressure sensor. (13) Return the removed covers to the original positions.

CAUTION

Replace dialysate pressure sensor board after turning off power SW of machine under care not to be charged with static electricity.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.6.1

8.6. Main board

Fig. 8.6-1 Main board

[General description and functions]


Main board is attached to right side of DENSO box. Machine operation is controlled by microprocessor contained in the board. Various set values are back-upped by RAM and EEROM. Version-up is enabled by replacing ROM with new one. A part of functions can be selected by operating DIP SW. Venous pressure sensor board can be attached (V.PRESS PCB). Also arterial pressure sensor board as option can be attached (A.PRESS PCB).

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.6.2

[Connection]
CN1: Connect to CN413 of (sub-board] through harness. CN2: Connect to [blood leak sensor] through harness. CN3: Connect to CN602 of [bubble sensor board] through harness. CN4: Not connected. CN5: Connect to CN408 of [sub-board] through harness. CN6: Not connected. CN7: Connect to [photo-sensor board of Bellowphragm pump A] through harness. CN8: Connect to [photo-sensor board of Bellowphragm pump B] through harness. CN9: Connect to CN502 of [DSUB board] through harness. CN10: Connect to CN2 of [operation panel board] through harness. CN11: Connect to CN701 of [arterial pressure sensor board] through harness. CN12: Connect to V3,V4,V9,V10,V12 and V17 through harnesses. CN13: Connect to CN701 of [venous pressure sensor board] through harness. CN14: Connect to CN701 of [dialysate pressure sensor board] through harness. CN15: Connect to LS2 and FS through harnesses. CN16: Connect to harness of LS3. CN17: Connect to LAMP through harness. CN18: Connect to harness of CLV. CN19: Connect to PS, FSW and LS1 through harnesses. CN20: Connect to V5,V6,V7 and V8 through harnesses. CN21: Connect to V2,V11,V13,V14 and V15 through harnesses. CN22: Connect to CN1 of [operation panel board] through harness. TM1: Connect to temperature sensors (T1,T2 and T3) and concentration sensor.

[Adjustment]
See 10.16. Adjustment of concentration.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.6.3

[Replacement] Main board


(1) Turn on power SW of machine and record all set values. (2) Turn off power SW of machine. (3) Remove right cover from upper part of machine and right cover from DENSO box. (4) Disconnect all connectors. (5) Remove venous pressure sensor board; and arterial pressure sensor board as option, if provided, as per the procedure specified in "8.7. Venous and arterial pressure sensor boards". (6) Remove main board under pressing spacer lock used to fix it. (7) Prepare new main board. (8) Attach the new board and check correct locking. (9) Connect all the disconnected connectors. (10) Turn on power SW of machine. (11) Input all the set values recorded at step (1). (12) Adjust concentration as per the procedure specified in "10.16. Adjustment of concentration". (13) Check that various analog sensor values indicated in LCD are correct. (14) Check correct function of machine. (15) Return the removed covers to the original positions.

ROM
(1) (2) (3) (4) (5) (6) (7) (8) (9) Turn on power SW of machine and record all set values. Turn off power SW of machine. Remove right cover from upper part of machine and right cover from DENSO box. Remove ROM using tool specially designed for the removal. Prepare new ROM. Attach the new ROM under care that it is positioned correctly and pins of it are not damaged. Turn on power SW of machine. If ROM of different version is attached, initialization alarm is actuated. Input all the set values recorded at step (1). procedure specified in "10.16. Adjustment of concentration". (11) Check correct function of machine. (12) Return the removed covers to the original positions.

(10) If EEPROM initialization alarm is actuated, adjust concentration as per the

Fig. 8.6-2 Attachment of ROM

CAUTION
Replace main board or ROM after turning off power SW of machine under care not to be charged with static electricity. After the board replacement, re-input set values. It should be noted that the values found in maintenance modes 5 and 6 are specific to machine.
NIPRO SURDIAL (SEM) ver. OR2102-9902

8.7.1

8.7. Venous and arterial pressure sensor boards

Fig. 8.7-1 Venous/arterial pressure sensor board

[General description and function]


Venous or arterial pressure sensor board is mounted on main board placed at right side of DENSO box. Venous or arterial pressure in venous or arterial line is monitored by the board in cooperation with pressure sensor (PSD701). Venous and arterial pressure sensor boards are same design. Application to venous or arterial pressure monitoring is determined by difference of mounting positions of the both boards on main board and difference of piping routes to pressure inlet port. Generally, only venous pressure is monitored. If arterial pressure monitoring is required, use of arterial pressure sensor ASSY (option) is required.

[Connection]
When used as venous pressure sensor: CN701: Connect to CN13 of [main board] through harness. When used as arterial pressure sensor: CN701: Connect to CN11 of [main board) through harness.

[Adjustment]
See 10.2. Adjustment of venous and arterial pressure sensors.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.7.2

[Replacement]
(1) Turn off power SW of machine. (2) Remove right cover from upper part of machine and right cover from DENSO box. (3) Disconnect all connectors. (4) Remove snapper to take out tube. (5) Remove venous or arterial pressure sensor board under pressing spacer lock used to fix it. (6) Prepare new venous or arterial pressure sensor board. (7) Attach the new board and check correct locking. (8) Attach tube and snapper. (9) Connect all the disconnected connectors. (10) Turn on power SW of machine. (11) Adjust venous or arterial pressure sensor as per the procedure specified in 10.2. Adjustment of venous and arterial pressure sensors. (12) Check correct indication of venous or arterial pressure. (13) Return the removed covers to the original positions.

CAUTION

Replace venous or arterial pressure sensor board after turning off power SW of machine under care not to be charged with static electricity. Connect tube to left side of sensor and expose right side of it to atmosphere. If connected to the right side, measured pressure is incorrect. After connecting tube to sensor, attach snapper. If not attached, tube may be disconnected by pressure application, resulting in blood inflow. Connect tubes for venous and arterial pressure monitoring to respective boards correctly.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.8.1

8.8. DSUB board

Fig. 8.8-1 DSUB board

[General description and functions]


DSUB board is placed at right side of DENSO box. The board is used to control blood pump flow rate at power failure when battery (VR501) as option is provided. The board is provided with connector (CN501) to communicate data with external monitor as option. The board is also used as junction to power connector of external monitor as option. The board is additionally used as junction to buzzer board.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.8.2

[Connection]
CN501: Connect to external monitor (option). CN502: Connect to CN9 of [main board] through harness. CN503: Connect to CN402 of [sub-board] through harness. CN504: Connect to CN810 of [buzzer board] through harness.

[Adjustment]
Not required

[Replacement]
(1) Turn off power SW of machine. (2) Remove right cover from upper part of machine and right cover from DENSO box. (3) If battery (option) is provided, disconnect its connector. (4) Disconnect all connectors. (5) Remove knob from BP BACKUP volume. (6) Remove spacer used to fix DSUB board and nuts used to fix the volume to take out the board. (7) Prepare new DSUB board. (8) Attach the new board. (9) Return the removed knob to BP BACKUP volume. (10) Connect all the disconnected connectors. (11) If battery (option) is used, connect its connector. Turn BP BACKUP volume to check correct blood pump operation under no power supply. (12) Turn on power SW of machine. (13) Check correct function. (14) Return the removed covers to the original positions.

CAUTION

Replace DSUB board after turning off power SW of machine under care not to be charged
with static electricity.

After the replacement, set BP BACKUP volume to OFF position. DO NOT miss parts used to fix. Spacer is fixed by nuts at its rear side.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.9.1

8.9. Buzzer board

Fig. 8.9-1 Buzzer board

[General description and function]


Buzzer board is placed at right side of DENSO box. The board is used to alert by buzzing (BZ801).

[Connection]
CN801: Connect to CN504 of [DSUB board] through harness.

[Adjustment]
Not required

[Replacement]
(1) (2) (3) (4) (5) (6) (7) (8) (9)

Turn off power SW of machine. Remove right cover from upper part of machine and right cover from DENSO box. Disconnect all connectors. Remove buzzer board under pressing spacer lock used to fix it. Prepare new buzzer board. Attach the new board and check correct locking. Connect all the disconnected connectors. Turn on power SW of machine. Make error to check buzzing.

(10) Return the removed covers to the original positions.

CAUTION

Replace buzzer board after turning off power SW of machine under care not to be charged with static electricity.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.10.1

8.10. Sub-board

Fig. 8.10-1 Sub-board

[General description and function]


Sub-board is placed at left side of DENSO box. The board is provided with drivers for blood, syringe and Viscous pumps (BP PCB,SP PCB, VPa PCB, VPb PCB). The board is used to operate AC driven components such as pump and heater.

[Connection]
CN401: Connect to P1 and P2 through harnesses. CN402: Connect to CN503 of [DSUB board] through harness. CN403: Connect to harness of BPC. CN404: Used to connect AC power supply to sub-board through harness. CN405: Used to connect AC power supply to sub-board through harness. CN406: Connect to FAN. CN407: Connect to SSR and relay for heater control through harnesses. CN408: Connect to CN5 of [main board] through harness. CN409: Connect to Vpa and VPb through harnesses. CN410: Connect to SP through harness. CN411: Used to connect DC+24V supply to sub-board through harness. CN412: Used to connect DC+5V, DC+12V and DC-12V supplies to sub-board through harnesses. CN413: Connect to CN1 of [main board] through harness. CN414: Connect to BATTERY. CN415: Connect to BP through harness.

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.10.2 TAB401: Connect to frame through harness. TAB402: Components are attached to the board of the part of AC110V machine. Not applicable when rating is AC230V. TAB403: Components are attached to the board of the part of AC110V machine. Not applicable when rating is AC230V.

[Adjustment]
Not required

[Replacement]

Sub-board
(1) Turn off power SW of machine. (2) Disconnect power plug from receptacle. (3) Remove left cover from upper part of machine and left cover from DENSO box. (4) Disconnect all connectors. DO NOT disconnect TAB402, TAB403 and CN416 when provided. (5) Remove sub-board under pressing spacer lock used to fix it. (6) Prepare new sub-board. Two types of the board shown below are available; one is for AC100V and other is for 230V. Select the type conforming to the rating. Board No. is silk screen printed at upper right side of sub-board. SBY0157 is for AC230V and SBY0163 is for AC110V. Unnecessary Nos. are soldered. And also, concerning SBY0163 for AC110V, components are attached to CN416.

SBY0157 for AC230V


(7) Attach the new board and check correct locking. (8) Connect all the disconnected connectors. (9) Insert power plug in receptacle. (10) Turn on power SW of machine. (11) Check correct function of machine. (12) Return the removed covers to the original positions.

SBY0163 for AC110V

Fig. 8.10-2 Two types of sub-board

NIPRO SURDIAL (SEM) ver. OR2102-9902

8.10.3

Fuse
(1) Turn off power SW of machine. (2) Disconnect power plug from receptacle. (3) Remove left cover from upper part of machine and left cover from DENSO box. (4) Remove fuse. (5) Prepare new fuse. Select fuse to be replaced by referring to the following purposes and
ratings of fuses:

FUSE402: AC250V 3A FUSE404: AC250V 3A FUSE405: AC250V 5A

P1 pump protection P2 pump protection BP driver protection

FUSE403: AC250V 3A RY402 detect power failure relay protection

(6) Attach the new fuse. (7) Insert power plug in receptacle (8) Turn on power SW of machine (9) Check correct function of machine. (10) Return the removed covers to the original positions.

CAUTION

Replace sub-board or fuse after disconnecting power plug from receptacle. Only power SW OFF is insufficient to avoid danger because AC supply placed at left side in DENSO box is not disconnected, potentially causing electric shock. Handle sub-board under care not to be charged with static electricity. Power rating of sub-board should conform to that of power source of machine. Otherwise, fire and/or machine failure may be caused. Use fuse of correct rating. Otherwise, fire and/or machine failure may be caused.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.1.1

9. MAINTENANCE MODES
9.1 Display

LCD is used to display various dialysis related information and setting screens by 40 letters in each of 4 lines. Many information and setting screens are available and desired screen can be displayed by SW operation on operation panel.

Fig. 9.1-1 Screen flow

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.1.2

CAUTION

LCD is easily scratched. DO NOT press or scratch its surface by pencil point or hard tool. Glass is mainly used in LCD. DO NOT apply mechanical impact. DO NOT press operation panel SW by pencil point or hard tool.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.2.1

9.2. Maintenance modes


Dialysis related data indication, alarm setting, alarm release, machine function monitoring or concentration calibration can be executed in selected maintenance mode. 10 maintenance modes are available. There are mode to indicate information only, mode enabling setting at any time required and mode enabling setting under specified condition only.
MODE 1 2 3 4 5 6 7 8 9 10 DATA ALARM SETTING ALARM FORCED OFF FUNCTION MONITOR CONCENTRATE DATA CHARACTERISTICS SYSTEM SERVICE FREQUENCY MACHINE FILLING PROCEDURE ALARM HISTORY TITLE Indication only Enabled setting at any time required Enabled setting at any time required Enabled setting under specified condition Enabled setting under specified condition Enabled setting under specified condition Enabled setting under specified condition Indication only Enabled setting at any time required Indication only SETTING

Indication only
Mode to indicate information only. Setting is disabled.

Enabled setting at any time required


Mode to enable setting or maintenance operation at any time required with exception that the setting and maintenance operation are disabled according to specified operating condition of machine.

Enabled setting under specified condition


Mode protected against setting change. The protect can be released by the following procedure to enable setting. The protect released condition is called maintenance mode free. (1) Select maintenance mode 7 in maintenance mode menu screen.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.2.2 (2) Press [F-3] or [F-4] SW to display 20. MAINTENANCE MODE FREE.

(3) Press [F-1] SW to set to ON.

(4) Operation SW's [+]/[-] are displayed above [F-1][F-2] and items enabled to be set blink.

DANGER

Setting or setting change of item to be maintained during dialysis is danger. DO NOT set or change item to be maintained under the condition that patient is connected to machine. DO NOT change set value without direction of doctor or technician. Protect modes 2, 4, 5, 6 and 7 to prevent malfunction of machine always by setting 20.MAINTENANCE MODE FREE to OFF. Release the protection only when setting change is required.

CAUTION

If ROM of different version is loaded, some maintenance values may be initialized. In such case, reset the values. When ROM or main board is replaced, or CPU is reset, set maintenance mode values followed by checking them. It is recommended to record maintenance mode values. Use supplement of this service manual to record them.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.3.1

9.3. Maintenance mode menu screen


In maintenance mode screen, selection and registration of desired maintenance mode are enabled. Press [F-3] or [F-4] SW to changeover screen and press [F-1] or [F-2] SW to select maintenance mode as desired. Press [F-5] SW to display rinse operation or dialysis operation screen. If alarm is actuated, alarm screen is displayed.

Fig. 9.3-1 Maintenance mode menu screen

INDICATION MODE1: DATA MODE2: ALARM SETTING MODE3: ALARM FORCED OFF MODE4: FUNCTION MONITOR MODE5: CONCENTRATE DATA MODE6: CHARACTERISITICS MODE7: SYSTEM MODE8: SERVICE FREQUENCY MODE9: MACHINE FILLING PROCEDURE MODE10: ALARM HISTORY

DESCRIPTION Dialysis related data indication mode Alarm setting mode Alarm release mode Function monitor and manual operation mode Concentration calibration mode Instrument adjustment mode System setting mode Operation frequency indicating mode Air purge mode Alarm history indicating mode

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.4.1

9.4. Maintenance mode 1 screen


In maintenance mode 1 screen, real time indication of various data during dialysis operation is executed. In the mode, only data indication is enabled and setting is disabled. Press [F-3] or [F-4] SW to check data. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.4-1 Maintenance mode 1 screen


INDICATION 1. TREATED BLOOD VOLUME 2. TREATMENT TIME 3. SN AVG.BLOOD FLOW 4. SN STROKE VOLUME 5. BLD REAL VALUE 6. BLD AUTO ZERO 7. BLD SENS VALUE 8. BLD RED RIGHT 9. BLD GREEN RIGHT 10.VCa UF LEAK NEG.PR 11.VCb UF LEAK NEG.PR 12.VCa UF LEAK POS.PR 13.VCb UF LEAK POS. PR 14.CONDUCTIVITY 15.CONCENTRATION 16.VENOUS PRESSURE 17.ARTERIAL PRESSURE 18.DIALYSATE PRESSURE 19.TMP 20.UF.COEFF 21.TEMP.SET VALUE 22.TEMPERATURE (T3) 23.TEMPERATURE (T2) DESCRIPTION Treated blood volume indication Executed dialysis time indication Mean SN blood flow rate indication One stroke SN1 blood volume indication Leaked blood level indication Bood leak auto zero value indication Blood leak SENS value indication Blood leak red value indication Blood leak green value indication Leaked closed circuit resulted A chamber negative pressure indication Leaked closed circuit resulted B chamber negative pressure indication Leaked closed circuit resulted A chamber positive pressure indication Leaked closed circuit resulted B chamber positive pressure indication Conductivity indication Concentration indication Venous pressure indication Arterial pressure indication Dialysate pressure indication TMP indication UFRP indication Set temperature indication Dialysate temperature T3 indication Heater outlet temperature T2 indication

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.5.1

9.5. Maintenance mode 2 screen


In maintenance mode 2 screen, alarm value can be set. When setting is protected, indication only is enabled and setting is disabled. Press [F-3) or [F-4] SW to check data. If setting is not protected, press [F-1] or [F-2] SW to change set value. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.5-1 Maintenance mode 2 screen


INDICATION 1.V.PRESS MAX RANGE 2.V.PRESS MIN RANGE 3.A.PRESS MAX RANGE 4.A.PRESS MIN RANGE 5.D.PRESS MAX RANGE 6.D.PRESS MIN RANGE 7.TMP(I) 8.CONDUCTIVITY MAX 9.CONDUCTIVITY MIN 10.TEMPERATURE MAX 11.TEMPERATURE MIN 12.BLOOD LEAK 13.UF MAX AGAINST BLF 14.SN HIGH/LOW RANGE 15.SN STROKE TIME 16.VCS UF LEAK TEST 17.UF LEAK VALUE a 18.UF LEAK VALUE b 19.UF LEAK TIME c 20.UF LEAK TIME d 21. RESET SW TIME 22.DIALYSIS SW TIME 23 BYPASS SW TIME DESCRIPTION Auto-setting of upper venous pressure alarm limit Auto-setting of lower venous pressure alarm limit Auto-setting of upper arterial pressure alarm limit Auto-setting of lower arterial pressure alarm limit Auto-setting of upper dialysate pressure alarm limit Auto-setting of lower dialysate pressure alarm limit Setting of TMP alarm limit Setting of upper conductivity alarm limit Setting of lower conductivity alarm limit Setting of upper dialysate temperature alarm limit Setting of lower dialysate temperature alarm limit Setting of blood leak alarm limit Setting of upper UF alarm limit Setting of upper and lower SN alarm limits Setting of SN stroke time alarm limit Setting of leak detection level from closed circuit Setting of leak detection level a from closed circuit Setting of leak detection level b from closed circuit Setting of leak detection time c from closed circuit Setting of leak detection time d from closed circuit Setting of no reset SW press alarm time Setting of no dialysis SW press alarm time Setting of no bypass SW press alarm time

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.6.1

9.6. Maintenance mode 3 screen


In maintenance mode 3 screen, alarm can be released and resumed at any time required. Press [F-3] or [F-4] SW to shift blink and press [F-1] or [F-2] SW to change setting. By selecting ON, alarm is released and by selecting OFF, alarm is resumed. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.6-1 Maintenance mode 3 screen


INDICATION 1.V.PRESS 2.A.PRESS 3.D.PRESS/TMP 4. TEMPERATURE 5.AIR BUBBLE 6.BLOOD LEAK 7. SYSTEM ALL 8.SYSTEM E02 9.SYSTEM E03 10. SYSTEM E04 11.SYSTEM E07 12. SYSTEM E08 13.SYSTEM E11 14.SYSTEM E12 15.SYSTEM E15 16.SYSTEM E16 17.SYSTEM E26 18.SYSTEM E33 19.SYSTEM E34 20.SYSTEM E35 21. SYSTEM E45 DESCPIPTION Release of upper/lower venous pressure alarm limits. Release of upper/lower arterial pressure alarm limits Release of upper/lower dialysate pressure alarm limits and TMP alarm limit Release of upper/lower dialysate temperature alarm limits Release of bubble alarm Release of blood leak alarm and blood leak sensor error Release of all alarms registered in system Release of T1 temperature sensor error Release of T2 temperature sensor error Release of T3 temperature sensor error Release of upper limit set by dialysate pressure limiter Release of lower limit set by dialysate pressure limiter Release of upper limit set by venous pressure limiter Release of lower Iimit set by venous pressure limiter Release of upper limit set by arterial pressure limiter Release of lower limit set by arterial pressure limiter Release of short water supply alarm Release of viscous pump (Vpa) drive error Release of viscous pump (Vpb) drive error Release of VP aspiration volume set by limiter Release of deaeration tank AS2 error

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.6.2

WARNING

DO NOT release any alarm during dialysis. Otherwise, machine failure may be caused.

CAUTION

Monitor machine functions continuously during release of any alarm.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.7.1

9.7. Maintenance mode 4 screen


In maintenance mode 4 screen, functions of components such as those assembled in fluid (dialysate) line can be monitored and valves, pumps, clamp and fan can be operated manually. If setting is protected, indication only is enabled and setting is disabled. Press [F-3] or [F-4] SW to check data. If not protected, press [F-3] or [F-4] SW to shift blink and press [F-1] or [F-2] SW to operate component. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.7-1 Maintenance mode 4 screen (function monitoring)

Fig. 9.7-2 Maintenance mode 4 screen (manual operation)

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.7.2
FUNCTION WHEN SYMBOL IS MARKED WITH O IN DRAWING During power supply During power supply During power supply During power supply During power supply During power supply During power supply During power supply During power supply During power supply During power supply During power supply During power supply Shaded photo-sensor Shaded photo-sensor Shaded photo-sensor Shaded photo-sensor Coupler being set Overload detection Being clamped Pressurized water Float drop Air detection Burnt out Burnt out Original point detection. Rotation signal Original point detection Rotation signal CW rotation CCW rotation CW rotation CCW rotation

SYMBOL

COMPONENT

NOTE

V2~V4 Two way electromagnetic valve V10~V17 V5~V9 P1 P2 CLV FAN VPA VPa VPB VPb H BP SP PH1 PH2 PH3 PH4 LS2 LS3 LS5 PS FS FSW P1F P2F Three way electromagnetic valve Deaeration pump Dialysate pump Clamp Fan A chamber viscous pump A chamber viscous pump B chamber viscous pump B chamber viscous pump Heater Blood pump Syringe pump Viscous pump position sensor Viscous pump position sensor Viscous pump position sensor Viscous pump position sensor Coupler SW Syringe overload SW Clamp position sensor Pressure SW Flow sensor Float SW P1 pump fuse P2 pump fuse

DANGER
DO NOT operate any component manually under the condition that patient is connected to

machine to avoid his/her injury.

CAUTION
Manual operation of any component should be performed by engineer familiar with its

construction and function.


DO NOT operate P1 or P2 pump without priming with fluid. Empty run may caused the

pump failure.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.8.1

9.8. Maintenance mode 5 screen


In maintenance 5 screen, dialysate concentration can be set and information related to dialysate preparation in dialysis mode selected in setting. Il screen can be indicated. 9.8.1. When dialysis mode BC1 or BC2 is set Currently set dialysis mode is found in screen title. If setting is protected, indication only is enabled and setting is disabled. Indicated items are common to both BC1 and BC2, while value different from other can be set. Press [F-3] or [F-4] SW to check data. If not protected, press [F-1] or [F-2] SW to change setting. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.8-1 Maintenance mode 5 screen (when BC1 mode is set)


INDICATION 1.VCa W+A RATIO 2.VCb W+A RATIO 3.THEORETICAL W+A 4.VCa W+B RATIO 5.VCb W+B RATIO 6.THEORETICAL W+B 7.CHAMBER 8.MODE 9.COND. W+A+B 10.CONC. W+A+B DESCRIPTION Setting of [water : A] ratio in A chamber Setting of [water : A] ratio in B chamber Setting of theoretical [water + A] concentration Setting of [water : A] ratio in A chamber Setting of [water : A] ratio in B chamber Setting of theoretical [water + B] concentration Indication of functioning chamber. Disabled setting Setting of dialysate preparation function at calibration Calibration of [water + A + B] conductivity Calibration of [water + A + B] concentration

CAUTION

Both BC1 and BC2 are bicarbonate dialysis modes. If different dialysate preparation information is registered, it is recommended to record that the information is registered in which mode BC1 or BC2 after checking concentrate type to be used.

See Concentration Calibration Procedure.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.8.2 9.8.2. When dialysis mode AC is set Currently set dialysis mode is found in screen title. If setting is protected, indication only is enabled and setting is disabled. Press [F-3] or [F-4] SW to check data. If not protected, press [F-1] or [F-2] SW to change setting. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.8-2 Maintenance mode 5 screen (when AC mode is set)

INDICATION 1.VCa W+AC RATIO 2.VCb W+AC RATIO 3.THEORETICAL W+AC 4.CHAMBER 5.MODE 6.COND. W+AC 7.CONC. W+AC

DESCRIPTION Setting of [water : AC] ratio in A chamber Setting of [water : AC] ratio in B chamber Setting of theoretical [water + AC] concentration Indication of functioning chamber. Disabled setting Setting of dialysate preparation function at calibration Calibration of [water + AC] conductivity Calibration of [water + AC] concentration

CAUTION

AC is acetate dialysis mode. It is recommended to record registered information after checking concentrate type to be used. See Concentration Calibration Procedure.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.9.2
INDICATION 18.BPF(OFFSET) 19.ROLLING TUBE TYPE 20.HEATER CONT.COEFFICIENT Kt 21.HEATER CONT.COEFFICIENT Kp 22.HEATER CONT.COEFFICIENT Ki 23.HEATER CONT.COEFFICIENT Kd 24. HEATER CONT.COEFFICIENT Kl 25. HEATER CONT.COEFFICIENT Kb 26.HEATER CONT.COEFFICIENT Lt1 27.HEATER CONT.COEFFICIENT Lt2 28.HEATER CONT.COEFFICIENT Lt3 29.TEMP.CORR.COEFFICIENT 30. T1 OFFSET 31.T2 OFFSET 32.13 OFFSET DESCRIPTION Indication of offset value to calibrate blood pump flow rate (instrument specific value) Setting of blood pump intended to be used Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in heater control Indication of coefficient in temperature calibration of conductivity Indication of T1 determined by input value added T1 temperature sensor value Indication of T2 determined by input value added T2 temperature sensor value Indication of T3 determined by input value added T3 temperature sensor value

DANGER
Change of setting other than calendar year, month and day; and clock hour and minute in maintenance mode 6 may cause significant changes of UF volume, conductivity (concentration) and dialysate temperature resulting in giving serious injury to patient. DO NOT change setting by any person other than engineer familiar with the setting. After setting change, perform trial test of instrument to check its normal function. Perform UF test to check that UF error is within specified range. Measure conductivity to check that difference between the measured one and set one is within specified range. Check that measured blood pump flow rate is within 5% of set flow rate. Check also that temperature control is normal and dialysate temperature is stabilized.

WARNING
DO NOT change "19.ROLLING TUBE TYPE" during dialysis to avoid trouble.

CAUTION
BP flow rate calibration (GAIN and OFFSET) values are those determined using standard NIPRO blood line. When blood line of smaller sized or other manufacturer is used, adjustment is required. Also when blood line of smaller sized is used, adjust gap between BP rotors and housing.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.10.1

9.10. Maintenance mode 7 screen


In maintenance mode 7 screen, system and option settings can be performed. If setting is protected, indication only is enabled and setting is disabled. Press [F-3] or [F- 4] SW to check data. If not protected, press [F-1] or [F-2) SW to change setting. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.10-1 Maintenance mode 7 screen


INDICATION 1.SYRINGE RATE DESCRIPTION Setting of cross-sectional area ratio of syringe intended to be used to NIPRO syringe as 100%. When the ratio is 50%, infusion rate is doubled. Setting to ON when blood pump SW is turned off at return to normal after power failure Setting of time required to reset upper/lower alarm limits of venous, arterial and dialysate pressure during dialysis Setting of dilution rate of Na hypochlorite in disinfection/rinse Setting of dilution rate of acetic acid in acid rinse Limiting of allowable upper venous pressure alarm limit Limiting of allowable lower venous pressure alarm limit Limiting of allowable upper arterial pressure alarm limit Limiting of allowable lower arterial pressure alarm limit Limiting of allowable upper dialysate pressure alarm limit Limiting of allowable lower dialysate pressure alarm limit Setting of initial value of target UF volume Setting of initial value of UFR Setting of minimum UFR Setting to ON when acid rinse port (option) is used Setting to ON when arterial pressure sensor (option) is used Selling to ON when back-up unit (option) for power failure is used Setting to ON when syringe pump is used Setting to ON when operated in console mode Setting to ON when protected setting in MODE 2 and

2. BP STOP AFTER POWER FAIL 3.V,A,DP ALARM RESETTING 4.DISINFECTION RATIO 5.ACID RATIO 6.VP MAX ALARM LIMIT 7.VP MIN ALARM LIMIT 8 AP MAX ALARM LIMIT 9.AP MIN ALARM LIMIT 10.DP MAX ALARM LIMIT 11.DP MIN ALARM LIMIT 12.INITIAL UF GOAL 13.INITIAL UF RATE 14. MINIMUM UF 15.ACID PORT OPTION 16.ART. PRESSURE OPTION 17.BACK UP OPTION 18.SYRINGE OPTION 19.CONSOLE MODE 20 MAINTENANCE MODE FREE

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.10.2

WARNING
Connection of each option should conform to its setting. Otherwise, disabled function of the option or malfunction of machine may be caused.

CAUTION
Set each item value after full understanding its function.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.11.1

9.11. Maintenance mode 8 screen


In maintenance mode 8 screen, service frequency of each component is indicated by operation time or operation frequency. Press [F-3] or [F-4] SW to check data. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.11-1 Maintenance mode 8 screen


INDICATION 1. POWER 2.FAN 3.P1 4.P2 5.CLV 6.V2 7.V3 8.V4 9.V5 10.V6 11.V7 12.V8 13.V9 14.V10 15.V11 16.V12 17.V13 18.V14 19.V15 20.V17 DESCRIPTION Indication of power connection time of machine Indication of fan operation time Indication of deaeration pump P1 operation time Indication of dialysate pump P2 operation time Indication of clamp operation frequency Indication of two way electromagnetic valve (EMV) V2 operation frequency Indication of two way EMV V3 operation frequency Indication of two way EMV V4 operation frequency Indication of three way EMV V5 operation frequency Indication of three way EMV V6 operation frequency Indication of three way EMV V7 operation frequency Indication of three way EMV V8 operation frequency Indication of three way EMV V9 operation frequency Indication of two way EMV V10 operation frequency Indication of two way EMV V11 operation frequency Indication of two way EMV V12 operation frequency Indication of two way EMV V13 operation frequency Indication of two way EMV V14 operation frequency Indication of two way EMV V15 operation frequency Indication of two way EMV V17 operation frequency

CAUTION
Once zero cleared, the cleared operation time or frequency can not be resumed.

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.12.1

9.12. Maintenance mode 9 screen


In maintenance mode 9 screen, air purge process can be executed according to instruction indicated in the screen. Press [F-1] SW to execute air purge process. Repress [F-1] SW to forcibly stop the process. Press [F-5] SW to display maintenance mode menu screen though the display is disabled during air purgeing.

Fig. 9.12-1 Maintenance mode 9 screen (during waiting)

Fig. 9.12-2 Maintenance mode 9 screen (during execution)

Fig. 9.12-3 Maintenance mode 9 screen (during stop)

NIPRO SURDIAL (SEM) ver. OR2102-9902

9.13.1

9.13. Maintenance mode 10 screen


In maintenance mode 10 screen, history of up to 20 latest alarms is displayed. Press [F-3] or [F-4] SW to check data. Press [F-1] SW to clear the history. Press [F-5] SW to display maintenance mode menu screen.

Fig. 9.13-1 Maintenance mode 10 screen

INDICATION

DESCRIPTION Indication of day/month/year of alarm actuation Indication of time of error actuation Indication of error No. of actuated alarm

/ :
E

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.1.1

10. ADJUSTMENT
10.1. Measurement of chamber capacity and adjustment of silicone volume
10.1.1. Measurement of chamber capacity In the cases such as replacement of viscous chamber or viscous pump, silicone oil volume in the chamber may be changed. Measure the chamber capacity based on the below procedures and input the measured result into machine.
(1) Required devices Electronic balance ( 0.2g as minimum unit) Container (to be placed on the balance (2000mL) Forceps (2) Turn on power SW of machine. (3) Turn on [DIALYSIS] SW to shift to dialysis wait process. (4) Release setting protection by selecting ON of 20.MAINTENANCE MODE FREE in maintenance mode 7 screen. (5) Set machine operation to console mode by selecting ON of 19.CONSOLE MODE in maintenance mode 7 screen.

(6) By pressing [PRIP.] SW to shift to preparation process, machine is started from console mode. (7) Check dialysate temperature stabilization. (8) Isolate air discharge line of AS1 from drain line by clamping upper tube of AS1 by forceps or the like.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.1.2

(9) Measure RO water weight discharged from each of A and B chambers when each of [V5 and V6] and [V7 and V8] are ON using electronic balance under referring to data displayed in maintenance mode 4 screen to derive each of the A and B chamber capacities respectively. Calculate mean value of 3 measurements for each of the chambers.

A chamber capacity : when V5 and V6 are ON B chamber capacity: When V7 and V8 are ON

Fig. 10.1-1 Measurement of chamber capacity


(10) Each of the capacities should satisfy the below requirement. If not so, adjust silicone oil volume in each of the chambers followed by performing the measurement 3 times for each of them. 345ml(g) chamber capacity 375ml(g) (11) Input the chamber capacities in maintenance mode 6 screen. A chamber capacity : 9.VCa VOLUME (mL) B chamber capacity : 10.VCb VOLUME (mL)

(11) Turn on [DIALYSIS] SW to shift to dialysis wait process. (12) Release console mode by selecting OFF of 19.CONSOLE MODE in maintenance mode
7 screen.

(13) Protect setting by selecting OFF of 20.MAINTENANCE MODE FREE in maintenance


mode 7 screen.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.1.2. Adjustment of silicone oil volume


Add silicone oil to the chamber when the capacity is larger than specified volume and aspirate from the chamber when smaller than the volume based on the following procedures:
(1) Required devices: Silicone tube (ID: 6mm. length: 200~300mm) Syringe (applicable to the silicone tube) Silicone oil (when added) Forceps (2 or more) Radio pliers

Tissue paper or the like used to wipe off excess silicone oil (2) Turn on power SW of machine. (3) Turn on [DIALYSIS] SW to shift to dialysis wait process. (4) Release setting protection by selecting ON of "20.MAINTENANCE MODE FREE maintenance mode 7 screen. (5) Clamp plug snapper plug. (6) Turn on V4 manually in maintenance mode 4 screen.

provided tube

placed at top of viscous pump using forceps and remove

6 from the plug by twisting latch along tooth line using radio pliers to remove the

(7) Use silicone oil filled syringe 2 to add the oil to the chamber and use empty syringe from the chamber. (8) Connect one end of silicone tube

to aspirate it

3 to the syringe
clamped using

and other end to tube

forceps. (9) Release the clamping and remove all air together with small volume of the oil from the silicone tube

by retracting the syringe plunger.

(10) Add or aspirate required volume of the oil to or from the chamber under referring to the syringe calibration by taking correction of the removed oil volume in the step (9) into consideration followed by re-clamping the tube 5. Fig. 10.1-2 Adjustment of silicone oil volume

NIPRO SURDIAL (SEM) ver. OR2102-9902

(11) Remove the silicone tube and re-attach the plug (12) Fix the plug securely with the snapper 6 and remove the forceps. (13) Turn off V4 manually in maintenance mode 4 screen. (14) Operate machine to measure the chamber capacity and input the measured result in
maintenance mode 6 screen.

(15) Protect setting by selecting OFF of "20.MAINTENANCE MODE FREE" in maintenance mode
7 screen.

DANGER

Take care for air not to be included in silicone oil to prevent conductivity (concentration) error or

UF error potentially causing patient injury.

WARNING

Use silicone oil of acceptable quality, viscosity, purity and the like specified by manufacturer. DO

NOT use any other silicone oil.


Add or aspirate silicone oil to or from chamber when the chamber capacity is larger or smaller

than specified volume.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.2. Adjustment of venous and arterial pressure sensors


10.2.1. Adjustment of venous pressure sensor
(1) Required devices Manometer (calibrated) Syringe Three way stopcock Air tube (ID: about 4mm) Precision screwdriver (2) Remove right cover from upper operation part and right cover from DENSO box to expose venous pressure sensor board. PV PCB: Venous pressure sensor board PA PCB: Arterial pressure sensor board (option)

Fig. 10.2-1 Venous and arterial pressure sensor boards (3) Turn on power SW of machine. (4) Turn on [DIALYSIS) SW to shift to dialysis wait process. (5) Release setting protection by selecting ON of 20.MAINTENANCE MODE FREE
maintenance mode 7 screen.

(6) Set 6.V.PRESS OFFSET value to 0mmHg in maintenance mode 6 screen. Record the
offset value to re-input it at the last step.

(7) Display venous pressure value in dialysis screen or maintenance mode 1 screen.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.2.2

(8) Connect syringe and manometer to venous pressure inlet port as shown in the below Fig.

Fig. 10.2-2 Adjustment of venous pressure sensor

(9) Expose manometer to atmosphere and try to indicate venous pressure of 0mmHg in the screen
by turning VR701.

(10) Pressurize manometer to 300mmHg by advancing syringe plunger and try to indicate venous
pressure of 300mmHg in the screen by turning VR702.

(11) Repeat the above steps (9) and (10) until each of 0mmHg and 300mmHg is indicated.
Manometer reading 0 mmHg 300 mmHg Adjusting means PV_PCB VR701 PV_PCB VR702

(11) Input 6.V.PRESS OFFSET value recorded at the above step (6) in maintenance mode 6
screen.

(12) Protect setting by selecting OFF of 20.MAINTENANCE MODE FREE in maintenance mode 7
screen.

(13) Disconnect the syringe and manometer. (14) Return the removed covers to the original positions.
10.2.2. Adjustment of arterial pressure sensor (option) Adjust arterial pressure sensor based on the same procedures as the case of the venous pressure sensor. Offset value and adjusting means are as follows.
Offset value

Maintenance mode 6 screen


Adjusting means Manometer reading

7.A.PRESS OFFSET

Adjusting means

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.3. Adjustment dialysate pressure sensor


(1) Required devices Manometer (calibrated) Syringe Three way stopcock Air tube (ID: about 4mm) Silicone tube (ID: about 5mm) Precision screwdriver Forceps (2) Remove left cover from upper part of machine and left cover from lower part of it to expose dialysate pressure sensor Assy.

PD PCB : Dialysate pressure sensor board PD Assy : Dialysate sensor Assy

Fig. 10.3-1 Dialysate sensor board and dialysate sensor Assy


(3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Release setting protection by selecting ON of 20.MAINTENANCE MODE FREE in maintenance mode 7 screen. (6) Set 8.D.PRESS OFFSET value to 0mmHg in maintenance mode 6 screen and record the offset value to re-input at the last step. (7) Indicate dialysate pressure value in dialysis screen or maintenance mode 1 screen.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.3.2
(8) Clamp tube connected to one side of dialysate pressure sensor using forceps and connect syringe and manometer to tube connected to other side of the sensor as shown in the below Fig.

Fig. 10.3-2 Adjustment of dialysate pressure sensor


(9)

Expose manometer to atmosphere and try to indicate dialysate pressure of 0mmHg in the screen by turning VR701. Pressurize manometer to 300mmHg by advancing syringe plunger and try to indicate dialysate pressure of 300mmHg in the screen by turning VR702. Repeat the above steps (9) and (10) until each of 0mmHg and 300mmHg is indicated.

(10)

(11)

Manometer reading 0 mmHg 3 0 0 mmHg

Adjusting means PD_PCB PD_PCB VR701 VR702

(11) Input 8.D.PRESS OFFSET value recorded at the above step (6) in maintenance mode

6 screen.
(12) Protect setting by selecting OFF of 20.MAINTENANCE MODE FREE in maintenance mode

7 screen.
(13) Remove the syringe and manometer. (14) Return the removed covers to the original positions.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.4.1

10.4. Adjustment of temperature sensor


Connect resistor for calibration to terminal bed on main board and adjust temperature sensor by input from operation panel.
(1) Required device Resistor for calibration (115.54 at 40.0C) (2) Remove right cover from upper operation part and right cover from DENSO box to expose terminal bed TM1 on main board.

MAIN PCB : Main board TM1 : Terminal bed

Fig. 10.4-1 Main board and terminal bed TM1


(3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Release setting protection by selecting ON of "20.MAINTENANCE MODE FREE" in maintenance mode 7 screen. (6) Adjust the sensor at 40.0 C only. (7) Take out temperature sensor to be adjusted from terminal bed and connect to resistor for calibration as shown in Figs. 10.4-1 and 10.4-2.

Fig. 10.4-2 Connection to resistor for calibration

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.4.2

(8) Adjust by checking indicated value when the resistor is connected. Check the value in maintenance mode 6 screen and input offset value.

31.T1 OFFSET 32.T2 OFFSET 33.T3 OFFSET

40.1c: 40.1c: 40.2c:

0.0 0.0 0.0

3 3: Set offset value

1: Item name 2: Currently recognized temperature

(9) Remove the resistor from terminal bed and attach temperature sensor followed by

connecting it as shown in Fig. 10 4-1 under care of correct color matching of harness (W: white; R: red).
(10) Protect setting by selecting OFF of 20.MAINTENANCE MODE FREE in maintenance

mode 7 screen.
(11) Return the removed covers to the original positions.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.5.1

10.5. Adjustment of pressure SW


Apply pressure to pressure SW to adjust pressure when the SW is ON by turning adjusting screw based on the following procedures.
(1) Required devices Manometer (calibrated) Syringe Three way stopcock Air tube (ID: about 4mm) Silicone tube (ID: about 5mm) Forceps Spanner (8mm) Hexagon wrench (4mm)

(2) Remove right cover from lower part of machine to expose pressure SW. (3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Clamp tube connected to one side of pressure SW using forceps and connect syringe and manometer to tube connected to other side of it as shown in the below Fig.

Fig. 10.5-1 Adjustment of pressure SW


(6) Check ON or OFF of pressure SW in maintenance mode 4 screen.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.5.2
(7) Apply pressure to pressure SW by gradual advancement of syringe plunger. (8) Adjust pressure SW so that it is ON at a pressure between 514.5 and 588mmHg (0.7 and 0.8 kgf/cm2) by loosening set nut using spanner followed by turning adjusting screw. ON of pressure SW can be found by ON of PS in maintenance mode 4 screen or clicking of micro SW. (9) After the adjustment, tighten the set nut. This may cause small change of pressure when pressure SW is ON. Recheck the ON pressure. (10) Then retract the syringe plunger gradually to check that pressure SW is OFF at a pressure of 441 mmHg (0.6kgf/cm2).

SW function ON OFF

ON/OFF pressure 514.5 to 566 mmHg (0.7 to 0.6kgf/cm2) 441 mmHg( 0.6kgf/cm2)

(11) Remove the syringe and manometer and return the pressure SW connection to the original. (12) Return the removed cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.6.1

10.6. Adjustment of blood leak sensor


Turn volume of blood leak sensor under checking value indicated in screen based on the following procedures.

NOTE

Clean the Blood leak detector, before you calibrate it. Blood leak detector which is not cleaned detect blood leak even it does not happen blood leak.

(1) Required device Precision screwdriver (2) Remove left cover from lower part of machine to expose blood leak sensor. (3) Remove 2 screws used to fix the sensor and take out it without disconnecting tubes. (4) Remove the sensor cover.

Fig. 10.6-1 Blood leak sensor


(5) (6) (7) (8) (9) Turn on power SW of machine. Turn on [DIALYSIS] SW to shift to dialysis wait process. Press [PREP.] SW to shift to preparation process. Water may be used instead of concentrate. Check water temperature stabilization. Adjust R43 to try to set 8.BLD RED LIGHT value to 055 in maintenance mode 1 screen.

(10) Adjust R47 to try to set 7.BLD SENSE VALUE to 80055 in maintenance mode 1 screen. (11) Repeat the above steps (9) and (10) until each of 055 and 80055 is set. (12) Attach the removed sensor cover to the sensor to return it to the original position. (13) Return the removed left cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.7.1

10.7. Adjustment of reducing valve


Adjust actuating pressure of reducing valve by turning adjusting screw of the valve by measuring manometer pressure after priming completion based on the following procedures.
(1) Required devices Manometer (calibrated) Connection means (tubes and T-pipe) Forceps

(2) Remove right cover form lower part of machine to expose reducing valve. (3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Press [PREP.] SW to shift to preparation process. Water may be used instead of concentrate. (6) Check water temperature stabilization. (7) Clamp tube connected to IN of reducing valve V10 using forceps and disconnect silicone tube from the valve to connect it to manometer through connection means.

E 10.7-1 Connection of reducing valve to manometer


(8) After lifting manometer sufficiently, release the clamping followed by lowering it to align with reducing valve horizontally. (9) Check visually water discharge from machine and set the manometer indication to 250 to 270mmHg (0.34 to 0.37kgf/cm2) during stop of the discharge by loosening set nut followed by turning adjusting screw of reducing valve. (10) Tighten the set nut. This may cause small change of the set value. Check the value again.

Adjusting value Reducing valve 250 to 270mmHg(0.34 to 0.37kgf/cm2)

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.8.1

10.8. Adjustment of priming end pressure


Adjust priming end pressure by checking pressure at priming end by turning relief valve 1 based the following procedures.
(1) Required devices Manometer (calibrated) Connection means (tubes and T-pipe) Forceps Monkey wrench

(2) Remove right cover from lower part of machine to expose relief valve 1. (3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Press [PREP.] SW to shift to preparation process. Water may be used instead of concentrate. (6) Check water temperature stabilization. (7) Clamp tube connected to IN of V11 using forceps and disconnect silicone tube from V1 connect it to manometer through connection means.

Fig. 10.8-1 Connection of manometer (8) After lifting manometer sufficiently, release the clamping followed by lowering it to align chamber horizontally. (9) Check visually water discharge from machine and set the manometer indication to 698 to 735mmHg (0.95 to 1.00kgf/cm2) during stop of the discharge by loosening set nut of valve 1 followed by turning the valve. Take care for silicone tube not to be twisted. (10) Tighten the set nut. This may cause small change of the set value. Check the value again.

Adjusting value Priming end pressure 698 to 735mmHg (0.95 to 1.00kgf/cm2)

(11) (12)

Remove the manometer and the connection means to return to the original condition. Return the removed cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.9.1

10.9. Adjustment of closed circuit pressure


Adjust closed circuit pressure based on the following procedures.
(1) Required devices Blood line Dialyzer

(2) Remove left cover from lower part of machine to expose relief valve 2. (3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Connect concentrate to machine. (6) Press [PREP.] SW to shift to preparation process. (7) Set dialyzer and blood line to machine. (8) Check stabilization of dialysate temperature and conductivity and release of bypassing. (9) Full open NV2 to run dialysate at the maximum flow rate. (10) Adjust blood pump flow rate so that dialysate pressure becomes 0mmHg. (11) Loosen set nut of relief valve 2 and turn the valve body to adjust dialysate flow rate indicated by flow mater to 600mL/min. Take care for silicone tube not to be twisted. (12) Tighten the set nut. This may cause small change of the dialysate flow rate. Check the flow rate again. (13) Turn NV2 to adjust dialysate flow rate to 500mL/min. (14) Return the removed cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

(13) (14)

Protect setting by selecting OFF of 20.MAINTENANCE MODE FREE in maintenance mode screen. Return the removed cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.10.2

10.10.2. Adjustment by Discharge Flow Time


Adjust priming flow rate of water by turning control valve NV1 by measuring discharge flow time which is substantially same as priming flow rate. Follow the below procedures.
(1) Required device Stopwatch (2) Remove right cover from lower part of machine to expose NV1. (3) Turn on power SW of machine. (4) Turn on [DIALYSIS] SW to shift to dialysis wait process. (5) Release setting protection by selecting ON of 20.MAINTENANCE MODE FREE in maintenance mode 7 screen. (6) Set machine operation to console mode by selecting ON of 19.CONSOLE MODE in maintenance mode 7 screen.

(7) By pressing [PRIP.] SW to shift to preparation process, machine is started from console mode. (8) Check dialysate temperature stabilization. (9) Measure discharge flow time from start of discharge to end of discharge. Discharge flow is intermittent.

Fig. 10 10-1 Discharge Flow Time


(10) Adjust discharge flow time from 29 to 31sec by turning adjusting screw of NV1. Turn NV1 right crease the rate of discharge, turn NV1 left increase the rate of discharge.

Adjusting valve Priming flow time 29 ~ 31sec

(11) Turn on [DIALYSIS] SW to shift to dialysis wait process.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.12.1

10.12. Adjustment of gaps between housing and rotor rollers of blood pump
10.12.1. Adjustment by water head pressure Adjust the gaps by application of water head pressure to blood line set to blood pump based following procedures.
(1) Required devices Blood line Water receiver Water bottle Hexagon wrench (1.5mm) (2) Turn off power SW of machine. (3) Set rolling tube of blood line to blood pump and connect water bottle to inflow end of t?? (4) Prime blood line with water from the bottle by gravity flow of water. (5) After the priming completion, turn rotor of blood pump manually to occlude rolling tube of 3 rollers of the rotor.

Fig. 10.12-1 Adjustment of gaps between housing and rotor rollers of blood pump
(6) Lift the bottle 1 so that height of water level in it from center of the rotor becomes above Open outflow end of the rolling tube above water receiver. (7) Loosen set screw 3 of the roller by hexagon wrench 2. Turn shaft 4 by screw repeat

narrowing and widening of gap between the roller and housing of blood pump set the gap to allow water discharge from the outflow end. (8) Tighten the set screw 3 at a moment when the water discharge is stopped by turning 4 slowly. (9) Repeat the above procedures until all the gaps are adjusted.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.11.1 10.11. Adjustment of bypass flow rate During V9 is OFF (that is, dialysate line is bypassed), adjust dialysate flow rate by turning adjustment screw of pinch cock 1 (PC1) under monitoring flow rate indication of flow meter based on the following procedures. (1) Remove left cover form lower part of machine to expose PC1. (2) Turn on power SW of machine. (3) Turn on [DIALYSIS] SW to shift lo dialysis wait process. (4) Press [PREP.] SW to shift to preparation process. Engage coupler with bypass connect
dialysate concentrate is not used, connect dumm? resistor to conductivity sensor.

(5) Check stabilization of dialysate temperature and concentration. (6) Check that [BYPASS] SW is OFF (released bypassing and dialysate flow through coupler (7) Adjust dialysate flow rate to 500ml/min by turning adjusting screw of NV2 under monitor
dialysate flow rate indication of flow meter.

(8) Turn on [BYPASS] SW to bypass dialysate line. (9) Adjust bypass flow rate to 500ml/min by turning adjusting screw of PC1 under monitor
bypass flow rate indication of flow meter. Tube should be centered at PC1 and plates of should be aligned correctly.

Fig. 10.11-1 Correct and incorrect conditions of pinch cock 1 (PC1) (10) Turn off [BYPASS] SW and wait for 1 to 2 minutes. (11) Turn on and off [BYPASS] SW several times to check that dialysate flow rate is 500ml/min
when bypassed and not bypassed.

(12) Apply adhesive or the like to adjusting screw of PC1 for fixing. (13) Return the removed cover to the original position.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.12.2

10.12.2. Adjustment using syringe and manometer


Adjust the gaps by application of water pressure to blood line set to blood pump by syringe based or the following procedures.
(1) Required devices Blood line Water receiver Syringe Manometer Hexagon wrench (1.5mm) (2) Turn off power SW of machine. (3) Set rolling tube of blood line to blood pump. (4) Prime the tube with water. (5) After the priming completion, turn rotor of blood pump manually to occlude the tube by one of ???rollers of the rotor.

Fig. 10.12-2 Adjustment of gaps between housing and rotor rollers of blood pump
(6) Connect syringe 1 and manometer 5 to inflow end of the tube and open outflow end o?? above water receiver. (7) Loosen set screw 3 of the roller by hexagon wrench 2. Turn shaft 4 by screwdriver repeat narrowing and widening of gap between the rotor and housing of blood pump. Initially turn the shaft to stop water discharge from the outflow end when a water pressure 0.2kgf/cm2 is applied to the tube through the inflow end by operating the syringe. (8) Tighten the set screw 3 at a moment when the water discharge is started under the pressure application by turning the shaft 4 slowly. (9) Repeat the above procedures until all the gaps are adjusted.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.13.1

10.13. Adjustment of blood pump flow rate


10.13.1 In case of you use Standard Rolling Tube ( 8 X 12mm) Adjust blood pump flow rate by setting blood line to blood pump to input GAIN and OFFSET values of blood pump flow rate by referring to actually measured flow rate (discharged water weight or volume/min) in maintenance mode 6 screen based on the following procedures.
(1) Required devices Electronic balance or measuring cylinder Stopwatch (2) Turn on power SW of machine. (3) Turn on [DIALYSIS] SW to shift to dialysis wait process. (4) Release setting protection by selecting ON of 20.MAINTENANCE MODE FREE in maintenance mode 7 screen. (5) Set blood line to blood pump.

(6) Place electronic balance or measuring cylinder below outflow end of the line to measure
discharged water weight or volume.

Fig. 10.13-1 Adjustment of blood pump flow rate


(7) Set blood pump flow rate to 50ml/min and operate the pump. (8) Measure the discharged water weight or volume per one minute. Flow rate (g/min or ml/min) = measured weight or volume time (one minute) value to try to adjust the actual flow rate to about 50ml/min. Decrease (increase) the value if the actual flow rate is higher (lower) than the set flow rate.

(9) Input 18.BPF(OFFSET) value of blood pump flow rate in maintenance mode 6 screen

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.13.2 (10) Set blood pump flow rate to 400ml/min and operate the pump. (11) Measure actually discharged water weight or volume/minute (ml/min). (12) Input 17.BPF(GAIN) value of blood pump flow rate in maintenance mode 6 screen value
t?? to adjust the actual flow rate to about 400ml/min. Decrease (increase) the value if the a?? flow rate is higher (lower) than the set flow rate.

(13) Repeat the above procedures (7) to (12) until each of the actual flow rates of 50 and
400ml/min is obtained.
Blood pump flow rate 50ml/min 400ml/min Adjusting value Maintenance mode 6 screen 18.BPF(OFFSET) Maintenance mode 6 screen 17.BPF(GAIN)

(14) Protect setting by selecting ON of 20.MAINTENANCE MODE FREE in maintenance


mode ?? screen.

10.13.2 In case of you use Small Rolling Tube ( 6.35X 9.75mm)


(1) See chapter 10.12 and adjust gaps between housing and rotor rollers by using Small

Rolling Tube.
(2) Change NORMAL to SMALL at [19.ROLLING TUBE TYPE] in Maintenance mode 6

screen (Max blood flow change 500ml/min to 350mI/min)


(3) Set 2 pies of Tube holder (See Chapter 6.2 of Operation Manual) at inlet and outlet of

???? pump unit.


(4) See chapter 10.13.1 and adjust blood pump flow as same method. But adjust 50mL/min

to OFFSET and adjust 250mL/min to GAIN.

Blood pump flow rate 50 mL/min 250 mL/min

Adjusting value Maintenance mode 6 screen 18.BPF(OFFSET) Maintenance mode 6 screen 17.BPF(GAIN)

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.14.1

10.14. Leak test and gap adjustment of clamp


10.14.1. Leak test of clamp
Leak test of clamp is performed to find whether gap adjustment of the clamp is required based on the following procedures. (1) Required devices (2) Manometer Connection means Forceps Syringe Stopwatch Turn on power SW of machine.

(3) Turn on [DIALYSIS] SW to shift to dialysis wait process. (4) Set manometer, connection means and syringe to clamp as shown in Fig. 10.14-1 (a). (5) Turn on BUBBLE SW to operate the clamp after bubble introduction into venous tube. (6) Set manometer indication to 150mmHg by pressurizing the venous tube using the syringe followed by clamping line between the syringe and manometer using forceps to maintain the pressure as shown in Fig. 10.14-1(b). (7) After the clamping, leave to stand for about 90 seconds to record the pressure A. (8) Additionally leave to stand for about 120 seconds to record the pressure B. (9) Additionally clamp line between the clamp and manometer using forceps as shown in Fig. 10.14-1(c). (10) After the clamping, additionally leave to stand for about 120 seconds to record the pressure C. (11) Derive leak value from the recorded pressures A, B and C. If the value is not acceptable, adjust the clamp gap and re-check. Leak value (mmHg) = Leak value from clamp and test setup Leak value from test setup =(AB)(BC)
Acceptable range Clamp leak value 1.00mmHg

Fig. 10.14-1

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.14.2 10.14.2. Gap adjustment of clamp If leak from the clamp is detected, adjust the clamp gap based on the following procedures.
(1) Required devices Spanner (x 2) (2) Remove left cover from upper part of machine to expose clamp. (3) Turn off power SW of machine. (4) Push rear side 1.3 to 1.5mm. (5) Release the push and loosen 3 set bolts using spanner to make the solenoid free. (6) Adjust the gap to 1.3 to 1.5mm by loosening set screw (nut) screw

1 of the clamp by finger to check me gap. In general, the gap is adjusted to 2 used to fix solenoid to upper part of the clamp 3 followed by turning adjusting

4. If venous tube of smaller diameter is used, reduce the gap of 1.3 to 1.5mm. The

adjustment is enabled only in direction to reduce the gap. To enlarge the gap, turn the adjusting screw 4 followed by shifting the solenoid 5. (7) Under solenoid (5) fits to screw (3) correctly, tighten the 3 set bolts 2 to fix the solenoid to upper pad of the clamp using spanner. (8) Push rear side of the clamp by finger to check the adjusted gap.
Adjusting value Clamp gap 1.3 to 1.5mm ( 1)

1: Change based on venous tube diameter if different from standard.

(9) After checking correct gap adjustment, tighten the set screw (nut) 3. (10) Perform the leak test again to check whether the gap is correctly adjusted. (11) Return the removed cover to the original position.

Fig. 10.14-2 Adjustment of clamp gap


NIPRO SURDIAL (SEM) ver. OR2102-9902

10.15.1

10.15. Adjustment of bubble sensor


Adjust bubble sensor by setting water filled blood line to bubble detector followed by turning volume of bubble sensor board based on the following procedures.
(1) Required devices Blood line Tester Precision screwdriver Micro-syringe

(2) Remove left cover from upper part of machine to expose bubble sensor board. (3) Set tester to DC range and connect test PIN602 of bubble sensor board to (+) probe of the tester and PIN601 of the board to (-) probe of it. (4) Set standard tube filled with about 37C water to bubble detector and adjust voltage indication of the tester to the maximum by turning VR601. (5) Check that the adjusted voltage is 1.6 to 1.8V. (6) Remove the tube from the detector and check that the voltage indication is reduced to 0.5V. (7) Set blood line to blood pump. (8) Operate the pump at a flow rate of 200mL/min to re-circulate 37C water through the line. (9) Turn on BUBBLE DETECTOR SW and introduce 0.05ml of bubble into the line using microsyringe under care for the bubble is not divided. Check that bubble alarm is actuated when the bubble passes through the detector sensor. (10) Return the removed cover to the original position.

Fig. 10.15-1 Bubble sensor board

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.16.1

10.16. Adjustment of conductivity and concentration


Adjust dilution ratio of dialysate concentrate intended to be used to input the adjusted ratio and calibrate indicated dialysate conductivity and concentration by manipulation of operation panel based on the following procedures.

Fig. 10.16-1 Adjusting and calibrating process

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.16.2

10.16.1. Selection of dialysis mode to be adjusted


Select dialysis mode: bicarbonate 1 (BC1), bicarbonate 2 (BC2) or acetate (AC) to be adjusted in setting II screen. Concentrate data of the selected mode is displayed in maintenance mode 5 screen.

(1) Turn on power SW of machine. (2) Turn on [DIALYSIS] SW to shift to dialysis wait process. (3) Press [F- 5] SW to display setting II screen. (4) Press [F- 3] or [F-4] SW to blink 4.MODE. (5) Press [F-1] or [F-2] SW to select dialysis mode. (6) Check the selected mode.

BC1(BICARBONATE1) : Bicarbonate dialysis mode 1 BC2(BICARBONATE2) AC(ACETATE) :Bicarbonate dialysis mode 2 :Acetate dialysis mode

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.16.3

10.16.2. Input of concentration data


Input basic concentrate data in maintenance mode 5 screen. When bicarbonate dialysis mode (BC1 or BC2) is selected in setting II screen, input A and B concentrate data. When acetate dialysis mode (AC) is selected, input acetate concentrate data. See the data specified in the operation manual. Also set concentration to theoretical value derived from the concentrate data in setting II screen.

Input of data of bicarbonate concentrate


(1) Check that current process is dialysis wait. (2) Press [F-5] SW to display maintenance mode 5 screen.

(3) Press [F- 3] or [F-4] SW to blink item to be inputted. (4) Press [F- 1] or [F-2] SW to input data. (5) Input the following values by referring to concentrate data specified in the operation

manual. 1.VCa W+A RATIO 2.VCb W+A RATIO 3.THEORETICAL W+A(mmol/L 4.VCa W+B RATIO 5.VCb W+B RATIO : Dilution ratio of A concentrate in chamber a : Dilution ratio of A concentrate in chamber b : Theoretical concentration of diluted A concentrate : Dilution ratio of B concentrate in chamber a : Dilution ratio of B concentrate in chamber b

6.THEORETICAL W+B(mmol/L) : Theoretical concentration of diluted B concentrate (6) Press [F-5] SW to display setting II screen.

(7) Press [F-3] or [F-4] SW to blink item to be inputted. (8) Press [F-1] or [F-2] SW to input data. (9) Input the following values by referring to concentrate data specified in the operation

manual. 1.BC(W+A+B) : Final concentration (theoretical concentration of diluted A concentrate + that of diluted B concentrate) 2.BC(W+A) : Theoretical concentration of diluted A concentrate

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.16.4

Input of data of acetate concentrate


(1) Check that current process is dialysis wait. (2) Press [F-5] SW to display maintenance mode 5 screen.

(3) Press [F-3] or [F-4) SW to blink item to be inputted. (4) Press [F-1] or [F-2] SW to input data. (5) Input the following values by referring to concentrate data specified in the operation manual.

1.VCa W+AC RATIO : Dilution ratio of acetate concentrate in chamber a 2.VCb W+AC RATIO : Dilution ratio of acetate concentrate in chamber b 3.THEORETICAL W+AC(mmol/L) : Theoretical concentration of diluted acetate

concentrate (6) Press [F-5] SW to display setting II screen.

(7) Press [F-3] or [F-4] SW to blink item to be inputted. (8) Press [F-1] or [F-2] SW to input data. (9) Input the following value by referring to concentrate data specified in the operation manual. 3.AC(W+AC) : Theoretical concentration of diluted acetate concentrate

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.16.5 10.16.3. Adjustment of dilution ratio

Adjustment of dilution ratio of bicarbonate A concentrate


Adjust the dilution ratio when bicarbonate dialysis mode is selected. Set dialysate preparation components to water and A concentrate only in maintenance mode 5 screen and operate machine in preparation or preparation end process. Measure concentration of prepared dialysate specimen after prepared dialysate conductivity is stabilized. If required, change the dilution ratio to adjust.
(1) Check that current process is dialysis wait. (2) Press [F-5] SW to display maintenance mode 5 screen. (3) Press [F-3] or [F-4) SW to blink 8.MODE. (4) Press [F-1] or [F-2) SW to set W+A for selecting water and A concentrate only as dialysate

preparation components.
(5) Connect the concentrate to concentrate connector. (6) Press [PREP.] SW to shift to preparation process to start dialysate preparation. (7) Wait until conductivity (mS/cm) is stabilized under monitoring 9.COND. W+A+B. (8) After the stabilization, take dialysate specimen from sampling port. (9) Measure the specimen concentration (mmol/L) using separately available concentration meter.

It is possible to derive the concentration by converting conductivity measured using any conductivity meter to the concentration by referring to conversion Table.
(10) Check that the measured concentration (mmol/L) is equal to theoretical concentration of diluted

A concentrate (3.THEORETICAL W+A). If not equal, change dilution ratios of 1.VCa W+A RATIO and 2.VCb W+A RATIO to adjust them.
Calculate each of the changed dilution ratios roughly by application of the below formula: New dilution ratio = [Theoretical concentration of diluted A concentrate X Current dilution ratio] [Measured concentration] New dilution ratio: dilution ratio after adjustment Input of

1 into 1.VCa W+A RATIO and 2.VCb W+A RATIO AA. AA 1 in 1.VCa W+A RATIO and 2.VCb W+A RATIO BB.BB

Current dilution ratio: dilution ratio before adjustment Indication of

Measured concentration : Concentration measured using meter (mmol/L) Theoretical concentration of diluted A concentrate A: (mmol/L) Indicated in 3.THEORETICAL W+A (11) After the adjustment, wait until indicated conductivity is stabilized. (12) Take dialysate specimen again from sampling port to measure its concentration using

concentration meter. Repeat the adjustment until the measured concentration (mmol/L) becomes equal to theoretical concentration of 3 THEORETICAL W+A.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.16.8 10.16.4. Calibration of indicated conductivity and concentration After adjustment completion of each dilution ratio and conductivity stabilization, calibrate indicated conductivity and concentration in maintenance mode 5 screen.

Calibration of indicated conductivity and concentration of bicarbonate dialysate


(1) Check completion of the adjustment and stabilization of dialysate conductivity. (2) Press [F- 5] SW to display maintenance mode menu 5 screen. (3) Press [F-3] or [F-4] SW to blink 9.COND. W+A+B. (4) Take dialysate specimen from sampling port. (5) Measure the specimen conductivity (mS/cm) using conductivity meter not assembled in
SURDIAL

(6) Press [F -1] or [F-2] SW to input the measured conductivity into 9.COND. W+A+B. (7) Convert the measured conductivity to concentration (mmol/L) by referring to conversion Table
when available. If not available, measure the specimen concentration using separately available concentration meter.

(8) After obtaining the concentration, press [F-3] or [F-4] SW to blink 10.CONC. W+A+B. (9) Press [F-1] or [F-2] SW to input the obtained concentration into 10.CONC. W+A+B.

Calibration of indicated conductivity and concentration of acetate dialysate


(1) (2) (3) (4) (5) (6) (7)
Check completion of the adjustment and stabilization of dialysate conductivity. Press [F-5] SW to display maintenance mode menu 5 screen. Press [F-3] or [F-4] SW to blink 6.COND. W+AC. Take dialysate specimen from sampling port. Measure the specimen conductivity (mS/cm) using conductivity meter not assembled in SURDIAL. Press [F-1] or [F-2] SW to input the measured conductivity into 6.COND. W+AC Convert the measured conductivity to concentration (mmol/L) by referring to conversion Table when available. If not available, measure the specimen concentration using separately available concentration meter.

(8) (9)

After obtaining the concentration, press [F-3] or [F-4] SW to blink 7.CONC. W+AC Press [F- 1] or [F-2] SW to input the obtained concentration into 7.CONC. W+AC.

NIPRO SURDIAL (SEM) ver. OR2102-9902

10.17.1

10.17. Adjustment of contrast


Adjust contrast of LCD by turning volume of operation panel board under observing the LCD based on the following procedures. (1) Required device Precision screwdriver (2) Remove left cover from upper part of machine to expose operation panel board. (3) Turn on power SW of machine. (4) When back-light is OFF, press any SW of operation panel to turn on the light. (5) Adjust contrast of LCD by turning VR1 of operation panel board using precision
screwdriver under observing the LCD. Take care for tip of the screwdriver not to be touched with any part and for your hand not to be contacted with resistor of high temperature.

(6) Return the removed cover to the original position.

Fig. 10.17-1 Operation panel board

NIPRO SURDIAL (SEM) ver. OR2102-9902

11.1.1

11. DIP SW'S


11.1. Operation of dip SW's
Special functions are enabled by operating dip SWs. (1) Remove right cover from upper part of machine and right cover from DENSO box to exp dip
SWs placed on main board.

MAIN PCB : Main board DIPSW : Dip SW

Fig. 11.1-1 Main board and dip SW's (2) Check Nos. of dip SWs; SW1 to SW4 from the left in sequence. (3) By shifting each dip SW to UP, it is turned on and by shifting it to DOWN, it is turned off.

NIPRO SURDIAL (SEM) ver. OR2102-9902

11.2.1

11.2. Functions of dip SW's


11.2.1. Function of dip SW1 By turning on dip SW1, skip of rinse mode is enabled. By pressing [RINSE] SW during execution of rinsing operation, currently executed rinsing mode is skipped. Keep dip SW1 OFF during any process other than execution of maintenance.

11.2.2. Dip SW2 By turning on dip SW2, indication of adjustment mode in LCD is enabled for use by the manufacturer only. Keep dip SW2 OFF during any process other than execution of maintenance.

11.2.3. Dip SW3 By turning on dip SW3, service frequency indicated in maintenance mode menu 8 screen can be cleared to 0. Turn on dip SW3 to display maintenance mode menu 8 screen for indicating [CLEAR] SW and for blinking service frequency. Select item to be cleared and press [CLEAR] SW. Keep dip SW3 OFF during any process other than execution of maintenance.

11.2.4. Dip SW4 Not used

CAUTION

Usable functions are limited to those operated by SW1 and SW3. DO NOT operate SW2 because it is intended to be used by manufacturer only. Keep all dip SW's OFF during any process other than execution of maintenance.

NIPRO SURDIAL (SEM) ver. OR2102-9902

12.1.1

12. VARIOUS DATA


12.1. Loaded CPU and ROM

Symbol

Designation

CPU type

ROM type

ROM indication

CPU

Main board

PD784031

HN27C4096AG-10

1 2 3

Type name Version No. Registered date

2 The last letter of version No. indicates language (E: ENGLISH; S: SPANISH).

Fig. 12.1-1 ROM

NIPRO SURDIAL (SEM) ver. OR2102-9902

12.2.1

12.2. Wiring in POWER BOX

Fig. 12.2-1 Wiring in POWER BOX

NIPRO SURDIAL (SEM) ver. OR2102-9902